Professional Documents
Culture Documents
2 7elect.wpd
Tasks
Objectives
General Information
Cations Anions
+ –
Na 142 mEq/L Cl 103 mEq/L
+ –
K 4 mEq/L HCO 3 27 mEq/L
2+ 2-
CA 5 mEq/L HPO4 2 mEq/L
2+ 2-
Mg 2 mEq/L SO4 1 mEq/L
others 1 mEq/L organic acids 5 mEq/L
Total 154 mEq/L proteins 16 mEq/L
Total 154 mEq/L
Note: Depending on the concentration of the specimen, a urine sample for Na/K analysis
may need to be diluted (often x 20 dilution is adequate).
Readout in mEq/L x dilution factor x # of liters in the 24 hour collection = mEq/24 hr.
Clinical Significance
1. Na+ , K+, Cl– ,& CO2 – Please refer to the chemistry lecture guide & textbook. for additional
information on these electrolytes and associated terms. Additional information on CO2 is
available in the pH/ABG lab in this manual.
2. Li – Although it can be found naturally in drinking water, the electrolyte / mineral lithium is
not normally found in serum in detectable amounts. Lithium carbonate salts are used to treat
and prevent manic depression. Lithium levels are monitored to avoid toxicity. Therapeutic
range = 0.5-1.5 mmol/L (mEq/L). Toxicity may occur at levels >1.5, while little therapeutic
value is obtained at low levels.
Methods of Determination
Solutions containing alkali ions diluted with an internal standard (usually lithium or
cesium) are aspirated into the flame. The atoms emit light at one or more distinctive
wavelength:
Na+ – 589 nm
K+ – 768 nm
Li+ – 761 nm
Cs+ – 852 nm
Since the emission energy varies with flame temperature and other conditions, there is
a need to monitor these variations with an internal standard. Lithium (or cesium)
solution is continuously being aspirated along with the sodium and potassium in the
patient sample, control sera, or Na/K standard. Changes in emission energy of the
sodium/potassium due to flame temperature variation, aspiration rate, etc. would also
occur to the emission energy of the lithium, therefore the concentrations of Na+ and K+
are determined by comparing their emission ratios with that of the Li internal standard.
Measures the amount of energy absorbed by Na and K rather than emitted. This technique
is more sensitive but less convenient than flame photometry.
2. The ground state atoms will absorb monochromatic light generated by a hollow
cathode tube containing the element being measured.
3. The amount of monochromatic light being absorbed (by the specimen element)
is proportional its concentration.
a. Titration methods
{ Whitehorn Titration Method – The chloride present in a protein-free filtrate is
precipitated as silver nitrate solution. The surplus silver nitrate is then titrated with
thiocyanate using ferric ammonium sulfate as the indicator.
{ Schales and Schales Titration Method – The sample is titrated with a standardized
mercuric nitrate solution in the presence of diphenylcarbazone as indicator.
Chloride ions (Cl– ) in the sample combine with mercuric ions (Hg ++ ) to form
mercuric chloride (HgCl2 ), a colorless, soluble, but only slightly ionized compound.
As long as chloride ions are present, mercuric ions preferentially react with them.
At the end point, when all chloride ions have been complexed, the excess mercuric
ions produce a violet color in the presence of diphenylcarbazone.
So, if the concentration of a solution is unknown, it can be found by measuring the volume of the
unknown solution that will read with a measured amount of a solution of known concentration
(called a standard solution). This process is known as titration.
The point at which equal concentrations of the standard and the unknown are present is called
the end point of the titration. Various means of detecting the end point are used. Sometimes the
formation of a precipitate indicates that the end point has been reached. A change in color of one
of the reacting solutions can also indicate the end point. The common method is through the use
of an indicator solution which is a third solution added in the titration procedure (in addition to the
standard and unknown).
The procedure for chloride determination based on the method of Schales and Schales
incorporates a mercuric titration using diphenylcarbazone as indicator.
{ Manual Method - The Sigma Diagnostics Chloride method is based upon the
quantitative displacement of thiocyanate by chloride from mercuric thiocyanate.
The liberated thiocyanate forms a red complex with ferric ions with an absorbance
Total CO2 content is a measure of the bicarbonate and the dissolved CO2 and carbonic acid.
The carbon dioxide is then absorbed into sodium hydroxide, the residual gases are
brought to the same constant volume and the second pressure measured. The drop
in pressure in mm can be multiplied by factors to bring the results to volumes percent
or mEq/L.
b. Van Slyke-Cullen Method – Acid is added to plasma and the CO2 produced in this
reaction is liberated by means of a partial vacuum and measured. Caprylic alcohol is
used to prevent foaming.
c. Colorimetric & Enzymatic methods of carbon dioxide measurement are available. Sigma
Diagnostics #131 is an enzymatic endpoint reaction measured in the ultraviolet
wavelength range.
4. Anion Gap
Although a true decreased anion gap rarely occurs, it has been noted in some multiple
myeloma patients. Possibility of instrument error should be eliminated before reporting
a decreased anion gap.
Procedures
1. Obtain chloride standard, controls, and unknowns and warm them to room temperature.
2. Thoroughly clean all four electrodes with silver polish, rinse with distilled water and buff with
tissue. Be certain no residue remains between the indicator electrodes at their common
mounting post. Avoid getting skin oils on the electrodes.
3. After cleaning, place a vial filled with @ 4 mL of Chloridometer Acid Reagent (or 4 mL of acid
solution and 4 drops of gelatin reagent) in the vial holder. Set the RANGE switch to LOW,
the TITRATION switch to AUTO and raise the holder so that the electrodes are immersed
and the stirrer begins.
4. If a reading does not appear after 30 seconds, re-rinse the electrodes and re-titrate using
a fresh vial. Do this until a reading is obtained.
6. Pipet 100 :L of the 100 mEq/L standard into the acid reagent. The addition of a sample with
a chloride concentration greater than 30 mEq/L will automatically reset the display and begin
the titration. If the titration endpoint displayed is 100 mEq/L ± 2 mEq, continue with
unknowns. If standard's concentration endpoint is unsatisfactory after several attempts,
contact instructor.
7. To analyze controls and unknowns, pipet 100 :L of sample into the acid reagent after the
endpoint is reached and the number is recorded, a new sample may be added. Analyze in
duplicate, and record results.
8. As the solution becomes whiter with precipitated silver chloride, the delay before a titration
begins will increase until a titration cannot be initiated. May need to replace solution after
10-15 analysis when using 100 microliter size sample.
9. Lower the vial, replace with fresh vial and, after raising the new vial, proceed as described
in step c above.
10. For each new vial, the titration procedures is the same. The electrodes only need to be
cleaned if there are visible deposits. When the instrument is not in use, leave the electrodes
immersed in distilled water and set the TITRATION switch to STANDBY or the RANGE
switch to POWER OFF.
Principle
Specimen
Serum or heparinized plasma. Promptly separate specimen from cells or clot to prevent shifts in
ionic equilibria. Chloride in serum or heparinized plasma is stable for 1 week refrigerated or
frozen.
Reagents
See product literature for contents of Sigma Diagnostics Chloride Reagent . DANGER: Reagent
may be fatal or cause blindness if swallowed. Vapor harmful. Combustible. Wear suitable
protective clothing, gloves and eye / face protection. In case of contact with eyes, flush with large
volume of water and seek medical advise. See product insert for additional information.
Deterioration: Chloride Reagent is not suitable for use its absorbance at 460 nm is more than
0.300.
Manual Procedure
2. Label tubes for REAGENT BLANK, STANDARD(S), CONTROL(S) and PATIENT SAMPLES.
3. Pipet 3.0 mL of Chloride Reagent into each tube and warm to assay temperature.
5. Mix gently by inversion and incubate for 5 minutes. (NOTE: The color is stable for at least
1 hour provided the temperature of the reaction mixture remains the same.)
Calculations
A SAMPLE - A BLANK
Chloride (mEq/L) of sample = ______________ X Concentration of Standard
A STANDA RD - A BLANK
Limitations
Procedure is linear from 50 to 130 mEq/L. If sample exceeds linearity, dilute with deionized water,
repeat assay and multiply result by dilution factor.
Expected Values
The expected range of chloride concentration was determined to be from 97 - 106 mEq/L. This
range is similar to reference values reported in the literature. It is strongly recommended that each
laboratory establish its own expected range characteristic for the population of the local area.
Principle
Phosphoenol pyruvate carboxylase (PEPC) catalyzes the reaction between phosphoenol pyruvate
and carbon dioxide (bicarbonate) to form oxalacetate and phosphate ion. In the second step of
the reaction, catalyzed by malate dehydrogenase (MDH), oxalacetate is reduced to malate with
simultaneous oxidation of equimolar amount of reduced nicotinamide adenine dinucleotide
(NADH) to (NAD). This results in a decrease in absorbance at 380 nm which is directly
proportional to the CO2 level.
Specimen
Serum or heparinized plasma. Ideally, venous blood should be collected and handled
anaerobically. After prompt separation from cells or clot, specimen should be kept tightly
stoppered. CO2 content of blood is stable for 1 hour when stored at 2-4°C under anaerobic
conditions.
Reagents
Reagents should be reconstituted according to directions. See product literature for contents of
Sigma CO2 REAGENT A, CO2 REAGENT B, and CO2 DILUENT. These reagents are for In Vitro
diagnostic use only. *DANGER: They contain sodium azide which is toxic if ingested any may
react with lead and copper plumbing for form highly explosive metal azides. On disposal flush with
large volume of water. Avoid contact and inhalation of CO2 REAGENT B.
Note: Avoid contamination of the reagents with CO2 . Do not blow into pipets, since breath
contains a high content of CO2 . Do not leave bottles open unnecessarily, since CO2
from air can contaminate the reagent.
Deterioration: Reagents are not suitable for use if the absorbance at 380 nm of freshly combined
reagent, is less than 1.000.
1. Prepare CO2 Sample Start Reagent according to instructions. (Each bottle does 5 tests.)
4. Pipet 2 mL of CO2 Sample Start Reagent into each tube and warm to assay temperature.
8. Subtract the absorbance of the STANDARD and SAMPLE from the absorbance of the
BLANK to obtain change in absorbance due to CO2 content.
Calculations
- A SAMPLE
A BLANK
Carbon Dioxide (mEq/L) of sample = ____________
X Concentration of Standard
A BLANK - A STANDA RD
Limitations
Expected Values
See product insert and information reported in the literature. It is strongly recommended that each
laboratory establish its own expected range characteristic for the population of the local area.
Name_________________________________
Date_________________________________
Electrolyte Worksheet
I. Chloride Sigma # 461
Wavelength _____________
Linearity _____________ Spectrophotometer Used ______________________
Identification Absorbance Abs test - Abs blank Concentration (units)
Blank
Standard
____________
Standard
____________
Control 1
____________
Control 2
____________
Quality Control
Your Results Controls’ range of expected results. In control? Yes /
No
Level 1
ID______________
Level
2ID_______________
Accepting Patient Results? Reason
1) 2) avg.
1) 2) avg.
1) 2) avg.
III. Carbon Dioxide
Calculations:
Quality Control
Your Results Controls’ range of expected In control? Yes /
results. No
Level 1 ID______________
Level 2ID_______________
Accepting Patient Results? Reason
C. Using the information provided, calculate the AG for the following patient.
D. Calculations
Name
Date_________________________________
Study Questions
Instructions: Legibly write your answers in the space provided. Unless otherwise indicated, each
question is worth one point. Using lecture notes, reading assignments and information presented
in this lab, answer the following questions.
2. Using flame photometry methodology, what is the relationship between the intensity of color
(wavelength) produced and the amount of element being measured?
4. What other compound can be used in place of lithium as an internal standard in flame
photometry?
6. Which plasma electrolyte, in high or low concentrations, may cause cardiac arrest? List two
causes of increased and decreased plasma concentrations of this electrolyte. (2 points)
8. List the four things that must be present to perform a titration procedure. (2 points)
11. Briefly describe the anion gap. Include what causes it and how it is used. (3 points)
The sodium, potassium, and CO2 are measured by ion selective electrodes. The chloride is measured by
the ferric thiocyanate method. Upon reviewing the results, the technician performs a quick written calculation
and decides that the specimen should be re-analyzed, believing that the results are not valid.
2. Provide the simple calculation that can be used to check the validity of these results.
5. What are some possible causes of an increased abnormal anion gap value?
7. How can the anion gap be used to evaluate the validity of electrolyte analysis?
A 58 year old woman with a history of hypertension arrived in the ER complaining of nausea, oliguria, and
vomiting. The following are electrolyte results obtained.