Professional Documents
Culture Documents
Application
This document summarizes important information for users of normal incidence shear
wave transducers such as the Panametrics-NDT V150-V157 and V220-V222 series.
These transducers are most commonly used with ultrasonic thickness gages, flaw
detectors, or pulser/receivers for measurement of shear wave velocity in test materials,
often in connection with calculation of elastic modulus as described in this application
note: Elastic Modulus Measurement
Background
Normal incidence shear wave transducers are single element contact transducers that
introduce shear waves directly into the test piece without the use of refractive mode
conversion. By definition, these waves propagate perpendicular to the test surface
while particle motion is parallel to the surface. The transducers use piezoelectric
elements cut perpendicular to the electrical poling direction so that when pulsed they
vibrate in shear (transverse) mode rather than longitudinal (compressional) mode when
an excitation voltage is applied, as seen in the concept drawing below.
(b) Coupling
A high viscosity shear wave couplant such as Olympus SWC (U8770277) must be
used to couple a normal incidence shear wave into the test piece. Standard ultrasonic
couplants will not work, because they are liquids or gels. A basic property of a liquid is
that it does not support a shear stress. Thus, low and medium viscosity liquids such as
conventional ultrasonic couplants will not transmit shear waves. For best results it is
important to use a very thin layer of the SWC high viscosity couplant, with firm coupling
pressure. The suggested procedure is to put a very small amount of couplant on the
transducer and spread it into a thin layer with a razor blade or straight edge, then
couple onto the test piece and further wring out the couplant by rotating the transducer.
You will typically see the echo amplitude increase as the couplant layer becomes
progressively thinner.
ultrasonic test frequencies, but care should be taken with transducer selection and
instrument setup.
http://www.olympus-ims.com/en/applications/normal-incidence-shear-wavetransducers/