This tutorial will cover the calibration of Q1 and Q3 using the PPG Standard solution and the PPG
3000 Standard solution for the Q TRAPTM System and the 4000 Q TRAPTM System.
A method for manual calibration will be described as well as an automatic method for PPG
optimization and calibration. The method described here uses direct infusion via the syringe pump
in positive ion mode using the TurboIonSpray source or the TurboVTM Source. The PPG
solutions are available in the Mass Spectrometer Test Kit P/N 401936.
Quadrupole Calibration
2.1
When you are back to the Hardware Configuration Editor, click Activate Profile. The computer
will establish communication with the devices in the profile and checkmark will turn green. Close
the editor.
Quadrupole Calibration
2.2
In this window, define the syringe diameter and flow rate. For a 1 mL syringe, the syringe
diameter is 4.61 mm (see table in Table of Hamilton Syringe Diameters for other diameters). Click
Start Syringe Pump once the parameters have been adjusted. Use a flow rate of 5-10 uL/min for
infusion of the calibrant.
Note: If the flow rate or syringe diameter must be changed once the syringe has been started, enter
the new value and click Set Flow Rate. Changes will not take effect until you click Set Flow Rate.
For the 4000 Q TRAPTM System: Fill the 1 mL syringe with the appropriate PPG solution and
place it in the external syringe pump. For calibration of the Q1 and Q3 quadrupoles in positive
mode, use the diluted PPG Standard solution (see Dilution of Standard PPG Solution). For
calibration of Q1 and Q3 in negative mode, use the PPG 3000 Standard solution. Set the
diameter on the syringe pump to 4.61 and the flow rate to 5-10 L/min.
Quadrupole Calibration
2.3
Set the Declustering Potential (DP) on the Compound tab; start with a value of 110. Alternatively,
individual settings for DP can be set for each PPG mass. Right-click the gray area beside the
masses and select Declustering Potential from the menu. Enter the values 110 respectively in the
rows, these values can now be separately adjusted for each PPG ion as each DP value is
optimized.
Quadrupole Calibration
Quadrupole Calibration
2.4
Select the PPGs Pos from the Standard menu. Ensure the Positive polarity is checked, then select
the two methods that were created in the previous section for the Q1 Method and the Q3 Method.
A method must be selected for LIT Method, any LIT method can be chosen as it wont be used in
this tutorial (a method for LIT calibration using PPG will be created in Section 3.8 in the tutorial
on the Calibration of the Linear Ion Trap Scan Speeds).
Quadrupole Calibration
Click the Reference button to view the reference table for the PPG positive calibration.
You can select which PPG ions to use during calibration by checking/unchecking the Use boxes.
Click Update Ref to save any changes, then click Close.
Note: The 4000 Q TRAPTM System does not require mass dependent parameter optimization.
Click OK to close the Tuning Options window.
Quadrupole Calibration
2.5
Quadrupole Calibration
The PPG masses will appear in the bottom right pane, right-click on the pane and select Open File
to view all of the PPGs in one window.
Ensure each PPG ions meets its intensity specification, as reported in the Appendix.
Quadrupole Calibration
2.6
Choose the reference file for this calibration from the Standard menu: PPGs Pos. Notice that the
quadrupole, the resolution and the polarity used in the data collection are automatically entered
into this window.
You can directly edit the Reference Table Editor by clicking Edit.
Quadrupole Calibration
10
The first pane that opens is the Mass Calibration Report pane. The top graph illustrates the mass
shift in the PPG ions since the last calibration. If the mass spectrometer is mass calibrated, the
typical mass shifts seen are 0.1 to 0.2 amu.
The middle pane illustrates the peak width at half height.
The bottom pane shows the changes in intensity compared to the last calibration.
Quadrupole Calibration
11
The second pane that opens is the Mass Calibration Results pane. This is a numerical
representation of the calibration results. Check the Peak Width column in the report and verify that
the peak widths are between 0.6 0.8 for all of the PPG ions.
If the data spectrum looks good and the mass shifts arent too large, click the Replace Calibration
button on the Mass Calibration Report pane. If all the mass shifts fall between the two dotted lines
shown on the graphical report, the instrument does not need to be recalibrated.
The following message is displayed, click Yes to replace the instrument calibration.
Quadrupole Calibration
12
The values for the new calibration are now shown in the bottom of the mass calibration report. The
report shown below is for the Q TRAPTM System
Note: If there is a drastic change in mass or intensity of one of the fragment ions, it is important to
determine why before this ion is used in calibration. Click No and refer to the mass calibration
report shown above. In the Found Mass column, determine the mass of the ion that is off. Next,
observe the quality of this ion in the raw data spectrum. If the incorrect ion was chosen, the Search
Range used in the LIT Mass Calibration may need to be widened or narrowed to allow the
software to find the correct ion. The software will use the centroid of the most intense peak in the
search range for calibration.
Quadrupole Calibration
13
2.7
Select the appropriate calibration standard from the Standard list. Select the quadrupole and
resolution to be optimized and calibrated. Check the Mass Calibrate on Success button to ensure
the calibration is updated after the resolution offsets have been adjusted.
For the Q TRAP System only, the Update Mass Dependent Parameters box can also be checked
to determine the optimal values for CEP and CXP at the low and high masses as part of the
resolution optimization process. Extrapolated values will be applied in experiments where the CEP
and CXP is not specified.
The Peak Search Parameters can be used to control where the automatic tool will look for the
peak to be calibrated on. Use Search range settings of 0.5 amu to ensure the program doesnt pick
another strong isotope in the window. Threshold is best left at the default value of 200 cps.
Quadrupole Calibration
14
Select the polarity to be calibrated. Select only one polarity at a time as the calibration solutions
are different for each polarity. A direct link to the Reference Table Editor is available from this
window by clicking Edit.
Click Start to begin automatic calibration.
Note: A text file is generated and displayed during the optimization process. This file records all
changes as it goes through the process. View the text file and the data file, as they are being
generated, to follow the progress of the optimization you may need to minimize windows. If an
ion is not found, abort the optimization. Expand the search range and re-start the process.
If the specifications are not reached within 20 iterations, the automated calibration will be aborted.
If it is able to reach the specifications, it will note successful completion in the text file and on the
screen.
Quadrupole Calibration
15
Before saving the Resolution Optimization results, review the generated text report carefully to
ensure that the desired sensitivity and resolution was obtained, and that the isotope peak used for
calibration was correct.
Click Yes to accept the Resolution Optimization results. Now the Q1 and Q3 are both calibrated in
positive mode.
Calibrate Q1 and Q3 in Negative Mode
Change solutions from PPG Standard to PPG 3000 and repeat the process described above to
calibrate Q1 and Q3 in negative mode.
Quadrupole Calibration
16
2.8
Appendix
Using the dilution solvent prepared above, perform a 1:50 dilution of the Standard PPG
solution.
For accuracy, it is recommended that a 1:10 dilution is prepared, followed by a 1:5 dilution.
Diameter
(mm)
1000
4.61
500
3.26
250
2.3
100
1.46
50
1.031
Quadrupole Calibration
17
Q1 (cps)
Q3 (cps)
FWHH (amu)
59.050
2.0 E6
1.0 E6
0.6 - 0.8
175.133
2.0 E6
1.0 E6
0.6 - 0.8
616.464
2.5 E6
1.0 E6
0.6 - 0.8
906.673
1.5 E7
8.0 E6
0.6 - 0.8
1254.925
2.0 E6
1.0 E5
0.6 - 0.8
1545.134
5.0 E5
N/A
0.6 - 0.85
Q1 (cps)
Q3 (cps)
FWHH (amu)
44.998
N/A
N/A
0.6 - 0.8
585.385
N/A
N/A
0.6 - 0.8
933.636
1.0 E6
2.0 E5
0.6 - 0.8
1223.845
N/A
N/A
0.6 - 0.8
1572.097
1.0 E5
N/A
0.6 - 0.8
Quadrupole Calibration
18
Q1 (cps)
Q3 (cps)
FWHH (amu)
559.050
3.0 E7
2.0 E7
0.6 - 0.8
175.133
1.0 E7
1.0 E7
0.6 - 0.8
616.464
1.5 E7
1.0 E7
0.6 - 0.8
906.673
2.0 E7
2.0 E7
0.6 - 0.8
1254.925
3.0 E6
3.0 E6
0.6 - 0.8
1545.134
1.0 E6
5.0 E5
0.6 - 0.8
2010.5
1.5 E6
1.0 E6
0.6 - 0.85
2242.6
1.0 E6
8.0 E5
0.6 - 0.85
Q1 (cps)
Q3 (cps)
FWHH (amu)
44.998
N/A
N/A
0.6 - 0.8
585.385
N/A
N/A
0.6 - 0.8
933.636
2.0 E7
1.0 E7
0.6 - 0.8
1223.845
N/A
N/A
0.6 - 0.8
1572.097
1.0 E5
N/A
0.6 - 0.8
1863.3
N/A
N/A
0.6 - 0.85
2037.4
3.0 E7
N/A
0.6 - 0.85
Quadrupole Calibration