Professional Documents
Culture Documents
discussions, stats, and author profiles for this publication at: http://www.researchgate.net/publication/50867783
CITATIONS
4 AUTHORS, INCLUDING:
Abdelkader Boubetra
Max Feinberg
6 PUBLICATIONS 77 CITATIONS
SEE PROFILE
SEE PROFILE
In Europe, the Drinking Water Directive of the European Commission indicates which methods (most of
which are CEN/ISO-standardized methods) should be used for the analysis of microbiological parameters
(European Commission, Environment, Council Directive 98/83/EC of 3 November 1998). According to the
Directive, alternative methods may be used, providing it can be demonstrated that the results obtained are at
least as reliable as those produced by the methods specified. The prerequisite for the routine use of any
alternative method is to provide evidence that this method performs equivalently to the corresponding
reference method. In this respect, the ISO 16140 standard (ISO, ISO 16140. Microbiology of Food and Animal
Feeding StuffsProtocol for the Validation of Alternative Methods, 2003) represents a key issue in generating such
a procedure based on an interlaboratory study. A new statistical tool, called the accuracy profile, has been
developed to better interpret the data. The study presented here is based upon the enumeration of Escherichia
coli bacteria in water. The reference method may require up to 72 h to provide a confirmed result. The aim of
this publication is to present data for an alternative method by which results can be obtained in 18 h
(Colilert-18/Quanti-Tray) based upon defined substrate technology (DST). The accuracy profile is a statistical
and graphical decision-making tool and consists of simultaneously combining, in a single graphic, expectation tolerance intervals (-ETIs) and acceptability limits (). The study presents the validation criteria
calculated at the three levels of contamination used in the trial for a equal to 80% and a equal to 0.3 and
combines the accuracy profiles of Escherichia coli for a of 0.3 log10 unit/100 ml, a of 0.4 log10 unit/100
ml, and a of 80% or 90%. Several interesting conclusions can be drawn from these data. The accuracy profile
method has been applied to the validation of the Colilert-18/Quanti-Tray method against reference method ISO
9308-1 (ISO, ISO 9308-1. Water QualityDetection and Enumeration of Escherichia coli and Coliform Bacteria.
Part 1. Membrane Filtration Method, 2000), using a of 80% and a of 0.4; the alternative method can be
validated between 1.00 and 2.05 log10 units/100 ml, equivalent to 10 to 112 CFU/100 ml.
result is the mode of a statistical distribution of values, but
around this mode there are values with lower probabilities.
Ideally, each MPN should be expressed with a confidence interval according to this fact. As microbiological methods are
empirical (also called direct) analytical techniques, the measurand is highly dependent upon the operating procedure.
Therefore, the measurand, as defined by a reference
method, can be somewhat different when determined by an
alternative method. For practical purposes, however, and to
undertake the validation, it is necessary to make this approximation. This situation is rather typical when alternative and
reference methods are compared.
Taking into account the points highlighted above, the prerequisite for the commercial retail and routine use of any
alternative method is to provide evidence that this method
performs equivalently to the corresponding reference method.
To provide such evidence, the manufacturers of alternative
proprietary kits, the food and beverage industry, the public
health services, and other authorities require a reliable and
commonly agreed upon procedure for the validation of such
alternative methods. In this respect, the ISO 16140 standard is
a key issue in generating such a procedure (13).
As suggested in its title, ISO 16140 (13) separately proposes
validation protocols for both quantitative and qualitative methods (1). An interesting requirement of the standard for quan-
Until now, there has been little formal guidance on procedures for adopting alternative methods for determining levels
of microbes in water. From a metrological point of view, the
first step in developing a procedure is to define the measurand.
The measurand itself may be simply defined as the quantity
intended to be measured (in reference 16, see section 2.3).
Due to the nature of microorganisms and the well-recognized
concept of CFU, the currently accepted method for defining
the measurand and for ensuring traceability in microbiology
consists of using a reference (or official) method. This is usually
a historic method which has been standardized and is recognized as reliable by the community of microbiologists and
regulatory bodies. We must keep in mind, however, that microbiological methods, even if they have the status of reference
methods, are based in counts of discrete units, and hence their
uncertainties have intrinsic, unavoidable, stochastic components that make their values higher than the uncertainties of
most chemical methods. This is still more accentuated in the
most probable number (MPN) methods, where the calculated
titative method validation is the organization of an interlaboratory study in accordance with the recommendations of ISO
5725 (15a). Some years ago, microbiologists were reluctant to
participate in collaborative studies due to the assumed instability of samples. Much improvement has occurred with regard
to stabilization of samples, and interlaboratory studies are now
commonly used for proficiency testing programs. This classical
collaborative approach can now also be applied to method
validation in microbiology; sample instability and organizational constraints are no longer an issue (2, 4, 12).
The classical statistical strategy employed for the interpretation of data in many validation procedures is based largely on
null-hypothesis testing. This type of analysis aims to demonstrate that an alternative method does not produce results
significantly different from those of the reference method. This
strategy presents many drawbacks that have been described
extensively in recent publications (7, 18). The most striking
observation was that the more uncertain the results obtained
by an alternative method are, the easier the validation is. For
this reason, a new statistical tool, called the accuracy profile,
has been developed to better interpret the data of a validation
study, so that misleading conclusions are avoided. This methodology has been extensively applied to chemical analytical
methods and, more recently, in the field of food microbiology (8).
The study presented here is based upon the enumeration of
Escherichia coli bacteria in water. The choice to use the accuracy profile for this parameter is based largely upon recent
changes introduced by new European regulations and subsequently by regulation 2073/2005 of the Commission of the
European Communities and its amendments for microbiological criteria (3).
The objective of these regulations is to ensure that drinking
water is free of pathogens such as viruses, protozoa, and bacteria. Waterborne pathogens cause diseases such as hepatitis,
giardiasis, and dysentery. The analysis of water for the presence of specific harmful viruses, protozoa, and bacteria is timeconsuming and thus expensive. In addition, not all analytical
laboratories are equipped and approved to proficiently perform the required testing. Water testing for the presence of
specific organisms is therefore limited to investigating specific
waterborne disease outbreaks.
E. coli and coliform bacteria are a broad class of bacteria
found in the environment and also in the feces of humans and
other animals. Therefore, the presence of coliform bacteria
and, in particular, E. coli in drinking water may indicate the
presence of harmful, disease-causing organisms. For this reason, the enumeration of E. coli cells in water is increasingly
used to assess water quality.
The current reference method has several limiting factors, in
particular, time, as it may require up to 72 h to provide a
confirmed result. The aim of this publication is to present data
for an alternative method by which results can be obtained in
18 h. In the context of a validation study, a collaborative study
was organized and data were collected according to the guidelines of ISO 16140 (13). The interpretation of these data using
the accuracy profile approach is presented, and the fitness for
purpose of this alternative method versus the reference
method is ascertained.
3361
Add the contents of one blister pack to a 100-ml room temperature water
sample in a sterile vessel.
Cap the vessel, and shake it until the reagent is dissolved.
Pour the sample-reagent mixture into a Quanti-Tray and seal it in an
IDEXX Quanti-Tray sealer.
Incubate the sealed tray at 36 2C for 18 to 22 h.
Read the results according to the result interpretation table, and count the
number of positive wells.
No confirmations are needed. The most probable number (MPN) can be calculated from the number of positive wells (see the Appendix) or read in the table
provided with the trays to convert the number of positive wells to MPN format.
The values in this table agree with those calculated with the FDAs Bacteriological Analytical Manual (BAM) method (19), used in the Appendix.
Experiment. (i) Experimental design. The experimental design used for the
interlaboratory study is described in ISO 16140 (see reference 13, section 6.3.3
and Annex H). The aim of the design is to comparatively determine the performance characteristics (accuracy and precision) of an alternative method against
the corresponding reference method. The design consists of at least eight participating laboratories producing usable results. The first step of the interlaboratory study is to select a single well-mixed representative water sample. The
water sample selected for this study was artificially contaminated with E. coli
strain ESC.1.131, an environmental strain isolated from water. Samples were
contaminated at three nominal levels (expressed as CFU/100 ml): control, 0
(sterile); low, between 1 and 10; medium, between 10 and 50; and high, between
50 and 200. The zero-contamination level was prepared for control purposes only
and was not included in the calculations. Prior to inoculation, the absence of E.
coli and coliform bacteria in the water sample was confirmed by the organizing
laboratory according to NF EN ISO 9308-1 (15).
Each batch of sample was divided into 100-ml aliquots that were transferred to
sterile vials, which were subsequently closed and sealed with tape. Each vial was
individually well mixed prior to shipment to a participant. Each participant
received two samples per contamination level and was required to make duplicate analyses of each sample. A total of 16 enumerations were performed by each
laboratory.
The stability of inoculated samples was determined over a 3-day period using
a prototype sample stored at 4 2C. The results of the stability study are
presented in Table 1. Following log transformation, data were subjected to a
one-way analysis of variance (ANOVA), and stability was found to be satisfactory
3362
BOUBETRA ET AL.
1
2
3
Medium
High
2
11
12
9
8
3
64
67
65
66
66
63
100
110
120
100
100
120
2
kL
s2kr
(1)
(v) For each level, calculate the gross average [ xk] of measurements made with
the alternative method. (vi) For each level, calculate the absolute bias according
to equation 2:
xk Tk (see step 2)
(2)
(3)
(viii) For each level, calculate the differences between the limits of the tolerance
interval and the target value T(k) according to equation 4.
xk kkM skR Tk
(4)
On the horizontal (x) axis, plot the target values [T(k)] in decimal log units
(logs).
On the vertical (y) axis, simultaneously plot the bias (equation 2), the
acceptability limits (), and the tolerance interval limits (equation 4), all
expressed in log10 units.
Laboratory
Reference method
Alternative method
Duplicate 1
Duplicate 2
Duplicate 1
Duplicate 2
Low
A
B
C
D
E
F
G
H
I
J
K
5
19
10
14
14
9
9
4
9
13
15
10
10
8
12
11
6
13
11
12
9
8
9
18
14
10
15
8
12
15
18
9
9
10
14
11
14
11
3
9
7
11
8
12
Medium
A
B
C
D
E
F
G
H
I
J
K
39
72
65
67
52
30
55
33
53
46
45
29
52
64
66
56
50
51
44
54
65
52
53
62
95
48
34
62
56
88
66
59
59
36
53
78
78
38
89
59
50
62
74
48
A
B
C
D
E
F
G
H
I
J
K
105
139
105
146
80
70
131
89
130
130
90
112
142
124
124
78
90
136
90
133
112
74
202
145
201
200
50
130
145
130
118
202
118
130
118
145
200
50
130
130
200
202
130
165
High
a
The study consisted of 3 levels of artificially contaminated samples, 11 laboratories, and duplicate measurements obtained with the reference and alternative methods. Level 0 was excluded from calculations.
3363
All calculations in this study were performed using Microsoft Excel, and
specific worksheets were prepared for this purpose. These worksheets may
be downloaded from http://www.paris.inra.fr/metarisk/downloads/software
_programs/excel_templates). The interpretation of the accuracy profile is as
follows. If across the validation domain of the method, all -expectation tolerance intervals (-ETIs) are included within the acceptability limits, the method
is declared valid over this range. Where any -ETI value exceeds one of the
acceptability limits, the method is not valid and the validity domain must be
reduced.
This can be interpreted as follows. According to the definition, a -ETI should
contain, on average, % of the predicted future results. Therefore, the analyst
and end user can be confident that % of future results will fall between the
limits of this interval. As long as this percentage is included in the acceptability
limits, the analyst can be confident that his measurements are comparable to
those obtained by the reference method, with an acceptance of log10 units.
RESULTS
Reference material preparation. Within the scope of this
validation study, the Colilert-18 method was submitted for
validation according to the ISO 16140 procedure (13). Results
were collected during a collaborative study organized by the
Institut Scientifique dHygie`ne et dAnalyse (ISHA, Massy,
France) and supervised by the AFNOR Certification Board.
Raw data (expressed as numbers of CFU/100 ml) are presented in Table 2.
Since microbial counts are not normally distributed, it was
decided to transform the data into decimal logarithms, as is
typical with such analyses. For each level of contamination,
the reference value, T (or target value), was calculated as the
median result obtained with the reference method. For the
low, medium, and high contamination levels, target values in
CFU/100 ml and their corresponding log10 values (in parentheses) were 12 (1.079), 64 (1.806), and 120 (2.079), respectively. These values are somewhat different from the expected
nominal values, demonstrating that the reference method
and/or reference sample preparation procedure also represents a source of uncertainty. Counts obtained for the alternative method were also transformed into log10 units.
FIG. 1. Linearity check for Escherichia coli data. Medians are indicated as vertical dashed lines and correspond to the target (T) values. In the
equation, Alt and Ref are abbreviations for the alternative and reference method names.
3364
BOUBETRA ET AL.
Parameter
Equation
Target value
No. of participants (I)
Avg of the level (alternative)
Repeatability SD (sr)
Between-lab SD (sL)
Reproducibility SD (sR)
Low
Medium
High
1.000
11
1.024
0.141
0.092
1.716
11
1.771
0.093
0.081
2.049
11
2.142
0.099
0.141
0.168
0.123
0.172
1.367
0.173
1.376
0.127
1.396
0.178
0.794
1.254
1.601
1.941
1.902
2.382
Bias
0.024
0.055
0.093
0.206
0.254
0.115
0.225
0.147
0.333
0.3
0.3
0.3
0.3
0.3
0.3
The accuracy profile method has been applied to the validation of the Colilert-18/Quanti-Tray against the reference
method ISO 9308-1 (15). Using a of 80% and a equal to
0.4, the alternative method can be validated between 1.00 and
2.05 log10 units/100 ml, equivalent to 10 to 112 CFU/100 ml.
The bias which is shown in Fig. 2 may be largely due to the
difference in the principles of enumeration between the refer-
3365
FIG. 2. Accuracy profiles of alternative methods for a equal to 80% and 90% and a equal to 0.3 or 0.4 log10 unit. The arrow indicates
a possible upper limit of quantification (ULOQ). TI, tolerance interval.
18 method, compared to the reference method with an acceptance level of 0.3 log10 units/100 ml, is appropriate for the
application domain ranging from 1.5 up to 2.1 log10 units/100
ml.
Some questions raised by this study remain. (i) The lowest
and highest target values used in the study were, respectively,
approximately 10 and 110 coliform bacteria in 100 ml. When
dealing with distribution water, actual control samples may
contain lower contamination levels below 1 CFU/100 ml. As it
is not possible to guarantee sample stability at such a low level
of contamination, this was not selected for the interlaboratory
study. This may be used as a criticism of collaborative studies
within the field of water microbiology. However, it must be
remembered that collaborative studies based on calibrated
samples are commonly used for proficiency testing schemes in
order to control laboratory trueness. (ii) Data were collected in
the framework of an interlaboratory study, and TIs were calculated by including a laboratory effect; it may explain why they
FIG. 3. Accuracy profile after correction of log-transformed data with a 2% correction factor.
3366
BOUBETRA ET AL.
APPENDIX
Definitions of terms. Definitions of terms used herein are as follows.
(i) is the chosen percentage of values that will be included in the
calculated interval. Notice that has a meaning here different from
that in the test of the hypothesis, where usually means the probability
of accepting a given alternative hypothesis. (ii) A expectation tolerance interval (-ETI) is defined as an interval that covers an average
percentage of a variable distribution. For instance, a -ETI can be
claimed to contain 90% of future measurements, on average. A -ETI
can be expressed as x kM sr, where kM is the coverage factor,
given by the equation
kM Qt v,
1
2
1
I J G2
H1
JH1
2
where H sL2 /sr2 sR2 /sr2 1, SR
is the reproducibility variance, and
2
Sr is the repeatability variance. The number of degrees of freedom, ,
is given by the formula
H 12
1
i
H
1
J
J
I1
IJ
(iii) The acceptability criterion, , is defined as the maximum acceptable difference between a result obtained by the alternative method
and the reference (true) value given by the reference method. Since we
are dealing with logarithms, this difference can be comprehended as a
ratio. For example, the fact that the acceptability criterion () is equal
to 0.3 log10 unit means that results (logarithms) by the alternative
method, log(Nalt), that are inside the interval log(result by the reference method) 0.3 [or log(Nref) 0.3] are acceptable. (iv) The
interval log(Nref) the acceptability criterion is named the acceptability interval. Notice that since 100.3 is equal to 0.501 and 100.3 is
equal to 1.995, an alternative expression for 0.3 could be 0.501
Nalt/Nref 1.995. The acceptability criterion (or the acceptability
interval) must be defined for each microbiological criterion selected.
Most probable number calculation. The MPN is the number that
makes the observed outcome most probable and the solution for
(concentration) in the following equation:
j1
g jm j
1 expmj
t jm j
j1
U
U
t
t
ln
ln
m
tg
vd
tg
where U denotes the amount of sample used as the unit (for example,
100 for the MPN/100 ml), v denotes the volume of each well (ml), and
d denotes the dilution (ml original sample/ml inoculated dilution).
REFERENCES
1. Anonymous. 2002. AFAQ AFNOR certification. Exigences relatives aux
etudes preliminaire et collaborative menees par un laboratoire expert. Internal report, revision 7. AFAQ AFNOR, Saint-Denis, France.
2. Augustin, J. C., and V. Carlier. 2006. Lessons from the organization of a
proficiency testing program in food microbiology by interlaboratory comparison: analytical methods in use, impact of methods on bacterial counts and
measurement uncertainty of bacterial counts. Food Microbiol. 23:1 138.
3. Commission of the European Communities. 2005. Commission regulation
(EC) no. 2073/2005 of 15 November 2005 on microbiological criteria for
foodstuffs. European Commission, Brussels, Belgium. http://eur-lex.europa
.eu/LexUriServ/LexUriServ.do?uriOJ:L:2005:338:0001:0026:EN:PDF.
4. Corry, J. E. L., B. Jarvis, S. Passmore, and A. Hedges. 2007. A critical review
of measurement uncertainty in the enumeration of food micro-organisms.
Food Microbiol. 24:1 230-253.
5. Eckner, K. F. 1998. Comparison of membrane filtration and multiple-tube
fermentation by the Colilert and Enterolert methods for detection of waterborne coliform bacteria, Escherichia coli, and enterococci used in drinking
and bathing water quality monitoring in southern Sweden. Appl. Environ.
Microbiol. 64:8 3079-3083.
6. European Commission, Environment. 1998. Council directive 98/83/EC of 3
November 1998 on the quality of water intended for human consumption.
European Commission, Brussels, Belgium.
7. Feinberg, M. 2007. Validation of analytical methods based on accuracy
profiles. J. Chromatogr. A 1158:174183.
8. Feinberg, M., D. Sohier, and J. F. David. 2009. Validation of an alternative
method for counting Enterobacteriaceae in foods based on accuracy profile.
J. AOAC Int. 92:2 527-537.20
9. Hubert, P. K., et al. 2004. Harmonization of strategies for the validation of
quantitative analytical procedures. A SFSTP proposalpart I. J. Pharm.
Biomed. Anal. 36:579586.
10. Hubert, P., et al. 2007. Harmonization of strategies for the validation of
quantitative analytical procedures. A SFSTP proposalpart II. J. Pharm.
Biomed. Anal. 45:7081.
11. Hubert, P., et al. 2007. Harmonization of strategies for the validation of
3367