Opinion Paper
Mario Plebani*, Laura Sciacovelli, Ada Aita, Michela Pelloso and Maria Laura Chiozza
Introduction
Quality specifications, frequently referred to as performance criteria, represent the level of performance
required to facilitate clinical decision-making [1]. Several
strategies to set analytical quality specifications have been
promulgated for more than 30 years, and finally a consensus on the hierarchy of models has been achieved as a
result of the Stockholm Conference [2]. Analytical quality
specifications represent fundamental criteria for measuring, assuring and maximizing the reliability of laboratory
results. However, at the Stockholm Conference, Walter G.
Guder presented a lecture on the influence on analytical
quality specifications of pre-analytical factors, underlining that despite the documented quality of analytical
process, results are sometimes distrusted because of noncompliance with the clinical case or previous results from
the same or other laboratories. When these cases were
analyzed in a quality assurance project, it turned out that
non-analytical errors explained more than 60% of the
cases, of which variables in the pre-analytical phase contributed to more than half of the cases [3].
In the last 14years a growing body of evidence has
been accumulated on the relevance of the pre-analytical
phase, its vulnerability and impact on the overall quality
of laboratory information. Although the seminal concept
of the brain-to-brain laboratory loop was described more
than four decades ago, awareness and consensus on
the importance of extra-analytical aspects in laboratory
quality are a recent achievement. In fact, a wide consensus has been achieved in the last few years on the need to
assure quality and safety in the so-called brain-to-brain
loop, in view of the fact that the quality of laboratory
information is strongly affected by the coordinated and
correlated management of all procedures and processes
of the total testing process (TTP). Under this perspective,
quality in laboratory medicine should be defined as the
guarantee that each and every step in the TTP is correctly
performed, thus ensuring valuable decision making and
effective patient care [4]. Achieving consensus for the
comprehensive definition of errors in laboratory testing
Unauthenticated
Download Date | 5/19/15 5:07 AM
944Plebani etal.: Performance criteria and quality indicators for the pre-analytical phase
has been a milestone in the effort to reduce errors and
improve patient safety in laboratory medicine [5].
According to this concept, all phases and activities of
the testing cycle should be assessed, monitored and
improved in order to decrease the total error rates and
thereby improve patient safety. Most laboratory-related
diagnostic errors are due to defects in extra-analytical
phases, including inappropriate test request and/or
result acknowledgment and interpretation. Therefore, in
the interests of patients, any direct or indirect negative
consequence related to a laboratory test must be considered, irrespective of which step is involved and whether
the error depends on a laboratory professional (e.g.,
calibration or testing error) or a non-laboratory operator
(e.g., inappropriate test request, error in patient identification and/or blood collection).
Unauthenticated
Download Date | 5/19/15 5:07 AM
Plebani etal.: Performance criteria and quality indicators for the pre-analytical phase945
Quality indicators
Preliminary results
Table 1 reports the data collected for all QIs of very high priority: errors are reported as a median value calculated on
all percentage results and sigma values. As an example, we
also report data showing the dispersion of results around
the median level (Figures 13). The variability and changes
in data over time can be explained by the heterogeneity and
progressive increase in the number of clinical laboratories
participating in the program. In addition, some laboratories have changed the type and/or increased the number of
QIs to be implemented on an annual basis. Figure 1 shows
the data collected on misidentification errors, a category
that considers both misidentified samples (on total number
of samples) and patients (on total number of requests), as
well as unlabeled samples (on total number of samples)
and errors concerning patient identification. On the lefthand side are data collected by all laboratories involved
in the program, while the right-hand side shows data from
one clinical laboratory (Lab 1). Figure 2 shows the data
on requests with erroneous data entry (on total number
of requests), only for the errors concerning the tests erroneously added but not requested namely added tests.
Finally, Figure 3 shows the data on hemolyzed samples
(on total number of samples). Some differences occurred
over time, in both the group of all laboratories (left-hand
side of the figure), and in the single laboratory (righthand side). On investigating the reasons for the increase
in errors in 2011 for Lab 1, it was found that the method
employed to identify hemolysis had changed from visual
detection to the use of the automated serum index; some
corrective measures taken led to a significant decrease in
the subsequent 2-year period. Likewise, the wide variability observed in the group of all laboratories may have been
due to differences in methods used for identifying hemolysis, namely visual detection and automated serum index.
On the basis of this evidence, it was decided to split the
data collected by laboratories using visual detection and
those using automated serum index. Finally, Table 2 provides as an example the proposed quality specifications
for three QIs based on the data collected, and according to
the proposal by Fraser etal. [25] that they should be classified into three levels: optimum, desirable and minimum.
As a minimum and maximum value, we adopted the 25
and 75 percentile, respectively. The decision to propose
three performance levels, although they are based on the
state-of-the-art, and not on biological variability as originally suggested by Fraser, was made in order to encourage
laboratories to gradually improve their performance and
that no more 25% of the laboratories results have an unsatisfactory performance.
Unauthenticated
Download Date | 5/19/15 5:07 AM
946Plebani etal.: Performance criteria and quality indicators for the pre-analytical phase
Table 1QIs results collected from 2009 to 2013.
Year
Results, %
(median)
Sigma value
(median)
Misidentification errors
Errors concerning patient
identification
Misidentified samples
Misidentified patients
Unlabelled samples
2009
2010
0.083
0.040
4.576
4.739
2011
2012
2013
0.057
0
0.010
4.656
5.040
5.040
2009
2010
2011
2012
2013
0.220
0.140
0.130
0.050
0.101
4.248
4.429
4.512
4.536
4.411
Wrong container
2009
2010
2011
2012
2013
0.050
0.040
0.020
0.020
0.038
4.791
4.853
5.027
4.932
4.860
Inappropriate sample
anticoagulant volume ratio
With insufficient sample
volume
2009
2010
0.100
0.041
4.583
4.849
2011
2012
2013
0.074
0.050
0.140
4.667
4.734
4.468
2009
0.010
5.054
2010
0.013
5.015
2011
0.010
4.338
0.8
Percentage
0.6
0.4
0.2
0
2009 2010 2011 2012 2013
All laboratories
Lab. 1
0.8
Percentage
Quality indicators
Misidentification errors
0.6
0.4
0.2
0
2009 2010 2011 2012 2013
2012
2013
0.010
0.020
4.966
4.500
Contaminated sample
2013
2014
0
0.150
3.258
4.429
Sample hemolyzed
Hematology/coagulation
Chemistry
Immunology
2009
2010
2011
2012
2013
0.850
0.660
0.630
0.270
0.440
3.887
3.949
3.975
4.175
4.093
Samples clotted
Hematology/coagulation
Chemistry
Immunology
2009
2010
2011
2012
2013
0.206
0.050
0.060
0.090
0.179
4.293
4.739
4.739
4.429
4.378
Discussion
The identification, implementation and monitoring of
QIs represent a fundamental tool for complying with
All laboratories
Lab. 1
Figure 2Requests with erroneous data entry (test added but not
requested): graph showing statistical measures (median, upper and
lower quartiles, minimum and maximum data values).
Unauthenticated
Download Date | 5/19/15 5:07 AM
Plebani etal.: Performance criteria and quality indicators for the pre-analytical phase947
Haemolyzed samples
Percentage
Conclusions
0
2009 2010 2011 2012 2013
All laboratories
Figure 3Hemolyzed samples: graphical representation of statistical measures (median, upper and lower quartiles, minimum and
maximum data values).
Table 2Quality specifications for three QIs.
Quality indicators Value
Misidentification
errors
Test transcription
errors (added tests)
Sample hemolyzed
Percentage
Sigma
Percentage
Sigma
Percentage
Sigma
Quality specifications
Minimum Desirable Optimum
(25th (Median)
(75th
percentile)
percentile)
0
4.54
0
4.26
0.120
3.84
0.010
5.04
0.070
4.59
0.440
4.09
0.040
5.25
0.240
4.74
0.852
4.39
References
based on the state-of-the-art should therefore be considered an essential preliminary step in arousing the awareness in clinical laboratories of the need to measure and
improve their performances in extra-analytical QIs. The
data collected from several laboratories worldwide have
provided valuable insight on the state-of-the-art both,
especially as they were obtained using a harmonized list of
QIs and with a homogeneous reporting system. The expression of the data as both a percentage and in sigma metrics
may allow clinical laboratories enhance their appreciation
of the quality level for each indicator and prioritize corrective actions and improvement initiatives. The classification of the quality specifications for available QIs into
three levels, optimum, desirable and minimum represents
the translation to the pre-analytical phase of a proposal
already adopted for evaluating analytical performances
1. Fraser CG. Quality specifications. In: CG Fraser, editor. Biological variation: from principles to practice. Washington DC: AACC
Press, 2001:2966.
2. Kenny D, Fraser CG, Hyltoft Petersen P, Kallner A. Consensus
agreement: conference on strategies to set global quality
specifications in laboratory medicine. Scand J Clin Lab Invest
1999;59:585.
3. Guder WG. Prenalytical factors and their influence on analytical
quality specifications. Scand J Clin Lab Invest 1999;59:54550.
4. Plebani M. Quality indicators to detect pre-analytical errors in
laboratory testing. Clin Biochem Rev 2012;33:858.
5. ISO/TS 22367:2008. Medical laboratories Reduction of error
through risk management and continual improvement. Geneva,
Switzerland: International Organization for Standardization,
2008.
6. Bonini P, Plebani M, Ceriotti F, Rubboli F. Errors in laboratory
medicine. Clin Chem 2002;48:6918.
7. Plebani M. The detection and prevention of errors in laboratory
medicine. Ann Clin Biochem 2010;47:10110.
8. Plebani M, Carraro P. Mistakes in a stat laboratory: types and
frequency. Clin Chem 1997;43:134851.
Unauthenticated
Download Date | 5/19/15 5:07 AM
948Plebani etal.: Performance criteria and quality indicators for the pre-analytical phase
9. Carraro P, Plebani M. Errors in a stat laboratory: types and
frequencies 10years later. Clin Chem 2007;53:133842.
10. Plebani M. Errors in clinical laboratories or errors in laboratory
medicine? Clin Chem Lab Med 2006;44:7509.
11. Lippi G, Becan-McBride K, Behlov D, Bowen RA, Church S,
Delanghe J, etal. Preanalytical quality improvement: in quality
we trust. Clin Chem Lab Med 2013;51:22941.
12. Holman JW, Mifflin TE, Felder RA, Demers LM. Evaluation of an
automated preanalytical robotic workstation at two academic
health centers. Clin Chem 2002;48:5408.
13. Da Rin G. Pre-analytical workstations: a tool for reducing laboratory errors. Clin Chim Acta 2009;404:6874.
14. Wallin O, Sderberg J, Van Guelpen B, Stenlund H, Grankvist K,
Brulin C. Preanalytical venous blood sampling practices demand
improvement a survey of test-request management, test-tube
labelling and information search procedures. Clin Chim Acta
2008;391:917.
15. Sderberg J, Brulin C, Grankvist K, Wallin O. Preanalytical errors
in primary healthcare: a questionnaire study of information
search procedures, test request management and test tube
labelling. Clin Chem Lab Med 2009;47:195201.
16. Kemp GM, Bird CE, Barth JH. Short-term interventions on wards
fail to reduce preanalytical errors: results of two prospective
controlled trials. Ann Clin Biochem 2012;49:1669.
17. Shahangian S, Snyder SR. Laboratory medicine quality
indicators: a review of the literature. Am J Clin Pathol
2009;131:41831.