STANDARD COMPOSITION:

Servo Control Mode: air temp,skin temp;


Instant Parameter Displaying: set temp, air temp ,skin temp,heater power
rate;
Audible & Visual Alarms: power failure, sensor failure, higher & lower
temp, air& skin temp override,fan failure,system failure;
>37 Preset Mode: power failure reset memory, alarm silence, keypad
lock, self-test function;
Hood with dual slope: 6 hand-ports; left & right cabinets, top-shelf;
RS-232 Output Connector (Use for network system ).
SPECIFICATION
POWER
Model B-2000-1E...............................................................................................................................110 / 120 Vac,
50Hz / 60Hz
Model B-2000-2E...............................................................................................................................220 / 240 Vac,
50Hz / 60Hz
Power Input ....................................................................................................................................................................
400VA
Refer to IEC60601-1, this equipment belongs to Class, Type BF. Its operation is continuous.
ALARM
Power Alarm
Power Failure.................................................Activates if primary power to the Incubator fails or the power wire is
accidentally disconnected from the wall receptacle.
Fan Alarm
Fan Failure.........................................................................................................................Activates if Fan motor
malfunctions.
Sensor Alarm
Air Sensor Failure...............................................................................................Air Sensor malfunction, refer unit to
service.
Sensor Out of Position................................................................Activated when Sensor is not in correct position to
perform calibration or continue operation.
Skin Sensor Disconnected........................................................Activated during Skin Control Mode when Skin Sensor
is disconnected from Temperature Controller.
Skin Sensor Failure................................................................................................Activated when Skin Sensor
malfunctions.

Deviation Alarm
High Air Temperature.......................................................Activated if displayed Air Temperature is 1.5above set
point.
Low Air Temperature........................................................Activated if displayed Air Temperature is 2.5below set
point.
High Skin Temperature...................................................Activated if displayed Skin Temperature is 1.0above set
point.
Low Skin Temperature.....................................................Activated if displayed Skin Temperature is 1.0below set
point.
Over Temperature Alarm
Over Air Temperature......................................................Activated when displayed Air Temperature was approach
38 for set temperature 37, or approach 40 for set temperature >37 in Air Mode.
Over Skin Temperature..................................Activated when displayed Skin Temperature was approach 38 for set
temperature 37, or approach 39 for set temperature >37 in Skin Mode.
System Alarm
System Failure ..................................................................................Activated if Temperature Control System
malfunctions.
TEMPERATURE CONTROL RANGE AND CORRELATIVE SPECIFICATIONS
Air Temperature........................................2037,for set temperature 37, 3739,for set temperature
>37
Skin Temperature......................................2037,for set temperature 37, 3738,for set temperature
>37
Temperature Rise
Time ..................................................................................................................................................... 30min
Temperature
Variability....................................................................................................................................................... 0.5
Temperature
Uniformity......................................................................................................................................................... 1
Skin Sensor
Precision.......................................................................................................................................................... 0.3
ENVIRONMENT REQUIREMENT
Environment Temperature
Operation Temperature Range.............................................................................................................................+20
+30
Store Temperature Range......................................................................................................................................-40
+55
Environment Humidity
Operation Humidity

Range............................................................................................................................................30%75%
Store Humidity
Range...................................................................................................................................................10%100%
Atmospheric Pressure Range........................................................................................................................... 700hPa
1060hPa
OTHER SPECIFICATION
Noise Level Within Hood Environment.................................................................... 55dB(Environment noise below
45dB)
Carbon Dioxide (CO2) Level within the Hood.......................................................Less than 0.5% when a mixture of
CO2 in air is delivered at 750ml/min at a point 10cm above the center of the mattress.
Air Velocity Over Mattress..................................................................................................................................less
than 0.35m/s
Infant Bed Max
Load...................................................................................................................................................................10kg
Transfusion Shelf Max
Load........................................................................................................................................................2kg
Cabinet Max
Load ....................................................................................................................................................................... 6kg

FUNCTION OF
Make the alarm to be silence for 10 minutes by press this key, following:
Deviation Alarm

SILENCE KEY

Sensor Alarm
Fan Alarm
FACTORY DEFAULT SETTINGS
Temperature Control Mode.......................................................................................................................Air
Temperature Mode
Air Set
Temperature.....................................................................................................................................................................3
2
Skin Set
Temperature...................................................................................................................................................................34

Temperature
Units............................................................................................................................................................................
NOTE: Open Access Panel or Access Door or the use of supplies or other equipment within the Incubator, which can
alter the air flow pattern, may effect temperature uniformity, temperature variability, the correlation of the Incubator
temperature reading to center mattress temperature and skin temperature.

Beijing julongsanyou Technology Co.,Ltd


HEAD OFFICE:
Address:No.19Deshanli,changxindian,FengtaiDistrict,Beijing
p.c:100072
Tel: 010-83841872 / 83861661
Fax: 010-83861661-807
Mobille: 13301395455 / 15601209824
Company email: bjjlsy@bjjlsy.com
Sales email:sales1@bjjlsy.com

sales3@bjjlsy.com

sales4@bjjlsy.com

You might also like