You are on page 1of 6

PANEL OVERVIEW Trigger

72-6802 User Manual Controlsthetriggermode,triggerlevel,trigger couplingsource,triggerslope,andalternate triggeringmode.Acceptstheexternaltriggerinput. AcceptstheCH1andCH2inputsignalsandground wire.Controlstheinputsignalcouplingmode.

PANEL OVERVIEW
Front Panel Overview

Controls Input

Terminals

Display Controls

1 CAL Output Generatestheprobecompensationsignal;2Vpp, 1kHz,positivesquarewave.Forprobe compensationdetails,seepage26. 2 INTEN Knob Controlsthebrightnessofalightspotortracein thedisplay. Detaileddescriptionsofeachblockstartfromthenextpage. Main Display Showsthewaveformsofinputsignals. Display Controls LCD Display Horizontal Controls Vertical Controls Controlspoweron/off,displayconfiguration,and theprobecompensationoutput. Showstheverticalscale,horizontalscale,XY displaymode,andwaveformfrequencies. Controlsthehorizontalscale,horizontalposition, sweeplength,andx10magnification. Controlstheverticalscale,verticalposition,display mode,CH2inversion,andalternatedisplaymode. 3 FOCUS Knob 4 TRACE Point Controlsthefocus(sharpness)ofthewaveformsin thedisplay.

Controlsthealignmentofthehorizontaltracein ROTATION parallelwithgraticulelines. Automaticallyadjuststhehorizontalscaletoan appropriaterangeaccordingtotheinputsignal.

5 AUTO Key

TIMEBASE

6 POWER Switch

Turnsonorofftheoscilloscopesmainpower. Whenthepowerison,theLEDlights.

13

14

PANEL OVERVIEW

72-6802 User Manual

LCD Display

Horizontal Controls

1 CH1

ShowstheCH1verticalscale.

Vertical Scale

1 Horizontal Controlsthehorizontalpositionoftracesandlight POSITION spots. Knob 2 10 MAG Switch 3 SWP VAR Magnifiesthehorizontalscalebyafactorof10.For horizontalmagnificationdetails,seepage34. Adjuststhehorizontalscale. Attheminimumposition,thehorizontalscale becomes2.5timeswiderthantheoriginalvalue selectedbytheTIME/DIVknob.Forexample,ifthe originalscaleis1mS/DIV,theadjustedscale becomes2.5mS/DIV. Atthemaximum(CAL)position,thereisno changeinthehorizontalscale. 4 TIME/DIV Knob Controlsthehorizontalscalefrom0.2s/divto0.5 s/divin20steps. WhenXYisselected,theoscilloscopeworksin theXYmode.ForXYmodedetails,seepage33.

2 Horizontal Showsthehorizontalscale. Scale 3 X-Y Mode 4 Signal 5 CH2 Vertical Scale Whenturnedon,indicatesthattheXYmodeis activated.ForXYmodedetails,seepage33. Showsthewaveformfrequency.Forfrequency measurementdetails,seepage32. ShowstheCH2verticalscale.

Knob

Frequency

15

16

PANEL OVERVIEW

72-6802 User Manual DUAL ADD TheCH1andCH2signalsaredisplayed simultaneously. TheCH1andCH2signalsareaddedor subtracted,andthentheresultis displayed.ForCH1/CH2 addition/subtractiondetails,seepage32.

Vertical Controls

5 VAR Knob

Adjuststheverticalscale. Attheminimumposition,theverticalscale becomes2.5timeswiderthantheoriginalvalue selectedbytheVOLTS/DIVknob.Forexample,if theoriginalscaleis1mV/DIV,theadjustedscale becomes2.5mV/DIV.

1 VOLTS/DIV ControlstheCH1/CH2verticalscalefrom 1mV/DIVto5V/DIVin12steps. Knob 2 Vertical Knob POSITION Controlstheverticalpositionoftracesandlight spotsforCH1/CH2. 6 CH2 INV Switch

Atthemaximum(CAL)position,thereisno changeintheverticalscale. InvertstheCH2inputsignalvertically.Whenthe verticalmodeswitchisintheADDposition,CH2is subtractedfromCH1insteadofbeingaddedto CH1.ForCH1/CH2addition/subtractiondetails, seepage32.

3 ALT/CHOP SelectshowtheCH1andCH2signalsappear inthe display,intheDUALtracemode. Switch CHOP Thetwosignalsarechoppedand displayedsimultaneously(generally usedinslowsweep,1ms/DIVor slower). Thetwosignalsaredisplayed alternatively(generallyusedinfast sweep,0.5ms/DIVorfaster).

ALT

4 Vertical MODE Switch

SelectsCH1andCH2displaymodes.

CH1/CH2 TheCH1orCH2signalisdisplayed independently.

17

18

PANEL OVERVIEW

72-6802 User Manual TV-V Theoscilloscopetriggerswhena verticalvideosynchronizationsignal appears.Fortriggeringonthefield, select2ms/DIVasthehorizontalscale; fortriggeringontheframe(two interlacedfields),5ms/DIV. Theoscilloscopetriggerswhena horizontalvideosynchronization signalappears.Fortriggeringonthe line,select10us/DIVasthehorizontal scale.UsetheSWPVARknobto controlthenumberofwaveforms.

Trigger Controls

TV-H

1 Trigger ALT Whenpressed,theTRIGALTswitchconstantly togglesthetriggersourcebetweenCH1andCH2 Switch signal,sothatbothsignalscanbeclearlyviewed. TheTRIGALTswitchworkswhenthevertical modeisintheDUALpositionandalsothetrigger sourceisintheCH1orCH2position. Note

Note

ForTVVandTVHtrigger,the oscilloscoperespondsonlyto negativepolaritysignals.

TheTRIGALTswitchdoesnotworkwhenthe ALT/CHOPswitchisintheCHOPposition. ThefrequencycounterintheLCDdisplaydoes notworkwhentheTRIGALTswitchispressed. 3 Trigger

Theoscilloscopecannottrigger inputsignalswhentheir frequenciesarelessthan25Hz.

Changesthetriggerlevelvertically.

2 Trigger MODE Switch

TheTRIGGERMODEswitchselectswhenthe oscilloscoperespondstothetriggerconditions. AUTO NORM

LEVEL Knob ThetriggerlevelmovesupwhenthetriggerLEVEL knobisturnedclockwise. Thetriggerlevelmovesdownwhenthetrigger LEVELknobisturnedcounterclockwise.

Theoscilloscopesweepsregardlessof theexistenceoftriggerconditions. Theoscilloscopesweepsonlywhena triggerconditionoccurs.

19

20

PANEL OVERVIEW 4 Trigger SLOPE Switch Selectsthetriggeringslope. Wheninthe+position( ),theoscilloscope triggerswhenthepositiveslopeofthetrigger sourcesignalcrossesthetriggerlevel. Whenintheposition( ),theoscilloscope triggerswhenthenegativeslopeofthetrigger sourcesignalcrossesthetriggerlevel. 1 CH 1 (X) Input Terminal 2 AC/GND/ 5 Trigger Switch Selectsthesignalonwhichtheoscilloscopesweeps. SOURCE CH1 CH2 LINE CH1signalbecomesthetrigger source. CH2signalbecomesthetrigger source. ACpowerlinesignalbecomesthe triggersource.Usefulwhentheinput signalsynchronizeswiththepower linefrequency. Theexternaltriggerinput(TRIGIN terminal)signalbecomesthetrigger source. 3 GND DC

72-6802 User Manual

Input Terminals

AcceptstheCH1inputsignal.InXYmode,the CH1inputsignalbecomestheXaxis.

Selectsthecouplingmodefortheinputsignal. AC GND TheoscilloscopeblockDCcomponents includedintheinputsignal. Showstheground(zerovolt)levelon thedisplay.Thismodeisonlyfor checkingthereferencelevel;input signaldoesnotappearonthedisplay. Theoscilloscopedisplaysallofthe inputsignal.

DC Switch

Terminal Input

Acceptsagroundwire.TheGNDterminalis connectedtotheoscilloscopemainframe. AcceptstheCH2inputsignal.InXYmode,the CH2inputsignalbecomestheYaxis.

EXT

4 CH 2 (Y) Terminal

6 TRIG IN

Terminal

Acceptsanexternaltriggersourcesignal.The signalbecomesactivewhenthetriggerSOURCE switchisintheEXTposition. Inputimpedance:1M//25pF


21

ForXYmodedetails,seepage33.

22

PANEL OVERVIEW

72-6802 User Manual

Rear Panel Overview

SETUP
Default Settings
Beforepoweringuptheoscilloscope,setupthefrontpanelas follows. POWER Switch Off Center Center CH1 INTEN Knob FOCUS Knob Vertical MODE Switch

1 FUSE & Line HoldstheACmainsfuseandselectstheACline voltage,115Vor230V. Voltage Selector Input 3 CH1 2 AC Power Connector OUTPUT OutputstheCH1signal;approximately20mV/DIV whenterminatedwith50. AcceptstheACpowercord.

ALT/CHOP Switch CH 2 INV Switch CH1/CH2 Vertical Knob POSITION CH1/CH2 VOLTS/DIV Knob

ALT(

Terminal 4 Z AXIS INPUT Terminal

Acceptsanexternalintensitymodulation(Zaxis) signal;1kHzsquarewave,DC2MHz.Positive goingreducesintensity.Fordetailedspecifications, seepage39.

Released(Inversiondisabled) Centerofthedisplay

ForAClinevoltageselectionandfusereplacement,seepage37.

50mV/DIV

23

24

You might also like