The CLF10 sc and CLT10 sc analyzers are best suited for
static applications where sample pH, flow, temperature and
chlorine concentration are stable. Hach recommends these
analyzers for trending in dynamic applications where these
parameters vary. Be sure to consult a Hach application
expert to match the best instrument to your application.
Drinking Water The CLF10 sc and CLT10 sc analyzers
can be used in applications where waste stream
management is a challenge, such as residual chlorine
monitoring in ground water systems and the distribution
system. Additionally, these analyzers can be used for
process control in at-the-plant applications where real-time
continuous results are beneficial.
Power Plants The CLF10 sc and CLT10 sc analyzers can
be used for the control of disinfection processes in boiler
operations and cooling systems.
General Industrial The CLF10 sc and CLT10 sc analyzers
can be used to monitor chlorine residual to prevent
biological build-up in applications serving various industrial
processes (feed water), HVAC operations (cooling water), or
in food and beverage applications.
Wastewater The total chlorine analyzer, CLT10 sc, can be
used to monitor chlorine residual and control chlorination
processes in wastewater treatment. For this application,
Hach recommends using the acidification/cleaning kit to
ensure continuous operation without clogging.
PW
IW
Specifications*
Chlorine Sensor
Measurement Range
0 to 10 ppm
Lower Limit of Detection (LOD)
30 ppb (0.03 ppm) or lower
Limit of Quantitation (LOQ)
90 ppb (0.09 ppm) or lower
Resolution
0.001 ppm (1 ppb)
Accuracy
Free Chlorine:
3% of the reference test** (DPD) at constant pH
less than 7.2 (0.2 pH unit)
10% of the reference test** (DPD) at stable pH
less than 8.5 (0.5 pH unit from the pH at calibration)
Total Chlorine:
10% of the reference test** (DPD) at stable pH
less than 8.5 (0.5 pH unit from the pH at calibration)
20% of the reference test** (DPD) at stable pH
greater than 8.5
Repeatability
30 ppb or 3%, whichever is greater
Operating Temperature
0 to 45C (32 to 113F)
Response Time
Free Chlorine: 140 seconds or less for 90% change (T90)
at a stable temperature and pH
Power Requirements
12 Vdc 10%, at 100 mA maximum (supplied by sc controller)
Sampling Time
Continuous
Mounting
Flat, vertical surface
Interferences
Free Chlorine: Monochloramine, chlorine dioxide, ozone,
and chalk deposits
Connections
Sample Line: 1/4-inch OD
Material
Corrosion-resistant materials, fully-submersible
(stainless steel, PVC, silicon rubber and polycarbonate)
Warranty
1-year warranty on the electrode body, includes the electronics
Pressure Limit
0.5 bar, no pressure impulses and/or vibrations
Panel Dimensions
Length 482.6mm (19 in.) x Width 495.3mm (19.5 in.) x Depth
151.2mm (5.95 in.) (with panel-mounted components)
Weight
Approximately 5.5 kg (12 lbs)
(panel and empty panel-mounted components only)
Sample pH
4-9
Controller Platform
sc controller models
Temperature Compensation
Internal temperature sensor
Waterproof Rating
Current rating for Sc100/1000/200 controllers, gateway, and
sensors IP65 (NEMA 4X)
Storage Temperature
Sensor: 0 to 50C (32 to 122F) dry, without electrolyte
Certification
CE / ETL, EMC
Shipping Weight
Approximately 9.1 kg (20 lbs)
Power Requirements
12 Vdc, 30 mA maximum (supplied by controller)
Principle of Operation
Amperometry is an electrochemical technique that measures the change in current resulting from chemical reactions taking place
on the electrodes. The generated current is proportional to the analyte concentration. A typical amperometric sensor consists of
two dissimilar electrodesan anode and a cathode (i.e. silver/platinum or copper/gold, respectively).
Typically, the electrodes are covered with a membrane cap containing electrolyte, providing for better selectivity of the analysis.
Additionally, a small constant electrical voltage is applied across the electrodes.
Below is a general schematic of the reduction-oxidation reaction taking place in a simple 2-electrode amperometric system:
Cathode (working electrode):
HOCl + H+ + 2e Cl- + H2O
(reduction of hypochlorous acid)
Anode (reference electrode):
Cl- + Metal Metal-Cl + e
(oxidation of the anodic material)
In a three-electrode amperometric system, such as used in the CLF10 sc and CLT10 sc, the anode is essentially split into two
partsa reference and an auxiliary (or counter) electrode. These systems are always supported by special electrical circuit
directing the voltage between all electrodes. The three-electrode design generally makes the measurements more stable and
provides longer life for the working and reference electrodes.
Engineering Specifications
1.
2.
3.
4.
5.
Dimensions
The analyzer should be installed in
an accessible location. It can be
mounted on a flat, vertical surface
(such as a wall, panel, stand, etc.).
It should allow for access for any
checking or maintenance. Sample
flow should meet the specifications
on previous page.
6.
Ordering Information
CLF10 sc FREE CHLORINE SENSOR, sc100 CONTROLLER,