Professional Documents
Culture Documents
\EDIC@ EasyBloodGas
MANUAL STANDBY
CONFIRM?
TEST COMPONENTS
1
2
3
4
5
PUMP
VALVE
SAMPLER SWITCHES
DOOR SWITCH
REAGENT READER
REPLACE COMPONENTS
TEST FLUIDICS
1
2
3
4
5
1 FLOW
2 BUBBLE DETECTOR CAL
3 PUMP CAL
REAGENT MODULE
ELECTRODES
PUMP TUBING
SENSOR MODULE/SAMPLER
VALVE MODULE
SENSOR STATUS
ELECTRODE mVs
1 ELECTRODE mVs
2 BAROMETRIC PRESSURE
3 TEMPERATURE
1 CALIBRANT A mVs
2 CALIBRANT B mVs
PATIENT RESULTS
LAST CONTROL
1 LAST
2 ALL
3 ENTER ID#
1 LEVEL 1
2 LEVEL 2
3 LEVEL 3
QC RESULTS
PRINT QC STATS
1 LAST CONTROL
2 PRINT QC STATS
3 PLOT QC CHARTS
1
2
3
4
REFERENCE LIMITS
PLOT QC CHARTS
1
2
3
4
5
1
2
3
4
DIAGNOSTICS
1
2
3
4
TEST COMPONENTS
TEST FLUIDICS
SENSOR STATUS
PRIME FLUIDS
DAILY CLEANER
DAILY CLEANER
SCREENS
HOME MENU
1
2
3
4
5
*
*
CALIBRATE
DAILY CLEANER
SECOND MENU
CALIBRATE
CALIBRATION
SCREENS
SECOND MENU
1
2
3
4
5
6
STANDBY
REPLACE COMPONENTS
DIAGNOSTICS
STORED INFORMATION
SETUP MENU
ADJUST DISPLAY
LIFT SAMPLER TO
ANALYZE
HOME MENU
1
2
3
4
5
ANALYZE SAMPLE
ANALYZE QC
CALIBRATE
DAILY CLEANER
SECOND MENU
LEVEL 1
LEVEL 2
LEVEL 3
ALL LEVELS
pH
PCO2
STORED INFORMATION
1
2
3
4
5
6
PATIENT RESULTS
QC RESULTS
REFERENCE LIMITS
CALIBRATION DATA
REAGENT MODULE
PRINTOUTS
PO2
QC LEVEL 1
QC LEVEL 2
QC LEVEL 3
NORMAL
CRITICAL
CALIBRATION DATA
1 SLOPES
2 mVs
3 PUMP/BUBBLE DETECTORS
USER OPTIONS
1
2
3
4
5
ANALYZE QC
1
2
3
4
pH
PCO2
PO2
ALL CHARTS
DATE/TIME
LANGUAGE
CORRELATION
BAROMETRIC PRESSURE
SECURITY CODE
REAGENT MODULE
LEVEL 1
LEVEL 2
LEVEL 3
PROFICIENCY
1 STATUS
2 CONCENTRATIONS
CONFIGURATION
1 CAL FREQUENCY
2 AUTO STANDBY
3 PO2 /PCO2 /BP
CALIBRATE
PRINTOUTS
1
2
3
4
5
6
CALIBRATION
SCREENS
USER OPTIONS
CONFIGURATION
PRINTER OPTIONS
PATIENT LIMITS
QC LIMITS
PATIENT INFORMATION
8 HRS
OFF
mmHg
4 TEMPERATURE
5 Hb
C
g/dL
PRINTER OPTIONS
1 PRINTER
2 CALD VALUES
3 PRINTOUTS (x2)
ON
ON
OFF
DAILY CLEANER
DAILY CLEANER
SCREENS
SECURITY CODE
* * * *
SECOND MENU
1
2
3
4
5
6
STANDBY
REPLACE COMPONENTS
DIAGNOSTICS
STORED INFORMATION
SETUP MENU
ADJUST DISPLAY
CONFIRM?
SETUP MENU
1
2
3
4
5
6
1
2
3
4
5
USER OPTIONS
CONFIGURATION
PRINTER OPTIONS
SET REFERENCE LIMITS
PATIENT INFORMATION
DELETE DATA
QC LEVEL 1
QC LEVEL 2
QC LEVEL 3
NORMAL
CRITICAL
Default off
ADJUST DISPLAY
1 BRIGHTNESS
2 BRIGHTNESS
3 REVERSE
PATIENT INFORMATION
1
2
3
4
5
6
PATIENT ID
PATIENT TEMP
Hb
FIO2
TIME DRAWN
OPERATOR ID
OFF
OFF
OFF
OFF
OFF
OFF
DELETE DATA
1
2
3
4
5
ALL
ALL
ALL
ALL
ALL
QC LEVEL 1
QC LEVEL 2
QC LEVEL 3
QC LEVELS
PATIENT RESULTS
Table of Symbols
Important information
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
Sampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
Valve Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Printer Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
CPU Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Pre-Amp Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Pump Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Contact Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Display Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Valve Module Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Rear Housing Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Replacement List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Troubleshooting
Troubleshooting Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Troubleshooting Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
AC Power Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
DC Power Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Display Assembly Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Sensor Module Heat Control Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Valve Module Heat Control Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Printer Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Pump Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Valve Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Electrode Signal Input Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Serial Communication Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Repair
P R O D U C T
D E S C R I P T I O N
System Overview
he repairs and procedures in this manual are intended for authorized field service
engineers. Instructions describe the repair or replacement of all modular internal
components. It is recommended that the field service engineer also refer to the Operators
Manual for detailed instructions on operating and troubleshooting the analyzer to determine
if repair or replacement is necessary for an internal component. The Replacement List
indicates assemblies and components available to the field service engineer. Refer to the
Repair section of this manual for the CPU schematic. Before performing any service, the
reagent module must be removed.
P R O D U C T
D E S C R I P T I O N
Sensor Module
Contains heat control circuitry, sampler position
sensors, bubble detectors, door sensor, sampler, and
compression plate. The electrodes (PO2, PCO2, pH,
reference) are housed in the sensor module.
Sampler
Attached to sensor module. Contains sample probe
with probe wiper. The sampler is removable from the
sensor module as a replacement part.
P R O D U C T
D E S C R I P T I O N
Valve Module
Controls reagent (from reagent module) or air
selection. The valve module contains a ceramic valve,
motor, position detector and reagent preheater. There
are no serviceable parts inside the valve module.
Printer Assembly
Contains the printer mechanism, printer software, and
printer electronics. Prints analysis, calibration and
diagnostic data.
CPU Assembly
Contains transformer, software, inputs (electrodes,
barometric pressure, room temperature), motor
control circuitry, sensor module heat control, and 2
serial ports.
P R O D U C T
D E S C R I P T I O N
Pre-amp PC Board
Connects contact housing to the CPU assembly and
amplifies electrode signals.
Pump Assembly
Contains peristaltic pump with wire connector.
P R O D U C T
D E S C R I P T I O N
Contact Housing
Contains all contacts from the electrodes, as well as
the contact housing cable for connection to the CPU
assembly.
Display Assembly
(Analyzer operator interface.) Contains operator
keypad and LCD display.
P R O D U C T
D E S C R I P T I O N
P R O D U C T
D E S C R I P T I O N
Replacement List
The items on this list are available for purchase from
Medica Corporation.
Description
*Access Door
Catalog Number
(manufactured before April 2001)
6509
Access Door
6541
Barcode Reader
6520
6514
6516
Compression Plate
6508
6517
6538
CPU Assembly
6539
Display Assembly
6540
Operators Manual
6501
pH Electrode
6201
PCO2 Electrode
PO2 Electrode
6202
Pump Assembly
6513
Reagent Module
6101
Reference Electrode
6204
Printer Assembly
6510
Pump Tubing
6504
Sampler
6306
Sensor Module
6506
Serial Cable
6518
Service Manual
6502
Valve Module
6507
6524
6525
Pre-Amp PC Board
6536
6203
T R O U B L E S H O O T I N G
Troubleshooting Overview
This section describes troubleshooting the mechanical
and electrical functions of the EasyBloodGas
analyzer. Medica recommends that dealers maintain
an inventory of operational test assemblies for
troubleshooting purposes. These test assemblies can
substitute the questionable assemblies to confirm
failures. Only those items listed on the Replacement
List are available for purchase. For information on
troubleshooting the fluidic system and electrodes,
refer to the EasyBloodGas Operators manual.
Troubleshooting Guide
Determine the problem and corrective action needed.
Action Steps
Problem/Description
Problem Source
Power malfunction
Defective display
AC power verification
DC power verification
Display assembly verification
EPROM verification
Printer malfunction
Printer verification
No pump movement
Pump verification
No valve movement
Valve verification
Serial cable
Transmission Protocol
T R O U B L E S H O O T I N G
AC Power Verification
When the power is first turned on, the EasyBloodGas
analyzer will beep, and the display will light up. If
this fails to happen, follow the verification procedure
in this section. Review this entire section before
replacing any components. First confirm that the
power cord is inserted securely inside the AC power
cord socket.
Warning: Careless handling of AC power,
especially 220V, can be hazardous or
fatal. Only qualified personnel should
work with AC power.
Fuse Holder
EasyBloodGas
100-115~VAC
0.8A
50/60 Hz
220~VAC
0.4A
50/60 HZ
Max pump
working pressure:
15PSI
NOTE:
For 100-115 VAC
operation, the
voltage select
switch marked
110 and 220 must
be in the 110
position.
004434-001 R0
!
SERIAL NO.
ETL LISTED
CONFORMS TO
UL STD 3101-1 C
78559 CERTIFIED TO
CAN/CSA STD C22.2 NO. 1010.1
\EDIC@
110
220
220
110
or
Factory-set Power Voltage Setting
3
F0.8 or F0.4
T R O U B L E S H O O T I N G
Test Point
Table 1
X1
X4
X5
DC Power Verification
Voltage
+ 5.00V
+ 0.15V
+ 24.00V + 5.00V
X6
+ 19.00V + 4.00V
- 19.00V + 4.00V
X7
+ 5.00V
X8
GND Reference
X9
- 12.00V + 0.25V
+ 12.00V + 0.25V
Table 2
X2
GND Reference
X3
+ 8.5V
Table 3
X10
+ 0.25V
+ 2.00V
TP9
GND Reference
TP10
TP12
+ 2.000V + .003V
- 12.00V + 0.25V
TP13
- 5.00V
TP14
+ 5.00
TP15
+ 15.00V + 1.00V
+ 0.25V
+ 0.25V
T R O U B L E S H O O T I N G
11
T R O U B L E S H O O T I N G
SECOND MENU
1
2
3
4
5
6
STANDBY
REPLACE COMPONENTS
DIAGNOSTICS
STORED INFORMATION
SETUP MENU
ADJUST DISPLAY
DIAGNOSTICS
1
2
3
4
TEST COMPONENTS
TEST FLUIDICS
SENSOR STATUS
PRIME FLUIDS
SENSOR STATUS
1 ELECTRODE mVs
2 BAROMETRIC PRESSURE
3 TEMPERATURE
12
T R O U B L E S H O O T I N G
SECOND MENU
1
2
3
4
5
6
STANDBY
REPLACE COMPONENTS
DIAGNOSTICS
STORED INFORMATION
SETUP MENU
ADJUST DISPLAY
DIAGNOSTICS
1
2
3
4
TEST COMPONENTS
TEST FLUIDICS
SENSOR STATUS
PRIME FLUIDS
SENSOR STATUS
1 ELECTRODE mVs
2 BAROMETRIC PRESSURE
3 TEMPERATURE
Printer Verification
SECOND MENU
1
2
3
4
5
6
STANDBY
REPLACE COMPONENTS
DIAGNOSTICS
STORED INFORMATION
SETUP MENU
ADJUST DISPLAY
SETUP MENU
1
2
3
4
5
USER OPTIONS
CONFIGURATION
SET REFERENCE LIMITS
PATIENT INFORMATION
DELETE DATA
CONFIGURATION
1
2
3
4
5
6
13
CAL FREQUENCY
AUTO STANDBY
PRINTER
PO2 /PCO2 /BP
TEMPERATURE
Hb
CONFIRM?
8HRS
OFF
ON
mmHg
C
g/dL
T R O U B L E S H O O T I N G
Pump Verification
For suspected pump problems, proceed to the
SECOND MENU and select DIAGNOSTICS,
TEST COMPONENTS, then PUMP. The screen
displays ROTATING.... Look for pump rotation.
When full pump rotation occurs, FULL
ROTATION? is displayed. Press YES and PASS
is displayed.
SECOND MENU
1
2
3
4
5
6
STANDBY
REPLACE COMPONENTS
DIAGNOSTICS
STORED INFORMATION
SETUP MENU
ADJUST DISPLAY
DIAGNOSTICS
1
2
3
4
TEST COMPONENTS
TEST FLUIDICS
SENSOR STATUS
PRIME FLUIDS
1
2
3
4
5
Valve Verification
SECOND MENU
TEST COMPONENTS
1
2
3
4
5
6
PUMP
VALVE
SAMPLER SWITCHES
DOOR SWITCH
REAGENT READER
STANDBY
REPLACE COMPONENTS
DIAGNOSTICS
STORED INFORMATION
SETUP MENU
ADJUST DISPLAY
DIAGNOSTICS
1
2
3
4
TEST COMPONENTS
TEST FLUIDICS
SENSOR STATUS
PRIME FLUIDS
TEST COMPONENTS
1
2
3
4
5
PUMP
VALVE
SAMPLER SWITCHES
DOOR SWITCH
REAGENT READER
14
T R O U B L E S H O O T I N G
Pin Type
RS-232
Pin Assignments:
Pin 1
Pin 2
TxD
Pin 3
RxD
Pin 4
RTS
Pin 5
CTS
Pin 7
Signal Ground
Transmission Rate
Bit Configuration
Bit Parity
None
Control Codes
None
15
R E P A I R
Repair Overview
To replace EasyBloodGas assemblies, disassemble the
analyzer in the order described in this section, referring to the photographs for assistance. After rear
housing removal, the printer assembly and display
assembly can be removed without removing the CPU.
Access to all other assemblies requires removing the
CPU board.
Before performing major repairs, print the following
settings from the SECOND MENU by selecting
STORED INFORMATION, then
PRINTOUTS.
USER OPTIONS, CONFIGURATION,
PRINTER OPTIONS,
PATIENT LIMITS, QC LIMITS,
PATIENT INFORMATION
SECOND MENU
1
2
3
4
5
6
STANDBY
REPLACE COMPONENTS
DIAGNOSTICS
STORED INFORMATION
SETUP MENU
ADJUST DISPLAY
4
STORED INFORMATION
1
2
3
4
5
6
PATIENT RESULTS
QC RESULTS
REFERENCE LIMITS
CALIBRATION DATA
REAGENT MODULE
PRINTOUTS
Required Tools
Only 4 tools are required to remove or install the
EasyBloodGas electronic and mechanical assemblies:
Phillips screwdriver, #2 magnetic
Phillips screwdriver, #1
Needle nose pliers
Flash Device Removal Tool (provided with software
updates)
16
R E P A I R
17
R E P A I R
18
R E P A I R
R E P A I R
Printer
Assembly
P5
User Options
Configuration
QC Limits
Patient Limits
QC Results
P1
P2
P13
P7
P6
20
R E P A I R
User Options
Configuration
QC Limits
Patient Limits
QC Results
Before changing the flash device on the EasyBloodGas, follow these steps:
SECOND MENU
Press STANDBY
MANUAL STANDBY
333
SERVICE MENU
REPLACE FLASH?
Press yes
CONFIRM?
Press yes
Disconnect power
21
R E P A I R
After changing the flash device on the EasyBloodGas, power up and follow
these steps:
SECOND MENU
Press STANDBY
MANUAL STANDBY
777
QC MENU
CONFIRM?
Press yes
22
R E P A I R
23
R E P A I R
Before changing the Pre-amp PC board on the EasyBloodGas, follow these steps:
SECOND MENU
Press STANDBY
MANUAL STANDBY
Press 333
SERVICE MENU
REPLACE PRE-AMP?
Press yes
CONFIRM?
Press yes
Disconnect power
Pre-amp PC Board/Contact
Housing Removal
Remove the screw which connects the grounding strap
to the pre-amp PC board.
Remove the 6 screws from the pre-amp PC board.
Remove the pre-amp PC board with the contact housing attached from the front housing. The contact
housing can be removed from the pre-amp PC board
by gripping both sides and pulling apart.
24
R E P A I R
After changing the Pre-amp PC board on the EasyBloodGas, power up and follow
these steps:
SECOND MENU
Press STANDBY
MANUAL STANDBY
777
QC MENU
CONFIRM?
Press yes
Pre-amp PC Board/Contact
Housing Installation
Align the opening on the pre-amp PC board with the
CPU board mounting post. This will align the screw
holes. Insert the 6 screws to secure. Do not
overtighten. Connect the grounding strap to the preamp PC board.
25
R E P A I R
26
R E P A I R
27
R E P A I R
Sampler Removal
If the sampler requires replacement, it can be
removed from the sensor module. Using a coin or
fingers, remove the retaining screw from the sensor
module.
Sampler Installation
Install the new sampler (includes sample probe and
wiper) in the sensor module. Insert the sampler
alignment pin into the sensor module. Rotate the
sampler into the closed position.
Install the screw mount (includes spring) into the
sampler. Replace the screw and tighten.
28
R E P A I R
In-field Reprogrammability/
Software Updates
If software updates are required, follow the
instructions provided with the new software.
29
N O T E S
30
004866-001 R4
\EDIC@