Professional Documents
Culture Documents
Revision History
Manual Version
Software Version
2.0
Revision date
Changes
October 2001
3.0
1.2
January 2002
4.0
1.31
December 2002
5.0
1.60
December 2004
6.0
1.70
May 2005
English only!
7.0
1.71
Juni 2006
cobas branding
8.0
1.71
Juni 2006
9.0
1.72
May 2007
10.0
1.74
June 2008
Edition notice
cobas b 121 system
In the course of 2005 the Roche OMNI C system was rebranded under the Roche
Diagnostics professional IVD user brand cobas.
Systems with a serial number of 5000 or above are cobas b 121 system.
Systems with a serial number up to 4999 are Roche OMNI C systems.
Copyright
2008, Roche Diagnostics GmbH. All rights reserved
The contents of this document may not be reproduced in any form or communicated
to any third party without the prior written consent of Roche Diagnostics.
While every effort is made to ensure its correctness. Roche Diagnostics assumes no
responsibility for errors or omissions which may appear in this document.
Subject to change without notice.
Brands
COBAS, COBAS B, COBAS BGE LINK, LIFE NEEDS ANSWERS, ROCHE OMNI,
AUTOQC, ROCHE MICROSAMPLER, COMBITROL and AUTO-TROL are
trademarks of Roche.
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
1
Contact addresses
Manufacturer
Edition
Revision 10.0, June 2008
First edition: October 2001
REF/No. 03260992001
Roche Diagnostics
2
June 2008
Instructions for Use Version 10.0
Table of contents
Revision History
Edition notice
Copyright
Brands
Contact addresses
Edition
Table of contents
Preface
How to use this manual
Where to find information
Conventions used in this manual
1
1
1
1
2
2
3
5
5
5
5
Part A
1 Safety information
Important information
Operating safety information
A-5
A-6
A-9
A-10
A-12
A-13
A-15
Installation
Shutdown
A-21
A-36
4 Specifications
Performance data
Default and input values
Sample throughput
Sample volume
Sample types
Calibrations
Environmental parameters
Product data
Printer
Screen
Barcode scanner
A-47
A-49
A-49
A-49
A-50
A-50
A-51
A-53
A-53
A-54
A-54
Calibration - general
B-37
8 Operating modes
B-43
B-43
B-49
B-61
Part C
9 Maintenance
Maintenance - general
Decontamination
Daily
Weekly
Semi annual
Sample-dependent maintenance procedures
Unscheduled
Additional maintenance procedures
Maintenance overview
Maintenance scheduler
Troubleshooting
C-5
C-5
C-7
C-8
C-9
C-10
C-17
C-30
C-31
C-32
Part D
10 Troubleshooting
Troubleshooting - general
System stops
List of system stops
Electrode status
Appendix
Operation
B-19
B-20
B-26
B-28
B-30
B-31
B-32
7 Calibration
Maintenance
2 General description
Introduction
General notes
Measurement and calibration procedure
Safety instructions for specific dangers
System description
D-5
D-5
D-7
D-14
Part E
Part B
11 List of consumables
Order information
5 Measurement
Preanalytics
Measuring procedure
B-5
B-8
Index
E-5
Part F
6 Quality control
B-17
B-17
Index
F-3
June 2008
3
Roche Diagnostics
4
June 2008
Instructions for Use Version 10.0
Preface
This manual has detailed descriptions of cobas b 121 system features and general
operational concepts, specification functions and use of controls, operating
techniques, emergency procedures, product labeling and maintenance procedures.
To help you find information quickly, there is a table of contents at the beginning of
the book and each chapter. In addition, a complete index can be found at the end.
Helping to locate and interpret information in this manual the following symbols are
used:
Symbol
Description
Procedural step
List item
Cross-reference
Call up of screen
Note
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
5
Symbol
Description
Attention, Warning, Danger
All sections / passages that are marked with this symbol describe
procedures and/or indicate conditions or dangers that could damage
or lead to a malfunction in the cobas b 121 system, and which
therefore should never be attempted and/or must be observed to
avoid potential injuries (to patients, users and third parties).
Risk of infection!
All sections and parts of texts that are marked with this symbol
describe procedures that may involve risk of infection.
IVD symbols
The IVD symbols are used in accordance with DIN EN 980:2003, ISO 15223:2000 (as
per the In Vitro Diagnostics Directive 98/79/EC), and DIN EN ISO 780:1997.
Symbol
Description
Store upright
Risk of infection!
(according to the standard DIN EN 61010-2-101:2002) (Instrument)
Risk of infection!
(according to the standard DIN ISO 15223-1:2005)(Consumables)
Catalogue number
Roche Diagnostics
6
June 2008
Instructions for Use Version 10.0
Symbol
Description
Caution (refer to accompanying documents). Please refer to safetyrelated notes in the manual, accompanying this instrument.
Please consult instructions for use
Other symbols
Description
Electrodes:
This date indicates the limit of the maximum storage time of
an electrode. The electrode must be installed in the instrument
no later than the imprinted date.
If the installation takes place on the imprinted date, it still falls
within the specifications. The calculation of the Install
before date is based on the production date of the elctrode.
Danger symbol: "Irritant" (on the label and the packaging of
C3 Fluid Pack)
Rating: Although not corrosive, momentary, longer-lasting, or
repeated contact with skin or mucous membrane may result in
inflammation. Danger of sensitization during contact with
skin (when classified with R 43).
Caution: Avoid contact with eyes and skin, do not inhale
vapors.
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
7
Abbreviations
Definition
ANSI
AQC
BG
Blood gas
CLIA
CLSI
cond
Conductivity
CSA
C1
Calibration Solution 1
C2
Calibration Solution 2
C3
Fluid Pack
dBA
DIL
Diluent
DNS
EC
European community
e.g.
EN
European standard
FMS
Hct
Hematrocrit
HIV
HW
Hardware
i.e.
ISE
IVD
LCD
LIS
LJ
Levey Jennings
Roche Diagnostics
8
MAC
MC
Measuring chamber
MCon
Measuring contact
MSDS
MSS
MV
Mean value
June 2008
Instructions for Use Version 10.0
Abbreviation
Definition
PP
Peristaltic pump
QC
Quality control
REF
Reference solution
SIP
SD
Standard deviation
SDC
SO2
Oxygen saturation
tHb
Total hemoglobin
UL
VDE
e For writing the measuring, calculated and input values see Chapter 8 Operating modes >
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
9
Roche Diagnostics
10
June 2008
Instructions for Use Version 10.0
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-45
June 2008
1 Safety information
Table of contents
Safety information
The information provided in this chapter is essential for the safe, trouble-free
operation of the instrument and must be read and understood by the user.
In this chapter
Chapter
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
A-3
1 Safety information
Table of contents
Roche Diagnostics
A-4
June 2008
Instructions for Use Version 10.0
1 Safety information
Important information
Important information
These Instructions for Use contain vital warnings and safety information.
This instrument is intended to be used only for the specialized purpose described in
the instructions. The most important prerequisites for use, operation, and safety are
explained to ensure smooth operation. No warranty or liability claims will be covered
if the machine is used in ways other than those described or if the necessary
prerequisites and safety measures are not observed.
The instrument may be operated only by persons whose qualifications enable them to
comply with the safety measures that are necessary during operation of the
instrument.
Suitable protective equipment, like laboratory clothing, protective gloves, protective
goggles and if necessary mouth protectors, must be worn to prevent direct contact
with biological working materials. In addition, a face mask is required if there is a
risk.
Adjustments and maintenance performed with covers removed and power connected
may be attempted only by a qualified technician who is aware of the associated
dangers.
Instrument repairs are to be performed only by the manufacturer or qualified service
personnel.
Only accessories and supplies either delivered by or approved by Roche are to be used
with the instrument. These items are manufactured especially for use with this
instrument and meet the highest quality requirements.
Operation of the instrument with solutions whose composition is not consistent with
that of the original solutions can negatively affect the long-term measurement
accuracy. Deviations in the composition of the solutions can also decrease the service
life of the electrodes.
For safety reasons, quality control measurements must be performed daily. Since the
measurements of the instrument depend not only on the correct characteristic
function, but also on a series of marginal conditions (e.g. pre-analysis), results
obtained from the instrument should be submitted for an expert opinion before
taking additional measures based on the supplied measurements.
Caution (refer to accompanying documents)!
Please refer to safety-related notes in the manual accompanying this instrument.
WARNING
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
A-5
1 Safety information
IEC/EN 61010-1:2001
IEC/EN 61010-2-101:2002
It was delivered from the factory in flawless condition with regards to safety features.
In order to preserve this condition and ensure safe operation, the user must respect
the notices and warnings that are contained in these Instructions for Use.
WARNING
Roche Diagnostics
A-6
If objects or liquids enter the internal areas of the instrument, remove the
instrument from its power supply and allow an expert to check it thoroughly
before using it again.
The power cord must be plugged into a grounded power receptacle. When using an
extension cord, make sure it is properly grounded.
Any rupture of the ground lead inside or outside the instrument or a loose ground
connection may result in hazardous operating conditions for the operating personnel.
Intentional disconnection of the grounding is not permitted.
The instrument is not suitable for operation with a direct current power supply.
Use only the original power plug delivered with the cobas b 121 system.
June 2008
Instructions for Use Version 10.0
2 General description
Table of contents
General description
In this chapter
Chapter
June 2008
A-7
2 General description
Table of contents
Roche Diagnostics
A-8
June 2008
Instructions for Use Version 10.0
2 General description
Introduction
Introduction
Figure A-1
The cobas b 121 system is a modular analyzer for measuring blood gases, electrolytes,
total hemoglobin, oxygen saturation and hematocrit in whole blood, serum, plasma,
acetate and bicarbonate containing dialysis solutions, and QC materials.
It is possible to complete database procedures or to make simultaneous adjustments
during measurement or calibration.
The individual, mutually independent operating modes are defined as follows:
o
Database: data about patients, measurement, calibration, QC, and the instrument
Info: Roche info, version number, fill levels, help, sensor report
BG (pH, PO2, PCO2) / ISE (Na+, K+, Cl, Ca2+) and tHb/SO2
A new fluid calibration system eliminates the need for expensive calibration gases.
This change results in easier handling, a smaller footprint, and reduced costs.
An easily understood "Touch Screen" interface facilitates easy operation and saves
costly and time-consuming user training.
The patented electrodes are completely maintenance-free, and only require a very
small sample volume.
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
A-9
2 General description
General notes
General notes
Application area
The instrument is designed to measure BG / ISE / tHb/SO2 in whole blood. The
accuracy of measurement values is checked accordingly.
In order to achieve accurate measurements of recommended aqueous control
solutions (with regards to deviations from biological samples), choose the proper
components and make the corresponding corrections in the QC measurement mode.
The accuracy of measurement values of undefined aqueous solutions cannot be
guaranteed (e.g. due to the possibility of interfering components and/or missing or
insufficient buffer systems, and/or differences in ionic strength and diffusion
potential when compared to biological samples).
Operating instructions
The instrument should be powered on at all times!
If the instrument will remain turned off for longer than 24 hours, it is necessary to
carry out shutdown procedures.
e For more information, see Chapter 3 Installation and shutdown!
Description
Button "Info"
More functions
Roche Diagnostics
A-10
June 2008
Instructions for Use Version 10.0
2 General description
General notes
Buttons
Description
Confirm
Cancel
Activate / deactivate
Start action
e For more information, see Chapter 8 Operating modes, section Buttons on page B-47!
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
A-11
2 General description
Calibration procedure
tHb and SO2 are factory calibrated. Ambient air and a zero point solution are used to
calibrate oxygen. The remaining parameters are calibrated using two solutions, which
are mixed in various ratios. Gas containers are not used. The use of at least two
calibration points for each measurement variable and constant internal monitoring of
the calibrations ensures the accuracy and precision of the measurement values.
Measurement evaluation
The results from the instrument must be carefully examined by a clinical-medical
specialist who will take the patient's clinical condition into consideration before any
clinical decisions are reached based on the test results.
In order to ensure the quality of the measurement results, complete a quality control
test on 3 levels (low, normal, high) after each electrode exchange, after each exchange
of solutions and packs and after startup of the instrument.
Additionally complete a quality control test on one level between two automatic
2P calibrations. The levels have to be alternated (low, normal, high).
e For more information, see Chapter 6 Quality control!
Roche Diagnostics
A-12
June 2008
Instructions for Use Version 10.0
2 General description
Safety instructions for specific dangers
Decontamination
The purpose of this procedure is to minimize risk when handling items that were in
contact with biological samples.
Roche recommends following a decontamination procedure in addition to
regulations specific to the laboratory.
These decontamination procedures should be performed periodically to minimize the
risk of infections (incl. hepatitis virus and HIV).
Always wear gloves!
e For more detailed information about decontamination, see Chapter 9 Maintenance!
Handling solutions
Store the instrument solutions according to the specified packaging requirements.
The solutions should be adapted to the ambient temperature before use.
The shelf life of the solutions is limited.
Please read the bottle label and the packaging for the correct storage temperature and
the maximum shelf life.
CAUTION! DO NOT FREEZE!
If frozen, the solution's concentration may change and cause calibration errors!
ATTENTION
Roche Diagnostics
Instructions for Use Version 10.0
Do not use damaged C3 Fluid Packs! Do not mix the individual components!
June 2008
A-13
2 General description
Handling electrodes
Store the electrodes according to the packaging specifications.
The shelf life of the electrodes is limited.
See the electrode label and the packaging for the correct storage temperature and the
maximum shelf life.
NOTE
Roche Diagnostics
A-14
June 2008
Instructions for Use Version 10.0
2 General description
System description
System description
Visual identification
C
F
D
E
Printer cover
Pump
Flap
Contrast setting
Bottle compartment
Screen
Figure A-2
Screen
All information (results, error messages, warnings, etc.) is displayed on the screen.
The screen consists of a 5.7" colour LCD that is covered with a touch-sensitive film
("touch screen").
Printer
The low-noise 2" thermal printer with integrated paper cutter for roll paper is located
underneath the printer cover.
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
A-15
2 General description
System description
Measurement chamber
The measurement chamber with the electrodes and the tHb/SO2 module are located
beneath the instrument cover.
The electrodes are flow-through electrodes with a visible sample channel.
The tHb/SO2-module is an optical sensor module for determining the levels of total
hemoglobin (tHb) and oxygen saturation (SO2) in whole blood.
Pump
A peristaltic pump transports the sample and the operating fluids inside the
instrument.
Flap
When opening the flap, notice two definitive locking positions:
o
Bottle compartment
The calibration solutions and the waste container are located behind the bottle
compartment cover.
Power supply
This unit contains the power switch and the power connector.
Position of the power switch
Figure A-3
Roche Diagnostics
A-16
June 2008
Instructions for Use Version 10.0
2 General description
System description
Reverse side
Interface
Only data processing units manufactured according to the standards IEC 950
(UL1950) may be attached to the interface connections!
COM 1
Barcode Scanner
COM 2
Network connection
Figure A-4
COM 1*
RS 232 Schnittstelle
COM 2*
RS 232 Schnittstelle
Barcode
Network
*TIP:
Always use a filter adapter when using the serial interface. Order this part from your
customer service representative!
COM 1
COM 2
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
A-17
2 General description
System description
Barcode scanner
Figure A-5
Scanning of solution data (type, lot number, expiration date current composition,
etc.)
Press the button on the underside to activate the scanner! A beeping sound and a brief
illumination of the LED on the upper side indicate the successful scanning of the barcode.
ATTENTION
e For further information, see the enclosed manual for the PS2 hand-held scanner (included
in scope of delivery).
Figure A-6
Identification label
A
A
Identification label
Figure A-7
Roche Diagnostics
A-18
June 2008
Instructions for Use Version 10.0
In this chapter, the software-guided installation and shutdown of the instrument are
described step by step. The sequence of the steps described must be strictly followed.
In this chapter
Chapter
June 2008
A-19
Table of contents
2. Remove the V2 and V9 tube ends from the guides .................................. A-37
3. Insert the shutdown kit and connect it .................................................... A-37
4. Remove the shutdown kit .......................................................................... A-38
5. Remove the waste container ...................................................................... A-38
6. Return the V2 and V9 tube ends to the guides ........................................ A-38
7. Remove the pump tube ............................................................................. A-39
8. Remove the electrodes ............................................................................... A-39
9. Remove needle and fill port ...................................................................... A-40
10. Remove the printer paper ....................................................................... A-40
11. Remove the tube under V1 and V2 ........................................................ A-41
12. Shutting down the AutoQC module ...................................................... A-41
13. Uninstall the AutoQC module ................................................................ A-42
Roche Diagnostics
A-20
June 2008
Instructions for Use Version 10.0
Installation
Location
For best results, choose a suitable, level location that is not subject to direct sunlight
for the device.
When installing an instrument that was stored in a cool room or was transported at
low temperatures, be aware that condensation may have formed and could cause
disturbances to the instrument. Be sure that the instrument is climatized for at least
one hour at room temperature before beginning operation.
The following conditions must be fulfilled:
o
Ambient temperature: + 15 C to + 33 C
An atmospheric pressure of 462 mmHg (61.63 kPa) corresponds to approx. 4.000 m above
sea level.
Use a stable and level work surface (max. 1 incline with bottles installed).
Relative humidity:
O
20 to 95 % (at +15 C to 31 C)
Allow sufficient free space around the instrument for air circulation and the
electrical connections.
After setting up the instrument at a location that meets the necessary conditions,
execute the following steps to ensure the instrument is ready for operation:
Roche Diagnostics
Instructions for Use Version 10.0
Check that the instrument and all accessories are present by comparing contents
with the delivery packing slip. If anything is missing, inform your Roche
representative immediately.
Inspect the instrument and the accessories for damage. If damage is noted, inform
the transportation company immediately. Retain the packaging material and
products as evidence for the damage claim.
June 2008
A-21
Installation
Accessories
The following parts are delivered as standard equipment with the instrument:
1 barcode scanner
1 shutdown kit
1 power cord
2 pump tubes
1 drip tray
2 needles
1 paper roll
2 MCon
Installation
1. Attach power cord and barcode scanner
1 Plug the power cord into the power supply (Figure A-8/1).
2 Connect the barcode scanner and. if necessary, the network cable (Figure A-8/2)
to the respective interface(s) (Figure A-8/2) on the reverse side of the instrument.
(2) Interfaces
Figure A-8
Docking mechanism
Figure A-9
Roche Diagnostics
A-22
June 2008
Instructions for Use Version 10.0
ATTENTION
Attempt to install and operate an AutoQC module only if the instrument is prepared for use
with the AutoQC module. The valves V12 and V13 must be present in the bottle compartment (see Figure A-10)!
Not prepared
Prepared
Figure A-10
Check to be sure that the power cord is connected to the power supply. After engaging the
AutoQC module, it will no longer be possible to connect the cord to the power supply!
ATTENTION
1 Open the AutoQC cover and remove the transport safety band and both safety
screws (see Figure A-11/1).
The safety screws are intended to fix the longitudinal and transversal slides and should
only be re-inserted whenever the AutoQC module will be transported.
2 Remove the red plastic relief clamp by lifting from the AutoQC valve V17, which is
located toward the back of the AutoQC unit on the inside (see Figure A-11/2).
B
A
(1)
A
(2)
B
Figure A-11
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
A-23
Installation
3 On the right side of the instrument. remove the gray plastic cover from the
AutoQC docking port. Close the AutoQC cover and push the AutoQC module
into the tracks on the instrument until it locks into place. The module's docking
port will slide into the opening on the side wall on the instrument and lock into
place.
Figure A-12
4 Connect both tube ends in the bottle compartment to the docking part according
to Figure A-13.
Figure A-13
5 Connect the cable of the AutoQC module to the jack on the reverse side of the
instrument.
Roche Diagnostics
A-24
June 2008
Instructions for Use Version 10.0
Figure A-14
3. Switch on
Switch the instrument on and wait until the program has loaded completely and has
started. The instrument is in the "System stop" mode. Before beginning the start up
procedures, you must select the language with which the instrument will be operated,
set the date and time, verify/activate the parameters tHb and SO2, check the barometer value and if the AutoQC module is activated.
Should a hardware problem occur during the booting process, the system is stopped and
an error message is given.
The Customer Service should be informed if the error occurs again.
4. Select language
1 Press:
h [Setup] > [Instrument] > [Language]
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
A-25
Installation
Ventilation tube
Figure A-15
Roche Diagnostics
A-26
June 2008
Instructions for Use Version 10.0
A
A
Figure A-16
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
A-27
Installation
B
C
Pump closed
A
Tension lever
Pump head
Linear clamp
Figure A-17
Figure A-18
3 Rotate the fill port holder so that the underside shows on the top (see
Figure A-19/1) and then push this over the needle (see Figure A-19/2).
Do not bend the needle when pushing it up!
4 Snap the fill port holder evenly onto the axis (see Figure A-19/3).
Roche Diagnostics
A-28
June 2008
Instructions for Use Version 10.0
(1)
(2)
(3)
Figure A-19
Printer lever
Figure A-20
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
A-29
Installation
6 Close the printer cover and feed the paper outward through the slit in the printer
cover.
Figure A-21
1 To do this. place the bottle tool on the screw cap of the C1 or C2 (Figure A-22/1).
2 Press the grips together and press the transparent disk downward Figure A-22/2).
3 Rotate the transparent disk in the clockwise direction. Stop when you notice (after
a short distance) resistance (Figure A-22/3).
(1)
(2)
(3)
Figure A-22
Roche Diagnostics
A-30
June 2008
Instructions for Use Version 10.0
Figure A-23
5 The instrument recognizes the correct solution and checks the expiration date.
On the screen, the respective bottle starts blinking. If the bottle has passed the
expiration date, the screen displays a warning.
6 Confirm with this button.
7 Insert the bottle completely into the appropriate position, following the
instructions on the docking mechanism (see Figure A-24/2). The bottles are
opened automatically.
8 Close the docking mechanism.
Check the positions of the bottles by matching the labels on the docking
mechanisms to the labels on the bottles (see Figure A-24/3).
(1)
(2)
Figure A-24
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
A-31
Installation
Figure A-25
Make sure that no air bubbles have formed in the inner electrolytes of the electrodes (see
Figure A-26). If there are air bubbles between the contact pin and the membrane, there
will not be effective electrical conduction. Result: calibration and measurement errors!
4 If necessary, remove air bubbles by holding the electrode vertically and by tapping
lightly with your fingernail against the electrode body (see Figure A-26).
Figure A-26
5 Insert the electrodes, beginning at the right and proceeding left according to the
colour code.
6 Push all electrodes slightly to the right so that they are lined up together without
gaps.
Roche Diagnostics
A-32
June 2008
Instructions for Use Version 10.0
Figure A-27
2 Place the tube into the tube guide slot (seeFigure A-28/2).
Be sure that the tube lies precisely in the guide slot. Otherwise it may be pinched, thereby
preventing aspiration of the reference solution.
Result: calibration and measurement errors!
B
(1)
Locking lever
Reference electrode
(2)
Figure A-28
3 Affix the white plug at the end of the reference electrode tube (see Figure A-29).
Push the plug in completely.
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
A-33
Installation
A
A
Figure A-29
Figure A-30
6 Scan the barcodes located on the inner packaging of each electrode or enter the
barcodes manually with the help of the keyboard.
Always save the inside packaging materials! Whenever the instrument is taken out of
service and then brought back into service, the barcodes of the remaining functional
electrodes must be scanned again!
Figure A-31
June 2008
Instructions for Use Version 10.0
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
A-35
Shutdown
Shutdown
Less than 24 hours
If the instrument will be needed within the next 24 hours, press:
h [More functions] > [System] > [Tools] > [Software shutdown]
Switch off the instrument.
description!
1. Remove bottles
Open the bottle compartment cover and remove bottles C1, C2 and C3.
Roche Diagnostics
A-36
June 2008
Instructions for Use Version 10.0
Ventilation tube V9
Figure A-32
3 Place both ends of the tube onto an absorbent pad, such as a paper towel or gauze
(see Figure A-33).
A
A
Figure A-33
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
A-37
Shutdown
Figure A-34
2 Start the procedure for emptying the tubes by pressing the [Start] button.
Ventilation tube
Figure A-35
June 2008
Instructions for Use Version 10.0
B
C
Tension lever
Pump head
Linear clamp
Figure A-36
A
A
Locking lever
Figure A-37
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
A-39
Shutdown
3 Carefully remove the fill port holder from the needle (see Figure A-38/2).
4 Remove the needle.
5 Push the needle to the left and into the sealer. Then carefully pull it upward
(see Figure A-38/3).
(1)
(2)
(3)
Figure A-38
Roche Diagnostics
A-40
June 2008
Instructions for Use Version 10.0
A
A
Figure A-39
If available (option):
12. Shutting down the AutoQC module
1 Open the cover and remove the AutoQC ampoule block.
2 The carriage in the AutoQC module moves to the service position.
3 Close all covers.
After successfully shutting down the instrument, it will be in the system stop mode (shut
down). Only a renewed installation procedure can alter this status.
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
A-41
Shutdown
A
A
Unlocking knob
Figure A-40
2 Remove both tube ends in the bottle compartment from the docking port.
A
A
Tube ends
Figure A-41
3 Unlock the AutoQC module by pulling the unlocking knob (see Figure A-40) on
the reverse side of the instrument and remove the AutoQC module from the
tracks on the instrument.
Roche Diagnostics
A-42
June 2008
Instructions for Use Version 10.0
4 Insert the red plastic relief clamp for the AutoQC valve V17.
A
A
Figure A-42
5 If you plan to transport the instrument, be sure to unplug the power cord, the
scanner, and the network cable. Then install the transport safety device for the
AutoQC module.
A
A
Figure A-43
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
A-43
Shutdown
Roche Diagnostics
A-44
June 2008
Instructions for Use Version 10.0
4 Specifications
Table of contents
Specifications
In this chapter, the performance data, as well as product and environmental data are
described.
In this chapter
Chapter
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
A-45
4 Specifications
Table of contents
Roche Diagnostics
A-46
June 2008
Instructions for Use Version 10.0
4 Specifications
Performance data
Performance data
Measured parameters
specified
Precision(a) (b)
(within-run)
(day-day)
0 - < 60 mmHg
60 - 140 mmHg
4.0 mmHg
0.533 kPa
0 - < 60 mmHg
60 - 200 mmHg
9.0 mmHg
1.200 kPa
Specified range
for
PO2
PCO2
B/Q
pH
Na
B/Q
SD < 2 mmHg
SD < 3 mmHg
4 mmHg
15 - 80 mmHg
2.5 mmHg
SD < 6 mmHg
8 mmHg
0.533 kPa
0.333 kPa
1.066 kPa
SD < 0.020
SD < 0.035
0.06
6.8 - 7.6
SD < 0.008
SD < 0.015
0.02
SD < 0.015
SD < 0.030
0.04
SD < 6 mmol/L
8 mmol/L
2.5 mmol/L
Cl-
4 - < 15 mmHg
Accuracy(a)
SD < 6 mmol/L
SD < 8 mmol/L
10 mmol/L
0.5 mmol/L
0.2 mmol/L
SD < 1 mmol/L
1.4 mmol/L
SD < 3 mmol/L
SD < 4 mmol/L
8.0 mmol/L
70 - 130 mmol/L
4.0 mmol/L
SD < 6 mmol/L
12.0 mmol/L
Table A-1
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
A-47
4 Specifications
Performance data
specified
Ca2+
(within-run)
(day-day)
0.2 mmol/L
0.1 mmol/L
0.5 mmol/L
0.8016 mg/dL
0.4008 mg/dL
2.0040 mg/dL
10 - < 20%
SD < 3%
SD < 5%
6.0%
20 - 60%
SD < 1.5%
SD < 2.5%
3.0%
< 60 - 80%
SD < 3%
SD < 5 %
6.0%
SD < 0.030
SD < 0.050
0.060
0.200 - 0.600
SD < 0.015
SD < 0.025
0.030
SO2
Q
Baro
Accuracy(a)
for
tHb
B/Q
Hct
Precision(a) (b)
Specified range
(c)
SD < 0.030
SD < 0.050
0.060
3 - < 6 g/dL
6 - 18 g/dL
> 18 - 25 g/dL
SD < (2.1% + 0.04 mmol/L) SD < (2.1%+0.07 mmol/L) < (7.1%-0.48 mmol/L)
3 - < 12 g/dL
12 - 25 g/dL
SD < 3.0%
SD < 3.3%
< 3.0%
SD < 3.0%
SD < 3.3%
< 3.0%
SD < 3.0%
SD < 3.3%
< 3.0%
50 - < 60%
60 - 100%
SD < 1.0%
----------------------------
< 2.0%
50 - 100%
SD < 1.0%
SD < 1.1%
< 2.0%
----------------------------
----------------------------
Table A-1
(a) Accuracy and standard deviations: specified for the standard temperature range of 15 - 31 C!
(b) According to CLSI within the specific range
(c) When used within the range of 300...450 mmHg, the barosensor must be calibrated by technical support before initial start up!
B....whole blood
Q....aqueous QC material
Roche Diagnostics
A-48
S...serum or plasma
June 2008
Instructions for Use Version 10.0
4 Specifications
Default and input values
P50 Adult
Default values
15.0 g/dL
150.0 g/L
9.3 mmol/L
26.7 mmHg
not editable
not editable
not editable
not editable
3.56 kPa
21.5 mmHg
P50 Fetal
2.87 kPa
FIO2
0.21
0.1 - 1.0
0.1 - 1.0
0.84
0.7 - 2.0
0.7 - 2.0
Patient temperature
37.0 C
2.0 - 44.0 C
2.0 - 44.0 C
98.6 F
35.6 - 111.0 F
35.6 - 111.0 F
Table A-2
Sample throughput
Activated / installed
electrodes
[samples / hours]
Syringe
Capillary
BG - tHb/SO2
30
30
BG - ISE - tHb/SO2
30
30
Table A-3
Sample volume
If the sample contains a higher Hct concentration, a sample volume of up to 70 L is
required!
Activated / installed electrodes
BG - ISE - tHb/SO2
[L](a)
(capillary position)
(syringe position)
[L](b)
[L](c)
68
90
60
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
A-49
4 Specifications
Sample types
Sample types
o
Whole blood
Serum
Plasma
QC material
Calibrations
Calibrations
Time intervals
Length (typical)
[min]
Sys cal
every 24 hours
(alternatively 8, 12 or 24 hours)
< 15
1P cal
<2
2P cal
every 12 hours
(alternatively 4, 8, 12 hours)
<7
Warm-up phase
< 25
Warm-up phase
<2
as needed
< 27
Electrode exchange
Table A-4
(a) incl. calibration
Roche Diagnostics
A-50
June 2008
Instructions for Use Version 10.0
4 Specifications
Environmental parameters
Environmental parameters
Temperature / humidity / stability
Instrument
Operating conditions
Ambient temperature
+15 C to +33 C
Sea level
-400 m to +4000 m
Relative humidity
Temperature
-20C to 50 C
Humidity
Shock resistance
< 30 g
Electrodes
Operating conditions
Ambient temperature
Relative humidity
+15 C to +33 C
20 - 95%, if T > +15 C to +31 C
20 - 90%, if T > +31 C to < +33 C
Temperature
+15 C to +30 C
Humidity
Roche Diagnostics
Instructions for Use Version 10.0
Temperatur
+2 C to +30 C
Humidity
20 to 80 % (not condensed)
over a period of 3 days
Shock resistance
< 50 g
June 2008
A-51
4 Specifications
Environmental parameters
Solutions
Operating conditions
Ambient temperature
+15C to +33 C
Relative humidity
Temperature
Relative humidity
20 to 95%
(a) storage time contains transportation time and the storage from delivery from the factory!
Temperatur
Relative humidity
20 - 95%
Shock resistance
< 30 g
Description
C3 Fluid Pack
Solution pack
QC material
Storage conditions in original packaging
COMBITROL TS+
AUTO-TROL TS+
COMBITROL TS+
up to 24 months at 2C to 8 C
AUTO-TROL TS+
up to 24 months at 2C to 8 C
or
COMBITROL TS+
AUTO-TROL TS+
Roche Diagnostics
A-52
AUTO-TROL TS+
June 2008
Instructions for Use Version 10.0
4 Specifications
Product data
Product data
Electrical data
Mains voltage range:
Frequency:
50/60 Hz
Required power:
150 W
Overvoltage category:
II
Contamination level:
Width:
35.4 cm
Dimensions
Instrument
AutoQC-Modul
Heigth:
46.7 cm
Depth:
41.0 cm
Width:
19.5 cm
Heigth closed:
Depth:
39.5 cm
Weight
cobas b 121 system (instrument):
Printer
Roche Diagnostics
Instructions for Use Version 10.0
Type:
Resolution:
12 dots/ mm
Full graphics:
Printing speed:
8 mm / sec
Paper width:
58 mm
Paper length:
ca. 30 m
June 2008
A-53
4 Specifications
Screen
Screen
Type:
Format:
5.7 inch
Resolution:
Type:
Reading speed:
up to 45 scans/sec.
Resolution:
0.1 mm
Reading distance:
up to 5 cm
Barcode scanner
Reading width:
up to 8 cm
Codabar
Code 39
Code 128
EAN-8
EAN-13
EAN-128
Interleaved 2 of 5
UPC-A
UPC-E
(a) Further available barcode types can be programmed in accordance with the enclosed manual of the
PS2 hand-held scanner (included in scope of delivery).
Roche Diagnostics
A-54
June 2008
Instructions for Use Version 10.0
Operation
Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-35
June 2008
5 Measurement
Table of contents
Measurement
In this chapter, all information necessary for carrying out measurements is described.
In this chapter
Chapter
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-3
5 Measurement
Table of contents
Roche Diagnostics
B-4
June 2008
Instructions for Use Version 10.0
5 Measurement
Preanalytics
Preanalytics
Sample collection
Follow the usual applicable safety precautions when drawing blood samples. When
handling blood samples, there always exists the danger of transmission of HIV, hepatitis B
and C viruses or other pathogens transmissible by blood. Employ suitable blood sampling
techniques in order to reduce risk to personnel.
Suitable protective equipment, like laboratory clothing, protective gloves, protective
goggles and if necessary mouth protectors, must be worn to prevent direct contact with
biological working materials. In addition, a face mask is required if there is a risk.
e Please refer to CLSI document M29-A3, "Protection of Laboratory Workers from
Sample acquisition
Only qualified personnel may perform the collection of blood needed for analytical
purposes.
The puncture site may never be squeezed! Mixing the blood sample with tissue fluid may
lead to the premature onset of clotting despite sufficient heparinization of the sample
collection containers! Incorrect sample collection or the use of an unsuitable sample
collection container may lead to errors and discrepancies in the measurement values.
e For detailed information about blood sampling, storage and handling, see for example
CLSI document H11-A4, "Procedures for the collection of arterial blood specimes;
Approved Standard (Fourth Edition 2004)".
Acceptable anticoagulants
The only clot inhibitors that may be used for analyses in the cobas b 121 system are
heparin salts. Other clot inhibitors, such as EDTA, citrate, oxalate, fluoride, and
ammonium-based materials have significant influence on the blood's pH and other
parameters and may not be used for this reason.
Sample collection especially for tHb, SO2 and Hct measurement
Whole blood, especially for the analysis of tHb, SO2 and Hct, must be thoroughly
mixed immediately before analysis in order to achieve consistent distribution of red
blood cells and plasma before insertion of the sample.
Carefully rotate the sample about two axes using your hand or a mechanical device or
insert a metal disk or ball in the syringe before collecting the sample. Shortly before
using the sample, carefully shake the syringe. The up and down motion of the disk or
ball inside the syringe cylinder ensures consistent mixing.
e Refer to CLSI document C46-A "Blood gas an pH analysis related measurements;
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-5
5 Measurement
Preanalytics
Syringes
If using another manufacturer's product with liquid heparin as a clot inhibitor, the
collection container should not be larger than required for the blood volume. This
will minimize the effects of the clot inhibitor on the thinning of the blood. The use of
plastic syringes is common, but there are cases when the use of plastic syringes is not
appropriate, for example, when PO2 values are expected to be outside the normal
range. If very high PO2 values are expected, the sample should be analyzed as quickly
as possible after the specimen collection.
Use only heparinized syringes. Improper use of syringes with liquid heparin will affect the
parameters, especially the ISE parameters!
Capillary tubes
The capillary tubes must have a minimum volume of 100 L.
Capillary tubes with ceramic sealing caps should not be used because the fracture that
forms when opening the capillary can damage the fill port of the instrument. Use only
capillary tubes with heat-treated ends to avoid damage to the instrument. When using
stirring rods like those offered by a few manufacturers, remove these rods before
inserting the sample in order to avoid clogging the instrument.
Roche MICROSAMPLER
The Roche MICROSAMPLER, which consists of two capillary tubes (115 L each) in
a plastic container, is ideally suited to atraumatic arterial blood collection.
Each laboratory should document the permissibility of sample containers that are
used. These products vary from manufacturer to manufacturer and sometimes from
lot to lot.
The use of sample containers or clot inhibitors other than those manufactured by Roche
Diagnostics may lead to adulteration of the samples and errors and differences in the
measurement values.
Roche developed a specialized sample collection container for this purpose and
recommends its use for this reason.
Clot catcher
The use of a clot catcher is recommended to prevent clogging of the sample path during measurement of critical blood, for example, when sampling blood of newborns,
blood from ear lobes, and blood from heels.
The clot catcher prevents blood clots and tissue particles from entering the instrument.
The clot catcher cannot be used in "Syringe mode" and not be connected in
"Capillary mode".
ATTENTION
Roche Diagnostics
B-6
June 2008
Instructions for Use Version 10.0
5 Measurement
Preanalytics
Sample handling
Whole blood
Withdraw whole blood samples using heparinized syringes, capillaries, or the
Roche MICROSAMPLER. Analyze the samples as soon as possible after sampling.
Remove air bubbles from the sample collection container immediately after the sampling procedure.
Immediately after withdrawing the sample with syringes, thoroughly mix the sample
with anticoagulant. This can be done by rolling the sample between both hands or
shaking. Properly label the samples, following the standard documentation procedure.
o
Samples with a PO2 level above 200 mmHg (26 kPa) should be collected in a glass
container if the measurement can not be performed within 15 minutes.
When using capillaries analyze samples for tHb, SO2 and Hct measurements immediately
after sampling to ensure correct and accurate measurement results.
Despite proper sampling procedures, errors can arise in the blood gas analysis:
o
due to insufficient mixing of the sample following removal and before the
measurement
due to ambient air contamination caused by air bubbles that are not removed
following removal of the sample
Serum
After the appearance of spontaneous clotting, process the sample in a centrifuge to
separate the cellular, solid components and the fibrin from the watery serum. Transfer
the serum to a suitable sample container and seal.
If it is necessary to store the sample, close the sample container tightly and cool it to
4 - 8 C. If a sample has been cooled, warm it to room temperature (15 - 33 C) before
analysis.
Plasma
Plasma samples are obtained by centrifuging heparinized whole blood, during which
the cellular components of the blood are removed from plasma.
Complete the analysis as quickly as possible.
If it is necessary to store the sample, close the sample container tightly and cool it to
4 - 8 C. If a sample has been cooled, warm it to room temperature (15 - 33 C) before
analysis.
Plasma samples older than 1 hour must be re-centrifuged in order to remove fibrin
clumps that may have formed.
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-7
5 Measurement
Measuring procedure
Measuring procedure
QC measurements must be performed in their entirety (i.e., all three QC levels must be
measured). Omitting QC measurements or ignoring QC measurement results may lead to
incorrect patient measurements, which may result in incorrect clinical decisions, possibly
endangering the patient's health.
e For detailed information, see Chapter 6 Quality control!
Mandatory input
Figure B-1
You have the option of measuring samples from syringes (without needles), ampoules
and capillaries.
Roche Diagnostics
B-8
June 2008
Instructions for Use Version 10.0
5 Measurement
Measuring procedure
Syringe mode
Improper heparinization of syringes with liquid heparin may cause false results. ISE
parameters are particularly susceptible.
Figure B-2
If the opening of the needle is covered by the fill port, the aspiration is blocked and the
sample is rejected!
If using an ampoule instead of a syringe (during a QC measurement), be sure that the
ampoule does not make contact with the fill port when inserting the needle into the
ampoule. The sharp edge can cause damage to the fill port.
The clot catcher cannot be used in "Syringe mode".
Figure B-3
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-9
5 Measurement
Measuring procedure
Figure B-4
4 Upon display of the instruction "Close flap", remove the syringe and close the flap.
5 The measurement starts.
Never open the flap during measurement. Doing this will result in rejection of the sample!
Capillary mode
1 Open the flap to the labelled capillary position (completely open position).
The aspiration process is started.
Open the flap slowly to avoid splattering small drops of liquid that are on the needle tip.
Figure B-5
Roche Diagnostics
B-10
June 2008
Instructions for Use Version 10.0
5 Measurement
Measuring procedure
2 Insert the capillaries or the Roche MICROSAMPLER into the fill port.
The end of the capillary must be open, otherwise the aspiration process is blocked and the
sample is rejected.
WARNING
Figure B-6
Figure B-7
4 Upon display of the instruction "Close flap", remove the capillary and close the
flap.
5 The measurement starts.
Never open the flap during measurement. Doing this will result in rejection of the sample!
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-11
5 Measurement
Measuring procedure
Data input
During measurement, various patient, operator, and sample-specific data may be
entered. You may use a standardized or a user-specific form when entering data.
Scanning of patient and operator data is possible!
The user must carry out a plausibility check for all barcode data read in and displayed by
the instrument!
You can create a user-specific form that contains the desired parameters by selecting
h [Setup] > [Displays & reports] > [Measuring data] > [Input values]
e See Reference Manual, Chapter Setup!
Press the [More functions] button to select the predefined form. This form remains
the standard until a new form is selected.
A
More functions
Figure B-8
Use the [line up] / [line down] buttons to select the entry that you would like to
modify or use your finger to select the appropriate line directly.
By pressing these buttons, you can complete the following steps:
Input data
Edit the underlying entry
Go to additional views
Start a printout
Roche Diagnostics
B-12
June 2008
Instructions for Use Version 10.0
5 Measurement
Measuring procedure
Figure B-9
Instructions for Use and to the respectively titled chapters in the Reference Manual!
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-13
5 Measurement
Measuring procedure
Roche Diagnostics
B-14
June 2008
Instructions for Use Version 10.0
6 Quality control
Table of contents
Quality control
For safety reasons, quality control measurements must be carried out on a daily basis.
In this chapter, all steps are described that are necessary for a successful QC
measurement.
In this chapter
Chapter
June 2008
B-15
6 Quality control
Table of contents
Roche Diagnostics
B-16
June 2008
Instructions for Use Version 10.0
6 Quality control
Quality control - general
WARNING
General QC concept
Roche Diagnostics always strives to ensure the highest quality standards for its
products. This quality awareness is the result of a sense of responsibility toward the
customer and the well-being of the patient.
The quality control is an important element of this claim. Aqueous blood gas/
electrolyte QC materials, such as COMBITROL TS+, AUTO-TROL TS+, etc., are
offered to ensure that the instrument provides measurements of high quality to
protect customers or their patients.
In order to ensure the quality of the measurement results, complete a quality control
test on 3 levels (low, normal, high) after each electrode exchange, after each exchange
of solutions and packs and after startup of the instrument.
Additionally complete a quality control test on one level between two automatic
2P calibrations. The levels have to be alternated (low, normal, high).
For example (2P calibration interval: 12 hours):
24 hours
--------
2P-Cal.
Level 1
2P-Cal.
Level 2
2P-Cal.
Level 3
2P-Cal.
Level 1
Figure B-10
NOTE
Complete at least two quality control tests on different levels once daily or more often
in accordance with local regulations.
Run quality control tests ideally prior to sample measurements.
A quality control program includes the analysis of sample materials with known
ranges of expected values and the comparison of these values with analyzer results.
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-17
6 Quality control
General QC concept
COMBITROL TS+
The target areas listed in the package text should be taken as 2SD areas
(SD = standard deviation) (e.g. for PO2, 2SD = 12 mmHg, 1SD = 6 mmHg).
The QC measurement results within the target value range 2SD are acceptable.
If QC measurement results fall outside the target value range 3SD, the parameter
must be locked!
QC measurement results that are greater than the target value 2SD, but less than the
target value 3SD, cause QC warning and must be treated accordingly .
Roche Diagnostics
B-18
June 2008
Instructions for Use Version 10.0
6 Quality control
Important information concerning the analysis of QC measurement results
With software version 1.31 and later, a Multirule violation automatically generates a
parameter lockout, as long as the default settings were not changed.
e See section Multirules on page B-28!
e See section QC consequences on page B-30!
The evaluation depends upon which SD areas are featured in the QC measurement
results:
o
Measured value is larger than target value 2SD, but less than target value 3SD
Consequence: A "QC Warning" is assigned to the parameter.
The user must now analyze the QC measurement results in accordance with
applicable regulations or repeat the measurement.
Call up the QC statistics in the QC database to aid in the analysis.
e See Chapter 8 Operating modes, section QC data on page B-58!
If a second measurement is greater than target value 2SD, but less than target
value 3SD, the parameter is not locked, but must not be used for further patient
measurements.
To eliminate the error, replace the electrode and/or contact technical support.
WARNING
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-19
6 Quality control
Material setup
Material setup
Take the lot number, expiration date, sample type, and target values (ranges) as well as the
corresponding barcodes from the text included in the recommended QC material.
Figure B-11
Use the [line up] / [line down] button to select the material to be changed or to select
a new "QC material".
Using the barcode scanner to
enter data
Use the barcode scanner to enter the material code found on the packaging insert.
The material code contains the information for the material, the proper level, lot
number, expiration date, and sample type.
Scan in two barcodes (the BG and tHb/SO2 code and the ISE code) for the set ranges.
The instrument automatically assigns these.
Manual entry
Use the [Data input] button to manually enter lot number, expiration date, sample
type and target values.
Complete and save the entry by pressing this button twice.
Roche Diagnostics
B-20
June 2008
Instructions for Use Version 10.0
6 Quality control
Material setup
Inserted mats
(1)
(2)
Figure B-12
4 Place the mat in the assigned position (A-F). When fully inserted only the rubber
mat will be seen, the glass ampoules will no longer be visible (see Figure B-12/2).
5 Repeat the same process for all additional mats.
6 Close the cover of the AutoQC module.
If only 3 full ampoules are left in a mat for a QC material, it is labeled as used.
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-21
6 Quality control
Material setup
Figure B-13
1 Use the [line up] / [line down] buttons to select the mat to be defined (A-F).
2 The selected material/level combination is assigned to this mat location by
pressing [Material].
[New mat]
An existing and defined mat is replaced by a new one from the same batch (e.g. in case
of insufficient or empty ampoules). The number of ampoules is reset to 20.
The selected material can be deleted from the mat position.
Figure B-14
This screen shows detailed information about the AutoQC material placed on this
mat.
Roche Diagnostics
B-22
June 2008
Instructions for Use Version 10.0
6 Quality control
Material setup
Press the [Details] button again to change to the mat assignment screen.
Figure B-15
Setting QC times
Depending on the selected material, this function is used to select the start time(s) for
the AutoQC measurement(s) and/or the time for performing a manual QC
measurement. After reaching the set time, a note appears in the message window.
Figure B-16
Up to 16 time entries per day can be made to initiate when an AutoQC measurement
should be started. Up to 6 materials (= 6 mats for the AutoQC) can be specified for
each time.
A little marker (small magenta-colored triangles) on the time scale indicates the
defined start time(s).
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-23
6 Quality control
Material setup
Figure B-17
1 Select the material/level combination from the list (this list contains all material/
level combinations that were created under [Setup] > [QC Material] > [AutoQC
Material] or [Setup] > [QC Material] > [Set ranges]).
2 Confirm the selection.
Roche Diagnostics
B-24
June 2008
Instructions for Use Version 10.0
6 Quality control
Material setup
Measurement retries
The number of repeated measurements (none=0, 1, 2) that are allowed in the case of
an incorrect measurement can be adjusted here. A faulty measurement is present if
the measurement lies outside the display range or no measurements can be output
due to a suction fault.
Figure B-18
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-25
6 Quality control
QC measurement
QC measurement
In order to ensure the quality of the measurement results, complete a quality control
test on 3 levels (low, normal, high) after each electrode exchange, after each exchange
of solutions and packs and after startup of the instrument.
Additionally complete a quality control test on one level between two automatic
2P calibrations. The levels have to be alternated (low, normal, high)
e See section General QC concept on page B-17!
Manual QC measurement
Make sure, that the ampoules do not come directly out of the refrigerator or from a hot
spot. They should rest for approximately 24 hours to slowly adapt to room temperature
(25 C, 77 F).
Figure B-19
Roche Diagnostics
B-26
June 2008
Instructions for Use Version 10.0
6 Quality control
QC measurement
Figure B-20
When inserting the needle into the ampoule, it is absolutely necessary to avoid contact
between the ampoule and the filling port. The sharp edge can cause damage to the fill
port.
Figure B-21
AutoQC measurement
The AutoQC measurement can be performed in programmed or manual mode.
To start measurement of the control material, activate the corresponding AutoQC
material (AUTO-TROL TS) and the selected level (e.g. level 1).
Start the AutoQC measurement by pressing [AutoQC].
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-27
6 Quality control
Multirules
Multirules
The evaluation of QC results is based on the Westgard(a) rules and their interpretation
for blood gas analysis(b). The Multirule process was derived from these rules. It
permits early detection of random and systematic errors associated with the
measuring device and its operation.
The Multirules procedure can only be applied in connection with a suitable control
material (e.g. COMBITROL TS+, AUTO-TROL TS+).
ATTENTION
The Multirule procedures produce the best results when 3 QC measurements with
randomly selected levels are completed per series (time between two 2-point
calibrations). A minimum of 2 QC measurements / series or 6 QC measurements /
3 series is required.
The QC concept expects Multirules rule 1 and 2 to be activated.
Press the following buttons to check the settings:
h [Setup] > [Parameters] > [Miscellaneous settings] > [Multirules]
Figure B-22
Select additional desired rules in the left window and assign it to the corresponding
parameter which is listed in the right window under "Parameters".
NOTE
The activation of range 2SD automatically deactivates all other rules (rules 1-6).
(a) James O. Westgard, et al: A Multi-Rule Shewhart Chart for Quality Control in Clinical Chemistry.
Clinical Chemistry, Vol. 27, No.3, 1981
(b) Elsa F. Quam BS, Lorene K. Haessig BS, Marlene J. Koch BS: A Comprehensive Statistical Quality
Control Program for Blood Gas Analyzers. Journal of Medical Technology 2:1 January 1985
Roche Diagnostics
B-28
June 2008
Instructions for Use Version 10.0
6 Quality control
Multirules
NT
NL
standard deviation
Rule
Description
1. 12
2. 13
3. (2 von 3)2
4. 22
5. 61
6. 9m
2SD range
Table B-1
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-29
6 Quality control
QC consequences
QC consequences
By default, the QC consequence "QC error" should be assigned to all parameters.
Press the following buttons to set or check the assigned QC consequences:
h [Setup] > [Parameter] > [Miscellaneous settings] > [QC Consequences]
Figure B-23
Roche Diagnostics
B-30
June 2008
Instructions for Use Version 10.0
6 Quality control
QC unlock
QC unlock
QC warning
Performing a QC measurement using the same material/level combination removes a
QC warning.
QC error
A locked parameter may be unlocked only if the cause of the lockout is known and the
error is corrected (e.g., timeout or measurement of wrong ampoule).
page C-20!
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-31
6 Quality control
QC troubleshooting
QC troubleshooting
Description of the current problem
After a QC measurement, one or more parameters are assessed as "not OK" (QC
warning or QC block).
The relevant parameters are shown accordingly in the "Ready" screen. When the
respective parameter button is pressed, a status report appears.
e See Chapter 10 Troubleshooting, section Electrode status!
The QC problem can only be solved by a correct QC measurement within the range if
the same material / level combination is measured.
Classification of QC problems
Group A
The cause is an aspirating or positioning problem with the QC sample. In this case,
usually more than one parameter is affected. A cause belonging to Group A can be
identified in the data base under "QC data" when an error message appears for the
corresponding parameter instead of a result.
Group B
The cause is a QC result that exceeds the target value range.
A cause belonging to Group B can be identified in the data base under "QC data"
when there is a QC result, but it exceeds the target value range.
If the manual QC measurement is "OK", clean the sample port, needle and
washing plate and compare the ampoule status in the AutoQC mat detail
screen with the actual availability of the ampoules in the AutoQC module.
e See Figure B-17 on page B-24!
June 2008
Instructions for Use Version 10.0
6 Quality control
QC troubleshooting
It must be checked whether the target value ranges under [Setup] > [QC
material] > [QC material] > (select appropriate material) > [Ranges]
correspond to the target value ranges stated in the package insert.
It must be checked whether before use the QC ampoules have been stored for
at least 24 hours at room temperature or in the AutoQC module.
If the error persists, call up an "Internal cleaning of sample path" ([System] >
[Wash & clean] > [Automatic routines]). Then, carry out a blood sample
measurement in order to wet the liquid paths.
If the error persists, replace the affected electrode/sensor. If all parameters are
affected at the same time, replace the reference electrode.
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-33
6 Quality control
QC troubleshooting
Roche Diagnostics
B-34
June 2008
Instructions for Use Version 10.0
7 Calibration
Table of contents
Calibration
In this chapter
Chapter
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-35
7 Calibration
Table of contents
Roche Diagnostics
B-36
June 2008
Instructions for Use Version 10.0
7 Calibration
Calibration - general
Calibration - general
The cobas b 121 system employs a patented method that allows the simultaneous
calibration of PCO2, pH, Na+, K+, Ca2+ and Cl electrodes while using only two
calibration solutions (C1 calibration solution 1 and C2 calibration solution 2).
The chemical properties of the solutions and the concentration of their components
make the system insensitive to environmental influences during storage and use.
Figure B-24
The progress bar visible in the calibration progress indicator is divided into the
following sections:
Roche Diagnostics
Instructions for Use Version 10.0
The gray area indicates the maximum time how long the respective calibration
may last.
The yellow area in the bar indicates the estimated duration of the calibration. An
average value is used as the basis. This area does not change during the calibration.
The green marking above the bar indicates the minimum possible duration of the
respective calibration.
June 2008
B-37
7 Calibration
Calibration - general
Automatic calibrations
The following calibrations are automatically initiated and performed by the analyzer.
System calibration
Every 8, 12 or 24 hours (default) which includes the following:
o
The user can set a permanent start time for the system calibration. This enables
completion of calibration tasks while the instrument is not in use or when the workload in
the laboratory or station is smaller.
e See Reference Manual, chapter Setup!
Roche Diagnostics
B-38
June 2008
Instructions for Use Version 10.0
7 Calibration
Calibration - general
System calibration
Conductivity calibration
1P calibration
2P calibration incl. O2
2P O2 calibration
2P calibration without O2
Press:
h [More functions] > [System]
Figure B-25
Using the [line up] / [line down] buttons you may now select the desired calibration.
Start the selected calibration by pressing this button.
Conductivity calibration
Calibrate the conductivity system with the C1 and C2 calibration solutions in order to
determine the actual mixing ratio in combination with the mixing system.
1P calibration
This calibration is a 1 point calibration (incl. O2) of all parameters.
2P calibration incl. O2
This calibration is a 2 point calibration of all parameters.
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-39
7 Calibration
Calibration - general
2P O2 calibration
This calibration is a 2 point calibration for the PO2 electrode.
2P calibration without O2
This calibration is a 2 point calibration of all parameters except PO2.
Roche Diagnostics
B-40
June 2008
Instructions for Use Version 10.0
8 Operating modes
Table of contents
Operating modes
In this chapter
Chapter
June 2008
B-41
8 Operating modes
Table of contents
Roche Diagnostics
B-42
June 2008
Instructions for Use Version 10.0
8 Operating modes
Operating modes - general
Analyzer
Setup
Instrument settings
Database
Info
User interface
All information (results, error messages, warnings, etc.) is displayed on the screen.
The screen consists of a 5.7" colour LCD that is covered with a touch-sensitive film
("touch screen").
For example
C
D
Act. status
Status line
Current time
More functions
Info
Figure B-26
e For more detailed information refer to the respective chapters in these Instructions for Use
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-43
8 Operating modes
User interface
Parameter
Parameter Display on the "Ready" screen
Depending on the settings and the status of the instrument, the parameter buttons
may have the following appearance:
Parameter activated and ready
Parameter temporarily deactivated (but calibrated)
Parameter activated with QC warning
Parameter temporarily deactivated with QC warning
Parameter not ready (not calibrated)(a)
Parameter not ready (due to QC lock)
Parameter not ready due to remote lock
Parameter permanently deactivated (under "Setup")
(a) Pressing the parameter button produces a status report
Additional
o
Operator ID
Sample type
Blood type
Specimen ID
Puncture site
ALLEN test
A/F (adult/fetal)
Pat ID
Last name
First name
Middle initials
Date of birth
Gender
Pract. Pat ID
Height
Weight
Roche Diagnostics
B-44
Insurance code
Suffix
Title
Maiden name
Ethnic origin
Address
Phone no.
Diagnosis
Medication
Diet
Admission date
Admission time
Discharge date
Discharge time
Admission status
Location
Diagnose code
Relig. denom.
June 2008
Instructions for Use Version 10.0
8 Operating modes
User interface
Marital status
Isolation status
Patient language
Hospital institute
Dosage cat.
Remark
Physician
Remark
Acceptor
Date drawn
Time drawn
Place drawn
Danger code
Clinic info
Container
24 hr. urine
Ventilation mode
VT
MV
PIP
Ti
Te
SRATE
ARATE
PEEP
MAP
Flow rate
PO2
PCO2
pH
Na
K
Sodium concentration
Potassium concentration
Cl
Ca
Calculated parameters
Chloride concentration
2+
Calcium concentration
Hct
Hematocrit
tHb
SO2
Baro
Air pressure
H+
cHCO3
ctCO2(P)
ctCO2(B)
BE
BEact
BEecf
BB
Buffer bases
ctO2
pHst
Standard pH value
cHCO3 st
PAO2
RI
Respiratory index
nCa2+
June 2008
B-45
8 Operating modes
User interface
SO2(c)
AaDO2
a/AO2
AG
Anion gap
MCHC
Osm
Osmolality
Hct(c)
P/F Index
PaO2/FIO2 ratio
Qs/Qt
Qt
avDO2
OER
Calculation after the second measurement if the following conditions are fulfilled:
PAO2t
RIt
AaDO2
a/AO2
pHt
Input parameters
Roche Diagnostics
B-46
PCO2
PO2
H+t
P50
FIO2
tHb
Total hemoglobin
Temperatur
Patient temperature
June 2008
Instructions for Use Version 10.0
8 Operating modes
User interface
Buttons
Buttton
Description
button "Info"
More functions
confirm
cancel
activate / deactivate
start
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-47
8 Operating modes
User interface
Buttton
Description
data input
details
Roche Diagnostics
B-48
June 2008
Instructions for Use Version 10.0
8 Operating modes
Analyzer mode
Analyzer mode
The "Analyzer" operating mode contains parameter information (for example,
"Ready"), system settings, and the QC measurement. The highest level of this
operating mode is the "Ready" screen.
"Ready" screen
The Ready screen is the central starting point for all operations. The instrument is
usually in the "Ready" state.
Figure B-27
If a "mandatory input" field has been modified (in the "Setup" operating mode),
measurement can begin only after entering the data associated with this input field.
Every possible input value can be defined as a "mandatory input".
NOTE
In the following example, the Pat ID has been defined as a mandatory entry.
Figure B-28
Press this button and enter the Pat ID. The measurement can then be started.
If the measurement is equipped with password protection, the "Ready" screen is
covered by the password window but the parameter section remains visible
(parameter information).
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-49
8 Operating modes
Analyzer mode
Figure B-29
Press this button and enter the required password. The measurement can then be
started.
When a mandatory input or password is activated and the flap is opened without
completing the input, one of the following messages appears:
"Complete mandatory inputs" or "Close flap and enter password."
The system section can be reached directly and only from the "Ready" screen.
This occurs by pressing the button [More functions]. This button calls up a window
with which the following functions may be activated:
B
More functions
Roche Diagnostics
B-50
June 2008
Instructions for Use Version 10.0
8 Operating modes
Analyzer mode
System
The following main menus are available:
Figure B-31
Quick access
Using these functions, you can start the following actions or change the following
settings:
Figure B-32
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-51
8 Operating modes
Analyzer mode
QC measurement
This function helps start a quality control measurement.
e For more detailed information, see Chapter 6 Quality control!
Setup
Use this function to make the following settings:
Figure B-33
You can activate the desired setting directly by pressing the respective line on the screen.
e For an exact description of this operating mode refer to the Reference Manual, Chapter
Setup!
Roche Diagnostics
B-52
June 2008
Instructions for Use Version 10.0
8 Operating modes
Analyzer mode
Database
Use this function to retrieve the following data:
Figure B-34
You can activate the desired database directly by pressing the respective line on the
screen.
All
All data of the selected database are listed.
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-53
8 Operating modes
Analyzer mode
The completed line is marked, the cursor moves to the next line.
search
This function enables the search according to search criteria in ascending or descending order.
sort
Date / Time (Only for measurement, calibration, QC, and instrument data)
Up arrow:
the records are arranged in ascending order (oldest date is at the top)
Down arrow:
the records are arranged in descending order (youngest date is at the top).
Sort criteria:
Up arrow:
Down arrow:
Sort criteria:
Up arrow:
Down arrow:
Sort criteria:
Up arrow:
the records are arranged alphabetically in ascending order (for example, from top to
bottom - Z->A)
Down arrow:
QC statistics
Figure B-35
Roche Diagnostics
B-54
Export data
8 Operating modes
Analyzer mode
Delete data
A list of all available forms is displayed (userspecific and standard). This selection remains the
defined standard until a new form is selected.
e See the Reference Manual, chapter Setup
Printer cover
Figure B-36
A
B
Front side
Reverse side
Figure B-37
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-55
8 Operating modes
Analyzer mode
Figure B-38
chapter Database!
Patient data
Figure B-39
Roche Diagnostics
B-56
June 2008
Instructions for Use Version 10.0
8 Operating modes
Analyzer mode
Measuring data
Figure B-40
Calibration data
Figure B-41
Starting this function, the instrument displays an overview of the saved calibration
data.
Every line displays a short record of a calibration and contains the date, time, type of
calibration, as well as the parameter's condition after the calibration.
Press these buttons to scroll through all parameters.
Select the marked entry - the results screen of this calibration is then
displayed.
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-57
8 Operating modes
Analyzer mode
QC data
Figure B-42
Starting this function, the instrument displays an overview of the saved QC data.
This screen shows you, based on selected query criteria, all QC materials that were
measured up to this point, the level, lot numbers, and the date on which the QC files
were begun.
After you have selected and completed an entry, press the [Zoom] button to receive all
available information on the completed QC file. Every line shows the date, time,
operator ID (when available), and the corresponding status of the available
parameters.
Press these buttons to scroll through all parameters.
Select the marked entry - the results display of the selected QC data are
displayed.
Instrument data
Figure B-43
Roche Diagnostics
B-58
June 2008
Instructions for Use Version 10.0
8 Operating modes
Analyzer mode
Info
Use this function to retrieve the following information:
Figure B-44
Roche info
This displays the name, address, telephone number, and e-mail address that you
defined under:
h [Setup] > [Instrument] > [Roche info]
e See Reference Manual, chapter Setup!
Version numbers
This lists the software version.
Fill levels
Use this function to check the fill level.
e See Chapter 9 Maintenance, section Check fill levels on page C-7!
Help
Use this function to retrieve online-help information.
Figure B-45
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-59
8 Operating modes
Analyzer mode
Sensor status
e See Chapter 10 Troubleshooting, section Sensor status on page D-16!
Counter overview
Figure B-46
Roche Diagnostics
B-60
Sample counter:
Number of measurements
QC counter:
Number of QC measurements
Cleaning counter:
June 2008
Instructions for Use Version 10.0
8 Operating modes
Software overview
Software overview
Figure B-47
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
B-61
8 Operating modes
Software overview
Part 1
Part 2
Figure B-48
Roche Diagnostics
B-62
June 2008
Instructions for Use Version 10.0
Maintenance
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-3
June 2008
9 Maintenance
Table of contents
Maintenance
In this chapter, all maintenance work is described that is necessary for trouble-free
operation of the instrument.
In this chapter
Chapter
June 2008
C-3
9 Maintenance
Table of contents
Roche Diagnostics
C-4
June 2008
Instructions for Use Version 10.0
9 Maintenance
Maintenance - general
Maintenance - general
After use, components of the instrument, including tubing, waste container, fill port, etc.,
contain biological fluids and represent a possible infectious risk.
Handle these components with care and according to regulations surrounding potentially
infectious material.
Avoid contact with skin. Always wear gloves! Danger of infection!
Decontamination
General information
The purpose of this procedure is to minimize the risk of infections when replacing
items that were in contact with blood.
Perform these decontamination procedures regularly.
Roche recommends following a decontamination procedure in addition to
regulations specific to the laboratory.
Use only liquid disinfectant such as protein remover (Roche deproteinizer) or an
alcohol-based (about 70%) surface disinfectant.
Do not spray disinfectant directly onto the instrument because this could cause
malfunctions in the electronics.
Do not use any type of bleaching agent. Exception: Roche Deproteinizer
Do not attempt to decontaminate any part of the instrument before shutting it down and
unplugging it from the power source.
ATTENTION
Before plugging the instrument back in and turning it on, always wait
15 minutes to allow the disinfectant to evaporate Danger of fire and explosion!
For safety reasons, only authorized technical service personnel may decontaminate the
power pack!
Touch screen
Tubing paths
on page C-17!
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
C-5
9 Maintenance
Decontamination
Touch screen
e See section Cleaning the screen on page C-28!
Tubing paths
e See section Decontaminate the tubing paths on page C-26!
Recommended decontaminant
Do not use any type of bleaching agent. Exception: Roche Deproteinizer!
Surfaces
Avoid contact with skin. Always wear gloves! Danger of infection!
Protein remover
(Roche Deproteinizer)
Potential dangers
Due to the alkaline and oxidizing character of this preparation, we cannot rule out
local irritation to the skin, eyes, and mucous membranes.
Roche Diagnostics
C-6
June 2008
Instructions for Use Version 10.0
9 Maintenance
Daily
Daily
Check fill levels
Check daily the fill levels of the solutions (C1 calibration solution 1, C2 calibration
solution 2) and the waste container (W waste container).
Perform visual checks or select:
h [Info] > [Fill levels]
Figure C-1
Exchange empty bottles, bottles whose usage date has expired, and full waste water
bottles.
e See section Exchanging the solutions on page C-10!
The printer paper is heat sensitive on one side only. Please make sure that you insert the
paper roll correctly.
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
C-7
9 Maintenance
Weekly
Weekly
Clean needle and fill port
Handle these parts with care danger of injury!
Always wear gloves! Danger of infection!
Figure C-2
2 Open the flap to the labelled capillary position (completely open position).
3 Use a moist (Deproteinizer for example) paper towel or gauze to clean the fill port
and remove any deposits on the needle (see Figure C-3/1+2).
4 Close the flap to the syringe position (half-opened position).
5 Clean the tip of the needle (see Figure C-3/3).
(1)
(2)
(3)
Figure C-3
Roche Diagnostics
C-8
June 2008
Instructions for Use Version 10.0
9 Maintenance
Semi annual
Semi annual
Replacement of the peristaltic pump tubes
While in the "Analyzer" operating mode, press:
h [More functions] > [System] > [Tools] > [Tubing exchange] > [Replace PP
tubing]
Follow the instructions on the screen and proceed as follows:
1 Open the top cover.
2 Open the peristaltic pump's clear plastic cover (tension lever).
3 Push the linear clamp (white plastic part) upwards (seeFigure C-4/1).
4 Remove the complete tubing set (tubing holder and tubing) (see Figure C-4/2).
B
C
(1)
A
Tension lever
Pump head
Linear clamp
(2)
Figure C-4
5 Check if the five rollers on the peristaltic pump rotate easily. If defective, please
contact Roche Technical Support.
6 Place the new tube around the rolling wheel.
7 Close the acrylic glass cover (tension lever). This presses the tubing holder into the
sealer.
8 Close the top cover.
The tubes may drip after disconnection. Remove excess fluids with a clean, absorbent
cloth.
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
C-9
9 Maintenance
Use of calibration solutions that were not produced by Roche may lead to an
invalidation of the instrument's guarantee.
All reagents are suitable only for diagnostic in-vitro use!
After replacing a solution, a quality control measurement must be performed with all
3 levels (low, normal, high). Make sure that the results agree with the target values.
e See Chapter 6 Quality control!
C3 fluid pack
Depending on the rate of measurement and/or the on-board stability, this fluid pack
should be exchanged every 4-6 weeks. The screen displays the appropriate
information.
Reuse of this bottle can lead to errors during calibration! Insert another bottle that has not
yet expired.
Roche Diagnostics
C-10
Cleaning solution
Reference solution
June 2008
Instructions for Use Version 10.0
9 Maintenance
Sample-dependent maintenance procedures
Implementation:
Figure C-5
2 Open the docking mechanism and remove the bottles that are to be exchanged
(see Figure C-6).
Figure C-6
5 The instrument recognizes the correct solution and checks the expiration date.
The respective bottle blinks on the screen.
If the bottle has passed the expiration date, the screen displays a warning.
Reuse of this bottle can lead to errors during calibration!
Insert another bottle that has not yet expired.
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
C-11
9 Maintenance
If inserting a bottle that has already been partially used, press this button and
enter the fill level.
Use the markings on the label to estimate the fill level of a bottle that is partially full.
8 A numerical keyboard appears on the display. Enter the correct fill level in % and
confirm with the [OK] button.
9 Close the bottle compartment cover. The solutions are automatically aspirated
upwards (detection in the flap).
Waste material
Remove the waste container (W waste container)
1 Open the bottle compartment cover. The bottle exchange image appears on the
display (see Figure C-7).
Figure C-7
Roche Diagnostics
C-12
June 2008
Instructions for Use Version 10.0
9 Maintenance
Sample-dependent maintenance procedures
Figure C-8
2 Open the screw cap by pressing the grips together and rotating them counter
clockwise (Figure C-9).
Figure C-9
3 When removing the screw cap, make sure that the green element inside the bottle
is not moved or removed (Figure C-10).
Figure C-10
Empty the waste material and decontaminate the container according to local regulations
(hazardous waste material!).
Flush the waste container cap with plenty of water.
4 Screw the cap back onto the bottle. The cap must be screwed shut until completely
closed!
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
C-13
9 Maintenance
Remove the sticker from the empty bottle of C1 solution (see Figure C-11).
This sticker may not be reused discard immediately.
position.
Roche Diagnostics
C-14
June 2008
Instructions for Use Version 10.0
9 Maintenance
Sample-dependent maintenance procedures
3 Carefully remove the fill port holder from the needle (see Figure C-12/2).
(1)
A
(2)
Axis
Figure C-12
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
C-15
9 Maintenance
4 Place the new fill port holder (with integrated fill port) over the needle. It is easier
if you turn the holder around so that the bottom side is showing on top (see
Figure C-13).
Figure C-13
Figure C-14
Roche Diagnostics
C-16
June 2008
Instructions for Use Version 10.0
9 Maintenance
Unscheduled
Unscheduled
Cleaning the bottle compartment
1 Open the bottle compartment cover. The bottle exchange image appears on the
display.
e See Figure C-7 on page C-12!
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
C-17
9 Maintenance
Unscheduled
Drip tray
Wash plate
Figure C-15
1 Open the flap to the labelled capillary position (completely open position).
2 Remove the sample drop cup and clean it according to local regulations (see
Figure C-16/1).
3 Open the bottle compartment cover.
4 Remove the red tube connector from the wash plate (see Figure C-16/2)
5 Push the wash plate down (to unlock) and pull it out (see Figure C-16/3). The
wash plate is locked when the plug is sticking out and cannot be removed.
(1)
(2)
(3)
Figure C-16
6 It is best to clean the wash plate under running water, then, dry well.
7 Push the wash plate back in completely and attach the tube plug.
8 Push the drip tray in.
9 Close the flap.
10 Close the bottle compartment cover.
Roche Diagnostics
C-18
June 2008
Instructions for Use Version 10.0
9 Maintenance
Unscheduled
printer lever
Figure C-17
Figure C-18
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
C-19
9 Maintenance
Unscheduled
Figure C-19
A
A
locking lever
MCon
Figure C-20
Roche Diagnostics
C-20
June 2008
Instructions for Use Version 10.0
9 Maintenance
Unscheduled
WARNING
6 Prevent the formation of gas bubbles in the inner electrolyte of the electrodes
(see Figure C-21/1).
If necessary, remove air bubbles by holding the electrode vertically and by tapping
lightly with your fingernail against the electrode body (see Figure C-21/2).
(1)
(2)
Figure C-21
Figure C-22
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
C-21
9 Maintenance
Unscheduled
Roche Diagnostics
C-22
After inserting a new pH electrode, the pH parameter may change to the non-calibrated
status during the first hours. The pH electrode is not defective. To remedy the situation,
start a "Calibration for "Ready".
June 2008
Instructions for Use Version 10.0
9 Maintenance
Unscheduled
Figure C-23
A
A
locking lever
reference electrode
Figure C-24
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
C-23
9 Maintenance
Unscheduled
A
A
white plug
Figure C-25
Figure C-26
If the tube does not lie precisely inside the guides, it may become pinched and thereby
prevent aspiration of the reference solution, resulting in calibration and measurement
errors.
7 Affix the white plug at the end of the reference electrode tube (see Figure C-25).
Push the plug in completely.
June 2008
Instructions for Use Version 10.0
9 Maintenance
Unscheduled
Removing obstructions
For removing obstructions in the sample path, proceed as follows:
First determine whether the electrodes have any obstructions and replace them, if
necessary.
Figure C-27
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
C-25
9 Maintenance
Unscheduled
Roche Diagnostics
C-26
June 2008
Instructions for Use Version 10.0
9 Maintenance
Unscheduled
Figure C-28
7 Empty the container according to local regulations and fill the shutdown kit
halfway with distilled water.
8 Replace the shutdown kit to the position of C3.
9 Connect the tubes of the set with the connectors from C1 and C2 (Figure C-28).
10 Press this button to begin cleaning the tubes.
11 Remove the shutdown kit.
Remove the remaining fluid from the tubes by briefly holding the tubes vertically, allowing
the fluid to run back into the container.
page C-26!
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
C-27
9 Maintenance
Unscheduled
After 30 seconds, the display changes back into its active condition.
Surfaces
Do not attempt to decontaminate any part of the instrument before shutting it down and
unplugging it from the power source.
ATTENTION
Before plugging the instrument back in and turning it on, always wait 15 minutes to allow
the disinfectant to evaporate Danger of fire and explosion!
For safety reasons, only authorized technical support personnel may decontaminate the
power pack!
Regularly decontaminate all outside surfaces of the instrument, including all covers
(for example, printer cover, bottle compartment cover, top cover). For these tasks, use
the decontaminant in accordance with local regulations.
Very dirty surfaces should first be cleaned with swabs or paper towel of gauze that
have been soaked in distilled water. Spray all removable covers (top cover, bottle
compartment cover) with surface disinfectant and then wipe them down with swabs
or paper towel of gauze. Allow some time for solutions to soak for proper cleaning
effects.
Never spray parts that cannot be removed or that are inside the instrument!
ATTENTION
Roche Diagnostics
C-28
June 2008
Instructions for Use Version 10.0
9 Maintenance
Unscheduled
Figure C-29
By pressing the [Test] button, you can check if the entire (black) area is active as a
touch-sensitive surface (see Figure C-30).
Figure C-30
By pressing the [Calibrate] button, you can use a pencil or other pointed object (but
which is not too hard, to avoid scratching the surface) to touch the white points in the
upper left and lower right corners. After release, the instrument will accept the exact
position.
From this time on, the instrument will use the touched points to calculate the offset
between the displayed pixels and the touch screen. After a point has been accepted,
the arrow disappears. The point itself remains visible and active (pressing the position
again re-establishes the point) (see Figure C-31).
Figure C-31
After leaving the window, the new correction values take effect.
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
C-29
9 Maintenance
Annual maintenance
Decontamination/cleaning :
Screen
Surfaces
Tube paths
Pump tube
Needle seal
Sample port
Pump head
Test/check:
Roche Diagnostics
C-30
AutoQC needle
Carry out quality control measurement with all 3 levels (low, normal, high)
June 2008
Instructions for Use Version 10.0
9 Maintenance
Maintenance overview
Maintenance overview
While in the "Analyzer" operating mode, press:
h [More functions] > [System] > [Tools] > [Maintenance]
The overview shows a list of all maintenance entries.
Figure C-32
The following warnings are default entries and can neither be deactivated nor
modified:
o
Annual service
Replace PP tubing
Decontaminate screen
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
C-31
9 Maintenance
Maintenance scheduler
Maintenance scheduler
Additional services can be added to the list in the menu.
h [Setup] > [Times & intervals] > [Maintenance scheduler]
starting with the top level of the analyzer mode.
e For the exact description, see the Reference Manual, chapter Setup!
Roche Diagnostics
C-32
June 2008
Instructions for Use Version 10.0
Troubleshooting
10
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-3
June 2008
10 Troubleshooting
Table of contents
Troubleshooting
In this chapter, all fault messages, their causes and remedies are described. These are
also displayed directly on the instrument screen. All messages are arranged according
to info number.
In this chapter
Chapter
10
June 2008
D-3
10 Troubleshooting
Table of contents
Roche Diagnostics
D-4
June 2008
Instructions for Use Version 10.0
10 Troubleshooting
Troubleshooting - general
Troubleshooting - general
fter use, components of the cobas b 121 system, including tubing, waste container, fill
port, etc., contain biological fluids and represent therefore a possible infectious risk.
Handle these components with care and according to regulations surrounding potentially
infectious materials.
Suitable protective equipment, like laboratory clothing, protective gloves, protective
goggles and if necessary mouth protectors, must be worn to prevent direct contact with
biological working materials. In addition, a face mask is required if there is a risk of
splashes. Suitable disinfection and sterilisation procedures must be applied.
A diagnostics function is available to facilitate fault removal. Pressing this button lists
suggestions for the fault removal.
e An exact description of this function can be found in the Reference manual in chapter
System stops
For example:
Figure D-1
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
D-5
10 Troubleshooting
System stops
General information
During situations when the proper function of the analyzer is not possible, System
stops are displayed on the screen. The purpose is to display the errors, to remove the
cause of the System stops and to bring the analyzer back to "Ready".
A stop of all running actions will occur if danger for the user (e.g. by an overflow of
the waste while opening the docking mechanism of the waste container) or for the
analyzer exists or the proper operation of the analyzer is not possible because of
technical problems.
If proper operation is not possible, but the complete control of the functions of the
analyzer is available (e.g. temperature alarm, measuring chamber cover open, flap
open, fill level alarm) all running actions will be finished and Mix 1 will be aspirated
into the measuring chamber; the resulting measuring data will be marked resp.
discarded.
Unrecoverable System stops remain on the screen and fulfill an emergency program if
possible, so that the analyzer stays operating (wetting of the electrodes / keeping the
tubing free).
It is possible to access the menu "System" if no automatic actions are currently
performed. When changing to the menu "System" during a "System stop", the "System
stop" will not be terminated, but reactivated when the menu "System" is closed again.
If e.g. there is a change from a "Temperature system stop" to the menu "System" and
the menu "System" is closed again, the analyzer will display the system stop
"Temperature" again. In the case a system calibration has been started in the menu
"System", it can not be conducted immediately because the "system stop" has to be
removed first.
After all "system stops" have been removed, "system stop consequence actions" will be
executed. There is a common list of consequence actions, which impedes double
actions. This means that if e.g. "washing" was activated several times, it is then
performed only once.
Possible consequence actions are:
Roche Diagnostics
D-6
Warm-up
Prepare solution C1
Prepare solution C2
Wash
Wash AutoQC
Conductivity calibration
System calibration
Aspirate Mix1
June 2008
Instructions for Use Version 10.0
10 Troubleshooting
List of system stops
Remedy
h [More functions] > [System] > [Test] > [Control sensors] > [Monitoring sensors]
Consequence actions
Check cable
Wash
Remedy
h [More functions] > [System] > [Test] > [Control sensors] > [Monitoring sensors]
Consequence actions
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
D-7
10 Troubleshooting
Remedy
during a measurement
during a calibration
Close flap
h [More functions] > [System] > [Test] > [Control sensors] > [Monitoring sensors]
o
Wash
Cause
Remedy
Wash
Cause
Remedy
Consequence actions
Consequence actions
Remedy
Roche Diagnostics
D-8
June 2008
Instructions for Use Version 10.0
10 Troubleshooting
List of system stops
Remedy
System calibration
Consequence actions
Remedy
h [More functions] > [System] > [Test] > [Control sensors] > [Monitoring Sensors]
o
Cause
Remedy
Aspirate Mix1
The docking mechanism of Fluid Pack C3 has been opened (micro switch
activated)
Consequence actions
Remedy
h [More functions] > [System] > [Test] > [Control sensors] > [Monitoring sensors]
Consequence actions
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
D-9
10 Troubleshooting
Remedy
Consequence actions
The solutions C1, C2 and/or C3 are empty (below alarm level) or are set to
"empty"
Remedy
Consequence actions
Remedy
h [More functions] > [System] > [Test] > [Control sensors] > [Monitoring sensors]
Consequence actions
Roche Diagnostics
D-10
June 2008
Instructions for Use Version 10.0
10 Troubleshooting
List of system stops
Remedy
Check PP tube
Check under:
h [More functions] > [Test] > [Valves & Aggregates] > [Peristaltic pump]
if values inside the following limits are displayed:
Consequence actions
If the displayed values are outside of the limits, perform a FMS volume
determination and correct the FMS volume value.
Aspirate Mix1
Conductivity calibration
The sample detection with sample sensors (SS1 and SS2) failed
Wash
Cause
Remedy
Cause
Remedy
Consequence actions
Remedy
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
D-11
10 Troubleshooting
The actually measured waste container fill level differs by more than 4 cm from
the calculated/set value
Remedy
The waste container fill level must be set roughly (+/- 4 cm) corresponding to the
actual fill level in the waste container
Wash
Cause
The internal flash memory has less than 8 KB space left for saving additional data
Remedy
Delete data records (database entries, protocols, log data) in order to free up
additional memory
Consequence actions
The PCMCIA card has less than 8 KB space left for saving additional data
Remedy
Delete data records from the PCMCIA card in order to free up additional memory
capacity.
In order to actually free up additional memory, the functions "Delete data" and "Optimize
database" have to be activated in this order
ATTENTION
Insert the PCMCIA card in a PC with a suitable slot, copy or import the data, then
delete the data from the card or format the card
h [More functions] > [System] > [Test] > [PC Components] > [PCMCIA card] >
[More functions]
e See Reference Manual, chapter System section Test > PC Components > PCMCIA card!
Roche Diagnostics
D-12
June 2008
Instructions for Use Version 10.0
10 Troubleshooting
List of system stops
The data access onto objects in the analyzer area failed, the correct operation of
the instrument can not be guaranteed
Remedy
Cause
Remedy
Cause
Remedy
Cause
Remedy
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
D-13
10 Troubleshooting
Electrode status
Electrode status
If a parameter is not ready (not calibrated), the symbol will be crossed out with grey
and red lines (see below).
Figure D-2
After you press the parameter button, a status report will give you information about
why the parameter could not be calibrated (see below).
e The exact description of the short texts can be found in the Reference Manual, chapter
Troubleshooting.
Figure D-3
Pressing this button lists suggestions for the fault removal (see below).
Figure D-4
Roche Diagnostics
D-14
June 2008
Instructions for Use Version 10.0
10 Troubleshooting
Electrode status
Figure D-5
e An exact description of this function can be found in the Reference Manual, chapter
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
D-15
10 Troubleshooting
Electrode status
Sensor status
For additional information about the sensor status, press:
h [Info] > [Sensor status]
Press [line up] / [line down] and select a sensor from the list on the left.
The following information about the selected sensor will be displayed:
Figure D-6
Roche Diagnostics
D-16
June 2008
Instructions for Use Version 10.0
Appendix
11
June 2008
11 List of consumables
Table of contents
List of consumables
In this chapter, all nessary consumables and order numbers are listed.
In this chapter
Chapter
11
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
E-3
11 List of consumables
Table of contents
Roche Diagnostics
E-4
June 2008
Instructions for Use Version 10.0
11 List of consumables
Order information
Order information
To measure the respective parameter following products are required:
Electrodes
Parameter
Cl-
Na+
K+
Ca2+
PCO2
PO2
pH
tHb
SO2
Chloride Electrode
03111571180 (BP1729)
Sodium Electrode
03111598180 (BP1730)
Potassium Electrode
03111628180 (BP1731)
CalciumElectrode
03111644180 (BP1732)
PCO2 Electrode
03111679180 (BP1733)
PO 2 Electrode
03111695180 (BP1734)
pH Electrode
03111717180 (BP1735)
Reference Electrode
03111873180 (BP2081)
MCon
03144062001 (BP2613)
SCon
03260909184 (BP2608)
Tabelle E-1
(a) Dummy or electrode has to be used for proper filling of the measuring chamber
Roche Diagnostics
Instructions for Use Version 10.0
Has to be installed
Dummy or electrode has to be used for proper filling of the measuring chamber
June 2008
E-5
11 List of consumables
Order information
Solutions
Paramenter
-
2+
Cl
Na
PCO2
PO2
pH
tHb
SO2
C3 Fluid Pack
03144038001 (BP2655)
C1 Calibration Solution
03144046001 (BP2661)
C1 Calibration Solution
03144020001 (BP2662)
W Waste
03144054001 (BP2663)
Ca
Tabelle E-2
Has to be installed
QC material
Parameter
-
Na
PCO2
PO2
pH
tHb
SO2
COMBITROLTS+. Level 1
03321258001 (BP9109)
COMBITROLTS+. Level 3
03321274001 (BP9111)
Cl
2+
Ca
Tabelle E-3
(a) Only for USA available!
Roche Diagnostics
E-6
Can be used
June 2008
Instructions for Use Version 10.0
11 List of consumables
Order information
Accessories
Parameter
-
Na
PCO2
PO2
pH
tHb
SO2
Deproteinizer
03110435180 (BP0521)
Ampoule Adapter
03066762001 (BP1938)
Clot Catcher
03112012180 (BP2243)
Cl-Cleaning Kit
03112098035 (BP2276)
Printer Paper
03113353180 (HP0070)
Roche MICROSAMPLER.
with accessories
03113663160 (US0600)(a)
Roche MICROSAMPLER.
w/o. accessories
03113671160 (US0601)(b)
Cl
2+
Ca
Tabelle E-4
(a) Only for USA available
(b) Only for USA available
Roche Diagnostics
Instructions for Use Version 10.0
Can be used
Do not use!
June 2008
E-7
11 List of consumables
Order information
Roche Diagnostics
E-8
June 2008
Instructions for Use Version 10.0
Index
12
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-3
June 2008
Index
Index
Numerics
1P calibration, B-39
1P calibration (1P cal), B-38
2P calibration (2P cal), B-38
2P calibration incl. O2, B-39
2P calibration without O2, B-40
2P O2 calibration, B-40
Bottles
Insertion of bottles, A-30
Remove bottles, A-36
Brands, 1
Buttons, B-47
Important buttons on the screen, A-10
Bypass
Cleaning, C-27
Abbreviations, 8
Accessories, A-22
List of consumables, E-7
Adjusting the screen, C-28
Analyzer mode, B-49
"Ready" screen, B-49
Database, B-53
Calibration data, B-57
Data export, B-55
Description of the buttons and their function, B-54
Instrument data, B-58
Measuring data, B-57
More functions of the database, B-54
Patient data, B-56
QC data, B-58
Info, B-59
QC measurement, B-52
Quick access, B-51
Setup, B-52
System, B-51
Annual maintenance, C-30
Application area, A-10
Automatic calibrations, B-38
1P calibration (1P cal), B-38
2P calibration (2P cal), B-38
Recalibration, B-38
System calibration, B-38
AutoQC
AutoQC measurement, B-27
Inserting AutoQC mats, B-21
Material assignment, B-22
Setting QC times, B-23
C1 calibration solution 1
Exchange, C-10
C2 calibration solution 2
Exchange, C-10
C3 fluid pack
Exchange, C-10
Calculated parameters, B-45
Calculated parameters at the patients
temperature, B-46
Calibration, B-35
Automatic calibrations, B-38
1P calibration (1P cal), B-38
2P calibration (2P cal), B-38
Recalibration, B-38
System calibration, B-38
Calibration progress indicator, B-37
General, B-37
User activated calibrations, B-39
1P calibration, B-39
2P calibration incl. O2, B-39
2P calibration without O2, B-40
2P O2 calibration, B-40
Calibration for "Ready", B-39
Conductivity calibration, B-39
System calibration, B-39
Calibration data, B-57
Calibration for "Ready", B-39
Calibration procedure, A-12
Calibrations, A-50
Capillary mode, B-10
Capillary tubes, B-6
Check fill levels, C-7
Check printer paper, C-7
Checking the barometer value, A-26
Classification (according IEC/ISO), A-53
Clean needle and fill port, C-8
Cleaning the bottle compartment, C-17
Cleaning the bypass, C-27
Cleaning the drip tray and wash plate, C-17
Cleaning the measurement chamber, C-25
B
Barcode scanner, A-18
Specifications, A-54
Bottle compartment, A-16
Cleaning, C-17
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
F-3
Index
F
D
Data export
Database, B-55
Data input
Measuring pocedure, B-12
Database, B-53
Calibration data, B-57
Data export, B-55
Description of the buttons and their function, B-54
Instrument data, B-58
Measuring data, B-57
More functions of the database, B-54
Patient data, B-56
QC data, B-58
Decontamination, A-13, C-5
Decontaminate the surfaces, C-28
Decontaminate the tubing paths, C-26
Recommended decontaminant, C-6
Sample port module, C-5
Surfaces of the instrument, C-6
Touch screen, C-6
Tubing paths, C-6
Default and input values, A-49
Dimensions, A-53
Disposal of waste water, bottles, electrodes, and
the instrument, A-13
Document information, 1
Drip tray
Cleaning, C-17
E
Edition, 2
Edition notice, 1
Electrodes
Handling electrodes, A-14
Insert electrodes, A-32
Insertion of the reference electrode, A-33
List of consumables, E-5
Roche Diagnostics
F-4
G
General description, A-7
General notes, A-10
H
Help, B-59
How to Use This Manual, 5
I
Important information, A-5
Info, B-59
Counter overview, B-60
Fill levels, B-59
Help, B-59
Roche info, B-59
Sensor status, B-60
Version numbers, B-59
Input parameters, B-46
Measurement information, B-45
Patient information, B-44
Installation, A-21, A-22
Accessories, A-22
Attach power cord and barcode scanner, A-22
Attach pump tube, A-28
AutoQC module, A-27
Begin installation routines, A-35
June 2008
Instructions for Use Version 10.0
Index
Maintenance, C-3
Additional maintenance procedures, C-30
Annual maintenance, C-30
Annual maintenance, C-30
Daily, C-7
Check fill levels, C-7
Needle
Clean needle, C-8
Remove needle and fill port, A-40
Roche Diagnostics
Instructions for Use Version 10.0
June 2008
F-5
Index
O
Obstructions
Removing obstructions, C-25
Operating instructions, A-10
Operating modes, B-41
Analyzer mode, B-49
Database, B-53
Info, B-59
QC measurement, B-52
Quick access, B-51
Setup, B-52
System, B-51
General, B-43
Software overview, B-61
User Interface, B-43
Operating safety information, A-6
Operation
Calibration, B-35
Measurement, B-3
Quality control, B-15
Operators Manual
version, 1
Order information, E-5
Accessories, E-7
Electrodes, E-5
QC material, E-6
Solutions, E-6
P
Parameter, B-44
Additional calculated parameters, B-46
Calculated parameters, B-45
Calculated parameters at the patients
temperature, B-46
Input parameters, B-46
Input parameters - measurement information, B-45
Input parameters - patient information, B-44
Measured parameters, B-45
Parameter appearance on the "Ready" screen, B-44
Patient data, B-56
Performance data, A-47
Measured parameters, A-47
Plasma
Sample handling, B-7
Power supply, A-16
Preanalytics, B-5
Sample collection, B-5
Sample collection containers, B-6
Sample handling, B-7
Preface, 5
Printer, A-15
Check printer paper, C-7
Replacing printer paper, C-19
Roche Diagnostics
F-6
Specifications, A-53
Product data, A-53
Classification (according IEC/ISO), A-53
Dimensions, A-53
Electrical data, A-53
Weight, A-53
Pump, A-16
Remove the pump tube, A-39
Replacement of the peristaltic pump tubes, C-9
Q
QC concept
General, B-17
QC consequences, B-30
QC data, B-58
QC error, B-31
QC material
List of consumables, E-6
Material setup, B-20
Inserting AutoQC mats, B-21
Material assignment - AutoQC materials, B-22
Setting QC times, B-23
Temperature / humidity / stability, A-52
QC measurement, B-26, B-52
AutoQC measurement, B-27
Manual QC measurement, B-26
QC troubleshooting, B-32
QC unlock, B-31
QC error, B-31
QC warning, B-31
QC warning, B-31
Quality control, B-15
General, B-17
General QC concept, B-17
Important information concerning the analysis of
QC measurement results, B-19
Material setup, B-20
Inserting AutoQC mats, B-21
Material assignment - AutoQC materials, B-22
Setting QC times, B-23
Multirules, B-28
QC consequences, B-30
QC measurement, B-26
AutoQC measurement, B-27
Manual QC measurement, B-26
QC troubleshooting, B-32
QC unlock, B-31
QC error, B-31
QC warning, B-31
Quick access, B-51
June 2008
Instructions for Use Version 10.0
R
Ready screen, B-49
Recalibration, B-38
Reference electrode
Replacement, C-23
Remove needle and fill port, A-40
Remove the electrodes, A-39
Remove the printer paper, A-40
Remove the waste container (W waste container), C-12
Removing obstructions, C-25
Replacement of the electrodes and the MCon, C-20
Replacement of the peristaltic pump tubes, C-9
Replacement of the reference electrode, C-23
Replacing printer paper, C-19
Results display, B-13
Reverse side, A-17
Interface, A-17
Warning and identification labels, A-18
Revision history, 1
Roche info, B-59
Roche MICROSAMPLER, B-6
S
Safety information, A-3
Important information, A-5
Safety instructions for specific dangers, A-13
Decontamination, A-13
Disposal of waste water, bottles, electrodes, and
the instrument, A-13
Handling electrodes, A-14
Handling samples, A-13
Handling solutions, A-13
Sample collection, B-5
Acceptable anticoagulants, B-5
Especially for tHb, SO2 and Hct measurement, B-5
Sample acquisition, B-5
Sample collection containers, B-6
Capillary tubes, B-6
Clot catcher, B-6
Roche MICROSAMPLER, B-6
Syringes, B-6
Sample handling, B-7
Plasma, B-7
Serum, B-7
Whole blood, B-7
Sample port module
Decontamination, C-5
Sample throughput, A-49
Sample types, A-50
Sample volume, A-49
Samples
Handling samples, A-13
Scheduler
Roche Diagnostics
Instructions for Use Version 10.0
Index
Index
W
Warning and identification labels, A-18
Wash plate
Cleaning, C-17
Waste material, C-12
Empty the waste container, C-13
Installing the waste container, C-14
Remove the waste container (W waste container), C-12
Using the empty C1 calibration solution 1 bottle
as W waste container, C-14
Whole blood
Sample handling, B-7
T
Temperature / humidity / stability, A-51
Electrodes, A-51
Instrument, A-51
QC material, A-52
Solutions, A-52
Touch screen
Adjusting the screen, C-28
Decontamination, C-6
Troubleshooting, D-3
Electrode status, D-14
Sensor status, D-16
General, D-5
QC troubleshooting, B-32
System stops, D-5
List of system stops, D-7
Tubing paths
Decontamination, C-6, C-26
U
User activated calibrations, B-39
1P calibration, B-39
2P calibration incl. O2, B-39
2P calibration without O2, B-40
2P O2 calibration, B-40
Calibration for "Ready", B-39
Conductivity calibration, B-39
System calibration, B-39
User interface, B-43
Buttons, B-47
Parameter, B-44
V
Version numbers, B-59
Roche Diagnostics
F-8
June 2008
Instructions for Use Version 10.0