You are on page 1of 79

Made in Russia

02

VIBROACOUSTIC AND INFRARED APPARATUS

VITAFON-2

Exploitation Manual
and
User Guide

Registered in the State Medical Equipment Log


Produced in accordance with the agreement with the inventor and under his supervision
by NPF MicroEPM Ltd., St-Petersburg, Russia

CONTENT
EXPLOITATION MANUAL......................................................................................... 3
General information............................................................................................................ 3
Functions of Vitafon-2........................................................................................................ 3
Technical data..................................................................................................................... 3
Complexity of kit................................................................................................................ 4
Safety rules.......................................................................................................................... 4
Structure of the apparatus................................................................................................... 4
Functional capabilities of Vitafon-2................................................................................... 5
Preparation of the apparatus and the operation procedure.................................................. 7
Preparation for the work and switching on............................... 7
Programming of time settings ................................. 7
Connection of converters.............................. 7
Programming of parameters, start of procedure, interruption of procedure ................... 7
Programming of parameters for the use of single vibrophone................................ 8
Programming of parameters for the use of IR-matrix.................... 9
Programming of parameters for the use of two vibrophones................. 9
Programming of parameters for the use of one vibraphone and IR-matrix................. 10
Programming of parameters for the auto start of procedures................. 11
Automatic modes with periodic start.................... 12
Exploitation, transportation and storage rules.................................................... 13
Possible malfunctions and methods of their removal......................................... 13
USER GUIDE.............................................................................................................. 14
Physiological mechanism of vibroacoustic effect.................. 14
Treatment zones ..................... 15
Healing effect of infra-red emanation ....................................................... 15
Duration of treatment procedures .................. 15
Choosing of parameters MODE and ENERGY................... 16
Time and duration of the treatment course................. 16
Habitation to the vibroacoustic treatment .................. 16
Combination with drug therapy ..................................... 16
Sensations .......................... 16
Rules of application of vibrophones .............................. 16
Contraindications ............................... 17
Alphabetical list of indications .................. 17
Literature and video materials on vibroacoustic therapy..................... 77
GUARANTEES............................................................................................................ 78
List of pictures:
Picture 1. Treatment zones C, 10, G, M, M1, M2, M3, 4 ................................... 79
Picture 2. VITAFON-2 apparatus and its parts.......................................................................... 5
Picture 3. Types of average power change cycles of the IR- matrix................................... 6
Picture 4. Types of frequency change cycles of microvibration..................................... 6
Picture 5. Application of vibrophiones with the edge of 90 to each other................................ 16
Picture 6. Application of the vibroplate for the treatment of generalized osteochondrosis....... 26
Picture 7. Application of the vibroplate for the prevention of bedsores................................. 40
Picture 8. Treatment zones , 1, , 2................................................................................... 61
Picture 9. Treatment zones D1, D2, D3, E1, E10, E12, E2, E20, E22, E3, E30, E32, E4, , F ........... 79
ATTENTION!
The present manual contains information and recommendations, which should be followed to make application of
the VITAFON-2 apparatus correct, safe and effective. Before using VITAFON-2 read attentively the present Exploitation Manual and the User Guide.
To be checked while purchasing:
working in all modes;
complicity of the kit;
distributors stamp on the Act of Acceptance and Guaranty Coupon.

-2-

EXPLOITATION MANUAL
GENERAL INFORMATION
Vibroacoustic and Infrared Apparatus VITAFON-2 is a medical device produced in accordance with the Russian
State Standard GOST R 50444-92 and certified.
Its safety characteristics answer the requirements of the Russian State Standard GOST R 50267.0-92/
IEC 601-1:1988 for the equipment of class 2, type BF, non-water resistant. According to the degree of potential risk,
the apparatus belongs to the class 2a, GOST R 51609-2000; according to the possible consequences of the fault during
the exploitation process to the class B; according to the perceived mechanical effects to the class 2,
GOST R 50444-92. The type of climatic performance NF, category 4.2, GOST 15150-69.
FUNCTIONS OF VITAFON-2
The apparatus is to be used for prevention and treatment of the diseases, caused by dysfunctions of capillary blood
and lymph flows. Via its removable vibroacoustic converters VITAFON-2 produces microvibration of continuously
changing frequency and via its IR-converter impulse infrared emanation. Vibroacoustic and infrared therapy, applied
together or separately, produce local increase of capillary blood and lymph flows leading to detectable healing effect
of inflammation diseases and traumas.
VITAFON-2 is to be used in hospitals, other medical institutions and sanatoriums. It can be also effectively applied
by a patient at home, after consultation with a doctor and under his supervision.
TECHNICAL DATA
Supply voltage, V........ 220 (110) + 10%
Power supply frequency, Hz...
50-60
Power consumption, V, not more than.
20
Duration of continuous work, hours, not less than..
8
Average term of service, years, not less than......
5
Weight, net, kg, not more than........
3,25
Overall dimensions, mm, not more than..... 18519585
Setting Parameters
Quantity of modes of vibroacoustic and / or infrared treatment
(later on treatment mode or procedure mode)......................................14
Duration of frequency change cycle and duration of average power change cycle of the IR-matrix:
Duration of Frequency Duration of Average Power
Change Cycle, sec.
Change Cycle, sec.
Modes 1and 6
150 + 15
300 + 30
Modes 2 and 4; AUTO 1AUTO 12
150 + 15
150 + 15
Mode 3
300 + 30
150 + 15
Mode 5
740 + 60
740 + 60
Mode 7
108 + 10
216 + 10
Mode 8
67 + 6

Note Mode 8 can be used only with the point converter


Treatment Mode

Range of microvibration frequency changing:


lower limit F1, Hz, within .................
upper limit Hz, not more than ...........................................
Duration of impulse frequency modulations (IM) for Modes 3 and 4,
sec., within ........
Microvibrations amplitude at the lowest possible frequency of the range,
m: max Amax, within
single vibrophone ..................
double vibrophone ................
vibroplate ......................................................................................
point vibrophone ...................
min Amin, not more than.................

38 + 5
20 000
0,8 1,2
16 28
10 22
26
10 30
0,2 Amax

Values of microvibrations amplitude according to the given parameters of ENERGY and WEIGHT:
WEIGHT
> 85 kg
< 15 kg

ENERGY 4
Amax
(0,5 0,8)Amax

ENERGY 3
(0,6 0,9)Amax
(0,3 0,6)Amax

ENERGY 2
(0,3 0,6)Amax
(0,2 0,4)Amax
-3-

ENERGY 1
0,2Amax
0,2Amax

Infrared emanation power:


min, mW, not more than .
3,0
max, mW, within ..... 20,0 40,0
Length of infrared wave of IR- matrix, m, within...... 0,95+0,04
COMPLEXITY OF KIT
A complete kit of the VITAFON-2 apparatus should include the following items:
Main block of the apparatus..... 1
Protective case for converters.... 3
Kit of removable converters:
single vibrophone.......... 1
double vibrophone..... 2
vibroplate... 1
IR-matrix....... 1
point vibrophone...................
Fixation devices:
middle.... 1
small .............
small mini..............
waist .........
combined.......
urological...................................................................
neck...................................................................
Exploitation Manual and Users Guide.... 1
Consumer packaging..... 1
Note If in this list quantity of particular item is not indicated, it means that this kit does not include the item.

SAFETY RULES
ATTENTION! The switched on apparatus has dangerous alternating voltage. While using, follow general safety
rules for exploitation of electrical devices.
IT IS FORBIDDEN TO WIPE A SWITCHED DEVICE WITH A WET NAPKIN. Keep the main body of the device dry.
IT IS FORBIDDEN TO USE THE APPARATUS IN THE BATHROOMS OR IN THE SHOWER CABINS.
IT IS FORBIDDEN TO DEMOUNT, REPAIR OR MODERNIZE THE APPARATUS BY ONES OWN OR
WITH THE HELP OF UNPROFESSIONAL SPECIALISTS. When during the exploitation any fault occurs, contact
the repair services, indicated in the Guarantee Coupon.
ATTENTION! FOR FIRE SAFETY BEFORE USING THE DEVICE MAKE SURE THAT THE SOCKET IS IN
ORDER AND THE PLUG AND CORD OF THE APPARATUS ARE UNDAMAGED.
Keep the apparatus from falling down and hitting by hard surfaces. Do not lay converters cords along.
If the extension cord is needed to switch on the apparatus, it is recommended to use the manufactured extension cord.
STRUCTURE OF THE APPARATUS
The VITAFON-2 apparatus (Picture 2) consists of the main block and removable converters: the vibrophones and
the IR-matrix.
Working surfaces of the single or double vibrophones and the vibroplate are the membranes. The work of the
vibrophones is accompanied with the continuous sounding of the membranes, changing within the sound range. Succession of the frequency change repeats in cycles during the work of the apparatus.
Working surface of an IR-matrix is a metal disc with equally located holes containing build-in optical elements.
Semiconductor radiators of invisible infrared radiation are used as optical elements, from which one central and three
equally distant optical elements are equipped with red light indicators. Lightening of the four red lights during the
exploitations process indicates that IR-matrix is in order.
Working surfaces of the converters are covered with a special alloy, which has decorative coating.

-4-

Cord
and plug

POWER
switch
Light-emitting
diode indicator

Main block of
the apparatus

VITAFON-2

Display

reset

Button
RESET

Button TIME

Buttons for
programming
of parameters

Button START

Button ENTER

TIME

LEFT port
for converters
switching

START

ENTER

Double
vibrophone

RIGHT port
for converters
switching

Single
vibrophone

IR-matrix

Vibroplate

Point
vibrophone

Picture 2 VITAFON-2 apparatus and its parts

FUNCTIONAL CAPABILITIES OF VITAFON-2


1) Application at a time of one or two removable converters from the converter set. Automatic definition of the
converter applied on the display;
2) Manual choose of necessary treatment modes by special buttons (results are shown on the display);
3) Automatic change of average power of infrared impulses produced by the IR-matrix (for the dependence of the
infrared impulses power on the different modes see Picture 3);
4) Continuous automatic change of microvibration frequency produced by vibrophones (for the dependence of the
microvibration frequency change on the different modes see Picture 4);
5) Option of manual choose of microvibration amplitude by the ENERGY parameter, corrected according to the
patients weight and height;
6) Automatic start of 1-3 procedures of one type (with the same MODE, ENERGY, WEIGHT and HEIGHT parameters) within one day for modes 1-8; single programming of duration and starting time for every of these procedures;
7) Performance of procedures with periodical stops and starts (modes AUTO 1, AUTO 2, AUTO 3, AUTO 5,
AUTO 8, AUTO 12);
8) Option of choosing patients weight (WEIGHT) within the range of 3 120 kg, with the step of 1 kilo;
9) Option of choosing patients height (HEIGHT) within the range of 50 200 centimeters, with the step of 1 centimeter;
10) Option of choosing procedure duration (DURATION) within the range of 1 60 minutes, with the step of 1
minute;
11) Countdown and indication on the display of the time, left before the end of the procedure;
12) Keeping of the current time indication and the working modes lately used during electrical supply disorders or
when the apparatus was switched off for less than a month;
13) Change of a particular parameter for one unit by single pressing of the buttons , ; rapid change of a
particular parameter by continuous pressing of one of these buttons;
14) Short sound signals confirming pressing of buttons during data entry process; double sound signal when minimal
or maximal possible values of a particular parameter are entered;
15) Short sound and light signals when the procedure is near to be over;
16) Light indication of electrical connection;
17) Option of current time indication (after pressing of the button TIME).

-5-

, mW
40

Modes 1; 6. DFC=300 sec.

, mW
40

Pmax

Pmax

20

20
3
Pmin

3
Pmin

T, sec

41 63

, mW
40
Pmax

95

255
T2

90
T1

395

300

243

150

T, sec
0

, mW
40
Pmax

Modes 2; 3; 4. DFC=150 sec.

20
Pmin

Mode 5. DFC=740sec.

495
T4

390
T3
DFC

577 643
T5
T6 T7 T8

Mode 7. DFC = 216 sec.

20
T, sec

3
0

40

75

110

190

150

225

260

3
Pmin

300

T, sec

108

54

162

216

Picture 3 Types of average power change cycles of the IR-matrix

F, Hz
FUp 9730

Mode 1. DFC=150 sec.

Mode 5. DFC=740 sec.


T1=180c; T2=150c; T3=120c; T4=90c; T5=75c; T6=55c; T7=45c; T8=25c

F, Hz
9730

243
30
FL1 38

, sec
0

41

95

63

F, Hz

T,

40
0

150

T1

T2

DFC

T4

T3

T5

T6 T7 T8

Mode 2. DFC=150 sec.

FUp 9730

F, Hz
9730

243

Mode 6. DFC=150 sec.

2430

30
FL1 38

T, sec
40

0
F, Hz
FUp
973

75

305
40

150

110

Mode 3. DFC=300 sec.

im

F, Hz

243
30
FL1 38
40

75

110

150

190

225

260

300

F, Hz

F, Hz

im

30
FL1 38
75

13,5

27

40,5

54

67,5

81

94,5

108

Mode 8. DFC=67 sec.

110

T, sec

40

T, sec

40

150

1200

243

95

T, sec
0

Mode 4. DFC=150 sec.

FUp
973

63

Mode 7. DFC=108 sec.

9730
F0 x 192
F0 x 64
F0 x 48
F0 x 16
F0 x 12
F0 x 4
F0= 34

T, sec
0

T, sec

41

150

33,5

Picture 4 Types of frequency change cycles of microvibration

-6-

67

PREPARATION OF THE APPARATUS AND THE OPERATION PROCEDURE


1. Preparation for the work and switching on
Make sure that the apparatus and its converters are not damaged. Before the first useing, wipe membranes with a
napkin to disinfect them. For this moisten a napkin with a 3% hydrogen peroxide solution and wring it out. Later on
do it when necessary.
Connect the apparatus to mains. Close the switch POWER. After that the light indicator will be switched and on
the display one will see the note about converters connected. If there are no connected converters, the note on the display will be as follows:
L OFF
R OFF
2. Programming of time settings
To read up current time data, press the button TIME. After that on the display one will see the note CURRENT
TIME in the upper line and indication of hours, minutes and seconds in the lower line. For example, the note on the
display can be as follows:
CURRENT TIME
16 : 25 25
To finish current time data reading up, press the button TIME again.
To adjust current time data, press the button TIME and hold it for 4 seconds. After that the note on the display
will be as follows:
CURRENT TIME
16 < 25 12
Using buttons and , set the right indication of hours (e.g. 18) and confirm the data by pressing the button ENTER. The note on the display will be as follows:
CURRENT TIME
18 : 25 < 23
Then, using buttons and , set the right indication of minutes (e.g. 00) and press the button RESET to
make the indication of seconds equal to zero. Confirm the data by pressing the button ENTER. The note on the display will be as follows:
CURRENT TIME
18 : 00 00
3. Connection of converters
Converters can be connected to both ports of the main block, LEFT and RIGHT. If there is only one single
vibrophone, connected to the LEFT port of the main block, the note on the display will be as follows:
L 1 VIBROPHONE
R NO
If there is only vibroplate, connected to the RIGHT port of the main block, the note on the display will be as follows:
L NO
R VIBROPLATE
If there are two converters connected e.g. double vibrophone is connected to the LEFT port and IR-MATRIX
is connected to the RIGHT port the note on the display will be as follows:
L 2 VIBROPHONES
R IR-MATRIX
If there is a point converter connected to the LEFT or RIGHT port, the note on the display will be as follows:
SWITCHED ON
POINT VIBROPHONE
No other converter should be connected to the main block together with the point converter.
4. Programming of parameters, start of procedure, interruption of procedure
Connect necessary converters to the main block see the treatment scheme and confirm the data on the converters connected by pressing the button ENTER. The note on the display will be as follows:
START MANUAL <
* START AUTO
Use buttons and to choose necessary start mode.

-7-

Now one can press the button START to start the treatment procedure, if there is no need to adjust MODE,
ENERGY, WEIGHT, HEIGHT and DURATION parameters. In such case, these parameters will be taken from the
last procedure performed.
To enter new parameters, press the button ENTER instead of button START. Examples of parameters programming for different types of converters are given below.
To interrupt or discontinue the procedure, press the button RESET. After that the note on the display will be as
follows:
SESSION
INTERRUPTED
In 30 seconds this note will be automatically changed to the data about the converters connected.
5. Programming of parameters for the use of single vibraphone
For example, the following parameters are indicated in the particular treatment scheme:
single vibrophone, manual start, MODE 4, ENERGY 3, weight of the patient 75 kg, height 180 cm., treatment
DURATION 35 minutes.
Connect the single vibrophone to the left port of the main block. After that the note on the display will be as follows:
L 1 VIBROPHONE
R OFF
Press the button ENTER. The note on the display will be as follows:
START MANUAL <
* START AUTO
Press the button ENTER. The note on the display will be as follows:
MODE
2
Using buttons and , choose MODE 4. Press the button ENTER. The note on the display will be as follows:
ENERGY
2
Using buttons and , choose ENERGY 3. Press the button ENTER. The note on the display will be as
follows:
WEIGHT
75
WEIGHT and HEIGHT of the patient, as well as the treatment DURATION, are to be entered subsequently according to the same principle. The notes on the display will be as follows:
HEIGHT
180
..

DURATION
35
..

DATA ENTERED.
START?
Press the button START. The apparatus will start the procedure. The note on the display will be as follows:
SESSION ENDS
IN 35 : 00
The figures on the display indicate minutes and seconds left to the procedure completion.
When the treatment session is almost over, one will see yellow and green blinking of the light indicator and hear
short sound signals. The longer sound signal indicates that the procedure is over. After that the note on the display will
be as follows:
SESSION
FINISHED
In 30 seconds or after pressing of the button ENTER, this note will change to the data on the converters connected:
L 1 VIBROPHONE
R OFF

-8-

6. Programming of parameters for the use of IR-matrix


For the treatment by IR- matrix, one should enter only the data on the MODE and treatment DURATION according to the particular treatment scheme. For example, the following parameters are indicated in the particular treatment
scheme:
IR-matrix, manual start, MODE 5, DURATION 25 minutes.
Connect the IR- matrix to the left port of the main block. After that the note on the display will be as follows:
L IR-MATRIX
R OFF
Press the button ENTER. The note on the display will be as follows:
START MANUAL <
* START AUTO
Press the button ENTER. The note on the display will be as follows:
MODE
2
Using buttons and , choose MODE 5. The notes on the display will be as follows:
DURATION
25
DATA ENTERED.
START?
Press the button START. The apparatus will start the procedure. The note on the display will be as follows:
SESSION ENDS
IN 25 : 00
The figures on the display indicate minutes and seconds left to the procedure completion.
When the treatment session is almost over, one will see yellow and green blinking of the light indicator and hear
short sound signals. The longer sound signal indicates that the procedure is over. After that the note on the display will
be as follows:
SESSION
FINISHED
In 30 seconds or after pressing of the button ENTER, this note will change to the data about the converters connected:
L IR-MATRIX
R OFF
7. Programming of parameters for the use of two vibrophones
For example, the following parameters are indicated in the particular treatment scheme:
converters in use: L double vibrophone, ENERGY 3; R vibroplate, ENERGY 2, manual start, MODE 2,
patients weight 80 kg, height 180 m., DURATION 10 minutes.
Connect the double vibrophone to the LEFT port of the main block and the vibroplate to the RIGHT port. After
that the note on the display will be as follows:
L 2 VIBROPHONES
R VIBROPLATE
Press the button ENTER. The note on the display will be as follows:
START MANUAL <
* START AUTO
Press the button ENTER. The note on the display will be as follows:
MODE
4
Using buttons and , choose MODE 2. Press the button ENTER. The notes on the display will be as follows:
ENERGY
L=2 <
R=4
Using buttons and , choose ENERGY 3. Press the button ENTER. The notes on the display will be as
follows:
ENERGY
L=3
R=4<

-9-

Using buttons and , choose ENERGY 2. Press the button ENTER. The notes on the display will be as
follows:
WEIGHT
75
WEIGHT and HEIGHT of the patient, as well as the treatment DURATION, are to be entered subsequently according to the same principle. The notes on the display will be as follows:
HEIGHT
180
DURATION
10
DATA ENTERED.
START?
Press the button START. The apparatus will start the procedure. The note on the display will be as follows:
SESSION ENDS
IN 10 : 00
The figures on the display indicate minutes and seconds left to the procedure completion.
When the treatment session is almost over, one will see yellow and green blinking of the light indicator and hear short
sound signals. The longer sound signal indicates that the procedure is over. After that the note on the display will be as follows:
SESSION
FINISHED
In 30 seconds or after pressing of the button ENTER, this note will change to the data about the converters connected:
L 2 VIBROPHONES
R VIBROPLATE
8. Programming of parameters for the use of one vibraphone and IR-matrix
For example, the following parameters are indicated in the particular treatment scheme:
Converters in use: L double vibrophone, ENERGY 2, R IR-matrix, manual start, MODE 5, patients
weight 80 kg, height 180cm., DURATION 6 minutes.
Connect the double vibrophone to the LEFT port of the main block and the IR-matrix to the RIGHT port. After
that the note on the display will be as follows:
L 2 VIBROPHONES
R IR-MATRIX
Press the button ENTER. The note on the display will be as follows:
START MANUAL <
* START AUTO
Press the button ENTER. The note on the display will be as follows:
MODE
2
Using buttons and , choose MODE 5. Press the button ENTER. The notes on the display will be as follows:
ENERGY
4
Using buttons and , choose ENERGY 2. Press the button ENTER. The notes on the display will be as
follows:
WEIGHT
75
WEIGHT and HEIGHT of the patient, as well as the treatment DURATION, are to be entered subsequently according to the same principle. The notes on the display will be as follows:
HEIGHT
180
DURATION
6
DATA ENTERED.
START?
- 10 -

Press the button START. The apparatus will start the procedure. The note on the display will be as follows:
SESSION ENDS
IN 6 : 00
The figures on the display indicate minutes and seconds left to the procedure completion.
When the treatment session is almost over, one will see yellow and green blinking of the light indicator and hear
short sound signals. The longer sound signal indicates that the procedure is over. After that the note on the display will
be as follows:
SESSION
FINISHED
In 30 seconds or after pressing of the button ENTER, this note will change to the data about the converters connected:
L 2 VIBROPHONES
R IR-MATRIX
9. Programming of parameters for the auto start of procedures
Procedures, for which MODES 1-8 are prescribed, can be performed in automatic regime (START AUTO). In
automatic mode it is possible to program parameters for 1-3 daily procedures (treatment sessions) of the same type
(with the same MODE, ENERGY, WEIGHT and HEIGHT parameters) with different treatment duration and different starting time for each procedure.
Connect the converters prescribed. Then, using buttons and , choose auto start option. The note on the
display will be as follows:
START MANUAL
* START AUTO
<
Press the button ENTER. Then, following the same principles, enter prescribed parameters of MODE, ENERGY, WEIGHT and HEIGHT parameters. The note on the display will be as follows:
* SESSION 1 START
000
Now the apparatus is ready for the programming of the duration and starting time of the first treatment session.
For example, the first session should be 15 minutes long and start at 09-30 a.m.
Using buttons and , programme the hour (09) when the session should start. The note on the display
will be as follows:
* SESSION 1 START
0900
Press the button ENTER and programme the figure of starting time minutes (30) for the first treatment session.
The note on the display will be as follows:
* SESSION 1 START
9:30
Press the button ENTER and, using buttons and , programme the prescribed procedure DURATION.
The note on the display will be as follows:
*

DURATION 1
15
Press the button ENTER. The note on the display will be as follows:
*
START?
DATA ENTRY ON?
Now the starting time and duration of the first treatment session are programmed. If you are to perform only one
session a day, it is enough to press the button START and the apparatus will automatically start the treatment procedure at the time entered. If you are to perform one more daily treatment session, after the pressing of button ENTER
you should enter the data for the second treatment session following the same algorithm, as described above for the
first session data entry. Then, accordingly, you may enter the data for the third daily treatment session.
If after programming of the starting time of one of the treatment sessions, the note on the display is as follows:
TIME BUSY
SESSION No
it means that the starting time you chose is already busy with some other previously programmed treatment session.
Press the button ENTER and programme correct data on hours and minutes of the following treatment session.
After programming of the parameters for the necessary number of treatment sessions, the note on the display is as
follows:
- 11 -

DATA ENTERED.
START?
Press the button START. The note on the display will be as follows:
* TIME HH:
SESSION No HH:
The upper line indicates the current time (hours and minutes); the lower line shows the number of the next treatment session ( 1, 2 or 3) and the time rest before its beginning.
At the time indicated for the beginning of this session, you will hear a sound signal and the treatment session will
begin. The note on the display will be as follows:
*
SESSION ENDS
IN :
The figures on the display indicate minutes and seconds left to the procedure completion. Yellow light indicator
will turn green to show that the procedure is in process.
When the treatment session is almost over, one will see yellow and green blinking of the light indicator and hear
short sound signals. The longer sound signal indicates that the procedure is over. After that the note on the display will
be as follows:
*
TIME HH:
SESSION No HH:
The upper line indicates the current time (hours and minutes); the lower line shows the number of the next treatment session and the time rest before its beginning.
The symbol * on the display shows that the apparatus either works in automatic mode or is under programming
process.
When the apparatus works in the automatic mode, all its buttons are blocked. If necessary, working process can
be discontinued by pressing of the button RESET.
10. Automatic modes with periodic start
The modes AUTO 1, AUTO 2, AUTO 3, AUTO 5, AUTO 8 and AUTO 12 differ from the abovementioned modes
1-8 in a way that the treatment sessions automatically start with the programmed breaks till the button RESET is pressed or
the apparatus disconnected. For these modes the duration of every treatment session is equal 3 minutes but the breaks
between these sessions, according to the particular mode, can last from 3 to 36 minutes:
AUTO 1 3 minutes of work, 3 minutes of break, 3 minutes of work, 3 minutes of break etc.;
AUTO 2 3 minutes of work, 6 minutes of break, 3 minutes of work, 6 minutes of break etc.;
AUTO 3 3 minutes of work, 9 minutes of break, 3 minutes of work, 9 minutes of break etc.;
AUTO 5 3 minutes of work, 15 minutes of break, 3 minutes of work, 15 minutes of break etc.;
AUTO 8 3 minutes of work, 24 minutes of break, 3 minutes of work, 24 minutes of break etc.;
AUTO 12 3 minutes of work, 36 minutes of break, 3 minutes of work, 36 minutes of break etc.;
For the dependence of total dose of vibroacoustic therapy on the parameters of MODE and DURATION see the
following table:
MODE
AUTO 1
AUTO 2
AUTO 3
AUTO 5
AUTO 8
AUTO 12

Periodicity of
start
1 time
in 6 minutes
1 time
in 9 minutes
1 time
in 12 minutes
1 time
in 18 minutes
1 time
in 27 minutes
1 time
in 39 minutes

Total dose of vibroacoustic therapy in minutes,


when one treatment session lasts
1 hour 2 hours 4 hours 6 hours 12 hours 14 hours 24 hours
30

60

120

180

360

420

720

21

42

81

120

240

282

480

15

30

60

90

180

210

360

12

21

42

60

120

141

240

15

27

42

81

96

162

12

21

30

57

66

111

Connect the converters prescribed. Then, using buttons and , choose auto start option. The note on the
display will be as follows:
START MANUAL
* START AUTO<
Press the button START. Using buttons and , choose the necessary working mode, for example,
AUTO 2. The note on the display will be as follows:
- 12 -

MODE
AUTO 2
WEIGHT and HEIGHT of the patient, as well as ENERGY are to be entered subsequently according to the same
principle. The note on the display will be as follows:
* DATA ENTERED.
START?
Press the button START. The apparatus will start the procedure. The note on the display will be as follows:
*
SESSION ENDS
IN 03 : 00
The figures on the display indicate the time (hours and minutes) left before the break. During the break the note
on the display will be as follows:
*
TIME HH:
AUTO 2
The upper line indicates the current time (hours and minutes); the lower line shows the chosen working mode.
When the break is over, the next treatment session will start automatically and will last for 3 minutes, then, after
another break, this cycle will be repeated again and again till the button RESET is pressed or the apparatus disconnected.
If the button RESET was pressed, the apparatus will not keep the last programmed parameters and is ready for
new programming of the parameters. If the apparatus was disconnected from the mains, the last programmed parameters will be kept and, after switching on, the apparatus will continue its work according to the last entered program.
EXPLOITATION, TRANSPORTATION AND STORAGE CONDITIONS
The apparatus is to work properly for a long time in the following conditions: air temperature: + (25+10) ; relative humidity: 45 80 %.
To keep the device in working condition, one should ship and store it duely packed, with air temperature being
within (-10 / +40 ) and relative humidity not more than 80 %. To prevent damage of liquid-crystalline display, it is
not advisable to utilize the device than the air temperature is less than -10 .
After a long-distance shift or cooling, the device is to stay in the normal conditions (room temperature) unswitched
for 4 hours.
For disinfection the main body and removable converters are to be wiped with a napkin slightly moistened with a
3% hydrogen peroxide solution.
During long breaks between treatment sessions the plug should be switched off from the socket, the converters
detached from the main body and put into their protective cases. The apparatus and its parts should be stored in their
initial package.
POSSIBLE MALFUNCTIONS AND METHODS OF THEIR REMOVAL
Possible malfunction
Probable reason
Method of removal
There is no light on the display No voltage in the socket, to Connect the apparatus to the operwhen the apparatus is switched on. which the apparatus is connected. able socket.
When the apparatus is on, the Built-in clock accumulator has Connect the apparatus to the
clock shows wrong indication of run down after long storage.
mains and wait for 67 hours till
current time.
the accumulator is charged.
Correct the time settings, following the instructions from the Exploitation Manual.
The way how the apparatus
works does not correspond to the
one, described in the Exploitation
Manual and to the functional abilities of the apparatus.

1. Serious disturbances in the net,


which caused loosing of specific
settings of the apparatus.
2. Bad contact in the electrical
socket.

When the apparatus is switched


on, press the button in a round hole
at the bottom of the main block
with a thin stick of insulating material.

After the button START was During the process of parameters Press the button ENTER. Connect
pressed, the following note ap- entering, the converters were un- unconnected converters to the main
pears on the display:
connected from the main block. block. One will see on the display
the information about the convertCONNECTION
ers, connected to the main block.
ERROR
If the described methods of malfunctions removal were not effective, contact the repairing service of the manufacturer.
- 13 -

USER GUIDE
A new branch of medicine appeared in 1994 the treatment by dosed microvibration effect. Due to its high effectiveness and relatively low cost, vibroacoustic therapy has become widely popular very fast.
Treatment methods described in the Users Guides for vibroacoustic apparatuses are based on the effect of increase of blood and lymph flows in the zones of vibroacoustic therapy. They are resulted from clinical and medical
examinations on vibroacoustic therapy made before the middle of 2002.
At the same time more and more facts connected with vibroacoustic therapy were found out, which could not be
explained just by the effect of increase of blood and lymph flows. During the attempts to find valid explanation it
was noticed that every living organism has its own microvibrational background. It appeared to be that all muscle
fibres are continuously vibrating, even while sleeping. Thus the assumption was made that microvibration is one of
the resources of every living organism.
This assumption became the base of resource conception of the processes, which take place in a living organism. It was a surprise that this conception gave logical explanations to many paradoxes and united previously
fragmentary pieces of knowledge. This resulted in clear understanding how one can raise the immunity, defeat the
diseases and postpone infirmity and ageing. These knowledge were expounded in the book of the group of Russian
scientists Resources of an organism immunity, health, longevity /1/.
One can find in the book strong proofs that vibroacoustic resource support is necessary not only for the medical
treatment purposes but also for general increasing of physical and mental human abilities. It is needed by grownups, children, babies.
Practical recommendations, which one will find in this book, allow raising ones immunity, slowing down ageing and reducing morbidity without serious changes in the life-style.
ATTENTION!
Before using VITAFON-2 you should perform medical examination and consult with your doctor about
your diagnosis, indications and contraindications.
Recommendations and instructions are based on clinical data of June 2002.
Recommendations and instructions are based on clinical trails performed at:
- Moscow I.M.Sechenov Medical Academy;
- Moscow Regional Research Clinical Institute;
- St-Petersburg I.I.Mechnikov Medical Academy;
- S.M.Kirov Military Medical Academy (St-Petersburg);
- State Medical University No.1 (St-Petersburg);
- G.I.Turner Russian Research Institute of Pediatric Orthopedia (St-Petersburg);
- Municipal Children hospital No.1 (St-Petersburg).
Vitafon-2 apparatus can be applied in the same cases and for the same purposes that Vitafon and Vitafon-IR.
Results of clinical trails of these apparatus are taken into consideration during preparation of the present guide.
Physiological mechanism of vibroacoustic effect
Microvibrations are present in every living organism and play there the same role as Brownian motion in abiotic environment. In the case when Brownian motion stops (e.g. when the temperature equals absolute zero) the chemical reactions stop. And all physiological and biochemical processes stop when no microvibration take place. Microvibrations
important and indispensable resource of the organism, comparable to food and oxygen. /1/.
A plant receives the microvibration energy from the wind, rain and animals activity. Animals and people have to
activate microvibrations by themselves. The sources of microvibrations are the activities of different muscle fibers,
performed even when the organism is resting or sleeping. While sleeping, the organism has to reduce background
muscle activity at least twice; it helps the organism to restore muscle resources. This reduction makes the lymph outflow from the tissues lower and, as a result, causes morning edemas, constraint in movements, sometimes painful sensations. Not mobile style of life essentially reduces the general level of microvibrations in an organism. This, sooner or
later, leads to different diseases.
The voice, usually used for the communication, satiates the lung and head with microvibrations. A baby cries so often and loud because crying helps him to saturate his body with microvibrations. Quite often it is enough to rock the
cradle to calm him. In the state of the deep rest, the cats start to purr; by doing this they compensate the decrease in
microvibration background. There is a proverb, that laughing prolongs our lives. Now we know that this saying has
serious physiological reasons.
Vibroacoustic therapy allows increasing of microvibration background of a human organism. This is the reason
why this kind of therapy is so effective and the field of its application is so broad. Vibroacoustic apparatuses use the
range of frequencies and vibrational amplitude which are equal to the natural microvibrations, produced by any living
organism.
- 14 -

Treatment zones
As a rule, vibroacoustic treatment should cover several treatment zones within a single procedure. The main treatment zones usually are the zone of pathology and the zones which dont have their own muscle tissue kidneys,
liver and the spinal column. Recourses of the whole human organism depend on the intensity of microvibrations in
these zones. /1/.
Every pathology causes redundant quantity of affected cells and metabolic products, utilization of which is a necessary condition of recovery. Lymphatic and venous outflows as well as activity of lymphatic knots and kidneys play the
most important role in this utilization process. Coming into blood, slags break acid and alkanine balance. Buffer characteristics of blood and activities of kidneys are to prevent this break. When kidneys are not active enough to keep the
balance, to preserve blood parameters an organism increases arterial pressure and decreases local blood outflow not to
receive from tissues products of dis-balance. So, treatment of the kidney zone is the main part of vibroacoustic
therapy. It causes activation of kidneys functions, prevention of arterial pressure increase and local blood outflow decrease. Kidney zone is usually treated in accordance with different methods of vibroacoustic therapy.
Vibroacoustic therapy at the liver zone provides acceleration of slag detoxication and utilization and considerably
diminishes allergic manifestations.
Vibroacoustic therapy at the zone of the spinal column is also very important. A particular spine zone often innervates pathology area, even a small oedema or pinch of a nerve root often causes a disease. Vibroacoustic treatment of
the particular spine zone accelerates blood supplement of suppressed nerve fibres and helps to remove an oedema or
intervertebral disc decompession.
More than this, vibroacoustic therapy at the zone of the spinal column increases the quantity of ancestral stem cells,
coming into blood (Patent of Russian Federation 2166924). Ancestral stem cells play the main role in the process of
rejuvenation of bodily tissues.
Vibroacoustic therapy at the zone of pathology provides acceleration of slag detoxication, diminishes oedemas, intensifies immune reactions and raises efficiency of drug treatment.
Healing effect of infra-red emanation
Therapeutic effect of infrared emanation is based on acceleration of microcirculation in body tissues, activation of
natural physical and biochemical processes in an organism and reduction of inflammation process.
The frequency of IR-matrix is continuously changing in accordance with a given cycle program from 3 mW to
40 mW. More than it, the steel-plate of IR-matrix also has built-in IR-converters, which produce little temperature
decrease in the application zone of body (2-3o). This causes acceleration of micro-shivering of constrictors and improvement of local lymphatic and venous outflow.
IR-matrix uses IR-diodes with a narrow emanation line and the average wave length of 0,95m. Treatment by
IR-emanation of this kind produces anti-inflammation effect.
Duration of treatment procedures
During vibroacoustic treatment of pathology zone more slags come out through lymphatic and venous outflow. The
longer treatment takes place, the more breaking acid and alkanine balance slags come out from the pathology zone.
This increases activity of kidneys and may lead to compensatory increase of pressure. So, in majority of cases total
daily local treatment of pathology zone should continue less than total daily treatment of kidney zone.
To prevent large inflammation process in cases of chronic diseases (it may be caused by quick inflammation mediators income through blood to the pathology zone) duration of treatment should be prolonged continuously for 1-3
minutes every day, starting from 3-10 minutes. For long and severe diseases initial duration and dose should be little
and are to increase continuously. Exacerbation of chronic process usually occurs in 3-7 days after the treatment start,
in some cases it can take up to 2 weeks. So, the dose should be prolonged continuously within the term of 1-3 weeks in
depending on duration and severity of the disease.
Treatment can be dosed not only by its duration but also by capacity, given by the ENERGY parameter. This
should be used for the treatment of highly sensitive patient, having manifestations of exacerbation already after 2-3
minutes of vibroacoustic treatment. In such cases the ENERGY is to be increased one step per every week, starting
from the first one.
Manifestations of exacerbation are normal and show positive dynamics of the treatment. However, they produce negative
psychological reaction in patient and on this reason should be minimized. One should take into account that probability of exacerbation start becomes lower if a daily dose of treatment is received via several short procedures (2-6 per day).
Oppositely, there are the cases when exacerbation of chronic process is to be provoked. This should be done by
long single local treatments in the pathology zone performed with a break of several days between the procedures. For
every treatment the ENERGY parameter should be 1-2 steps more or treatment duration is to be 2-3 times longer than
a normal one. If no effect occurs, both the ENERGY and treatment duration are to be increased.
MODE is also an important characteristic, concerning range of treatment frequencies. MODEs differ by their infrasound energy and by treatment toughness. All in all there are 6 different MODEs: 1, 2, 5, 7, 3, 4 and 6.

- 15 -

Choosing of parameters MODE and ENERGY


MODE and ENERGY parameters are to be chosen depending on the treatment zone.
Bowels are the zone, the most sensitive to microvibrations of all kinds. In a bowels zone MODEs 3, 4, 6, and ENERGY steps 3 and 4 are not to be applied. The only exception is a full bowels dyscenisia after surgeries. Still the
treatment in these MODEs is to be brief and accompanied with vibroacoustic therapy of low intensity.
MODEs 3, 4, 6 accelerate lymphatic outflow and are to be applied at extremities for the treatment of traumatic oedemas and hematomas. MODEs 3, 4 and 6 can be briefly applied for the treatment of internal organs for the achievement of anti-inflammation effect in chronic pathologies.
For the treatment of internal organs MODEs 1, 2, 5 and 7 are to be applied.
MODE 8 is applied only together with a point converter for the shaking of stones in the pelvis. This can be done
only in hospital. This MODE can be also applied at home for the treatment of biologically active points on extremities.
Time and duration of the treatment course
Procedures are to be performed daily or with a break of 1-2 days. Detectable healing effect usually occurs when the
course is over. Duration of the course depends on the degree of severity, duration and other characteristics of a disease
and varies from several days to several months. It is an interesting fact that the healing effect will be stored and accumulated if the treatment by VITAFON-2 takes place periodically.
ATTENTION! If vibroacoustic therapy takes place simultaneously in accordance with several different schemes
for the treatment of different diseases, total duration of a single treatment session should not exceed 1 hour. Breaks
between different treatment courses should not be less than 5 days, if there is no special note in the treatment scheme.
Habitation to the vibroacoustic treatment
Habitation for the vibroacoustic treatment was never registered. Still there is a kind of dependence on vibroacoustic
therapy for some chronic diseases, e.g. hypertension, ademona and some other. It means that in some time (usually in
some months) after the healing effect is reached and the treatment is over the result achieved continuously degrades. In
such cases treatment courses are to be repeated periodically or the so-called supplementary treatment is to be performed (1-3 times a week).
Combination with drug therapy
Vibroacoustic treatment can be effectively combined with the drug therapy. However, one should take into consideration
that drug efficiency improves within the scope of VITAFON-2 action, and in some cases it may become necessary to reduce
initial drug dose. Many drugs have different side-effects and patients can feel badly while taking drugs.
Sensations
During the treatment procedure, sensations are generally absent or there is a pleasant feeling of warmth or vibration. In the case of a chronic disease, an aggravation of pain usually occurs. It is normal and shows positive dynamics.
The aggravation of pain can be considerably reduced by little continuous prolongation of session duration and, if necessary, by the use of anti-inflammatory of bactericidal drugs.
Rules of application of vibrophones
IR-matrix and point converter should be applied directly to the patients body. Membranes surfaces are to be applied to the body covered with a gauze, paper or cotton towel. The vibroplate should be located under the back of a
laying patient; its surface should be covered with thin moisture-proof material.
All converters should be fixed with an elastic bandage, special fixing equipment for the apparatuses from the Vitafon family or can be simply supported by hands and pressed by the patients own weight. Membranes and converters surfaces should be applied closer to the body. The maximum sensation of vibration at the low frequency range
may serve as a criterion of the right application. It is not advisable to press converters with a too much effort, because
it will result in the cessation of microvibration and, consequently, of the curative action.
Converters can be placed at any distance from each other (if there is no special note in the treatment scheme) within
one plane or at any angle. It is not advisable to locate vibrophones with their converters exactly towards each other, if
the distance between two converters is less than 10 cm. In this case they will suppress each others energy and the
healing effect will be diminished.
For the treatment of joints and in some other cases it is recommended to apply the vibrophones with the edge of 90 to
each other. (Picture 5). In this case the action of the apparatus at the intersection of the vibrophones axes is maximal.

90

Picture 5. Application of vibrophones with the edge of 90 to each other.


- 16 -

Contraindications
VITAFON-2 should not be applied in the following cases:
- in body areas of oncologic pathology;
- in pregnancy;
- in body areas affected by thrombophlebitis;
- in body areas affected by apparent atherosclerosis;
- during acute infectious diseases;
- during body fever;
- where implanted pacemakers are active.
Due to the fact that no treatment schemes are worked out yet, it is forbidden to locate vibrophones directly onto the
heart area.
If particular body areas having stones (gall and urinary bladder etc.) become part of the treatment zone, the therapy
should be performed only after doctors permission and under his control.
Alphabetical list of indications
Adenoma (benign prostatic hyperplasia)..........32
Arthritis, arthrosis, rheumatic polyarthritis...............44
Bedsores (prevention and treatment)............................................................ 39
Bronchitis .................53
Burns ................59
Chronic gastroduodenitis (gastritis)..............66
Chronic hepatitis...............64
Complications of infantile cerebral paralysis................70
Complications of scholiosis..............46
Constipation..........................40
Cystitis ............. 31
Diabetic angiopathy.................. 76
Dislocations and strains................ 49
Encopresis (children faeces incontinence)................42
Enuresis (children urine incontinence)..................... 42
Essential hypertension...............54
Fractures and posttraumatic stiffness of joints..............18
Frostbites and supercooling.............. 61
Furunculosis, carbunculosis, acne ............74
Glaucoma..................74
Hematomas and injuries....................75
Hemorrhoid...............62
Impotence..................26
Insomnia ...................36
Lymphostasis of lower extremities...............68
Maxillary sinusitis.................49
Osteochondrosis, radiculitis, generalized steochondrosis ..............................20
Otitis and sensorineural deafness .............60
Periodontitis, paradontosis, stomatitis, rehabilitation after dental surgery...........43
Prostatitis...................... 28
Pyelonephritis................... 37
Resource support of debilitated patients...................................................69
Rhinitis (running nose)......................50
Serous mastitis, lactostasis............... 34
Tonsillitis (sore throat)..............52
Tracheitis...............67
Traumas of spinal column.....................................................................................46
Trophic ulcers .................. 35
Vegetovascular dystonia...........68
Wounds and sutures .....................48

- 17 -

FRACTURES AND POSTTRAUMATIC JOINT CONTRACTURES


VITAFON-2 is used to accelerate fracture healing, to increase the strength of connective tissue, to prevent complications, to remove swelling and itching within areas covered with fixed dressing, and to restore joint activity. One
should start treatment with the apparatus not earlier than in 3-4 days after trauma and after prior treatment. VITAFON-2 may be applied to areas where there are metal inserts. The apparatus can be used in combination with other
means of treatment, including plaster bandage and Yelizarovs apparatus.
A patient should lie supine.
. TREATMENT OF EXTREMITIES FRACTURES

A pillow should be placed under the fractured extremity to elevate it higher than the heart level and to accelerate
blood outflow.
Treatment procedures should be performed at the Zone and at the fracture zone. At the fracture zone vibrophones
should be located as follows:
- In the case of plaster fixation, one vibrophone is to be placed directly over the plaster, the other upper to the
bandage on the smooth body area, nearest to the fracture.
- In the case of metal inserts fixation, one vibrophone is to be placed lower to the fracture, the other 3-4 cm. up from it.
1. Treatment of uncomplicated extremities fractures
1. Treatment should be performed in accordance with the scheme 1, if doctor does not prescribe some other scheme.
Daily number of procedures 3 (once in the morning, once in the noon and once in the evening).
2. Manual start.
3. Duration of the treatment course 21 days.
The treatment should start as follows:
Enter correct data. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the fracture zone.
Start the session.

Day
1-21

Scheme 1. Treatment of uncomplicated extremities fractures


L Double vibrophone
R Double vibrophone
Number of
Energy
procedure
Duration,
Mode
L
R
minutes
Zone
Fracture zone
1-3
2
3
4
20

Treatment of femoral neck fractures and complicated extremities fractures in automatic mode
1. Treatment should be performed in accordance with the scheme 2, if doctor does not prescribe some other scheme.
Daily number of procedures 3 (once in the morning, once in the noon and once in the evening). Morning and
noon procedures consist of 2 (first) sessions, the evening one of two sessions.
2. Session 1 Auto start, session 2 Manual start.
3. Duration of the treatment course until the full recovery.

Number
of procedure

Scheme 2. Treatment of femoral neck fractures and complicated extremities fractures;


procedures in automatic mode
Session 1, auto start
Session 2, manual start
Data entry
Data adjustment
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
Energy
Energy
Duration,
Duration,
Mode
Mode
L
R
L
R
minutes
minutes
Zone Fracture zone
Fracture zone
1
2
3
4
30
2
2
3
4
30
3
2
3
4
30
2
4
4
30
Session 1. Enter correct data for the Session 1 in the automatic mode. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the fracture zone.
Start the session by pressing the button START. One will see on the screen indications of treatment time and duration for the first procedure. The apparatus should not be switched off. The apparatus will start the first, second and
third procedure at the automatic mode.

- 18 -

Session 2. Enter correct data for the Session 2 in the automatic mode. Converters should be fixed at the fracture zone: one
vibraphone should be located at the same place that during the session 1, another one at the area, nearest to the fracture.
Start the session in the manual mode. When the procedure is over, programme the apparatus for the auto start of the
first procedure of the next day. The apparatus should not be switched off.
Treatment of femoral neck fractures and complicated extremities fractures in periodic start mode
1. Treatment should be performed in accordance with the scheme 3, if doctor does not prescribe some other scheme.
2. Auto start.
3. Duration of the treatment course until the end of the rest cure.
Scheme 3. Treatment of femoral neck fractures and complicated extremities fractures; procedures in periodic start mode
L Double vibrophone
R Double vibrophone
Day
Energy
Duration of
Mode
procedure,
L
R
hours
Zone
Fracture zone
1
AUTO 12
3
6
2
AUTO 12
3
12
3
AUTO 12
3
3
12
4
AUTO 12
3
3
14
5
AUTO 12
3
3
14
6
AUTO 12
3
4
14
7
AUTO 8
3
4
12
Then the scheme of the Day 7 should be repeated till the end of the rest cure.
The treatment should start as follows:
Enter correct data. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the fracture zone.
Start the session. If after the treatment time, indicated in the treatment scheme, the procedure is finished by pressing of the button RESET, the parameters programmed will be lost; if it is finished by switching off of the apparatus,
the parameters of the last procedure will be kept.
B. TREATMENT OF FRACTURES IN THE THORAX AREA (RIBS, COLLAR-BONES)
1. Treatment should be performed in accordance with the scheme 4, if doctor does not prescribe some other scheme.
Daily number of procedures 3.
2. Manual start; for the bed patients Auto start.
3. Duration of the treatment course 21 days.
Scheme 4. Treatment of fractures in the thorax area
L -Double vibrophone
R Double vibrophone
Number of
Day
Energy
procedure
Duration,
Mode
L
R
minutes
Zone
Fracture zone
1-21
1-3
2
3
1
10
A patient should lie supine.
The treatment should start as follows:
Enter correct data. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the fracture zone.
Start the session (if Auto start option is chosen, the first, second and third procedures will start in automatic mode;
the apparatus should not be switched off).
. TREATMENT OF POSTTRAUMATIC STIFFNESS OF JOINTS

For the treatment of posttraumatic stiffness of joints the only one converter double vibraphone is to be applied.
Vibrophones are placed at the joint zone at the angle of 90o to each other. During first 5 minutes of the procedure, special joint gymnastics should be made (long banding of the injury with little pressure 20-30 sec., then long opening).
For the treatment of posttraumatic stiffness of joints the following parameters should be used:
- MODE 2;
- ENERGY 4;
- DURATION 10 minutes.
- 19 -

OSTEOCHONDROSIS, RADICULITIS, GENERALIZED


OSTEOCHONDROSIS
1. Treatment should be performed in accordance with the schemes 5-9, if doctor does not prescribe some other
scheme. Daily number of procedures 2 (once in the morning, once in the evening).
2. Manual start.
Duration of the treatment course for the treatment of osteochondrosis and radiculitis of the spinal column -21 days,
for generalized osteochondrosis 3 months.
A patient should lie supine on a soft bad.
During the first 10 days, a painful area should be treated with 5% iodine solution before going to bed. It is essential
to keep the painful area warm throughout the course of treatment.
It is recommended to treat only one kind of osteochondrosis and radiculitis at a time.
If one takes hypotensional drugs and during the treatment his blood pressure becomes too low, the dose is to be
corrected.
Treatment must be accompanied with the daily monitoring of arterial pressure. In the case of elevation of arterial pressure, the treatment must be discontinued. In the case of giddiness or weakness, the procedure duration should be reduced.
In the case of cardiac arrhythmia of after infarction treatment of the thorax area of the spine can be performed only
after consulting with a doctor and under his supervision.
In the case of mild atherosclerosis the treatment is to be performed only at the zone K, excluding any spine zone.
In the case of kidney stones the treatment can be performed only after consulting with a doctor and under his supervision.
To achieve stable effect one should undergo 2-4 courses of treatment. The interval between treatment courses is 2 weeks.
Treatment of osteochondrosis and radiculitis of the jugular area of the spine should be performed in accordance with the scheme 5.
Every procedure consists of 3 sessions. A pillow should be placed under patients neck and shoulders.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone D1.
Start the session. When the session is over start the session 2.
Session 2. Enter correct data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone D2.
Start the session. When the session is over start the session 3.
Session 3. Enter correct data for the Session 3. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 1.
Start the session.
Treatment of osteochondrosis and radiculitis of the thorax area of the spine is to be performed in accordance
with the scheme 6.
Every procedure consists of 3 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 10.
Start the session. When the session is over start the session 2.
Session 2. Enter correct data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 12.
Start the session. When the session is over start the session 3.
Session 3. Enter correct data for the Session 3. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 2.
Start the session.
Treatment of osteochondrosis and radiculitis of the lumbar area of the spine is to be performed in accordance with the scheme 7.
Every procedure consists of 3 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 22.
Start the session. When the session is over start the session 2.
- 20 -

- 21 -

Duration,
minutes

Mode

Duration,
minutes

Mode

Duration,
minutes

Mode

Number of procedure

Scheme 5. Treatment of osteochondrosis and radiculitis of the jugular area of the spinal column
Session 1
Session 2
Session 3
Data entry
Data adjustment
Data adjustment
L Double vibrophone
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
R Double vibrophone
Day
Energy
Energy
Energy
L
R
L
R
L
R
Zone Zone
Zone
Zone
Zone Zone

D1

D2

1
1
1
3
1
3
1
3
1
3
1
3
1
3
1
2
1
3
1
3
1
3
1
3
1
3
1
3
1
2
3
1
3
2
3
1
3
2
3
1
3
2
2
2
3
1
3
2
3
1
3
2
3
1
3
1
5
3
1
4
5
3
1
4
5
3
1
4
3
2
5
3
1
4
5
3
1
4
5
3
1
4
1
1
3
1
4
1
3
1
4
1
3
1
4
4
2
1
3
1
4
1
3
1
4
1
3
1
4
1
2
3
1
5
2
3
1
5
2
3
1
5
5
2
2
3
1
5
2
3
1
5
2
3
1
5
1
5
3
1
5
5
3
1
5
5
3
1
5
6
2
5
3
1
5
5
3
1
5
5
3
1
5
7
Break
1
1
3
6
1
3
2
6
1
3
6
8
2
1
3
2
6
1
3
6
1
3
2
6
1
2
3
6
2
3
2
6
2
3
6
9
2
2
3
2
6
2
3
6
2
3
2
6
1
5
3
7
5
3
2
7
5
3
7
10
2
5
3
2
7
5
3
7
5
3
2
7
1
1
3
7
1
3
2
7
1
3
7
11
2
1
3
2
7
1
3
7
1
3
2
7
1
2
3
8
2
3
2
8
2
3
8
12
2
2
3
2
8
2
3
8
2
3
2
8
1
5
3
8
5
3
2
8
5
3
8
13
2
5
3
2
8
5
3
8
5
3
2
8
14
Break
1
1
3
9
1
3
2
9
1
9
15
2
1
3
2
9
1
3
9
1
3
2
9
1
2
3
9
2
3
2
9
2
9
16
2
2
3
2
9
2
3
9
2
3
2
9
1
5
3
10
5
3
2
10
5
10
17
2
5
3
2
10
5
3
10
5
3
2
10
1
1
3
10
1
3
2
10
1
10
18
2
1
3
2
10
1
3
10
1
3
2
10
1
2
3
11
2
3
2
11
2
11
19
2
2
3
2
11
2
3
11
2
3
2
11
1
5
3
11
5
3
2
11
5
11
20
2
5
3
2
11
5
3
11
5
3
2
11
21
Break
For the preventive treatment, perform this scheme beginning from the Day 8.

- 22 -

Duration,
minutes

Mode

Duration,
minutes

Mode

Duration,
minutes

Mode

Number of procedure

Scheme 6. Treatment of osteochondrosis and radiculitis of the thorax area of the spinal column
Session 1
Session 2
Session 3
Data entry
Data adjustment
Data adjustment
L Double vibrophone
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
R Double vibrophone
Day
Energy
Energy
Energy
L
R
L
R
L
R
Zone Zone
Zone Zone
Zone Zone

10

12

2
1
1
3
1
3
1
3
1
3
1
3
1
3
1
2
1
3
1
3
1
3
1
3
1
3
1
3
1
2
3
1
3
2
3
1
3
2
3
1
3
2
2
2
3
1
3
2
3
1
3
2
3
1
3
1
5
3
1
4
5
3
1
4
5
3
1
4
3
2
5
3
1
4
5
3
1
4
5
3
1
4
1
1
3
1
4
1
3
1
4
1
3
1
4
4
2
1
3
1
4
1
3
1
4
1
3
1
4
1
2
3
1
5
2
3
1
5
2
3
1
5
5
2
2
3
1
5
2
3
1
5
2
3
1
5
1
5
3
1
5
5
3
1
5
5
3
1
5
6
2
5
3
1
5
5
3
1
5
5
3
1
5
7
Break
1
1
3
6
1
3
2
6
1
3
6
8
2
1
3
2
6
1
3
6
1
3
2
6
1
2
3
6
2
3
2
6
2
3
6
9
2
2
3
2
6
2
3
6
2
3
2
6
1
5
3
7
5
3
2
7
5
3
7
10
2
5
3
2
7
5
3
7
5
3
2
7
1
1
3
7
1
3
2
7
1
3
7
11
2
1
3
2
7
1
3
7
1
3
2
7
1
2
3
8
2
3
2
8
2
3
8
12
2
2
3
2
8
2
3
8
2
3
2
8
1
5
3
8
5
3
2
8
5
3
8
13
2
5
3
2
8
5
3
8
5
3
2
8
14
Break
1
1
3
9
1
3
2
9
1
9
15
2
1
3
2
9
1
3
9
1
3
2
9
1
2
3
9
2
3
2
9
2
9
16
2
2
3
2
9
2
3
9
2
3
2
9
1
5
3
10
5
3
2
10
5
10
17
2
5
3
2
10
5
3
10
5
3
2
10
1
1
3
10
1
3
2
10
1
10
18
2
1
3
2
10
1
3
10
1
3
2
10
1
2
3
11
2
3
2
11
2
11
19
2
2
3
2
11
2
3
11
2
3
2
11
1
5
3
11
5
3
2
11
5
11
20
2
5
3
2
11
5
3
11
5
3
2
11
21
Break
For the preventive treatment, perform this scheme beginning from the Day 8.

- 23 -

Duration,
minutes

Mode

Duration,
minutes

Mode

Duration,
minutes

Mode

Number of procedure

Scheme 7. Treatment of osteochondrosis and radiculitis of the lumbar area of the spinal column
Session 1
Session 2
Session 3
Data entry
Data adjustment
Data adjustment
L Double vibrophone
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
R Double vibrophone
Day
Energy
Energy
Energy
L
R
L
R
L
R
Zone Zone
Zone Zone
Zone Zone

22

30
1
1
3
2
3
1
3
2
3
1
3
2
3
1
2
1
3
2
3
1
3
2
3
1
3
2
3
1
2
3
2
3
2
3
2
3
2
3
2
3
2
2
2
3
2
3
2
3
2
3
2
3
2
3
1
5
3
2
4
5
3
2
4
5
3
2
4
3
2
5
3
2
4
5
3
2
4
5
3
2
4
1
1
3
2
4
1
3
2
4
1
3
2
4
4
2
1
3
2
4
1
3
2
4
1
3
2
4
1
2
3
2
5
2
3
2
5
2
3
2
5
5
2
2
3
2
5
2
3
2
5
2
3
2
5
1
5
3
2
5
5
3
2
5
5
3
2
5
6
2
5
3
2
5
5
3
2
5
5
3
2
5
7
Break
1
1
3
6
1
3
3
6
1
3
6
8
2
1
3
3
6
1
3
6
1
3
3
6
1
2
3
6
2
3
3
6
2
3
6
9
2
2
3
3
6
2
3
6
2
3
3
6
1
5
3
7
5
3
3
7
5
3
7
10
2
5
3
3
7
5
3
7
5
3
3
7
1
1
3
7
1
3
3
7
1
3
7
11
2
1
3
3
7
1
3
7
1
3
3
7
1
2
3
8
2
3
3
8
2
3
8
12
2
2
3
3
8
2
3
8
2
3
3
8
1
5
3
8
5
3
3
8
5
3
8
13
2
5
3
3
8
5
3
8
5
3
3
8
14
Break
1
1
3
9
1
3
3
9
1
9
15
2
1
3
3
9
1
3
9
1
3
3
9
1
2
3
9
2
3
3
9
2
9
16
2
2
3
3
9
2
3
9
2
3
3
9
1
5
3
10
5
3
3
10
5
10
17
2
5
3
3
10
5
3
10
5
3
3
10
1
1
3
10
1
3
3
10
1
10
18
2
1
3
3
10
1
3
10
1
3
3
10
1
2
3
11
2
3
3
11
2
11
19
2
2
3
3
11
2
3
11
2
3
3
11
1
5
3
11
5
3
3
11
5
11
20
2
5
3
3
11
5
3
11
5
3
3
11
21
Break
For the preventive treatment, perform this scheme beginning from the Day 8.

Session 2. Enter correct data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 3.
Start the session. When the session is over start the session 3.
Session 3. Enter correct data for the Session 3. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 30.
Start the session.
- 24 -

Duration,
minutes

Mode

Duration,
minutes

Mode

Duration,
minutes

Mode

Number of procedure

Scheme 8. Treatment of osteochondrosis and radiculitis of the lumbar-sacrum area of the spinal column
Session 1
Session 2
Session 3
Data entry
Data adjustment
Data adjustment
L Double vibrophone
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
R Double vibrophone
Day
Energy
Energy
Energy
L
R
L
R
L
R
Zone Zone
Zone Zone
Zone Zone

30

32

4
1
1
3
2
3
1
3
2
3
1
3
2
3
1
2
1
3
2
3
1
3
2
3
1
3
2
3
1
2
3
2
3
2
3
2
3
2
3
2
3
2
2
2
3
2
3
2
3
2
3
2
3
2
3
1
5
3
2
4
5
3
2
4
5
3
2
4
3
2
5
3
2
4
5
3
2
4
5
3
2
4
1
1
3
2
4
1
3
2
4
1
3
2
4
4
2
1
3
2
4
1
3
2
4
1
3
2
4
1
2
3
2
5
2
3
2
5
2
3
2
5
5
2
2
3
2
5
2
3
2
5
2
3
2
5
1
5
3
2
5
5
3
2
5
5
3
2
5
6
2
5
3
2
5
5
3
2
5
5
3
2
5
7
Break
1
1
3
6
1
3
3
6
1
3
6
8
2
1
3
3
6
1
3
6
1
3
3
6
1
2
3
6
2
3
3
6
2
3
6
9
2
2
3
3
6
2
3
6
2
3
3
6
1
5
3
7
5
3
3
7
5
3
7
10
2
5
3
3
7
5
3
7
5
3
3
7
1
1
3
7
1
3
3
7
1
3
7
11
2
1
3
3
7
1
3
7
1
3
3
7
1
2
3
8
2
3
3
8
2
3
8
12
2
2
3
3
8
2
3
8
2
3
3
8
1
5
3
8
5
3
3
8
5
3
8
13
2
5
3
3
8
5
3
8
5
3
3
8
14
Break
1
1
3
9
1
3
3
9
1
9
15
2
1
3
3
9
1
3
9
1
3
3
9
1
2
3
9
2
3
3
9
2
9
16
2
2
3
3
9
2
3
9
2
3
3
9
1
5
3
10
5
3
3
10
5
10
17
2
5
3
3
10
5
3
10
5
3
3
10
1
1
3
10
1
3
3
10
1
10
18
2
1
3
3
10
1
3
10
1
3
3
10
1
2
3
11
2
3
3
11
2
11
19
2
2
3
3
11
2
3
11
2
3
3
11
1
5
3
11
5
3
3
11
5
11
20
2
5
3
3
11
5
3
11
5
3
3
11
21
Break
For the preventive treatment, perform this scheme beginning from the Day 8.

Treatment of osteochondrosis and radiculitis of the lumbar-sacrum area of the spine is to be performed in
accordance with the scheme 8.
Every procedure consists of 3 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 30.
Start the session. When the session is over start the session 2.
Session 2. Enter correct data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 32.
Start the session. When the session is over start the session 3.
Session 3. Enter correct data for the Session 3. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 4.
Start the session.
Treatment of generalized osteochondrosis is to be performed in accordance with the scheme 9.
Connected to the left port vibroplate should be formed as one raw (see Picture 6) and put along the spinal column; the
upper end of it should be in the middle of the neck. Converter, connected to the right port, should be fixed at the Zone .
After the course is over, one should have a break of 1-3 months. After the break the treatment course should be repeated from the day 8 to the day 90.
Scheme 9. Treatment of generalized systemic osteochndrosis
L Vibroplate
R Double vibrophone
Daily number
Day
Energy
of procedures
Duration,
Mode
L
R
minutes
Spinal column Zone
1
2
2
1
3
5
2
2
2
3
7
3
2
2
2
3
7
4
2
2
3
10
5
2
2
3
3
10
6
2
2
3
13
7
2
2
3
3
13
8
2
2
3
15
9
2
2
3
3
17
10
2
2
3
19
11
2
2
3
3
20
12
2
2
3
22
13
2
2
3
3
25
14
2
2
3
27
3
3
30
15-90
2
2
3
30
Every second day treatment procedures at the spinal column
should not be performed.

Picture 6. Application of the vibroplate for the treatment of generalized osteochondrosis

- 25 -

IMPOTENCE
1. Treatment should be performed in accordance with the scheme 10, if doctor does not prescribe some other scheme.
Daily number of procedures 2 (once in the morning, once in the evening).
2. Manual start.
3. Duration of the treatment course 21 days.
To make the effect achieved stable, the treatment course can be repeated after a break of 2 6 months.
A patient should lie supine. If necessary, a small pillow can be placed under the pelvis to keep the horizontal position.
Every procedure consists of 3 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port the first vibrophone perineum zone (between the anus and the root of the penis),
another one bladder zone (lower abdomen).
Start the session. When the session is over start the session 2.
Session 2. Enter correct data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the zone, indicated in the treatment scheme, either 3 or 4.
Start the session. When the session is over start the session 3.
Session 3. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the zone, indicated in the treatment scheme, either 32 or 30.
Start the session by pressing of the button START, for the main parameters of the sessions 2 and 3 are the same.

- 26 -

6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21

3
3
3
3
3
3
3
3
3
3
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4

3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3

4
3
4
3
4
3
4
3
4
3
4
3
4
3
4
3
4
3
4
3
4
3
4
3
4
3
4
3
4
3
4
3
4
3
4
3
4
3
4
3
-

- 27 -

1
1
1
1
1
1
1
1
1
1
2
2
2
2
2
2
2
2
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
-

1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5

3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3

30
32
30
32
30
32
30
32
30
32
30
32
30
32
30
32
30
32
30
32
30
32
30
32
30
32
30
32
30
32
30
32
30
32
30
32
30
32
30
32
-

1
1
1
1
1
1
1
1
1
1
2
2
2
2
2
2
2
2
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
-

Duration,
minutes

5
5
5
5
6
6
6
6
7
7
7
7
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8

Energy at
the Zone Zone

Energy

1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5

Energy

Mode

5
5
5
5
6
6
6
6
7
7
7
7
8
8
8
8
9
9
9
9
10
10
10
10
11
11
11
11
12
12
12
12
13
13
13
13
14
14
14
14
15
15

Mode

3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3

Energy at
the Zone Zone

Session 3
Repetition of the program entered
L Double vibrophone
R Double vibrophone
Energy
L
R

Duration,
minutes

1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5

Mode

1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2

Perineum
Zone
and blad
der zone

Duration,
minutes

Number of procedure

Day
1

Scheme 10. Treatment of impotence


Session 1
Session 2
Data entry
Data adjustment
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
Energy
Energy
L
R
L
R

5
5
5
5
6
6
6
6
7
7
7
7
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8

PROSTATITIS
1. Treatment should be performed in accordance with the scheme 11 (Standard course), if doctor does not prescribe
some other scheme. Daily number of procedures 1.
2. Manual start.
3. Duration of the treatment course 22 days.
A patient should lie supine. If necessary, a small pillow can be placed under the pelvis to keep the horizontal position.
During the first 7 days treat the lower area of the abdomen once a day with iodine solution (if the tolerance to iodine is good, treatment area is to be equivalent to two palms of the patients hands).
Every procedure consists of 2 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port the first vibrophone perineum zone (between the anus and the root of the penis),
another one bladder zone (lower abdomen).
Start the session. When the session is over start the session 2.
Session 2. Enter correct data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 4.
Start the session.

Day

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22

Scheme 11. Treatment of prostatitis. Standard Course


Session 1
Session 2
Data entry
Data adjustment
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
Energy
Energy
Duration,
Duration,
L
R
L
R
Mode
Mode
Zone
Perineum and minutes
Zone Zone minutes

bladder zones

4
1
2
1
5
1
2
2
3
2
2
1
6
2
2
2
4
5
2
1
7
5
2
2
5
1
2
1
8
1
2
2
6
2
3
2
6
2
3
2
7
5
3
2
7
5
3
2
8
1
3
2
8
1
3
3
5
2
3
2
9
2
3
3
6
5
3
3
7
5
3
3
7
1
3
3
8
1
3
3
8
2
3
3
9
2
3
3
9
5
3
3
10
5
3
3
10
1
3
3
11
1
3
3
9
2
3
3
12
2
3
3
9
5
3
3
13
5
3
3
9
1
3
3
14
1
3
3
9
2
3
3
15
2
3
3
10
5
3
3
16
5
3
3
10
1
3
3
17
1
3
3
10
2
3
3
18
2
3
3
10
5
3
3
19
5
3
3
10
1
3
3
20
1
3
3
10

In the case if the standard treatment course was not efficient, the patient should undergo the intensified course.
1. Treatment should be performed in accordance with the scheme 12, if doctor does not prescribe some other scheme.
Daily number of procedures 2 (once in the morning, once in the evening).
2. Manual start.
3. Duration of the treatment course 21 days.
Every procedure consists of 3 sessions.

- 28 -

2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21

- 29 -

Duration,
minutes

Mode

Duration,
minutes

1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5

Mode

1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2

Scheme 12. Treatment of prostatitis. Intensive course


Session 1
Session 2
Session 3
Change of converters and
Data entry
Data adjustment
data adjustment
L Double vibrophone
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
R IR-matrix
Energy
Energy
Energy
L
R
L
R
L
R
Perineum
Zone
Zone
Zone
and bladder
Zone Energy
Zone

zone
3
3
10
1
3
3
1
5
1
3
Perineum 5
3
3
10
1
3
30
1
5
1
3
Bladder
5
3
3
11
2
3
32
1
5
2
3
Perineum 6
3
3
11
2
3
4
1
5
2
3
Bladder
6
3
3
12
5
3
3
2
6
5
3
Perineum 7
3
3
12
5
3
30
2
6
5
3
Bladder
7
3
3
13
1
3
32
2
6
1
3
Perineum 8
3
3
13
1
3
4
2
6
1
3
Bladder
8
3
3
14
2
3
3
2
7
2
3
Perineum 8
8
3
3
14
2
3
30
2
7
2
3
Bladder
3
3
15
5
3
32
2
7
5
3
Perineum 8
8
3
3
15
5
3
4
2
7
5
3
Bladder
3
3
16
1
3
3
2
8
1
3
Perineum 8
8
3
3
16
1
3
30
2
8
1
3
Bladder
3
3
17
2
3
32
2
8
2
3
Perineum 8
8
3
3
17
2
3
4
2
8
2
3
Bladder
3
3
18
5
3
3
2
9
5
3
Perineum 8
3
3
18
5
3
30
2
9
5
3
Bladder
8
3
3
19
1
3
32
2
9
1
3
Perineum 8
8
3
3
19
1
3
4
2
9
1
3
Bladder
3
3
20
2
3
3
2
9
2
3
Perineum 8
8
3
3
20
2
3
30
2
9
2
3
Bladder
3
4
10
5
3
32
2
6
5
3
Perineum 8
8
3
4
10
5
3
4
2
6
5
3
Bladder
3
4
11
1
3
3
2
9
1
3
Perineum 8
8
3
4
11
1
3
30
2
9
1
3
Bladder
3
4
12
2
3
32
2
7
2
3
Perineum 8
8
3
4
12
2
3
4
2
7
2
3
Bladder
3
4
13
5
3
3
2
9
5
3
Perineum 8
3
4
13
5
3
30
2
9
5
3
Bladder
8
3
4
14
1
3
32
2
8
1
3
Perineum 8
8
3
4
14
1
3
4
2
8
1
3
Bladder
3
4
15
2
3
3
2
9
2
3
Perineum 8
8
3
4
15
2
3
30
2
9
2
3
Bladder
3
4
15
5
3
32
2
9
5
3
Perineum 8
8
3
4
15
5
3
4
2
9
5
3
Bladder
3
4
15
1
3
3
2
9
1
3
Perineum 8
8
3
4
15
1
3
30
2
9
1
3
Bladder
3
4
15
2
3
32
2
9
2
3
Perineum 8
3
4
15
2
3
4
2
9
2
3
Bladder
8
3
4
15
5
3
3
2
9
5
3
Perineum 8
3
4
15
5
3
30
2
9
5
3
Bladder
8
Duration,
minutes

Mode

Number of procedure

Day

Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port the first vibrophone perineum zone (between the anus and the root of the penis),
another one bladder zone (lower abdomen).
Start the session. When the session is over start the session 2.

Session 2. Enter correct data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the zones 3, 30, 32 or 4.
Start the session. When the session is over start the session 3.
Session 3. Change the converters and adjust data for the Session 3. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port the first vibrophone perineum zone (between the anus and the root of the penis),
another one bladder zone (lower abdomen).
Start the session.
General recommendations: Intensified course is not to be performed if the standard one is not finished. During
cold seasons, a patient should stay in the warm chamber for not less than an hour after the treatment procedure.
Throughout the course of treatment it is necessary to keep the pelvis area warm (long underwear, long coat etc.).
The treatment of acute prostatitis should be combined with taking antibiotics (prescribed by a doctor) and performed under medical supervision.
In the course of treatment of chronic prostatitis, an aggravation of pain sensations usually occurs and continues for
5-10 days. In rare cases, a transient decline of potency may be experienced on days from 2 to 4 after the treatment
start. On days from 8 to 14 the potency steadily increases. In such cases antibiotics are required.
If prostatitis is associated with a bladder problem, products of cleaning and reparation processes may be discharged
with urine. After 10-14 days urine becomes clear, pain subsides, the frequency of urination substantially decreases and
potency improves.
In case of chronic prostatitis 2-4 courses of treatment are required. A break after the first course should be 3-5 days,
after the second and the third courses 2-3 weeks.

Duration,
minutes

Mode

Duration,
minutes

Mode

Duration,
minutes

Scheme 13. Treatment of prostatitis. Preventive course


Session 1
Session 2
Session 3
Repetition of the program
Data entry
Data adjustment
entered
L Double vibrophone
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
R Double vibrophone
Energy
Energy
Energy
L
R
L
R
L
R
Perineum
Zone
Zone
Zone
Zone
Zone
and bladder

4
zone
1
3
4
15
2
3
2
10
2
3
2
10

Mode

Number of weekly
procedures

For preventive treatment VITAFON-2 should be used once a week.


1. Preventive treatment should be performed in accordance with the scheme 13 (Preventive course), one procedure
every week, if doctor does not prescribe some other scheme.
2. Manual start.
A patient should lie supine.
Every procedure consists of 3 sessions.

Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port the first vibrophone perineum zone (between the anus and the root of the penis),
another one bladder zone (lower abdomen).
Start the session. When the session is over start the session 2.
Session 2. Enter correct data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 3.
Start the session. When the session is over start the session 3.
Session 3. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 4.
Start the session by pressing of the button START, for the main parameters of the sessions 2 and 3 are the same.

- 30 -

2
3
4
5
6
7
8
9
10
11
12
13
14

1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2

1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2

L Double vibrophone
R IR-matrix
Energy
Duration,
L
R
Zone Bladder minutes

zone
2
7
3
7
3
7
3
7
3
8
3
8
3
8
3
8
3
8
3
8
3
8
3
8
3
9
3
9
3
9
3
9
3
9
3
9
3
9
3
9
3
9
3
9
3
9
3
9
3
9
3
9
3
9
3
9

Mode

Data entry

1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2

Session 2
Change of converters
and data adjustment
L Double vibrophone
R Double vibrophone
Energy
Duration,
L
R
Zone
Bladder minutes

zone
2
1
5
3
1
5
3
2
5
3
2
5
3
2
7
3
2
7
3
2
7
3
2
7
3
3
9
3
3
9
3
3
9
3
3
9
3
3
10
3
3
10
3
3
10
3
3
10
3
3
10
3
3
10
3
3
10
3
3
10
3
3
10
3
3
10
3
3
10
3
3
10
3
3
10
3
3
10
3
3
10
3
3
10

Session 3
Data adjustment

Mode

Mode

Session 1

Number of procedure

Day

CYSTITIS
In the acute condition the apparatus is used in combination with drug therapy. In remission the apparatus can be
used without taking anti-inflammatory agents: it will be enough to treat an abdomen with 5% iodine solution before
going to bed during 10 days (in case of good tolerance).
1. Treatment should be performed in accordance with the scheme 14, if doctor does not prescribe some other scheme.
Daily number of procedures 2 (once in the morning, once in the evening).
2. Manual start.
Scheme 14. Treatment of cystitis

1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2

L Double vibrophone
R Double vibrophone
Energy
Duration,
L
R
minutes
Zone Zone

4
3
1
5
3
2
5
3
2
5
3
2
5
3
2
5
3
2
5
3
2
5
3
2
5
3
3
5
3
3
5
3
3
5
3
3
5
3
3
5
3
3
5
3
3
6
3
3
6
3
3
6
3
3
6
3
3
6
3
3
6
3
3
6
3
3
6
3
3
6
3
3
6
3
3
6
3
3
6
3
3
6
3
3
6

Duration of the treatment course 14 days. The course of treatment may be repeated after 2 weeks. For preventive
purposes the course may be performed once or twice a year.
During treatment one should be in the supine position. The bladder should be empty.
Every procedure consists of 3 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the bladder zone (lower abdomen).
Start the session. When the session is over start the session 2.
Session 2. Change the converter and adjust the data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the bladder zone (lower abdomen).
Start the session. When the session is over start the session 3.
Session 3. Enter correct data for the Session 3. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 4.
Start the session.
- 31 -

ADENOMA (BENIGN PROSTATIC HYPERPLASIA)


The apparatus should be used according to the recommendations of the urologist for the treatment of benign
prostatic hyperplasia of the first (sometimes also the second) stage (residual urine not more than 150 mm) to reduce
painful sensations in the perineum zone and to reduce urination frequency.
Contraindication: prostate cancer.
The first treatment course the preparative one should be performed in accordance with the scheme 15. After a break
of one week the second treatment course the basic one should be performed in accordance with the scheme 16. To
strengthen the effect achieved, it is recommended to repeat the basic course after a break of two weeks. Then the patient should perform supportive course in accordance with the scheme 17.
During cold seasons, a patient should stay in the warm chamber for not less than an hour after the end of the treatment procedure. Throughout the course of treatment it is necessary to keep the pelvis area warm (warm clothes etc.).
Vibroacoustic therapy can be applied in combination with the drug treatment. In the case of specific infection, the
drug treatment course prescribed by the urologist should precede vibroacoustic treatment.
A patient should lie supine. If necessary, a small pillow can be placed under the pelvis to keep the horizontal position and to enhance blood circulation. It is advisable not to urinate before the treatment.
PREPARATIVE TREATMENT COURSE

1. Treatment should be performed in accordance with the scheme 15. Daily number of procedures 1.
2. Manual start.
3. Duration of the treatment course 15 days.
Every procedure consists of 2 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- Connected to the left port at the zone ;
- Connected to the right port the first vibrophone perineum zone (between the anus and the root of the penis),
another one bladder zone (lower abdomen).
Start the session. When the session is over start the session 2.
Session 2. Enter correct data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port as indicated in the treatment scheme, at the Zone 3 or 4.
Start the session.
Scheme 15. Treatment of prostate gland adenoma. Preparative course
Session 1
Session 2
Data entry
Data adjustment
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
Day
Energy
L
R
Duration,
Duration,
L
R
Mode
Mode
minutes
Zone Perineum and minutes
Energy at
Zone Energy

bladder zones
the Zone
1
1
2
1
10
2
2
3
1
5
2
2
2
2
10
2
2
4
1
5
3
2
3
3
10
2
3
3
1
10
4
2
3
3
10
2
3
4
1
10
5
2
3
3
15
2
3
3
2
10
6
2
3
3
15
2
3
4
2
10
7
2
3
3
15
2
3
3
2
10
8
2
3
3
15
2
3
4
2
10
9
2
3
3
15
2
3
3
2
10
10
2
3
3
15
2
3
4
3
10
11
2
3
3
15
2
3
3
2
10
12
2
3
3
15
2
3
4
3
10
13
2
3
3
15
2
3
3
2
10
14
2
3
3
15
2
3
4
3
10
15
2
3
3
15
2
3
3
2
10

- 32 -

BASIC TREATMENT COURSE

1. Treatment should be performed in accordance with the scheme 16. Daily number of procedures 1.
2. Manual start.
3. Duration of the treatment course -15 days.
Every procedure consists of 2 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone 4;
- connected to the right port the first vibrophone perineum zone (between the anus and the root of the penis),
another one bladder zone (lower abdomen).
Start the session. When the session is over start the session 2.
Session 2. Enter correct data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 3.
Start the session.

Day
Mode
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15

3
3
3
4
4
4
4
4
4
4
4
4
4
4
4

Scheme 16. Treatment of prostate gland adenoma. Basic course


Session 1
Session 2
Data entry
Data adjustment
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
Energy
Energy
Duration,
Duration,
L
R
L
R
Mode
minutes
Zone Perineum and minutes
Zone
Zone
4
bladder zones

3
1
2
10
2
2
2
10
1
2
10
2
2
2
10
1
3
10
2
3
2
10
1
3
10
2
3
2
10
1
3
15
2
3
2
10
2
3
15
2
3
2
10
2
3
15
2
3
2
11
2
3
15
2
3
2
12
2
3
15
2
3
2
13
2
3
15
2
3
2
14
2
3
15
2
3
2
15
2
3
15
2
3
2
15
2
3
15
2
3
2
15
2
3
15
2
3
2
15
2
3
15
2
3
2
15
SUPPORTIVE TREATMENT COURSE

1. Treatment should be performed in accordance with the scheme 17, once-twice a week.
2. Manual start.
The procedures should be performed at the same way as during the basic treatment course.

Mode
4

Scheme 17. Treatment of prostate gland adenoma. Supportive course


Session 1
Session 2
Data entry
Data adjustment
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
Energy
Energy
L
R
L
R
Duration,
Duration,
Mode
Perineum
minutes
minutes
Zone Zone
Zone
and bladder

3
4
zones
2
3
15
2
3
2
15

- 33 -

1
2
3
4-12
13
14

1
2
1
2
1
2
1
2
1
2
1
2

- 34 -

Duration,
minutes

Mode

Duration,
minutes

Mode

Duration,
minutes

Scheme 18. Treatment of serous mastitis and lactostasis


Session 1
Session 2
Session 3
Repetition of the
Data entry
Data adjustment
program entered
L Double vibrophone
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
R Double vibrophone
Energy
Energy
Energy
L
R
L
R
L
R
Zone Zone
Zone Zone
Zone Inflammation

D1

zone
1
2
1
5
1
2
1
5
1
3
1
1
3
2
5
1
3
2
5
1
3
2
2
3
2
5
2
3
2
5
2
3
2
2
3
2
5
2
3
2
5
2
3
2
5
3
2
5
5
3
2
5
5
3
2
5
3
2
5
5
3
2
5
5
3
2
1
3
2
5
1
3
2
5
1
3
2
1
3
2
5
1
3
2
5
1
3
2
2
3
2
5
2
3
2
5
2
3
2
2
3
2
5
2
3
2
5
2
3
2
5
3
2
5
5
3
2
5
5
3
2
5
3
2
5
5
3
2
5
5
3
2

Mode

Day

Number of procedure

SEROUS MASTITIS, MANIFESTATIONS OF LACTOSTASIS


Attention! It is forbidden to apply the apparatus for the treatment of suppurative mastitis.
It is recommended to start the treatment course by VITAFON-2 under the supervision of a doctor according to his
prescriptions.
1. Treatment and preventive therapy should be performed in accordance with the scheme 18, if doctor does not prescribe some other scheme. Daily number of procedures 2 (once in the morning, once in the evening).
2. Manual start.
3. Duration of the treatment course 14 days.
A patient should lie supine.
Every procedure consists of 3 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone D1.
Start the session. When the session is over start the session 2.
Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 1.
Start the session by pressing the button START for the main parameters for the sessions 1 and 2 are the same.
When the session is over start the session 3.
Session 3. Enter correct data for the Session 3. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port inflammation zone (for the treatment of lactostasis and prevention of mastitis one
vibrophone at each breast, for the treatment of mastitis both vibrophones at the area of induration localization).
Start the session.

1
3
5
5
6
8
8
8
5
5
3
3

1
2
3
4-19
20
21

1
2
1
2
1
2
1
2
1
2
1
2

- 35 -

Duration,
minutes

Mode

Duration,
minutes

Mode

Duration,
minutes

Scheme 19. Treatment of trophic ulcers


Session 1
Session 2
Session 3
Change of converters and data Repetition of the program enData entry
adjustment
tered
L Double vibrophone
L Double vibrophone
L Double vibrophone
R IR-matrix
R Double vibrophone
R Double vibrophone
Energy
Energy
Energy
L
R
L
R
L
R
Zone
Zone
Zone
Zone
Ulcer zone

D2 or 3

1 or 4
1
1
10
1
1
10
1
3
1
6
1
3
1
6
2
1
10
2
1
10
2
3
2
7
2
3
2
7
5
1
10
5
2
10
5
3
2
8
5
3
2
8
2
2
10
2
2
15
2
3
2
10
2
3
2
10
5
1
10
5
1
15
5
3
2
8
5
3
2
8
1
1
10
1
1
10
1
3
2
6
1
3
2
6

Mode

Day

Number of procedure

TROPHIC ULCERS
Attention! Do not use any ointments during the treatment with the apparatus. The treatment should not be
performed in body zones affected by trombophlebitis.
1. Treatment should be performed in accordance with the scheme 19, if doctor does not prescribe some other scheme.
Daily number of procedures 2 (once in the morning, once in the evening).
2. Manual start.
Duration of the treatment course 21 days. If necessary, the course of treatment may be repeated after a week.
A patient should lie supine. Before a treatment procedure, treat the edges of the ulcer with iodine solution, in 3 cm.
of diameter.
Treatment zones are K, the ulcer and tissues nearby; zones D2 and E1 for ulcers on upper extremities and zones E3
and E4 for ulcers on lower extremities.
For the treatment of the ulcer zone itself, IR-matrix should be placed exactly on the ulcer. For disinfecting one should
moisten a napkin with 3% hydrogen peroxide solution, wring it out and wash the IR-matrix before and after the treatment.
Vibrophones are applied to the edges of the affected area (in 1-2 cm. from the ulcer) covered with a gauze, paper or
cotton towel.
Every morning procedure consists of 1 session, every evening procedure of 3 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed at the ulcers zone.
Start the session. When the session is over start the session 2.
Session 2. Change the converters and adjust data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port for ulcers on upper extremities at the Zone D2, for ulcers on lower extremities at the
Zone 3.
Start the session. When the session is over start the session 3.
Session 3. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port for ulcers on upper extremities at the Zone 1, for ulcers on lower extremities at the
Zone 4.
Start the session by pressing of the button START, for the main parameters of the sessions 2 and 3 are the same.

INSOMNIA
1. Treatment should be performed in accordance with the scheme 20, variation 1 or 2 (if the patient suffers also from
thorax osteochondrosis or hypertension). Daily number of procedures 1.
2. Manual start.
The procedure is to be performed before going to bed and should be started as follows:
Enter correct data. Converters should be fixed as follows:
Variation 1:
- at the Zone ;
Variation 2:
- connected to the left port at the Zone ;
- connected to the right port at the Zone D1.
Start the session.
Scheme 20. Treatment of insomnia. Variation 1
L Double vibrophone, R no
L
Duration,
Mode
Energy
minutes
Zone
2
3
15
Scheme 20. Treatment of insomnia. Variation 2
L Double vibrophone, R Double vibrophone
Energy
Duration,
Mode
L
R
minutes
Zone
Zone D1
2
3
1
15

- 36 -

PYELONEPHRITIS
In the case of kidney stones (not more than 4 mm. in size) the treatment can be performed only after consulting
with a doctor and under his supervision.
1. Treatment should be performed in accordance with the schemes 21 or 22, if doctor does not prescribe some other
scheme. Daily number of procedures 2 (once in the morning, once in the evening).
2. Manual start.
3. Duration of the treatment course 14 (21) days.
A patient should lie supine.
Vibroacoustic treatment should be applied in combination with drug therapy prescribed by the doctor. In the case of
nephroptosis, vibrophones should be applied 3-5 cm. below.
Every procedure consists of 2 sessions.
Scheme 21. Treatment of pyelonephritis
Session 1
Session 2
Data entry
Data adjustment
L Single vibrophone
L Double vibrophone
Number of
R IR-matrix
R No
Day
procedure
Energy
Energy
Duration,
Duration,
Mode
Mode
L
R
L
minutes
minutes
Zone
Zone
1

1
2
1
1
2
1
1
6
1
1
1
3
1
1
7
2
2
1
1
4
1
1
8
1
1
1
5
1
1
9
3
2
1
1
6
1
1
10
1
5
2
2
5
2
6
4
2
2
2
3
2
2
6
1
5
2
4
5
2
7
5
2
2
2
5
2
2
7
1
5
2
6
5
2
8
6
2
2
2
7
2
2
8
1
5
2
7
5
2
9
7
2
2
2
7
2
2
9
1
2
2
7
2
2
10
8
2
2
2
7
2
2
10
1
2
2
7
2
3
6
9
2
2
2
7
2
3
6
1
2
2
7
2
3
7
10
2
2
2
7
2
3
7
1
2
2
7
2
3
8
11
2
2
2
7
2
3
8
1
2
2
7
2
3
9
12
2
2
2
7
2
3
10
1
2
2
7
2
3
11
13
2
2
2
7
2
3
12
1
2
2
7
2
3
13
14
2
2
2
7
2
3
14
1
2
2
7
2
3
15
15
2
2
2
7
2
3
16
1
2
2
7
2
3
17
16
2
2
2
7
2
3
18
1
2
2
7
2
3
20
17-21
2
2
2
7
2
3
20

- 37 -

Session 1. Enter correct data for the Session 1. Both converters should be fixed at the Zone .
During treatment course a vibrophone and an IR-matrix are to be replaced with each other for every new
procedure.
Start the session. When the session is over start the session 2.
Session 2. Change of converters and data adjustment for the Session 2. Converter should be fixed at the Zone .
Start the session.
After the end of the treatment one should stay in the warm chamber for at least 30 minutes.
In the case of chronic pyelonephritis and in order to prevent complications it is advisable to perform 2-3 courses of
treatment with an interval of 7 days after each finished course.
Repeated courses of treatment should take place at the stage of remission in accordance with the scheme 22 with no
drugs taken.

Day

1
2
3
4
5
6
7
8
9
10
11
12
13
14

Scheme 22. Treatment of pyelonephritis in the remission


Session 1
Session 2
Data entry
Data adjustment
L Single vibrophone
L Double vibrophone
Number of
R IR-matrix
R No
procedure
Energy
Energy
Duration,
Duration,
Mode
Mode
L
R
L
minutes
minutes
Zone
Zone
1

2
2
2
3
2
2
10
1
2
2
3
5
2
2
15
1
2
2
2
7
2
2
20
1
2
2
2
7
2
3
22
1
2
2
2
7
2
3
24
1
2
2
2
7
2
3
26
1
2
2
2
7
2
3
28
1
2
2
2
7
2
3
30
1
2
2
2
7
2
3
30
1
2
2
2
7
2
3
30
1
2
2
2
7
2
3
30
1
2
2
2
7
2
3
30
1
2
2
2
7
2
3
30
1
2
2
2
7
2
3
30

- 38 -

BEDSORES (PREVENTION AND TREATMENT)


Prevent the contact of wet discharges with the apparatusess converters during the treatment procedures.
1. Preventive treatment of bedsores
1. Treatment should be performed in accordance with the schemes 23 or 24, if doctor does not prescribe some other
scheme. Daily number of procedures in accordance with the scheme 23 3 (once in the morning, once in the noon
and once in the evening).
2. Automatical start.
3. Duration of the treatment course 3 weeks. The treatment course should be repeated after a break of 5 days.
The vibroplate should be formed in two raws (see Picture 7) and put under the pelvis zone and lower part of the
back; the upper end of it should be fixed at the Zone .

Picture 7. Application of the vibroplate for the prevention of bedsores.


Treatment in accordance with the scheme 23 should be performed as follows:
For the days 1 3 one should every day enter correct data for three daily procedures in the automatical mode.
When all the necessary parameters are entered, start the session. Then the apparatus will be automatically switched on
and off according to the parameters entered.
The parameters of the treatment procedures for days 4 21 are the same. After data entry for the day 4 the apparatus should not be disconnected from the mains. The apparatus will be automatically switched on and off.

Day
1
2
3
4-21

Scheme 23. Preventive treatment of bedsores


L Vibroplate; R No
Number of
Energy
Duration,
procedures Mode
L
minutes
Zone and pelvis zone
3
6
1
40
3
6
2
40
3
6
3
40
3
2
4
40

Scheme 24. Preventive treatment of bedsores; periodical start mode


L Vibroplate; R No
Energy
Day
Duration,
Mode
L
hours
Zone and pelvis zone
1
Auto 8
1
14
2
Auto 8
2
14
3
Auto 8
3
14
4-21 Auto 8
4
14

Start the session. If after the treatment time, indicated in the treatment scheme, the procedure is finished by pressing of the button RESET, the parameters programmed will be lost; if it is finished by switching off of the apparatus,
the parameters of the last procedure will be kept.
2. Treatment of bedsores
For the treatment of existing bedsores, the procedures should be performed right at the bedsores zones to quicken
the process of discharge of metabolism products and at the liver zone to improve detoxication from the slags, which
can come into blood together with the lymph from the bedsore zone.
1. Treatment should be performed in accordance with the scheme 25, if doctor does not prescribe some other scheme.
Daily number of procedures 3 (once in the morning, once in the noon and once in the evening).
- 39 -

2. Automatical start.
3. Duration of the treatment course 2 weeks.
Every procedure consists of 2 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
connected to the left port at the Zone 4 right. (Both vibraphones are to be placed at the same zone, not far from
each other);
- connected to the right port at the Zone 32.
Start the session. When the session is over start the session 2.
Session 2. Enter correct data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the zone of two biggest bedsores (one vibraphone per particular bedsore area).
Start the session.

Day

1
2
3
4
5
6
7-14

Scheme 25. Treatment of bedsores


Session 1
Session 2
Data entry
Data adjustment
L Double vibrophone
L Double vibrophone
Daily number
R Double vibrophone
R Double vibrophone
of procedures
Energy
Energy
Duration,
L
R Duration,
L
R
Mode
Mode
Zone
Zone minutes
Zone Inflammation minutes
4 right 32

zone
3
2
3
3
3
2
3
3
5
3
2
3
3
5
2
3
3
7
3
2
3
3
7
2
3
3
10
3
2
3
3
9
2
3
3
13
3
2
3
3
11
2
3
3
15
3
2
3
3
13
2
3
3
17
3
2
3
3
15
2
3
3
20
Then the patient should continue treatment according to the Scheme 23 or 24.

CONSTIPATION
1. Treatment should be performed in accordance with the scheme 26, if doctor does not prescribe some other scheme.
Daily number of procedures 2 (once in the morning, once in the evening).
2. Manual start.
3. Duration of the treatment course 21 days.
A patient should lie supine. If necessary, a small pillow can be placed under pelvis to keep the horizontal position.
In the case of a single constipation, it will be enough to perform 3-4 procedures.
If constipation lasts more than 4 days it is advisable to give an enema before the treatment in order not to traumatize bowels.
In the case of chronic constipation it is recommended to perform three full treatment courses with a break of 2-3
days after every finished course.
If vibroacoustic treatment by VITAFON-2 takes place in complex with drug therapy, a daily drug dose of purgatives is to be gradually reduced from the third day of treatment.
Every procedure consists of 3 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 3.
Start the session. When the session is over start the session 2.
Session 2. Enter correct data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the zone, indicated in the treatment scheme, either 1 or 2.
Start the session. When the session is over start the session 3.
Session 3. Enter correct data for the Session 3. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the zone, indicated in the treatment scheme 32, 30, 3 or 4.
Start the session.
- 40 -

2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21

1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2

- 41 -

1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5
1
1
2
2
5
5

Duration,
minutes

Energy

Session 3
Data adjustment
L Double vibrophone
R Double vibrophone
L
R
Energy at
the Zone

3
32 1
3
30 1
3
3
1
3
4
1
3
32 1
3
30 1
3
3
1
3
4
1
3
32 1
3
30 1
3
3
1
3
4
1
3
32 1
3
30 1
3
3
1
3
4
1
3
32 1
3
30 1
3
3
1
3
4
1
3
32 1
3
30 1
3
3
1
3
4
1
3
32 1
3
30 1
3
3
1
3
4
1
3
32 1
3
30 1
3
3
1
3
4
1
3
32 1
3
30 1
3
3
1
3
4
1
3
32 1
3
30 1
3
3
1
3
4
1
3
3
Zone

Mode

Duration,
minutes

Energy

Zone

Mode

Duration,
minutes

Mode

Number of procedure

Day
1

Scheme 26. Treatment of constipation


Session 1
Session 2
Data entry
Data adjustment
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
Energy
L
R
L
R
Energy
at the
Zone Zone
Zone

3
1
3
1
3
1
3
1
3
3
1
3
1
3
1
3
2
3
3
2
3
1
3
2
3
1
3
3
2
3
1
3
2
3
2
3
3
5
3
1
4
5
3
1
3
3
5
3
1
4
5
3
2
3
3
1
3
1
4
1
3
1
3
3
1
3
1
4
1
3
2
3
3
2
3
2
4
2
3
1
4
4
2
3
2
4
2
3
2
4
4
5
3
2
5
5
3
1
4
4
5
3
2
5
5
3
2
4
4
1
3
2
5
1
3
1
4
4
1
3
2
5
1
3
2
4
4
2
3
2
5
2
3
1
4
4
2
3
2
5
2
3
2
4
4
5
3
2
6
5
3
1
3
5
5
3
2
6
5
3
2
3
5
1
3
2
6
1
3
1
3
5
1
3
2
6
1
3
2
3
5
2
3
2
6
2
3
1
3
5
2
3
2
6
2
3
2
3
5
5
3
2
7
5
3
1
3
5
5
3
2
7
5
3
2
3
5
1
3
2
7
1
3
1
3
6
1
3
2
7
1
3
2
3
6
2
3
2
7
2
3
1
3
6
2
3
2
7
2
3
2
3
6
5
3
2
8
5
3
1
3
6
5
3
2
8
5
3
2
3
6
1
3
2
8
1
3
1
3
7
1
3
2
8
1
3
2
3
7
2
3
2
8
2
3
1
3
7
2
3
2
8
2
3
2
3
7
5
3
2
9
5
3
1
3
7
5
3
2
9
5
3
2
3
7
1
3
2
9
1
3
1
3
8
1
3
2
9
1
3
2
3
8
2
3
2
10
2
3
1
3
8
2
3
2
10
2
3
2
3
8
5
3
2
10
5
3
8
5
3
2
10
5
3
8

3
3
3
3
3
3
3
3
4
4
4
4
4
4
4
4
5
5
5
5
5
5
5
5
6
6
6
6
6
6
6
6
7
7
7
7
7
7
7
7
7
7

ENURESIS (children urine incontinence)


1. Treatment should be performed in accordance with the scheme 27, if doctor does not prescribe some other scheme.
Daily number of procedures 3 (for the children younger than 10 years) or 2 (for the children older than 10 years).
2. Manual start.
3. Duration of the treatment course 14 days. To achieve stable effect one should perform 2-3 courses of treatment
with a break of two weeks after every finished treatment course.
A patient should lie supine on a soft bad.
The treatment should start as follows:
Enter correct data. Converters should be fixed as follows:
- connected to the left port at the Zone 3;
- connected to the right port at the bladder zone (lower abdomen).
Start the session.

Day

1-5

6-10
11-14

Number of
Mode
the procedure
1
2
3
1
2
3
1
2
3

1
1
1
3
3
2
4
4
2

Scheme 27. Treatment of enuresis


L Double vibrophone
Duration, minutes
R Double vibrophone
Energy
Up to 5
From 5 up to From 10 years
L
R
years old
10 years old
old and older
Zone 3
Bladder zone
1
1
3
5
10
1
2
3
5
1
2
4
5
10
1
2
3
5
10
1
2
4
5
2
2
4
5
10
1
3
3
5
10
1
3
4
5
2
3
4
5
10

ENCOPRESIS (children faeces incontinence)


Treatment by VITAFON-2 goes well with other types of physiotherapy.
1. Treatment should be performed in accordance with the scheme 28, if doctor does not prescribe some other scheme.
Daily number of procedures 3 (for the children younger than 10 years) or 2 (for the children older than 10 years).
2. Manual start.
3. Duration of the treatment course 15 days. To achieve stable effect one should perform 2-3 courses of treatment
with a break of two weeks after every finished treatment course.
A patient should lie supine on a soft bad.
The treatment should start as follows:
Enter correct data. Converters should be fixed as follows:
- connected to the left port at the Zone 3;
- connected to the right port at the Zone 4.
Start the session.

Day

1-5
6-10
11-15

Number of
the procedure

Mode

1
2
3
1
2
3
1
2
3

1
1
1
3
3
2
4
3
2

Scheme 28. Treatment of encopresis


L Double vibrophone
R Double vibrophone
Energy
Up to 5
years old
L
R
Zone 3
Zone 4
1
2
3
1
2
3
1
2
4
2
2
3
2
2
4
2
2
4
1
3
4
2
3
4
2
3
4

- 42 -

Duration, minutes
From 5 up to From 10 years
10 years old old and older
5
5
5
5
5
5
5
5
5

5
5
7
7
8
8

PERIODONTITIS, PARADONTOSIS, STOMATITIS,


RECOVERY AFTER DENTAL SURGICAL OPERATIONS
VITAFON-2 is used to ameliorate pain and swelling, reduce dental loosening via blood supply restoration in the inflammation zone; it also shortens time of healing and helps to avoid complications after medical manipulations, including dental surgery.
PERIODONTITIS, PARADONTOSIS, STOMATITIS, RECOVERY AFTER DENTAL SURGICAL OPERATIONS

In the case of periodontitis and paradontosis, the apparatus should be applied as an additional therapy during the
traditional treatment course. It should be used after removing of subgingival plague.
In the case of stomatitis, mucous tunic of the oral cavity should be previously treated with the Lugol solution.
1. Treatment should be performed in accordance with the scheme 29, if doctor does not prescribe some other scheme.
Daily number of procedures 2 (once in the morning, once in the evening).
2. Manual start.
3. Duration of the treatment course 10-14 days.
Scheme 29. Treatment of periodontitis, paradontosis, stomatitis; recovery after dental surgical operations
L Double vibrophone
R Double vibrophone
Number of
Energy
Day
procedure
Duration,
L
R
Mode
minutes
Zone Inflammation zone
1
2
3
1
5
1
2
2
3
1
7
1
2
3
2
7
2
2
2
3
2
10
1
2
3
2
10
3
2
2
3
2
10
4-10
1
2
3
3
10
(4-14)
2
2
3
3
10
The treatment should start as follows: Enter correct data. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the pathology zone (in the case of periodontitis and paradontosis in the zone of
inflamed gum tissues; in the case of stomatitis at the projection of inflamed mucous tunic to the face surface; in
the case of surgical operation in the operation zone).
Start the session.
RECOVERY AFTER TOOTH REPLACEMENT, TOOTH CANAL CLEANING OR TOOTH FILLING

The apparatus can be used in order to prevent complications after making a prosthetic appliance for the teeth (tooth
replacement), filling a tooth or root canal cleaning. In such case the procedures by VITAFON-2 should start in 6 hours
after completion of the treatment mentioned.
1. Treatment should be performed in accordance with the scheme 30, if doctor does not prescribe some other scheme.
Daily number of procedures 2 (once in the morning, once in the evening).
2. Manual start.
3. Duration of the treatment course 2-3 days.
Scheme 30. Treatment after tooth replacement, root canal cleaning or tooth filling
L Double vibrophone
R Single vibrophone
Number of
Day
Energy
procedure
Duration,
Mode
L
R
minutes
Zone
Dental root zone
1
2
3
2
5
1
2
2
3
2
10
1
2
3
2
15
2-3
2
2
3
2
15
The treatment should start as follows: Enter correct data. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port over the root of a bad tooth.
Start the session.
- 43 -

ARTHRITIS, ARTHROSIS, RHEUMATIC POLYARTHRITIS


Duration of the treatment course for one joint is three weeks. In case of neglected disease it is advisable to perform
3-4 treatment courses with an interval of 1-2 weeks.
If a patient feels strong pain during the treatment, treatment duration is to be reduced. It is not recommended to
treat several joints at a time.
If after treatment procedures pressure elevation, giddiness or weakness occurs, the treatment is to be interrupted for
a diagnosis confirmation.
Vibrophones are applied to a joint at its most flat part at an angle of approximately 90 to each other. (Picture 5).
A patient should lie supine. Before the procedure, a joint should be treated with an anti-inflammation agent (if you
apply iodine solution, keep in mind that it can be used once a day, right before the start of treatment and only if there
are no traces of previously applied iodine on the skin).
ARTHRITIS

1. Treatment should be performed in accordance with the scheme 31. Daily number of procedures 2.
2. Manual start.
3. Duration of the treatment course 21 days.
Every morning procedure consists of one (second) session, every evening one of two sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the joint zone.
Start the session. When the session is over start the session 2.

Day

1
2
3
4-18
19
20
21

Number of procedure

Session 2. Change the converters and enter correct data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the joint zone.
Start the session.

1
2
1
2
1
2
1
2
1
2
1
2
1
2

Scheme 31. Treatment of arthritis


Session 1
Data entry

Mode
1
2
5
1
2
5
5

L Double vibrophone
R IR-matrix
Energy
Duration,
L
R
minutes
Zone Joint zone
3
3
7
3
3
8
3
3
9
3
3
10
3
3
10
3
3
10
3
3
10

- 44 -

Mode
1
1
2
2
5
5
1
1
2
2
5
5
5
5

Session 2
Change of converter
and data adjustment
L Double vibrophone
R Double vibrophone
Energy
Duration,
L
R
minutes
Zone Joint zone
3
3
10
3
3
10
3
4
10
3
4
15
3
4
10
3
4
15
3
4
15
3
4
15
3
4
15
3
4
15
3
4
15
3
4
15
3
4
15
3
4
15

ARTHROSIS AND RHEUMATIC POLYARTHRITIS

1
2
3
4-18
19
20
21

- 45 -

Duration,
minutes

Mode

Duration,
minutes

1
1
2
2
5
5
1
1
2
2
5
5
5
5

Mode

1
2
1
2
1
2
1
2
1
2
1
2
1
2

Scheme 32. Treatment of arthrosis, rheumatic polyarthritis


Session 1
Session 2
Session 3
Repetition of the
Data entry
Data adjustment
program entered
L Double vibrophone
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
R Double vibrophone
Energy
Energy
Energy
L
R
L
R
L
R
Zone
Zone
Zone
Zone
Zone
Joint

3 (D2)

32 (1)

Zone
3
2
3
1
3
2
3
1
3
4
7
3
2
3
1
3
2
3
1
3
4
8
3
2
3
2
3
2
3
2
3
4
9
3
2
3
2
3
2
3
2
3
4
10
3
2
4
5
3
2
4
5
3
4
10
3
2
4
5
3
2
4
5
3
4
12
3
2
4
1
3
2
4
1
3
4
10
3
2
4
1
3
2
4
1
3
4
15
3
2
4
2
3
2
4
2
3
4
15
3
2
4
2
3
2
4
2
3
4
15
3
2
4
5
3
2
4
5
3
4
15
3
2
4
5
3
2
4
5
3
4
15
3
2
4
5
3
2
4
5
3
4
15
3
2
4
5
3
2
4
5
3
4
15
Duration,
minutes

Mode

Day

Number of procedure

Treatment should be performed in accordance with the scheme 32. Daily number of procedures 2 (once in the
morning, once in the evening).
1. Manual start.
2. Duration of the treatment course 21 days.
For the treatment of lower extremities joints (hip, knee, ankle, small foot joints etc.) treatments zones are as follows:
, 3, 32 and a painful joint itself.
For the treatment of upper extremities joints (shoulder, elbow, wrist, small hand joints etc.) treatments zones are as follows: , D2, 1 and a painful joint itself.
Every procedure consists of three sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 3 (D2).
Start the session. When the session is over start the session 2.
Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 32 (1).
Start the session by pressing of the button START, for the main parameters of the sessions 1 and 2 are the same.
When the session is over start the session 3.
Session 3. Enter correct data for the Session 3. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the joint zone.
Start the session.

COMPLICATIONS OF SCOLIOSIS
1. Treatment should be performed in accordance with the scheme 33, if doctor does not prescribe some other scheme.
Daily number of procedures 2 (once in the morning, once in the evening).
2. Manual start.
3. Duration of the treatment course 10-14 days.
A patient should lie supine on a soft bed.
The treatment should start as follows:
Enter correct data. Converters should be fixed to a painful spine area symmetrically relative to the spinal column at
the distance of 3-5 cm. from it.
Start the session.

Day

1-2
3-10
(3-14)

Number of procedure

Scheme 33. Treatment of complications of scholiosis

1
2
1
2

L Double vibrophone
R Double vibrophone
Energy
Mode

Painful zone
2
2
2
2

3
3
3
3

3
3
3
3

Duration,
minutes
15
15
20
20

TRAUMAS OF THE SPINAL COLUMN


1. Treatment should be performed in accordance with the scheme 34, if doctor does not prescribe some other scheme.
Daily number of procedures 3 (once in the morning, once in the noon and once in the evening).
2. For the session 1 Manual start or, in the case of bed patients Auto start; for the session 2 Manual start.
3. Duration of the treatment course 4 months.
To reduce inflammation process, the trauma zone should be treated with iodine solution (2-3 cm.). One should keep
in mind that iodine treatment might be performed only if there are no traces of previously applied iodine on the skin.
The duration of iodine treatment course 5-10 days.
It is usually very difficult to apply converters in the cases of spine traumas, especially during the first 30 days. One
can fix at the patient a set of three joint converters. Choosing of the necessary converter will be performed by the
change of cords switched on.
Every procedure consists of two sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left and right ports at the trauma zone, symmetrically concerning the spinal column within the
fracture zone to treat the maximum area.
Start the session (if Auto start option is chosen, the first, second and third procedures will start in automatic mode;
the apparatus should not be switched off).
Session 2. Change of right converter and data adjustment for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port, the vibroplate at the zone of probable or existing bedsores.
Start the session in the manual mode. When the session is over, adjust the data for auto start of the first procedure
of the next day, if it is to be performed in the automatic mode. The apparatus should not be switched off.

- 46 -

1
2
3
4
5
6
7
8
9
10
11-31

Number of procedure

Day

1
2
3
1
2
3
1
2
3
1
2
3
1
2
3
1
2
3
1
2
3
1
2
3
1
2
3
1
2
3
1
2
3

32-34
35-61

1
2
3

62-64
65-91

1
2
3

92-94
95-120

1
2
3

Scheme 34. Treatment of spinal cord traumas


Session 1, manual or auto start
Session 2, manual start
Change of converter
Data entry
and data adjustment
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Vibroplate
Energy
Energy
Duration,
Duration,
Mode
Mode
L
R
L
R
minutes
minutes
Zone Bedsores zone
Trauma zone
1
1
1
3
5
1
1
3
2
1
1
3
2
3
2
10
1
1
1
3
5
1
1
4
2
1
1
4
2
3
2
15
1
1
1
4
5
1
1
4
2
1
1
5
2
3
2
20
1
1
1
5
5
1
1
5
2
1
1
6
2
3
2
25
1
1
1
6
5
1
1
6
2
1
1
6
2
3
2
30
1
1
1
6
5
1
1
7
2
1
1
7
2
3
2
35
1
1
1
7
5
1
1
7
2
1
1
8
2
3
2
40
1
1
1
8
5
1
1
8
2
1
1
8
2
3
2
40
1
1
1
8
5
1
1
9
2
1
1
9
2
3
2
40
1
1
1
9
5
1
1
9
2
1
1
10
2
3
2
40
1
1
1
10
5
2
1
10
2
1
2
10
2
3
2
40
Break
1
2
2
11
5
1
1
11
2
2
2
11
2
3
2
40
Break
1
2
2
12
5
1
1
12
2
2
2
12
2
3
2
40
Break
1
2
2
14
5
1
1
14
2
2
2
14
2
3
2
40

- 47 -

WOUNDS AND SUTURES


VITAFON-2 is used to shorten treatment, accelerate forming of epidermis and to prevent complications and scars.
Treatment should start in 24 hours after appearing of the wound and its prior surgical treatment (pus and dirt removal, washing the wound with a Furacillin solution, treatment of wound edges with iodine solution and sterile bandage). If there is a bandage, vibrophones should be placed directly to the wound area on the bandage.
A patients position depends on his condition and a wound location.
Vibrophones may be applied directly to the dressing over the wound or to the suture.
1. Treatment should be performed in accordance with the scheme 35, if doctor does not prescribe some other scheme.
Daily number of procedures 1-2.
2. Manual start.
3. Duration of the treatment course 7 days.
The treatment should start as follows:
Enter correct data. Converters should be fixed at the wound zone.
Start the session.

Day
1
2
3-7

Scheme 35. Treatment of wounds and surgical sutures


L Double vibrophone
R Double vibrophone
Daily number
Energy
of procedures
Duration,
Mode
L
R
minutes
Wound zone
1-2
1
2
2
5
1-2
2
2
2
10
1-2
5
2
2
15

Treatment of complicated wounds and large surgical sutures consists of two or three daily procedures in accordance with the severity of the patients state. Every procedure consists of 2 sessions. During the session 2, location of
vibrophones is to be changed concerning a wound area. Duration of the treatment in the case of complicated wounds
should last till the full recovery.
1. Treatment should be performed in accordance with the scheme 36, if doctor does not prescribe some other scheme.
Daily number of procedures 2-3.
2. Manual start.
3. Duration of the treatment course 14 days or until the full recovery.
Every procedure consists of 2 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the wound zone.
Start the session. When the session is over start the session 2.
Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the wound zone, location of vibrophones is to be changed concerning a wound area.
Start the session by pressing of the button START, for the main parameters of the sessions 1 and 2 are the same.

Day

1
2
3-14

Scheme 36. Treatment of complicated wounds and surgical sutures


Session 1
Session 2
Data entry
Repetition of the program entered
L Double vibrophone
L Double vibrophone
Daily number
R Double vibrophone
R Double vibrophone
of procedures
Energy
Energy
Duration,
Duration,
L
R
L
R
Mode
Mode
minutes
minutes
Zone
Wound
Zone
Wound

zone

zone
2
1
3
2
5
1
3
2
5
2-3
2
3
2
7
2
3
2
7
2-3
5
3
2
9
5
3
2
9

- 48 -

DISLOCATIONS AND STRAINS


One should start treatment with the apparatus not earlier than 8-10 hours after injuring and after prior treatment (the
joint should be in its original position).
1. Treatment should be performed in accordance with the scheme 37, if doctor does not prescribe some other scheme.
Daily number of procedures 2.
2. Manual start.
3. Duration of the treatment course 7-21 days.
A patient should lie supine.
Treatment should be performed directly at the trauma zone, vibrophones are applied to the smoother patches of the
area. In the case with an extremity trauma, the extremity should be up, higher than the heart level.
When the procedure is over, a patient should lie quietly for not less than half an hour.
To prevent inflammation, it is advisable to treat the trauma zone with iodine solution from the 3-d up to the 7-th
days of the treatment.
The treatment should start as follows:
Enter correct data. The converters should be fixed at the trauma zone. Start the session.

Day

1
2
3-7
(3-21)

Scheme 37. Treatment of dislocations and strains


L Double vibrophone
R Double vibrophone
Number of
Energy
procedure
Duration,
Mode
L
R
minutes
Trauma zone
1
1
2
2
20
2
2
3
3
20
1
5
3
3
20
2
4
3
3
30
1
4
4
4
30
2
4
4
4
30

MAXILLARY SINUSITIS
Attention! In the case of enlarged adenoids, the treatment must be performed under the doctors supervision.
In the case of acute maxillary sinusitis, clean the Highmore sinuses before using the apparatus (to be done by otolaryngologist).
The apparatus is also to be used exactly after surgical treatment. The apparatus can be used in combination with the
drug therapy prescribed by the doctor.
For the treatment of chronic maxillary sinusitis it is recommended to undergo 2-3 treatment courses with a break of
two weeks after every finished course.
In preventive purposes the treatment course must be repeated once-twice a year.
During the first 6-8 days of the treatment, the thorax should be treated with iodine solution before going to bed
(5% iodine solution, treatment square about two palms).
During treatment sessions a vibrophone and an IR-matrix are to be replaced with each other for every new procedure.
1. Treatment of maxillary sinusitis should be performed during remission in accordance with the scheme 38, if doctor
does not prescribe some other scheme. Daily number of procedures 2 (once in the morning, once in the evening).
2. Manual start.
3. Duration of the treatment course 14 days.
While the treatment of zones D1 and K, a patient should lie supine.
While the treatment of zone G, a patient should sit or recline.
Every procedure consists of 3 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed at the Zone G.
Start the session. When the session is over start the session 2.
Session 2. Enter correct data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone D1.
Start the session. When the session is over start the session 3.
Session 3. Enter correct data for the Session 3. Converter should be fixed at the Zone .
Start the session.
- 49 -

1
2
3
4
5
6
7
8
9
10
11
12
13
14

Number of procedure

Day

1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2

Scheme 38. Treatment of maxillary sinusitis


Session 1
Session 2
Session 3
Change of converters and
Change of converters and
Data entry
data adjustment
data adjustment
L Single vibrophone
L Double vibrophone
L Double vibrophone
R IR-matrix
R Double vibrophone
R No
Energy
Energy
Energy
Duration,
Duration,
Duration,
Mode L
Mode
Mode
R
L
R
L
minutes
minutes
minutes
Zone G
Zone Zone D1
Zone
1
2
5
1
2
5
2
3
3
3
2
3
7
2
2
6
2
2
6
2
3
3
5
2
3
7
5
2
6
5
2
6
2
3
3
5
2
3
7
1
2
7
1
2
7
2
3
3
7
2
3
7
2
2
7
2
2
7
2
3
3
7
2
3
7
5
2
9
5
2
9
2
3
3
7
2
3
7
1
2
9
1
2
9
2
3
3
7
2
3
7
2
2
12
2
2
12
2
3
3
7
2
3
7
5
2
12
5
2
12
2
3
3
7
2
3
7
1
2
12
1
2
12
2
3
3
7
2
3
7
2
2
12
2
2
12
2
3
3
7
2
3
7
5
2
12
5
2
12
2
3
3
7
2
3
7
1
2
12
1
2
12
2
3
3
7
2
3
7
2
2
12
2
2
12
2
3
3
7
2
3
7

RHINITIS (RUNNING NOSE)


With a timely treatment, it is possible to prevent the progression of the disease. In the course of treatment, a heavy
nasal discharge may sometimes occur for the period of 1-2 days. If on the 14th day of treatment nasal discharges are
not over, the patient should consult with an allergologist.
In the case of allergic rhinitis, the apparatus should be used only at the final stage of treatment in combination with
antiallergic preparations prescribed by the doctor.
Visible symptoms can be over already after 3-4 days of treatment but it is recommended to finish the full course in
order to prevent symptomatic relapse.
1. Treatment of rhinitis should be performed at the early stage of the disease in accordance with the scheme 39, if
doctor does not prescribe some other scheme. Daily number of procedures 2 (once in the morning, once in the
evening).
2. Manual start.
3. Duration of the treatment course 14 days.
Patients position:
- For the treatment at the Zone G a patient should seat or recline;
- For the treatment at the Zones , and 4 right a patient should lie supine.
Every procedure consists of 2 sessions.
- 50 -

Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone .
Start the session. When the session is over start the session 2.
Session 2. Enter correct data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone G;
- connected to the right port at the Zone 4 right (both vibrophones should be located near by).
Start the session.
It is recommended to perform a one-day preventive course in accordance with the scheme 40 (two sessions once
in the morning and once in the evening).

Day

1
2
3
4
5
6
7-14

Scheme 39. Treatment of rhinitis


Session 1
Session 2
Data entry
Data adjustment
L Double vibrophone
L Double vibrophone
Number of
R Double vibrophone
R Double vibrophone
procedure
Energy
Energy
Duration,
Duration,
L
R
L
R
Mode
Mode
minutes
minutes
Zone
Zone Zone
Zone G
4 right
1
1
1
1
5
1
1
2
4
2
1
1
1
10
1
1
2
5
1
2
2
1
10
2
1
2
6
2
2
2
1
10
2
1
2
7
1
5
3
1
10
5
2
2
8
2
5
3
1
10
5
2
2
9
1
1
3
1
15
1
2
2
10
2
1
3
1
15
1
2
2
10
1
2
3
1
15
2
2
2
10
2
2
3
1
15
2
2
2
10
1
5
3
1
15
5
2
2
10
2
5
3
1
15
5
2
2
10
1
1
3
2
15
1
2
2
10
2
1
3
2
15
1
2
2
10

Number of
procedure

1
2

Scheme 40. Preventive treatment of rhinitis


Session 1
Session 2
Data entry
Data adjustment
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
Energy
Energy
Duration,
Duration,
L
R
L
R
Mode
Mode
minutes
Zone Zone minutes
Zone
Zone

G
4 right
2
3
2
15
5
2
2
10
2
3
2
15
5
2
2
10

- 51 -

TONSILLITIS (SORE THROAT)


1. Treatment should be performed in accordance with the scheme 41, if doctor does not prescribe some other scheme.
Daily number of procedures 2 (once in the morning, once in the evening).
2. Manual start.
3. Duration of the treatment course 14 days.
Patients position:
- For the treatment at the Zone a patient should seat or recline;
- For the treatment at the Zones and a patient should lie supine.
Every procedure consists of 2 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone .
Start the session. When the session is over start the session 2.
Session 2. Enter correct data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone as close to the tonsils as possible. Patients head should be slightly
thrown back.
Start the session.
In acute tonsillitis, use the apparatus in combination with bactericidal preparations and under doctors supervision.
Before every procedure remove pus from the tonsils and treat the tonsils with Lugol solution.
In the case of chronic tonsillitis, the treatment should be performed at the stage of remission. To prevent exacerbation of chronic tonsillitis, the treatment course should be performed twice-trice a year.

Day

1
2
3
4
5
6
7-14

Number of
procedure

1
2
1
2
1
2
1
2
1
2
1
2
1
2

Scheme 41. Treatment on tonsillitis (sore throat)


Session 1
Session 2
Data entry
Data adjustment
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
Energy
Energy
Duration,
Duration,
Mode
Mode
L
R
L
R
minutes
minutes
Zone Zone
Zone Zone
1
1
1
5
1
1
2
5
1
1
1
5
1
1
2
6
2
2
1
10
2
2
2
6
2
2
1
10
2
2
2
6
5
2
2
10
5
2
2
8
5
2
2
10
5
2
2
8
1
3
2
10
1
3
2
10
1
3
2
10
1
3
2
10
2
3
2
10
2
3
2
10
2
3
2
10
2
3
2
10
5
3
2
10
5
3
2
10
5
3
2
10
5
3
2
10
1
3
2
10
1
3
2
10
1
3
2
10
1
3
2
10

- 52 -

BRONCHITIS
VITAFON-2 is used to shorten treatment, prevent complications and consolidate the effects of treatment after recovery.
The apparatus can be used in combination with the drug therapy, prescribed by the doctor.
In the case of arrhythmia, the application of vibrophones to the zones E1 and F must be excluded.
The apparatus can be used in combination with the drug therapy. During the treatment, coughing may get worse. In
such case one can reduce the duration of zones C and C1 treatment. It is essential that throughout the course of treatment the thorax should be kept warm.
During first 5 days of the treatment, a patient should treat the thorax area with iodine solution before going to bed
(5% iodine solution, treatment square about two palms).
For the treatment of chronic bronchitis it is recommended to undergo 2-3 treatment courses with a break of one
week after every finished course.
1. Treatment should be performed in accordance with the scheme 42, if doctor does not prescribe some other scheme.
Daily number of procedures 2 (once in the morning, once in the evening).
2. Manual start.
3. Duration of the treatment course 15 days.
Every procedure consists of 3 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone .
Start the session. When the session is over start the session 2.
Session 2. Enter correct data for the Session 2. Converters should be fixed at the zones and 10.
Start the session. When the session is over start the session 3.
Session 3. Converters should be fixed as follows:
- connected to the left port at the Zone 1;
- connected to the right port at the Zone F.
Start the session by pressing of the button START, for the main parameters of the sessions 2 and 3 are the same.

1
2
3
4-14
15

1
2
1
2
1
2
1
2
1
2

Mode

- 53 -

Duration,
minutes

L Double vibrophone
R Double vibrophone
Energy
L
R
Zone Zone

10
1
1
1
3
1
1
1
3
2
2
2
5
2
2
2
5
5
2
2
7
5
2
2
7
2
2
2
8
2
2
2
8
1
2
2
8
1
2
2
8

Session 3
Repetition of the
program entered
L Double vibrophone
R Double vibrophone
Energy
L
R
Zone
Zone F
1
1
1
1
3
1
1
1
3
2
2
2
5
2
2
2
5
5
2
2
7
5
2
2
7
2
2
2
8
2
2
2
8
1
2
2
8
1
2
2
8

Mode

L Double vibrophone
R Double vibrophone
Energy
L
R
Zone Zone

1
2
1
5
1
2
1
5
2
2
2
10
2
3
2
10
5
3
2
10
5
3
2
15
2
3
2
15
2
3
2
15
1
3
2
15
1
3
2
15

Duration,
minutes

Data adjustment

Duration,
minutes

Data entry

Mode

Day

Number of procedure

Scheme 42. Treatment of bronchitis


Session 1
Session 2

ESSENTIAL HYPERTENSION
1. Treatment should be performed in accordance with the schemes 43-48, if doctor does not prescribe some other
scheme. Daily number of procedures 2 (once in the morning, once in the evening).
2. Manual start.
3. Duration of the treatment course 35-90 days.
Attention! If in the course of treatment arterial pressure increases, the treatment should be discontinued for checking of vibrophones position and diagnosis confirmation.
In the case of nephrolithiasis (stones less than 4 cm.) the treatment should be performed under the doctors supervision according to his recommendations. If the patient has stones in one kidney (less than 4 cm. in size), the treatment
should be performed only in the zone of healthy kidney (both vibrophones are to be applied at the zone K from the
side of the healthy kidney). Any disease of the only kidney is a strict contraindication against self-treatment. In this
case application of vibroacoustic therapy must be approved by the doctor. If the patient has stones of more than 4 cm.
in both kidneys, treatment by VITAFON-2 may be applied only after surgical removal of stones in hospital.
Contraindications:
- Acute inflammatory process in the kidneys (such manifestations as blood pressure increase or blood in urine after
vibroacoustic therapy).
- If the patient treats the only kidney he/she has, it is strictly forbidden to undergo vibroacoustic treatment by his/her
own initiative. In such cases vibroacoustic treatment should be indicated only by the doctor.
A patient should lie supine or sit. In the case of falling of a kidney the patient must lie supine during the treatment
and vibrophone is to be placed in the zone of falling kidney (2-3 cm. lower from the K zone).
The column Drug dose in the Schemes 45-48 show the general tendency of daily drug dose reduction. Drug dose
may be reduced only if stable healing effect was achieved during previous procedures. Full daily dose is the dose prescribed by the doctor for usual taking. The minimal dose is the drug dose reduction of which leads to elevation of optimal arterial pressure. Optimal pressure can be measured by the satisfactory state of the patients health. During the
treatment, the drug dose should be corrected to keep higher systolic pressure within the range of 145160 mm. of mercury, if doctor does not prescribe some other scheme.
It is not recommended to reduce the drug dose fast. To keep arterial pressure at the optimal level, the dose should
be gradually reduced up to the achievement of desirable clinical effect together with prolongation of single VITAFON-2 session. After some time it will become possible to keep optimal pressure by vibroacoustic procedures only.
This period continues for 16 weeks or more depending on the severity and age of hypertension. After full drug
withdrawal arterial pressure should continue to reduce for 110 mm. of mercury for every week of vibroacoustic treatment (usually 2 -4 mm. of mercury weekly) depending on the severity and age of hypertension.
During treatment according to the schemes 4548 it should be taken into consideration that vibroacoustic therapy
reinforces effect of antihypertensive drugs. So, patients receiving vibroacoustic treatment in combination with drug
therapy should obey the following rules:
- daily control of arterial pressure (in the morning and in the evening);
- daily drug dose should be taken during the day in equal parts or as it is recommended by the doctor;
- full drug withdrawal can take place only on the condition that arterial pressure does not increase higher than
160 mm. of mercury when no drugs are taken;
- it should be taken into consideration that the effect of vibroacoustic therapy becomes obvious in 1-2 hours after the
first procedure and is effective from 6 to 24 hours; more than it, the healing effect accumulates with every new procedure performed.
After completion of the full treatment course in accordance with the schemes 4348 it is necessary to perform supportive preventive procedures one six times a week. Preventive procedures are described at the end of each scheme.
Total daily duration of these supportive procedures is 20-60 min. They can be divided into 1, 2 or 3 treatment sessions.
The most important point of preventive treatment is to choose such weekly duration of the kidney zone treatment that
will keep arterial pressure at its optimal level. Optimal is the level of satisfactory state of the patients health. If preventive treatment was discontinued for more than a month, duration of a treatment session should be prolonged gradually. Sessions, missed due to the circumstances, can be compensated by elevation of duration or frequency of other
sessions performed on the same day or on the next days.
If arterial pressure of the patient elevates for the first time and is not more than 140/90 mm. of mercury, it is
recommended to perform a single procedure according to the scheme 23. Duration of this single procedure should be
10 minutes.
To make the effect achieved stable, this procedure may be repeated within next 2-4 days, treatment duration being
elevated up to 15-20 min.
In the case of hypertension of the stage 1, the treatment is to be performed in accordance with the schemes 43
and 44 with no simultaneous drug therapy. If arterial pressure comes to its maximum in the morning, the treatment
should be performed in the evening. If maximum pressure occurs in the afternoon or in the evening, the treatment
should be performed in the morning.

- 54 -

In the case of stable elevation of arterial pressure (hypertension of stage 2-3, arterial pressure more than
160/100 mm. of mercury), the treatment should be performed in accordance with the scheme 45, in combination with
drugs prescribed by the doctor. In the case if after 4 weeks of treatment the effect achieved is not a sufficient one and the
full drug dose is still required, it is recommended to continue vibroacoustic treatment in accordance with the scheme 46.
In the case of systolic arterial hypertension (old people hypertension) when elevation only of systolic arterial
pressure takes place (systolic arterial pressure higher than 160 mm. of mercury, diastolic pressure less than 90 mm. of
mercury), the treatment should be performed in accordance with the scheme 47.
In the case of nephrogenetic hypertension, the treatment should be performed in accordance with the scheme 48.
In the case of physical overload, stress or supercooling it is recommended to perform an additional treatment
procedure according to the scheme 23. The duration of such procedure should be 15-20 minutes.
It is also possible to prolong a scheduled treatment session for 15-20 min.
This procedure can be also performed to remove high arterial pressure caused by taking unusual doses of alcohol. In such case it should be performed not earlier than in 6 hours after taking alcohol. Then 1-2 additional sessions
with total duration of 90 min. are to be performed (the first of them should last for not less than in 40 min., the second
is to be performed in 6-7 hours after the first).
Scheme 43. Treatment of hypertension of stage 1, arterial pressure = 125-139 mm of mercury
L Double vibrophone
Number of
Day
Energy
procedure
Mode
Duration, minutes
Zone
1
2
3
10
1
2
2
3
10
1
2
3
10
2
2
2
3
15
1
2
3
10
3-35
2
2
3
20
Then regular preventive procedures are to be performed once-trice a
week, per 20-30 minutes daily within 1-2 procedures. The quantity and
duration of sessions should be chosen in a way to support the blood
pressure on its optimal level.
Scheme 44. Treatment of hypertension of stage 1, arterial pressure = 140-159 mm of mercury
L Double vibrophone
Number of
Day
Energy
procedure Mode
Duration, minutes
Zone
1
2
3
10
1
2
2
3
10
1
2
3
10
2
2
2
3
15
1
2
3
10
3
2
2
3
20
1
2
3
10
4-7
2
2
3
25
1
2
3
10
8-21
2
2
3
30
1
2
3
10
22-28
2
2
3
35
1
2
3
10
29-60
2
2
3
40
Then regular preventive procedures are to be performed once-trice a
week, per 30-50 minutes daily within 1-2 procedures. The quantity and
duration of sessions should be chosen in a way to support the blood
pressure on its optimal level.

- 55 -

Scheme 45. Treatment of hypertension of stages 2 3


L Double vibrophone
Number of
Day
Energy
Duration,
procedure Mode
Drug dose
minutes
Zone
1
2
3
10
1
Full daily dose
2
2
3
10
1
2
3
10
2
Full daily dose
2
2
3
15
1
2
3
10
3
Full daily dose
2
2
3
20
1
2
3
15
4-7
Full daily dose
2
2
3
25
1
2
3
15
8-14
3/4 of daily dose
2
2
3
30
1
2
3
20
15-21
1/2 of daily dose
2
2
3
30
1
2
3
20
22-28
1/4 of daily dose
2
2
3
35
1
2
3
20
No drugs or
29-90
minimal dose
2
2
3
40
Then regular preventive procedures are to be performed twice-trice a week, per
40-60 minutes daily within 1-2 procedures. The quantity and duration of sessions
should be chosen in a way to support the blood pressure on its optimal level.
Scheme 46. Treatment of hypertension of stages 2 3. Intensified course
L Double vibrophone
Number of
Day
Energy
Duration,
procedure
Mode
Drug dose
minutes
Zone
1
2
3
20
29-35
Full daily dose
2
2
3
40
1
2
3
25
36-42
3/4 of daily dose
2
2
3
40
1
2
3
30
43-49
1/2 of daily dose
2
2
3
40
1
2
3
35
50-56
1/4 of daily dose
2
2
3
40
No drugs or
1
2
3
40
57-60
minimal dose
2
2
3
40
1
2
3
45
No drugs or
61-67
minimal dose
2
2
3
40
No drugs or
1
2
3
45
68-90
minimal dose
2
2
3
45
Then regular preventive procedures are to be performed 2-4 times a week, per
40-60 minutes daily within 1-2 procedures. The quantity and duration of sessions
should be chosen in a way to support the blood pressure on its optimal level.

- 56 -

Scheme 47. Treatment of systolic hypertension


L Double vibrophone
Number of
Day
Energy
Duration,
the procedure
Mode
minutes
Drug dose
Zone
1
2
3
10
1
Full daily dose
2
1
2
3
10
2
Full daily dose
2
2
3
5
1
2
3
10
3-4
Full daily dose
2
2
3
10
1
2
3
10
5-6
Full daily dose
2
2
3
15
1
2
3
10
7-8
3/4 of daily dose
2
2
3
20
1
2
3
15
9-10
3/4 of daily dose
2
2
3
20
1
2
3
20
11-12
3/4 of daily dose
2
2
3
20
1
2
3
20
13-14
3/4 of daily dose
2
2
3
25
1
2
3
20
15-21
1/2 of daily dose
2
2
3
30
1
2
3
20
22-28
1/2 of daily dose
2
2
3
35
1
2
3
20
29-35
1/4 of daily dose
2
2
3
40
1
2
3
25
36-42
1/4 of daily dose
2
2
3
40
1
2
3
30
43-49
No drugs or minimal dose
2
2
3
40
1
2
3
35
50-56
No drugs or minimal dose
2
2
3
40
1
2
3
40
57-70
No drugs or minimal dose
2
2
3
40
1
2
3
20
71-77
No drugs or minimal dose
2
2
3
40
1
78-90
No drugs or minimal dose
2
2
3
40
Then regular preventive procedures are to be performed 2-4 times a week, per 40-60 minutes daily
within 1-2 procedures. The quantity and duration of sessions should be chosen in a way to support the
blood pressure on its optimal level.

- 57 -

Scheme 48. Treatment of nephrogenic hypertension


L Double vibrophone
Number of
Day
Energy
Duration,
the procedure Mode
minutes
Drug dose
Zone
1
2
3
5
1
Full daily dose
2
1
2
3
7
2
Full daily dose
2
1
2
3
10
3
Full daily dose
2
1
2
3
10
4
Full daily dose
2
2
3
5
1
2
3
10
5-6
Full daily dose
2
2
3
10
1
2
3
10
7-8
Full daily dose
2
2
3
15
1
2
3
10
9-10
3/4 of daily dose
2
2
3
20
1
2
3
15
11-12
3/4 of daily dose
2
2
3
20
1
2
3
20
13-14
1/2 of daily dose
2
2
3
20
1
2
3
20
15-21
1/2 of daily dose
2
2
3
25
1
2
3
20
22-28
1/2 of daily dose
2
2
3
30
1
2
3
20
29-35
1/4 of daily dose
2
2
3
35
1
2
3
20
36-42
1/4 of daily dose
2
2
3
40
1
2
3
25
43-49
No drugs or minimal dose
2
2
3
40
1
2
3
30
50-56
No drugs or minimal dose
2
2
3
40
1
2
3
35
57-63
No drugs or minimal dose
2
2
3
40
1
2
3
40
64-70
No drugs or minimal dose
2
2
3
40
1
2
3
20
71-77
No drugs or minimal dose
2
2
3
40
1
78-90
No drugs or minimal dose
2
2
3
40
Then regular preventive procedures are to be performed 3-6 times a week, per 20-60 minutes daily within
1-2 procedures. The quantity and duration of sessions should be chosen in a way to support the blood pressure on its optimal level.

- 58 -

BURNS
VITAFON-2 is applied for blood supplement reconstruction in the tissues burned by high temperature, aggressive
chemicals, electricity or ionic radiation.
For the treatment of burns of degrees 1 and 2 (total square less than 50 m2 and skin integrity survived) vibrophones should be applied to an affected area covered with a gauze towel. Before the application of vibrophones the skin
around the burn (2 cm.) should be treated with iodine solution.
1. Treatment should be performed in accordance with the scheme 49, if doctor does not prescribe some other scheme.
Daily number of procedures 1-3.
2. Manual start.
3. Duration of the treatment course 7 days.
The treatment should start as follows:
Enter correct data. Converters should be fixed at the burned zone in such a way as to treat much of affected area.
Start the session.
Scheme 49. Treatment of burns of degrees 1 and 2 (total square less than 50 m2 and skin integrity survived).
L Double vibrophone
R Double vibrophone
Daily
Day
number of
Energy
Duration,
procedures
Mode
L
R
minutes
Burned zone
1
1-3
1
2
2
10
2
1-3
2
2
2
15
3-7
1-3
5
2
2
15
For the treatment of large burns of degrees 1 and 2 (total square more than 50 cm2) and burns of degrees 3 and
4 VITAFON-2 should be applied as an integral additional part of the complex treatment in hospital.
1. Treatment should be performed in accordance with the scheme 50. Daily number of procedures 3-4.
2. Manual start.
3. Duration of the treatment course until full recovery, not less than 14 days.
With more severe burns, vibrophones should be applied to an affected area, without pressing them, over 4-5 layers
of a loose bandage.
In the case of necessity, before the treatment affected area should be treated by an anti-septic agent.
Every procedure consists of 2 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the burned zone.
Start the session. When the session is over start the session 2.
Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port burned zone, which received no or less treatment during session 1.
Start the session by pressing of the button START, for the main parameters of the sessions 1 and 2 are the same.
Scheme 50. Treatment of burns of degrees 1, 2, 3 and 4
Session 1
Session 2
Repetition of the
Data entry
program entered
L Double vibrophone
L Double vibrophone
Daily
R Double vibrophone
R Double vibrophone
Day number of
procedures
Energy
Energy
Duration,
Duration,
L
R
L
R
Mode
Mode
minutes
minutes
Zone
Burned
Zone Burned

zone

zone
1
3-4
1
3
3
10
1
3
3
10
2
3-4
2
3
3
15
2
3
3
15
3-14
3-4
5
3
3
15
5
3
3
15

- 59 -

1
2
3-4
5-15
(5-30)

1
2
1
2
1
2
1
2

Duration,
minutes

Mode

Duration,
minutes

Mode

Duration,
minutes

Scheme 51. Treatment of otitis and sensorineural deafness


Session 1
Session 2
Session 3
Repetition of the
Data entry
Data adjustment
program entered
L Double vibrophone
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
R Double vibrophone
Energy
Energy
Energy
L
R
L
R
L
R
Zone Zone
Zone Zone
Zone Zone

D1
1
D2

1
2
1
5
1
1
1
3
1
1
1
3
1
2
1
5
1
1
1
3
1
1
1
3
2
2
1
10
2
1
1
5
2
1
1
5
2
3
2
10
2
1
2
5
2
1
2
5
5
3
2
10
5
1
2
7
5
1
2
7
5
3
2
15
5
1
2
7
5
1
2
7
2
3
2
15
2
1
2
7
2
1
2
7
2
3
2
15
2
1
2
10
2
1
2
10

Mode

Day

Number of procedure

OTITIS AND SENSORINEURAL HEARING DISORDER


For the treatment of otitis the apparatus is used at the early stage of the disease and in order to prevent the disease
after having caught cold.
1. Treatment should be performed in accordance with the scheme 51, if doctor does not prescribe some other scheme.
Daily number of procedures 2 (once in the morning, once in the evening).
2. Manual start.
3. Duration of the treatment course for the treatment of otitis 15 days. Painful sensations can be over after 2-5
days but it is still recommended to finish the full treatment course. In the case of acute inflammation, the apparatus
may be used only in combination with anti-inflammatory preparations and under doctors supervision.
Duration of the treatment course for the treatment of sensorineural hearing disorder 30 days. In the case of
necessity, one should perform 2-3 treatment courses with a break of 2 weeks after each finished course.
Every procedure consists of 3 sessions.

Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone .
Start the session. When the session is over start the session 2.
Session 2. Enter correct data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone D1;
- connected to the right port at the Zone 1 from both sides.
Start the session. When the session is over start the session 3.
Session 3. Converters should be fixed as follows:
- connected to the left port at the Zone D2;
- connected to the right port at the Zone from both sides.
Start the session by pressing of the button START, for the main parameters of the sessions 2 and 3 are the same.

Picture 8. Treatment zones , 1, , 2

- 60 -

FROSTBITES AND SUPERCOOLING


The apparatus is applied in order to achieve restoration of blood supplement in the bodily tissues affected by coldness; to remove edema and inflammatory manifestations after heating of the frostbitten areas.
After frostbite it is forbidden to apply direct heat or to rub affected areas for several hours. Warming up should be
caused by natural blood supplement of frostbitten areas so, vibroacoustic therapy must be started as soon as possible.
If necessary, treat the affected skin with alcohol or iodine solution and make a sterile bandage before the treatment.
A patient should lie supine or sit during the treatment (depending on the location and severity of the frostbite).
Treatment of local supercooling and frostbites of degree 1
1. Treatment should be performed in accordance with the scheme 52, if doctor does not prescribe some other scheme.
Daily number of procedures 3-4.
2. Manual start.
3. Duration of the treatment course 7 days.
The treatment is to be performed right at the frostbitten area.
The treatment should start as follows:
Enter correct data. Converters should be fixed in order to perform the treatment not only at the frostbitten area, but
also 5-7 m. around this area.
Start the session.
Scheme 52. Treatment of frostbites and supercooling of degree 1
L Double vibrophone
R Double vibrophone
Daily number
Day
Energy
of procedures
Duration,
Mode
L
R
minutes
Inflammation zone
1
3-4
1
3
3
20
2
3-4
2
3
3
30
3-7
3-4
5
3
3
30

Day

Daily number
of procedures

Treatment of frostbites of degrees 2 and more


1. Treatment should be performed in accordance with the scheme 53, if doctor does not prescribe some other scheme.
Daily number of procedures 3-4.
2. Manual start.
3. Duration of the treatment course 10-14 days (until the full recovery).
The treatment is to be performed at the zone K, at the border of frostbitten area and right at the frostbitten area.
Every procedure consists of 2 sessions.

1
2
3-10 (3-14)

3-4
3-4
3-4

Scheme 53. Treatment of frostbites of higher degrees


Session 1
Session 2
Data entry
Repetition of the program entered
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
Energy
Energy
L
R
L
R
Duration,
Duration,
Mode
Mode
Border of
minutes
minutes
Frostbitten
Zone frostbitten
Zone
zone
zone
1
2
3
15
1
2
3
15
2
3
3
15
2
3
3
15
5
3
3
15
5
3
3
15

Session 1. Enter correct data. Converters should be fixed as follows:


- connected to the left port at the Zone ;
- connected to the right port at the border of the frostbitten area.
Start the session. When the session is over start the session 2.
Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the border of the frostbitten area.
Start the session by pressing of the button START, for the main parameters of the sessions 1 and 2 are the same.
If the frostbitten area is large, location of converters should be changed for every new procedure.
- 61 -

HEMORRHOID
If chronic constipation occurs, a patient should alternate treatment courses for hemorrhoid and constipation.
1. Treatment should be performed in accordance with the scheme 54, if doctor does not prescribe some other scheme.
Firstly a patient should undergo the treatment of course 1 in accordance with the scheme 54, then, after a break of
2-3 days course 2 in accordance with the scheme 54. Daily number of procedures 2 (once in the morning, once
in the evening).
2. Manual start.
3. Duration of every treatment course 14 days.
A patient should lie supine. If necessary, a small pillow can be placed under pelvis.
If necessary, the course of treatment may be repeated every three months.
Scheme 54, course 1. Every procedure consists of 3 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 1.
Start the session. When the session is over start the session 2.
Session 2. Enter correct data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port during the morning procedure at the Zone 3, during the evening procedure at the
Zone 32.
Start the session. When the session is over start the session 3.
Session 3. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port during the morning procedure at the Zone 30, during the evening procedure at the
Zone 4;
Start the session by pressing the button START for the main parameters for the sessions 2 and 3 are the same.
Scheme 54, course 2. Every procedure consists of 3 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 1.
Start the session. When the session is over start the session 2.
Session 2. Enter correct data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone D1.
Start the session. When the session is over start the session 3.
Session 3. Enter correct data for the Session 3. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port during the morning procedure at the Zone 30, during the evening procedure at the
Zone 32.
Start the session.

- 62 -

2
3
4
5
6
7
8
9
10
11
12
13
14

1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2

- 63 -

Duration, minutes

Mode

Duration, minutes

Mode

Duration,
minutes

Mode

Number of procedure

Day
1

Scheme 54. Treatment of hemorrhoid. Treatment course 1


Session 1
Session 2
Session 3
Data entry
Data adjustment
Repetition of the program entered
L Double vibrophone
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
R Double vibrophone
Energy
Energy
Energy
L
R
L
R
L
R
Morning
Morning
Zone Zone
Zone
Zone 3.
Zone
Zone 30.

Evening

Evening
Zone 32
Zone 4
1
3
1
3
1
3
2
3
1
3
2
3
1
3
1
3
1
3
2
3
1
3
2
3
2
3
1
3
2
3
2
3
2
3
2
3
2
3
1
3
2
3
2
3
2
3
2
3
5
3
1
4
5
3
2
3
5
3
2
3
5
3
1
4
5
3
2
3
5
3
2
3
1
3
1
4
1
3
2
3
1
3
2
3
1
3
1
4
1
3
3
3
1
3
3
3
2
3
2
5
2
3
2
4
2
3
2
4
2
3
2
5
2
3
3
4
2
3
3
4
5
3
2
5
5
3
2
4
5
3
2
4
5
3
2
5
5
3
3
4
5
3
3
4
1
3
2
6
1
3
2
4
1
3
2
4
1
3
2
6
1
3
3
4
1
3
3
4
2
3
2
6
2
3
2
4
2
3
2
4
2
3
2
6
2
3
3
4
2
3
3
4
5
3
2
7
5
3
2
5
5
3
2
5
5
3
2
7
5
3
3
5
5
3
3
5
1
3
2
7
1
3
2
5
1
3
2
5
1
3
2
7
1
3
3
5
1
3
3
5
2
3
2
8
2
3
2
5
2
3
2
5
2
3
2
8
2
3
3
5
2
3
3
5
5
3
2
8
5
3
2
5
5
3
2
5
5
3
2
8
5
3
3
5
5
3
3
5
1
3
2
9
1
3
2
6
1
3
2
6
1
3
2
9
1
3
3
6
1
3
3
6
2
3
2
9
2
3
2
6
2
3
2
6
2
3
2
9
2
3
3
6
2
3
3
6

2
3
4
5
6
7
8
9
10
11
12
13
14

1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2

Duration,
minutes

Mode

Duration, minutes

Mode

Duration,
minutes

Mode

Number of procedure

Day
1

Scheme 54. Treatment of hemorrhoid. Treatment course 2


Session 1
Session 2
Session 3
Data entry
Data adjustment
Data adjustment
L Double vibrophone
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
R Double vibrophone
Energy
Energy
Energy
L
R
L
R
L
R
Morning
Zone 30.
Zone Zone
Zone Zone
Zone

D1

Evening
Zone 32
1
3
2
7
1
3
1
3
1
3
2
1
3
2
7
1
3
1
3
1
3
3
2
3
2
9
2
3
1
3
2
3
2
2
3
2
9
2
3
1
3
2
3
3
5
3
2
9
5
3
1
4
5
3
2
5
3
2
9
5
3
1
4
5
3
3
1
3
2
10
1
3
1
4
1
3
2
1
3
2
10
1
3
1
4
1
3
3
2
3
2
11
2
3
1
5
2
3
2
2
3
2
11
2
3
1
5
2
3
3
5
3
2
12
5
3
1
5
5
3
2
5
3
2
12
5
3
1
5
5
3
3
1
3
2
12
1
3
1
6
1
3
2
1
3
2
12
1
3
1
6
1
3
3
2
3
2
12
2
3
1
6
2
3
2
2
3
2
12
2
3
1
6
2
3
3
5
3
2
12
5
3
1
7
5
3
2
5
3
2
12
5
3
1
7
5
3
3
1
3
2
12
1
3
1
7
1
3
2
1
3
2
12
1
3
1
7
1
3
3
2
3
2
12
2
3
1
8
2
3
2
2
3
2
12
2
3
1
8
2
3
3
5
3
2
12
5
3
1
8
5
3
2
5
3
2
12
5
3
1
8
5
3
3
1
3
2
12
1
3
1
9
1
3
2
1
3
2
12
1
3
1
9
1
3
3
2
3
2
12
2
3
1
9
2
3
2
2
3
2
12
2
3
1
9
2
3
3

4
4
6
6
7
7
8
8
9
9
10
10
11
11
12
12
12
12
12
12
12
12
12
12
12
12
12
12

CHRONIC HEPATITIS
Treatment with the VITAFON-2 apparatus should be prescribed by the doctor and take place under his control
without any drug therapy performed simultaneously.
1. Treatment should be performed in accordance with the scheme 55, if doctor does not prescribe some other scheme.
Daily number of procedures 1-2.
2. Manual start.
3. Duration of the treatment course 6 months.
A patient should lie supine.
During the treatment course a vibrophone and an IR-matrix are to be replaced with each other for every
new procedure.
Every procedure consists of 2 sessions.

- 64 -

Session 1. Enter correct data for the Session 1. Converters should be fixed in accordance with the procedure:
- either at the Zone ;
- or at the Zone F right. (Both converters are to be placed at the same zone, not far from each other.)
Start the session. When the session is over start the session 2.
Session 2. Enter correct data for the Session 2. Converters should be fixed in accordance with the procedure:
- either at the Zone ;
- or at the Zone 4 right. (Both converters are to be placed at the same zone, not far from each other.)
Start the session.
Scheme 55. Treatment of chronic hepatitis
Session 1
Session 2
Data entry
Data adjustment
L Single vibrophone
L Single vibrophone
Number of
Day
R IR-matrix
R IR-matrix
procedure
L R Duration,
L
R Duration,
Mode
Mode Energy
Energy
minutes
minutes
Zone
Zone
1
2
2

5
2
3

10
1
2
2
2

5
2
3

10
1
2
2

5
2
3

10
2
2
2
2

5
2
3

10
3
1
2
3
F right
5
2
3
M4 right
5
1
2
2

5
2
3

10
4
2
2
2

5
2
3

10
1
2
2

5
2
3

10
5
2
2
2

5
2
3

10
6
1
2
3
F right
7
2
3
M4 right
7
1
2
2

10
2
3

15
7-11
2
2
2

10
2
3

15
12
1
2
3
F right
10
2
3
M4 right
10
1
2
2

10
2
3

15
13-16
2
2
2

10
2
3

15
17
1
2
3
F right
13
2
3
M4 right
13
Then continue procedures of days 13-17 within 6 months. Procedures may be missed but not more
often than once a week and in the days when there is no procedure at the zones F right and 4 right.
After missing, procedures of the previous day should be performed.
After 6 months of treatment, the patient should undergo a blood test for RNA of hepatitis virus.
If after the treatment course of 6 months no remission occurs, after a break of two weeks the patient, under the doctors supervision, should start individual intensified treatment course, chosen by the doctor.
After the remission, the patient should start supportive course to prevent recidivation. Supportive treatment course
should be performed in accordance with the scheme 56 and should last for 17 weeks. Then the supportive course
should be repeated once-twice a year.
Scheme 56. Treatment of chronic hepatitis. Supportive course
Day of the week
Monday
Tuesday
Wednesday
Thursday
Friday
Saturday
Sunday

1; 2

Number of the week


5
6-8
9-11

B
B
B

- 65 -

12-14

15-17
B

Scheme .

Time
of the day

Morning
Evening

Session 1
Session 2
Data entry
Data adjustment
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
Energy
Energy
Duration,
L
R
L
R
Mode
Mode
minutes
Zones
Zone
Zone Zone
F right and 4 right
2
3
2
15
2
3
3
2
3
2
15
2
3
3

Time
of the day

8-00
9-00
20-00
21-00

Mode
2
2
2
2

Scheme B.
L Double vibrophone
R Double vibrophone
Energy
L
R
Zones
Zone
F right and 4
right
3
3
3
3
3
3
3
3

Duration,
minutes
10
10

Duration,
minutes
10
10
10
10

CHRONIC GASTRODUODENITIS (GASTRITIS)


Treatment with the VITAFON-2 apparatus is adjunctive to the drug therapy, should be prescribed by the doctor and
takes place under his control. One should keep in mind that acoustic vibrations accelerate drug effect, so doses and
duration of drug treatment may be reduced.
At the stage of remission the apparatus is used for preventive purposes without any drug treatment.
1. Treatment should be performed in accordance with the scheme 57, if doctor does not prescribe some other scheme.
Daily number of procedures 1.
2. Manual start.
3. Duration of the treatment course 15-20 days. If necessary, the course of treatment may be repeated after 1-2
weeks.
Procedures should be performed before a meal of in 2 hours after a meal.
Every procedure consists of 2 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the painful stomach zone.
Start the session. When the session is over start the session 2.
Session 2. Enter correct data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 20.
Start the session.

Day

1-2
3-4
5-15 (5-20)

Scheme 57. Treatment of chronic gastrodeudenitis (gastritis)


Session 1
Session 2
Data entry
Data adjustment
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
Energy
Energy
Duration,
Duration,
Mode
Mode
L
R
L
R
minutes
minutes
Zone Stomach zone
Zone Zone 20
1
2
2
5
2
3
2
5
2
3
2
10
2
3
2
7
2
3
2
15
2
3
2
10

- 66 -

TRACHEITIS
The VITAFON-2 apparatus is used to shorten treatment and prevent complications. The therapy by VITAFON-2
can be applied together with drug treatment.
During the treatment course coughing can get worse. In such case, one can reduce treatment duration at the zones
and 10.
It is essential that throughout the course of treatment the thorax should be kept warm.
1. Treatment should be performed in accordance with the scheme 58, if doctor does not prescribe some other scheme.
Daily number of procedures 2 (once in the morning, once in the evening).
2. Manual start.
3. Duration of the treatment course 15 days.
Patients position:
- For the treatment at the Zones and a patient should lie supine;
- For the treatment at the Zones , 10, 1 and D2 a patient should sit or recline with a pillow under the neck and
shoulders.
During the first 5 days of the treatment, the thorax should be treated with iodine solution before going to bed (5%
iodine solution, treatment square about two palms).
Every procedure consists of 3 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone .
Start the session. When the session is over start the session 2.
Session 2. Enter correct data for the Session 2. Converters should be fixed at the zones and 10.
Start the session. When the session is over start the session 3.

1
2
3
4-14
15

1
2
1
2
1
2
1
2
1
2

- 67 -

Duration,
minutes

Session 3
Data adjustment
L Double vibrophone
R Double vibrophone
Energy
L
R
Zone Zone
1
D2
1
1
1
2
1
1
1
2
2
2
2
3
2
2
2
3
5
2
2
4
5
2
2
4
2
2
2
5
2
2
2
5
1
2
2
3
1
2
2
3

Mode

Duration,
minutes

Mode

Duration,
minutes

Scheme 58. Treatment of tracheitis


Session 1
Session 2
Data entry
Data adjustment
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
Energy
Energy
L
R
L
R
Zone Zone
Zone
Zone

10
1
2
1
5
1
1
1
3
1
2
1
5
1
1
1
3
2
2
1
10
2
2
2
5
2
3
1
5
2
2
2
5
5
3
1
10
5
2
2
7
5
3
1
15
5
2
2
7
2
3
1
15
2
2
2
8
2
3
1
15
2
2
2
8
1
3
1
10
1
2
2
5
1
3
1
10
1
2
2
5

Mode

Day

Number of procedure

Session 3. Enter correct data for the Session 3. Converters should be fixed as follows:
- connected to the left port at the Zone 1;
- connected to the right port at the Zone D2.
Start the session.

LYMPHOSTASIS OF LOWER EXTREMITIES


1. Treatment should be performed in accordance with the scheme 59, if doctor does not prescribe some other scheme.
Daily number of procedures 1.
2. Manual start.
3. Duration of the treatment course 14 days (then procedures should be performed once-trice a week).
If necessary, the treatment course can be repeated after a break of 1 3 weeks.
A patient should lie supine. If necessary, a small pillow can be placed under shins.
It is recommended not to get up after the procedure till the end of the night. While sleeping, legs should be slightly
elevated (e.g. by several layers of a blanket).
During the treatment course a vibrophone and an IR-matrix are to be replaced with each other for every
new procedure.
Every procedure consists of 3 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 32.
Start the session. When the session is over start the session 2.
Session 2. Change the converters and enter correct data for the Session 2. One converter should be fixed at the zone of
right popliteal space, the other one at the zone of left popliteal space.
Start the session. When the session is over start the session 3.
Session 3. Enter correct data for the Session 3. Converters should be fixed symmetrically at the groin zone.
Start the session.

1-2
3-5
6-14
Then oncetrice a week

15

Duration,
minutes

Duration,
minutes

Duration,
minutes

Day

Scheme 59. Treatment of lymphostasis of lower extremities


Session 1
Session 2
Session 3
Change of converters
Data adjustment
Data entry
and data adjustment
L Double vibrophone
L Single vibrophone
L Single vibrophone
R Double vibrophone
R IR-matrix
R IR-matrix
Energy
Energy
Energy
L
R
L
R
L
R
Mode
Mode
Mode
Zone Zone
Popliteal zone
Groin zone

32
1
3
3
10
4
2
3
1
2
3
2
3
3
12
4
3
4
2
3
4
5
3
3
15
4
3
5
5
3
5
5

VEGETOVASCULAR DYSTONIA
Patients having thyroid gland diseases should apply the apparatus only according to the endocrinologists recommendations and under his supervision.
Treatment courses should be performed when necessary with the break of two weeks after every finished course.
1. Treatment should be performed in accordance with the scheme 60, if doctor does not prescribe some other
scheme. Daily number of procedures 2 (once in the morning, once in the evening).
2. Manual start.
3. Duration of the treatment course 14 days.
Every procedure consists of 3 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone D1.
Start the session. When the session is over start the session 2.
Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone D2.
Start the session by pressing of the button START, for the main parameters of the sessions 1 and 2 are the same.
When the session is over start the session 3.

- 68 -

1
2
3-5
6-13
14

1
2
1
2
1
2
1
2
1
2

Duration,
minutes

Mode

Duration,
minutes

Mode

Duration,
minutes

Scheme 60. Treatment of vegetovascular dystonia


Session 1
Session 2
Session 3
Repetition of the
Data entry
Data adjustment
program entered
L Double vibrophone
L Double vibrophone
L Double vibrophone
R Double vibrophone
R Double vibrophone
R Double vibrophone
Energy
Energy
Energy
L
R
L
R
L
R
Zone Zone
Zone Zone
Zone Zone

D1

D2

2
1
3
2
1
2
1
3
1
2
1
3
1
2
3
2
2
3
2
2
2
1
5
2
2
1
5
2
2
3
3
2
3
3
5
3
1
7
5
3
1
7
2
3
3
3
2
3
3
2
3
1
9
2
3
1
9
2
3
3
3
2
3
2
1
3
1
7
1
3
1
7
2
3
3
2

Mode

Day

Number of procedure

Session 3. Enter correct data for the Session 3. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 2.
Start the session.

RESOURCE SUPPORT OF DEBILITATED PATIENTS


Resource support of the organism is one of the main components of successful patients cure. Microvibration is an
important and indispensable resource, needed by the body tissues even while sleeping. Periodical start mode is the
most effective for the resource support of the organism because little single doses provide gradual organisms reactions and allow receiving of the dose needed. Liver and kidneys are the most important organs, which supply the
whole organism with resources so, they are the first to be treated.
1. Treatment should be performed in accordance with the scheme 61, if doctor does not prescribe some other scheme.
2. Automatical start.
3. Duration of the treatment course until the end of rest cure.
The treatment should start as follows:
Enter correct data. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 4 right. (Both vibraphones are to be placed at the same zone, not far
from each other).
Start the session. If after the treatment time, indicated in the treatment scheme, the procedure is finished by pressing of the button RESET, the parameters programmed will be lost; if it is finished by switching off of the apparatus,
the parameters of the last procedure will be kept.
Scheme 61. Resource support of debilitated patients
L Double vibrophone
R Double vibrophone
Day
Energy
Duration of the
Mode
L
R
procedure
Zone
Zone 4 right
1
AUTO 12
2
1
24 hours
2
AUTO 12
3
2
6+6 hours **
3
AUTO 12
2
1
24 hours
4
AUTO 12
3
2
6+6 hours **
5
AUTO 12
2
1
24 hours
6
AUTO 12
3
2
6+6 hours **
Then continue procedures of the Day 6 till the end of the rest cure.
** When the total duration of the procedure is 12 hours, the dose should be received within 2 sessions per 6 hours each;
the first one is to be performed in the morning, right after waking up, the second in the evening, before going to bed.
- 69 -

COMPLICATIONS OF INFANTILE CEREBRAL PARALYSIS


The treatment of complications of infantile cerebral paralysis should be performed according to the doctors recommendations and under his supervision.
1. Treatment should be performed in accordance with the scheme 62, if doctor does not prescribe some other scheme.
Daily number of procedures 2 (once in the morning, once in the evening).
2. Manual start.
3. Duration of the full treatment course 91 days (the full treatment course consists of three sub-courses per one
week each with a break of two weeks after every finished sub-course).
A patient should lie supine during the treatment.
Every procedure consists of 3 sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the zones D1, D2 or D3.
Start the session. When the session is over start the session 2.
Session 2. Enter correct data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the zones 1, 10, 12, 2, 20, 22 or 3.
Start the session. When the session is over start the session 3.
Session 3. Enter correct data for the Session 3. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the zones 2, G or 1.
Start the session.
During the two-week breaks between sub-courses and after the last one it is recommended to treat contraggerated
muscles daily. Zones of application of two double vibrophones are to be indicated by neuropathologist. The vibrophones are to be placed along the muscle. During single session 1-3 muscles can be treated.

The following parameters should be applied for this procedure:


MODE 4; ENERGY 4; DURATION 5-15 minutes.

- 70 -

3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21

2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2

- 71 -

2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2

Duration,
minutes

5
5
5
5
5
5
5
5
5
5
5
5
5
5
8
8
8
8
8
8
8
8
8
8
8
8
8
8
12
12
12
12
12
12
12
12
12
12
12
12
12
12

Session 3
Data adjustment
L Double vibrophone
R Double vibrophone
L
R
Energy
at the Zone
Zone
3
2
3
5
3
G
2
5
3
2
3
5
3
1
1
5
3
2
3
5
3
G
2
5
3
2
3
5
3
1
1
5
3
2
3
5
3
G
2
5
3
2
3
5
3
1
1
5
3
2
3
5
3
G
1
5
3
2
3
5
3
1
2
8
3
2
3
5
3
G
1
8
3
2
3
5
3
1
2
8
3
2
3
5
3
G
1
8
3
2
3
5
3
1
2
8
3
2
3
5
3
G
1
8
3
2
3
5
3
1
2
8
3
2
3
5
3
G
1
10
3
2
3
5
3
1
2
10
3
2
3
5
3
G
1
10
3
2
3
5
3
1
2
10
3
2
3
5
3
G
1
10
3
2
3
5
3
1
2
10
3
2
3
5
3
G
1
10
Energy

Energy

Mode

Duration,
minutes

Energy

Mode
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2

Session 2
Data adjustment
L Double vibrophone
R Double vibrophone
L
R
Energy
at the Zone
Zone
3
1
1
3
1
1
3
10
1
3
10
1
3
12
1
3
12
1
3
2
1
3
2
1
3
20
1
3
20
1
3
22
1
3
22
1
3
3
1
3
3
1
3
1
1
3
1
1
3
10
2
3
10
2
3
12
2
3
12
2
3
2
2
3
2
2
3
20
2
3
20
2
3
22
2
3
22
2
3
3
2
3
3
2
3
1
2
3
1
2
3
10
2
3
10
2
3
12
2
3
12
2
3
2
2
3
2
2
3
20
2
3
20
2
3
22
2
3
22
2
3
3
2
3
3
2

Mode

1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2

Session 1
Data entry
L Double vibrophone
R Double vibrophone
L
R
Energy
at the Zone
Zone
3
D1
1
5
3
D2
1
5
3
D3
2
5
3
D1
1
5
3
D2
1
5
3
D3
2
5
3
D1
1
5
3
D2
1
5
3
D3
2
5
3
D1
1
5
3
D2
1
5
3
D3
2
5
3
D1
1
5
3
D2
1
5
3
D3
1
8
3
D1
1
8
3
D2
2
8
3
D3
1
8
3
D1
1
8
3
D2
2
8
3
D3
1
8
3
D1
1
8
3
D2
2
8
3
D3
1
8
3
D1
1
8
3
D2
2
8
3
D3
1
8
3
D1
1
8
3
D2
1
12
3
D3
1
12
3
D1
2
12
3
D2
1
12
3
D3
1
12
3
D1
2
12
3
D2
1
12
3
D3
1
12
3
D1
2
12
3
D2
1
12
3
D3
1
12
3
D1
2
12
3
D2
1
12
3
D3
1
12

Duration,
minutes

Number of procedure

Day

Scheme 62. Treatment of complications of infantile cerebral paralysis. Treatment course 1

38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56

- 72 -

5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5

Duration,
minutes

Energy

Session 3
Data adjustment
L Double vibrophone
R Double vibrophone
L
R
Energy
at the
Zone
3
2
3
5
3
G
2
5
3
2
3
5
3
1
2
5
3
2
3
5
3
G
2
5
3
2
3
5
3
1
2
5
3
2
3
5
3
G
2
5
3
2
3
5
3
1
2
5
3
2
3
5
3
G
2
5
3
2
3
5
3
A1
3
8
3
2
3
5
3
G
3
8
3
2
3
5
3
1
3
8
3
2
3
5
3
G
3
8
3
2
3
5
3
1
3
8
3
2
3
5
3
G
3
8
3
2
3
5
3
1
3
8
3
2
3
5
3
G
3
10
3
2
3
5
3
A1
3
10
3
2
3
5
3
G
3
10
3
2
3
5
3
1
3
10
3
2
3
5
3
G
3
10
3
2
3
5
3
1
3
10
3
2
3
5
3
G
3
10
Zone

Mode

Duration,
minutes

5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5

Energy

5
5
5
5
5
5
5
5
5
5
5
5
5
5
8
8
8
8
8
8
8
8
8
8
8
8
8
8
12
12
12
12
12
12
12
12
12
12
12
12
12
12

Session 2
Data adjustment
L Double vibrophone
R Double vibrophone
L
R
Energy
at the
Zone
3
1
3
5
3
1
3
5
3
10
3
5
3
10
3
5
3
12
3
5
3
12
3
5
3
2
3
5
3
2
3
5
3
20
3
5
3
20
3
5
3
22
3
5
3
22
3
5
3
3
3
5
3
3
3
5
3
1
3
8
3
1
3
8
3
10
3
8
3
10
3
8
3
12
3
8
3
12
3
8
3
2
3
8
3
2
3
8
3
20
3
8
3
20
3
8
3
22
3
8
3
22
3
8
3
3
3
8
3
3
3
8
3
1
3
12
3
1
3
12
3
10
3
12
3
10
3
12
3
12
3
12
3
12
3
12
3
2
3
12
3
2
3
12
3
20
3
12
3
20
3
12
3
22
3
12
3
22
3
12
3
3
3
12
3
3
3
12
Zone

Energy

Zone

Mode
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5

Mode

37

1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2

Session 1
Data entry
L Double vibrophone
R Double vibrophone
L
R
Energy
at the
Zone
3
D1
2
3
D2
2
3
D3
3
3
D1
2
3
D2
2
3
D3
3
3
D1
2
3
D2
2
3
D3
3
3
D1
2
3
D2
2
3
D3
2
3
D1
2
3
D2
2
3
D3
1
3
D1
1
3
D2
2
3
D3
1
3
D1
1
3
D2
2
3
D3
1
3
D1
1
3
D2
2
3
D3
1
3
D1
1
3
D2
2
3
D3
1
3
D1
1
3
D2
1
3
D3
1
3
D1
2
3
D2
1
3
D3
1
3
D1
2
3
D2
1
3
D3
1
3
D1
2
3
D2
1
3
D3
1
3
D1
2
3
D2
1
3
D3
1

Duration,
minutes

36

Number of procedure

Day

Scheme 62. Treatment of complications of infantile cerebral paralysis. Treatment course 2

73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91

- 73 -

5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5

Duration,
minutes

5
5
5
5
5
5
5
5
5
5
5
5
5
5
8
8
8
8
8
8
8
8
8
8
8
8
8
8
12
12
12
12
12
12
12
12
12
12
12
12
12
12

Session 3
Data adjustment
L Double vibrophone
R Double vibrophone
L
R
Energy
at the Zone
Zone
2
2
3
5
2
G
2
5
2
2
3
5
2
1
2
5
2
2
3
5
2
G
2
5
2
2
3
5
2
1
2
5
2
2
3
5
2
G
2
5
2
2
3
5
2
1
2
5
2
2
3
5
2
G
2
5
2
2
3
5
2
A1
3
8
2
2
3
5
2
G
3
8
2
2
3
5
2
1
3
8
2
2
3
5
2
G
3
8
2
2
3
5
2
1
3
8
2
2
3
5
2
G
3
8
2
2
3
5
2
1
3
8
1
2
3
5
1
G
3
10
1
2
3
5
1
A1
3
10
1
2
3
5
1
G
3
10
1
2
3
5
1
1
3
10
1
2
3
5
1
G
3
10
1
2
3
5
1
1
3
10
1
2
3
5
1
G
3
10
Energy

Energy

Energy

Duration,
minutes
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
8
8
8
8
8
8
8
8
8
8
12
12
12
12
12
12
12
12
12
12
12
12
12
12

Session 2
Data adjustment
L Double vibrophone
R Double vibrophone
L
R
Energy
Mode
at the Zone
Zone
4
2
1
3
4
2
1
3
4
2
10 3
4
2
10 3
4
2
12 3
4
2
12 3
4
2
2
3
4
2
2
3
4
2
20 3
4
2
20 3
4
2
22 3
4
2
22 3
4
2
3
3
4
2
3
3
4
2
1
3
4
2
1
3
4
2
10 3
4
2
10 3
4
2
12 3
4
2
12 3
4
2
2
3
4
2
2
3
4
2
20 3
4
2
20 3
4
2
22 3
4
2
22 3
4
2
3
3
4
2
3
3
4
1
1
3
4
1
1
3
4
1
10 3
4
1
10 3
4
1
12 3
4
1
12 3
4
1
2
3
4
1
2
3
4
1
20 3
4
1
20 3
4
1
22 3
4
1
22 3
4
1
3
3
4
1
3
3

Mode

72

1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2

Session 1
Data entry
L Double vibrophone
R Double vibrophone
L
R
Energy
Mode
at the Zone
Zone
4
2
D1
3
4
2
D2
3
4
2
D3
3
4
2
D1
3
4
2
D2
3
4
2
D3
3
4
2
D1
3
4
2
D2
3
4
2
D3
3
4
2
D1
3
4
2
D2
3
4
2
D3
3
4
2
D1
3
4
2
D2
3
4
2
D3
3
4
2
D1
3
4
2
D2
3
4
2
D3
3
4
2
D1
3
4
2
D2
3
4
2
D3
3
4
2
D1
3
4
2
D2
3
4
2
D3
3
4
2
D1
3
4
2
D2
3
4
2
D3
3
4
2
D1
3
4
1
D2
3
4
1
D3
3
4
1
D1
3
4
1
D2
3
4
1
D3
3
4
1
D1
3
4
1
D2
3
4
1
D3
3
4
1
D1
3
4
1
D2
3
4
1
D3
3
4
1
D1
3
4
1
D2
3
4
1
D3
3

Duration,
minutes

71

Number of procedure

Day

Scheme 62. Treatment of complications of infantile cerebral paralysis. Treatment course 3

FURUNCULOSIS, CARBUNCULOSIS, ACNE


The VITAFON-2 apparatus can be applied for the treatment of furunculosis, carbunculosis and acne at the earlier
stages of their formation (reddening, swelling) or after their surgical d-bridement / drain of abscess only on the assumption of regular pus outflow.
1. Treatment should be performed in accordance with the scheme 63, if doctor does not prescribe some other scheme.
Daily number of procedures 1.
2. Manual start.
3. Duration of the treatment course 10 (15) days.
The treatment should start as follows:
Enter correct data in accordance with the treatment scheme. If necessary, clean the focus of affection before using the
apparatus (remove the pus, treat the surface with alcohol or iodine solution). Apply vibrophones at the affected area
covered with a sterile gauze towel. Start the session.
Scheme 63. Treatment of furunculosis, carbunculosis and acne
L Single vibraphone; R No
Energy
Day
Duration,
Mode
L
minutes
Inflammation zone
1
1
1
7
2
2
2
10
3-10 (3-15)
5
2
15

GLAUCOMA
The treatment should be performed according to the recommendations of ophthalmologist and under his supervision.
The healing effect occurs due to increase of blood outflow and anti-edematous action.
1. Treatment should be performed in accordance with the scheme 64, if doctor does not prescribe some other scheme.
Daily number of procedures 2 (once in the morning, once in the evening).
2. Manual start.
3. Duration of the treatment course 84 days (12 weeks).
A patient should sit.
Every procedure consists of 3 sessions.
Session 1. Enter correct data for the Session 1. The vibraphone connected to the left port should be fixed at the Zone .
Start the session. When the session is over start the session 2.
Session 2. Enter correct data for the Session 2. Converters should be fixed as follows:
- connected to the left port at the Zone 2;
- connected to the right port during the morning procedure at the left eye zone. One vibraphone should be placed at
the temple zone, another at the eye (a closed eyelid should be covered with a gauze towel). Both vibrophones are to
be supported by hands.
During the evening procedure the converter connected to the right port is not in use.
Start the session. When the session is over start the session 3.
Session 3. Enter correct data for the Session 3. Converters should be fixed as follows:
- connected to the left port at the Zone 2;
- connected to the right port during the morning procedure at the right eye zone. One vibraphone should be placed
at the temple zone, another at the eye (a closed eyelid should be covered with a gauze towel). Both vibrophones are
to be supported by hands.
During the morning procedure the converter connected to the right port is not in use.
Start the session.

- 74 -

1
2
3
4
5
6
7
8-28

1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2

29-35
36
37-56

1
2
1
2

57-63
64
65-84

1
2
1
2

2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2

Session 3
Data adjustment
L Double vibrophone
R Double vibrophone
Energy
L
R
Zone Right eye
2
zone
1
3
1
3
2
1
3
2
1
3
2
1
3
2
1
3
2
1
3
2
1
3
2
Break
1
3
2
1
3
2
Break
1
3
2
1
3
2

Duration,
minutes

Mode

Duration,
minutes

Mode

Duration,
minutes

Mode

Number of procedure

Day

Scheme 64. Treatment of glaucoma


Session 1
Session 2
Data entry
Data adjustment
L Double vibrophone
L Double vibrophone
R No
R Double vibrophone
Energy
Energy
L
L
R
Zone
Left eye
Zone
2
zone
2
3
10
2
3
3
2
3
11
2
1
3
2
3
12
2
3
2
4
2
3
13
2
1
4
2
3
14
2
3
2
4
2
3
15
2
1
4
2
3
16
2
3
2
5
2
3
17
2
1
5
2
3
18
2
3
2
5
2
3
19
2
1
5
2
3
20
2
3
2
5
2
3
20
2
1
5
2
3
20
2
3
2
5
2
3
20
2
1
5
2
3
20
2
3
2
5
2
3
20
2
1
5
Break
Break
2
3
20
2
3
2
5
3
20
1
2
3
20
2
3
2
5
2
3
20
2
1
5
Break
Break
2
3
20
2
3
2
5
3
20
1
2
3
20
2
3
2
5
2
3
20
2
1
5

3
3
4
4
4
4
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5

HEMATOMAS AND INJURIES


Vibrophones are applied directly to the injured area. In the case of deep hematoma, two vibrophones are placed at
the angle of 900 and directed toward a hematoma (Picture 5).
In cases of inveterate hematomas, procedure duration may be increased up to 40 minutes and the course of treatment may be increase up to 3-4 weeks.
In cases of hematomas on the head or on any internal organs, the treatment can be performed only after consultation with a doctor and under his supervision.
1. Treatment should be performed in accordance with the scheme 65. Daily number of procedures 2.
2. Manual start.
3. Duration of the treatment course 1-6 (10) days.
The treatment should start as follows:
Enter correct data. Start the session.

- 75 -

1
2
3
4
5
6
(6-10)

1
2
1
2
1
2
1
2
1
2
1
2

Duration,
minutes

Mode

Duration,
minutes

Mode

Duration,
minutes

Mode

Number of
procedure

Day

Scheme 65. Treatment of hematomas and injuries


L Double vibrophone
R Double vibrophone
Injuries of extremities Injuries of inner organs
Injuries of head
Energy
Energy
Energy
L R
L
R
L
R
Trauma
Zone Trauma
Zone Trauma
zone

zone

zone
1
3 3
10
1
3
2
5
1
3
1
1
3 3
10
1
3
2
6
1
3
1
2
3 3
15
2
3
2
7
2
3
1
2
3 3
15
2
3
2
8
2
3
1
5
3 3
20
5
3
2
9
5
3
1
5
4 4
20
5
3
2
10
5
3
1
5
4 4
20
5
3
2
10
5
3
1
5
4 4
20
5
3
2
10
5
3
1
5
4 4
20
5
3
2
10
5
3
1
5
4 4
20
5
3
2
10
5
3
1
5
4 4
20
5
3
2
10
5
3
1
5
4 4
20
5
3
2
10
5
3
1

1
2
9
10
10
12
10
12
10
12
10
12

1-3
4-6
7-21

1
2
1
2
1
2

- 76 -

Duration,
minutes

Mode

Duration,
minutes

Scheme 66. Treatment of diabetic angiopathy


Session 1
Session 2
Data entry
Data adjustment
L Double vibrophone
L Double vibrophone
R Double vibrophone
R IR-matrix
Energy
Energy
L
R
L
R
Zone
Inflammation
Zone
32
zone
1
2
1
5
1
1
3
1
2
2
5
1
2
3
2
2
2
10
2
2
4
2
3
2
10
2
2
4
5
3
2
10
5
2
5
5
3
2
15
5
2
5

Mode

Day

Number of procedure

DIABETIC ANGIOPATHY
The treatment should be prescribed by the doctor and take place under the doctors supervision.
1. Treatment should be performed in accordance with the scheme 66, if doctor does not prescribe some other scheme.
Daily number of procedures 2 (once in the morning, once in the evening).
2. Manual start.
3. Duration of the treatment course 21 days. If necessary, the treatment course can be repeated after the break of 10 days.
One should lie supine on a soft bad.
During treatment course a vibrophone and an IR-matrix are to be replaced with each other for every new
procedure.
Every procedure consists of two sessions.
Session 1. Enter correct data for the Session 1. Converters should be fixed as follows:
- connected to the left port at the Zone ;
- connected to the right port at the Zone 32.
Start the session. When the session is over start the session 2.
Session 2. Enter correct data for the Session 2. Fixation of the converters at the zone of affected vessels should be prescribed by the doctor (e.g. feet, shins, thighs, buttocks, forearms, hands, groins, knees, elbows etc.).
Start the session.

LITERATURE AND VIDEO MATERIALS ON VIBROACOUSTIC THERAPY


1. Vasilyev .E., Kovelenov .Y., Kovlen D.V., Ryabchuk F.N., Fedorov V.. Resources of an organism
immunity, health, longevity. St-Petersburg, Vita Nova, 2004. 416 pages, pictures
This book is written by the group of Russian doctors and physicists. This revolutionary book came into existence after
the invention and study of the new resource of any living organism in 2002 -2003. This resource appeared to be not a mystical ground but evident and measurable physical phenomenon. This newly-discovered resource appeared to be as important
and indispensable to any human being as food, water and oxygen. The lack of it definitely proves that the organism is dead!
To form and keep this resource, a human organism daily spends as much energy as by 2-3 hours of active physical work.
The very attractive characteristic of this resource is that it can be received by an organism not only by consuming of food,
water and oxygen but also by some other quite easy methods. It is quite interesting that some of these methods can be found
among religious rules and ceremonies. Still the most effective methods became available only in the present time.
A human organism is managed so perfectly that the best way one can help it is the help by resources. Even first applications of the resource support method have brought to the defeat over chronic hepatitis and , chronic impaired
cardial function, serious achievements in sports, successes in the treatment of hypertension, glaucoma, osteochondrosis, arthrosis, burns and serious traumas.
The book in a popular way describes the new resource and opportunities, connected to it. It is also interesting because it contains the whole complex of practical recommendations which allow raising the immunity, physical and
mental abilities, defeating the diseases and postponing infirmity and ageing without serious changes in a life-style.
2. Vibroacoustic therapy. Collection of materials of 2003-2004 St-Petersburg, Vita Nova, 2005. 192 pages,
pictures
The book presents materials of the papers read at the Third National Research-Practical Conference on Vibroacoustics in Medicine (St-Petersburg, June 2004) as well as clinical reports, methodological recommendations and doctors
manuals prepared in 2003-2004.
3. Vibroacoustics in Medicine. Collection of the reports made at the conferences on vibroacoustic therapy.
St-Petersburg, Vita Nova, 2003. 240 pages, pictures
The book presents materials of the papers read at the Second National Research-Practical Conference on Vibroacoustics in Medicine (St-Petersburg, June 6-7, 2002). The goal of the conference was to generalize 8 years of experience in active application of vibroacoustic therapy in the medical practice.
Scientific editor of the book O.I. Efanov, Professor, Doctor of Medical Sciences, Academician of the Academy of
Medical and Technical Sciences, member of the International Academy of Informatization.
4. Vibroacoustic Therapy in a Complex Treatment of Patients. Doctors' Manual. Scientific editor Dr. of
Medicine, Assistant Professor .I. Kurtov. St-Petersburg, Vita Nova, 2003. 96 pages, pictures
The Manual contains modern conception on the role, played by vibroacoustic therapy in the complex treatment of
patients in cardiology, urology, gynaecology, surgery, traumatology and orthopedics.
5. F.N. Ryabchuk, V.A. Aleksandrova. Application of the Vibroacoustic and Infra-Red Effect for the Treatment of Children with Somatic Pathologies. Doctors' Manual. Reviewer N.V. Orlova, Professor of Pediatric Department 1 of Medical Academy of Postgraduate Education, St-Petersburg St-Petersburg: Publishing House of
Medical Academy of Postgraduate Education, 2003. 36 pages
The book summarizes the experience of application of the apparatuses Vitafon and Vitafon-IR in the children's
department 2 of Medical Academy of Postgraduate Education, St-Petersburg.
6. .F.Eryshev and others. Application of "Vitafon" Apparatus for the Treatment of Alcohol Abstinence
Syndrome. Doctors' and Patients' Manual. St-Petersburg, Vita Nova, 2003. 16 pages
The book summarizes the experience of application of the Vitafon apparatus for the treatment of alcohol abstinence syndrome in V.M.Behterev St-Petersburg Psychoneurological Institute and Vasileostrovsky Narcological Dispensary of St-Petersburg.
7. Y.S. Astahov and others. Application of "Vitafon" Apparatus for the Treatment of Open-angle Glaucoma.
Patients' Manual. St-Petersburg, Vita Nova, 2003. 8 pages, pictures
The book describes the treatment scheme and method of application of the "Vitafon" apparatus for the treatment of
open-angle glaucoma. The work was made at the Ophthalmologic Department St-Petersburg I.P.Pavlov State Medical
University.
8. Vibroacoustic resource support of the organism new approach to health. Popular science film. 42 minutes, formats: VHS, DVD.
One can buy these books at the book shops and drug stores of different towns and cities.
It is also possible to order these books by post via cash on delivery. Send your orders to the following address:
Vitafon Ltd., Saint-Petersburg 198096, post office box 28, Russia.
E-mail: info@vitafon.ru
Internet sites: www.vitafon.ru, www.microepm.ru, www.vsegdazdorov.ru.
Phone numbers: +7 (812) 783-3602; 747-2648; 747-2627; +7 (495) 118-8788; 788-4049.
- 77 -

GUARANTEES
The VITAFON-2 apparatus is certified in a proper way, answers all technical requirements and corresponds to
the sample approved.
The warranty period is 12 months from the purchasing date, on the condition that exploitation and storage rules,
given in the present User Manual, are followed and the Acceptance Act contains the manufactures stamp. The warranty storage period is 2 years from the manufacturing date.
If purchasing date is not indicated or there is no distributors stamp in the Acceptance Act and Guarantee Coupon,
the guarantee term begins from the manufacturing date.
Repair, change or return of the purchased apparatus is performed in accordance with the Russian legislation and
Consumers Remedy Law. If an apparatus is out of order within the warranty term, its owner can repair or change it for
free on the condition that the Guaranty Coupon is presented. The apparatus can be changed in the place where it was
bought.
Guarantee is not valid in the following cases:
- if there is no Guarantee Coupon;
- if stamps on the apparatus are damaged;
- if exploitation and storage rules were not followed by the user;
- if the apparatus is physically damaged.
In these cases the apparatus can be repaired at the expense of its owner.
Manufacturer is not to be blamed in cases of damages, produced in the main block by water or other liquids, substances or objects inside. The manufacturer is also not responsible for damages caused by natural conditions, fire or
other reasons not related to the manufacturer.
To find the nearest repairing center or get any other information on the exploitation of VITAFON-2, use the following phone numbers:
Byelorussia
Columbia
Estonia
Kazakhstan
Latvia
Lithuania
New Zealand
Russia
Serbia and
Montenegro
Ukraine

Minsk
Bogota
Tallinn
Alma-Ata
Riga
Vilnius
Auckland
Moscow
St-Petersburg
Beograd
Kharkov

+375 17
+57 1
+372
+7 327
+371
+370 5
+64 9
+7 495
+7 812
+381 11

233-98-97
250-18-09
621-36-88
254-67-00,
714-48-09
270-35-26
520-38-10
118-87-88,
783-36-02,
262-26-35,
334-06-44
+380 57 719-77-96

vibroacustica@yahoo.com
trico@hot.ee
254-57-52

788-40-49
747-26-48
328-30-82,

MANUFACTURERS LEGAL ADDRESS:


VITAFON Ltd.
23, Ogorodniy per.,
St-Petersburg, 198097, Russia

- 78 -

info@vitafon.co.nz
rtk@tlms.ru
info@vitafon.ru
nezelo@eunet.yu

- 79 -

You might also like