Cyclic Fatigue Resistance of RaCe and Mtwo Rotary Files

in Continuous Rotation and Reciprocating Motion


Sekar Vadhana, BDS, Balasubramanian SaravanaKarthikeyan, MDS, Suresh Nandini, MDS,
and Natanasabapathy Velmurugan, MDS
Abstract
Introduction: The purpose of this study was to evaluate
and compare the cyclic fatigue resistance of RaCe (FKG
Dentaire, La Chaux-de-Fonds, Switzerland) and Mtwo
(VDW, Munich, Germany) rotary les in continuous rota-
tion and reciprocating motion. Methods: A total of 60
new rotary Mtwo and RaCe les (ISO size = 25, taper
= 0.06, length = 25 mm) were selected and randomly
divided into 4 groups (n = 15 each): Mt
c
(Mtwo NiTi les
in continuous rotation), R
c
(RaCe NiTi les in continuous
rotation), Mt
r
(Mtwo NiTi les in reciprocating motion),
and R
r
(RaCe NiTi les in reciprocating motion). A cyclic
fatigue testing device was fabricated with a 60

angle of
curvature and a 5-mm radius. All instruments were
rotated or reciprocated until fracture occurred. The
time taken for each instrument to fracture and the
length of the broken fragments were recorded. All the
fractured les were analyzed under a scanning electron
microscope to detect the mode of fracture. The
Kolmogorov-Smirnov test was used to assess the
normality of samples distribution, and statistical anal-
ysis was performed using the independent sample t
test. Results: The time taken for the instruments of
the Mt
r
and R
r
groups to fail under cyclic loading was
signicantly longer compared with the Mt
c
and R
c
groups (P < .001). Scanning electron microscopic obser-
vations showed that the instruments of all groups had
undergone a ductile mode of fracture. The length of
the fractured segments was between 5 and 6 mm, which
was not statistically signicant among the experimental
groups. Conclusions: Mtwo and RaCe rotary instru-
ments showed a signicantly higher cyclic fatigue resis-
tance in reciprocating motion compared with continuous
rotation motion. (J Endod 2014;40:995999)
Key Words
Cyclic fatigue, nickel-titaniumles, reciprocation, rotation
N
ickel-titanium les (NiTi) are commonly used in current endodontic practice. NiTi
les offer many advantages over stainless steel les such as exibility and elasticity
(1, 2). Despite these advantages, NiTi instruments appear to have a high risk of
separation (3). One of the reasons for the fracture of NiTi instruments is torsional
or cyclic fatigue (4, 5). Torsional fatigue occurs when the tip of the instrument
binds in the canal while the shank continues to rotate (3, 5). Cyclic fatigue occurs
when the instrument continues to rotate freely in a curvature, and at the point of
maximum exure, tension/compression cycles are generated until fracture occurs
(5). Increasing the resistance to le separation has been the main goal for ensuring
safety during endodontic instrumentation.
Conventional NiTi rotary endodontic les are manufactured by machining starting
wire blanks that are in the superelastic austenitic phase (6). Under stress, it changes to
the martensitic phase. One of the unique properties of the martensitic phase is that it has
excellent resistance to fatigue (7). The stress-induced transformation to the martensitic
phase is reversible (8). The temperature at which the martensitic phase gets trans-
formed to the austenitic phase is called the austenitic nish temperature. The higher
the austenitic nish temperature, the longer the le remains in the martensitic phase.
This phase transformational behavior and microstructure of NiTi can be optimized us-
ing thermomechanical processing. This ultimately increases the mechanical and fatigue
properties of the le (9). Another method of improving cyclic fatigue resistance is by
electropolishing the les. This surface treatment improves the surface smoothness,
thereby delaying the initiation of surface cracks (10).
An alternative method of increasing cyclic fatigue resistance is the use of rotary NiTi
instruments in reciprocating motion (11). Two reciprocating systems are currently
available: Reciproc (VDW, Munich, Germany) and WaveOne (Dentsply Maillefer, Bal-
laigues, Switzerland). Reciprocating NiTi les have a better cyclic fatigue resistance
when compared with that of continuous rotary NiTi les (12). Studies have been con-
ducted to evaluate the use of various rotary NiTi les including Mtwo (VDW), K3, Pro-
Taper (Dentsply Maillefer, Ballaigues, Switzerland), and Twisted Files (Sybron Endo,
Orange, CA) in reciprocating motion, and it has been proved that these les possess
better cyclic fatigue resistance (1214). Mtwo les, which have a cross-section similar
to that of Reciproc les (15), have improved cyclic fatigue resistance in both continuous
and reciprocating motion (12). RaCe les (FKG Dentaire, La Chaux-de-Fonds,
Switzerland) have a triangular cross-section with distinct positive cutting angles
(16). RaCe les have improved cyclic fatigue resistance in continuous rotation when
compared with that of ProTaper and Helix rotary les (17). However, to date, the effect
of reciprocating motion on RaCe les has not been studied. The null hypothesis is that
there is no difference in the cyclic fatigue resistance of RaCe and Mtwo rotary les using
continuous rotation and reciprocating motion.
From the Department of Conservative Dentistry and Endodontics, Meenakshi Ammal Dental College and Hospital, Meenakshi Academy of Higher Education and
Research, Maduravoyal, Chennai, India.
Address requests for reprints to Dr SureshNandini, Department of Conservative Dentistry andEndodontics, Meenakshi Ammal Dental College and Hospital, Meenakshi
Academy of Higher Education and Research, Alapakkam Main Road, Maduravoyal, Chennai 600 095, Tamil Nadu, India. E-mail address: nandini_80@hotmail.com
0099-2399/$ - see front matter
Copyright 2014 American Association of Endodontists.
http://dx.doi.org/10.1016/j.joen.2013.12.010
Basic ResearchTechnology
JOE Volume 40, Number 7, July 2014 Cyclic Fatigue Life of RaCe and Mtwo Files 995
Materials and Methods
A total of 30 new rotary Mtwo (VDW, Munich, Germany) and 30
RaCe instruments (FKG Dentaire SA, La Chaux-de-Fonds, Switzerland)
(ISO tip size = 25, taper = 0.06, length = 25 mm) were selected. All
the instruments were previously inspected under an optical stereo mi-
croscope (Zoom Stereo Binocular Microscope [ZSM-111], Hicksville,
NY); with 20magnication for any visible signs of deformation. None
of the instruments were discarded. All the les were then subjected to
cyclic fatigue testing.
Cyclic Fatigue Testing Device
A static cyclic fatigue testing device was custom fabricated for
this study (Fig. 1A and B). It consisted of a main metal frame
made of iron to which an articial canal system and a support for
the handpiece were being attached. The canal system, which simu-
lated a root canal, consisted of 2 adjustable metal frames made of
brass that can accommodate any instrument to its exact size and ta-
per. It was constructed with a 60

angle of curvature. The curvature


started at 5 mm from the tip of the canal. The WaveOne handpiece
was mounted over the support, which also ensured the correct posi-
tioning and placement of les to the same appropriate depth for all
the samples.
Cyclic Fatigue Test
Sixty samples were randomly divided into 4 groups (n = 15) ac-
cording to the type of rotary les and rotary motions used.
Group Mt
c
Fifteen Mtwo instruments were allowed to rotate in continuous
rotation (CW) motion using a WaveOne motor set at continuous rotation
mode with recommended torque control settings and at a constant
speed of 300 rpm.
Group Mt
r
Fifteen Mtwo instruments were allowed to rotate in reciprocating
motion (counterclockwise [CCW] = 170 and CW = 50) using a Wave-
One motor set on the WaveAll mode with recommended torque control
settings and at a constant speed of 350 rpm (11).
Group R
c
Fifteen RaCe instruments were allowed to rotate in continuous
rotation motion using a WaveOne motor set on the continuous rotation
mode with recommended torque control settings and at a constant
speed of 600 rpm.
Group R
r
Fifteen RaCe instruments were allowed to rotate in reciprocating
motion (CCW = 170 and CW = 50) using a WaveOne motor set on
the WaveAll mode with recommended torque control settings and at
a constant speed of 350 rpm (11). Glycerin (Glycerin Pure; AB Enter-
prises, Mumbai, India) was used as a lubricant after the use of each le
during instrumentation in this study. Instruments were rotated or recip-
rocated until fracture occurred. To obviate human errors, cyclic fatigue
Figure 1. (A and B) The custom fabricated static cyclic fatigue testing device. (C) Stereomicroscopic image of a new RaCe le. (D) Stereomicroscopic image of a
fractured RaCe le depicting the fracture at the D5D6 regions.
Basic ResearchTechnology
996 Vadhana et al. JOE Volume 40, Number 7, July 2014
testing was performed by viewing the rotating or reciprocating les
under a dental operating microscope (Seiler Microscope, St Louis,
MO) at a magnication of 25to precisely determine the time of frac-
ture. All les were tested by 1 operator while the other operator was
simultaneously operating the stopwatch. The length of the broken frag-
ments was measured using a vernier caliper under a dental operating
microscope (25 magnication).
Scanning Electron Microscopic Analysis
The fractured fragments of each instrument were collected from
all the groups. The fractured surfaces of the les were examined under
a scanning electron microscope (FEI Quanta FEG 200; Hillsboro, OR)
to determine the modal characteristics of fracture (Fig. 2AD).
Stereomicroscopic Analysis
A few samples were randomly selected from each group and
observed under an optical stereo microscope (40) before and after
the failure of the respective instruments.
Statistical Analysis
The mean and standard deviation of the time to fracture and the
fractured fragment length were calculated for all the experimental
groups. The Kolmogorov-Smirnov test was used to assess the normality
of sample distribution. Both intragroup and intergroup comparisons
of cyclic fatigue resistance of the samples were performed using the
independent sample t test (SPSS version .17; SPSS, Chicago, IL). A P
value of <.05 was used as a criterion for statistical signicance.
Results
Descriptive statistics for the 4 experimental groups are listed in
Table 1. It was observed that the cyclic fatigue resistance of the RaCe
and Mtwo NiTi rotary les increased signicantly when operated in
reciprocation compared with the continuous rotation mode. It was
also observed that Mtwo les showed signicantly more resistance to
fracture under cyclic loading than RaCe les. The mean length of the
fractured segments was observed as 56 mm, which was not statistically
signicant among the groups.
Scanning Electron Microscopic Analysis Results
Scanning electron microscopic analysis of fractured surfaces of all
the groups revealed craterlike formation along with numerous dimples
and microbubbles. These features indicate that the instrument had un-
dergone a ductile mode of fracture, which is predominantly observed in
cyclic fatigue failure.
Discussion
This study compared the cyclic fatigue resistance of RaCe and
Mtwo rotary les in continuous and reciprocating motion and also as-
sessed the mode of fracture under scanning electron microscopy. This
study rejects the proposed null hypothesis and has shown that both
Mtwo and RaCe rotary les have improved cyclic fatigue resistance in
Figure 2. Scanning electron microscopic images of fractured les showing microbubbles, craters, valleys, and dimples. RaCe le used in continuous rotation (A)
and in reciprocation motion (B). Mtwo le used in continuous rotation (C) and in reciprocation motion (D).
Basic ResearchTechnology
JOE Volume 40, Number 7, July 2014 Cyclic Fatigue Life of RaCe and Mtwo Files 997
reciprocating motion when compared with continuous rotation. Mtwo
rotary les were selected because they have a cross-section similar to
Reciproc les (12), and RaCe les were chosen because there are no
studies using these les in reciprocating motion to date. Cyclic fatigue
testing can be performed using a static or dynamic model (15, 18).
In a static model, the instrument does not move axially. This creates
alternate compressive and tensile stresses in a particular area of the
instrument, leading to premature failure. The dynamic model
incorporates cyclic axial movement, which provides a better clinical
simulation and increases the lifespan of rotary les, but the
amplitude, speed of pecking motion, and axial movement are purely
subjective in clinical practice. The ability to constrain the les in a
precise trajectory is also difcult in dynamic testing (18).
There is no evidence for testing rotary les under standardized
specications (3, 19). Hence, a nontooth static model of a
standardized articial canal was used in this study, which minimizes
the inuence of other variables of le separation other than cyclic
fatigue.
Earlier cyclic fatigue studies have been conducted using various
angles of curvatures such as 30

, 60

, and 90

. A canal of 30

does
not constrain the le properly, whereas a curvature of 90

incorporated
more stresses to the le (13, 20). The center of curvature in most of the
studies was between 5 and 7 mm from the tip of the instrument (4).
Thus, in this study, an articial canal was fabricated with a 60

of cur-
vature with the maximum point of curvature at 5 mm from the tip of the
instrument.
In this study, fatigue life of the instruments was evaluated by time
and not by the number of cycles to fracture (NCF). The time to fracture is
easier to record than NCF. In continuous rotation motion, the assess-
ment of the number of cycles to fracture can be obtained by simply
multiplying the rotational speed by the time elapsed until fracture
occurred. However, in reciprocation motion, the NCF can be deter-
mined only by knowing the amplitude of the oscillating motion with a
constant time unit, which is not provided by the manufacturers (15).
Both Race and Mtwo rotary les showed better resistance to cyclic
fatigue in reciprocation compared with continuous rotation motion in
this study. The fatigue life is determined by the number of times the
crack opens and closes with each cycle. In continuous rotation motion,
a 360

rotation is completed in 1 cycle, which leads to increased stress


levels and more widening of surface cracks.
Tensile stresses are concentrated at a particular point of the instru-
ment in continuous rotation, which culminates to the fracture of the in-
strument. However, in reciprocating motion using the WaveAll mode
(CCW = 170 and CW= 50), it takes 3 cycles to complete one 360

rota-
tion (12). The stresses are distributed to 3 points around the working
portion of the le, minimizing the opening of surface cracks (15). In
our study, the cyclic fatigue life of RaCe les in reciprocation was 1.9
times more than that of continuous rotation, whereas Mtwo les per-
formed 2.9 times better in reciprocation than in continuous rotation.
This increase in the cyclic fatigue life of RaCe les in reciprocation
compared with continuous rotation was not proportional compared
with that of Mtwo les. The reason could be attributed to the difference
in the rotational speeds of RaCe les when used in reciprocation (350
rpm) and continuous rotation (600 rpm). Further studies are required
to conclude the role of rotational speeds in the cyclic fatigue life of rotary
les in reciprocating motion.
In our study, Mtwo les performed better than RaCe rotary NiTi
les in both continuous and reciprocating motion. Previous studies
have shown controversial results regarding the cyclic fatigue life of
Mtwo and RaCe in continuous rotation when compared with other les
(17, 1921). In reciprocating motion, Mtwo has been proven to have a
similar cyclic fatigue life to that of Twisted Files and Reciproc and
superior resistance compared with WaveOne (12).
Resistance to cyclic fatigue depends on various factors such as diam-
eter, metal mass, exibility, cross-sectional shape, and surface nish of ro-
tary les (22). RaCe les have an electropolished surface nish and good
exibility and thus are expected to possess a better cyclic fatigue life (23).
In our study, it showed a reduced fatigue life in comparison with Mtwo.
This could be caused by the presence of alternating cutting edges
in RaCe les. Although the alternating spiraled and nonspiraled seg-
ments of RaCe les increase its exibility, maximum bending occurs
at these segment junctions. Because exural stresses are conned to
the uted portion of the working area and are not distributed evenly
throughout the length of the le, fatigue occurs more rapidly in RaCe
les (22). The stereomicroscopic analysis of the new RaCe les in
our study showed that the junction of the spiraled and nonspiraled seg-
ments exists at the D5D6 region (Fig. 1C). The synergistic effect of the
presence of such a junction at 5 mm of the RaCe le (which corre-
sponds to the maximum curvature of the articial canal) and the static
mode of testing could have resulted in premature failure of RaCe les.
Figure 1D shows that the failure occurred at the segment junctions.
Electropolishing potentially removes the residual stresses on the instru-
ment and increases the fatigue life (10). However, the role of electro-
polishing in the cyclic fatigue resistance of the instruments cannot be
considered in this study because the cross-sections of both the le
systems (RaCe and Mtwo) were different.
Another determining factor for the cyclic fatigue life of les is the
shape of the le at its circumference. The forces required to disrupt the
molecular attraction can be decreased by providing a greater mass of
rounded surface following the cutting edge of les at its circumference
(22). At the radial extremity, RaCe les have a more angular surface
because of their triangular cross-section in comparison with that of
Mtwo les. This might lead to increased stresses in a small area, which
ultimately resulted in cyclic fatigue failure with less tension.
Scanning electron microscopic observations have shown that all
specimens of both the groups depicted the presence of microbubbles.
Circular abrasion marks, dimples, and craterlike formation conrmed
that all instruments had undergone cyclic fatigue.
On analyzing the data regarding the length of the fractured frag-
ments, there was no statistical difference in the mean length of all the
instruments tested. In this study, almost all the instruments fractured
between D5 and D6. The fractured length of the instruments ranged
TABLE 1. Intra- and Intergroup Comparison of Mean and Standard Deviation of Time Taken to Fracture of Instruments and Length of Fractured Instruments
Groups
Continuous Reciprocating
n = 15 Time SD Length SD Time SD Length SD
Mtwo 283.27 58.346
A
5.533 0.3994
a
815.73 82.893
B
5.633 0.4419
a
RaCe 173.67 49.844
A
5.467 0.4419
a
337.67 39.811
B
5.500 0.4629
a
P value <.001 .668 <.001 .426
A signicant difference in P values is marked as
A
and
B
(P < .05). An insignicant difference in P values is marked as
a
and
a
(P > .05).
Basic ResearchTechnology
998 Vadhana et al. JOE Volume 40, Number 7, July 2014
from4.55.5 mm, which conrmed the positioning of the les in a pre-
cise trajectory. Further studies may be required on the canal centering
ability and cutting efciency of RaCe and Mtwo rotary NiTi les in recip-
rocating motion.
Conclusion
Within the limitations of this study, Mtwo and RaCe rotary les
were more resistant to cyclic fatigue in reciprocating motion when
compared with continuous rotation motion. Mtwo les performed bet-
ter than RaCe les in both continuous and reciprocating motion.
Acknowledgments
The authors deny any conicts of interest related to this study.
References
1. Taschieri S, Necchi S, Rosano G, et al. Advantages and limits of nickel-titanium in-
struments for root canal preparation: a review of the current literature. Schweiz
Monatsschr Zahnmed 2005;115:10005.
2. Parashos P, Messer HH. Rotary NiTi instrument fracture and its consequences.
J Endod 2006;32:103143.
3. Plotino G, Grande NM, Cordaro M. A review of cyclic fatigue testing of nickel-
titanium rotary instruments. J Endod 2009;35:146976.
4. Sattapan B, Nervo GJ, Palamara JE, et al. Defects in rotary nickel-titanium les after
clinical use. J Endod 2000;26:1615.
5. Peters OA. Current challenges and concepts in the preparation of root canal systems:
a review. J Endod 2004;30:55967.
6. Thompson SA. An overview of nickeltitanium alloys used in dentistry. Int Endod J
2000;33:297310.
7. Paque F, Ganahl D, Peters OA. Effects of root canal preparation on apical geometry
assessed by micro-computer tomography. J Endod 2009;35:10569.
8. Shen Y, Zhou HM, Wang Z, et al. Phase transformational behavior and mechanical
properties of thermomechanically treated K3XF nickel titanium instruments.
J Endod 2013;39:91923.
9. Shen Y, Zhou H, Zheng Y, et al. Current challenges and concepts of the thermome-
chanical treatment of nickel titanium instruments. J Endod 2013;39:16372.
10. Praisarnti C, Chang JWW, Cheung GSP. Electropolishing enhances the resistance of
nickel-titanium rotary les to corrosion-fatigue failure in hypochlorite. J Endod
2010;36:13547.
11. Yared G. Canal preparation with only one Ni-Ti rotary instrument: preliminary ob-
servations. Int Endod J 2008;41:33944.
12. Pedulla E, Grande NM, Plotino G, et al. Inuence of continuous or reciprocating
motion on cyclic fatigue resistance of 4 different nickel-titanium rotary instruments.
J Endod 2013;39:25861.
13. Gambarini G, Rubini AG, Al Sudani D, et al. Inuence of different angles of recip-
rocation on the cyclic fatigue of nickel-titanium endodontic instruments. J Endod
2012;38:140811.
14. De-Deus G, Moreira EJ, Lopes HP, et al. Extended cyclic fatigue life of F2 ProTaper
instruments used in reciprocating movement. Int Endod J 2010;43:1063.
15. Lopes HP, Elias CN, Vieira MVB, et al. Fatigue life of Reciproc and Mtwo instruments
subjected to static and dynamic. J Endod 2013;39:6936.
16. Schafer E, Oitzinger M. Cutting efciency of ve different types of rotary nickel-
titanium instruments. J Endod 2008;34:198200.
17. Kim HC, Yum J, Hur B, et al. Cyclic fatigue and fracture characteristics of ground and
twisted nickel-titanium rotary les. J Endod 2010;36:14752.
18. Li UM, Lee BS, Shih CT, et al. Cyclic fatigue of endodontic nickel titanium rotary in-
struments: static and dynamic tests. J Endod 2002;28:44851.
19. Yao JH, Schwartz SA, Beeson TJ. Cyclic fatigue of three types of rotary nickel-titanium
les in a dynamic model. J Endod 2006;32:557.
20. Bhagabati N, Yadav S, Talwar S. An in vitro cyclic fatigue analysis of different end-
odontic nickel-titanium rotary instruments. J Endod 2011;12:34.
21. Plotino G, Grande NM, Melo MC, et al. Cyclic fatigue of NiTi rotary instruments in a
simulated apical abrupt curvature. Int Endod J 2010;43:22630.
22. McSpadden JT. Mastering Endodontic Instrumentation, 1st ed. Chattanooga, TN:
Cloudland Institute; 2007:516.
23. Anderson ME, Price JWH, Parashos P. Fracture resistance of electropolished rotary
nickeltitanium endodontic instruments. J Endod 2007;33:12126.