Professional Documents
Culture Documents
of IC Packages
Figure 2-1
1
4. Defining the
Test Parameters
2
Fixtures and probes add The HP 4291A can also directly
inductive, capacitive and resistive display any two of the twenty
parasitics that add error to the impedance parameters, is
measurement. The ability to specified to operate from 0.1
compensate for these parasitics is ohms to 50 kohms, has the
absolutely important, often the distinctive capability of
fixture parasitic values overwhelm performing compensation to
the conductor impedance values. remove probe parasitics, and can
Port extension is also very easily extend the measurement
critical, IC packages do not easily port.
mount onto conventional 7 mm
fixtures, and so a cable must be 6. The Probe
connected from the measurement Figure 6-2
port to a fixture/probe. Using traditional handheld probes
on today's high-density 7. Setup For a Swept
From these measurement technologies can limit Frequency Measurement
contraints, an instrument must measurement accuracy and
have the minimum capabilities: damage fragile devices. The It is understood that the user will
Cascade Microtech FPM-1X Fine have a firm understanding of the
1 MHz to 1.5 GHz frequency range
Microprobe achieves a level of initial test conditions:
Wide impedance range accuracy accuracy for characterizing
beyond 50 ohms high-speed MCMs and SMT Start Frequency
assemblies. Refer to Figure 6-1.
Direct swept impedance results The FPM-1X combines electrical Stop Frequency
(|Z|, L, C, R) and mechanical accuracy needed
for fine-pitch applications. By Test Signal Level
Error correction at probe tip minimizing capacitance and
Number of Points (NOP)
ground inductance, the FPM-1X
Cable extension enables state-of-the-art probing of The Start and Stop frequencies are
modules with pads as small as 1 based on the IC package operating
The HP 4291A Impedance mil.
Analyzer is designed to address all conditions.
of these issues. Refer to the A unique ground system provides The Test Signal Level should be
Figure 5-1. This chart shows the the low-inductance required to set at 1 volt so that the instrument
typical 10% accuracy zones of a effectively troubleshoot can avoid operating at the noise
4-terminal pair impedance high-density applications. Refer to floor.
analyzer, a vector network Figure 6-2. To minimize loading
analyzer and the HP 4291A over effects, the probes provide as little The number of measurement
frequency and impedance. as 0.15pF of tip capacitance. points (NOP) can be set at 801
points to gain the maximum
amount of swept information.
Later, the NOP can be reduced
once a swept area has been fully
characterized and no anomolies
are located. The key benefit of a
small NOP is a faster sweep rate.
Please note that when changing
the NOP, frequency or test signal
level of the HP 4291A, the
calibration correction factors have
Figure 6-1 to be internally recalculated. This
Figure 5-1
tends to reduce accuracy. For the
absolute best accuracy, always
3
perform calibration under the As an aside, the impedance Compensation is different from
exact test conditions needed, function should also be selected at calibration. Compensation
otherwise a SOLA calibration may this time. To measure the mathematically removes residual
have to be performed again. self-inductance of a conductor, set capacitance, resistance and
channel 1 to be Ls (series inductance from test fixtures
For the best accuracy, it is inductance) and set channel 2 to using an user defined Open, Short
recommended that the HP4291A be Rs (series inductance). and 50 ohm Load. In the
test environment be set with the compensation sub-menu, before
following conditions before the 8. Calibration and executing any of the
performance of calibration: Compensation compensation constant
Fixture Selection measurements, it is recommended
Point Average Factor There is a very strong distinction that the COMP POINT be changed
Test head between calibration and from [FIXED] to [USER]. This is
compensation. Refer to Figure 8-1. so that the compensation data
Fixture selection is found under Calibration determines the matches with the calibration data.
the MEAS menu. Fixture selection calibration reference-plane for all
measurements. The calibration Calibration is always performed at
optimizes the calibration modeling
constants are generated at the the test head APC connector with
for greater accuracy by using a
reference plane and are based on the SOLA standards while
different model number for each
the measurements of the Short, compensation is always
standard fixture. Selecting the
Open, Load and Air-capacitor performed at the d-u-t
fixture to the [None] mode allows
(SOLA) standards. The SOLA measurement point.
users to use their own unique
fixture or probe. standards are factory defined
9. The Probing and
standards. Calibration is
The Point Average Factor will performed in the HP 4291A CAL Measurement System
mathematically average acquired menu. While in the CAL menu,
data at calibration, compensation before executing any of the Essentially these are the steps
and at the measurement. Selection calibration constant that mount the HP4291A test
of a point average factor of 8 or measurements, it is recommended station to a Cascade Microtech
greater will reduce the noise that the CAL POINTS be changed Summit 9000 (or 10000) Analytical
present in low threshold signal from [FIXED] to [USER]. This Probe Station and fully prepare
measurements. permits a faster swept calibration. the instrument to perform a
The penalty is that if the test continuous swept measurement,
The choice of the test head will displaying self-inductance in this
signal level is changed then the
fundamentally affect the accuracy example. Refer to Figure 9-1. This
SOLA calibration must be
of a measurement, each testhead is not a trivial exercise; the key
performed again.
is optimized for a particular assumption is that the test is
impedance range. A general rule is being performed by an user with a
to use the low-impedance test HP 4291A Measurement Range @ 10% Accuracy high level of microprobing
head for device-under-test (d-u-t) ,P SH GD Q FH >
5 D QJ H
Ω@
expertise. The microscope and
impedances of 10 ohms or less, . .
micromanipulators
+3$
0 0 0 * *
)UH T>+ ] @
Figure 8-1
4
Figure 9-3
5
h. Perform fixture COMPENSATION 10. The Test Data 11. Final Thoughts
using Open, Load and Short. For
an OPEN, use the exact The HP 4291A should be It is clear that the electrical
point-to-point distance and displaying data corresponding to characterization of IC packages is
alignment of the IC package Ls and Rs as a function of an accomplished art. It requires
conductor to separate the frequency. Refer to Figure 10-1. In the patience of a saint, the fine
microprobe contacts in air from this particular measurement, the motor skills of gymnast, the
each other. For a LOAD, use the IC package sample is a quad flat craftsmanship of a watchmaker
50 ohm load on the Impedance pack, has copper leads, and is and the latest design knowledge of
Standard Substrate. For a SHORT, normally used with digital a IC packaging engineer. The
remove the substrate, use a sheet circuitry using a 50 MHz clock purpose of this solution note is
of clean bright copper as the rate. A start frequency of 50 MHz not to fully describe in great detail
short, making contact on the and a stop frequency of 500 were how to make a specific type of
copper sheet using the exact used in the test. The conductor measurement, although it may be
point-to-point distance of the IC has a length of about 6 mm. interpreted as such. Instead, the
package conductor to separate the Typical results were a series objective is give enough general
microprobes. inductance of 500 pH and a series information for people to feel
resistance of about 10 mohms, comfortable using the HP 4291A
i. Verify that COMPENSATION has
which fits in for theoretical Impedance-Material Analyzer with
been correctly completed by
values. There is some evidence a Cascade Microtech Probe
measuring the substrate 50 ohm
that the anomalies centered at 215 Station for this application. Other
load and the copper short. The
MHz may have been caused by the key related subjects that have not
measured results should be very
grounding probe wiper arm. been addressed here are testing
close to 50 ohms and 0 ohms. If
guidelines, computer modeling,
not, then perform
data analysis and conversion,
COMPENSATION again.
ground plane measurements and
j. Remove the Impedance Standard evaluating multiple pin array
Substrate and copper sheet from matrices. Each of these topics are
the insulated sheet. Place the IC worthy of deep exploration and
package on the insulated sheet are well beyond the scope of this
and mechanically secure it so that document. Instead, the following
no movement will occur. references are provided to
education and guide people
k. Maneuver the microprobes to the involved with IC package testing:
desired conductor and make firm
contact to it. Figure 10-1 "Standard Measurement Guideline
for Electronic Package Inductance
l. The test system is now ready for a An alternate way of characterizing and Capacitance Model
swept inductance measurement. an IC package is to look at the Parameters" by JEDEC Solid State
impedance magnitude and phase Products Engineering Council
data. Refer to Figure 10-2.
"High Performance Packaging
Solutions" by Dr. Eric Bogatin,
published by Integrated Circuit
Engineering
Figure 10-2
6
12. Recommended List of
System Components
HP 4291A Impedance/Material
Analyzer with Option 012 Low
Impedance Test Head (1)
7
H
United States:
Hewlett-Packard Company
Test and Measurement Organization
5301 Stevens Creek Blvd.
Bldg. 51L-SC
Santa Clara, CA 95052-8059
1 800 452 4844
Canada:
Hewlett-Packard Canada Ltd.
5150 Spectrum Way
Mississauga, Ontario
L4W 5G1
(905) 206 4725
Europe:
Hewlett-Packard
European Marketing Centre
P.O. Box 999
1180 AZ Amstelveen
The Netherlands
Japan:
Hewlett-Packard Japan Ltd.
Measurement Assistance Center
9-1, Takakura-cho, Hachioji-shi,
Tokyo 192, Japan
Tel: (81) 426 48 3860
Fax: (81) 426 48 1073
Latin America:
Hewlett-Packard
Latin American Region Headquarters
5200 Blue Lagoon Drive
9th Floor
Miami, Florida 33126
U.S.A.
(305) 267 4245/4220
Australia/New Zealand:
Hewlett-Packard Australia Ltd.
31-41 Joseph Street
Blackburn, Victoria 3130
Australia
131 347 ext. 2902
Asia Pacific:
Hewlett-Packard Asia Pacific Ltd
17-21/F Shell Tower, Times Square,
1 Matheson Street, Causeway Bay,
Hong Kong
(852) 2599 7070
© Copyright 1994
Hewlett-Packard Company
Data subject to change
Printed in U.S.A. 06/94
5962-9725E