06 (09/2003)

* VITALAB SELECTRA MARCA REGISTRADA DE MERCK.

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

KIT

PGINA

ALBUMINA ............................................................................................................... 01
ALT/TGP-UV ............................................................................................................. 02
AMILASE CNPG ....................................................................................................... 03
AST/TGO-UV ............................................................................................................ 04
CK-NAC ................................................................................................................... 05
CLORETOS COLORIMTRICO .............................................................................. 06
COLESTEROL 250 ................................................................................................. 07
COLESTEROL ENZ. LQUIDO ................................................................................ 08
COLESTEROL HDL ................................................................................................ 09
CREATININA ............................................................................................................ 10
DHL-UV .................................................................................................................... 11
FOSFATASE ALCALINA CINTICA .......................................................................... 12
FOSFATO-UV ........................................................................................................... 13
GAMA GT- CINTICA ............................................................................................... 14
GLICOSE ENZ. LQUIDA ......................................................................................... 15
GLUCOX 500 ........................................................................................................... 16
PROTENAS TOTAIS ............................................................................................... 17
TRIGLICRIDES 120 .............................................................................................. 18
TRIGLICRIDES ENZ. LQUIDO ............................................................................ 19
URATO 160 .............................................................................................................. 20
URATO ENZ. LQUIDO ............................................................................................ 21
URIA -UV ............................................................................................................... 22

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

Manual de Automao - Selectra 2

NDICE

TEST PARAMETERS A 37C


Name
:
Abbr. Name
:
Mode
:
Wavelength
:
Units
:
Decimals
:
Low Conc.
:
High Conc.
:
Calibrator Name
:
Repeat
:
Number
:
Concetration
:
Interval
:
Cut-off
:
Prozone check
:
Ref.male low
:
Ref. male high
:
Ref. female low
:
Ref. male high
:
Ref. ped. Low
:
Ref. ped. High
:
Control 1
:
Control 2
:
Correlat. Factor
:
Correlat. Offset
:

ALBUMINA
ALBD
Endpoint
620nm
g/dL
1
0
10
*
*
*
*
*
NO
NO
*
*
*
*
*
*
*
*
1.000
0.000

MONO MODE
Sample Blank
R1 bottle
Normal volume
Rerun volume

NO
25mL
300L
301L

:
:
:
:

Sample
Normal volume
Rerun volume
Incubation time
Low absorbance
High absorbance
R. Abs. L. Limit
R. Abs. H. Limit
Reagent Blank
Cal. Low Limit
Cal. High Limit
Factor

:
:
:
:
:
:
:
:
:
:
:

3L
2L
4.5min
-0.100
3.000
-0.100
0.500
yes
*
*
*

Parmetro a ser definido pelo usurio.

ALBUMINA
1666 testes
PREPARO DO REAGENTE DE USO
Vide Instrues de uso do kit.
CONTROLE DE QUALIDADE
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).

Manual de Automao - Selectra 2

ALBUMINA

VITALAB selectra marca registrada Merck.

REVISO:

06 (09/2003)

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

01

High absorbance
R. Abs. L. Limit
R. Abs. H. Limit
R. Abs. Deviation
Reagent Blank
Cal. Low Limit
Cal. High Limit
Factor

:
:
:
:
:
:
:
:

3.000
1.000
2.000
0.200
yes
*
*
- 1746

TEST PARAMETERS A 37C


Name
:
Abbr. Name
:
Mode
:
Wavelength
:
Units
:
Decimals
:
Low Conc.
:
High Conc.
:
Calibrator Name
:
Prozone check
:
Ref.male low
:
Ref. male high
:
Ref. female low
:
Ref. male high
:
Ref. ped. Low
:
Ref. ped. High
:

ALT/TGP
ALTD
Kinetic
340nm
UI/L
0
0
350
*
NO
*
*
*
*
*
*

Control 1
Control 2
Correlat. Factor
Correlat. Offset

:
:
:
:

*
*
1.000
0.000

CONTROLE DE QUALIDADE
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor
elevado (soro controle P - Doles).

:
:
:
:

NO
25mL
230L
248L

VITALAB selectra marca registrada Merck.

:
:
:
:
:

23 L
5L
77,230sec
10%
0.800

MONO MODE
Sample Blank
R1 bottle
Normal volume
Rerun volume
Sample
Normal volume
Rerun volume
Delay, min time
Linearity Limit
Low absorbance

Parmetro a ser definido pelo usurio.

ALT/TGP
400 testes
PREPARO DO REAGENTE DE USO
Vide Instrues de uso do kit.

REVISO:

Manual de Automao - Selectra 2

ALT/TGP UV

06 (09/2003)

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

02

High absorbance
R. Abs. L. Limit
R. Abs. H. Limit
R. Abs. Deviation
Reagent Blank
Cal. Low Limit
Cal. High Limit
Factor

:
:
:
:
:
:
:
:

3.000
-0.100
0.600
3.000
yes
*
*
3953

TEST PARAMETERS A 37C


Name
:
Abbr. Name
:
Mode
:
Wavelength
:
Units
:
Decimals
:
Low Conc.
:
High Conc.
:
Calibrator Name
:
Prozone check
:
Ref.male low
:
Ref. male high
:
Ref. female low
:
Ref. male high
:
Ref. ped. Low
:
Ref. ped. High
:

AMILASE
AMILD
Kinetic
405nm
UI/L
0
0
2000
*
NO
*
*
*
*
*
*

Control 1
Control 2
Correlat. Factor
Correlat. Offset

:
:
:
:

*
*
1.000
0.000

CONTROLE DE QUALIDADE
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor
elevado (soro controle P - Doles).

:
:
:
:

NO
25mL
250L
253L

VITALAB selectra marca registrada Merck.

:
:
:
:
:

5L
3L
77,230sec
15%
0.000

MONO MODE
Sample Blank
R1 bottle
Normal volume
Rerun volume
Sample
Normal volume
Rerun volume
Delay, min time
Linearity Limit
Low absorbance

Parmetro a ser definido pelo usurio.

AMILASE CNPG
240 determinaes
PREPARO DO REAGENTE DE USO
Vide Instrues de uso do kit.

REVISO:

Manual de Automao - Selectra 2

AMILASE CNPG

06 (09/2003)

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

03

High absorbance
R. Abs. L. Limit
R. Abs. H. Limit
R. Abs. Deviation
Reagent Blank
Cal. Low Limit
Cal. High Limit
Factor

:
:
:
:
:
:
:
:

3.000
1.000
2.000
0.200
yes
*
*
- 1746

TEST PARAMETERS A 37C


Name
:
Abbr. Name
:
Mode
:
Wavelength
:
Units
:
Decimals
:
Low Conc.
:
High Conc.
:
Calibrator Name
:
Prozone check
:
Ref.male low
:
Ref. male high
:
Ref. female low
:
Ref. male high
:
Ref. ped. Low
:
Ref. ped. High
:

AST/TGO
ASTD
Kinetic
340nm
UI/L
0
0
350
*
NO
*
*
*
*
*
*

Control 1
Control 2
Correlat. Factor
Correlat. Offset

:
:
:
:

*
*
1.000
0.000

CONTROLE DE QUALIDADE
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor
elevado (soro controle P - Doles).

:
:
:
:

NO
25mL
230 L
248 L

VITALAB selectra marca registrada Merck.

:
:
:
:
:

23L
5L
77,230sec
10%
0.800

MONO MODE
Sample Blank
R1 bottle
Normal volume
Rerun volume
Sample
Normal volume
Rerun volume
Delay, min time
Linearity Limit
Low absorbance

Parmetro a ser definido pelo usurio.

AST/TGO
400 testes
PREPARO DO REAGENTE DE USO
Vide Instrues de uso do kit.

REVISO:

Manual de Automao - Selectra 2

AST/TGO UV

06 (09/2003)

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

04

TEST PARAMETERS A 37C


Name
:
Abbr. Name
:
Mode
:
Wavelength
:
Units
:
Decimals
:
Low Conc.
:
High Conc.
:
Calibrator Name
:
Prozone check
:
Ref.male low
:
Ref. male high
:
Ref. female low
:
Ref. male high
:
Ref. ped. Low
:
Ref. ped. High
:

CKNAC
CKNACD
Kinetic
340nm
UI/L
0
0
1.700
*
NO
*
*
*
*
*
*

Control 1
Control 2
Correlat. Factor
Correlat. Offset

:
:
:
:

*
*
1.000
0.000

:
:
:
:

NO
25mL
250L
252L

:
:
:
:
:

5L
3L
130,389sec
10%
0.000

MONO MODE
Sample Blank
R1 bottle
Normal volume
Rerun volume
Sample
Normal volume
Rerun volume
Delay, min time
Linearity Limit
Low absorbance

High absorbance
R. Abs. L. Limit
R. Abs. H. Limit
R. Abs. Deviation
Reagent Blank
Cal. Low Limit
Cal. High Limit
Factor

:
:
:
:
:
:
:
:

3.000
0.000
0.800
3.000
yes
*
*
8095

Parmetro a ser definido pelo usurio.

CKNAC
250 determinaes
PREPARO DO REAGENTE DE USO
Vide Instrues de uso do kit.
CONTROLE DE QUALIDADE
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor
elevado (soro controle P - Doles).
VITALAB selectra marca registrada Merck.

REVISO:

Manual de Automao - Selectra 2

CK - NAC

06 (09/2003)

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

05

TEST PARAMETERS A 37C


Name
:
Abbr. Name
:
Mode
:
Wavelength
:
Units
:
Decimals
:
Low Conc.
:
High Conc.
:
Calibrator Name
:
Repeat
:
Number
:
Concetration
:
Interval
:
Cut-off
:
Prozone check
:
Ref.male low
:
Ref. male high
:
Ref. female low
:
Ref. male high
:
Ref. ped. Low
:
Ref. ped. High
:

CLORETOS
CLD
Endpoint
505nm
mEq/L
1
0
125
*
3
1
*
*
NO
NO
*
*
*
*
*
*

Control 1
Control 2
Correlat. Factor
Correlat. Offset

:
:
:
:

*
*
1.000
0.000

MONO MODE
Sample Blank
R1 bottle
Normal volume
Rerun volume

:
:
:
:

NO
25mL
300L
301L

Sample
Normal volume
Rerun volume
Incubation time
Low absorbance
High absorbance
R. Abs. L. Limit
R. Abs. H. Limit
Reagent Blank
Cal. Low Limit
Cal. High Limit
Factor

:
:
:
:
:
:
:
:
:
:
:

3L
2L
11.5min
-0.100
3.000
-0.100
0.150
yes
*
*
*

Parmetro a ser definido pelo usurio.

CLORETOS COLORIMTRICO
350 determinaes
PREPARO DO REAGENTE DE USO
Vide Instrues de uso do kit.
CONTROLE DE QUALIDADE
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor
elevado (soro controle P - Doles).

Manual de Automao - Selectra 2

CLORETOS COLORIMTRICO

VITALAB selectra marca registrada Merck.

REVISO:

06 (09/2003)

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

06

TEST PARAMETERS A 37C


Name
:
COLESTEROL250
Abbr. Name
:
COLD
Mode
:
Endpoint
Wavelength
:
505nm
Units
:
mg/dL
Decimals
:
0
Low Conc.
:
0
High Conc.
:
500
Calibrator Name
:
*
Repeat
:
3
Number
:
1
Concetration
:
*
Interval
:
*
Cut-off
:
NO
Prozone check
:
NO
Ref.male low
:
*
Ref. male high
:
*
Ref. female low
:
*
Ref. male high
:
*
Ref. ped. Low
:
*
Ref. ped. High
:
*
Control 1
Control 2
Correlat. Factor
Correlat. Offset

:
:
:
:

*
*
1.000
0.000

MONO MODE
Sample Blank
R1 bottle
Normal volume
Rerun volume

:
:
:
:

NO
25mL
300L
300L

Sample
Normal volume
Rerun volume
Incubation time
Low absorbance
High absorbance
R. Abs. L. Limit
R. Abs. H. Limit
Reagent Blank
Cal. Low Limit
Cal. High Limit
Factor

3L
2L
11.5min
-0.100
3.000
-0.100
0.300
yes
*
*
*

:
:
:
:
:
:
:
:
:
:
:

Parmetro a ser definido pelo usurio.

COLESTEROL 250
833 determinaes
PREPARO DO REAGENTE DE USO
Vide Instrues de uso do kit.
CONTROLE DE QUALIDADE
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor
elevado (soro controle P - Doles).

Manual de Automao - Selectra 2

COLESTEROL 250

VITALAB selectra marca registrada Merck.

REVISO:

06 (09/2003)

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

07

TEST PARAMETERS A 37C


Name
:
Abbr. Name
:
Mode
:
Wavelength
:
Units
:
Decimals
:
Low Conc.
:
High Conc.
:
Calibrator Name
:
Repeat
:
Number
:
Concetration
:
Interval
:
Cut-off
:
Prozone check
:
Ref.male low
:
Ref. male high
:
Ref. female low
:
Ref. male high
:
Ref. ped. Low
:
Ref. ped. High
:

COLEST.ENZ.
COLD
Endpoint
505nm
mg/dL
0
0
800
*
3
1
*
*
NO
NO
*
*
*
*
*
*

Control 1
Control 2
Correlat. Factor
Correlat. Offset

*
*
1.000
0.000

:
:
:
:

Sample
Normal volume
Rerun volume
Incubation time
Low absorbance
High absorbance
R. Abs. L. Limit
R. Abs. H. Limit
Reagent Blank
Cal. Low Limit
Cal. High Limit
Factor

:
:
:
:
:
:
:
:
:
:

3L
2L
5min
-0.100
: 3.000
-0.100
0.300
yes
*
*
*

Parmetro a ser definido pelo usurio.

COLESTEROL ENZIMTICO LQUIDO


666 determinaes
PREPARO DO REAGENTE DE USO
Vide Instrues de uso do kit.
CONTROLE DE QUALIDADE
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).

Manual de Automao - Selectra 2

COLESTEROL ENZ. LQUIDO

VITALAB selectra marca registrada Merck.

REVISO:

06 (09/2003)

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

08

TEST PARAMETERS A 37C


Name
:
COLESTEROL HDL
Abbr. Name
:
CHDL
Mode
:
Endpoint
Wavelength
:
505nm
Units
:
mg/dL
Decimals
:
0
Low Conc.
:
0
High Conc.
:
400
Calibrator Name
:
*
Repeat
:
*
Number
:
*
Concetration
:
*
Interval
:
*
Cut-off
:
NO
Prozone check
:
NO
Ref.male low
:
*
Ref. male high
:
*
Ref. female low
:
*
Ref. male high
:
*
Ref. ped. Low
:
*
Ref. ped. High
:
*
Control 1
Control 2
Correlat. Factor
Correlat. Offset
MONO MODE
Sample Blank
R1 bottle
Normal volume
Rerun volume
Sample
Normal volume
Rerun volume
Incubation time

:
:
:
:

*
*
1.000
0.000

:
:
:
:

NO
25mL
400 L
400 L

:
:
:

10L
5L
11.5min

Low absorbance
High absorbance
R. Abs. L. Limit
R. Abs. H. Limit
Reagent Blank
Cal. Low Limit
Cal. High Limit
Factor

:
:
:
:
:
:
:
:

-0.100
3.000
-0.100
0.300
yes
*
*
*

Parmetro a ser definido pelo usurio.

COLESTEROL HDL
40 determinaes
PREPARO DO REAGENTE DE USO
Vide Instrues de uso do kit.
OBSERVAES:
Utilizar como calibrador o padro que
acompanha o produto Colesterol HDL Doles.
Utilizar como amostra o sobrenadante
obtido na etapa de precipitao.
Vide instrues de uso do Colesterol HDL
- Doles.
Utilizar como reagente de cor o tampo
enzimas do Colesterol 250 Doles ou
Colesterol Enzimtico Lquido Doles.
CONTROLE DE QUALIDADE
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor
elevado (soro controle P - Doles).

Manual de Automao - Selectra 2

COLESTEROL HDL

VITALAB selectra marca registrada Merck.


REVISO:

06 (09/2003)

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

09

TEST PARAMETERS A 37C


Name
:
Abbr. Name
:
Mode
:
Wavelength
:
Units
:
Decimals
:
Low Conc.
:
High Conc.
:
Calibrator Name
:
Repeat
:
Number
:
Concetration
:
Interval
:
Cut-off
:
Prozone check
:
Ref.male low
:
Ref. male high
:
Ref. female low
:
Ref. male high
:
Ref. ped. Low
:
Ref. ped. High
:

CREATININA
CREAD
TwoPoint
505nm
mg/dL
1
0
10
*
*
*
*
*
NO
NO
*
*
*
*
*
*

Control 1
Control 2
Correlat. Factor
Correlat. Offset

:
:
:
:

*
*
1.000
0.000

:
:
:
:

NO
25mL
230L
243L

:
:
:

23L
10L
32.9sec

MONO MODE
Sample Blank
R1 bottle
Normal volume
Rerun volume
Sample
Normal volume
Rerun volume
Incubation time

Low absorbance
High absorbance
R. Abs. L. Limit
R. Abs. H. Limit
R. Abs. Deviation
Reagent Blank
Cal. Low Limit
Cal. High Limit
Factor

:
:
:
:
:
:
:
:
:

-0.100
3.000
-0.100
0.600
3.000
yes
*
*
*

Parmetro a ser definido pelo usurio.

CREATININA
2200 determinaes
PREPARO DO REAGENTE DE USO
Vide instrues de uso.
CONTROLE DE QUALIDADE
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).

Manual de Automao - Selectra 2

CREATININA CINTICA

VITALAB selectra marca registrada Merck.

REVISO:

06 (09/2003)

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

10

TEST PARAMETERS A 37C


Name
:
Abbr. Name
:
Mode
:
Wavelength
:
Units
:
Decimals
:
Low Conc.
:
High Conc.
:
Calibrator Name
:
Prozone check
:
Ref.male low
:
Ref. male high
:
Ref. female low
:
Ref. male high
:
Ref. ped. Low
:
Ref. ped. High
:

DHLUV
DHLD
Kinetic
340nm
UI/L
0
0
1.250
*
NO
*
*
*
*
*
*

Control 1
Control 2
Correlat. Factor
Correlat. Offset

:
:
:
:

*
*
1.000
0.000

:
:
:
:

NO
25mL
250 L
253 L

:
:
:
:

5L
2L
70,186sec
10%

MONO MODE
Sample Blank
R1 bottle
Normal volume
Rerun volume
Sample
Normal volume
Rerun volume
Delay, min time
Linearity Limit

Low absorbance
High absorbance
R. Abs. L. Limit
R. Abs. H. Limit
R. Abs. Deviation
Reagent Blank
Cal. Low Limit
Cal. High Limit
Factor

:
:
:
:
:
:
:
:
:

0.300
3.000
0.700
2.000
3.000
yes
*
*
- 8095

* Parmetro a ser definido pelo usurio.


DHLUV
160 determinaes
PREPARO DO REAGENTE DE USO
Vide Instrues de uso do kit.
CONTROLE DE QUALIDADE
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor
elevado (soro controle P - Doles).

Manual de Automao - Selectra 2

DHL - UV

VITALAB selectra marca registrada Merck.

REVISO:

06 (09/2003)

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

11

TEST PARAMETERS A 37C


Name
:
Abbr. Name
:
Mode
:
Wavelength
:
Units
:
Decimals
:
Low Conc.
:
High Conc.
:
Calibrator Name
:
Prozone check
:
Ref.male low
:
Ref. male high
:
Ref. female low
:
Ref. male high
:
Ref. ped. Low
:
Ref. ped. High
:
Control 1
:
Control 2
:
Correlat. Factor
:
Correlat. Offset
:
MONO MODE
Sample Blank
R1 bottle
Normal volume
Rerun volume
Sample
Normal volume
Rerun volume
Delay, min time
Linearity Limit
Low absorbance

FOSFATASE ALCALINA
FALC
Kinetic
405nm
UI/L
0
0
690
*
NO
*
*
*
*
*
*
*
*
1.000
0.000

:
:
:
:

NO
25mL
250 L
253 L

:
:
:
:
:

5L
2L
70,377sec
10%
0.000

High absorbance
R. Abs. L. Limit
R. Abs. H. Limit
R. Abs. Deviation
Reagent Blank
Cal. Low Limit
Cal. High Limit
Factor

:
:
:
:
:
:
:
:

3.000
0.000
1.200
3.000
yes
*
*
2764

* Parmetro a ser definido pelo usurio.


FOSFATASE ALCALINA CINTICA
400 determinaes
PREPARO DO REAGENTE DE USO
Vide Instrues de uso do kit.
CONTROLE DE QUALIDADE
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).
VITALAB selectra marca registrada Merck.

REVISO:

Manual de Automao - Selectra 2

FOSFATASE ALCALINA CINTICA

06 (09/2003)

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

12

TEST PARAMETERS A 37C


Name
:
Abbr. Name
:
Mode
:
Wavelength
:
Units
:
Decimals
:
Low Conc.
:
High Conc.
:
Calibrator Name
:
Repeat
:
Number
:
Concetration
:
Interval
:
Cut-off
:
Prozone check
:
Ref.male low
:
Ref. male high
:
Ref. female low
:
Ref. male high
:
Ref. ped. Low
:
Ref. ped. High
:

FOSFATO
FOSFD
Endpoint
340nm
mg/dL
1
0
20
*
3
1
*
*
NO
NO
*
*
*
*
*
*

Control 1
Control 2
Correlat. Factor
Correlat. Offset

:
:
:
:

*
*
1.000
0.000

MONO MODE
Sample Blank
R1 bottle
Normal volume
Rerun volume

:
:
:
:

NO
25mL
300 L
301 L

Sample
Normal volume
Rerun volume
Incubation time
Low absorbance
High absorbance
R. Abs. L. Limit
R. Abs. H. Limit
Reagent Blank
Cal. Low Limit
Cal. High Limit
Factor

:
:
:
:
:
:
:
:
:
:
:

3L
2L
11.5min
-0.100
3.000
-0.100
1.300
yes
*
*
*

* Parmetro a ser definido pelo usurio.


FOSFATO UV
350 determinaes
PREPARO DO REAGENTE DE USO
Vide Instrues de uso do kit.
CONTROLE DE QUALIDADE
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor
elevado (soro controle P - Doles).

Manual de Automao - Selectra 2

FOSFATO - UV

VITALAB selectra marca registrada Merck.


REVISO:

06 (09/2003)

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

13

TEST PARAMETERS A 37C


Name
:
Abbr. Name
:
Mode
:
Wavelength
:
Units
:
Decimals
:
Low Conc.
:
High Conc.
:
Calibrator Name
:
Prozone check
:
Ref.male low
:
Ref. male high
:
Ref. female low
:
Ref. male high
:
Ref. ped. Low
:
Ref. ped. High
:

GAMAGT
GGTD
Kinetic
405nm
UI/L
0
0
300
*
NO
*
*
*
*
*
*

Control 1
Control 2
Correlat. Factor
Correlat. Offset

:
:
:
:

*
*
1.000
0.000

CONTROLE DE QUALIDADE
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).

:
:
:
:

NO
25mL
240 L
249 L

VITALAB selectra marca registrada Merck.

:
:
:
:

12 L
3L
70,137sec
10%

MONO MODE
Sample Blank
R1 bottle
Normal volume
Rerun volume
Sample
Normal volume
Rerun volume
Delay, min time
Linearity Limit

Low absorbance
High absorbance
R. Abs. L. Limit
R. Abs. H. Limit
R. Abs. Deviation
Reagent Blank
Cal. Low Limit
Cal. High Limit
Factor

:
:
:
:
:
:
:
:
:

-0.100
3.000
-0.100
1.000
3.000
yes
*
*
2121

* Parmetro a ser definido pelo usurio.


GAMA GT CINTICA
400 determinaes
PREPARO DO REAGENTE DE USO
Vide Instrues de uso do kit.

REVISO:

Manual de Automao - Selectra 2

GAMA GT CINTICA

06 (09/2003)

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

14

TEST PARAMETERS A 37C


Name
:
Abbr. Name
:
Mode
:
Wavelength
:
Units
:
Decimals
:
Low Conc.
:
High Conc.
:
Calibrator Name
:
Repeat
:
Number
:
Concetration
:
Interval
:
Cut-off
:
Prozone check
:
Ref.male low
:
Ref. male high
:
Ref. female low
:
Ref. male high
:
Ref. ped. Low
:
Ref. ped. High
:

GLICOSE
GLICD
Endpoint
505nm
mg/dL
1
0
600
*
3
1
*
*
NO
NO
*
*
*
*
*
*

Control 1
Control 2
Correlat. Factor
Correlat. Offset

:
:
:
:

*
*
1.000
0.000

MONO MODE
Sample Blank
R1 bottle
Normal volume
Rerun volume

:
:
:
:

NO
25mL
300 L
301 L

Sample
Normal volume
Rerun volume
Incubation time
Low absorbance
High absorbance
R. Abs. L. Limit
R. Abs. H. Limit
Reagent Blank
Cal. Low Limit
Cal. High Limit
Factor

:
:
:
:
:
:
:
:
:
:
:

3L
2L
5min
-0.100
3.000
-0.100
0.300
yes
*
*
*

* Parmetro a ser definido pelo usurio.


GLICOSE ENZ. LQUIDA
Ref.A 1666 determinaes
Kito 4998 determinaes
PREPARO DO REAGENTE DE USO
Vide Instrues de uso do kit.
CONTROLE DE QUALIDADE
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).

Manual de Automao - Selectra 2

GLICOSE ENZ. LQUIDA

VITALAB selectra marca registrada Merck.


REVISO:

06 (09/2003)

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

15

TEST PARAMETERS A 37C


Name
:
Abbr. Name
:
Mode
:
Wavelength
:
Units
:
Decimals
:
Low Conc.
:
High Conc.
:
Calibrator Name
:
Repeat
:
Number
:
Concetration
:
Interval
:
Cut-off
:
Prozone check
:
Ref.male low
:
Ref. male high
:
Ref. female low
:
Ref. male high
:
Ref. ped. Low
:
Ref. ped. High
:

GLICOSE
GLICD
Endpoint
505nm
mg/dL
1
0
600
*
3
1
*
*
NO
NO
*
*
*
*
*
*

Control 1
Control 2
Correlat. Factor
Correlat. Offset

:
:
:
:

*
*
1.000
0.000

MONO MODE
Sample Blank
R1 bottle
Normal volume
Rerun volume

:
:
:
:

NO
25mL
300 L
301 L

Sample
Normal volume
Rerun volume
Incubation time
Low absorbance
High absorbance
R. Abs. L. Limit
R. Abs. H. Limit
Reagent Blank
Cal. Low Limit
Cal. High Limit
Factor

:
:
:
:
:
:
:
:
:
:
:

3L
2L
11.5min
-0.100
3.000
-0.100
0.300
yes
*
*
*

* Parmetro a ser definido pelo usurio.


GLUCOX 500
Ref.A 1666 determinaes
Kito 6666 determinaes
PREPARO DO REAGENTE DE USO
Vide Instrues de uso do kit.
CONTROLE DE QUALIDADE
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle N
- Doles) e outro soro controle de valor elevado
(soro controle P - Doles).

Manual de Automao - Selectra 2

GLUCOX 500

VITALAB selectra marca registrada Merck.


REVISO:

06 (09/2003)

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

16

TEST PARAMETERS A 37C


Name
:
Abbr. Name
:
Mode
:
Wavelength
:
Units
:
Decimals
:
Low Conc.
:
High Conc.
:
Calibrator Name
:
Repeat
:
Number
:
Concetration
:
Interval
:
Cut-off
:
Prozone check
:
Ref.male low
:
Ref. male high
:
Ref. female low
:
Ref. male high
:
Ref. ped. Low
:
Ref. ped. High
:

PROT.TOTAIS
PROTD
Endpoint
546nm
g/L
1
0
12
*
*
*
*
*
NO
NO
*
*
*
*
*
*

Control 1
Control 2
Correlat. Factor
Correlat. Offset

:
:
:
:

*
*
1.000
0.000

MONO MODE
Sample Blank
R1 bottle
Normal volume
Rerun volume
Sample

:
:
:
:

NO
25mL
250 L
252 L

Normal volume
Rerun volume
Incubation time
Low absorbance
High absorbance
R. Abs. L. Limit
R. Abs. H. Limit
Reagent Blank
Cal. Low Limit
Cal. High Limit
Factor

:
:
:
:
:
:
:
:
:
:
:

5L
3L
5min
-0.100
3.000
-0.100
0.350
yes
*
*
*

* Parmetro a ser definido pelo usurio.


PROTENAS TOTAIS
2000 determinaes
PREPARO DO REAGENTE DE USO
Misturar 10mL do reagente de trabalho de
Biureto com 08 gotas de soluo alcalina.
Aps o preparo o reagente permanece estvel
por 07 dias se mantido entre 2-8C.
CONTROLE DE QUALIDADE
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).

Manual de Automao - Selectra 2

PROTENAS TOTAIS

VITALAB selectra marca registrada Merck.


REVISO:

06 (09/2003)

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

17

TEST PARAMETERS A 37C


Name
:
Abbr. Name
:
Mode
:
Wavelength
:
Units
:
Decimals
:
Low Conc.
:
High Conc.
:
Calibrator Name
:
Repeat
:
Number
:
Concetration
:
Interval
:
Cut-off
:
Prozone check
:
Ref.male low
:
Ref. male high
:
Ref. female low
:
Ref. male high
:
Ref. ped. Low
:
Ref. ped. High
:

TRIGLICRIDES
TRIGD
Endpoint
505nm
mg/dL
1
0
1000
*
3
1
*
*
NO
NO
*
*
*
*
*
*

Control 1
Control 2
Correlat. Factor
Correlat. Offset

:
:
:
:

*
*
1.000
0.000

MONO MODE
Sample Blank
R1 bottle
Normal volume

:
:
:

NO
25mL
300 L

Rerun volume
Sample
Normal volume
Rerun volume
Incubation time
Low absorbance
High absorbance
R. Abs. L. Limit
R. Abs. H. Limit
Reagent Blank
Cal. Low Limit
Cal. High Limit
Factor

:
:
:
:
:
:
:
:
:
:
:

301 L
3L
2L
11.5min
-0.100
: 3.000
-0.100
0.150
yes
*
*
*

* Parmetro a ser definido pelo usurio.


TRIGLICRIDES 120
400 determinaes
PREPARO DO REAGENTE DE USO
Vide Instrues de uso do kit.
CONTROLE DE QUALIDADE
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).

Manual de Automao - Selectra 2

TRIGLICRIDES 120

VITALAB selectra marca registrada Merck.


REVISO:

06 (09/2003)

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

18

TEST PARAMETERS A 37C


Name
:
Abbr. Name
:
Mode
:
Wavelength
:
Units
:
Decimals
:
Low Conc.
:
High Conc.
:
Calibrator Name
:
Repeat
:
Number
:
Concetration
:
Interval
:
Cut-off
:
Prozone check
:
Ref.male low
:
Ref. male high
:
Ref. female low
:
Ref. male high
:
Ref. ped. Low
:
Ref. ped. High
:

TRIGLICRIDES
TRIGD
Endpoint
505nm
mg/dL
1
0
800
*
3
1
*
*
NO
NO
*
*
*
*
*
*

Control 1
Control 2
Correlat. Factor
Correlat. Offset

:
:
:
:

*
*
1.000
0.000

MONO MODE
Sample Blank
R1 bottle
Normal volume
Rerun volume

:
:
:
:

NO
25mL
300L
301L

Sample
Normal volume
Rerun volume
Incubation time
Low absorbance
High absorbance
R. Abs. L. Limit
R. Abs. H. Limit
Reagent Blank
Cal. Low Limit
Cal. High Limit
Factor

:
:
:
:
:
:
:
:
:
:
:

3L
2L
5min
-0.100
3.000
-0.100
0.150
yes
*
*
*

* Parmetro a ser definido pelo usurio.


TRIGLICRIDES ENZIMTICO LQUIDO
666 determinaes
PREPARO DO REAGENTE DE USO
Vide Instrues de uso do kit.
CONTROLE DE QUALIDADE
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).

Manual de Automao - Selectra 2

TRIGLICRIDES ENZ. LQUIDO

VITALAB selectra marca registrada Merck.


REVISO:

06 (09/2003)

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

19

TEST PARAMETERS A 37C


Name
:
Abbr. Name
:
Mode
:
Wavelength
:
Units
:
Decimals
:
Low Conc.
:
High Conc.
:
Calibrator Name
:
Repeat
:
Number
:
Concetration
:
Interval
:
Cut-off
:
Prozone check
:
Ref.male low
:
Ref. male high
:
Ref. female low
:
Ref. male high
:
Ref. ped. Low
:
Ref. ped. High
:

URATO160
URAD
Endpoint
505nm
mg/dL
2
0
14
*
3
1
*
*
NO
NO
*
*
*
*
*
*

Control 1
Control 2
Correlat. Factor
Correlat. Offset

:
:
:
:

*
*
1.000
0.000

MONO MODE
Sample Blank
R1 bottle
Normal volume
Rerun volume

:
:
:
:

NO
25mL
250L
253L

Sample
Normal volume
Rerun volume
Incubation time
Low absorbance
High absorbance
R. Abs. L. Limit
R. Abs. H. Limit
Reagent Blank
Cal. Low Limit
Cal. High Limit
Factor

:
:
:
:
:
:
:
:
:
:
:

5L
2L
11.5min
0.000
3.000
-0.100
0.300
yes
*
*
*

* Parmetro a ser definido pelo usurio.


URATO 160
640 determinaes
PREPARO DO REAGENTE DE USO
Vide Instrues de uso do kit.
CONTROLE DE QUALIDADE
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).

Manual de Automao - Selectra 2

URATO 160

VITALAB selectra marca registrada Merck.

REVISO:

06 (09/2003)

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

20

TEST PARAMETERS A 37C


Name
:
Abbr. Name
:
Mode
:
Wavelength
:
Units
:
Decimals
:
Low Conc.
:
High Conc.
:
Calibrator Name
:
Repeat
:
Number
:
Concetration
:
Interval
:
Cut-off
:
Prozone check
:
Ref.male low
:
Ref. male high
:
Ref. female low
:
Ref. male high
:
Ref. ped. Low
:
Ref. ped. High
:

URATO
URAD
Endpoint
505nm
mg/dL
2
0
14
*
3
1
*
*
NO
NO
*
*
*
*
*
*

Control 1
Control 2
Correlat. Factor
Correlat. Offset

:
:
:
:

*
*
1.000
0.000

MONO MODE
Sample Blank
R1 bottle
Normal volume
Rerun volume

:
:
:
:

NO
25mL
250 L
253 L

Sample
Normal volume
Rerun volume
Incubation time
Low absorbance
High absorbance
R. Abs. L. Limit
R. Abs. H. Limi
Reagent Blank
Cal. Low Limit
Cal. High Limit
Factor

:
:
:
:
:
:
:
:
:
:
:

5L
2L
5min
0.000
3.000
-0.100
0.300
yes
*
*
*

* Parmetro a ser definido pelo usurio.


URATO ENZIMTICO LQUIDO
640 determinaes
PREPARO DO REAGENTE DE USO
Vide Instrues de uso do kit.
CONTROLE DE QUALIDADE
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).

Manual de Automao - Selectra 2

URATO ENZ. LQUIDO

VITALAB selectra marca registrada Merck.


REVISO:

06 (09/2003)

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

21

TEST PARAMETERS A 37C


Name
:
URIAUV
Abbr. Name
:
UREIAD
Mode
:
TwoPoint
Wavelength
:
340nm
Units
:
mg/dL
Decimals
:
0
Low Conc.
:
0
High Conc.
:
300
Calibrator Name :
*
Repeat
:
3
Number
:
1
Concetration
:
*
Interval
:
*
Cut-off
:
NO
Prozone check
:
NO
Ref.male low
:
*
Ref. male high
:
*
Ref. female low
:
*
Ref. male high
:
*
Ref. ped. Low
:
*
Ref. ped. High
:
*
Control 1
:
*
Control 2
:
*
Correlat. Factor
:
1.000
Correlat. Offset
:
0.000
MONO MODE
Sample Blank
R1 bottle
Normal volume
Rerun volume

:
:
:
:

NO
25mL
300L
303L

Sample
Normal volume
Rerun volume
Delay, min time
Low absorbance
High absorbance
R. Abs. L. Limit
R. Abs. H. Limit
Reagent Blank
Cal. Low Limit
Cal. High Limit
Factor

:
:
:
:
:
:
:
:
:
:
:

3L
2L
32,70sec
0.600
3.000
1.000
2.300
yes
*
*
*

* Parmetro a ser definido pelo usurio.


URIA UV
666 determinaes
PREPARO DO REAGENTE DE USO
Vide Instrues de uso do kit.
CONTROLE DE QUALIDADE
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).

Manual de Automao - Selectra 2

URIA - UV

VITALAB selectra marca registrada Merck.

REVISO:

06 (09/2003)

Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br

22

You might also like