You are on page 1of 63

Basic Operation of the AccuTOF

And Getting Specification Data

3/9/2006

Basic Procedure (This is after you will get good evacuation value 3x10-5Pa or less)
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.

Cerate New Project file(Folder) (with Analysis list)


MCP conditioning (in MS tune manager)
Do Noise Coefficient of Avarager
Base Line Noise Check
Resolution test (tuning)
Make calibration file
Sensitivity Test
Exact Mass Test
Negative tune and making tune file
(other Attachment test)

(page-9 b. create new project)


(page-13 a-2 MCP conditioning & a-3)
(page-13 a-4 Optimum of base line noise)
(page-17 b Check base line noise level)
(page-18 c Resolution test)
(page-34 e Calibration)
(pare-43 f Sensitivity test)
(page-56 Exact Mass test)
(page-62)
(page-62)

**You have to do a-2 & a-3 every re-starting vacuum system.

1. Start AccuTOF
Drying Gas(to Flange)

Nebulizing Gas(to Probe)

Needle 1 (ESI/APCI/etc.)

Error LED
EVAC SW

Needle 2(for dual ESI or Corona ESI)


VENT SW
Heater (ESI/APCI/etc)
Power ON SW
Power OFF SW

Gas Flow Control valve


a.

Turn on Power ON SW (VENT SW LED will be lightning)

b.

Turn on EVAC SW (Must need N2 gas presser more than 600Mpa) > Start pumping down
EVAC SW LED will be stop the LED-blinking until evacuation presser 1.4x104Pa or less
Error LED will be light if revolution of turbo 1 & 2 reach 80% after next in 8 minutes.
Or pump-system will be something wrong.

c.
d.

Start the computer, printer and etc.


After light on the EVAC LED must wait 4095 minutes.
It is EVAC-cont-pb has timer for protect the MCP from spark of damage.
(Finally Presser will be reach to 1or3 x10 5 Pa or less) (at least 5x10-5pa)
(Reason: MCP is very sensitive spark of damage, easy to break by spark.)
(But! if Evacuation pressure condition is good that pressure reach to normal value
before 4095 minutes you can invalid timer on EVAC-cont-pb)
For example: After shot time vent for maintenance of orifice 1, 2 , ring lens and RP, TMP
or any short time electricity maintenance.
But either way to need check presser level (1to3 x 10-5pa)
How to Check presser is e. Check evacuation condition and MS monitor following

e.

Check evacuation condition


You can check evacuation presser on the MS-tune or MS-monitor.
(MS-tune is in the MassCenter main software , MS-monitor on desitop)
Open from

f.

or

After start pump down recommend to check on MS Monitor


(Check Vacuum Gage level and MCP Elapsed Time on Exhaust tab)
>(see a Detail of MS Monitor as below)

2. Shutdown AccuTOF (pump stop) operation.


a. Confirm LC or any solvent is not running.
b. Close Neblizing Gas and Drying Gas valves. (Operation on PC)
c. Close MS Tune Manager and MassCenter software (On PC)
d. Confirm N2 gas presser more than 600kPa.
(If N2 Gas presser dose not reach more than 600kPa, You can not vent evacuation.)
e. Push VENT switch on front panel.
VENT switch will be blinking next 10 minutes.
f. After 10 minutes VENT switch will be light.
g. You can see N2 gas that blow out from the Orifice 1.
h.
i.
j.
k.

Close N2 Gas line (If necessary)


Shutdown Computers and LC-system. (If necessary)
Push OFF switch on front panel. (If necessary)
Turn off Main Breaker (If necessary)

Now the AccuTOF Stop completely.


A.

Explanation of MassCenter
MassCenter Main program Every AccuTOF software run from her
(Expect Data-manager, MS-monitor)
Data Manager Program

Maintenance MassCenter files

Periodic Table Program


Isotopic Simulator Program
MS Monitor Program

Monitor every unit of AccuTOF Condition

Force System to Quit

If MassCenter freeze, use this Icon for quit


MassCenter software.
3

a. Detail of MS-Monitor

[General tab]
Unit select tab
Around panels door SW

Actual condition

Condition LED (You have red LED, when some function is wrong)
Also if you have red-LED, you can not choose any Instrument Mode on the MS-tune.
(See next page)
Can not choose any mode

[Ion Source tab]


There is flange open/close sensor
at the flange.

Condition LED
4

[Explain actual voltage of Orifice 1,2,and Ring lens]


Set point of Orifice 1,2 and Ring Lens Voltage is
,
,
on right pictures

Ring Lens

: Ring lens Voltage


(Potential difference between Orifice 1 and Ring lens)
: Orifice 1 Voltage
(Potential difference between Orifice 1 and Orifice2)
: Orifice 2 Voltage

Ion Guide

Orifice1

Orifice2

(Potential difference between Ion Guide Bias and Orifice2)


RP
Actual Voltage is :
Ring Lens is Ion Guide bias Voltage[Analyzer tab] + Orifice 1 + Orifice 2 + Ring Lens Set Voltage.
Orifice 1 is Ion Guide bias Voltage[Analyzer tab] + Orifice 2 + Orifice 1 Set Voltage.
Orifice 2 is Ion Guide bias Voltage[Analyzer tab] + Orifice 2 Set Voltage.

[Analyzer tab]

Sensor in Ion Guide Power Unit.


Ion Guide Bias Voltage

Condition LED
5

TMP

[Detector Tab]

If you have never do MCP conditioning


after run evacuation,

Condition LED

[Exhaust Tab]

Condition LED

If you do not have More then 4950min, you can not choose Instrument Mode on the MS-tune.

b. Detail of Force System to Quit

b-1.

Click here
Get MassCenter Process
b-2

You can see running process on here


b-3

Click here Terminate MassCenter System (process stop)

b-4
Click Get MassCenter Process again then ascertain the following window. Then click OK
(If you have process again,
Click Terminate MassCenter System also again.)

b-5

Click Close (close window)

B.

About MassCenter Software and opration.


Sample Inlet Control

Calibration

Tuning

Agilent1100 Managet

MS Tune Manager

Calib. Manager
Manager

Control Agilent1100

Tuning Parameter

Mass Assign Information

Agilent1100 Parameter

Single Acquisition

Make file of Calibration and


Mass Drift Compensation

MassCenter Main
(D ata Acquisition
on
PRO JEC T
MS Acquisition setting
Agilent1100 parameter

Calibration Data

MS Tune Parameter

Drift Compensation Data


Continuous Measurement

Analysis List
Acquistion Data
Quantity Data Quantity Project

etc .

Data Reduction
Chromatgrum

Spectrum Viewer

Chromatgram reduction

Spectrum reduction

Real Time Chromatgram Viewer

Real Time Spectrum Viewer

Quantity Manager
Make Quantity Project
Data measurement

Quantity Data Reduction

[MassCenter Main]
MassCenter Main is fundamental system on AccuTOF Software.
Following applications are started from MassCenter Main.
MS Tune Manager,
Mass Calibration Manager
Agilent1100 Manager
Chomatgrum Viewer
Spectrum Viewer.
Each parameters are saved in Project Folder. You can create your own new project and save every
parameters and data files. You can administer every data file and parameter in each project folder.
[For example] If you change any parameter, not affect any parameters in another project folder.
Note: Mass calibration files are administered in MassCenterMain. Accordingly, Calibration files are
common files every projects.
8

1. Start Mass Center Main software


a. Start MassCenter Main from desktop.

Double click

b. Create New Project


b-1. File > New

b-2. Open New window


Choose (double click) Project Wizard
Then OK

If you have had Created wizard files,


check Load Setting then choose from here
This case no choose then Next>
b-3. Create project name(TEST)
Choose or create save folder name
(C:/ACCUTOF)
then Next>

b-4. Choose Create New Project


Check Create Analysis list
(If you want create analysis list
same time, check here)
New Analysis List name
(Spec-test)

If you want to create new project from to have created project, check here and choose here.
(This mean, you want create new project almost same condition with you are having project.)
Then Next>
b-5. Choose your own parameter(Field)
Select here
This case(Spec-test) you should choose:
Sample Location,
Acquisition Data,
Acquisition Data Folder,
MS Acquisition Condition,
Agilent1100. (If customer has Aglent1100)
b-6. Finish Wizard
Click Finish

If you want save this setting condition, check save setting then put new name to here.
[Change Project]
Choose File > Open Project

10

If you have Agilent1100 LC-auto sampler and detector choose use in Agilent1100
Instrument > Configuration > Details tab
Module use in System
Auto Sampler : within Agilent1100
UV/VIS detector : within Agilent1100

2. Conditioning MassCenterMain Window

Analysis List
Project Name
Agilent 1100 status panel
Open Agilent1100 manager from here

Project View
MS Status Panel
Open MS Tune manager from here

Disk free space panel

Queue View
Displaying process in measurement

11

a.

Open Analysis List


a-1-1 File > Open

a-1-2 Choose Analysis List File name

Analysis List

Other Window
Choose your own to show window from here

3. MS tuning
a-1. Open MS Tune manager from MassCrenterMain.
[MS Tune manager Window]

12

Check evacuation presser It will be 1.**x10-5Pa or less.(it will be green bar)


a-2. MCP conditioning (It need to do, when first using after pumping down.)
a-2-1

If you cant choose MCP conditions menu, you should check vacuum condition on the MS monitor.

a-2-2

Set MCP conditioning parameter.


Voltage =2500v
Time=120min
Than Click start

a-2-3

Close window after finish conditioning.


If do not increase voltage on this window,
you should check door-SW or HV-PB.
Actual voltage
Close window

a-3. After finish MCP conditioning increase High Voltage of Flight-Tube, Reflectoron, and Pulser Unit by manually.
a-3-1. Every voltage set to0 then operate mode to Warm Up mode. (MCP=0v)
a-3-2. open monitor window and vacuum window watch base line and presser when increasing voltage.
If you have arcing in analyzer area, those may change. If you have arcing, decrease voltage immediately.
Or back to Evacuation Ready mode.
Also open MS monitor window then watch actual voltage.
a-3-3. Flight tube voltage: -300v step to -3000v. Leave and watch about 30-60sec each steps voltages.
a-3-4. Reflectorn voltage: 50v step to 300v. Leave and watch about 30-60sec each steps voltages.
a-3-5. +/- pulser voltages: 50v step to 300v. Leave and watch about 30-60sec each steps voltages.
a-3-6. Flight tube voltage: -100v step to -5000v. Leave and watch about 30-60sec each steps voltages.
a-3-7. Reflectorn voltage: 50v step to 600v. Leave and watch about 30-60sec each steps voltages.
a-3-8. +/- pulser voltages: 50v step to 600v. Leave and watch about 30-60sec each steps voltages.
a-3-9. Flight tube voltage: -50v step to -7000v. Leave and watch about 30-60sec each steps voltages.
a-3-10. Reflectorn voltage: 50v step to 900v. Leave and watch about 30-60sec each steps voltages.
a-3-11. +/- pulser voltages: 50v step to 778v. Leave and watch about 30-60sec each steps voltages.
a-3-12. set every voltage to the performance report setting then leave 30 minutes and watch base line/presser.
Conform NO ARCING in the Flight Tube area.

13

a-3-13. increase MCP voltage slowly 100v step to 2500v.


Do same way on Negative mode before operating negative analysis.
a-4. Optimum of base line noise
a-4-1. Choose Tools > Noise Coefficient of Averager

It is optimum of base line noise on the FASTFLIGHT.


a-4-2. Start process. Click Start

It is only to do, when install, move and replace the FASTFLIGHT.


It is not daily routine.
a-5. Open Tuning parameter and MS monitor window.
[For example : Window layout]

14

a-6. Open tune parameter windows


a-5-1 right click on each function icon then choose Display or Display New Window

a-7 Open MS monitor window


a-6-1 Click here

a-8. About mode menu

You can choose operate mode


by two ways

If you did not do MCP conditioning , you can not choose Warm-UP, Standby and Operate Mode
Also, if you did not pass time 4095minutes or more after pumping down.

15

[Explanation of each mode]


Evacuation Ready >>

Evacuation presser is 2x10-4Pa or less.


High volt (analyzer field and Ion Source) OFF
Temperature OFF
(Usually choose this mode , when the MassCenter is closing.)

Warm-UP

>>

(Daily Rest position)

High volt (analyzer field, Including MCP-voltage) ON


High volt (Ion Source) OFF
Temperature OFF
(Choose this mode , when short time close the MassCenter.
and usual rest operation (night time))

Standby

>>

(Daily Rest position)

High volt (analyzer field, Including MCP-voltage) ON


Temperature ON
High volt (Ion Source) OFF

Operate
Notes:

>>

Every High Volt and Temperature ON

N2 Gas valves are separate with this mode. Any time can do open/close.

a-8. About MS monitor window


a-8-1.

Change display MS Range from Setup

Right click then choose Setup

16

[Spectrum Monitor Parameter]


Start MS and End MS
Same as Scan speed on sector type MS

Averager Sampling Time. (ns)


(Same as data sampling points)

Short time is better for Mass accurate.


But can not available on wide scan mass range.
(Too much data point on one data)(too big data size)
Up to 2000 = 0.5ns (? not sure)

a-8-2.

Zoon up MS monitor

There are three way for zoom up.


1.

MS range and Intensity


Drag slanting then choose own zoom
up size.

2. Zoom Up Only Mass range


Drag horizontally
(Do not make window)
3. Zoom up Only Intensity
Drag Vertically
(Do not make window)

4.

Double click in window to back to initial Size

17

b.

Check Base line noise level (test-1)

b-1 set in each value as following.


Analyzer(Advanced) setting
Pusher volt.= 777.8v(or see performance sheet)
Pulling volt=-777.8v(or see performance sheet)
Flight Tube=-7000v
Detector Setting
= 2000v
ESI Ion Source Setting
Needle volt =

0v

Orifice 1 volt =

0v

Orifice 2volt =

0v

Ring Lens =

0v

Spectrum Monitor parameter


Range =100 to 1000
Spec. Recording Interval = 0.4 s
Data Sampling Interval = 0.5 ns
(other parameter will be any setting OK)

Take Operate Mode

Intensity count (up low < 100)

Noise width is 100 counts or less.


(Usually 50 to 60 counts)

18

C. Resolution Test
R>6000
Sample : Reserpine m/z=609+ (using concentration 50to100ppb with Methanol )
On infusion mode = use syringe pump
Pump Flow : 0.05 to 0.2 ml/min
(Never use on customers LC-pump for Resolution test = Never pour a Reserpine to LC-bottles)
Usually high concentration is better for resolution. (High peak intensity)
c-1. Set up Syringe Pump and connect to ESI Probe.
Syringe Pump
Syringe (5 or 10 ml)
Connect between syringe
and probe.
c-2.

Set syringe inside diameter if necessary.


(see Inside Diameter list on syringe pump)

c-3.

Set Base Tuning condition that from a last page of the performance report.
c-3-1.

Choose ESI+ mode

c-3-2.

Set condition from performance report.


[Analyzer setting General and Detail]
[Analyzer Advance Setting]

c-3-3.

Set condition
[ESI+ Ion Source Setting Voltage/Current and Temperature/Gas]
Needle 1 Voltage [V]

2000v

Ring Lens Voltage [V]

10v

Orifice 1 Voltage [V]

80v

Orifice 2 Voltage [V]

32v

[Temperature]

c-3-4

Desolvating Chamber [C]

250c (check temperature control)

Orifice 1 [C]

80c

Take to [Standby] mode

19

c-3-5

Set Detector Voltage (step up)


1000v > 1500v > 2000v > 2500v
(finally to see performance sheet)

c-3-6

check Probe condition


Adjust needle position
0.1 to 1 mm

(200ul/min)

3mm

(10ul/min)

0.1 mm

(1ml/min)

Pump flow on > Check probe tip to see sample coming.


c-3-7
c-3-8

Wait while increase temperature


Take [Operate] mode
Gas ON (ESI+ Ion Source Setting Menu)
Nebulizing Gas

ON (Gas Flow rate 1cc/min)

Dringing Gas

ON (Gas Flow rate 2.5-3.0 cc/min)

(Finally adjust gas flow by seeing ion intensity that increase highest)
c-9

Adjust Resolution and Intensity refer following Tune manual

AccuTOF: Peak Tuning


1.

General

Tuning is intended to introduce to the detector samples ionized by the ion source as efficiently as possible without any loss.
The sample sprayed from the ESI needle is dried by the N2 gas to be ionized, introduced to the vacuum from the orifice
1, and directed to the ion guide in the transfer field through the ring lens and orifice 2. Ions are advanced to the analyzer
field by the energy of the bias voltage at the ion guide, to which the bias voltage and high frequency voltage are applied.
They are adjusted by the focusing lens to maximize the volume to be forwarded to the orthogonal ion accelerator. They are
accelerated by the orthogonal accelerator toward the flight tube, reversed by the reflectron, and forwarded to the detector.
The mass number is determined by the time of the flight from the accelerator (start) to the detector (goal).
The TOF-MS determines the mass number solely by the time when the ions arrive at the detector (goal). As a result, if
the ions spread widely at the orthogonal ion accelerator (start), ions with the same mass number will have different start
positions, and arrive at the goal at different times, compromising resolution and sensitivity. For the analyzer field alone,
tuning the TOF-MS is intended to minimize the spread of ions in the vertical direction at the orthogonal ion accelerator
(horizontal spread is unrelated to the distance to the detector with reference to the direction of ion acceleration (downward
direction in the figure).
To achieve that goal, adjusting the focusing lens, pusher bias voltage, and suppress voltage is essential. (See 2-2 below
for details.)
20

AccuTOF Model Diagram

2.
2-1.

Description of Components
Orifice, Ring Lens, Ion Guide
Orifice1

Ring Lens

Orifice2

Ion Guide

760 Torr

2 Torr

3x103 Torr

(Ring Lens Voltage)


The ring lens is located between orifices 1 and 2. When a voltage is applied to the ring lens, ions that have passed
through the orifice 1 will efficiently be directed to the orifice 2. (See the figure above.) Thus, the ring lens voltage will
determine the detection sensitivity for the target ions. Optimizing the ring lens voltage is essential in detecting the target
substance at high sensitivity. Normally, the ring lens voltage is set low to detect ions with smaller molecular weight, and
high to detect ions of high mass compounds.
21

(Orifice 1 Voltage)
The orifice 1 voltage is the difference in potential between orifices 1 and 2. If the difference is great, ions that have
passed through the orifice 1 collide more frequently with vapor molecules in the area between orifices 1 and 2, resulting in
a hard condition where fragment ions are easily detected (called In-Source CID). Setting the voltage difference between
orifices 1 and 2 low will result in a mild condition where molecular ions are easily detected.
(Orifice 2 Voltage)
The orifice 2 voltage is the difference in potential between the ion guide and orifice 2. The potential difference is
intended to efficiently direct ions to the ion guide. When the voltage is increased while keeping constant the potential
difference between orifices 1 and 2, and the potential difference between the orifice 2 and ion guide is increased, the CID
may sometimes occur near the inlet of the ion guide.
(Ion Guide)
A bias voltage and a high frequency voltage of 3 MHz are applied to the ion guide. The bias voltage carries the ions to
the next space (lens and orthogonal ion accelerator). The high frequency voltage contains ions within the pole. The
AccuTOF, a quadrupole system with a high ion focusing power, features enhanced sensitivity and resolution, compared to
other types (hexode or octal system). When the frequency voltage is set high, high mass ions go through the ion guide
efficiently; when the voltage is set low, low mass ions pass efficiently; and the ions are transferred to the focusing lens at
the speed of the bias voltage.
[Optimizing Orifices 1 and 2, Ring Lens, and Ion Guide Levels]
To detect quasi-molecular ions of a target compound at high efficiency, set the ring lens voltage to 10-20 V, and the
orifice 2 voltage to 5-10 V. Introduce the sample by infusion (continuous injection from the pump) or flow injection (from
the manual injector or auto sampler) in a few flows. Monitoring the target ion intensity on the ion monitor, adjust the
orifice 1 voltage to search for the conditions where the target quasi-molecular ions are detected at the highest sensitivity.
We recommend that you adjust the orifice 1 voltage first, then optimize the ring lens voltage and orifice 2 voltage, and
fine-turn the orifice 1 voltage again. If you introduce the sample by flow injection, and optimize the ion source level, you
may perform a single experiment from the MS Tuning Manager. For analysis like In-Source CID, follow the same
steps. Set the orifice 1 voltage and ring lens voltage higher than the level appropriate for detection of quasi-molecular ions.
The [General] tab of the [Analyzer-Setting] panel has an item [Ion Guide High-Frequency Peaks Voltage]. Setting this
low will facilitate detecting low mass ion, eliminating high mass ions. Setting this high (up to 2500 V) will optimize
detection of high mass ions, eliminating low mass ions. Adjust the voltage level according to your applications. We
recommend setting this voltage to 1000-2000 V and 2500 V when you detect background ions and reserpine respectively.
The bias voltage of the ion guide is theoretically 27 V, although it varies among units. Changing this a great deal will
change the ion speed in the horizontal direction, driving ions to incorrect locations, and reducing the sensitivity. (See the
figure below.)
22

You need not change the bias voltage at the user, since it is optimized at the Factory or at the time of
installation.

Ion Guide Bias Voltage - Optimized

Ion Guide Bias Voltage - Inappropriate

The user needs to determine optimum voltage levels of the orifices 1 and 2, ring lens, and ion guide (RF) according
to the target samples and objective of analysis.

2-2.

Focusing Lens, Orthogonal Accelerator

(Focusing Lens)
The focusing lens is designed to efficiently direct to the orthogonal accelerator the ions that have traveled at the speed of
the ion guide bias voltage. An orifice with a horizontal hole is located between the lens and orthogonal accelerator, and the
lens will push the ions efficiently through the horizontal hole, creating a thin, flat ion packet at the orthogonal accelerator.

23

[Optimizing Focusing Lens Setting]


Use the parameters in the Details tab of Analyzer-Setting to set the focusing lens voltage. The parameters were set to the
optimum levels at the time of installation. Since the lens and analyzer of the AccuTOF are designed to resist contamination,
the voltages do not change very much.
However, they may change slightly over time, and the user needs to fine-tune the voltages occasionally.



Adjust the focus voltage in +/- 10 V steps to maximize the peak intensity and resolution.
Adjust the voltages of the condenser lens and quadrupole lens in 1-2 V steps to maximize the peak intensity and

resolution.

Adjust Right/Left in 0.5 to 1.0 V steps to maximize the peak intensity and resolution.
Adjust Top/Bottom in 0.1 to 0.5 V steps to maximize the resolution.


Right/Left and Top/Bottom are designed to introduce ions to the orthogonal ion accelerator. It is desirable to
adjust Right/Left and Top/Bottom within +/- 15 V and +/- 5 V respectively.
The above adjustment is related to the pusher bias, suppress, and Reflectron voltages. Optimize the focusing

lens setting in combination with these.
(Orthogonal Ion Accelerator)
The ion packet directed by the focusing lens to the orthogonal ion accelerator will arrive at the center of the pusher and
pulling plate. A voltage of +/-777.8 V will be applied simultaneously to the pusher and pulling plate, pushing the packet to
the flight tube with the accelerating energy. The pusher and pulling plate voltages will be reset to 0 V after the orthogonal
ion accelerator has projected the ions, accumulating another ion packet for next acceleration. The next packet will be
accelerated to the flight tube in the same manner. This process is repeated every 0.5 to 1.0 ns.
Another packet of ions is accumulated in the orthogonal accelerator while the pusher and pulling voltages are 0 V.
Creating an ion packet as flat as possible at this point will greatly affect the resolution and sensitivity. In theory, ions will
be forwarded from the focusing lens in parallel when the pusher and pulling voltages are 0 V. In actuality, however, the
voltages are expected to be fully 0 V when switched from +/-777.8 V to 0 V at high speed. Therefore, the AccuTOF
allows the operator to adjust the pusher bias and suppress voltage to offset the voltage at 0 V. This will assure ions as
perfectly flat as possible at the orthogonal ion accelerator.

24

[Optimizing Orthogonal Ion Accelerator Setting]


To adjust the orthogonal ion accelerator, go to Pusher Bias in the General tab of Analyzer-Setting in
Analyzer[Advance]-Setting.
You need not adjust the following in Analyzer[Advance]-Setting at all, since they are basic settings for ion flight. (except
for polarity change for Pos/Neg analysis).

Pusher voltage
Pulling voltage



Flight tube voltage

You need not adjust the suppress voltage most cases. When needed, adjust it in 0.05 to 0.1 V steps to maximize the
resolution. Adjust the voltage within +/-1.0 V at maximum.
Adjust Pusher Bias Voltage in the General tab of Analyzer-Setting when you adjust the focusing lens. Adjust it in 0.05 to
0.2 V steps to maximize the resolution and intensity.

2-3.
Reflectoron
(Reflectoron)
Ion packets accelerated from the orthogonal ion accelerator to the flight tube will be reversed the direction by the
Reflectoron toward the detector. Functions of Reflectoron are:
1.
The direction reversed along the way, ions with the same size will double the distance of flight (enhanced
resolution).
Passing through the Reflectoron, ions with the same mass number yet slightly different speed will be
2.
simultaneously arrive at the detector (enhanced sensitivity and resolution).

25

Reflectoron Model Diagram


[Optimizing Reflectoron Setting]
To adjust the Reflectoron voltage, go to Reflectoron Voltage in the Details tab of Analyzer-Setting.
You need not adjust this in most cases, since it is optimized at the time of installation. When needed, adjust
the voltage in 10 V steps to maximize the sensitivity and resolution.
Changing the Reflectoron voltage will have ions reverse the direction in front or deep inside the Reflectoron, changing
the distance of flight (= time of flight). As a result, the mass number display will significantly change. Make sure to
perform Mass Calibration when you changed the Reflectoron voltage.
Adjust the voltage in combination with other items of Analyzer and Analyzer[Advanced] Setting.
As discussed above, you need not adjust Analyzer and Analyzer[Advanced] Setting frequently. Use a peak you
are familiar with (e.g., peak from LC solvent) as an indicator for tuning, and perform peak tuning when the
resolution and sensitivity have significantly declined.
The next section will describe the tuning process in detail with reserpine as a sample. You need not use
the initial values as used in the instructions. You may use the initial conditions of the users system.
3.
Tuning from the factory setting
3-1. Set every parameter getting from performance sheet.(Analyzer setting, Ion source setting)
3-2. Tuning Procedure:
3-2-1. Tune the Top/Bottom potential in 0.1-0.5 V steps. Maximize peak height, not resolution (better keep
resolution ). The maximum should be within +/- 2V, otherwise there may be a problem.
3-2-2. Tune the Right/Left potential in 0.5-1.0 V steps. Maximize peak height, not resolution (better keep
resolution ). The maximum should be within +/- 10V, otherwise there may be a problem.
3-2-3. Tune the pusher bias in 0.2 volt steps. Maximize both intensity and resolution. Decrease the step size to
0.1 volt steps and then to 0.05 volts steps.
If you have not good resolution following step
3-2-4. Adjust both the Focus and Deflector (Condenser/Quadrupole )lens voltages. These potentials are I
nterrelated. Maximize both intensity and resolution. to tune the Focus potential in 10 V steps; set the
Condenser and Quadrupole lens potential to 0 initially and then tune in 1 V steps.
(Condenser and Quadrupole lens voltages are not big change the ion intensity.)
3-2-5. Repeat 3-2-1 to 3-2-3.
If you have not good resolution following step
3-2-5. Tune the reflectron in 20 volt steps. Look for good resolution.
3-2-6. Repeat 3-2-1 to 3-2-4.
If you have not good resolution following step
3-2-7. Set the Ion-Guide Bias voltage +/- 1v step then repeats 3-2-1 to 3-2-4.
26

Probably you can get resolution > 6000 by this procedure.


If you get resolution > go to C-10 at page 28
4. Tuning from initial condition.
4-1 Setup (Analyzer):

Peaks Voltage (RF): 2000 (maximum level).

Bias: 27 (or Performance sheet)

Pusher bias: 0.
Focus: -120

Condenser: 0
Quadrupole 0
Left/Right: 0
Top/Bottom: 0

Reflectron: 900v

Setup (Ion Source):


Ring lens: 5 to 10v (for reserpine)
Orifice 1: 30v (use 50 to 85V for Reserpine tuning -- see note 2)
Orifice 2: 5v

Needle: 2000 V. Current will be tens of nanoAmperes. If corona discharge occurs, the current will increase
to hundreds of nanoAmperes.

Drying gas: 2.5-3 on bubble meter.


Nebulizing gas: ~ 0.5 on bubble meter.

Needle position: inner needle protrudes 0.5 mm for flow rate of 200 ul/minute (see Note 3).

4-2. Tuning procedure


4-2-1. Tune the reflectron in 20 volt steps. Look for good resolution. (Not big reduce intensity)
4-2-2.

Tune the Top/Bottom potential in 0.1-0.5 V steps. Maximize peak height, not resolution (better keep
resolution ). The maximum should be within +/- 2V, otherwise there may be a problem.
4-2-2. Tune the Right/Left potential in 0.5-1.0 V steps. Maximize peak height, not resolution (better keep
resolution ). The maximum should be within +/- 10V, otherwise there may be a problem.
Tune the pusher bias in 0.2 volt steps. Maximize both intensity and resolution. Decrease
the step size to 0.1 volt steps and then to 0.05 volts steps.
If you have not good resolution following step
4-2-3. Adjust both the Focus and Deflector (Condenser/Quadrupole )lens voltages. These potentials are I nterrelated.
Maximize both intensity and resolution. to tune the Focus potential in 10 V steps; set the Condenser and
Quadrupole lens potential to 0 initially and then tune in 1 V steps.
4-2-2.

(Condenser and Quadrupole lens voltages are not big change the ion intensity.)
4-2-5. Repeat 3-2-1 to 3-2-3.
If you have not good resolution following step
4-2-5. Tune the reflectron in 20 volt steps. Look for good resolution.
4-2-6. Repeat 3-2-1 to 3-2-4.
If you have not good resolution following step
4-2-7. Set the Ion-Guide Bias voltage +/- 1v step then repeats 3-2-1 to 3-2-4.
27

Peaks Voltage (RF)

The peaks voltage affects the high-pass character of the RF ion guide. Decreasing this potential affects the low-m/z
cutoff. A very rough guide to the peaks voltage is given below (an approximation for easy recall):

Peaks
Voltag
e

Lowes
t m/z

500

50

1000

100

1500

150

2000

200

2500

250

A value of 600 volts is useful for small-molecule work (up to m/z 1000).
In-Source Fragmentation:
Reserpine fragments

For reserpine, drop the Peaks Voltage to 1000 and increase the orifice 1 potential from 30 V to 80 V. Adjust the ring lens
potential slightly to optimize fragmentation.
Definitions:

Orifice 1 refers to the potential difference between orifice 2 and orifice 1.

Orifice 2 refers to the potential difference between the ion guide and orifice 2. Do not adjust under normal
operation because this can induce fragmentation in the ion guide region.

Ring lens refers to the potential difference between the ring lens and orifice 1.

Coordinate system:

The x-direction is the direction of time-of-flight extraction (into the reflectron)

The y-direction is the direction of ion injection.


The z-direction is side-to-side (left and right when facing the ion source)

Focusing lenses:

Direction of ion motion into


reflectron region

The first lens is segmented into quadrants to steer the beam up and down and left and right:
+
-

28

The second lens is the einzel focusing lens, and the third lens is grounded.

Condenser: Average value of all elements of lens 1.


Quadrupole: Quadrupolar potential applied to lens 1 (+ on top/bottom, - on left/right: see above figure)

Right/left: Balance between top and bottom segments, steers beam vertically.
Top/Bottom: Balance between left and right segments, steers beam horizontally.

Use C:Program FilesJEOLMassCenterUtilitiesMS Monitor to see actual voltages on each segment (for
troubleshooting).
English Database:

Resides in c:Program FilesJEOLMass CenterApplicationsStringTable.mdb


Edit with: C:Program FilesJEOLMassCenterToolsMdbViewerMdbViewer.exe
Notes:

1.
The quadrupole resonating frequency is set with the configuration tool and the value is stored in the registry.
The correct value is set for each instrument and is listed in the parameter list in the final test document. If software is
reinstalled, this value will have to be set for each machine.
2.
Use an orifice 1 voltage of 50 V for tuning to give the strongest signal for the [M+H]+ on reserpine, but this
potential is high enough to induce fragmentation. For typical small molecules, an orifice 1 potential of approximately 30
V results in minimal fragmentation.
3.
At lower flow rates (5-10 microliters per minute), extend the needle to about 2 mm. If the desired flow rate is
lower than this, you will need to replace the ESI needle with the optional microspray needle or switch to the optional
nanoESI source.

c-10

Zoom up around m/z=609

c-11

Hard Copy on PCs monitor

d. Disconnect Syringe pump then connect LC with test-column. (For calibration and sensitivity test)
Test Column:(HP-LC column Zorbax Eclipse XDB-C18 2.1x50mm %-micron (HP P/N=960967-902))

29

Set flow rate to 0.2ml/min (Use LC controller or Agilent1100 controller on MS-Center)


Set Solvent B (Methanol) 100%

d-1. Agilent1100 controller operation.


Open Agilent1100 Menu
If you do not see Agilent 1100 status panel on
the desk top, open from : Instrument >
Instrument Status Panel > Agilent 1100
Click here

[Status Menu]

Click here to operation menu


[Binary Pump menu]
[General]

Flow rate setting


(set 0.1 or 0.2) in this case
Solvent %-rate setting
Set 100% now
(later , will set to 80 to 100% for
sensitivity test)
Different by sensitivity condition

30

[Details]
Set Solvent name etc

Unnecessary full in these menus for specification test


[Auto Sampler Menu]
[General]

Injection Volume [ul]


(Set 2.0 for sensitivity test)
Needle Wash [yes/no]
Use which vial for wash
(Check here and set 91 then
put in MeOH sample bottle at 91 location on Auto sampler)(for sensitivity test)
[Details]

31

Unnecessary change these menus for Test

[Detector Menu]
[General]

Detector wave setting


And Step
(Set 100-950 step=0.2
for detector work test)
Check pre-run

[Details]
(no need set any thing for test)

[Column Compartment]
[General]
(set 35 for test and check same as Left)

32

[Details]
(no need set any thing)

[Time Table Menu]

Parameter Window

Time Table view menu


a.

Time table

b.

Solvent graph

c.

Flow/Presser graph

d.

Detector Wave graph


Create/Add/Delete the time of tables. (Right click)

Set Stop time 10 to 30 minute for test. Different by pump flow and solvent A&B %.
(same as time on retention time in acquisition setting)
Another module are set to same as pump.(for test)

[ON/OFF Menu]
[Standard]
Check here is the module is ON.
All On/Off

33

[At Power on]


Set All on (for test)

Other : Instrument > Configuration..

Select each module.


Add/Delete
Recognize/Details

34

e.

Calibration

Sample :

TFA Na (Sodium Trufuoroacetate) (m/z > 3000) 100ppm with MeOH/H2O=70/30 %


PEG mix (m/z>2000) (10 to 100ppm with MeOH)

Use manual injection (Disconnect remote cable from manual-injector to LC)


LC flow 0.2ml/min (Set by Agilent1100 manager or manual set using LC module)
Tune condition of AccuTOF :
Same as Resolution test. (If necessary, adjust orifice 1,2 and ring lens for PEG.)
e-1 Acquire PEG(or TFA Na) that using single acquisition.
Acquire > Start Acquisition

e-2 Start Wizard


Next >
(If you have the acquisition setting file,
You can check and load here.)

e-3 Create new data name(acquisition Data)


If necessary, you can put in some
sample information and comment here.
Choose data saved folder.
(If there is no folder name,
you con create new folder at here now)
Next >
e-4 Select Acquisition Setting
This case choose
Specify new MS acquisition settings

Must blank Agilent1100 settings

35

NEXT >

e-5 Data Acquisition Setting


[m/z=range]
start = 100 end = 1000
[Stop Time]
2.00 min
[Spectrum Recording Interval]
0.5 s
[Data Sampling Interval]
0.5 ns
NEXT >
Never check Use Data Compression
e-6 Data Reduction
NEXT >
No need to do anything.

e-7 Choose real time display mode


TIC
Mass Spectrum
NEXT >

e-8 Finish then tart acquisition data.

e-9 If you have never had folder, you will get the message.

36

OK

e-10 Appear this window on the MS tune window.

Then Open Chromatgram , Spectrum Viewer and Start-Run Window automatically.

e-11 Click Start Run then inject PEG(or TFA Na) immediately. (5 to 10ml)
Check PEG (or TFA Na) that is detecting on spectrum monitor.
e-12 Wait Run time (2.0min)

or

If you want to stop in the run time, choose MassCenter Main Window then Stop button.
1

Or

37

Choose Quit Current Acquisition

OK

e-13. Open Calibration Maneger Menu


Instrument > Mass Calibration
(Mass Center Main Menu)

e-14.
Choose (1-1000)line
(This time)
Start Calibration Wizard

e-15. [Start Wizard]


NEXT >

38

e-16.

[Type]
Set file name that is going to make a calibration New File.
Or you have made file name, if you want update file. Choose from browse.
New File name (create here)
or choose Old file name
Choose 1000 or less
NEXT >

e-17. [How to create]


Choose Use Existing Data

Select you measured data name.


(See above k a-3, this case cal-1)
Set PEG(or TFA Na) detected time
NEXT >

e-18. [Calibration Parameters]


Select Mass Reference
(Used sample PEG
> Mass Reference PEG+Na+)
Used sample TFA Na
> Mass Reference YOKUDERUNA

Choose Auto
39

Curve Approximation 4 or 5
(See later explanation)
NEXT >

e-19. [Centroided spectrum]


Smoothing 3
Single Moving Average Method

Set Peak threshold level


(low and high mass side)

NEXT >

e-20. [Finish]
If you want to save this parameter setting,
check then create name to here.

Finish

e-21. See small window then open Calibration Window automatically

[See Next Page]

40

e-21. [Mass Calibration Viewer]


To be assigned Real data with in Reference date Automatically

Then, change parameter to minimum value here

Choose best setting for calibration value(above) to


minimum.

41

e-22. [Manual Peak Assignment]


1.

Remove Assignment
Calibrate > Remove All Assignment

2.

Choose you own peak


Choose some peak
(For example)

3.

Assign correct peak


Calibrate > Assign Peaks > Choose Correct Mass Number

4.

One or two more assign peaks by same way above 1 to 3 process.

5.

Re-Assignment
After Assignment 2 or 3 peaks, Choose Calibrate > Automatic Assignment

6.

Save New Calibrated name (or update calibrate)


File > Save As then Create New File

or
File > Update

42

7.

File > Exit then Return

Finish calibration

8.

Other : Change to other Calibration file on parameter


Choose you want to change calibration files line

Calibration > Register Data > Then choose you own calibration file.

43

f. Sensitivity Test (with continuity analysis function check)


Sample : Reserpine
Specification :

10pg s/n= 10 or better ( 5ppb 2ul inject 3 injections)

f-1 Set LC-MeOH=90% first to try to get data by this value then if necessary change 80 to 100%
f-2 If necessary re-tuning, connect syringe pump again, then do tuning best condition for Reserpine.
Then save tune condition.
(Use Reserpine 50 to 500 ppb)

(MS Tune Manager)


f-3 Make Acquisition Setting
1.

Instrument > MS Acquisition Setting

2.

Full in these setting menu.

1.

Click TOF

2.

Set Retention time


20 min
(Reserpine detect time)

3.

Data Acquisition
Start : 100
End : 1000
Spectrum Recording Interval : 0.5 s
Data Sampling Interval : 0.5 ns

4.

MS tune setting
Set MS tuning file that made as Resolution test or above a-1.
Browse > choose MS tune file

44

5.

Monitoring
Chromatogram : TIC
Mass Chromatogram : Start : 609.00
End : 609.50
Spectrum : Mass Spectrum

3.

Save Acquisition Setting name

f-4 Make LC setting for sensitivity test.


1.

Open Agilent 1100 manager

2.

Full in each parameters


[Binary Pump]
Flow rate : 0.2 ml/min
Solvent ratio MeOH : H2O
90 : 10
(Set B = 90%)

[Auto Sampler]
Injection Volume [ul] : 2 ul
Enable Automatic Needle Wash
Wash Vial Number : 81

45

[DAD detector]
Wavelength Range [nm] : 190-950
Step [nm] : 0.2
Auto Balance : Pre-run
(These are not necessary setting for
sensitive test.)

Enable analysis at UV Lamp off


Enable analysis at VIS Lamp off

[Column Compartment]
Left Heater : Control
Temperature : 35 c
Right Heater : Same as Left

[Time Table]
Stop Time and Post Time
Binary Pump : Unrestricted
Or
Measure Time
(same as Acq setting time)

[ON/OFF]
All ON
3.

Save LC condition file


Make your own name.

46

(For example: ReserpineLC)

f-5 Make Analysis List


1.

Open or Create Analysis List.


Open Analysis List from Project View
Choose Analysis List then Right click Open

If necessary do refresh viewer window

If you did not create Analysis List, create from File > New
Then create Analysis List by seeing

b Create New Project


at the 4. Start Mass Center Main software

f-6 Create Analysis List Parameter.


Right Click then choose Add Analysis

Number of Analyses
3

(this time)

47

Fill out each parameter (For example following)

(Your own, sample location, Acquisition Data, Acquisition Data Folder, MS Acquisition Conditions,
Agilent1100, each parameters.)
f-7 Start measurement
Analysis > Start

If you dont have any problem in parameter, you have this window.

Then start acquisition wizard.

NEXT >

Made Analysis list


Analyze Sample
Acquire MS Data
Acquire UV/VIS Data

48

Analysis line
NEXT >

Delete this process


Monitor Acquisition

NEXT >

Finish
Then LC Auto sampler pick sample up,
Start Acquisition automatically with
Agilent1100.

Stop acquisition after three times


measure samples.
or
Stop from Mass Center main window

49

g. Reduction
Open from Mass Center Main window
1. Choose Data Tug

2.

Select Data name then right click


Choose Open , to open data

After that Spectrum Viewer and


Chromatogram Viewer are opened automatically.
If you can not see, select Refresh

[Spectrum Viewer]
Maximum

50

[Chromatogram Viewer]

[Basic Operation on Viewer]

[Zoom up] Left drag for intensity, Mass Number or both


[Back to original size] Double click
[Peak label by manual] Alt + click peak (spectrum & chromatogram viewer)
[Delete Peak label]

Alt+ click labeled peak (Only labeled peak by manually)

[Peak label by auto]

Tool > Project Settings > Peak Label-tab


Then choose labeling menu
Check

> Label automatically


> Use threshold ..

then set %

Also there are some display setting in Project Settings


As necessary you can change this setting.

Printer header label etc..

51

[Create (Mass) Chromatogram]


1. Create from Chromatogram Viewer

Chromatogram > View > Chromatogram > open Chromatogram papameter

Select parameter

Set Mass Number range

Apply than Close


2. Create From Spectrum Viewer
Right button drag that you want to create mass number range.

52

[Create Mass Spectrum]


1.

Create From Spectrum Viewer

Spectrum > View > At >


Open Spectrum Parameter

Select Retention Time


or Scan number
Set you want to create Spectrum Range

[If you want to display subtraction


background Spectrum] Check here
and set Range.

Apply then Close

[Subtraction Spectrum]

Manual Peak m/z-number labeling


Alt+Click on Peak top you can move label using drag the label
Remove created label>>Alt+Click the Label
Background subtract from chromatogram viewer.
1. Shift + Right button drag (or click) at sample area (This operation does not create MS pectrum)
2. Then Ctrl + Right button Drag (or click) at background area.
3. Display Background subtracted Spectrum.

53

2.

Create from Chromatogram Viewer

Right button drag that you want to create R.T. number range.

[Show Mass Tables]


View > Table View Mode

Next or back page of Mass Tables

[Mass Spectrum Resolution Display]


1.

Zoom up Mass spectrum

2.

Spectrum > Mass Resolution

54

Apply then Close

[S/N calculate on chromatogram]


1.

Create TIC or Your own Mass Chromatogram

2.

Select Chromatogram > Signal to Noise Ratio

Select Noise Range

Apply then Close

55

[Save Disk Space]


If you need only to be shown spectrums and chromatograms, you can save just displayed data
then remove raw (huge) data for save disk space.
(No need whole raw data)
File > Save As > View Data

[Reduction Sensitive data]


1.

Open sensitive data.

2.

Make Mass Chromatogram (Mass range 609.00 to 609.50)

3.

Calculate S/N.

4.

3 times getting data then print out


(Make good window layout that looks like to same with factory data)(using Print Screen)

5.

If you cant reach to specifications, re-select noise range on Signal to Noise Ratio

6.

If you cant reach to specifications; re-tune , change LC setting or etc


Then re-measure data again.
Note: Slow LC rate better chromatogram.
Much water rate better peak shape chromatogram.

56

h.

Exact Mass Test

Sample : PEG600 with Reserpine ( Mix100ppb each) (inject from manual injector)
Spec. : 5ppm or less (Each mass number and Flight Tune voltage)
(At Flight Tube Voltage 7000v, 6999v, 7001v)
h-1. Tuning Best condition if necessary on PEG or Reserpine
(If necessary save tune setting)
Create tune file just only change Flight Tube Voltage on 6999v and 7001v

Change Flight Tube Volt.

Finally create three Tune files (Flight Tube Voltage at 6999v, 7000v and 7001v)
h-2. Measure Sample using Flight Tune Voltage at 6999v, 7000v, 7001v.
(Use Single acquisition will be useful)
Single acquisition >>>> see e Calibration page 34 process measurement.
h-3. Set(Open) 6999v(tune file) then measure. (Do not change other parameter)
h-4. Set(Open) 7000v(tune file) then measure. (Do not change other parameter)
h-5. Set(Open) 7001v(tune file) then measure. (Do not change other parameter)
h-6. Reduction for exact mass test.
1.

(Little be deferent of user exact mass analysis)

Open Data
Choose data name then Open

If necessary, do Refresh

57

2. Choose sample then make sample spectrum.


[Chromatogram Viewer]

Right drag then create (PEG & Reserpine) spectrum.


[Spectrum Viewer] (PEG with Peserpine spectrum)

3.

Internal Calibration
a.

choose M/Z Determine

Spectrum > M/Z Determine

b.

Set parameter

Usually only change Threshold level

Apply then Close

58

c.

Display Determine spectrum [looks bar type spectrum]

Then close parameter window


d.

Keep choosing Determine spectrum (to be surround by blue rectangular)

e.

Choose Change Mass Calibration

Tools > Change Mass Calibration

f.

Click Change on Mass Drift Compensation

Then choose Internal Mass Drift Compensation

59

g.

Start Internal Mass Drift Compensation Wizard

NEXT >

Choose Mass Reference Table or m/z Number


(This time choose File=Reserpine or 609.2811)

Automatic

Finish

h.

Open Intarnal Mass Drift Compensation Data window


Then select Reselpine Peak

(Change peak color to Red)

60

i.

Assign m/z number


Calibrate > Assign Peaks > 1: 609.2811

Then Reserpine Peak color change to blue


j.

Finish Mass Drift Calibration

File > Exit and Return

Confirm to be Changed Mass Drift Compensation


Internal Mass Drift Compensation Data

Then Close Change Mass Calibration window


Now m/a of Mass M/Z Determined Spectrum (Bar graph) has been changed
(If Sample with Internal reference are not same scan times,
reduction process will be little deference. See> AccuTOF Basic Operatin=other paper manual)
k.

Compare PEG Table between Theoretical and Measured number.


Open measured Table and Printout
View > Table View Mode

61

Table View Mode

File > Print

Theoretical PEG m/z number


437.23625
481.26246
525.28867
569.31488
613.34109
657.36730
701.39351
745.41972
789.44594
Calculate between measured m/z number and theoretical m/z number
(Measured m/z Theoretical m/z) divide theoretical m/z = error m/z
Then do test (reduction) on Flight Tube Voltage 6999v and 7001v by same way.
If not good value, try to re-tuning then calibration again.
Then measure again

62

Other
Function TEST

1. ESI NEG
1.

print out ESI pos tune condition.

2.

Change to ESI neg

On Tune Manager

Instrument > Ionization Mode > ESI Compare ESI pos tune file then set every voltage opposite priority (+/-)
except ion-source temperature, MCP voltage and Peak-frequency(Ion Guide volt.)
(same setting with POS)
3.

Connect syringe pump then monitor reserpine m/z=607 (M-H-)


Usually you can get resolution 5000 or good ( but, Negative mode not has a specification value)

4.

If necessary tuning then to save tune file.

2. Option (APCI, Nano-ESI, Dual-ESI etc.)


1.

Every Ion mode may be almost same setting with ESI mode except ION-source setting.

2.

APCI needle volt. = 4000 to 4500v, vaporizer temp 400 to 450C

3.

Nono-ESI pos needle volt=2000v, orifice 1 temp.=80C (for reserpice)


Then negative mode needle volt=1000v, orifice 1 temp=80C (for reserpine)

4.

Dual-ESI same as ESI (each needle)

[How to Back up data to DVD+RW device]


1)

Set DVD+RW media

2)

Format DVD
2-1. Open control panels then right-click on DVD drive.
2-2. Select Format (Format will be 2 minutes or less)

3)

Create a Folder using by windows explore

4)

You need to have Project folder in the DVD, when copy a data.
4-1.

If you do not have project create project by New project wizard


in Mass Center.
or
You need to copy whole project each.
(Can not copy only data, analysis list, tune file etc)

5)

Open Mass Center Data Manager software on desk top.

6)

Copy your own data to DVD using Mass Center Data Manager.
6-1.

Select data then click [copy]

6-2.

Select file of destination folder or project name.

6-3.

Paste (files will stat copying to

63

You might also like