You are on page 1of 16

Gradient Operation in HPLC

Dr. Shulamit Levin Medtechnica


levins@medtechnica.co.il levins@medtechnica.co.il http://www.forumsci.co.il/HPLC http://www.forumsci.co.il/HPLC http://shulalevin.tripod.com http://shulalevin.tripod.com

Gradient Operation in HPLC

Introduction - Optimizing Gradient Separations


The following diagram illustrates the cycle time parameters that are used in a typical gradient

Dr. Shulamit Levin Medtechnica


levins@medtechnica.co.il levins@medtechnica.co.il http://www.forumsci.co.il/HPLC http://www.forumsci.co.il/HPLC http://shulalevin.tripod.com http://shulalevin.tripod.com
% Organic

Final Hold
Gr ad ien t

Inject

Gradient Time Total Cycle Time

Re-equilibration

Outline
Introduction Strategies for Higher Throughput Gradient Separations to Achieve Maximum Throughput and Optimal Resolution system solutions method solutions

Typical Problems Encountered in Gradient Chromatography


Non-reproducible retention times Difficulties to transfer from analytical to narrowbore columns Long reequilibration times Long cycle times (injection to injection) More efficient analyses desired

Dr. Shulamit Levin, Medtechnica

Introduction - Options to Improve Sample Throughput


System Solutions:
Reduce Gradient Delay Volume Decrease Re-equilibration time time Reduce Injection Cycle time Modify Instrument Use Multiple Parallel Columns Adjust Detector Sampling Rate

Outline
Introduction Strategies for for Higher Throughput Gradient Separations to Achieve Maximum Throughput and Optimal Resolution
system system solutions reduce gradient delay volume decrease decrease re-equilibration re-equilibration time reduce injection cycle time method solutions

Method Solutions:
Use Shorter Gradients Use Higher Flow Rates Use Shorter Columns Use a Smaller Particle Size Decrease Re-equilibration Time Increase Temperature

Test Probe Structures


S O
0.20

Initial Separation and Conditions


C CH3 O

N N N N
H2C CH

Conditions: Column: Symmetry C18, 5 m, 4.6 X 50 mm Mobile phase: A=0.1% TFA in water, B=0.1% TFA in acetonitrile

0.18 0.16 0.14

OH

NH2

0.12 0.10 AU 0.08

1-hydroxy-7-aza-benzotriazole
H3C

methyl 3-amino-2-thiophenecarboxylate

C
Gradient: 0-60% B in 8 minutes Column temperature: 30.0 C Flow rate: 1 mL/min.

0.06 0.04

O S O NH2
H2N C

0.02 0.00 0.00 0.50 1.00 1.50 2.00 Minutes 2.50 3.00 3.50 4.00

Detector: 254 nm Injection volume: 1 L

4-methylbenzene sulfonamide

4-aminobenzophenone

A - 1-hydroxy-7-aza-benzotriazole B - 4-methylbenzene sulfonamide C - methyl 3-amino-2-thiophenecarboxylate D - 4-aminobenzophenone

H. Weller, Bristol-Myers Squibb Pharmaceutical Research Institute

Dr. Shulamit Levin, Medtechnica

Volumes in an HPLC System


System Volume
Waters 2690 <650 L
B D

Determination of System Precolumn Volume


Definition: Delay volume is the volume of plumbing between the point the gradient is formed and the inlet of the column.
Measured by linear gradient Measured by step gradient
1.0 0.8 0.6

Column Volume Column Detector


Height

1.0 0.8 0.6

Solvent delivery

Injector

Height
0.4 0.2 0.4 0.2

Proportioning Valve

Extra Column Volume

Volume (L)

Volume (L)

System components affecting dwell volume: Pump Gradient Mixers Injector

Effect of Precolumn Volume


Reducing Delay Volume
in theory
Final Hold

Gradient Shape and Precolumn Volume


Theoretical Gradient No column , 1.00 mL/min 3.9x150 mm, 1.00 mL/min 2.1x150 mm, 0.29 mL/min

real life

Final Hold

% Organic

% Organic

Gr

Inject

Gradient Time

Re-equilibration

Inject

Gr ad ien t

ad ie

nt

%B

Gradient Time

Re-equilibration

Total Cycle Time

Total Cycle Time

Delay Volume
0 5 10

Re-equilibration Time
15 Minutes 20 25 30

Dr. Shulamit Levin, Medtechnica

Outline
Introduction Strategies for Higher Throughput Gradient Separations to Achieve Maximum Throughput and Optimal Resolution
system solutions reduce gradient delay delay volume decrease re-equilibration time reduce injection cycle time time method solutions
Calculation of of Gradient Equilibration Volume
Re-equilibration Re-equilibration is is a a necessary necessary part part of of gradient gradient chromatography. chromatography. Both Both the the HPLC HPLC system system and and the the column column must must be be at at initial initial conditions conditions at at the the beginning beginning of of each each run run to to ensure ensure reproducible reproducible chromatographic chromatographic separations. separations. The The re-equilibration re-equilibration volume volume can can be be divided divided into into two two parts, parts, the the system system washout washout and and the the column column re-equilibration. re-equilibration. For For good good system/column system/column equilibration equilibration

Gradient Shape and Re-equilibration


Theoretical Gradient No column , 1.00 mL/min 3.9x150mm, 1.00 mL/min 2.1x150mm, 0.29 mL/min

%B

Delay Volume
0 5 10

Re-equilibration Time
15 Minutes 20 25 30

Column Re-equilibration
3.9 x 150 mm 2.1 x 150 mm 1.0 x 150 mm 0 1.0 mL/min 0.290 mL/min 0.066 mL/min 0
3xSystem 5xColumn

trr = = (3V (3VT T+ + 5V 5Vcc)/F )/F


where: where: t trr is is the the re-equilibration re-equilibration time time in in minutes, minutes, V VTT is is the the total total system system volume, volume, V Vcc is is the the column column volume volume in in mL mL F F is is the the flowrate flowrate in in mL/min. mL/min. column r column volume volume = = 0.7( 0.7( r22L/2 L/2) ) system system volume volume = = 650-3000 650-3000 L L

10

Volume, mL

10

20

30

40

Time, min

Dr. Shulamit Levin, Medtechnica

Reduction of Re-equilibration Time


Reduce Re-equilibration time, two Approaches:
Final Hold

Reduction of Re-equilibration Time


(Approach (Approach 1 1- Increase Increase Flow Flow Rate) Rate)
Final Hold Final Hold

1. increase flow rate


Gradient Time Re-equilibration Total Cycle Time % Organic
nt

Gr

Gr

nt die G ra

ie ad

Inject

ie ad

Inject

Gradient Time Total Cycle Time

Re-equilibration % Organic

nt

Inject

2. reduce column volume

Gradient Time

Re-equilibration

AU

Dr. Shulamit Levin, Medtechnica

% Organic % Organic

Total Cycle Time

1
Final Hold

2
Final Hold

Column: 3.9 X 50 mm Column volume (c.v.) =0.60 mL 5 minute gradient @ 1 mL/min instrument delay volume (d.v.) = 650 L gradient volume = g t x c.v. = 3

Column: 3.9 X 50 mm

Column volume (c.v.) =0.60 mL 5 minute gradient @ 2 mL/min instrument delay volume (d.v.) = 650 L g x c.v. = 3 gradient volume = t

Gr

Reduction of Re-equilibration Time (Approach 1 - Increase flow rate)


Symmetry C18
Alkylphenones
ad Gr

nt die G ra

ie ad

Inject

Gradient Time

Re-equilibration % Organic

Inject

Gradient TimeRe-equilibration Total Cycle Time

Total Cycle Time

Total re-equilibration time, rt Total re-equilibration time, rt = {3(0.65) + 5(0.7)(0.6)}/1 = {3(0.65) + 5(0.7)(0.6)}/2 = 4.0/1 = 4.0/2 = 4.0 min. = 2.0 min.

nt

re-equilibration time is reduced by 50%

Reduction of Re-equilibration Time


(Approach (Approach 2 2 -- Reduce Reduce Column Column Volume) Volume)

Final Hold

Final Hold

ad Gr

3.9 x 50 mm 40-100%B 3.0 min, 3 mL/min 4.5 min, 2 mL/min 9.0 min, 1 mL/min
0 2 4 Minutes 6 8

Inject

t ien

t ien

Gradient Time

Re-equilibration % Organic

Inject

Gradient Time

Re-equilibration

% Organic

Total Cycle Time

Total Cycle Time

Column: 2.1 X 50 mm Column volume (c.v.) = 0.170 mL 5 minute gradient @ 1 mL/min instrument delay volume (d.v.) = 650 L gradient volume = t g x c.v. = 0.85 Total re-equilibration time, t r = {3(0.65) + 5(0.7)(0.17)}/1 = 2.5 min.

Column: 2.1 X 20 mm Column volume (c.v.) = 0.069 mL 5 minute gradient @ 1 mL/min instrument delay volume (d.v.) = 650 L gradient volume = t g x c.v. = 0.35 Total re-equilibration time, t r = {3(0.65) + 5(0.7)(0.069)}/1 = 2.0 min.

re-equilibration time is reduced by 20%

Reduction of Re-equilibration Time (Approach 2 - Reduce Column Volume)


1.20

Outline
Introduction Strategies for Higher Throughput Gradient Separations to Achieve Maximum Throughput and Optimal Resolution
system solutions reduce gradient delay volume decrease re-equilibration re-equilibration time reduce injection cycle time method solutions

2.1 x 50 mm

1.10

1.00

2.1 x 20 mm

0.90

Conditions: Symmetry C18, 5 m Mobile phase: A=0.1% TFA in water, B=0.1% TFA in acetonitrile Gradient: 0-60% B in 5 minutes Column temperature: 30.0 C Detector: 254 nm Injection volume: 1 L Flow rate: 1 mL/min.

0.80

0.70

0.60 AU 0.50

0.40

-Increase throughput by approximately 25% by reducing column volume from 0.170 (50 mm length) to 0.069 (20 mm length).

0.30

0.20

0.10

0.00 0.00 0.50 1.00 1.50 2.00 2.50 Minutes 3.00 3.50 4.00 4.50 5.00

Faster Chromatography Shorter Cycle Time


Reducing Total Cycle Time
Reduce Cycle Times by:
Programming a system purge in the method which occurs during the injection of the sample or... Employing two columns and performing column switching.
0

Samples per day 320 +16%

AU

276
Column: 3.9 X 50 mm 3 mL/min 3 min. gradient

Minutes

24

Dr. Shulamit Levin, Medtechnica

Higher Throughput Through Column Switching


Column: Symmetry, C18, 5 L, 19 X 50 mm Flow Rate: 20 mL/min. Re-equilibration requires 5 column volumes = 150 mL = 7.5 min. Re-equilibration period = unused time

Summary - System Solutions


Reducing Gradient Delay Volume Use 0.12 mm (0.005") i.d. tubing instead of 0.25 mm (0.009") to reduce system volume; Shorten all tubing lengths; Reduce the extra-column volume in the auto-injector by employing a smaller loop

Inject Gradient Hold Re-equilibration


Column switching can reduce runtimes by approx. 30%

0
1 column 2 columns

10

12.5

20

0 hrs

10.5hrs

16.5hrs

Note: a second pump must be employed

Remove gradient mixers

Summary - System Solutions (cont'd)


Achieve Faster Gradient Chromatography By...
Reducing Re-equilibration Time Reduce Reduce column volume volume Increase flow rate Reducing Cycle Time Program injection injection to occur occur during during re-equilibration re-equilibration Implement Implement column switching switching

Outline
Introduction Strategies for Higher Throughput Gradient Separations to achieve maximum throughput and maximize resolution system solutions method solutions use use shorter shorter gradients gradients use use higher higher flow flow rates rates use shorter columns use shorter columns use use smaller smaller particle particle sizes sizes
increase temperature

Dr. Shulamit Levin, Medtechnica

Optimizing Separations
The following diagram illustrates the cycle time parameters that were optimized to achieve high throughput goals.
Final Hold
% Organic

Factors Influencing Resolution in Gradient RP-HPLC Separations...


Resolution, Resolution, Rs, Rs, between between two two closely closely resolved resolved analytes analytes in in gradient RP-HPLC is a function of column efficiency gradient RP-HPLC is a function of column efficiency N, N, selectivity selectivity a a,, and and the the retention retention factor: factor:

Rs =

t w

N 4

ln

1 Bct0 + 1

Efficiency Efficiency Selectivity Selectivity Retention Retention


Gr ad ien t

c = %B/minute =

% tg

1 B.
Re-equilibration

% . t0 + 1 tg

Inject

Gradient Time Total Cycle Time

Upon substitution of the actual variables ( %/tg (gradient time)) for c, gradient slope, one can see the relationship between gradient time and resolution, and....

What Factors Influence Gradient RP-HPLC Separations...


...further derivatization of this term shows the relationship between resolution and flow rate, F, and column length, L, or column volume, r22L/2.

Resolution Dependance on Gradient Time and Flow Rate for a Gradient Method
(Symmetry ,, 4.6 (Symmetry C C18 4.6 X X 50 50 mm, mm, 5 5 m) m) 18

Rs =
20

t w

N 4

ln

1 B. % tg

. . or2 .L/F + 1
t

Rs =

t w

N 4

ln

1 B. % . t0 + 1 tg
Rs

18 16 14 12 10 16 8 8 6 4 4 Gradient Time [min] 2 0 0.3 0.5 1 2 Flow [mL/min] 5 8 1

Efficiency Efficiency Selectivity Selectivity Retention Retention

} }

Efficiency Selectivity Retention


1 It is the effect of these variables, % F, tgg and L, that we will investigate. B . tg . t

Dr. Shulamit Levin, Medtechnica

1. Effect of changing gradient run time, tg 2. Effect of changing flow rate, F

. r2 .L/F + 1

Impact of Reducing Gradient Time (tg g) on Resolution


Conditions:
Absorbance

Summary Impact of Gradient Time on Resolution

Gradient Time: 8 minutes


Column: Symmetry C18, 5 m, 4.6 X 50 mm Mobile phase: A=0.1% TFA in water, B=0.1% TFA in acetonitrile Gradient: 0-60% B in noted gradient time Column temperature: 30.0 C Flow rate: 1 mL/min. Detector: 254 nm Injection volume: 1 L

Minutes

5.50

Gradient Time: 4 minutes


Absorbance

Resolution increases as gradient time increases Throughput decreases as gradient time increases

Minutes

3.50

Gradient Time: 2 minutes


Absorbance

-Longest gradient time provides best resolution -Shortest gradient time maximizes throughput
Minutes 2.50

Gradient Time: 1 minute


Absorbance

-Reducing just gradient time sacrifices resolution

Minutes

2.50

Impact of Flow Rate (F) on Resolution


Flow rate: 1 mL/min.
Absorbance

Summary Impact of Flow Rate on Resolution

Conditions: Column: Symmetry C18, 5 m, 4.6 X 50 mm Mobile phase: A=0.1% TFA in water, B=0.1% TFA in acetonitrile Gradient: 0-60% B in 4 minutes Column temperature: 30.0 C Detector: 254 nm Injection volume: 1 L

Minutes

5.00

Resolution goes through an optimum due to the combination of gradient expansion and decrease in plate count Optimum resolution is approximately 1 to 2 mL/min for most practical separation problems

Flow rate: 2 mL/min.


Absorbance

Minutes

3.50

Flow rate: 5 mL/min.


Absorbance

- Resolution goes through an optimum due to the combination of gradient expansion and decrease in plate count

Minutes

2.50

Dr. Shulamit Levin, Medtechnica

Resolution Dependance on both Flow Rate and Gradient Time for a Gradient Method
(Symmetry , 4.6 (Symmetry C C18 18 4.6 X X 50 50 mm, mm, 5 m)
Absorbance
1.20

Reduction of Cycle Time


Flow rate: 0.5 mL/min. Grad. time: 16 min.
Conditions: Column: Symmetry C18, 5 m, 4.6 X 50 mm Mobile phase: A=0.1% TFA in water, B=0.1% TFA in acetonitrile Column temperature: 30.0 C Detector: 254 nm Injection volume: 1 L

Rs =
2 0 1 8

t w

1 N ln % . . 4 t or2 .L/F + 1 B. tg

Efficiency Efficiency Selectivity Selectivity Retention Retention


16 14

Absorbance

} }

1.00

Flow rate: 1.0 mL/min. Gradient time: 8.0 min.

Minutes

14.00

12

10

Rs

16
8

0.60

Absorbance

4
2

Gradient Time [min] 2


0

3. Effect of changing gradient run time, tg, and flow rate, F

Flow rate: 2.0 mLmin. Gradient time: 4.0 min.

Minutes

9.00

Flow rate increased proportional to gradient time decrease. Elution pattern is maintained as cycle time is decreased resulting in an increase in throughput.

0.3 0.5 1 2 Flow [mL/min] 5 8 1

0.20

Absorbance

Flow rate: 5.0 mL/min. Gradient time: 2.0 min.

Minutes

3.50

Minutes

2.00

Summary Reduction of Cycle Time


Introduction Introduction

Outline
Strategies Strategies for for Higher Higher Throughput Throughput Gradient Gradient Separations Separations to achieve maximum throughput and maximize achieve throughput and maximize resolution resolution system system solutions solutions

To obtain the maximum sample throughput the gradient time must be adjusted inversely proportionally to the flow rate. As shown in the previous slide the sample throughput was increased by 800% upon increasing the flow rate to 5 mL/min. and reducing the gradient time to 2 minutes.

method method solutions use use shorter gradients gradients use higher flow use flow rates use use shorter columns columns use smaller particle use particle sizes sizes increase increase temperature temperature

Dr. Shulamit Levin, Medtechnica

10

The Number of Column Volumes per Minute Impacts Resolution

Impact of Column Length on Resolution


2 Approaches:

How Short is Too Short?


It is not the column length which influences the separation in so much as the number of gradient volumes moving across the column.

Approach 1: Gradient volume in not proportion to the column volume (gradient run time constant while changing the column length). Approach 2: Scale gradient volume in proportion to the column volume (change the gradient run time proporionally with the column length).

Dr. Shulamit Levin, Medtechnica

11

Column Volume to Gradient Volume Relationship (Approach 1) -Gradient volume not scaled scaled to column volume
50 mm column 20 mm column

Impact of Column Length on Resolution (Approach 1)

-Gradient Gradient volume volume not scaled scaled to to column column volume
,2.1 x 20 mm
2.1 x 30 mm
1.20 1.10

Conditions: Symmetry C18, 5 m Mobile phase: A=0.1% TFA in water, B=0.1% TFA in acetonitrile

2.1 x 50 mm

Gradient: 0-60% B in 5 minutes


Column temperature: 30.0 C Detector: 254 nm Injection volume: 1 L Flow rate: 1 mL/min.

Column volume = 0.5 mL 5 minute gradient @ 1 mL/min gradient volume = tg x f.r. = 5 Total volume = g.v./c.v. = 10 column vols.

Column volume = 0.2 mL 5 minute gradient @ 1 mL/min gradient volume = tg x f.r. = 5 Total volume = g.v./c.v. = 2 column vols.
AU

1.00

0.90

0.80

0.70

0.60

0.50

0.40

0.30

0.20

Maintain resolution by not scaling gradient volume proportionally to column volume. However maximun reduction of analysis time is not realized as when gradient volume is scaled.

0.10

0.00 0.00 0.50 1.00 1.50 2.00 2.50 Minutes 3.00 3.50 4.00 4.50 5.00

Column Volume to Gradient Volume Relationship (Approach 2) -Gradient Gradient volume volume scaled scaled to to column column volume volume
50 mm column 20 mm column

What Factors Influence Gradient RP-HPLC Separations...


L (column length) is varied. Gradient volume is scaled in proportion to the column volume.

Column volume = 0.5 mL

Column volume = 0.2 mL

5 minute gradient @ 1 mL/min


gradient volume = tg x f.r. = 5 Total volume = g.v./c.v. = 10 column vols.

2 minute gradient @ 1 mL/min


gradient volume = tg x f.r. = 2 Total volume = g.v./c.v. = 10 column vols.

Rs =

t w

N ln 4

1 B. % . t tg

. r2 .L/F + 1

Efficiency Selectivity

Retention
Terms are constant

}
12

Dr. Shulamit Levin, Medtechnica

Impact of Column Length on Resolution (Approach 2) -Gradient volume scaled to to column volume
2.1 x 10 mm Gradient time = 1 minute 2.1 x 20 mm Gradient time = 2 minutes 2.1 x 30 mm Gradient time = 3 minutes
1.20 1.10 1.00 0.90 0.80

Summary Impact of Column Length on Resolution

2.1 x 50 mm Gradient time = 5 minutes

Conditions: Symmetry C18, 3.5 m Mobile phase: A=0.1% TFA in water, B=0.1% TFA in acetonitrile Gradient: 0-60% B in noted time Column temperature: 30.0 C Detector: 254 nm Injection volume: 1 L Flow rate: 1 mL/min.

0.70 0.60 AU 0.50 0.40 0.30 0.20

-Reduce analysis time by >50%. -Trade-off: reduction in resolution

Maximum sample throughput is realized when the gradient volume is scaled proportionally to the column volume.

0.10 0.00 0.00 0.50 1.00 1.50 2.00 Minutes 2.50 3.00 3.50 4.00 4.50

Gradient Delay Time


Final Hold
Gr ad ien t
t en di ra

Final Hold

Impact of the Number of Column Volumes on Peak Shape


0.05

% Organic

% Organic

Inject

Gradient Time Total Cycle Time

Re-equilibration

Inject

Gradient Time Total Cycle Time

Re-equilibration

2.1 x 10 mm Peak Width for Peak # 2 = 0.0533

Column: 4.6 X 50 mm Column volume (c.v.) = 0.83 mL 5 minute gradient @ 1 mL/min instrument delay volume (d.v.) = 650 L gradient volume = tg x c.v. = 6 Total system volume = 0.7(c.v.) + d.v. = 1.2 mL

Column: 2.1 X 50 mm
0.04

Column volume (c.v.) = 0.17 mL 5 minute gradient @ 0.2 mL/min instrument delay volume (d.v.) = 650 L gradient volume = tg x c.v. = 6 Total system volume = 0.7(c.v.) + d.v. = 0.72 mL

0.03

Conditions: Symmetry C18, 5 m Mobile phase: A=0.1% TFA in water, B=0.1% TFA in acetonitrile Gradient: 0-60% B in 4 minutes Column temperature: 30.0 C Detector: 254 nm Injection volume: 1 L Flow rate: 2 mL/min.

AU 0.02

Column re-equilibration time = total vol./f.r. Column re-equilibration time = total vol./f.r. = 0.67/0.2 = 1.15/1 = 3.6 min. = 1.2 min.

0.01

gradient is delayed by a factor of 3


0.00 0.20 0.40 0.60 0.80 1.00 Minutes 1.20 1.40 1.60 1.80 2.00

Dr. Shulamit Levin, Medtechnica

13

Reducing the Effect of Gradient Delay Delay Volume


- Make Make gradient gradient steeper steeper by by increasing increasing the the flow flow rate rate or or decreasing decreasing the the gradient gradient
time time

Outline
Introduction Introduction Strategies Strategies for for Higher Higher Throughput Gradient Separations to achieve achieve maximum maximum throughput throughput and maximize resolution resolution system system solutions

0.05 0.04 0.03 AU 0.02 0.01 0.00 0.20

peak width for peak # 2 = 0.0533

0.40

0.60

0.80

1.00 Minutes

1.20

1.40

1.60

1.80

2.00

Increase flow rate (2 to 3 mL/min.)

Increase gradient slope by 50% (4 to 2 min. grad)

0.030 0.025 0.020 AU 0.015 0.010 0.005 0.000

peak width for peak # 2 = 0.0448

peak width for peak # 2 = 0.0501

method solutions use shorter gradients gradients use higher flow rates use shorter columns columns use smaller particle sizes increase temperature
2.00

0.20

0.40

0.60

0.80

1.00 Minutes

1.20

1.40

1.60

1.80

2.00

0.20

0.40

0.60

0.80

1.00 Minutes

1.20

1.40

1.60

1.80

Impact of Particle Size (dp) on Resolution

Comparison of Resolution Dependence on Particle Size


Symmetry C18, 3.5 m, 4.6 X 50 mm Max. Resolution @ 1 mL/min., 16 min. 20 gradient = 18.8 18
16 14 12 Rs Rs 10 8 6 4 2 0 0.3 2 0.5 4 Gradient Time [min] 8 6 4 2 0 0.3 2 0.5 1 Flow [mL/min] 4 Gradient Time [min]
AU

1.20 1.10 1.00 0.90 0.80 0.70 AU 0.60 0.50 0.40 0.30 0.20 0.10 0.00 0.00
1.00 0.90

Conditions:

Symmetry C18, 5 m, 4.6 X 50 mm Max. Resolution @ 1 mL/min., 16 min. 20 gradient = 16.2


18 16 14 12 16 10 8 8 16

Symmetry C18, 3.5 m, 2.1 x 50 mm Rs (peak A and B) = 11.97 Rs (peak B and C) = 8.79 Rs (peak C and D) = 7.35 C A
0.50 1.00 1.50 2.00

B
Minutes 2.50 3.00 3.50 4.00 4.50

Columns: Symmetry 1 C 8 , 5 m, 4.6 8, 3.5 m, X 50 mm and Symmetry 1C 4.6 X 50 mm Mobile phase: A=0.1% TFA in water, B=0.1% TFA in acetonitrile Gradient: 0-60% B in 4 minutes Column temperature: 30.0 C Detector: 254 nm Injection volume: 1 L Flow rate: 1 mL/min.

0.80 0.70 0.60 0.50 0.40 0.30 0.20

18, 5 m, 2.1 x 50 mm Symmetry C Rs (peak A and B) = 6.71 Rs (peak B and C) = 7.32 Rs (peak C and D) = 6.45 C

-Achieve increased resolution with the smaller particle size material in the same gradient time -Increase throughput and resolution with smaller particle size if flow rate is increased

B A
0.50 1.00 1.50 2.00 2.50 Minutes 3.00 3.50 4.00 4.50 5.00

1 Flow [mL/min]

1 5

1 8

0.10 0.00 0.00

Dr. Shulamit Levin, Medtechnica

14

Summary Impact of Particle Size on Resolution

Outline
Introduction Introduction Strategies Strategies for for Higher Higher Throughput Throughput Gradient Gradient Separations Separations to to achieve achieve maximum maximum throughput throughput and and maximize maximize resolution resolution system system solutions solutions method method solutions solutions use shorter gradients use higher flow rates use shorter columns use smaller particle sizes increase temperature

Resolution is increased as a result of using a smaller particle size. This is due to the increase in the number of theoretical plates. If the flow rate is increased as well as the particle size being decreased, an increase in sample throughput is realized with increasing resolution.

Impact of Temperature
Faster Separtation Narrower Peaks Lower Backpressure Lower Viscosity Greater Sensitivity-

Summary - Method Solutions


To obtain the fastest throughput:
increase flow rate decrease column volume decrease particle size scale gradient volume with decrease in column volume increase temperature to reduce viscosity of mobile phase allowing increases in flow rate

50 40 30 AU 25 C

2 3 Minutes

Dr. Shulamit Levin, Medtechnica

15

You might also like