Professional Documents
Culture Documents
M a n u a l
9 0 2 -3 0 0 1 4
R e v is io n
X 1
2 4
O c to b e r
1 9 9 7
GENERAL INFORMATION
NOTICES
Disclaimer Rheometric Scientific, Inc. has taken due care in the preparation of this
Instrument Although this document is intended to describe the initial production version of
Development the SR5, all information is based upon a prototype evaluation instrument.
Document This document, and RSI Orchestrator Online Help, are the only Rheometric
Propriety Scientific authorized publications concerning operation and calibration of this
instrument.
Copyright 1997, Rheometric Scientific, Inc.
Meaning
DANGER
WARNING
CAUTION
A situation that, if not avoided, will result in damage to either this instrument or
other inanimate property.
NOTE
SAFETY
Service Do not attempt to service this instrument, as it contains no user-serviceable
components.
Electrical Do not make physical contact with any component inside the instrument unless a
Hazards set of procedural steps specifically instructs you to do so. In such cases, follow
the procedure step-by-step.
Required While operating this instrument, you must wear eye protection that either meets
Equipment or exceeds ANSI Z87.1 standards. Additionally, wear protective clothing that has
been approved for protection against the materials under test and the test
temperatures.
Safety Notices The following safety notices that appear throughout this document are repeated
here for emphasis.
Page 8
CAUTION
POSSIBLE DAMAGE TO INSTRUMENT
The membrane air dryer must be connected between your house air and the Test
Station. Failure to do so will result in extensive damage to this instrument.
Page 10
CAUTION
POSSIBLE DAMAGE TO INSTRUMENT
ANVIL
Page 10
CAUTION
POSSIBLE DAMAGE TO INSTRUMENT
Read the operating and maintenance instructions that were supplied with your air dryer.
Failure to properly maintain the dryer, moisture-trap, and coalescing filters will result in
extensive damage to this instrument.
Pages 15 and 32
DANGER
POSSIBLE PERSONAL INJURY
POSSIBLE DAMAGE TO INSTRUMENT
Turn OFF the Circulator prior to removing or installing any Lower Fixture. The Fluid
Bath is an open vat of fluid that is sealed only when a Lower Fixture is installed. Failure
to turn OFF the Circulator prior to fixture removal and installation will cause the Fluid
Bath to overflow, resulting in possible injury and extensive damage to this instrument.
Page 32
WARNING
TOXICITY
In Table 5-1 below, SF 96-50 silicon fluid emits toxic fumes when heated to a
temperature greater than +140C. Use SF 96-50 silicon fluid above +140C only if an
externally-vented exhaust hood is installed above the circulator. Ensure that exhaust
airflow is adequate to expel fumes emitted from the circulator. Failure to heed this
warning may result in injury or death.
Page 37
DANGER
POSSIBLE PERSONAL INJURY
POSSIBLE DAMAGE TO INSTRUMENT
Turn OFF the Circulator prior to removing or installing the Peltier Fixture. The Fluid
Bath is an open vat of fluid that is sealed only when the Peltier Fixture is installed.
Failure to turn OFF the Circulator prior to fixture removal and installation will cause the
Fluid Bath to overflow, resulting in possible injury and extensive damage to this
instrument.
CONTENTS
1 INTRODUCTION ....................................................................................................................................... 7
1-1 OVERVIEW...................................................................................................................................... 7
1-1.1 Organization ........................................................................................................................... 7
1-1.2 Related Sources of Information .............................................................................................. 7
1-2 DESCRIPTION OF INSTRUMENT .................................................................................................. 7
1-2.1 General Principle .................................................................................................................... 7
1-3 SPECIFICATIONS ........................................................................................................................... 8
1-3.1 Base System .......................................................................................................................... 8
1-3.2 Environmental Control Systems ............................................................................................. 8
1-4 REQUIREMENTS ............................................................................................................................ 8
1-4.1 Pneumatic .............................................................................................................................. 8
2 COMPONENT OPERATION ..................................................................................................................... 9
2-1 MODULES ....................................................................................................................................... 9
2-2 REQUIRED READING: PREVENTING DAMAGE TO THE INSTRUMENT .................................. 10
2-2.1 Air Bearing ............................................................................................................................ 10
2-2.2 Air Dryer ............................................................................................................................... 10
2-3 TEST STATION COMPONENTS .................................................................................................. 11
2-3.1 Stage .................................................................................................................................... 12
2-3.2 User Interface ....................................................................................................................... 12
2-3.3 Anvil Components ................................................................................................................ 12
2-4 ENVIRONMENTAL CONTROLLER .............................................................................................. 13
2-4.1 Overview .............................................................................................................................. 13
2-4.2 Status Panel ......................................................................................................................... 13
3 FROM FIXTURE INSTALLATION TO SAMPLE LOADING ................................................................... 14
3-1 OVERVIEW.................................................................................................................................... 14
3-2 FIXTURE INSTALLATION AND REMOVAL .................................................................................. 15
3-2.1 Lower Fixture ........................................................................................................................ 15
3-2.2 Upper Fixture ........................................................................................................................ 17
3-3 ZEROING NORMAL FORCE......................................................................................................... 17
3-4 AUTOCAL ...................................................................................................................................... 18
3-4.1 Overview .............................................................................................................................. 18
3-4.2 Procedure ............................................................................................................................. 18
3-5 TRACK MODE ............................................................................................................................... 19
3-5.1 Overview .............................................................................................................................. 19
3-5.2 Procedure ............................................................................................................................. 19
3-6 ZEROING AND SETTING THE GAP............................................................................................. 19
3-6.1 Overview .............................................................................................................................. 19
3-6.2 Zeroing the Gap ................................................................................................................... 19
3-6.3 Setting the Gap .................................................................................................................... 20
3-7 SAMPLE LOADING GUIDELINES ................................................................................................ 20
3-7.1 Parallel Plate ........................................................................................................................ 20
1
INTRODUCTION
1-1 OVERVIEW
1-1.1 This manual describes instrument-specific features of the Rheometric Scientific
Organization Universal Stress Rheometer SR5. The following information is contained herein:
Chapter 1 Introduction
Provides an overview of your documentation and the instrument, and lists
instrument performance specifications.
Chapter 2 Component Operation
Identifies and describes components of the base instrument.
Chapter 3 From Fixture Installation to Sample Loading
Details, in sequence, procedures that you must perform prior to conducting a
test.
Chapter 4 Calibration
Contains instrument calibration procedures.
Chapter 5 Appendices
Add-on sections that describe Environmental Control Systems and various
options as they become available.
1-1.2
Related Sources
of Information
1-3 SPECIFICATIONS
1-3.1
Torque Range
Base System
Rotation Rate
Angular Resolution
5 to 2000 grams
1-3.2
Environmental Fluid Bath
Control Systems
Peltier
Furnace
1-4 REQUIREMENTS
1-4.1 Air Pressure Into the
Pneumatic Instrument
2.5 scfm
CAUTION
POSSIBLE DAMAGE TO INSTRUMENT
The membrane air dryer must be connected between your house air and the Test
Station. Failure to do so will result in extensive damage to this instrument.
2
COMPONENT OPERATION
2-1 MODULES
Throughout this manual, references are made to the following instrument modules:
Test Station
The Test Station (Figure 2-1) is the single chassis comprising the SR5 Base
System. It houses the Stress Head, Fluid Bath, and all support electronics.
TEST
STATION
ENVIRONMENTAL
CONTROLLER
Figure 2-1 Test Station and Environmental Controller
Environmental Controller
The Environmental Controller (Figure 2-1) contains all support electronics
necessary to operate optional environmental control systems. Although not
needed to operate the Fluid Bath, the Environmental Controller is supplied with
options such as Peltier.
Personal Computer (PC)
The Personal Computer (IBM -compatible PC) operates RSI Orchestrator
software. Please refer to the manufacturers literature for hardware-specific
details concerning your PC. For minimum PC requirements, refer to the
current revision of the RSI Orchestrator Startup Guide 902-30010.
CAUTION
POSSIBLE DAMAGE TO INSTRUMENT
ANVIL
2-2.2 This instrument is supplied with a membrane-type air dryer that blocks moisture
Air Dryer by passing air through a filter that blocks water molecules.
Moist air is continuously purged from the dryer through a series of purge holes.
The amount of purge can be adjusted by rotating the purge knob on the dryer.
Settings range from zero (no purge and, hence, no drying) to 3 (high purge and
maximum drying). We recommend a setting of 2. Do not set the purge below 2,
as this will not sufficiently dry the air into the instrument.
Depending upon the membrane dryer you receive, the dryer may require periodic
maintenance. Preceding and in line with the air dryer are a moisture-trap filter
and a coalescing filter. It is vital that you maintain both the moisture-trap and
coalescing filters in accordance with the instructions provided with the filters. In
general, maintenance of these filters usually involves no more than draining them
at the intervals specified by the manufacturer. Note the following caution.
CAUTION
POSSIBLE DAMAGE TO INSTRUMENT
Read the operating and maintenance instructions that were supplied with your air dryer.
Failure to properly maintain the dryer, moisture-trap, and coalescing filters will result in
extensive damage to this instrument.
10
USER INTERFACE
STAGE
ANVIL STOP
ANVIL
11
2-3.1 The stage is a motorized platform that supports the Stress Head. The stage can
Stage be manually raised and lowered using the Stage Control on the User Interface
2-3.2 The User Interface is a control panel that allows control and display of selected
User Interface Test Station functions, as described in Figure 2-3.
Stage Control
Press
To move Stage
Up slowly
Up quickly
Down slowly
Down quickly
Power
Illuminated when power is
applied to the instrument
Environmental Off
Press to unconditionally remove power from the Environmental Control
System.
Display
Temperature currently being measured by the Environmental Control
System
Gap between Upper and Lower Fixtures, relative to a zero reference
established when you zero the Gap (see paragraph 3-6 ZEROING AND
SETTING THE GAP on page 19)
2-3.3
Anvil
Components
Upper Test Fixtures mount onto the Anvil. As mentioned in paragraph 2-2.1 Air
Bearing (page 10), the Anvil is an extension of the air bearing that rotates during
testing. The Anvil Stop, essentially a pneumatic piston, steadies the Anvil
during Upper Fixture installation and sample loading. To hold the Anvil in its
current position, turn the Anvil Stop Control to the Lock position. To allow the
Anvil to rotate freely, turn the Anvil Stop Control to the Unlock position.
Fixture installation procedures are given in paragraph 3-2 FIXTURE
INSTALLATION AND REMOVAL on page 15.
12
2-4.2 Although all Environmental Control functions are initiated using RSI Orchestrator,
Status Panel the Environmental Controller provides the following monitoring on the
Environmental Control Status Panel (Figure 2-4).
13
3
FROM FIXTURE INSTALLATION
TO SAMPLE LOADING
3-1 OVERVIEW
This chapter contains procedures that must be performed prior to executing a test using RSI Orchestrator
software. Perform these procedures in the sequence presented, as outlined below:
Fixture Installation and Removal
Instructions are provided to install and remove Upper and Lower Fixtures used
in conjunction with the base (standard) instrument.
Zeroing Normal Force
A method is given to establish a zero reference normal force.
AutoCal
General information concerning AutoCal (Automatic Fixture Inertia Calculation)
is provided with instructions for implementation.
Track Mode
A procedure is given to use Track Mode, which allows the instrument to
provide optimum angular resolution.
Zeroing and Setting the Gap
Separate procedures are given for manually adjusting the gap, or allowing RSI
Orchestrator to adjust the gap for you.
Sample Loading Guidelines
Guidelines to use while loading samples are provided for each geometry.
Geometry constants and formulas used for data calculation are included in RSI
Orchestrator Online Help (see the Reference Guide in the Online Help Table
of Contents).
14
DANGER
POSSIBLE PERSONAL INJURY
POSSIBLE DAMAGE TO INSTRUMENT
Turn OFF the Circulator prior to removing or installing any Lower Fixture. The Fluid
Bath is an open vat of fluid that is sealed only when a Lower Fixture is installed. Failure
to turn OFF the Circulator prior to fixture removal and installation will cause the Fluid
Bath to overflow, resulting in possible injury and extensive damage to this instrument.
Refer to Figure 3-1 during the following procedures.
Installation
a. Raise the stage to maximum height (see paragraph 2-3.1 Stage on page 12).
b. Push the Lower Fixture fully into the base, aligning the four captive screws
with the holes in the base.
c. Tighten the four captive screws in the Lower Fixture (do not over-torque).
Removal
a. Raise the stage to maximum height (see paragraph 2-3.1 Stage on page 12).
b. Loosen the four captive screws in the Lower Fixture.
c. Pull the Lower Fixture from the base.
15
NOTE
Early production units are
equipped with eight mounting
holes.
If your unit is so
equipped, align the captive
screws on the Lower Fixture
with the holes shown (at left)
filled in black.
FRONT OF INSTRUMENT
CAPTIVE SCREWS
LOWER FIXTURE
16
3-2.2 The following procedure describes Upper Fixture installation for all Environmental
Upper Fixture Control options.
Refer to Figure 3-2 during the following procedure.
Installation
a. Raise the stage to maximum height
(see paragraph 2-3.1 Stage on page
12).
b. Hold the Anvil in its current position by
turning the Anvil Stop Control to the
Lock position (see paragraph 2-3.3
Anvil Components on page 12).
c. Align the tapped hole in the fixture with
the threaded rod in the anvil.
d. Fasten the fixture to the anvil by rotating
the threaded rod until it draws the
fixture fully upward into the anvil (hand
tighten only).
e. If you are about to load a sample, keep
the Anvil Stop Control locked;
otherwise, turn it to the Unlock position.
THREADED ROD
ANVIL
FIXTURE
17
3-4 AUTOCAL
3-4.1 AutoCal (Automatic Fixture Inertia Calculation) is a function that measures the
Overview mass of an Upper Fixture, and adds this mass to the mass of the air bearing and
its associated components. The inertia of the total mass is then calculated for
use in compensating for inertia during testing. The inertia value is stored with the
geometry information and can be used in all subsequent experiments using that
Upper Fixture.
During Dynamic testing, accurate angular displacement of the Upper Fixture can
be achieved only if the fixtures inertia is overcome at each peak deflection point.
Inertia compensation therefore becomes more important as the Dynamic test
frequency (and consequently inertia) increases.
AutoCal must be performed prior to the first time an Upper Fixture is used to run
tests. Subsequent use of the same fixture does not require AutoCal.
18
3-5.2 Prior to making contact between the Upper Fixture and the sample, activate
Procedure Track Mode by clicking the Track Mode function (Control menu) of RSI
The RSI Orchestrator Gap Control Panel function provides a single dialog box
that allows you to both zero and set the Gap.
3-6.2 If conducting tests at other than ambient temperature, ensure that the fixtures are
Zeroing the Gap at the starting or working temperature prior to proceeding.
a. Ensure that a sample is not loaded and the upper and lower fixtures are
clean.
b. Using the Stage Control on the User Interface (see paragraph 2-3.2 User
Interface on page 12), lower the stage to achieve a Gap of about 0.5 mm (as
judged visually).
c. Click either the Gap Control Panel function (Control menu), or the Gap
Control button
. The Gap Control dialog box is displayed.
d. Click Zero Fixture. An information form is displayed showing the duration
(Elapsed Time) of the zeroing operation. The present gap is displayed in the
Current Gap field, which updates at one second intervals. The following
events occur during the zeroing operation:
1. The stage lowers to achieve contact between fixtures.
2. Following contact, Current Gap is zeroed, and the information form is no
longer displayed.
e. In preparation for loading the sample, raise the Stage to maximum height by
clicking Send to Top.
f. Remain in the Gap Control Panel and proceed to the next paragraph Setting
the Gap.
19
The recommended Gap setting between Parallel Plates is between 0.5 and 2
millimeters.
3-7.1.2
Loading
Perform the following procedure to load samples using the Parallel Plate fixtures.
In general, Figure 3-4 shows properly loaded Calibration Samples, included in
the Calibration Kit (see paragraph 4-1.2 Calibration Kit on page 24), which are
representative of most samples. At the left is the PDMS Calibration Sample, at
the right is the 100 cP Calibration Sample. Note that samples such as PDMS
should bulge somewhat from the edges of the plates, while the 100 cP fluid
should taper downward from the bottom edge of the Upper Fixture.
UPPER FIXTURE
SAMPLE
LOWER FIXTURE
20
3-7.2 Prior to loading the sample, ensure that all previous procedures in this chapter
Cone and Plate have been performed (if applicable, as indicated within each procedure).
3-7.2.1
Gap Setting
The actual gap setting for each Cone and Plate fixture is stated on the
Certificate of Calibration that is included in the fixture kit. If no Certificate of
Calibration is available, contact our Technical Service department.
3-7.2.2
Test Temperatures
Use Cone and Plate for isothermal testing only. Transitions to elevated
temperatures cause expansion of the fixture. Since the nominal gap setting for
cone and plate geometry is 50 microns (0.05 millimeter), thermal expansion can
cause contact between fixtures, resulting in erroneous test data.
3-7.2.3
Loading
Perform the following procedure to load samples using Cone and Plate fixtures.
In general, Figure 3-5 shows properly loaded Calibration Samples, included in
the Calibration Kit (see paragraph 4-1.2 Calibration Kit on page 24), which are
representative of most samples. At the left is the PDMS Calibration Sample, at
the right is the 100 cP Calibration Sample. Note that samples such as PDMS
should bulge somewhat from the edges of the plates, while the 100 cP fluid
should taper downward from the bottom edge of the Upper Fixture.
UPPER FIXTURE
SAMPLE
LOWER FIXTURE
21
g. Release the Anvil by turning the Anvil Stop Control to the Unlock position.
3-7.3 Prior to loading the sample, ensure that all previous procedures in this chapter,
Couette except Zeroing and Setting the Gap, have been performed (if applicable, as
indicated within each procedure). You need not set the gap when using a
Couette.
The following sample loading procedure supports both standard and DIN
Couettes.
a. Raise the Stage to maximum height (see paragraph 2-3.1 Stage on page 12).
b. Activate Track Mode (please read paragraph 3-5 TRACK MODE on page
19).
c. Hold the Anvil in its current position by turning the Anvil Stop Control to the
Lock position (see paragraph 2-3.3 Anvil Components on page 12).
d. Pour the sample into the Cup (Lower Fixture).
e. Using the Stage Control buttons on
UPPER SURFACE
the User Interface (see paragraph 2OF BOB
3.2 User Interface on page 12),
UPPER SURFACE
lower the Bob into the Cup until the
OF CUP
sample level reaches the upper
surface of the Bob, and the following
criteria are met (see Figure 3-6):
1. The sample forms a meniscus
BOB
CUP
between the edges of the Bob
and Cup, but does not flow over
the upper surface of the Bob.
2. The upper surface of the Bob is
between zero and 5 millimeters
below the upper surface of the
SAMPLE
Cup. If this is not the case,
adjust the sample volume
accordingly.
f. Release the Anvil by turning the
Figure 3-6 Couette with Sample
Anvil Stop Control to the Unlock
Loaded
position.
22
4
CALIBRATION
4-1 OVERVIEW
4-1.1 This chapter contains calibration operations that you should perform. Table 4-1
Operations lists these operations in the required order of performance and the mandatory or
recommended frequency of performance.
Frequency of Performance
Torque Calibration
Mandatory:
During instrument installation
Following replacement of stress head
Recommended once per month
Spin Correction
Mandatory:
During instrument installation
Following movement and re-leveling of
the instrument
Recommended following Torque Calibration
Mandatory:
During instrument installation
Following replacement of stress head
Recommended once per month
23
4-1.2 This instrument is supplied with a Calibration Kit. It should contain the items
Calibration Kit listed in Table 4-2.
Table 4-2 Calibration Kit
Item
Calibration Fixture
Level
Monofilament Line
Pulley
Lock Screw)
Calibration Samples
(used to provide a
known viscosity)
Weights
24
4-2.2 Using the items provided in your Calibration Kit, perform Torque Calibration as
Procedure follows:
a. Raise the Stage to maximum height (see paragraph 2-3.1 Stage on page 12).
b. Release the Anvil by turning the Anvil Stop Control to the Unlock position
(see paragraph 2-3.3 Anvil Components on page 12). The Anvil must be free
to rotate.
c. Install the Calibration Fixture as shown in Figure 4-1. Note that the rod must
be positioned to the rear of the instrument, and must fit into the channel as
shown.
d. Insert the Pulley into the Test Station frame as shown in Figure 4-1, ensuring
that the flat machined into the Pulley shaft faces toward the left.
Secure the Pulley by tightening the setscrew in the access hole using a hex
wrench.
e. Prepare a piece of monofilament line with a knot at one end and a loop at the
other end.
f. Perform the following steps (refer to Figure 4-2):
1. Place the knotted end of the line through the slot in the Calibration Fixture,
and place the line in the groove of the pulley.
2. Hang the 50 gram weight (part number 613-02061) from the loop in the
line, ensuring that the weight hangs freely. The weight should cause the
Calibration Fixture to rotate counterclockwise (as viewed from the top of
the instrument), pulling the rod to rest against the left side of the channel
(as viewed facing the front of the instrument).
3. Adjust the line so it rests neither on the upper nor the lower surface of the
calibration fixture, as shown in Figure 4-2.
4. Ensure that the weight is free to hang without obstruction, and that the
weight is steady (not swinging from side-to-side).
g. Using RSI Orchestrator software, select Calibrate Instrument (Utilities
menu). The Transducer Setup form is displayed.
h. Click XducerCal. A calibration information form is displayed.
i. Click Torque Calibration. Yet another information form is displayed.
j. Click Ok. Calibration is complete when the following events have taken
place:
From the starting position, the Calibration Fixture should rotate clockwise,
moving the rod to the right. When calibration is complete, the rod should
return to the starting position, and the numbers in the Torque Calibration
Value and Motor Temp Reference fields should change value.
k. Exit the calibration by clicking Ok.
l. Remove the weight, line, pulley, and Calibration Fixture.
25
ROD
CALIBRATION FIXTURE
PULLEY
26
CALIBRATION FIXTURE
LINE
WEIGHT
CALIBRATION FIXTURE
UPPER SURFACE
LOWER SURFACE
LINE
27
e.
f.
f.
g.
h.
i.
j.
k.
Raise the Stage to maximum height (see paragraph 2-3.1 Stage on page 12).
Install a 25 mm Parallel Plate fixture.
Zero the Gap (see paragraph 3-6.2 Zeroing the Gap on page 19).
Load a quantity of the 10cP Calibration Sample from the Calibration Kit (see
paragraph 3-7.1 Parallel Plate on page 20). The quantity should be sufficient
to set a Gap of 1 millimeter.
Ensure that the Anvil Stop Control is in the Unlock position (see paragraph
2-3.3 Anvil Components on page 12). The Anvil must be free to rotate.
Set the Gap to 1 millimeter (see paragraph 3-6.3 Setting the Gap on page
20).
Using RSI Orchestrator software, select the Calibrate Instrument function
(Utilities menu). The Transducer Setup form is displayed.
Click SpinCorr. An information form is displayed, notifying you that Spin
Correction is about to begin.
Begin Spin Correction by clicking Ok. An information form is displayed,
notifying you that the instrument should not be disturbed during Spin
Correction.
Wait until Spin Correction is complete, as indicated by both of the following
events:
1. The information screen is no longer displayed.
2. The number in the Spin Correction field of the Transducer Setup form
changes value.
Click Ok.
Raise the stage to maximum height, then remove the Calibration Sample and
(if desired) Parallel Plate.
28
29
4-5.2 Using the items provided in your Calibration Kit, perform Normal Force
Procedure Calibration as follows:
a. Raise the Stage to maximum height (see paragraph 2-3.1 Stage on page 12).
b. Install the Calibration Fixture in the same manner as for the Torque
Calibration (see Figure 4-1). Do not install the pulley.
c. Ensure that the Anvil Stop Control is in the Unlock position (see paragraph
2-3.3 Anvil Components on page 12). The Anvil must be free to rotate.
d. Using RSI Orchestrator software, select Calibrate Instrument (Utilities
menu). The Transducer Setup form is displayed.
e. Click XducerCal. A Calibration information form is displayed.
f. Click NormalCal.
g. Click Zero. An information form is displayed, notifying you that Normal Force
is being zeroed. Before proceeding, wait until the Transducer Setup form is
again displayed.
h. Hang the 1000 gram weight from the hook that is mounted to the bottom of
the Calibration Fixture.
i. Click NormCal. A form is displayed, prompting you to enter the applied
weight.
j. Enter 1000.
k. Click Calibrate. Calibration is complete when the Transducer Setup form is
again displayed.
l. Exit the calibration by clicking Ok.
m. Remove the weight and Calibration Fixture.
30
5
APPENDICES
5-1 OVERVIEW
5-1.1 The Appendices in this chapter describe instrument options and provide
Purpose instructions that do not necessarily apply to operation and calibration of the base
instrument.
31
DANGER
POSSIBLE PERSONAL INJURY
POSSIBLE DAMAGE TO INSTRUMENT
Turn OFF the Circulator prior to removing or installing any Lower Fixture. The Fluid
Bath is an open vat of fluid that is sealed only when a Lower Fixture is installed. Failure
to turn OFF the Circulator prior to fixture removal and installation will cause the Fluid
Bath to overflow, resulting in possible injury and extensive damage to this instrument.
5-2.2 Fluid and electrical connections to the circulator are shown in Figure 5-1. At the
Circulator time this document was published, it was determined that the fluid hoses should
be pushed over the Circulator fluid nozzles and secured with hose clamps.
Connection
5-2.3 Depending on the desired operating range of the circulator, fill the circulator with
Filling the fluid as specified in Table 5-1.
Circulator
WARNING
TOXICITY
In Table 5-1 below, SF 96-50 silicon fluid emits toxic fumes when heated to a
temperature greater than +140C. Use SF 96-50 silicon fluid above +140C only if an
externally-vented exhaust hood is installed above the circulator. Ensure that exhaust
airflow is adequate to expel fumes emitted from the circulator. Failure to heed this
warning may result in injury or death.
Table 5-1 Circulator Operating Ranges and Fluids
Desired Circulator Operating Range
(bath range is slightly less)
Fluid
-20C
to +30C
+1C
to +99C
100% water
-5C
to +90C
to +130C
+40C
0C to +200C
(see WARNING above)
32
33
5-2.4 Prior to using RSI Orchestrator software to control circulator temperature, ensure
RSI Orchestrator that RSI Orchestrator Instrument Testing Limits are set to use the bath as
follows:
Fluid Bath
Configuration
a. Select the Instrument Configuration function (Utilities: Service menu).
The Instrument Options dialog box (Figure 5-2) is displayed.
34
35
5-3.2
Tentative Temperature Range
Specifications
Ramp Rate
Stability
0.1C
5-3.3 Thermal operating range is the Peltier system temperature range at a given
Selecting a fluid temperature. To select a thermal operating range, fluid temperature must
be set in accordance with the following guideline.
Thermal
NOTE
Operating Range
The temperature differentials given below are
estimated.
Assuming that the ambient temperature in the vicinity of the heat sink is +20C, a
differential (T) of -40C exists between the fluid temperature and the lower limit
of the thermal operating range. A T of +80C exists between the fluid
temperature and the upper limit of the thermal operating range.
To determine thermal operating range, add these differentials to the fluid
temperature. For example, at a fluid temperature of +20C, the thermal
operating range of the Peltier system is -20C to +100C, calculated as follows:
Lower Limit
Upper Limit
T + (+20C) =
-40C + (+20C) = -20C
T + (+20C) =
+80C + (+20C) = +100C
At a fluid temperature of +40C, the thermal operating range of the Peltier system
is 0C to +120C, calculated as follows:
Lower Limit
Upper Limit
T + (+40C) =
-40C + (+40C) = 0C
T + (+40C) =
+80C + (+40C) = +120C
36
5-3.4 The Peltier Fixture is a self-contained unit that simply mounts onto the base of
Installation the instrument.
DANGER
POSSIBLE PERSONAL INJURY
POSSIBLE DAMAGE TO INSTRUMENT
Turn OFF the Circulator prior to removing or installing the Peltier Fixture. The Fluid
Bath is an open vat of fluid that is sealed only when the Peltier Fixture is installed.
Failure to turn OFF the Circulator prior to fixture removal and installation will cause the
Fluid Bath to overflow, resulting in possible injury and extensive damage to this
instrument.
Refer to Figure 5-4 during the following procedure.
Installation
a. Raise the stage to maximum height (see paragraph 2-3.1 Stage on page 12).
b. Push the Peltier Fixture fully into the base, aligning the four captive screws
with the tapped holes in the base.
c. Tighten the four captive screws in the Peltier Fixture (do not over-torque).
d. Connect the cable to the receptacle.
Removal
a. Raise the stage to maximum height (see paragraph 2-3.1 Stage on page 12).
b. Disconnect the cable from the receptacle.
c. Loosen the four captive screws in the Peltier Fixture.
c. Pull the Peltier Fixture from the base.
NOTE
The Peltier fixture is a Lower Fixture for use only
with Parallel Plate or Cone and Plate Upper
Fixtures.
37
NOTE
Early production units are
equipped with eight mounting
holes.
If your unit is so
equipped, align the captive
screws on the Peltier Fixture
with the holes shown (at left)
filled in black.
FRONT OF INSTRUMENT
CAPTIVE SCREWS
PELTIER FIXTURE
RECEPTACLE
(connected to
Environmental
Controller)
CABLE
5-3.5 Supply fluid to the Peltier system by connecting the fluid source to the instrument
38
5-3.6 Prior to using RSI Orchestrator software to control circulator temperature, ensure
RSI Orchestrator that RSI Orchestrator Instrument Testing Limits are set to use the bath as
follows:
Peltier
Configuration
a. Select the Instrument Configuration function (Utilities: Service menu).
The Instrument Options dialog box is displayed.
39
5-4.1 When used with the Fluid Bath or Peltier Environmental Control Systems, the
Principle Humidity Cover helps prevent evaporation of samples and/or solvents. This
WELL
TRAP
(REMOVABLE)
COVER
(LEFT HALF)
COVER
(RIGHT HALF)
40
f. If you wish to purge the internal atmosphere of the Cover, connect a nitrogen
gas purge line to the connector (not shown) that is mounted on the Cover.
g. Place the Well onto the Cover.
h. Fill the Well with the sample solvent.
NOTE
In the next step, do not allow the Trap to touch
the Well. The Upper Fixture must be able to
rotate freely.
i. Push the Trap downward until it is submersed in the solvent that is contained
in the Well.
41
INDEX
Air Bearing
caution concerning instrument damage, 10
Air Dryer
caution concerning instrument damage, 10
Anvil
identification, 12
locking and unlocking, 12
Anvil Stop Control, 12
AutoCal (Automatic Fixture Inertia Calculation),
18
Cal Curve, 29
Calibration
Cal Curve Examination, 29
Kit, 24
list of required calibrations, 23
Normal Force, 30
Spin Correction, 28
Torque, 25
Cone and Plate
Gap setting, 21
loading samples, 21
test temperatures, 21
Couette
loading samples, 22
Environmental Controller
monitoring, 13
Fixtures
Lower, installation/removal, 15
Upper, installation and removal, 17
Fluid Bath
configuration, 34
filling the circulator, 32
principle, 32
Gap
setting, 20
zeroing, 19
Humidity Cover
principle of operation, 40
usage, 40
Instrument description, 7
Manual Organization, 7
Normal Force
Calibration, 30
Parallel Plate
Gap setting (recommended), 20
loading samples, 20
Peltier Option
circulator options, 39
configuration, 39
installation, 37
specifications, 36
thermal operating range, 36
Peltier System
principle of operation, 35
RSI Orchestrator Online Help, 7
Specifications, instrument, 8
Spin Correction, 28
Stage, 12
Test Station
component identification, 11
Torque
Calibration, 25
Track Mode, 19
User Interface, 12
42