MULTIGAS MONITOR

For Safety Reasons,


This Equipment Must Be Operated
By
Qualified Personnel Only

BEFORE EACH DAYS USAGE,


THE SENSITIVITY MUST
BE TESTED.
?
or
?
BUMP TEST
or
RESPONSE CHECK

CAUTION
DISCONTINUE USE AFTER LOW
BATTERY ALARM SOUNDS.
PROMPTLY RECHARGE
BATTERY AFTER EACH USE.
IF INSTRUMENT WILL NOT BE
USED FOR 30 DAYS OR MORE,
LEAVE BATTERY ON CHARGE
CONTINUOUSLY OR REMOVE
BATTERY FROM THE INSTRUMENT.
CONTINUOUS CHARGING WILL
NOT DAMAGE BATTERY.

WATCHMAN
MULTIGAS MONITOR
Operating Instructions Quick Reference
Turn ON..Press the ON/OFF button.
Turn OFF..Press and hold the ON/OFF button for five seconds.
After Turn On..Optional Fresh Air SetupPress the RESET(YES) button.
To bypass Fresh Air Setup, press the PAGE(NO) or wait five seconds.The measure page will appear after either selection.
Pump CheckAttach tany sample lines, probesor lile traps as appropriate.
Block inlet until pump alarm sounds, if no alarms sounds check the flow system.
Back Light..Press any button at any time to turn on the display backlight.
Other PagesPress the PAGE button to view the Battery Status page.
Press the PAGE button again to view the Time/Date page and other optional pages.
Return to Measure Page.Wait 10-15 seconds, or momentarily depress the ON/OFF button or press the PAGE button several times until the Measure Gases screen appers.
Reset Alarms.Press the RESET button to clear alarms. NOTE: Alarmscan only be cleared from page showing the alarm condition.

WARNING

If the alarm condition is indicated, evacuate personnel from the area and notify the safety officer or follow your companys safety procedures. Recharge or replace batteries
in a non-hazardous area to prevent potential ignition of combustible atmospheres. THE ULTIMATE USER MUST READ THE WATCHMAN INSTRUCTION
AND TECHNICIAL MANUALS BEFORE USE.fAILURE TO FOLLOW THE INSTRUCTIONS CAN RESULT IN SERIOUS PERSONAL INJURY OR DEATH.

MINE SAFETY APPLIANCES COMPANY

%LEL Calibrated for

PENTANE

DATA DOCKING WINDOW

DISPLAY

EXTERNAL ALARM
CONNECTOR

WATCHMAN

MULTIGAS MONITOR
% O2

COMB

ppp H2 S

ppm CO
Page

AUDIO ALARM

LED ALARM
WINDOW

PAGE
BUTTON

On/Off

ON/OFF
BUTTON

Reset

RESET
BUTTON

TM

SLIDE PANEL
HALF-WAY

3.6 VDC

814127

NOTE DESIGN
OF TRACK FOR
RE-ASSEMBLY

3.6 VDC

814127

Battery Pack Installation

COMB

%O 2

ppm CO

ppm H2S

Page

On/Off

Reset

Battery Pack Installation

COMB

%O 2

SELF TEST
Please wait
ppm H2S

ppm CO

Page

On/Off

Reset

Battery Pack Installation

COMB

%O 2

Version 3.3
ppm H2S

ppm CO

Page

On/Off

Reset

30
SECONDS

Battery Pack Installation

%O 2

COMB

Operating Beep ?
NO

YES
ppm H2S

ppm CO

Page

On/Off

Reset

Battery Pack Installation

COMB

22 : 30

%O 2

22 JAN 95

NO set time ? YES

ppm H2S

ppm CO

Page

On/Off

Reset

Battery Pack Installation

COMB

%O 2

13:45
NEXT
ppm H2S

ppm CO

Page

On/Off

Reset

Battery Pack Installation

COMB

%O 2

TIME SET
COMPLETE
ppm H2S

ppm CO

Page

On/Off

Reset

Battery Pack Installation

COMB

%O 2

19 DEC 95
NEXT
ppm H2S

ppm CO

Page

On/Off

Reset

Battery Pack Installation

COMB

%O 2

DATE SET
COMPLETE
ppm H2S

ppm CO

Page

On/Off

Reset

Battery Pack Installation

COMB

%O 2

POWER OFF
ppm H2S

ppm CO

Page

On/Off

Reset

Battery Pack Installation

COMB

%O 2

ppm CO

ppm H2S

Page

On/Off

Reset

SETTING
ALARM
LEVELS

Combustible Gas
High Alarm

Oxygen
Low Alarm
Oxygen
HIGH Alarm
Toxic Gas #1, 2, 3
High Alarm
Time Weighted Average (TWA)
Short Term Exposure Level (STEL)

TURN ON THE UNIT

COMB

%O 2

ppm CO

ppm H2S

Page

On/Off

Reset

CALIBRATION MODE

Calibration
Sequence

POWER
OFF BUTTON
PRESS THE
ON/OFF
PRESS AND HOLD
THE PAGE AND RESET BUTTONS

COMB

%O 2

ppm CO

ppm H2S

Page

On/Off

Reset

Calibration
Sequence

RELEASE THE BUTTONS AND


PRESS THE RESET BUTTON

COMB

%O 2

Calibrate now ?
NO
YES
ppm H2S

ppm CO

Page

On/Off

Reset

5
SECONDS

Calibration
Sequence

ALLOW FOR WARM-UP,


PRESS THE RESET BUTTON

COMB

%O 2

APPLY FRESH AIR


QUIT

RDY
ppm H2S

ppm CO

Page

On/Off

Reset

READY

FRESH AIR

Calibration
Sequence

OXYGEN
SENSOR
IS
ALL
SENSORS
ARE SPAN
ADJUSTED
ADJUSTED
TO 20.8
TO ZERO

COMB

%O 2

Adjusting zeros
Please wait
ppm H2S

ppm CO

Page

On/Off

Reset

Calibration
Sequence

COMB

0.0

%O 2

Apply gas

SKIP

RDY
ppm H2S

ppm CO

Page

On/Off

Reset

Calibration
Sequence

COMB

0.0

%O 2

SPAN
OK
ppm H2S

ppm CO

Page

On/Off

Reset

Calibration
Sequence

Span setpoint is % LEL

Pentane Simulant
COMB

58

%O 2

SPAN
OK
ppm H2S

ppm CO

Page

On/Off

Reset

Calibration
Sequence

Sensor Problem
COMB

%O 2

** OUT OF RANGE **
Key Ignored
ppm H2S

ppm CO

Page

On/Off

Reset

Calibration
Sequence

SKIP IF NOT USING 20.8 % O2

COMB

%O 2

Apply gas
SKIP

RDY
ppm H2S

ppm CO

Page

15.4

On/Off

Reset

Calibration
Sequence

COMB

%O 2

SPAN

15.4

OK
ppm H2S

ppm CO

Page

On/Off

Reset

Calibration
Sequence

Span setpoint is 300 ppm CO

COMB

SKIP
0

%O 2

RDY
Apply gas
ppm H2S

ppm CO

Page

On/Off

Reset

Calibration
Sequence

Span setpoint is 300 ppm CO

COMB

300

%O 2

OK
SPAN
ppm H2S

ppm CO

Page

On/Off

Reset

Calibration
Sequence

Span setpoint is 10 ppm H2S

COMB

%O 2

SKIP
Apply gas

ppm H2S

ppm CO

Page

RDY
0

On/Off

Reset

Calibration
Sequence

Span setpoint is 10 ppm H2S

COMB

%O 2

OK
SPAN

ppm H2S

ppm CO

Page

10

On/Off

Reset

Calibration
Sequence

COMB

%O 2

CALIBRATION
COMPLETE
ppm H2S

ppm CO

Page

On/Off

Reset

Calibration
Sequence

COMB

%O 2

Remove Cal Cap


RDY
ppm H2S

ppm CO

Page

On/Off

Reset

Calibration
Sequence

COMB

%O 2

Please wait
ppm H2S

ppm CO

Page

On/Off

Reset

Calibration
Sequence

COMB

%O 2

POWER OFF
ppm H2S

ppm CO

Page

On/Off

Reset

Calibration Completed
COMB

%O 2

ppm CO

ppm H2S

Page

On/Off

Reset

DISPLAY PAGES

DISPLAY PAGES
%O 2

COMB

BATTERY OK
3. 5 VOLTS
ppm H2S

ppm CO

Page

On/Off

Reset

DISPLAY PAGES
%O 2

COMB

12: 30

1 JAN 96
ppm H2S

ppm CO

Page

On/Off

Reset

DISPLAY PAGES
%O 2

COMB

0
PEAK
55

ppm H2S

ppm CO

Page

20.8
100

On/Off

Reset

DISPLAY PAGES
%O 2

COMB

STEL :
50

10
ppm H2S

ppm CO

Page

On/Off

Reset

DISPLAY PAGES
%O 2

COMB

0.3 hour TWA :


50
10
ppm H2S

ppm CO

Page

On/Off

Reset

DISPLAY PAGES
%O 2

COMB

NEXT LOG # is 12
SKIP
RECD
ppm H2S

ppm CO

Page

On/Off

Reset

DISPLAY PAGES
%O 2

COMB

Tag Data ?
NO
YES
ppm H2S

ppm CO

Page

On/Off

Reset

DISPLAY PAGES
%O 2

COMB

_ _ _ _ _ _ _ _ _ _ Tag
NO
CHANGE
ppm H2S

ppm CO

Page

On/Off

Reset

DISPLAY PAGES
%O 2

COMB

A-1._ _ _ _ _ _ _ Tag
NEXT
ppm H2S

ppm CO

Page

On/Off

Reset

DISPLAY PAGES
%O 2

COMB

A-1._ _ _ _ _ _ _ Tag
OK
CHANGE
ppm H2S

ppm CO

Page

On/Off

Reset

DISPLAY PAGES
%O 2

COMB

Now recording
data log # 12
ppm H2S

ppm CO

Page

On/Off

Reset

NORMAL DISPLAY
%O 2

COMB

0 % LEL
0

ppm H2S

ppm CO

Page

20.8
0

On/Off

Reset

%O 2

COMB

To turn OFF: hold


POWER for 5 sec.
ppm H2S

ppm CO

Page

On/Off

Reset

%O 2

COMB

POWER OFF
ppm H2S

ppm CO

Page

On/Off

Reset

COMB

%O 2

ppm CO

ppm H2S

Page

On/Off

Reset

? ???
?
?
?
?
?
?? ?

ANY QUESTIONS

????

Major Components

WAT C H M AN

TM

MULTIGAS MONITOR

CAUTION
DISCONTINUE USE AFTER LOW
BATTERY ALARM SOUNDS.
PROMPTLY RECHARGE
BATTERY AFTER EACH USE.
IF INSTRUMENT WILL NOT BE
USED FOR 30 DAYS OR MORE,
LEAVE BATTERY ON CHARGE
CONTINUOUSLY OR REMOVE
BATTERY FROM THE INSTRUMENT.
CONTINUOUS CHARGING WILL
NOT DAMAGE BATTERY.

WARNING
TO PREVENT
IGNITION OF GASES
RECHARGE IN NONHAZARADOUS LOCATION

1 2 3 4 5 6 7 8

VER
3.2

A
B

OFF
ON

A
B

PUMP AND DRIVE ASSEMBLY


814933

Supply 3.15 vdc to the pump leads. With no load,


the current should be 20 to 25 mA. At 4 H2O
pressure, the current should be 22 to 45 mA with
a flow rate of 1 to 1.4 lpm. Block the pump inlet
and the pump should be able to pull > 40 H2O
pressure with a current draw < 55 mA.

A
B

VER
3.3

CALIBRATION PROCEDURE
PUMP AND DRIVE ASSEMBLY
1. WITH THE PUMP VOLTAGE AT 4.5 VDC AND
THE LOAD SET AT 50 H2O, THE PUMP
CURRENT MUST BE < 55 MA..
2. AT 3 VDC AND ZERO LOAD, THE PUMP
CURRENT 14 MA..
3. WITH THE PUMP AT 3.2 VDC AND A LOAD OF
30 H2O,THE PUMP CURRENT MUST BE < 38 MA
AND THE FLOW RATE MUST BE 195 ML PER
MINUTE.

710459

710445

Cut Sample Line


and Insert

PARTS
BATTERY PACK
SAMPLE LINE 5 FOOT
SAMPLE LINE 10 FOOT
SAMPLE LINE 15 FOOT
SAMPLE LINE 25 FOOT
SAMPLE LINE 50 FOOT
PROBE, 20 PLASTIC
PROBE, 3 FOOT BRASS
PROBE, 3 FOOT PLASTIC
PROBE, 4 FOOT BRASS
LINE TRAP (REPLACED 74814)
WATER TRAP WITH FITTINGS
WATER TRAP, IN LINE
CHARGER, OMEGA, 120 AC
CHARGER, OMEGA,220 VAC
CHARGER, OMEGA, 110/220 VAC, FIVE UNIT
CHARGER, OMEGA, 8 TO 24 VDC
REMOTE ALARM, 90 DECIBELS
REMOTE ALARM, 110 DECIBELS
CALIBRATION KIT, RP W/ 1.5 lpm REGULATOR

814127
011354
011955
011912
011913
011958
486934
011961
073743
011960
710459
497199
710445
494716
495965
801759
800525
800992
800991
477150

PARTS
CALIBRATION GAS, LEL PENTANE SIMULANT / 15% OXYGEN
CALIBRATION GAS, LEL PENTANE SIMULANT / 15% OXYGEN /
60 PPM CARBON MONOXIDE
CALIBRATION GAS, 10 PPM HYDROGEN SULFIDE
CALIBRATION GAS, LEL PENTANE SIMULANT / 15% OXYGEN /
300 PPM CARBON MONOXIDE / 10 PPM HYDROGEN SULFIDE
CALIBRATION GAS, LEL PENTANE SIMULANT / 15% OXYGEN /
10 PPM HYDROGEN SULFIDE
MAIN BOARD
DISPLAY BOARD
SENSOR BOARD
PUMP AND DRIVE
PRESSURE SWITCH
COMBUSTIBLE SENSOR
OXYGEN SENSOR
CARBON MONOXIDE SENSOR
HYDROGEN SULFIDE SENSOR
SULFUR DIOXIDE SENSOR
NITROGEN DIOXIDE SENSOR
NITRIC OXIDE SENSOR

478192
478919
467898
804770
804769
802470
814131
814437
814933
635621
478537
480566
636240
636241
807476
807477
808350

PARTS
FLOW REGULATOR
O RING SAMPLE INLET CAP
FILTER, SAMPLE INLET (5)
INLET FILTER REPLACEMENT KIT

467896
632152
497200
469995