Item Description:

Three Phase slip ring induction motor coupled to DC shunt


generator, mounted on base plate with accessories.

Induction motor: 3, 5kW, 400V L-L (RMS) 50Hz, 1435 rpm (nominal), slip
ring induction motor. Star or delta connected (configurable either in
star or delta).
DC shunt machine: The shunt machine is used as loading unit and will be
used as DC generator. Armature voltage in the range 220-400V, 5kW, Speed
range 0 to 2000 rpm, shunt field separately excited. Shunt winding
rating: 220V to 400V range.
The induction machine should be of inverter grade and will be powered
from three phase inverter with switching frequency up to 20kHz.
The induction motor should run in field weakening mode with ratio up to
1.3 or machine should be designed to operate 0 to 2000rpm in both
directions. Please mention the field weakening ratio. The DC generator
should also be designed to run up to a speed of 2000rpm.
Coupling and Fitting: Both the induction motor and load unit should be
fitted with a mechanical coupling and mounted rigidly on a base plate.
The base plate should be fitted with anti-vibration pads to withstand
the vibrations. The base plate will be placed on the floor and will not
be fitted to the floor. The base plate should not move due to vibration
when the machines are mounted.
Mandatory Accessories: Following sensors need to be supplied with the
unit.
(i) Speed Encoder with at least 1024 pulses/rev or higher resolution.
(ii) Torque sensor fitted at the coupling of the two machines. Sensor
should be rated for 5HP machines.
Note that above accessories should be fitted to the above machine
set.
Safety: The above set will be used in the research lab so should meet
all the safety requirements for an experimental set up (as per
British/European/American standards).There should be safety couplingguard and guard to the exposed shaft portion if any. Base plate and
machine body should have provision to earth or grounding. All power
terminals should meet the safety standards for laboratory experimental
operation (as per British/European/American standards).
The machine set should have appropriate lifting hooks.
The motor generator set should be aligned properly and must be balanced.
The machine set should have good finish and elegant external look for
laboratory use. The quotation must be supplied with clear photographs of
the machine.
The electrical parameters (such as rotor and stator leakage inductance,
resistance and mutual inductance) or the blocked rotor and open circuit
test data must be supplied. The DC generator parameters such as motor
armature resistance, armature inductance are to be supplied. The moment
of inertia for the both machines is to be supplied.
Commissioning and assembly: Items can be shipped as a single unit with
machines fitted or two separate units. In the latter case the machines
should be fitted, balanced and commissioned by the supplier at the lab.
Care should be taken while packing and shipping. Items will not be
accepted if the unit is damaged during shipping. If the items are
damaged or shaft alignment is disturbed during transit, the complete
unit needs to be replaced. Under such conditions, repaired unit at the
customer site will not be accepted. The commercial quote must be
accompanied with the photograph of the unit and also detailed data
sheet.
A STANDARD WARRANTY OF 1 YEAR IS REQUIRED.