Possible Causes
Corrective Actions
Distal line kinked, IV - Check distal tubing
site clotted or
- Check IV site
positional, pressure - Reset pressure limit
limit set too low
Prox Occl A Proximal line
- Examine line A/B for kinks or
(B) at start upobstructed (line A or closed clamps
B container or syringe - Open any closed proximal clamps
disconnected or
- Open and close door
obstructed during start- For syringe, select BACKPRIME
up)
Prox Air A/B Proximal line
- Examine line A/B for kinks or
Air
obstructed (line A or closed clamps or air in cassette
B container
- Open any closed proximal clamps
disconnected or
- Press START
obstructed)
Prox Air A/B Empty container or - Check for clamps or empty
Backprime excess air in line
container
- Back prime into line B container
or syringe.
Distal Air
No Action
Alarm
Pump programming
incomplete
Line A/ B
VTBI
Complete
Cassette Test
Failure
VTBI is complete.
Lockout
Violation
SYRINGE PUMP
POLICY: PPG NUR 94
This policy and procedure guide and describes the procedure of intravenous
infusion or antibiotic administration by using different types of infusion
pump in Mouwasat Hospital.
DEFINITION:
Device used to administer medications or IV fluids slowly and accurately
over a desired length of time by using a syringe.
POLICY
Needs written order to use syringe pump.
The medications and in fluids which needs to be given via syringe
pump to be given as per guidelines to the medication.
ALARMS/ CAUSES
Syringe disengagement during
infusion syringe size 1
Cannot start
CORRECTIVE ACTION
Press the STOP switch to the buzzer.
Close the IV line.
Prime the IV line.
Open the IV line.
Press the START switch to restart infusion.
Silence the buzzer by pressing STOP switch
Close the IV line.
Load the TERUMO brand properly.
Reset the syringe and verify the plunger is
held firmly against the slider.
Prime the IV line.
Open the IV line.
Press the START switch to restart infusion.
Press the STOP switch to the buzzer.
Check for a kinked or clogged tubing.
Prime the line and make sure the plunger is
held firmly against the slider.
Eliminate the cause of occlusion.
Then restart the infusion.
Plug the pump into the AC or DC main as
soon as possible.
Check the flow rate if it is 0.0 ml/u
Reset the flow rate (less than 300 ml/u)
Reset delivery limit.
Make silence the buzzer.
Replace the syringe to continue the
infusion.