Professional Documents
Culture Documents
zen ceiling
Quick Guide Volume 1
Software VD10 and higher
Caution
21 CFR - Federal law restricts this device to sale
by or on the order of a physician.
Usability
BEFORE USING THIS SYSTEM
This system should be used only by employees
adequately trained in the use of this equipment.
Before using this system, the operator should be
thoroughly acquainted with the instructions for
use and safety recommendations provided in
this manual.
Failure to follow the instructions for use and
safety recommendations provided in this manual can cause serious injury to the patient, to the
operator or to other persons.
Original language
Version
This Quick Guide applies to system/software version VD10 and higher.
AXA4-100.622.24.02.02
Content Overview
System Overview (Exam Room).........8
System Operation ...........................16
Examination ...................................44
Postprocessing (Exam Room) ..........84
Quantification (Exam Room) .........102
IVUSmap.......................................130
Rotational Angiography, 3D ..........142
Troubleshooting ...........................174
AXA4-100.622.24.02.02
Peripheral Angiography.................166
Safety information:
Note
These texts contain tips, more detailed information, and important notes.
Caution/Warning
Cause
Possible consequences
Precautions or remedies
AXA4-100.622.24.02.02
Table of Contents
System Overview (Exam Room) .....................8
Overview Artis ceiling ...........................................8
AXA4-100.622.24.02.02
Examination ...............................................44
IVUSmap...................................................130
IVUSmap examination workflow .......................131
Step 1: Start IVUSmap and acquire scene...........133
Step 2: Mark the vessel segment of interest.......135
Step 3: Pullback ................................................137
Step 4: Review ..................................................141
3D ....................................................................146
3D DR ...............................................................148
3D DSA .............................................................150
DynaPBV Neuro and DynaPBV Body ...................154
3D CARD ...........................................................158
DR-DYNAVISION................................................162
DYNAVISION .....................................................164
Peripheral Angiography.............................166
PERISTEPPING ...................................................170
PERIVISION .......................................................172
Troubleshooting .......................................174
System messages..............................................174
Emergency operation........................................176
No unit movement possible! .............................178
Buffer full! - Memory full! ..................................178
Stopping movements........................................180
Patient rescue ...................................................182
Restarting .........................................................186
AXA4-100.622.24.02.02
Introduction
We welcome you as a user of an Artis Q/Q.zen
ceiling acquisition system from Siemens.
This Quick Guide is designed to give you quick
access to information on the basic operating
steps and the most important functions of your
system during routine examination work.
Note
This Quick Guide is only valid in combination
with the Instructions for Use and the safety
information contained therein.
Content Overview
The following topics are covered by this Quick
Guide:
Registering and positioning patients
Procedure for standard examinations
Image postprocessing and quantification in
Procedure for advanced examinations:
AXA4-100.622.24.02.02
(3)
(2)
(1)
(5)
(4)
AXA4-100.622.24.02.02
(3)
(2)
(4)
(1)
(5)
(6)
AXA4-100.622.24.02.02
(7)
(5)
(4)
(3)
(2)
10
Head Side
Right Side
AXA4-100.622.24.02.02
(2)
(2)
(2)
Tabletop neuro
Tabletop wide
Tabletop narrow
Tabletop long
Surgery table
Trumpf surgery table TruSystem 7500: Please refer to the Operator Manual of the manufacturer.
AXA4-100.622.24.02.02
(3)
11
(5)
(4)
(3)
(2)
(6)
12
AXA4-100.622.24.02.02
(4)
(1)
(2)
(3)
(5)
(5)
AXA4-100.622.24.02.02
13
Example of DCS 3
14
AXA4-100.622.24.02.02
ages.
The Assist screen shows position data, sys-
Display by pressing the icon on the touchscreen or by clicking the icon on the Examination task card in the control room.
AXA4-100.622.24.02.02
15
System Operation
Control consoles
(1)
(2)
(3)
(4)
Example
The control modules are equipped with a joystick and several keys.
16
module
on the front side of the emergency STOP
AXA4-100.622.24.02.02
System Operation
Triggering STOP in case of an emergency
You can stop unit movements and radiation at
any time, e.g. in the following cases:
If a malfunction of the system occurs.
If a unit movement causes an emergency sit-
immediately.
- All system drives are shut down and movements are stopped immediately. Movement can only be resumed when STOP is
canceled.
- Radiation is interrupted, fluoroscopy and
acquisition are stopped.
However, radiation can be released again if
you press the fluoroscopy/exposure release
pedal again, even without canceling STOP.
- Injections1 are interrupted, but can be continued without canceling STOP.
Canceling STOP
Pull the red emergency STOP button.
AXA4-100.622.24.02.02
STOP
17
floating tabletop
(with or without servo assistance)
additional control
console
18
AXA4-100.622.24.02.02
System Operation
Table movements
Note
The following also applies to the Trumpf surgery
table TruSystem 7500.
For operating of the additional control module
and additional movements, refer to the Operator Manual of the manufacturer.
Raising / lowering the table
Press the key for the required movement and
of the tabletop.
- The LED lights up.
- The tabletop can only be moved longitudinally when the brakes are released.
To enable tabletop movement in any direction
once again:
Press the key again.
AXA4-100.622.24.02.02
19
The movement of the tabletop can be configured by Siemens service in two ways:
Continuous holding
Note
The Fluoro pedal of the footswitch can also be
configured by Siemens Service to release the tabletop brakes (dual function).
Table with tilt
20
If a tilting patient table is installed, the tabletop with the patient can be tilted by 15 to
the head-up position (reverse Trendelenburg) or head-down position (Trendelenburg).
Lateral tilt (longitudinal axis)
If the table with lateral tilt is installed, the tabletop with the patient can be tilted sideways
by another 15.
Press this key to activate lateral tilt.
Note
If ISO tilting has been configured, the patient angle is maintained when the Isocenter key is activated.
Then, the patient table follows the C-arm angulation.
AXA4-100.622.24.02.02
System Operation
Moving the tabletop
Continuous holding has been configured
Press the panning knob down and hold it
down.
Move the tabletop (floating movement).
sideways.
- The tabletop follows the movement.
Tilting the tabletop
Caution
Patient table tilted
Patient slips from the table to the ground.
bletop.
Secure the patient sufficiently with the ac-
cessories.
Always keep the mattress fixed on the table-
AXA4-100.622.24.02.02
21
!
22
(View from
the rear)
Note
The movement direction depends on the console orientation. 24.
The speed of the movements depends on how
far you deflect the joystick: maximum speed
with maximum tilt.
Combined C-arm movements are possible when
the operating element is deflected diagonally,
e.g. 45.
AXA4-100.622.24.02.02
System Operation
C-arm movements
Take care to avoid possible collisions.
The correct console orientation is set. 24
Caution
Contrast media or blood may seep into the system.
Risk of system malfunction and danger of infection
AXA4-100.622.24.02.02
23
Orientation key
24
(Default position)
Control console at the foot end of the table-
top
Control console to the left of the tabletop
Control console at the head end of the table-
top
If your system is equipped with a laser positioner, it can be switched on and off by pressing the
orientation key for less than 3 s.
AXA4-100.622.24.02.02
System Operation
Stand movements
Take care to avoid possible collisions.
Positioning the C-arm longitudinally
Press one or both side buttons on the stand/
Right Side
counterclockwise
AXA4-100.622.24.02.02
Head Side
25
You can also save three most frequently used positions that you require for examinations.
Note
Use the Patient Transfer position for cardio pulmonary resuscitation (CPR).
Programmed stand, C-arm, and table movements are initiated as follows:
1. Select system or programmed position
26
Note
Information about the movement sequence is
shown on the Assist screen.
The speed of the movements depends on
how far you deflect the stand/C-arm joystick: maximum speed with maximum deflection.
If you release the stand/C-arm joystick or
AXA4-100.622.24.02.02
System Operation
Programmed movements
Selecting the target position
Press the required position key.
flect it to the backward until the target position has been reached.
Or, if hands free movement is configured:
Press the configured hands free pedal on the
footswitch.
AXA4-100.622.24.02.02
27
What is stored?
All data important for imaging geometry are
stored:
Stand and C-arm position
Table height and tilt (configurable by Sie-
mens Service)
SID, zoom stage, collimation and filter posi-
tions
Note
Programmed positions can be protected against
deletion or overwriting by Siemens Service.
Such positions are used for 3D examinations, for
example.
28
AXA4-100.622.24.02.02
System Operation
Storing programmed positions
Move the unit to the required position.
Set the primary collimator and the filter dia-
phragms.
Press the Store key.
stand/C-arm joystick.
Storing system positions with shortcut (direct
positions I, II, III)
Move the unit to the required position.
Press the store key.
AXA4-100.622.24.02.02
29
Note
Automap is not possible for images of other Siemens systems, other manufacturers or other
modalities such as CT/MR).
Automap is not possible when a reference image of plane B from a biplane system is selected
on a single plane system.
30
AXA4-100.622.24.02.02
System Operation
Automap
Moving to the system position of the reference image
Select the required reference image.
Press the System Positions key.
AXA4-100.622.24.02.02
31
FD
20x20
FD
Membrane keys for C-arm and FD movements
26x30 +
FD
30x40
Version 1
Version 2
32
tions)
C-arm orbital movement (RAO/
LAO angulations)
Longitudinal C-arm movement
scape)
FD lifting and lowering - setting
the SID
AXA4-100.622.24.02.02
System Operation
Image size and format
FD lift / Setting the SID
The image receptors are mounted on slides so
that the distance from the X-ray source to the
image receptor plane (SID) can be changed.
Take care to avoid possible collisions.
Press the rocker switch on the back of the
stand/C-arm joystick.
- Upper switch: increase SID
- Lower switch: decrease SID
Or
Press the membrane key on the flat detector.
AXA4-100.622.24.02.02
Note
33
(1)
(1)
(2)
(2)
34
Card collimator
Angio collimator
no finger filter
Additional keys
rectangular diaphragm
rectangular diaphragm
AXA4-100.622.24.02.02
System Operation
Keys on the collimator control module
With the keys on the collimator control module
you can select different collimator and filter
functions and set the image input format:
Open collimator leaves completely and re-
AXA4-100.622.24.02.02
Collimator/FD rotation
35
Zoom stage
36
AXA4-100.622.24.02.02
FD 30x40
48 cm
42 cm
32 cm
22 cm
16 cm
11 cm
System Operation
Manual override: Rotating the image clockwise
Press the top key and keep it pressed.
module.
- The zoom stage is reduced (= larger input
field) or enlarged (= smaller input field) by
one step.
- The size of the active input field is shown
on the Assist screen.
AXA4-100.622.24.02.02
37
Note
The behavior of the collimation when changing
the zoom stage can be configured in different
manners.
Please ask Siemens Service.
Note
38
AXA4-100.622.24.02.02
System Operation
Collimation and filters
Rectangular collimation
You can set the rectangular diaphragms with the
lower joystick on the collimator control module:
Deflect the collimator joystick in the required
direction.
- The position of the collimator leaves is
shown in the live/LIH image.
open vertically
open horizontally
close horizontally
close vertically
AXA4-100.622.24.02.02
39
phragm filters)
for example, for DSA aortic arch or peripheral
DSA of both legs
Cardiac
Angio
40
The wedge filters are set with the upper two joysticks on the Angio collimator control module.
The finger filter is also set with the upper two
joysticks on the Angio collimator control module.
The LED on the wedge/finger filter key on the
Angio collimator control module indicates the
selected filter diaphragm type.
AXA4-100.622.24.02.02
System Operation
Setting the wedge filters
The LED for the wedge filters is lit.1
Deflect the joystick in the required direction.
Right joystick/filter
rotate clockwise
move to the
right
move to the
left
rotate counterclockwise
move to the
left
rotate counterclockwise
move to the
right
rotate clockwise
Left joystick
Right joystick
rotate clockwise
move to the
right
move to the
left
rotate counterclockwise
move to the
left
rotate counterclockwise
move to the
right
rotate clockwise
AXA4-100.622.24.02.02
41
(2)
(1) Touchscreen
(2) Mouse joystick
Mouse joystick
(1) (2) (3)
42
Example
Note
You can display a tool tip for every button by
pressing the question mark button first.
AXA4-100.622.24.02.02
System Operation
Selecting a task card
the touchscreen.
- The task card you have selected automatically moves to the foreground.
Selecting a function
Press the button.
For easier operation, the possible joystick functions are displayed on the live
screen in the examination room. If a joystick function is selected, it is indicated
at the bottom of the control area.
AXA4-100.622.24.02.02
43
Examination
Default patient
An emergency patient is automatically registered each time the system is restarted. It allows
you to perform fluoroscopy and acquisition immediately after switching on the system as soon
as the imaging system is ready.
44
Note
Patient name, patient ID, date of birth and sex
are used for unique identification of a patient in
the databases and on archive media.
Note
If you want to register the whole patient list of
the day, you can preregister the patient.
Click Preregister after entering the patient
data.
- The patient is put in the Scheduler.
AXA4-100.622.24.02.02
Examination
Registering a patient
Taking over patient data from the RIS
The Scheduler will automatically be updated
with registration information at fixed intervals,
as configured in your system.
Press the Patient Browser (Num .) key on
tient (if known) and select the Patient position1 from the list (data shown in boldface
type are mandatory entries).
Registering a patient manually
Press the Patient Registration (Num 0) key
Study list.
Select the Patient position from the list.1
AXA4-100.622.24.02.02
45
Note
When the system is in BYPASS FLUORO mode,
or during radiation, the patient cannot be registered for examination.
In this case, click Preregister.
Caution
The system assigns patient orientation labels to
the images depending on patient position selected during patient registration.
Danger of treatment of the wrong side.
Do not acquire images with the patient posi-
tioned different from the patient position selected during patient registration.
46
Caution
Data collision between Sensis and Artis.
Patient registration on the Artis causes inconsistent patient data.
Register patient only by Sensis.
Note
If you are using the Sensis Information System,
you must select an examination type from the
Study selection list.
AXA4-100.622.24.02.02
Examination
Registering the patient and starting examination
Click Exam.
rect.
- The patient is registered.
on Sensis.
Enter the data (data shown in boldface type
AXA4-100.622.24.02.02
The Examination task card is placed in the foreground and you can start the examination immediately.
47
Note
During transfer and positioning of the patient,
you can block motorized movements in order to
prevent unintentional unit movements.
Note
48
AXA4-100.622.24.02.02
Examination
Preparing the patient and
equipment
Positioning the system for patient transfer
Press the System Positions key.
appropriate accessories.
Attach and position the required radiation
protection accessories.
AXA4-100.622.24.02.02
flect it to the backward until the target position has been reached.
49
Note
During examinations, always make sure that
there are no unwanted objects in the beam
path.
Note
50
Note
The ECG cables could become tangled with or be
broken by moving parts, e.g. the X-ray tube, during C-arm angulation.
Pay attention to the ECG cables during unit
movements.
AXA4-100.622.24.02.02
Examination
Setting the examination position
Press the System Positions key.
flect it to the backward until the target position has been reached.
Attaching the ECG
For ECG monitoring of the patient or ECG-gated
fluoroscopy and acquisition:
such a way that they do not appear in the
beam path.
Check the connection of the gating signal
AXA4-100.622.24.02.02
51
Isocenter
To ensure that you do not have to readjust the
table height during an examination with different angulations and to ensure optimum evaluation results, the organ of the patient, e.g. heart,
or region of interest must be positioned in the
isocenter of the C-arm.
Note
Depending on mattress thickness and patient
thickness, isocentric movements are restricted.
(3)
(2)
(4)
52
(1)
AXA4-100.622.24.02.02
Examination
Setting the isocenter
Positioning the patient approximately (without fluoroscopy)
Place the C-arm in frontal position, if it is not
terest is centered.
If necessary, activate the radiation field limi-
AXA4-100.622.24.02.02
53
Patient position
Caution
The system assigns patient orientation labels to
the images depending on patient position selected during patient registration.
Danger of treatment of the wrong side.
Do not acquire images with the patient posi-
tioned different from the patient position selected during patient registration.
Caution
Misusing the patient position to flip an image
Risk of misinterpretation and incorrect diagnosis
Always enter the patient position correctly.
54
Caution
Image flipped
Confusion between up/down/left/right can
cause incorrect diagnosis!
The examiner is responsible for using the functions and interpreting the images correctly and
the consequences resulting from it.
Restore the original image orientation at the
Note
If you want to flip images that have already been
acquired, you can do that on the PostProc task
card in the control room.
AXA4-100.622.24.02.02
Examination
Checking parameters
Checking/changing the patient position
Settings task card
AXA4-100.622.24.02.02
55
56
Note
Press the Acquisition Overrides button, to select a different scene length or measuring field.
AXA4-100.622.24.02.02
Examination
Selecting the acquisition program/exam set
Exam task card
Press this icon next to the currently selected
exam set.
- A selection of available exam sets and
assigned acquisition programs appears.
buttons.
Press the button of the required new acquisi-
tion program.
- The new acquisition program is activated
and the dialog is closed.
Or
exam sets.
Press the button of the required new exam
set.
AXA4-100.622.24.02.02
Example
57
Example
58
Example
AXA4-100.622.24.02.02
Examination
Changing the acquisition program in the current exam set
Exam task card
quisition program.
- The available acquisition programs of the
exam set are displayed.
Press the required acquisition program.
Or (if configured)
AXA4-100.622.24.02.02
59
Note
The fluoroscopy/roadmap programs are always
available in three versions, which usually differ
with regard to their dose.
Example
60
Example
Note
The fluoroscopy programs automatically change
to Roadmap programs when Roadmap is selected.
AXA4-100.622.24.02.02
Examination
buttons.
Press the required series description text.
roscopy/roadmap program.
Press the required fluoroscopy/roadmap pro-
gram.
Or (if configured)
AXA4-100.622.24.02.02
61
Note
You can store the last fluoroscopy (or roadmap)
as a scene.
Note
62
Note
Press this icon, if the message box for the fluoroscopy signal appears, or if you want to reset
the signal before intervening, for example.
Caution
If any radiation indicator lights up although you
have not pressed a button:
Danger due to radiation
Press the emergency STOP button.
Notify Customer Service.
AXA4-100.622.24.02.02
Examination
Caution
If poor image quality results from the set parameters:
Risk of poor image quality and unnecessary
exposure to radiation
Stop radiation immediately and check the pa-
Caution
Malfunction in the detector cooling system and
display of the message No X-Ray available in
... min.
Risk that the examination cannot be completed
When the countdown has elapsed, radiation will
no more be possible.
Complete the current examination within the
Fluoroscopy
Press the FLUORO pedal of the footswitch.
AXA4-100.622.24.02.02
63
Note
You can store any image currently displayed,
even during fluoroscopy, as a new image (Store
Monitor), or as a reference image.
You can also press the middle button of the
mouse joystick to store a reference image.
Note
64
If the system has been configured with two reference screens, you can store an image as a reference image and display it on the Reference or
Assist screen.
Click the icon to toggle the active screen.
Note
Using the CLEARstent image function (Image
task card) you can generate special reference
images from any scene or fluoroscopy scene acquired natively.
The stent/balloon marker should be clearly visible in the scene.
If the stent/balloon marker is not visible in the
current image, CLEARstent will not be possible.
AXA4-100.622.24.02.02
Examination
Acquisition
Caution
Acquiring thin objects
Risk of unnecessary radiation exposure due
to incorrect dose
Always perform fluoroscopy before acquisi-
tion.
Using the handswitch
Press the acquisition button (two stage) of
the handswitch.
Using the footswitch
Press the acquisition pedal of the foot-
switch.
The acquisitions (scenes) are stored as set in the
acquisition program and appear on the screen.
AXA4-100.622.24.02.02
65
In order to have a fast access to special acquisition program, an Alternative Acquisition can
be assigned to a footswitch pedal.
An alternative acquisition program can be used,
for example, to provide a low dose acquisition
program instead of fluoroscopy at critical situations where the fluoro dose is limited by the
maximal skin dose.
The configuration to enable/disable the alternative acquisition is done by assigning of one of
the footswitch pedals to this function.
Example
66
In the Exam Set and Program Editor, the alternative acquisition program is located between
the 4 acquisition programs and the 3 fluoro programs. If any acquisition progam is visible there,
then it is an alternative acquisition program assigned to the footswitch pedal.
Caution
Fluoro release foot pedal mixed up with the alternative acquisition release pedal.
Increased irradiation by performing acquisition
Be careful to use the correct foot pedal.
Check the labels and the position of the foot
pedals.
AXA4-100.622.24.02.02
Examination
Alternative acquisition
Alternative acquisition has been configured
to a footswitch pedal.
Select an exam set containing an alternative
acquisition program.
If an alternative acquisition program is available,
an indicator will be present near the acquisition
symbol on the touchscreen, on the Live screen
and on the Examination task card.
Perform fluoroscopy.
Note
For alternative acquisitions, always the measuring fields of the regular acquisition program are
used. The forecast parameters are displayed for
the regular acquisition program and not for the
alternative acquisition program.
Alternative acquisition pedal is used to start acquisition with automatic injection, although
power injectors are triggered exclusively by pedal "2".
Exposure to radiation without clinical benefit
If automatic injection is required, use the
handswitch, or the regular acquisition pedal (pedal "2") on the footswich to release Xray.
Start acquisition by means of the footswitch
AXA4-100.622.24.02.02
Caution
67
Note
To avoid motion artifacts, the patient should neither move during roadmap nor should the patient table nor the system be moved.
Roadmap phases
Phase 1
Phase 2
68
ed.
Indicators
Injection
OPAC
Review
AXA4-100.622.24.02.02
Examination
Roadmap
A Roadmap program is selected which has
quired.
AXA4-100.622.24.02.02
pressed.
- After fluoroscopy has begun a normal fluoroscopy image is displayed briefly, then the
display automatically switches to subtracted images.
- An injection indicator is displayed on the
live screen.
69
70
During Roadmap phase 3, anatomical background can be added to the subtraction image.
In the examination room, the anatomical background can be adjusted using the mouse joystick.
Vessel/catheter contrast with Roadmap
During Roadmap phase 3, but not during X-ray,
the vessel/catheter contrast can be adjusted for
Roadmap images.
The initial values for Vessel Contrast and Catheter Contrast are defined in the Roadmap program.
Note
Whenever a roadmap mask is restored with the
previous mask function, vessel/catheter contrast
will be reset to the default values from the roadmap program.
AXA4-100.622.24.02.02
Examination
Resetting roadmap
If motion artifacts arise, roadmap can be reset
(new mask):
Actuate the Reset Roadmap pedal (if config-
ured - example).
Or
the console.
Press this icon.
AXA4-100.622.24.02.02
71
Phase 3
72
AXA4-100.622.24.02.02
Examination
Advanced Roadmap using a DSA image
A Roadmap program is selected which has
movement
joystick and hold it in this position until the Carm position is reached.
Or press the hands free foot pedal.
When the position is reached the following message appears: Automatic run: Position reached
Collimation and zoom stage are set corresponding to the parameters of the reference image.
Performing fluoroscopy
Actuate the FLUORO pedal and keep it
pressed.
After fluoroscopy has begun a normal fluoroscopy image is displayed.
Then, the display automatically switches to a
subtracted image.
Performing Roadmap phase 3 (subtraction)
Press the fluoroscopy switch as often as re-
quired.
You will see a fluoroscopy image from which
the last filled image of phase 2 is subtracted.
AXA4-100.622.24.02.02
73
74
Note
The latest roadmap masks are kept during the
whole study and will be deleted when a new patient is registered.
AXA4-100.622.24.02.02
Examination
Previous mask Roadmap
Select roadmap and perform a roadmap
phase 1,2,3.
Move the C-arm and perform another road-
map dialog.
The following message appears:
Automatic run: Activate stand movement
Press the stand joystick down, deflect the
AXA4-100.622.24.02.02
75
CLEARstent
With the CLEARstent imaging functions, you
can emphasize fine structures, e.g. inflated
stents. There are the following possibilities for
CLEARstent:
Use the CLEARstent reference image func-
76
Original image
CLEARstent image
AXA4-100.622.24.02.02
Examination
CLEARstent Dynamic acquisition workflow
A CLEARstent Dynamic acquisition is performed in the following sequence:
1. Prepare the patient, stand, C-arm and table.
Check the injector.
2. Perform a brief fluoroscopy: Press fluoro pedal of footswitch.
Check/set SID, zoom stage, collimation, filters and measuring fields.
3. Select a CLEARstent Dynamic acquisition
program.
Ready for CLEARstent acquisition is displayed.
4. Start acquisition: Press the exposure release
button or pedal and keep it pressed.
- Acquisition phase 1 takes place.
- At the beginning of phase 2, an injection
indicator is displayed on the Live screen.
- The message Injection is displayed.
a CLEARstent Dynamic scene will be created, stored, and displayed on the live screen.
AXA4-100.622.24.02.02
77
Hold)
This mode is the default display on the live
screen when no other mode is selected.
The last fluoroscopic/roadmap image is displayed on the live screen after completion of
fluoroscopy. Reference images are displayed
on the reference screen.
Overlay Ref(erence) can be selected or de-
selected.
The (inverted) reference image is overlaid
into the live/LIH fluoroscopic image.
Note
Overlay Ref is only possible if the reference image was acquired with the same modality (AX,
no CT, MR), the same input format/zoom stage,
the same patient position, and the same portrait/landscape position with FD30x40.
78
During roadmap or DSA, the selected reference image is displayed on the Assist screen.
Note
The availability of display modes depends on the
screen configuration of your system.
AXA4-100.622.24.02.02
Examination
Display mode
Selecting the display mode
Settings task card
Image icon.
- The display mode is changed to Overlay
Reference, i.e. the overlay image is faded
in the image on the Live screen.
Setting the degree of Overlay Reference/
Fading in or out the reference image
Deflect the mouse joystick.
AXA4-100.622.24.02.02
79
Review modes
LIH (Last Image Hold): The last image of flou-
but, unlike Loop, the review direction is reversed after the last image (ping-pong effect).
Depending on scene type, different review
modes are available:
Available review
modes
80
LIH
Fluoro/Roadx
map scenes
DR/DSA/PERI
scenes
CARD scenes
Rotational
scenes
Report images
MFH
Loop
x
x
x
x
x
x
Note
You can change the Review Mode by selecting
View > Review Mode in the main menu in the
control room.
AXA4-100.622.24.02.02
Examination
Dynamic scene replay
Stopping the loop
Deflect the mouse joystick briefly.
Single step
Deflect the mouse joystick briefly.
AXA4-100.622.24.02.02
81
Representative
A scene is represented by a specific image in the
overview display.
Native: It is the center image for scenes ac-
quired natively.
It is an image with maximum fill (Max Fill im-
Note
If you want to use an image other than the automatically chosen image as scene representative,
you can replace it.
82
Note
The scene overview shows single images
(frames) of a scene.
To display the scene overview while the
scene directory is open, press the right button of the mouse joystick.
AXA4-100.622.24.02.02
Examination
Scene directory
Scenes
Press this icon to open the Image Player.
Reference images
AXA4-100.622.24.02.02
83
max.
bright value
highest absorption
stored value
range
lowest absorption
dark value
84
non-subtracted:
subtracted:
WC
Brightness (subtracted)
Window width (native)
WB
WW
Contrast (subtracted)
WC
Note
After acquisition, the image is displayed in accordance with the window values defined in the
exam set.
After loading again, the image is displayed using
the window values stored with the image.
AXA4-100.622.24.02.02
Window width/
Contrast -
Window width/
Contrast +
Window center/
Brightness -
To close windowing:
Press the right button of the mouse joystick.
AXA4-100.622.24.02.02
85
2480
1024
1024
FD
Screen
86
1024
1024
1024
1024
1920
2480
FD
AXA4-100.622.24.02.02
Screen
Control room
Examination room
AXA4-100.622.24.02.02
87
DSA postprocessing
Press this icon.
During a DSA exposure, the mask (= image without contrast medium) is set automatically. It
normally occurs after the second frame, but no
later than after 2 s.
88
Note
All pixelshift values are reset when you set a new
mask.
Note
In the case of DYNAVISION or 3D scenes, only
one mask image is usually available per acquisition position (exception: washout scene).
The mask therefore cannot be redefined for
DYNAVISION or 3D scenes.
AXA4-100.622.24.02.02
Native display
DSA functions
Native
Press this icon.
Subtracted
Press the icon again.
AXA4-100.622.24.02.02
89
Note
In the case of DYNAVISION or 3D scenes, only
one fill image and one mask image is usually
available per acquisition position.
Pixelshift may need to be carried out individually
for each acquisition position.
Use automatic pixelshift to shift a range of mask
and fill frames.
Motion artifacts
As a result of patient or unit movement, the images used for subtraction might not coincide
completely. If it happens, the anatomical background is not removed completely (motion artifacts).
90
Pixelshift variants
Automatic pixelshift:
You mark a region of interest (ROI). The computer automatically calculates the best result
for this region.
Manual pixelshift
AXA4-100.622.24.02.02
Starting pixelshift
Press this icon.
- The mask is shifted in such a way that subtraction is optimized for the marked
region.
AXA4-100.622.24.02.02
Automatic pixelshift
91
Scope of action
Each pixelshift action (automatic or manual) always applies to the current image and all images
up to the end of the scene.
Movement
Pixelshift
action
Movement
Pixelshift
action
Movement
Scene
Pixelshift
action
Note
If you scroll back and then perform pixelshift,
the pixelshift values you set (further back in the
scene) may be overwritten.
92
Note
Depending on the image content, the Flex image calculation may take a few seconds. It also
applies to scrolling.
Note
At the end of the scene, pixelshift is automatically deselected.
AXA4-100.622.24.02.02
Undoing pixelshift
Press this icon.
Terminating pixelshift
Press the right button of the mouse joystick.
Flexible pixelshift
Press this icon.
The mask is divided into squares like a chessboard. The optimum shift amount is determined
and applied automatically to each square.
Saving the Flex image
A Flex image can be stored using the Store
Monitor function.
You can also store the flexible pixelshifted image as a reference image.
You can also copy the flexible pixelshifted image
to the filmsheet.
Press the required icon.
AXA4-100.622.24.02.02
93
94
Note
The opacity function cannot be applied to DYNAVISION or 3D scenes.
Note
You can also store the image with maximum
contrast medium filling as a new image (Store
Monitor), or copy it to the filmsheet.
AXA4-100.622.24.02.02
DSA functions
Scroll to the starting image.
Bolus start
At the image in which the contrast agent bolus
is visible for the first time:
Press this icon.
Scrolling
The scene can be scrolled forward or up to one
image backward.
Scroll the scene image by image:
Deflect the mouse joystick briefly.
stick.
The image with maximum contrast medium filling is calculated, displayed and stored as a reference image.
AXA4-100.622.24.02.02
Max/Min Opac.
95
96
AXA4-100.622.24.02.02
be placed.
Drawing lines or arrows
Press the Line or Arrow button.
Click the image to fix the starting point (= ar-
Drawing polygons
Press the Polygon Line button.
Click in the image to set the starting point,
Drawing circles
Press the Circle button.
Click the detail of interest in the image and
AXA4-100.622.24.02.02
97
Distance calibration
To perform a measurement, e.g. distance, the
image must be calibrated.
Note
Calibrations performed in Quant are valid also
for measurements done in PostProc and vice
versa.
When you start measuring on a non-calibrated
image, the system automatically suggests a calibration method depending on the image angle.
The image angle is the geometric sum of the image's LAO/RAO angle and CRAN/CAUD angle:
Image angle = acos ( cos (LAO/RAO angle) * cos
(CRAN/CAUD angle) )
Example: With RAO = 14 and CRAN = 10, the
image angle is approximately 17.
98
AXA4-100.622.24.02.02
Caution
Incorrect calibration
Inaccurate results possible
The examiner is responsible for applying the correct calibration method.
Perform calibration carefully.
meric buttons.
AXA4-100.622.24.02.02
99
Note
The default TOD value is stated in the calibration
dialog. It can be changed by Siemens Service.
Note
The calibration object should be larger than the
object to be measured.
100
AXA4-100.622.24.02.02
AXA4-100.622.24.02.02
101
Note
Pay attention to the messages at the bottom of
the image.
102
Note
Catheter calibration is not recommended for
LVA. For QVA, catheter calibration should be
used with FD zoom stages only (not in full format).
The catheter calibration software is validated for
a catheter size of 7 French.
Using catheter calibration with catheter sizes of
5 French or smaller is not recommended because accuracy could be reduced.
AXA4-100.622.24.02.02
Catheter calibration
- The contour detection algorithm automatically detects the edges of the catheter.
Now enter the actual size of the catheter used
for calibration (in French units).
E.g.
Or
Press the OK button to accept.
AXA4-100.622.24.02.02
103
Note
The smaller the calibration object, the greater
the calibration error.
104
Note
You can move the circle by dragging the circle
center to the required position.
You can change the size of the circle by dragging
the circle line larger or smaller.
Check, if center remains at the same posi-
tion.
AXA4-100.622.24.02.02
Or
Press the OK button to accept.
AXA4-100.622.24.02.02
105
the shape of the coronary arteries in three dimensions based on two 2-dimensional images.
Note
106
Before you can perform a 2D quantitative vascular analysis, you must calibrate the image. 99
103
Note
The selected image plays an important role in accurate and reproducible contour detection.
We recommend selecting an image close to the
diastolic phase of the cardiac cycle.
You can display the results on the screen, film/
print and send them to another network node.
AXA4-100.622.24.02.02
AXA4-100.622.24.02.02
QCA or
QVA
107
Note
The center line must be marked in the direction
of blood flow as it influences the calculation of
the extent of stenosis.
Note
To achieve reproducible analysis results, it is advisable to start and end the segment at a characteristic, conspicuous point. Side branches are
suitable for this purpose.
108
Note
In the exam room, the contour is automatically
smoothed.
Note
You can film, print, or send a report in the same
way as normal images.
However, results of an analysis on the Quant
task card are only transferred to the film sheet if
the Report is stored.
AXA4-100.622.24.02.02
click.
Marking a vessel segment with bifurcation
Define the proximal starting point of the
AXA4-100.622.24.02.02
109
IZ3D
Two projections are sufficient to create an accurate 3D reconstruction in the case that:
110
terest.
There is at least a 30 difference between the
AXA4-100.622.24.02.02
rectory icon.
In the scene directory, scenes are marked for
IZ3D:
: already used
: already used and analyzed
Select the second image.
AXA4-100.622.24.02.02
111
Epipolarity lines are used to indicate the approximate region of the display where the vessel being analyzed within a 2D image. The lines
indicate the uppermost and lowermost boundaries of the screen region within which the vessel should fall.
112
AXA4-100.622.24.02.02
image.
Vessel contour detection 107
AXA4-100.622.24.02.02
113
Note
You can rotate, pan and window the 3D vessel
model via pressing the left, right, or middle
mouse joystick button respectively, and deflecting the mouse joystick.
The Optimal Projection indicates a system position that meets the following requirements:
For 3D vessel models that do not include a bi-
furcation, the Optimal Projection is the projection that shows the least foreshortening
of the analyzed segment.
For 3D vessel models that include a bifurca-
tion, the Optimal Projection is the projection that displays the greatest angle between
the branches of the bifurcation.
114
The Optimal Projection is automatically displayed when the 3D vessel model has been created.
You can select one of the following items:
A: ... or A: ... B: ... : angulation of the single
3D vessel model
AXA4-100.622.24.02.02
joystick and hold it in this position until the Carm position is reached.
Or press the hands free foot pedal.
AXA4-100.622.24.02.02
115
Note
116
AXA4-100.622.24.02.02
AXA4-100.622.24.02.02
117
Display in 3D model
118
AXA4-100.622.24.02.02
AXA4-100.622.24.02.02
119
With LVA (Left Ventricle Analysis), you can determine the ejection fraction and the wall motion of the left cardiac ventricle.
The LVA software is designed for analyses of the
left ventricle of the heart in a 30 RAO projection.
Note
For a single plane image, calibration is optional.
If LVA analysis is performed on an uncalibrated
image then the set of results will be limited.
Make sure that the calibration image and the
ventricle scene have the same geometric acquisition parameters.
Note
120
AXA4-100.622.24.02.02
As soon as you move the cursor inside the image, it changes shape to a pencil.
AXA4-100.622.24.02.02
121
LVA workflow
A suitable scene is loaded and displayed in
Depending on configuration (by Siemens Service), an LVA analysis is performed with or without automatic detection of the ED (end diastolic)
frame and the ES (end systolic) frame. Auto ED/
ES detection is only possible if the scene contains ECG data.
122
AXA4-100.622.24.02.02
AXA4-100.622.24.02.02
123
Note
124
AXA4-100.622.24.02.02
Or
Press the scroll back/forward button.
if necessary.
- Automatic contour detection 127
- Defining a contour manually 129
if necessary.
- Automatic contour detection 127
- Defining a contour manually 129
AXA4-100.622.24.02.02
Or
125
Note
The ventricle contours serve as the basis for calculating the results of the analysis. There are
two ways to define the contours for the ED and
ES: either manually or by marking the valve
plane and the apex followed by automatic contour detection.
126
Note
In the exam room, the contour is automatically
smoothed with Soft.
AXA4-100.622.24.02.02
valve.
- The points marking the valve plane are
indicated by squares.
Now define the position of the apex:
The mouse pointer has shape of a diamond.
Click the apex cordis.
AXA4-100.622.24.02.02
127
Note
At the end of the manual contour definition, the
valve plane will be indicated with a line connecting the upper and lower point of the aortic valve.
These points are indicated as small squares.
Note
128
Note
Pressing this icon you can reject the current
analysis and delete analysis graphics.
Note
You can film, print, or send a report in the same
way as normal images.
However, results of an analysis on the Quant
task card are only transferred to the film sheet if
the Report is stored.
AXA4-100.622.24.02.02
age.
Press this icon.
a double click.
The contour is drawn.
AXA4-100.622.24.02.02
129
IVUSmap
IVUS imaging
Intravascular Ultrasound (IVUS) is a catheterbased system that allows physicians to acquire
images of diseased vessels from inside the
artery.
IVUS provides measurements of lumen and vessel size, plaque area and volume, and the location of key anatomical landmarks.
IVUSmap
For obtaining a coregistration of an angiographic X-ray image and IVUS images, the IVUSmap
feature is available. The IVUSmap workflow
consists of acquiring and selecting a dedicated
X-ray image to be used for coregistration, determining the vessel centerline, and performing of
the pullback.
Review
130
IVUS system
Currently, the Volcano s5i IVUS system is supported.
Catheter types
Currently, the following IVUS catheters are supported:
Volcano Eagle Eye Gold catheter
Volcano Eagle Eye Platinum catheter
AXA4-100.622.24.02.02
IVUSmap
IVUSmap examination workflow
The IVUSmap feature can only be successfully
utilized when:
IVUSmap is properly licensed and config-
ured.
A supported IVUS system is powered on,
AXA4-100.622.24.02.02
3. Start IVUSmap
131
Note
132
Note
The user interface of the IVUS system and the
Artis IVUSmap application are both displayed in
the exam room.
If a shared display is configured, it is automatically switched between both screens, otherwise
both screens are visible simultaneously.
AXA4-100.622.24.02.02
IVUSmap
Step 1: Start IVUSmap and
acquire scene
Position the IVUS catheter in the vessel of in-
terest.
Perform the Ring Down on the IVUS system.
Then advance the IVUS catheter beyond the
lesion.
Press this button on the touchscreen.
frame by frame
Deflect the mouse joystick briefly.
- Left: one frame backward
- Right: one frame forward
Or
AXA4-100.622.24.02.02
133
Caution
User inadvertently marks the wrong vessel during Angio/IVUS registration.
It may result in mis-registration of the IVUS
and Angio images.
134
Note
In this step, make sure that you mark the two
points in the right sequence: first the marker at
the tip of the guiding catheter proximal vessel
location, then the distal vessel location.
Note
You may undo and repeat your marking.
AXA4-100.622.24.02.02
IVUSmap
Step 2: Mark the vessel
segment of interest
Press the Mark Vessel button.
ing catheter.
- The indicated position is marked as the
guide catheter position.
- The following message is displayed:
Select distal vessel location
Click in the image to mark the distal part of
the vessel.
- The indicated position is marked as the distal vessel position.
The following message is displayed: Select
next work step or correct centerline
sel centerline.
Deselect the Correct Centerline function.
AXA4-100.622.24.02.02
135
Caution
Movement of guiding catheter during transducer pullback.
It may result in mis-registration of the IVUS
and Angio images.
The user is responsible for ensuring the IVUS
136
Note
When selecting the Pullback workflow step and
the current stand position/table position differs
from the stand and table position of the ECG gated angio scene used for registration, the following message is displayed: Return system to
registration position (deflect joystick)
Deflect the stand joystick in any direction and
AXA4-100.622.24.02.02
IVUSmap
Step 3: Pullback
Note
The pullback should be stopped before the IVUS
catheter is pulled into the guiding catheter.
Press the Pullback button.
back device.
Press the Record button on the IVUS system.
pedal for (alternative) acquisition or fluoroscopy and keep it pressed until the end of
the pullback.
Stop X-ray acquisition: Release the foot-
switch pedal.
- The following message is displayed: Waiting for the End of Pullback ...
Press the Stop button on the IVUS system.
AXA4-100.622.24.02.02
137
138
AXA4-100.622.24.02.02
IVUSmap
The following message is displayed: Check if
dots are within brackets for reliable result.
Then select Review.
If the green dots are within the brackets:
Press the Review button.
If the green dots are not within the brackets,
When IVUS images are displayed, you can already start placing bookmarks.
AXA4-100.622.24.02.02
139
Note
If the registration between the X-ray angiographic image and the IVUS images does not
match exactly, you can correct it in the control
room.
140
AXA4-100.622.24.02.02
IVUSmap
Step 4: Review
In the last workflow step, you can review the
coregistrated angiographic image and the IVUS
images.
The Review workflow step is automatically selected.
Loop, single stepping
AXA4-100.622.24.02.02
141
Rotational Angiography, 3D
3D
Special 3D acquisition modes are available
which enable the generation of image sets suitable for 3D reconstruction on the syngo
Workplace. 146
Using syngo InSpace 3D, DynaCT and DynaCT
Cardiac the acquired data sets can be reconstructed to 3D images on the syngo Workplace.
You will find further information for reconstruction in the syngo Workplace Quick Guide
(Volume 2).
Note
3D examinations can be performed only in the
calibrated system positions.
DR-DYNAVISION
142
AXA4-100.622.24.02.02
Rotational Angiography, 3D
Preparations
Preparing the patient
Immobilize the patient as well as possible to
tion.
Inform the patient about the coming exami-
nation procedure.
Preparing the stand and table
Position the C-arm in Left Side or Right Side
position.
beam path.
Preparing the injection
Puncture and insert the catheter.
Connect the contrast medium injector and
program it.
Perform fluoroscopy and collimate.
Position the catheter in the target area.
Automatic injection
Set the X-ray delay on the injector.
Switch the injector to Armed.
Manual injection
Deselect synchronization.
AXA4-100.622.24.02.02
beam path.
143
Injection mode
Contrast medium can either be applied with the
help of an automatic injector or manually.
In general, injection with the injector produces
better results as the flow of the contrast medium
can be kept constant over the total duration of
the injection.
If the injector is set to Armed it will be trig-
gered.
It is recommended to set a delay time (X-ray
delay) of 0.5 to 1 s on the injector. With it, an
optimum contrast medium filling is achieved
already with the first projection.
The X-ray delay which is programmed at the
injector is used to delay the begin of the injection phase. During this delay time, there is
no image acquisition.
144
tem assumes that a manual injection is performed with no contrast medium applied and
starts the injection phase immediately.
For manual injection, it is recommended to
set an X-ray delay of 0 s.
Caution
Injector not connected (correctly) or not set to
armed (especially with PERIVISION, DYNAVISION or 3D examinations)
Risk of X-ray radiation applied without diagnostic use
Pay attention to the messages on the Assist
AXA4-100.622.24.02.02
Rotational Angiography, 3D
Automatic/manual control
The rotational examination sequence depends
on the Dyna Control or 3D Control setting in
the acquisition program:
If Auto is set in the acquisition program, then
Caution
Patient moves hand or arm into the way of the
moving C-arm.
Patient injury
It is the responsibility of the operator to en-
AXA4-100.622.24.02.02
the C-arm (and start of the acquisition) manually with the multifunctional button (blue
button on the hand switch).
- The injection must also be carried out manually.
145
3D
There are special acquisition modes intended to
create Angiographic Computed Tomography
(ACT) 3D images for various applications (vessels and bones).
The images will be acquired in equidistant angular spacing. Rotation is carried out with predetermined (calibrated) rotational ranges.
3D DR: Like in DR-DYNAVISION, images are
are acquired in mask run and fill run and subtraction is performed. 150
3D CARD: Images are acquired in several ac-
146
AXA4-100.622.24.02.02
Rotational Angiography, 3D
Centering the target volume in the isocenter
(3D)
When a 3D acquisition program was selected,
there are two possibilities:
1. The isocenter has already been set before selection of the 3D acquisition program. The region of interest is positioned in the isocenter.
- The LED on the Isocenter key is lit.
You can start immediately.
2. The isocenter has not been set. The region of
interest is not yet positioned in the isocenter.
- The LED on the Isocenter key is flashing.
The isocenter assistant helps you positioning
the region of interest in the isocenter.
beam path.
The following message is shown:
3D ISO: Lateral, activate stand movement
Press down the stand joystick and deflect it.
beam path.
AXA4-100.622.24.02.02
If the region of interest is positioned in the isocenter, press the isocenter key in order to skip
the isocenter assistant.
147
3D DR
3D DR examination sequence
Preparations
3D DR test phase
(6)
148
(4)
(7)
(5)
(3)
(1)
AXA4-100.622.24.02.02
Rotational Angiography, 3D
3D DR workflow
1. Prepare the patient, stand, C-arm and table.
Check the injector, if used.
2. Select a 3D DR acquisition program.
3. Position the region of interest in the isocenter.
- You can start immediately if the LED on the
Isocenter key is lit.
- Follow the isocenter assistant if the LED
on the Isocenter key is flashing.
4. Move C-arm to start position:
Press down the stand joystick and deflect it.
5. Perform the 3D DR test run to make sure that
no collision will occur.
6. Perform a brief fluoroscopy for automatic
setting of acquisition parameters: Press fluoro pedal of footswitch.
7. Automatic or manual injection if contrast
medium is used.
Start injector if manual injection is used.
8. Perform the 3D DR acquisition run:
- Start acquisition: Press the exposure
release button or pedal and keep it
pressed. Rotational run will start automatically, if Auto is programmed. Injection will
also be performed automatically if the
injector is set to Armed.
- Start acquisition. Then start rotational run
in time: Press the blue multifunctional
button of the handswitch, if Manual is
programmed. 145
AXA4-100.622.24.02.02
149
3D DSA
3D DSA examination sequence
Preparations
150
AXA4-100.622.24.02.02
Rotational Angiography, 3D
(2)
(4)
(6)
(8)
(5)
(9)
(7)
(3)
(1)
(4) Mask run with rotating C-arm for image acquisition without contrast medium
(5) Reversal in the start position of the test run
(6) Return run
(7) Injection at reversal point
(8) Fill run with rotating C-arm for image acquisition with contrast medium
(9) End of acquisition
AXA4-100.622.24.02.02
151
3D DSA workflow
A 3D DSA acquisition is performed in the following sequence:
1. Prepare the patient, stand, C-arm and table.
Check the injector, if used.
2. Select a 3D DSA acquisition program.
3. Position the region of interest in the isocenter.
- You can start immediately if the LED on the
Isocenter key is lit.
- Follow the isocenter assistant if the LED
on the Isocenter key is flashing.
4. Move C-arm to start position:
Press down the stand joystick and deflect it.
152
AXA4-100.622.24.02.02
Rotational Angiography, 3D
Ready for 3D is displayed.
7. Perform the 3D DSA mask run:
Start acquisition: Press the exposure release
button or pedal and keep it pressed.
Movement is performed automatically. Mask
images are acquired.
8. The return run takes place automatically.
9. Automatic or manual injection.
Start injector if manual injection is used.
11. Let go the exposure release when all rotational runs are complete.
(Acquisition is stopped automatically.)
12. Transfer of image data to the syngo
Workplace:
- Automatically immediately after the acquisition
- Manually as long as data is available on the
Artis
13. Slice reconstruction using InSpace Reconstruction on syngo Workplace.
14. Visualization using the InSpace task card on
syngo Workplace.
AXA4-100.622.24.02.02
153
from data acquired in a normal 3D DSA examination on Artis using a special acquisition
program.
3D DSA examination sequence 150
DynaPBV Neuro examination sequence
(2)
(6)
(7)
(9)
154
(11)
(8)
(1)
(5)
(3)
(4)
AXA4-100.622.24.02.02
Rotational Angiography, 3D
DynaPBV Neuro workflow
with manual injection (i.v. injection)
Preparations
Prepare the injector: Add the contrast and sa-
the touchscreen.
If C-arm is not in AP position
System message: 3D ISO: Frontal, activate
stand movement
Deflect stand joystick until C-arm reaches AP
position.
Adjust table to position region of interest in
isocenter.
eral position.
Adjust table height to position region of in-
terest in isocenter.
System message: 3D Start Position: Activate
stand movement
Deflect stand joystick until C-arm reaches
start position.
Test phase
System message: 3D Test Phase: Auto, activate stand movement
Deflect stand joystick until test run is fin-
ished.
System message: Test Phase: Start fluoro to
complete test phase
Briefly press fluoro pedal on footswitch.
AXA4-100.622.24.02.02
155
Injection
We recommend injecting intravenously about
2 s before starting the mask run. Typically, the
following injection may be applied:
60 ml @ 4-5 ml/s and after that 60 ml @ 4-
5 ml/s NaCl
Or (without dual-head injector):
80 ml @ 4-5 ml/s without NaCl
156
Return run
native
8s
5s
Bolus watching
10-15 s, max. 30 s
AXA4-100.622.24.02.02
Bolus reaches
sinus sagittalis
Fill run
8s
Rotational Angiography, 3D
Acquisition
System message: Ready for 3D
Start the i.v. injection: Press the acquisition
button or pedal.
Keep the acquisition button or pedal pressed
until the end of the fill run.
Mask phase
Press the blue multifunctional button of the
AXA4-100.622.24.02.02
Fill phase
157
3D CARD
3D CARD examination sequence
Preparations
3D CARD test phase
(2)
(4)
(6)
(8)
158
(5)
(9)
(1)
(7)
(3)
AXA4-100.622.24.02.02
Rotational Angiography, 3D
(5) Start of acquisition 3D run 2
(6) 3D acquisition run 2
(7) Start of 3D acquisition run 3
(8) 3D acquisition run 3
(9) End position
3D CARD workflow
A 3D CARD acquisition is performed in the following sequence:
1. Prepare the patient, stand, C-arm and table.
Check the injector, if used.
Check the ECG, if used.
3. Position the region of interest in the isocenter.
- You can start immediately if the LED on the
Isocenter key is lit.
- Follow the isocenter assistant if the LED
on the Isocenter key is flashing.
4. Move C-arm to start position:
Press down the stand joystick and deflect it.
5. Perform the 3D CARD test run to make sure
that no collision will occur.
6. Perform a brief fluoroscopy for automatic
setting of acquisition parameters: Press fluoro pedal of footswitch.
AXA4-100.622.24.02.02
159
9. Let go the exposure release when all rotational runs are complete.
(Acquisition is stopped automatically.)
160
AXA4-100.622.24.02.02
Rotational Angiography, 3D
3D acquisition programs
The acquisition programs for 3D acquisitions are
named as follows:
<acquisition time> <acquisition mode> (<body
part>) (<specification>)
<specification>
Care
AVG
Description
Bolus time in seconds
For native reconstructions
For subtracted reconstructions
DynaCT
Cardiac without ECG gating
Cardiac with ECG gating
Neuro parenchymal blood volume
Liver parenchymal blood volume
For examination of the head
For examination of other body
parts, e.g. abdomen
Low dose acquisition program
For Aortic ValveGuide
AXA4-100.622.24.02.02
Name component
<acquisition time>
e.g. 5 s
<acquisition mode>
DR
DSA
DCT
Card
Card ECG
Neuro-PBV
Liver-PBV
<body part>
Head
Body
161
DR-DYNAVISION
DR-DYNAVISION examination sequence
(2)
(6)
(4)
(7)
(5)
(3)
(1)
162
AXA4-100.622.24.02.02
Rotational Angiography, 3D
DR-DYNAVISION workflow
1. Prepare the patient, stand, C-arm and table.
Position the region of interest in the isocenter.
Check the injector, if used.
2. Perform a brief fluoroscopy in the start position to check the examination region: Press
fluoro pedal of footswitch.
3. Select a DR-DYNAVISION acquisition
program.
4. Perform the DR-DYNAVISION test run to
make sure that no collision will occur.
5. Perform a brief fluoroscopy for automatic
setting of acquisition parameters: Press fluoro pedal of footswitch.
AXA4-100.622.24.02.02
163
DYNAVISION
DYNAVISION examination sequence
(2)
(4)
(6)
(8)
(5)
(9)
(7)
(3)
(1)
164
AXA4-100.622.24.02.02
Rotational Angiography, 3D
DYNAVISION workflow
1. Prepare the patient, stand, C-arm and table.
Position the region of interest in the isocenter.
Check the injector, if used.
2. Perform a brief fluoroscopy in the start position to check the examination region: Press
fluoro pedal of footswitch.
3. Select a DYNAVISION acquisition program.
4. Perform the DYNAVISION test run to make
sure that no collision will occur.
5. Perform a brief fluoroscopy for automatic
setting of acquisition parameters: Press fluoro pedal of footswitch.
AXA4-100.622.24.02.02
165
Peripheral Angiography
PERISTEPPING
With a motor-driven stepping device, a linear examination region (e.g. legs, spine, intestine) can
be acquired step by step (in native display) with
PERISTEPPING (= DR Stepping). 170
PERIVISION
PERIVISION is the stepping method introduced
by Siemens with digital online subtraction. With
an injection, an angiography is created in stepping technique in subtracted display. 172
Moved component
If a motorized stepping device is installed, peripheral angiography can be performed.
A component is moved linearly according to the
system type:
The C-arm in case of Artis ceiling, .
Running direction
166
The running direction determined in the acquisition program can be selected according to the
blood flow direction. It always refers to the registered patient position.
for examining the leg arteries: head to feet
for examining the leg veins: feet to head
AXA4-100.622.24.02.02
Peripheral Angiography
Preparations
Transparency compensation
Transparency compensation is available to avoid
direct radiation and to compensate for the transparency differences in the pelvis-leg region.
Place the transparency compensation on the
feet.
Inform the patient about the coming exami-
nation procedure.
AXA4-100.622.24.02.02
Either run the fastening belt around the tabletop or pull it through underneath the positioning mattress and fasten it.
167
Automatic/manual control
The examination sequence depends on the
Stepping Mode setting in the acquisition program:
If Auto is set in the acquisition program, then
Caution
168
AXA4-100.622.24.02.02
Peripheral Angiography
Preparing the stand and table
Position the C-arm in Left Side or Right Side
position.
Adjust the SID.
image receptor.
Set the LAO/RAO angulation (max. 30).
program it.
Perform fluoroscopy and collimate.
(recommendation: 0 s).
Switch the injector to Armed.
Manual injection
Deselect synchronization.
AXA4-100.622.24.02.02
Automatic injection
169
PERISTEPPING
PERISTEPPING examination sequence
Preparations
Test phase
You determine the start position, all the acquisition positions and the end position.
You also check that the movement range is
free of obstacles and hazards.
Return phase
Injection if contrast medium is used
Fill phase
(2)
170
(1)
(2)
(4)
(6)
(6)
(3)
(5)
(7)
AXA4-100.622.24.02.02
Peripheral Angiography
PERISTEPPING workflow
1. Prepare the patient, stand, C-arm and table.
Check the injector, if used.
2. Perform a brief fluoroscopy in the start
position: Press fluoro pedal of footswitch.
Check/set SID, table height, collimation, filters and measuring fields.
3. Select a PERISTEPPING acquisition program.
AXA4-100.622.24.02.02
171
PERIVISION
PERIVISION examination sequence
Preparations
Test phase
You determine the start position, all the acquisition positions and the end position.
You also check that the movement range is
free of obstacles and hazards.
Mask phase
172
(2)
(1)
(2)
(4)
(3)
(6)
(7)
(4)
(5)
(6)
AXA4-100.622.24.02.02
Peripheral Angiography
PERIVISION workflow
1. Prepare the patient, stand, C-arm and table.
Check the injector, if used.
2. Perform a brief fluoroscopy in the start
position: Press fluoro pedal of footswitch.
Check/set SID, table height, collimation, filters and measuring fields.
3. Select a PERIVISION acquisition program.
AXA4-100.622.24.02.02
173
Troubleshooting
System messages
There are various types of system messages:
Feedback, e.g. DYNA test phase, manual
Instructions, e.g. Automatic run: Move the
table lift up
Error messages
Message lines
The line at the bottom of the images shows
shows tips on system operation on the foreground and positive feedback/states in the
background.
174
Note
Operating restrictions that only the Service Center can remedy are marked with "... SC ...".
Example: "NO XRAY: call SC"
AXA4-100.622.24.02.02
Troubleshooting
System messages are displayed:
on the Assist screen / message bar (2 lines at
bottom)
- Stand/table messages
- System status messages
on the monitor of the system console
various resources.
AXA4-100.622.24.02.02
175
Emergency operation
The Artis Q/Q.zen ceiling systems have been developed in such a way that the highest possible
system availability is guaranteed. In other
words, the examination is obstructed as little as
possible if, for instance, an error occurs due to a
fault.
If one component fails, the other continue to
work.
Therefore, the system has several operation
states, so that a failure of one component does
affect the whole system as less as possible, e.g.:
1. If an image acquisition system fails, the system goes into Bypass fluoroscopy
2. If the image evaluation system fails (console
in the control room), the system goes into
Backup mode
176
After switch-on, all components are poweredup. Some of them are faster, some are slower.
So various operation states pass:
Off -
Bypass
fluoroscopy No x-ray Continuous
fluoroscopy
only
Note
If images have been acquired in backup mode,
e.g. immediately after power-on, they must be
transferred from the buffer to the memory.
Then the following message is displayed:
Syncing database, ... scenes left
AXA4-100.622.24.02.02
Troubleshooting
Bypass fluoroscopy
In various system states, e.g. during switch-on
or switch-off or if a technical fault occurs or
when no user is logged-in, only bypass, that is,
continuous fluoroscopy without the imaging
system without LIH, is possible. It is not possible
to release acquisitions.
You can recognize it as follows:
The fluoroscopy image is only visible in the
examination room.
The last fluoroscopy image (LIH, Last Image
Backup mode
displayed.
Review is possible in backup mode. Other scenes
can be loaded via Scene + / Scene - but only
CARD, DR and DSA scenes are loaded.
AXA4-100.622.24.02.02
177
has responded.
Movements blocked?
Check the Block Movement function.
Patient?
178
STOP
button again .
AXA4-100.622.24.02.02
Troubleshooting
Message
Buffer for < 4 Acq Runs
No Buffer for the next Acq Run
No acq: No Buffer for the next Acq
Message
Free space in the memory is not sufficient (hard
disk is full).
Acquisition is blocked, fluoroscopy is still possible.
Delete (archived!) patients (if possible, pa-
AXA4-100.622.24.02.02
179
Stopping movements
All motorized movements stop when:
1. the operating element is released,
2. an end position is reached,
3. a basic position is reached (can be configured
by Siemens Service),
4. a collision sensor responds,
5. the collision computer reduces the speed until stop,
6. an emergency STOP button is pressed.
Note
In certain situations, the speed of unit movements is reduced for safety reasons.
180
AXA4-100.622.24.02.02
Troubleshooting
Resuming movement after collision (override)
Case 4.
Caution
Safety override active
Higher risk of collision
Risk of mechanical or personal damage
Execute unit movements with special care
with x.
Case 5.
Case 6.
AXA4-100.622.24.02.02
possible directions.
181
Patient rescue
E.g. in case of power failure, motoric system
movements are not available. However, you can
move parts of the system manually.
It is only possible to remove the patient man-
182
If necessary, you can move the tabletop of patient tables with manual tabletop movement
manually in the longitudinal and transverse direction.
(This is not possible on patient tables with motor-assisted tabletop movement)
Pull/push vigorously to overcome the braking
force.
AXA4-100.622.24.02.02
Troubleshooting
In order to be able to quickly response and move
the ceiling suspended C-arm away from the patient an additional clasp at the ceiling suspension has been installed.
(1)
(2)
Note
(3)
Once the belt has been released (3) the C-
arm can be easily pushed away from the patient towards the end of the table.
AXA4-100.622.24.02.02
183
Caution
Patient table tilted
Risk of patient sliding from the table
Secure the patient sufficiently with shoulder
In difficult cases, you may have to rescue the patient along with the mattress.
Lift the mattress at the foot end to unfasten
184
AXA4-100.622.24.02.02
Troubleshooting
After power recovery:
Place the belt back in the ceiling rail and
move the C-arm carefully back to the previous position until the belt is tightened again.
Close the hook again (1) and switch on the
system.
Please call the Siemens Service to ensure
AXA4-100.622.24.02.02
185
Restarting
If a message ... Restart imaging system appears, you must terminate the imaging system
application and then restart it.
Restart needs some time.
There are three possibilities for restarting:
1. Restart Application 187
In this case, only the console in the control
room is restarted.
Acquisition and fluoroscopy are still possible.
Recommended if there is a problem in the
control room only.
186
AXA4-100.622.24.02.02
Troubleshooting
Restart application or restart system
Select Options > End Session... from the
main menu.
- The End Session dialog box is displayed.
Click Restart Application...
AXA4-100.622.24.02.02
187
Caution
Resetting the image evaluation system
Risk of data loss
Only ever press the RESET button when the image evaluation is not ready.
Do not press the RESET button during an ac-
quisition.
Wait until all drive activity has come to an end
Note
188
AXA4-100.622.24.02.02
Troubleshooting
RESET image evaluation system
half a second.
- The image evaluation system (console in
the control room) is restarted and should
be ready again after a few minutes.
Switching off manually and power on again
tem console for more than 4 s and then release the button.
Wait until everything is off.
Wait for about 30 s.
Press the power-on pushbutton on the sys-
tem console for more than 0.5 s and then release the button.
AXA4-100.622.24.02.02
Switch-off the system, only if the imaging system can no longer be terminated by software.
189
Index
-
190
A
Acquisition 65
Acquisition frame rate
changing 59
Acquisition program
changing 57, 59
Acquisition programs
3D 161
Advanced Roadmap
performing 73
phases 72
Alternative acquisition 67
Alternative Acquisition
Program 66
Anatomical background 95
with Roadmap 70
Angio collimator
control module 34
Angiographic Computed
Tomography (ACT) 146
Angulations 23
Annotations 97
Aortic valve
in LVA 127
Apex
in LVA 127
Application
restart 187
Arrows
drawing 97
Artis zee ceiling
overview 8
Assist screen 15
Automap 31
Automatic contour detection
in LVA 127
Automatic control
in DYNA, 3D 145
in PERI 168
Automatic isocenter
calibration 101
B
Backup mode
emergency operation 177
Bifurcation
in QCA/QVA/IZ3D 109
Blocking radiation 48
Blocking unit movements 48
Brakes
releasing brakes for table 20
Brightness 84
Buffer 177
full! 178
Button
emergency STOP 16
Bypass fluoroscopy
display 177
emergency operation 177
C
Calibration
exam room 103
using the table-object
distance 99
Card collimator
control module 34
Cardio Pulmonary Resuscitation
(CPR) 26
C-arm
joystick 22
movements 10
moving 23
positioning
longitudinally 25
AXA4-100.622.24.02.02
Index
Creating
3D vessel model 111
Current roadmap mask 74
D
DCS 14
DCS extended 14
Default patient 44
Degree
of Overlay Ref 79
Dilatation timer 62
Direct position
keys 24
storing 29
Direct positions 26
Disk
full 179
Display
BYPASS FLUORO 177
of stent borders 118
Display Ceiling Suspension
(DCS) 14
Display modes
native 78
subtracted 78
Distances
drawing and measuring 101
DR-DYNAVISION 142, 162
examination sequence 162
workflow 163
DSA
native display 89
postprocessing in exam
room 88
DSA-DYNAVISION 142, 164
Dyna Control 145
DynaPBV Body 154
DynaPBV Neuro 154
examination sequence 154
injection 156
workflow 155
DYNAVISION 142, 164
examination sequence 164
workflow 165
E
ECG
attaching 51
AXA4-100.622.24.02.02
191
Index
Emergency operation 176
backup mode 177
bypass fluoroscopy 177
Emergency patient
registering 45
Emergency STOP button 16
locations 16
Epipolarity lines 112
Exam set 56
for IVUSmap 130
selecting 57
Examination sequence
3D CARD 158
3D DR 148
3D DSA 150
DR-DYNAVISION 162
DynaPBV Neuro 154
DYNAVISION 164
PERISTEPPING 170
PERIVISION 172
Examination-specific
messages 174
192
FD
rotation 36
FD lift 33
membrane keys 32
Fill phase
3D CARD 158
3D DR 148
3D DSA 150
PERISTEPPING 170
PERIVISION 172
Filter
joysticks 38
Filter diaphragms 40
Finger filter 40
key 35
setting 41
Flip
preselection 54
Fluoroscopy 63
changing the program 61
storing a scene 62
Fluoroscopy signal
resetting 62
Footswitch pedal
alternative acquisition 66
Frame rate
changing 59
H
Help
on touchscreen 42
I
ICStent reference image 64
Image
making it coincide with
mask 90
mirror/flip preselection 54
scrolling 83
Image angle 98
Image directory
of a scene 82
Image format
selecting 37
Image texts
predefined 97
Injection
3D CARD 158
3D DR 148
3D DSA 150
DynaPBV Neuro 156
PERISTEPPING 170
PERIVISION 172
preparing 143
Injection mode
rotational angiography 144
Input field 36
Input format
key 35
Intravascular Ultrasound
(IVUS) 130
ISO tilting 20
Isocenter
in rotational
angiography 143
key 18
setting 52, 53
stopping 19
IVUS imaging 130
IVUSmap 130
catheters 130
exam set 130
examination workflow 131
pullback 137
step 1 133
step 2 135
step 3 137
step 4 141
AXA4-100.622.24.02.02
Index
J
Joystick
C-arm 22
collimator, filter 38
patient table 18
stand 22
K
Key
orientation 24
Keys
on the C-arm control
module 24
on the collimator control
module 35
on the stand control
module 24
on the table control
module 18
L
Large Display 15
screen layout 15
Large display 14
Left Ventricle Analysis 120
Left Ventricular Analysis (LVA)
exam room 121
LIH 80
Lines
drawing 97
Live FLUO/LIH 78
Live screen 15
Lock tabletop movement
key 18
Loop
display mode 80
in LVA 129
Low dose acquisition 66
LVA 120
exam room 121
workflow 122
M
Manual contour correction
in QCA/QVA 109
Manual control
in DYNA, 3D 145
in PERI 168
Mask
making it coincide with
image 90
new 89
Mask phase
3D DSA 150
PERIVISION 172
Membrane keys
for C-arm and FD
movements 32
for FD lift 32
Memory
full! 178
Message bar
on Large Display 15
Message lines 174
MFH = Max. Fill Hold 80
Mirror
preselection 54
Mouse
joystick 42
Mouse joystick 42
Movement
after collision 181
resuming 180
stopped 180
Movements
blocking 48
C-arm 23
not possible 178
patient table 12
stand 25
stand, C-arm 10
Moving
C-arm to 3D vessel
model 115
Moving the mask 89
N
Native 89
display modes 78
Native display
in DSA 89
AXA4-100.622.24.02.02
IZ3D 106
3D vessel model 111
exam room 107
quad view 113
Stent planning 116
193
Index
New mask
setting 89
No unit movement possible! 178
Optimal Projection
for IZ3D 114
OR table 11
Orbital movements 23
Orientation
key 18, 24
Orientation key 24
OT/UT
conversion 23
Overlay Ref 78, 79
Overlay Reference
performing 79
setting the degree 79
Override 181
Overtable/undertable
conversion 23
Overview
Artis zee ceiling 8
patient table 11
194
P
Panning
exam room 87
with zoom 87
Panning knob
patient table 18
Patient
accepting data from the
RIS 45
accepting from Sensis 47
registering 45
registering in an
emergency 45
registering manually 45
rescuing in an
emergency 182
scene directory 83
Patient position 54
checking/changing 55
Patient rescue 182
Patient table
joystick 18
movements 12
moving 19
overview 11
panning knob 18
raising / lowering 19
Performing
Advanced Roadmap 73
Overlay Reference 79
PERI
running direction 166
Peripheral angiography 166
PERISTEPPING 166, 170
examination sequence 170
workflow 171
PERIVISION 166, 172
examination sequence 172
workflow 173
Phases
Advanced Roadmap 72
roadmap 68
Pixelshift 90, 91
automatic 91
flexible 93
manual 93
starting 91
Pointer
using 87
Polygons
drawing 97
Position
moving to a reference image
position 31
Postprocessing
DSA in exam room 88
Preparations
3D CARD 158
3D DR 148
3D DSA 150
peripheral angiography 167
PERISTEPPING 170
PERIVISION 172
rotational angiography 143
Pressure measurement 51
Previous roadmap mask 74
Programmed positions 26
key 24
moving to 27
storing 29
Pullback
for IVUSmap 137
AXA4-100.622.24.02.02
Index
Q
QCA 106
exam room 107
QCA 3D 106
3D vessel model 111
Quad view
in IZ3D 113
Quantitative Coronary
Analysis 106
Quantitative Vascular
Analysis 106
Quantitative Vascular Analysis
(QCA, QVA, IZ3D)
exam room 107
QVA 106
exam room 107
R
Radiation
blocking 48
Rectangular collimation 39
Reference image
displaying for system
position 31
fading in or out 79
moving to a position 31
scrolling 83
Reference screen 15
Registration
default patient 44
Replace mask
during roadmap 70
Replacing the mask 89
Report
generating and storing 109,
129
Representative 82
Rescue 182
RESET 189
Resource display 175
Restart
application or system 187
Restarting 186
Return phase
3D DSA 150
PERISTEPPING 170
Review
IVUSmap 141
Review modes 80
Review rate
setting 81
RIS
accepting patient data 45
Roadmap 68
anatomical background 70
changing the program 61
performing 69
replace mask 70
vessel/catheter contrast 70
with two system
positions 74
Roadmap phases 68
Roadmapping technique 68
Rotating
3D vessel model to C-arm
position 115
Rotating the FD 36
Rotational angiography 142
Running direction
in peripheral
angiography 166
S
Scene
image directory 82
scrolling 83
single step 81
Scene directory 83
scrolling 83
Scene overview 82
Screen layout
of the Large Display 15
Screens 15
Scrolling
in the directories 83
SCU reset 178
Sensis
accepting patient data 47
Servo assistance
patient table 18
Shortcut
storing a position 29
SID
setting 33
Single step 81
Skin dose
alternative acquisition 66
AXA4-100.622.24.02.02
Pulse rate
changing 61
195
Index
196
T
Table
movements 19
raising / lowering 19
tilting 20
Table control module 18
keys 18
Table lift
key 18
Table swivel
key 18
Table tilt
key 18
Tabletop
changing 12
moving longitudinally
only 19
moving with joystick 21
moving with panning
knob 20
versions 11
Task cards
on the touchscreen 42
Test phase
3D CARD 158
3D DR 148
3D DSA 150
PERISTEPPING 170, 172
Texts
predefined 97
TOD
performing a calibration 99
Toggle
display mode 80
Tooltip help
on touchscreen 42
Touchscreen 42
tooltip help 42
Touchscreen control 42
Touchscreen layout
configuration 43
Transparency
compensation 167
Trendelenburg
tilting the tabletop 20
Trolley
for control modules 13
Trumpf Surgery Table 11, 19
TSC 42
U
Unit computer
resetting 178
Unit movements
blocking 48
not possible 178
User guidance 174
V
Vessel contour detection
in QCA/QVA 107
AXA4-100.622.24.02.02
Index
Vessel/catheter contrast
with Roadmap 70
Wedge filter 40
key 35
setting 41
Window center 84
Window values
automatic assignment 85
manual setting 85
Window width 84
Windowing
exam room 85
Workflow
3D CARD 159
3D DR 149
3D DSA 152
CLEARstent Dynamic 77
DR-DYNAVISION 163
DynaPBV Neuro 155
DYNAVISION 165
IVUSmap 131
PERISTEPPING 171
PERIVISION 173
Z
Zero stop
key 24
Zoom
by a factor of 2 86
to acquisition size 86
Zoom format
key 35
Zoom stage 36
selecting 37
Zooming
exam room 87
AXA4-100.622.24.02.02
197
198
AXA4-100.622.24.02.02
Disclaimer
All the pictures of devices and the program user interface
shown in this Quick Guide are only examples. The available
functions depend on the type of system, the installed options,
and the current configuration.
Options have not been marked specially. The options will only
be available if the corresponding licenses have been purchased
and installed. The quotation text of your order is the sole reference for the functional scope of your system.
Other differences in detail may occur in your system due to
constant development and improvement.
Reproduction of images can cause loss of detail.
All names shown in figures are purely fictional. Similarities to
existing people are entirely coincidental.
Siemens reserves the right to change the implementation and
specifications of this document without prior notice.
For up-to-date information, please contact your local Siemens
representative.
AXA4-100.622.24.02.02
Notes
199
Legal Manufacturer
Siemens AG
Wittelsbacherplatz 2
DE-80333 Muenchen
Germany
Global Siemens
Headquarters
Siemens AG
Wittelsbacherplatz 2
80333 Muenchen
Germany
Global Siemens
Healthcare Headquarters
Siemens AG
Healthcare Sector
Henkestrasse 127
91052 Erlangen
Germany
Phone: +49 9131 84-0
www.siemens.com/healthcare