GS-8800 System Operation & Software Architecture


September 2005

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

GS-8800 Software
GS-8800 system provides 3 different software packages
1. Measurement S/W 2. Data Viewer S/W 3. Calibration S/W

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

GS-8800 Tools
GS-8800 system provides 4 type of tools
A. Test Plan Mover B. Database Generator

C. TestSetApplicationSetup

D. Path Effect Viewer

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

GS-8800 Software Architecture


Result Store Test Plan Generator Graph Viewer On going test result display Information Window

Additional Features

WTM and MoM Measurement Class Library Asset Class Library Drivers(MoM, IVI) Additional Features
GS-8800 Test Step 1

Wireless Test Manager


GS-8800 Test Step 2 GS-8800 Test Step 3

....

UI
Measurement
Spurious Measurement BER Measurement Set Switch Path Fader Control

....

Asset
8960 MoM Power Supply MoM Switch MoM IVI SA Driver IVI SigGen Driver Fader Driver

....
Class Library
Spirent API

Agilent VISA

IO Layer
8960 Test Set 66331B Power Supply 87130A Switch Driver E4445A PSA E4438C ESG E8247D PSG SR5500 Fader

Instrument

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

1. Measurement Software
Easy to use for measurement
Measurement Window
Test Step Results

Function Tabs

Graph Window
Graphical Results

At-a-Glance Yield and Test Result

Information Window Supplemental Window


Intermediate Test Step Results

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

1. Measurement Software
Define Test Plan->Set Spec/Param->Set Global Param->Run

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

1. Measurement Software
Typical Test Plan
<Format> Global Parameter Initiate Call Processing Test Case Test Case Test Case Test Case End Call

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

1. Measurement Software
Define a Test Plan: Drop Test Step into current Test Plan
Nested Looping

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

1. Measurement Software
Use Test Plan Generator Editor to create complex test plan

Check Test Plan

Use different tab for multi format test Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

1. Measurement Software
Set Test Specs and Parameters (if its necessary.)

Specs Table Parameters Table

Help Text
Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

1. Measurement Software
Ensure correct application (TA/LA) set

<Format> Global Parameter

TestSet Application
Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

1. Measurement Software
Set Global Parameters (if its necessary.)
Global Parameters Table

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

1. Measurement Software Looping (1)


Provides an Advance looping features 9 looping parameters available to iterate the test sequence
Example: Channel Number, GSM MS TX PL, GPRS MS TX PL Index, Supply Voltage, Temperature and etc

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

1. Measurement Software Looping (2)


Using parameters : Loop Start, Loop Step and Loop Stop

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

1. Measurement Software Looping (3)


Using Separated Sequence Definite the sequence in comma separated format

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

1. Measurement Software Looping (4)


Using Loop List to define multiple sequences Sequence can be defined in:
Simple Range : <Start>-<Stop> Complex Range : <Start>:<Step>:<Stop>

set to empty set to 1

Agilent Confidential

Back

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

1. Measurement Software
Run Test Plan: Click Run Button

Run Button

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

1. Measurement Software Information Window

Handy Calculator (1)

Test Plan Progress

Handy Calculator (2)


Agilent Confidential

Handy Calculator (3)

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

1. Measurement Software
Instrument address setup and verification

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

1. Measurement Software Tool Bar


Login Mode Setting Display Setting Help tools Export Test Plan Test Plan Generator MS TX Level Pair Fader Setup

To setup GPRS MS TX PL Index looping parameter

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

1. Measurement Software - Email Setup


Send test plan status acknowledgment through email

User.ini File

Options: Off, On, Abort, Fail, Abort&Fail


Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

1. Measurement Software - External/UUT Cable (1)


External Cable
Cable connects from RF IF Box to next interface box such as Shield Box

UUT Cable
Cable connect to UUT Change for each UUT model

RF Interface

UUT

Global Path External Cable Cal Offset

UUT Cable

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

1. Measurement Software - External/UUT Cable (2)


Setup External/UUT Cable Loss file in Set Global Parameter Tab

External/UUT Cable

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

1. Measurement Software - Customization


External control to Environmental Chamber and UUT
System.ini File

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

2. Result Data Viewer


Result data is saved in csv format and kept in /TestResults Using GS-8800 Data Viewer to browse the result data Enable further analysis using external tools such as Excel.

View result using Data Viewer


Agilent Confidential

View result using Excel

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

2. Result Data Viewer


Definable Result Output - items and the order to be output into csv result file
[WCDMA Result Data]

User.ini file

Date,Time,Test Title,Measured Value,Lower Limit,Upper Limit,Units,Pass/Fail, Channel, Item Number, Classification1, Classification3

Date Time 11/7/2003 4:41:01 PM 11/7/2003 4:42:03 PM 11/7/2003 4:42:03 PM 11/7/2003 4:42:03 PM

Test Title WCDMA Origination Passed TxSpur 9kHz<=f<150kHz TxSpur 9kHz<=f<150kHz TxSpur 9kHz<=f<150kHz Lower LimUpper LimUnits PASS PASS -36 dBm -36 dBm -36 dBm

RBW1kHz #1 19.20kHz RBW1kHz #2 49.10kHz RBW1kHz #3 11.80kHz Pass/Fail PASS PASS PASS PASS

Measured PASS -91.89 -92.43 -92.85

Channel ItemNumbClassification1 Classification3 9750:10700 10 CP 10 9750:10700 140 WB 140 140 WB 140 9750:10700 9750:10700 140 WB 140

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

2. Result Data Viewer Graphical Result


Graphical result data is saved in .gph extention and kept in /TestResults Using GS-8800 Data Viewer to browse the graphical result data

View results using Data Viewer

Double Click to display graph

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

2. Result Data Viewer Summary Report


A glance of the test plan result status Save as xxx_SR.csv format

xxx_SR.csv

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

3. System Calibration
- Measure a path loss and frequency characteristic between the measurement instrument and DUT.
Why do you measure a path loss and frequency characteristic ? - Correct the path loss and freq characteristic.

System components RF Interface Box Filter Box

DUT

Path loss
Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

3. System Calibration Key Points


1. Define the measurements paths 2. Define the reference point of the system 3. Define the reference analyzer 4. Reference analyzer calibrates SG
Reference Analyzer

RF Interface Box Filter Box System components


Agilent Confidential

Reference Point

DUT

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

3. System Calibration Items


1. Power sensor calibration 2. Reference signal measurements 3. Uplink path calibration 4. Downlink path calibration 5. Cable calibration

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

3. System Calibration Items Power Sensor


Zeroing and Calibration of power sensor to do first The power meter is the reference analyzer of the power in this system.

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

3. System Calibration Items - Reference Signal


Calibrate the SG level to use uplink path calibration Use the Reference analyzer Include the cable to connect to Reference Point
Reference Analyzer

RF Interface Box Filter Box System components

Reference Point

SG
Calibration Cable

Measure Power Level


Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

3. System Calibration Items Uplink Path


SG is connected to Reference Point Reference signal input to Reference point Analyzer measured SG Signal level through RF Interface box
Reference Analyzer

RF Interface Box Filter Box System components


Agilent Confidential

Reference Point

SG
Calibration Cable

Uplink Path

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

3. System Calibration Items Downlink Path


Reference Analyzer connect to Reference point

Reference Analyzer

RF Interface Box Filter Box System components


Downlink Path
Agilent Confidential

Reference Point

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

3. System Calibration Items External Cable


Measure loss of cable which connects System with DUT

Reference Analyzer

RF Interface Box Filter Box System components


Agilent Confidential

Cable

SG

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

3. GS-8800 Calibration Software


Easy to Calibrate the System
Parameter/ Sequence selection and editor Menu bar

Path Measurement Status Table

Graph Window

Measurement Result Table


Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

3. GS-8800 Calibration Software


Graphical operation instruction message

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

3. GS-8800 Calibration Software


Real Time cal result data Abort Cal

Cal data in tabular and graphical


Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

3. GS-8800 Calibration Software


Edit Calibration file CalData.cal

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

3. GS-8800 Calibration Software - Verification


GS-8800 provide 2 Step Verification Step1) Each Instruments Diagnostics by Measurement S/W
Instruments Self Test & Self Cal

Step2) System Verification by Calibration S/W (Veri.Mode)


RF I/F Path Loss/Switches, Cabling&Connection, Instruments Fail

Instruments Diagnostics
Agilent Confidential

System Verification

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

A. GS-8800 Tools Test Plan Mover


Duplicate and moving test plan between database file of the GS8800 RF DV System Upgrade database version
To launch File Navigator

Information of the database loaded Select test plan to be move Click >> button to start moving the test plan

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

B. GS-8800 Tools Database Generator


To duplicate original GS-8800_Default.mdb Requires GS-8800 USB Security Key
Location of the database to be generated To launch File Navigator

Name of the database to be generated Click Execute button to start generating database
Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

C. GS-8800 Tools TestSetApplicationSetup


Use to setup TestSetApplicationSetup.xml file in /TestData

A list of supported application for user selection

Agilent Confidential

AGILENT RESTRICTED DO NOT INCLUDE SLIDE TO EXTERNAL AUDIENCES WITHOUT NDA

D. GS-8800 Tools Path Effect Viewer


Define system path loss xml file Using Path Effect Service to edit path loss file By default, this configuration is disable in operator mode
Default Path name

Path Effect Viewer

Agilent Confidential