SPECTROPHOTOMETERS Atomic Absorption

AAS-210, Atomic Absorption Spectrophotometer


pressure lock ensures constant pressure and reliable contact. Multi-function autosampler features automatic standard sample preparation, automatic correction of sampling probe depth, automatic tracing and correction of liquid surface height in the sample vessel, with the sampling accuracy of 1% and reproducibility of 0.3%, realizing fully automation of graphite furnace analysis. Perfect safety protection measures: Alarm and automatic protection to fuel gas leakage, abnormal flow, insufficient air pressure and abnormal flame extinction in flame system. Alarm and protection function to insufficient carrier gas and protective gas pressure, insufficient cooling water supply and over-heating in graphite furnace system. Advanced and reliable electronic design: Adopting large-scale programmable logic array and Inter I2C bus technology. European type sockets and AMP adapters with high reliability to ensure long term reliability of the whole electronic system. Easy and practical analysis software: Easy-to-use MS analysis software is made under Windows operating system, realizing fast parameter setting and optimization. Automatic sample dilution, automatic curve fitting, automatic sensitivity correction. Automatic calculation of sample concentration (content), mean value, standard deviation and relative standard deviation calculation. Multi-elements determination in sequence to the same sample. Measured data and final results can be printed out and edited in Excel format.

Features: Innovated rich oxygen air-acetylene flame analysis technique: The patented flame analysis technique adopting rich oxygen air-acetylene flame as the substitution for nitrous oxide-acetylene flame for high temperature element analyses, such as Ca, AI, Ba, W, Mo, Ti, V, etc. Flame temperature is continuously adjustable between 2300-2950C, which makes it possible to choose the best atomization temperature for different elements. It features easy operation, low analysis cost and wide flame AAS analytical range. Rich oxygen flame will not pollute the environment and is not harmful to human bodies. It's a break-through in flame AAS analysis. Integrated flame/graphite furnace atomization system, changeable with flame emission burner: Automatically controlled changeover of the integrated flame and graphite furnace atomizer featuring easy operation and time saving eliminates human labor. A flame emission burner head can be installed to perform flame emission analysis to akali metals as K, Na etc. Accurate fully automated control system: Automatic 6-lamp turret, automatic adjustment of lamp current and optimization of light beam position. Automatic wavelength scanning and peak picking. Automatic spectral bandwidth changing. Automatic changeover between flame and graphite furnace operation, automatic optimization of position parameters, automatic ignition and automatic gas flow setting. Reliable fully automatic graphite furnace analysis: Adopting FUZZY-PID and dual curve mode light-controlled temperature control technique, temperature auto-correction technique, ensures fast heating, good temp. reproducibility & high analytical sensitivity. The temperature control accuracy is less than 1%. Graphite furnace with pneumatic control and

Comparison:
Characteristic Mass of Some Elements using rich oxygen air-C2H2 flame and other flame methods
Element
Ca Yb Eu AI Sr Sa Mo W Ga Sm La Sn

Wavelength Rich oxygen N20-C2H2 Air-C2H2 (nm) air-C2H2 flame flame flame
422.7 378.8 459.4 309.3 460.7 553.5 313.3 255.1 287.4 429.7 550.1 224.6 0.009 0.037 0.137 0.4 0.016 0.1 0.15 3.2 0.4 2.92 37.2 0.8 0.05 0.08 0.3 0.7 0.1 0.4 0.4 5.0 1.0 8.5 35.0 3.0 50 1.3 0.15 10.0 0.8 0.07 7.6 3.0

Atomic Absorption SPECTROPHOTOMETERS


SPECIFICATIONS
Wavelength range Wavelength accuracy Resolution 190-900nm Better than 0.25nm Two spectral lines of Mn at 279.5nm and 279.8nm can be separated with the spectral bandwidth of 0.2nm and valley-peak energy ratio less than 30%. 0.004A/30min The D2 lamp background correction capability at 1A is better than 30 times. The S-H background correction capability at 1.8A is better than 30 times. Motorized 6-lamp turret (Two high performance HCLs can be mounted on the turret to increase the sensitivity in flame analysis.) Wide pulse current: 0~25mA, Narrow pulse current: 0~10mA. 400Hz square wave pulse; 100Hz narrow square wave pulse + 400Hz wide square wave pulse. Single beam, Czerny-Turner design grating monochromator 1800 l/mm 277mm 250nm 0.1 nm, 0.2nm, 0.4nm, 1.2nm, auto switch over 10cm single slot all-titanium burner Corrosion resistant all-plastic spray chamber. High efficiency glass nebulizer with metal sleeve. sucking up rate: 6-7mL/min Room temperature-3000C 2000C/s 28mm (L) x 8mm (OD) Cd0.8*10-12g, Cu5*10-12g, Mo1*10-11g Cd3%, Cu3%, Mo4% R928 photomultiplier with high sensitivity and wide spectral range. Under Windows operating system Working curve auto-fitting; standard addition method; automatic sensitivity correction; automatic calculation of concentration and content. 1~99 times. automatic calculation of mean value, standard deviation and relative standard deviation Sequential determination of multi-elements in the same sample. With model function Measurement data and final analytical report printout, editing with Excel. 55 sample vessels and 5 reagent vessels Polypropylene 3ml for sample vessel. 20ml for reagent vessel 1l 1~99 times Accurate dual pump system. with 100I and 1 ml injectors. Cu: Characteristic concentration 0.025 mg/L, Detection limit0.006mg/L; Ba: Characteristic concentration 0.22mg/L AI: Characteristic concentration 0.4mg/L

Main Specification

Baseline stability Background correction

Lamp turret

Light Source System

Lamp current adjustment Lamp power supply mode Monochomator Grating

Optical System

Focal length Blazed wavelength Spectral bandwidth Burner Spray chamber

Flame Atomizer

Nebulizer Emission burner provided Temperature range Heating rate

Graphite Furnace

Graphite tube dimensions Characteristic mass Precision Detector Software Analytical method

Detection and Data Processing System

Reapeat times Multi-task function Condition reading Result printing

Standard RS-232 serial port communication Sample tray capacity Vessel material Vessel volume Minimum sampling volume Repeatable sampling times Sampling system

Graphite Furnace Autosampler

Characteristic Concentration and Detection Limit Function Expansion Dimensions and Weight

Air-C2H2 flame Rich oxygen Air-C2H2 flame

Hydride vapor generator can be connected for hydride analysis. Main unit Graphite furnace Autosampler 107X49x58cm,140kg 42X42X46cm, 65kg 40X29X29cm, 15kg

SPECTROPHOTOMETERS Atomic Absorption


AAS-110A/120A/130A, Flame/Graphite Furnace Atomic Absorption Spectrophotometer

Features: Innovated rich oxygen air-acetylene flame analysis technique (AAS-110A) The patented flame analysis technique adopting rich oxygen air-acetylene flame as the substitution for nitrous oxide-acetylene flame for high temperature element analyses, such as Ca, AI, Ba, W, Mo, Ti, V, etc. Flame temperature is continuously adjustable between 2300-2950C, which makes it possible to choose the best atomization temperature for different elements. It features easy operation, low analysis cost and wide flame AAS analytical range. Rich oxygen flame will not pollute the environment and is not harmful to human bodies. Its a break-through in flame AAS analysis. Integrated flame/graphite furnace atomization system, changeable with flame emission burner Automatically controlled changeover of the integrated flame and graphite furnace atomizer featuring easy operation and time saving eliminates human labor. A flame emission burner head can be installed to perform flame emission analysis to Alkali metals as K, Na etc. (AAS-110N120A) Accurate fully automated control system Automatic multi-lamp turret, automatic adjustment of lamp current and optimization of light beam position. Automatic wavelength scanning and peak picking Automatic spectral bandwidth changing

Automatic changeover between flame and graphite

furnace operation, automatic optimization of position parameters, and automatic ignition.

Reliable fully automatic graphite furnace analysis Adopting FUZZY-PID and dual curve mode light-controlled temperature control technique, temperature auto-correction technique, ensures fast heating, good temperature reproducibility and high analytical sensitivity. The temperature control accuracy is less than 1 %. Graphite furnace with pneumatic control and pressure lock ensures constant pressure and reliable contact. Perfect safety protection measures Alarm and automatic protection to fuel gas leakage, abnormal flow, insufficient air pressure and abnormal flame extinction in flame system; Alarm and protection function to insufficient carrier gas and protective gas pressure, insufficient cooling water supply and over-heating in graphite furnace system. Advanced and reliable electronic design Adopting large-scale programmable logic array and Inter 12C bus technology European type sockets and AMP adapters with high reliability to ensure long term reliability of the whole electronic system. Easy and practical analysis software Easy-to-use AAS analysis software is made under Windows operating system, realizing fast parameter setting and optimization. Automatic display of measured data, automatic calculation and analytical result automatic print out.

Atomic Absorption SPECTROPHOTOMETERS


SPECIFICATIONS
Wavelength range Wavelength accuracy Resolution 190-900nm 0.25nm Two spectral lines of Mn at 279.5nm and 279.8nm can be separated with the spectral bandwidth of 0.2nm and valley-peak energy ratio less than 30%. 0.004A/30min The D2 lamp background correction capability at 1A is better than 30 times. The S-H background correction capability at 1.8A is better than 30 times. ( only for AAS-110A/120A) 6-lamp turret (AAS-110A/120A), 4-lamp turret (AAS-130A) Auto-alignment, fully automated scan and peak-picking. Automatic adjustment and display. Wide pulse current: 0~25mA, Narrow pulse current: 0~10mA 400Hz square wave pulse; 100Hz narrow square wave pulse + 400Hz wide square wave pulse ( AAS-110A/120A ) Single beam, Czerny-Turner design grating monochromator 1800 l/mm 277mm 250nm 0.1 nm, 0.2nm, 0.4nm, 1.2nm, automatic change. 10cm single slot all-titanium burner Corrosion resistant all-plastic spray chamber. High efficiency glass nebulizer with metal sleeve. sucking up rate: 6-7mL/min Provided with AAS-110A/120A Room temperature-3000C 2000C/s 28mm (L) x 8mm (OD) Cd0.8*10-12g, Cu5*10-12g, Mo1*10-11g Cd3%, Cu3%, Mo4% R928 photomultiplier with high sensitivity and wide spectral range. Windows operating system Working curve auto-fitting; standard addition method; automatic sensitivity correction; automatic calculation of concentration and content. Maximum 20 times of repeat measurement, automatic calculation of mean value, standard deviation and relative standard deviation. Sequential mesurement of multi-elements in one sample. With model function Measurement data and final analytical report printout, editing with Excel.

Main Specification

Baseline stability Background correction

Lamp turret

Light Source System

Lamp current adjustment Lamp power supply mode Monochomator Grating

Optical System

Focal length Blazed wavelength Spectral bandwidth Burner Spray chamber

Flame Atomizer

Nebulizer Emission burner Temperature range Heating rate

Graphite Furnace

Graphite tube dimensions Characteristic mass Precision Detector Software Analytical method

Detection and Data Processing System

Reapeat times Multi-task function Condition reading Result printing

Standard RS-232 serial port communication

Characteristic Concentration and Detection Limit Function Expansion Dimensions and Weight

Normal Air-C2H2 flame Rich oxygen Air-C2H2 flame

Cu: Characteristic concentration 0.025 mg/L, Detection limit0.006mg/L; Ba: Characteristic concentration 0.22mg/L AI: Characteristic concentration 0.4mg/L (for AAS-110A )

Hydride vapor generator can be connected for hydride analysis. 1020 (L) x 490 (W) x 540 (H) mm (main unit), unpacked 80kg 420 (L) x 420 (W) x 460 (H) mm (graphite futnace), unpacked 50kg

SPECTROPHOTOMETERS Atomic Absorption


AAS-110B/120B/130B, Flame Atomic Absorption Spectrophotometer
Flame atomization system with flame emission burner A flame emission burner head can be installed to perform flame emission analysis to Alkali metals as K, Na etc. (AAS-110B/120B) Accurate fully automated control system Automatic multi-lamp turret, automatic adjustment of lamp current and optimization of light beam position. Automatic wavelength scanning and peak picking Automatic spectral bandwidth changing Automatic ignition Perfect safety protection measures Alarm and automatic protection to fuel gas leakage, abnormal flow, insufficient air pressure and abnormal flame extinction in flame system; Advanced and reliable electronic design Adopting large-scale programmable logic array and Inter 12C bus technology European type sockets and AMP adapters with high reliability to ensure long term reliability of the whole electronic system. Easy and practical analysis software Easy-to-use AAS analysis software is made under Windows operating system, realizing fast parameter setting and optimization. Automatic display of measured data, automatic calculation and analytical result automatic print out.
190-900nm 0.25nm Two spectral lines of Mn at 279.5nm and 279.8nm can be separated with the spectral bandwidth of 0.2nm and valley-peak energy ratio less than 30%. 0.004A/30min The D2 lamp background correction capability at 1A is better than 30 times. The S-H background correction capability at 1.8A is better than 30 times. (only for AAS-110B/120B) 6-lamp turret (AAS-110B/120B), 4-lamp turret (AAS-130B) Auto-alignment, fully automated scan and peak-picking. Automatic adjustment and display. Wide pulse current: 0~25mA, Narrow pulse current: 0~10mA 400Hz square wave pulse; 100Hz narrow square wave pulse + 400Hz wide square wave pulse ( AAS-110B/120B ) Single beam, Czerny-Turner design grating monochromator 1800 l/mm 277mm 250nm 0.1 nm, 0.2nm, 0.4nm, 1.2nm, automatic change. 10cm single slot all-titanium burner Corrosion resistant all-plastic spray chamber. High efficiency glass nebulizer with metal sleeve. sucking up rate: 6-7mL/min Provided with AAS-110A/120A R928 photomultiplier with high sensitivity and wide spectral range. Windows operating system Working curve auto-fitting; standard addition method; automatic sensitivity correction; automatic calculation of concentration and content. Maximum 20 times of repeat measurement, automatic calculation of mean value, standard deviation and relative standard deviation. Sequential mesurement of multi-elements in one sample. With model function Measurement data and final analytical report printout, editing with Excel.

Features: Innovated rich oxygen air-acetylene flame analysis technique (AAS-110B) The patented flame analysis technique adopting rich oxygen air-acetylene flame as the substitution for nitrous oxide-acetylene flame for high temperature element analyses, such as Ca, AI, Ba, W, Mo, Ti, V, etc. Flame temperature is continuously adjustable between 2300-2950C, which makes it possible to choose the best atomization temperature for different elements. It features easy operation, low analysis cost and wide flame AAS analytical range. Rich oxygen flame will not pollute the environment and is not harmful to human bodies. Its a break-through in flame AAS analysis.
Wavelength range Wavelength accuracy Resolution

SPECIFICATIONS

Main Specification
Baseline stability Background correction Lamp turret

Light Source System

Lamp current adjustment Lamp power supply mode Monochomator Grating Focal length Blazed wavelength Spectral bandwidth Burner Spray chamber Nebulizer Emission burner Detector Software Analytical method

Optical System

Flame Atomizer

Detection and Data Processing System

Reapeat times Multi-task function Condition reading Result printing

Standard RS-232 serial port communication Cu: Characteristic concentration 0.025 mg/L, Normal Air-C2H2 flame Characteristic Detection limit0.006mg/L; Concentration and Ba: Characteristic concentration 0.22mg/L Rich oxygen Detection Limit Air-C2H2 flame AI: Characteristic concentration 0.4mg/L (for AAS-110A) Hydride vapor generator can be connected for hydride analysis. Function Expansion Dimensions and Weight 1020 (L) x 490 (W) x 540 (H) mm (main unit), unpacked 80kg

Atomic Absorption SPECTROPHOTOMETERS


AAS-310/320, Flame Atomic Absorption Spectrophotometer
Features: High cost-effective flame AAS Reasonable design, adopting the same key parts as in high end instruments, ensures basic functions but less automation to provide an economic model for users Reliable integration of main unit with microprocessor Built-in microprocessor with necessary auto-control and data processing functions achieve high reliability of the instrument. Simple and easy operation Eye-catching digital display, multi-function data processing ability and fast function-key direct input realize easy and fast analysis.
SPECIFICATIONS
Wavelength range Wavelength accuracy 190-900nm 0.5nm Two spectral lines of Mn at 279.5nm and 279.8nm can be separated with the spectral bandwidth of 0.2nm and valley-peak energy ratio less than 30%. 0.005A/30min The D2 lamp background correction capability at 1A is better than 30 times.

Main Specification

Resolution Baseline stability Background correction

2 lamps are powered simultaneously ( one preheating )

Light Source System

Lamp current adjustment range: 0~20mA Lamp power supply mode Monochomator Grating Focal length Blazed wavelength Spectral bandwidth Adjustment Burner Spray chamber Powered by 400Hz square pulse Single beam, Czerny-Turner design grating monochromator 1800 l/mm 277mm 250nm 0.1 nm, 0.2nm, 0.4nm, 1.2nm, 4 steps Manual adjustment for wavelength and slit 10cm single slot all-titanium burner Corrosion resistant all-plastic spray chamber. High efficiency glass nebulizer with metal sleeve. sucking up rate: 6-7mL/min Manual adjustment mechanism for vertical, horizontal positions and the rotation angle of the burner Fuel gas leakage alarm R928 photomultiplier with high sensitivity and wide spectral range. Automatic adjustment of light source power. Light energy and negetive high-voltage aut-balance. LED display of energy and mesurment value, concentration direct reading Transient, time average, peak height, peak area. Integral time is selectable in the range of 0.1-19.9s. 0.1~99 Automatic calculation of mean, standard deviation and relative standard deviation. Repeating number is in the range of 1-99 Automatic curve fitting with 3~7 standards; Sensitivity auto-correction Measurement data, working curve, signal profile and analytical conditions can all be pronted out. Check current status of each function key Cu: Characteristic concentration 0.025 mg/L, Detection limit0.006mg/L;

Optical System

Flame Atomizer

Nebulizer Position adjustment Gas line protection Detector Electronic and micr-computer system Display mode Read mode

Detection and Data Processing System

Scale expansion Data processing mode Measurement mode Result printing Instrument self-check

Characteristic Concentration and Detection Limit Function Expansion Dimensions and Weight

Air-C2H2 flame

Hydride vapor generator can be connected for hydride analysis. 1020 (L) x 490 (W) x 540 (H) mm, 80kg unpacked

SPECTROPHOTOMETERS Atomic Absorption


How to choose yours AAS configuration:
Model 02-enriched Flame The number Auto flame*1 emission of HCL*2 Aligment
Yes Yes 6 Yes

Auto HP-HCL*3 Background correctiom sampler


Yes S-H, D2 Yes

PC Flow control rate control


Yes

Atomizer

AAS-210

AAS-110A

Yes

Yes

Yes

Yes

S-H, D2

No

Yes

AAS-120A

No

Yes

Yes

Yes

S-H, D2

No

Yes

AAS-130A

No

No

Yes

No

D2

No

Yes

AAS-110B AAS-120B AAS-130B AAS-320

Yes No No No

Yes Yes No No

6 6 4 4

Yes Yes Yes No

Yes Yes No No

S-H, D2 S-H, D2 D2 D2

No No No No

Yes Yes Yes No

Air-C2H2 Flame*4 02-enriched Flame*1 Automatic Graphite Furnace*4 Hydrid Generation*5 Air-C2H2 Flame 02-enriched Flame Manual Graphite Furnace Hydrid Generation Air-C2H2 Flame Manual Graphite Furnace Hydrid Generation Air-C2H2 Flame Manual Graphite Furnace Hydrid Generation Air-C2H2 Flame Manual 02-enriched Flame Hydrid Generation Air-C2H2 Flame Manual Hydrid Generation Air-C2H2 Flame Manual Hydrid Generation Air-C2H2 Flame Manual Hydrid Generation

Note: 02 -enriched flame*1 Our patentedair-C2H02-2 flame (Substitution for N-20C2H2 flame) The number of HCL*3 The number of HCLs could be loaded on the Turret High peformance HCL*3 Two high performence HCLs can be mounted on the turret to increase the sensitivity in flame analysis
Model Atomizer
Air-C2H2 Flame*4 02-enriched Flame*1 Graphite Furnace*4 Hydrid Generation*5 Air-C2H2 Flame 02-enriched Flame Graphite Furnace Hydrid Generation Air-C2H2 Flame Graphite Furnace Hydrid Generation Air-C2H2 Flame Graphite Furnace Hydrid Generation Air-C2H2 Flame 02-enriched Flame Hydrid Generation Air-C2H2 Flame Hydrid Generation Air-C2H2 Flame Hydrid Generation Air-C2H2 Flame Hydrid Generation

Elements
Ag, Au, Ba, Bi, Ca, Cd, Co, Cr, Cs, Cu, Fe, Ga, In, K, Li, Mg, Mn, Mo, Na, Ni, Pb, Pd, Pt, Rb, Rh, Sb, Sn, Sr, Te, Tl, Zn Be, Ca, Sr, Ba, Al, Ga, Si, Ge, Sn, Y, La, Sm, Eu, Yb, Ti, Zr, V, Cr, Mo, W Ag, Al, Au, Be, Bi, Cd, Co, Cr, In, Mn, Mo, Ni, Pb, Pd, Sb, Se, Sn, Sr, Te, Tl, V As, Se, Sb, Bi, Pb, Sn, Te, Ge, Hg Ag, Au, Ba, Bi, Ca, Cd, Co, Cr, Cs, Cu, Fe, Ga, In, K, Li, Mg, Mn, Mo, Na, Ni, Pb, Pd, Pt, Rb, Rh, Sb, Sn, Sr, Te, Tl, Zn Be, Ca, Sr, Ba, Al, Ga, Si, Ge, Sn, Y, La, Sm, Eu, Yb, Ti, Zr, V, Cr, Mo, W Ag, Al, Au, Be, Bi, Cd, Co, Cr, In, Mn, Mo, Ni, Pb, Pd, Sb, Se, Sn, Sr, Te, Tl, V As, Se, Sb, Bi, Pb, Sn, Te, Ge, Hg Ag, Au, Ba, Bi, Ca, Cd, Co, Cr, Cs, Cu, Fe, Ga, In, K, Li, Mg, Mn, Mo, Na, Ni, Pb, Pd, Pt, Rb, Rh, Sb, Sn, Sr, Te, Tl, Zn Ag, Al, Au, Be, Bi, Cd, Co, Cr, In, Mn, Mo, Ni, Pb, Pd, Sb, Se, Sn, Sr, Te, Tl, V As, Se, Sb, Bi, Pb, Sn, Te, Ge, Hg Ag, Au, Ba, Bi, Ca, Cd, Co, Cr, Cs, Cu, Fe, Ga, In, K, Li, Mg, Mn, Mo, Na, Ni, Pb, Pd, Pt, Rb, Rh, Sb, Sn, Sr, Te, Tl, Zn Ag, Al, Au, Be, Bi, Cd, Co, Cr, In, Mn, Mo, Ni, Pb, Pd, Sb, Se, Sn, Sr, Te, Tl, V As, Se, Sb, Bi, Pb, Sn, Te, Ge, Hg Ag, Au, Ba, Bi, Ca, Cd, Co, Cr, Cs, Cu, Fe, Ga, In, K, Li, Mg, Mn, Mo, Na, Ni, Pb, Pd, Pt, Rb, Rh, Sb, Sn, Sr, Te, Tl, Zn Be, Ca, Sr, Ba, Al, Ga, Si, Ge, Sn, Y, La, Sm, Eu, Yb, Ti, Zr, V, Cr, Mo, W As, Se, Sb, Bi, Pb, Sn, Te, Ge, Hg Ag, Au, Ba, Bi, Ca, Cd, Co, Cr, Cs, Cu, Fe, Ga, In, K, Li, Mg, Mn, Mo, Na, Ni, Pb, Pd, Pt, Rb, Rh, Sb, Sn, Sr, Te, Tl, Zn As, Se, Sb, Bi, Pb, Sn, Te, Ge, Hg Ag, Au, Ba, Bi, Ca, Cd, Co, Cr, Cs, Cu, Fe, Ga, In, K, Li, Mg, Mn, Mo, Na, Ni, Pb, Pd, Pt, Rb, Rh, Sb, Sn, Sr, Te, Tl, Zn As, Se, Sb, Bi, Pb, Sn, Te, Ge, Hg Ag, Au, Ba, Bi, Ca, Cd, Co, Cr, Cs, Cu, Fe, Ga, In, K, Li, Mg, Mn, Mo, Na, Ni, Pb, Pd, Pt, Rb, Rh, Sb, Sn, Sr, Te, Tl, Zn As, Se, Sb, Bi, Pb, Sn, Te, Ge, Hg

AAS-210

AAS-110A

AAS-120A

AAS-130A

AAS-110B AAS-120B AAS-130B AAS-320

Note: Air-C2H2 Flame & Graphite Furnace *4 This is integrative. automatically controlled changover of flame and graphite furnace atomizer. Hydride Generation*5 This is a optional accessory MRC.VER.01-10.12