Professional Documents
Culture Documents
Page 1 of 81
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
Page 2 of 81
11/17/2009
Page 3 of 81
Table of Contents
1
Introduction
1.1
Thank You
1.2
Getting the Most Out of This Manual
Installation
2.1
Installing More Languages
Setup
4.1
Database
4.1.1
4.1.2
4.2
General
4.2.1
4.2.2
4.2.3
4.2.4
4.2.5
4.2.6
4.3
Incentive
4.6.1
4.6.2
4.6.3
4.6.4
4.6.5
4.6.6
4.6.7
Variables
4.5.1
4.6
HIS
4.4.1
4.4.2
4.4.3
4.4.4
4.4.5
4.4.6
4.4.7
4.4.8
4.5
Ambient factors
Spirometer Detection
Save Options
Units
Language
User Identification
Spirometry
4.3.1
4.3.2
4.3.3
4.3.4
4.3.5
4.4
Location of Data
The Final Report Template
Current Incentive
Show incentive at this PC
Java missing -warning
Targeted FET
Incentive size
Data Communication
Success Criteria
Calibration
5.1
Flow Transducer and Calibration
Calibration Procedure and Acceptance
5.2
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 4 of 81
Measurement
7.1
Measurement Steps
7.2
Getting prepared for the Measurements
7.3
Starting the Measurement
7.4
Stopping the Measurement
7.5
Acceptance of the Measurement
7.6
The Flow and Duration Indicator
7.7
The Reproducibility Indicator
7.8
Session Status Summary
7.9
Deleting an Individual Measurement
7.10
Measurements with the Incentive
Analyzing Results
8.1
Curves
8.1.1
8.1.2
8.1.3
8.1.4
8.1.5
8.2
Numerical Results
8.2.1
8.2.2
8.3
8.4
9
Trend
Printing a Current Window
Interpretation
9.1
Ventilation Function
9.2
Bronchodilatation Test Result
9.3
Manual Interpretation Text
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 5 of 81
Export
12.2.1
12.2.2
12.3
XML Export
GDT Export
Logging
13 Keyboard Shortcuts
14 Drop Down Menus
14.1
In the Opening View
14.1.1
14.1.2
14.1.3
14.2
Database
Device
Help
Session
Measure
Device
Show
Help
15 Maintenance
16 Trouble Shooting
16.1
Leaks
16.2
Spirometer and Computer Connection
16.3
Operating Environment
16.4
Software
17 Certification
17.1
EC Declaration of Conformity
17.2
Certificate of Conformity
18 Accessories
19 Release Information
20 Additional Information
21 Disclaimer
22 Trademarks
23 References
24 APPENDIX: Variables
25 APPENDIX: Reproducibility Indicators
25.1
Introduction
25.2
VC, FVC and FEV1
25.3
FEF75, FEF50 and FEF25 (ERS 1993 only)
25.4
FEF25-75% (ERS 1993 only)
25.5
PEF (ATS/ERS 2005 only)
25.6
MVV (ATS/ERS 2005 only)
Figures
Figure 1-1. Spiro2000, Startup Screen.
Figure 3-1. Starting of the Spiro2000 program.
Figure 3-2. Spirometer not found - message.
Figure 3-3. User Identification screen after starting the Spiro2000 program.
Figure 3-4. The Opening View.
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 6 of 81
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
1
1.1
Page 7 of 81
Introduction
Thank You
We all at Medikro Oy wish to thank you for the purchase of the Medikro spirometer with Spiro2000 software. We trust it will
serve you well. In order to understand better the spirometer system and get most out of it we encourage you to study this
manual in detail. If you already have used a previous version of Spiro2000, you may want only to have a glance at the
additions made in this document (since version 1.7.2). The additions are flagged with green color. Please note also that the
whole text and structure of this document has been revised; these changes have not been flagged.
1.2
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 8 of 81
Installation
The detailed Installation Guide is provided with the Spiro2000 software. It contains a description of all procedures to be done
before starting the spirometry measurements.
2.1
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
3
3.1
Page 9 of 81
After you have started the program you will see the name of the program as well as the version number on the screen while
the program is being loaded. The program will search for the spirometer attached to your computer serial port and identify the
COM port. (The search can be omitted, if you do not use the software for making measurements, see the chapter 4.2.2
Spirometer Detection.)
In case the software scan cannot find the spirometer in any of the COM ports, the following message will be displayed:
This message will appear only once. The software will try to detect the spirometer in the background until it is found. The
personal information can be given and sessions can be studied even before the instrument has been found.
The following Log On screen will appear.
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 10 of 81
Figure 3-3. User Identification screen after starting the Spiro2000 program.
You have to type in at least 2 characters (initials of your name) in the "User name" field and then click Log On. For more
information about the user identification options, see the chapter 4.2.6 User Identification.
After the operator has entered the user name the program will activate the opening view.
The operation of Spiro2000 is easy and fast. All the key commands are push buttons that can be selected by a mouse click.
Alternatively, drop down menus are offered. At the opening view you can continue with buttons which appear active
(highlighted).
3.2
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 11 of 81
When you want to stop using the Spiro2000 software, select the Exit Button (Figure 3-5).
The User Identification screen (Figure 3-3) appears. Now select the Exit button and the program exits to Windows.
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
4
4.1
4.1.1
Page 12 of 81
Setup
Database
Location of Data
The operator can define the location of the data generated within a spirometry session. The location of following data can be
established:
Description of data
Result Database (DSN)
Reference Database (DSN)
Default data transfer folder
Calibration log file
Default value
Medikro_Spirometry_Result
Medikro_Spirometry_Reference (only for information, the location cant be changed)
C:\Medikro\DataTransfer
C:\Medikro\Calibration Log\SpirometryCalibration.xml
The program allows the operator to define databases also in the LAN environment.
4.1.2
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
4.2
4.2.1
Page 13 of 81
General
Ambient factors
The ambient conditions of the operating environment affect the results measured and calculated by the spirometer. Due to this
fact
room temperature,
atmospheric pressure and
relative humidity
should be recorded and entered into the program daily and also if they change notably during the day.
To enter these values (Figure 4-2) use the following command path:
Device
Settings
General
Ambient factors
4.2.2
Spirometer Detection
The default setting of the Spiro2000 program includes an automatic serial port scan for spirometer detection. If you are using
the software only for viewing database results and do not have a spirometer connected to your PC, you may wish the program
to omit the automatic search. To omit the automatic search (Figure 4-2) use the following command path:
Device
Settings
General
Spirometer Detection
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 14 of 81
Inactivate/activate the automatic search by clicking the Spirometer Detection check box.
4.2.3
Save Options
You have three choices to save the session data. The selection (Figure 4-3) is made using the following command path:
Device
Settings
General
Save options
Autosave. Will save all information of the session (personal information, measurement signals and results) to the database
without reminding you at the end of the session.
No Autosave. Will remind you at the end of the session to save all information of the session (personal information,
measurement signals and results) to the database.
No Save. No information of the session will be saved to the database. No reminder will be given.
4.2.4
Units
You may select the units you wish to use from the following two alternatives:
The software uses selected units for body weight and height and ambient temperature. Body height (and sometimes also body
weight) affects reference values. Accordingly, ambient conditions affect the results measured and calculated by spirometry.
To select the units (Figure 4-4) use the following command path:
Device
Settings
General
Units
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 15 of 81
4.2.5
Language
You have an option to select the language that your software is using. The amount of selectable languages depends on how
many languages have been installed on the system. More languages can be installed via separate language packs (see the
chapter 2.1 Installing More Languages).
To select the language (Figure 4-4) use the following command path:
Device
Settings
General
Language
and select the desired language from the list.
4.2.6
User Identification
To identify the user, the software displays a user identification screen, where user gives his/her user name (see Figure 3-3).
This is a default procedure and recommended for most operating environments. Optionally, the user identification can be
automated (disabled) by unselecting the check box. Use the following command path (see Figure 4-4):
Device
Settings
General
Check identification at startup
If the user identification is unchecked, a default user name will be automatically entered to the user identification screen that
is displayed for a short time. Default user name can be selected from the Default user name drop down list and it has the
following choices:
Windows user (the user name that is used when logged into the operating system)
Last user (the user name that has been used previously)
If the user identification is enabled, the selections above are available with an additional choice:
No name
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 16 of 81
If the user identification is enabled, the user has a possibility to modify the default name in the user identification screen.
4.3
4.3.1
Spirometry
Calibration Volume, Max Variation and Logging
Before you run the calibration, you should verify that the calibration volume specified in the program is the same as the
volume of the calibration syringe. You may change also maximum variation of the calibration gains and choose if calibration
is being logged. The settings (Figure 4-5) are found using the following command paths:
Device
Settings
Spirometry
Calibration syringe volume ml
Device
Settings
Spirometry
Max. calibration variation %
Device
Settings
Spirometry
Calibration logging
4.3.2
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 17 of 81
Device
Settings
Spirometry
Linearization file
The file is identified by the product code of the Flow Transducer (for example M9228) and the manufacturing lot number (for
example lot10.cfg). Please note that .cfg is the file extension (file type) of the linearization file. In case the lot number in the
package of the Flow Transducers is not the same as the chosen linearization file, select the right linearization file from the list
box. If the file is not listed, please select "New file" (at the top of the list) and follow the instructions given by the Spiro2000
software. The linearization file and its file name are provided in the Calibration File Diskette within the Flow Transducers
Box.
4.3.3
There are generally only minor differences between the American Thoracic Society (ATS) 1994 [1] and the European
Respiratory Society (ERS) 1993 [2] recommendations. The views of both the ATS and ERS were joined in 2005 [3-5].
The standard selection affects primarily
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 18 of 81
4.3.4.1
FEV1 %ref
x >= -2 SD
70 <= x < -2 SD
60 <= x < 70
50 <= x < 60
35 <= x < 50
x < 35
Sovijarvi 1994. With this choice the ventilation function is interpreted according to Sovijarvi 1994 [6].
The interpretation classifies the per phase spirometry results according to the variable that has the lowest value compared to
the reference value. The exceptions are VC and FVC, of which the classification is accomplished according to the variable
having a higher value compared to the reference value.
Ventilation function
Normal
Mild decrease
Moderate decrease
Severe decrease
Critical decrease
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 19 of 81
4.3.4.4
4.3.4.5
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 20 of 81
4.3.4.7
Reference Set
The reference set represents the predicted pulmonary function values for the patient. To change the reference set selection (14
alternative reference sets in Spiro2000 v. 1.8), use the following command path (Figure 4-6):
Device
Settings
Spirometry
Reference Set
Select the appropriate reference set to change the default value of the set. The set includes the predicted values for both
genders and all age groups. There are several choices for the predicted values of different national or ethnic groups. The
current reference sets in Spiro2000 are:
Name of Ref. Set
Bergl_Z
Brazil
China
Cra_Knu
ECSC_P
ECSC_PZ
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 21 of 81
ECSC_Z
Forche2
Gulsvik_Z
Hedenstrom_Z
Japan
Knudson
NHANES III_W
Vil_Koi
European Coal and Steel Community for adults, Zapletal for children
Austrian reference values: Forche for adults and children
Norwegian reference values: Gulsvik for adults, Zapletal for children
Swedish reference values: Hedenstrom for adults, Zapletal for children
Japanese reference values
Knudson for adults and children
NHANES III (Hankinson) for 8-80 years old. Wang for children below 8 years. *)
Finnish reference values: Viljanen for adults, Koillinen for children
*) The only reference set that takes the ethnic group into account. The ethnic groups mentioned in the references are related
to the ethnic groups used in Spiro2000 in a following way:
Reference
NHANES III
Wang
4.3.5
4.4
4.4.1
HIS
Hospital Information System Interface
Spiro2000 software can be connected to a variety of Hospital Information Systems (HIS). There are two different interfaces:
GDT and XML. GDT (Gerte-Daten-Trger) is tailored for German standard interface for device data. XML is a general
interface and can be applied to almost any HIS.
You can select the Hospital Information System Interface using the following command path:
Device
Settings
HIS
HIS System
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 22 of 81
The options are "None", "GDT" and "XML". Depending on your selection, some of the HIS related options (shown in the
Figure 4-7 and explained below) are not in use.
4.4.2
4.4.3
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 23 of 81
persons already existing in the database. Similarly, incoming sessions' information will replace the sessions which exist in the
database. (Sessions not found in the input document remain untouched in the database.)
You can select the import mode using the following command path (Figure 4-7):
Device
Settings
HIS
Import Mode
4.4.4
4.4.5
4.4.6
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 24 of 81
Device
Settings
HIS
Identification Criteria
4.4.7
4.4.8
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 25 of 81
HIS
Export Session Data
4.5
4.5.1
Variables
Variables for display and print
Variables can be individually selected for display or print (there are altogether 55 variables, see APPENDIX: Variables).
To change the setting (Figure 4-8), use the following command path:
Device
Settings
Variables
Note: By choosing variables for display you can also affect which manoeuvre buttons are active. For example, if none of TV
manoeuvre variables (TV, FR or MV) is chosen, TV manoeuvre button is deactivated (grayed). The same principle is applied
also to VC, FEVC, FIVC, FVC and MVV manoeuvres.
4.6
Incentive
An incentive screen is a motivator for children (and if need be, for any test subject) to exhale with maximal effort for the
spirometry examination. An interactive 3-dimensional animation is displayed during the test and it operates by responding to
the performed exhale. The animation is a Java applet, which is run within the web browser and uses a TCP/IP protocol for
communication. To work with the incentive, there has to be a TCP/IP protocol and a Java virtual machine installed and
configured for use. We recommend that your system administrator install the communication setup.
Setup of the incentive screen has been made into the dialogue sheet Incentive (Figure 4-9). In most cases, it is enough to
check the settings once and start using the incentive without further modifications. All the settings that are related to the
incentive can be found at:
Device
Settings
Incentive
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 26 of 81
At the incentive dialogue sheet (Figure 4-9) the following settings can be adjusted:
4.6.1
Current Incentive
This is a main selector from which the incentive is taken into use. The default value is none, which means that incentive is
not in use. Currently there is one incentive, the Frog. Freddie the Frog is an animated frog that spends its time at its favorite
pond and waits for strong exhalations that will make it leap to a lily pad on the opposite side of the pond. If the exhalation is
not effective enough, Freddie splashes into the water instead.
4.6.2
4.6.3
4.6.4
Targeted FET
A target time of the exhalation (Forced Expiratory Time) in seconds.
4.6.5
Incentive size
Size of the incentive screen in pixels. A default value of 416 x 320 is suitable for most computers since bigger size requires
more performance for the computer.
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Data Communication
4.6.6.1
Page 27 of 81
Port number of the web server. A web server is needed to display and run the incentive at a network. If the port number is
reserved, the number is flagged red and an error message will be displayed when exiting Settings (by selecting OK). In that
case, a new free HTTP server port can be found by clicking the Find a free port button. It is also possible to edit the port
number manually.
4.6.6.2
4.6.6.3
4.6.6.4
Host Name
Host name of the measurement workstation. Can be edited, if Automatic Host Name is unchecked. This may also be the
numeric IP Address of the measurement workstation (e.g. 192.168.64.142). If the incentive is displayed only in a local
measurement workstation, also the values LocalHost or 127.0.0.1 can be used. Please consult your system administrator.
4.6.6.5
The whole address to a net page, where the incentive applet is displayed. In order to view the incentive at external computers,
their browsers have to make a connection to this address (you can copy this address via the clipboard). If the incentive is
selected to open at the measurement computer, the browser will be opened automatically in this address.
4.6.6.6
Incentive buffer
Length of the communication buffer in seconds. A default value of 0.3 s is recommended for most computers and networks.
By increasing this value, it makes the communication more reliable but adds more delay between the exhalation and the
animation.
4.6.7
Success Criteria
4.6.7.1
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 28 of 81
Note that in Current session best and Prev session best, the comparison limit is determined as a percentage of the current
session best or previous session best, respectively. Each selected variable can have an individual percentage.
Also, note that the comparison is always phase related. It means that post phase measurements are always compared to the
post phase measurements and pre phase measurements are always compared to the pre phase measurements.
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 29 of 81
Calibration
Calibration forms an essential part of good laboratory practice.
Medikro spirometers should be calibrated daily as outlined in the ATS and ERS recommendations (see the chapter 23
References). Typically calibration is performed when the spirometer system is switched on. In addition, calibration should be
repeated when starting to use a new lot of Flow Transducers and in case of there is a notable change during the day in
environmental conditions (ambient temperature, pressure or humidity).
Using a calibration log file is recommended and ensures the reliability of your results. You can select the location of this file
as defined in the chapter 4.1.1 Location of Data. Other calibration settings are explained in the chapter 4.3.1 Calibration
Volume, Max Variation and Logging .
5.1
Flow Transducers are manufactured to high precision and it is not necessary to calibrate the spirometer system with each
Flow Transducer separately. It is recommended, however, that you perform a new calibration when starting to use a new lot
of Flow Transducers. The lot number is identified at the label of the Flow Transducer package (Figure 5-2).
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
5.2
Page 30 of 81
Attach the spirometer with the Flow Transducer into the Outlet Port Adapter (Figure 5-4) of your Calibration Syringe.
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 31 of 81
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 32 of 81
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 33 of 81
6.1
Personal Information
All spirometry measurements will start with the identification of the patient. Personal information of a new patient can be
entered into a database via keyboard or previous session information can be retrieved from the database. Database contains
the following information:
6.2
At least the following fields must be filled before the measurement can be performed:
Date of Birth
Personal ID or Patient Code
Gender
Reference set
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 34 of 81
Height
Weight
Date of Birth: Enter the data as instructed - 2 digits for the day, 2 digits for the month and 4 digits for the year. The order of
the data entry is taken from the regional settings of the PC to reflect the local date format.
Age: The program will calculate the age of the patient automatically.
Personal ID: Enter the ID of the patient. In some countries the social security code is used for this purpose.
For example: 1234
Patient Code: Open field for any alphanumeric code used by the establishment / clinic / hospital to identify the patient.
For example: A1B2C3D4
Gender: Select from the following options: Male / Female. No information option is just for indicating, that the selection
has not been done, yet.
Ethnic Group: Indicates the ethnic origin of the person. Use the drop down menu for a proper selection. Ethnic group
information does not change the calculation criteria for lung function, expect when NHANES III_W is chosen for the
persons reference set (see the chapter 4.3.4.7 Reference Set)
Reference Set: Select the appropriate reference set. The set represents the predicted pulmonary function values for the
patient. The selection is valid only during the current session. When the session is closed, the default reference set becomes
valid again. The default value of the reference set can be changed as explained in the chapter 4.3.4.7 Reference Set.
Height: Enter the height in centimeters, or in feet and inches.
For example: 177 (centimeters) or 5 (feet) and 6 (inches).
Weight: Enter the weight in kilograms (kg) or in pounds (lb).
For example: 56 (kg) or 135 (lb.)
BMI: The program will calculate automatically the Body Mass Index. The formula for BMI is as follows:
BMI = weight (kg) / height2 (m2).
Profession: Enter the profession of the patient.
Smoking: Select from the following options: Yes / No / No information.
Illness: The recording of relevant illnesses of the patient.
Medication: The recording of relevant medication of the patient.
Current problem: The recording of any current medical problem.
Comment: Enter any additional observations.
Co-operation: Select from the following options: Good / Satisfactory / Poor / No information.
Operator: Indicates the operator identification entered at Log On phase of the program.
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 35 of 81
Ventilation function: Indicates the status of the ventilation function. The program will calculate this after the measurements
have been completed.
Result of Bronchodilatation Test: Shows the comparative result of the pre- and post phase tests. The program will calculate
this, if post measurements have been accomplished.
Interpretation Text: Interpretation of the session results by consulting doctor. Free text format (Note: only the five first
rows can be printed in a report).
6.3
The selection of the person is made in the list box called "Persons in the database", which includes a list of the names of all
patients in alphabetical order according to family name and then first name. The name of a patient can be searched by
scrolling down the name list or by using an automatic name search function. The name search is straightforward. Enter one or
more first letters of the last name into the Last name field. The list of names displays now the names starting with these
specified letters. The name search is more accurate the more letters are used in the search (Figure 6-4).
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 36 of 81
You can also use person's Date of birth, Personal ID and/or Patient Code as search criteria. The search fields include
Date of birth: Enter one or more first numbers of the date in an yyyymmdd format.
For example: 19550228
Personal ID: Enter one or more first alphanumeric characters of the personal ID.
For example: 1234
Patient code: Enter one or more first alphanumeric characters of the code.
For example: A1B2C3D4
The search result is shown in the Persons in the database field.
After the person has been selected by highlighting the person's name, pressing Choose the Person's Session button will open
a choose session window.
This window will show all previous sessions in time sequence (Figure 6-5).
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 37 of 81
The operator can select either a new session or a possible previous session.
If the operator selects a new session, all personal information from the latest session (if it exists) will be displayed in the
Personal Information window. The operator can now update the information and start a new measurement by pressing the
Meas. button (Figure 6-6) in the Personal Information window.
If the operator selects a previous session, all personal data and measurement results of the session are displayed in the
Personal Information window. The operator can now inspect the previous measurement results and curves by pressing first
the Meas. button (Figure 6-6) in the Personal Information window
and then by selecting the needed result or curve layout by pressing one of the buttons in "Show Curves" or "Show Results"
button groups (Figure 6-7).
For more detailed information, see the chapter 8 Analyzing Results. The Final Report of each session can also be
investigated.
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
7
7.1
Measurement
Measurement Steps
1
2
3
4
5
6
7.2
Page 38 of 81
The patient data is now displayed on the top of the screen in the Title Bar. All functional buttons are in active state. The key
features of the Measurement Window are:
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 39 of 81
Explain the measurement and prepare the patient for the manoeuvre before the actual measurement. The measurement should
also be demonstrated to the patient. Pay attention to correct posture with head elevated, complete inhalation, Flow Transducer
position and complete exhalation. All manoeuvres should be started and finished with resting tidal breathing. The actual
maximal exhalation effort of FEVC and FVC should last at least 6 seconds (3 s for children aged < 10 yrs, ATS/ERS 2005).
MVV manoeuvre should be started only after the patient is breathing maximally and the manoeuvre should last 12 -15
seconds (ATS 1994 & ATS/ERS 2005) or 15 seconds (ERS 1993). It is recommended to allow the patient to perform 1-2
sample manoeuvres before the measurement.
7.3
There are following choices displayed as buttons on the lower left side of the screen:
Button
TV
VC
FEVC
FIVC
FVC
MVV
Manoeuvre
Tidal Volume
Slow Vital Capacity
Forced Expiratory Vital Capacity (only expiratory results will be calculated)
Forced Inspiratory Vital Capacity (only inspiratory results will be calculated)
Forced Vital Capacity (FEVC+FIVC = both expiratory and inspiratory results will be calculated)
Maximum Voluntary Ventilation
If you want to start by only monitoring the signal before any of the manoeuvres, you can accomplish it via a menu option (see
the chapter 14.2.2.1 Start).
Note: by choosing variables for display (see the chapter 4.5.1 Variables for display and print) you can also affect which
manoeuvre buttons are active. E.g. if none of TV manoeuvre variables (TV, FR or MV) is chosen, the TV manoeuvre button
is deactivated (grayed). The same principle is applied also to VC, FEVC, FIVC, FVC and MVV manoeuvres.
The pressing of any manoeuvre buttons starts the measurement. Click the appropriate manoeuvre button when the patient is
ready. The program allows the patient to complete the measurement in 60 seconds. (Well trained patients can complete
several manoeuvre types during the 60 second measurement time. You can start a new manoeuvre immediately after another
without stopping the measurement by clicking the respective button.)
7.4
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 40 of 81
The measurement can be stopped with the Stop button before the 60 seconds are finished. The time of the measurement can
be monitored from the volume/time graph on the screen (Figure 7-5) or from the timer. The measurement will be stopped
automatically after 60 s, if not stopped by the button.
After stopping the measurement, the program will show instantly the FEVC, FIVC or FVC envelope curve, slow VC or MVV
curves as well as the measurement results in the measurement window. This numerical output includes Reference values
(Ref), results of the current measurement (Curr), best results of the session (Best) and the difference between the Current and
Best result (Curr - Best).
7.5
You can accept the measurement or reject it based on the manoeuvre start (extrapolated volume), duration (see Figure 7-7)
and possible manoeuvre artefacts (submaximal effort, cough, glottis closure). The Reproducibility Indicator (see the
chapter 7.7 The Reproducibility Indicator) may help in evaluating the magnitude of the artefacts. For more details of
accepting the manoeuvre see [4].
You can start a new measurement with the same or a new manoeuvre as soon as the patient is ready.
7.6
It indicates the volume change in 1 second (ATS 1994 or ATS/ERS 2005 criteria selected) or in 0.5 seconds (ERS 1993
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 41 of 81
criteria selected). When there is no horizontal bar to indicate a volume change in specified time, the no change criterion to
end or reverse the expiratory / inspiratory manoeuvre is reached.
The time of the last expiration is also shown during FEVC and FVC manoeuvres. The expiration should last at least 6
seconds (3 s for children aged < 10 yrs, ATS/ERS 2005).
7.7
If the criteria are not fulfilled the program will flag the results in red. For more details of reproducibility, see APPENDIX:
Reproducibility Indicators and [4].
7.8
7.9
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 42 of 81
7.10
When the measurement is started, (see 7.3 Starting the Measurement) the incentive applet will set up to a stand-by state and
the frog will be seen from a side view (Figure 7-12). Now it is time to start the study.
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 43 of 81
During the measurement, the frog will react to the ventilation performance by crouching down at the deep inhalation and
making a leap at the consecutive exhalation. Note that in order to get the frog to crouch and leap the inhalation must be deep
enough, at least 50 % of the current sessions pre phase best FVC (if the current session does not have FVC, the previous
sessions pre phase best FVC is used instead. If even that does not exist, the reference value is used). The goal for the jump is
the opposite lily pad, which corresponds a successful FVC (Figure 7-13). If the FVC fails, the frog will splash into the water
(Figure 7-14). Small tabs at the upper part of the screen are called success flags and indicate by color if the criteria of the
variables have been fulfilled.
A tab with a dark yellow background means that the value has not been calculated yet.
A tab with a bright yellow background means that the measured value fulfils the criterion.
A tab with a bright red background means that the measured value does not fulfill the criterion.
Note that the spirometry software makes a prediction for the FVC in the middle part of the exhalation. According to the
prediction, the frog will land in the water on the opposite lily pad. However, it is possible that the frog will splash into the
water even if the FVC succeeds. This can happen if the exhalation gets better at the latter half of exhalation.
A clear indicator of a successful FVC is that the frog will jump up and down and raise its hands after the leap (Figure 7-13).
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 44 of 81
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 45 of 81
Analyzing Results
You can analyze the results as graphs or as numeric results, or as a combination of both. According to your choice, select any
of the buttons VT, VT6, FV or Envel. under Show Curves, or any of Meas., Best or Trend under Show
Results (Figure 8-1).
8.1
Curves
The data of the session is displayed as curves and tables (Figure 8-2, Figure 8-4, Figure 8-5, Figure 8-3).
8.1.1
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 46 of 81
8.1.3
8.1.4
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 47 of 81
8.1.5
Figure 8-6. Draw a marquee around the interesting view area. Applies also to VT6 window.
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 48 of 81
Figure 8-7. After drawing a marquee the curve is magnified for more detailed viewing.
To return to the original graph size select Device / Normalize Graph command or click the VT, FV, Envel. or VT6 button
(Show Curves group).
8.2
Numerical Results
The following commands will represent the measurement results in numerical formats:
8.2.1
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 49 of 81
8.2.2
Best results of the Pre- and the Post- measurements are compared to Reference Values. Also the Post - Pre difference is
shown.
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 50 of 81
The best results are also displayed as a Histogram. It will show the 95 % normal range for each measured variable as a green
bar. The best Pre- result is indicated with a blue circle and the best Post- result with a red square. The histogram results will
give direct indication on the severity of a limited lung function.
8.3
Trend
Trend allows the inspection of the trend of each individual variable in a graphical (Figure 8-10) and numeric table (Figure
8-13) form.
The operator can select one or several variables for the trending inspection from the Choose Variables field (Figure 8-11).
The selected variables and their dimensions are shown on the top left hand side of the trend graph. Several variables can be
selected by pressing SHIFT while making a mouse click.
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 51 of 81
The operator can also select the sessions to be excluded from the trending inspection. The exclusion of the sessions can be
selected from the Exclude Sessions field (Figure 8-11). Several sessions can be selected by pressing SHIFT while making a
mouse click. New results are calculated automatically for immediate trending inspection.
X-axis variable can be selected from the appropriate field (Figure 8-11). There are three choices:
The "% ref" button (Figure 8-12) activates the inspection of results in relation to reference (predicted) values. The results are
expressed as a percentage (%) of reference value, provided the selected variable has a reference value.
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 52 of 81
Figure 8-13. A separate floating window shows trending in a numeric table presentation.
The Table button opens a floating window that contains a numeric table of trending data (Figure 8-13). A drop down
selector Variables in the upper left corner of the window shows the list of variables that the operator has chosen from the
previous window (Figure 8-11). By using this selection, the operator can choose the variable of which trend data is being
studied at a time.
8.4
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
9
9.1
Page 53 of 81
Interpretation
Ventilation Function
Spiro2000 interprets the severity of lung function as described in the chapter 4.3.4.1 Ventilation Function Interpretation. The
result is shown in the Personal Information window (Figure 9-1) and in the first page of Final Report (Figure 10-4).
9.2
9.3
Figure 9-1. Ventilation Function, Bronchodilatation Test Result and Manual Interpretation Text
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 54 of 81
personal information, showing the patient name and ID is selectable (on every page)
smoking history (page 1)
clinical information (page 1)
best results as tables and histogram (pages 1)
reproducibility of the measurements (pages 1-3), selectable for the first page
numerical values of the individual Pre- and Post- measurements, showing the curves is selectable (pages 2-3)
numerical values of the selected trending variables (from page 4 forward)
interpretation (page 1)
envelope curves of the individual Pre- and Post- measurements (page 2)
flow-volume curves of the individual Pre- and Post- measurements (page 3)
On the first page, the user can select two graphs to display by selecting two of the following check boxes (at the upper left
corner of the report page):
best curves
trending graph
VT6 curves
10.1
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 55 of 81
You may choose, which page opens first, when viewing the report, see the chapter 4.1.2 The Final Report Template
Use <<Prev, 1, 2, 3, 4 and Next>> buttons in the upper part of the screen to change the page of the final report. The scroll bar
is available on the right margin of the window to allow scrolling the final report page on the screen.
10.2
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 56 of 81
10.3
10.3.1
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 57 of 81
C:\Program Files\Medikro\Temp\Report\. The Final Report contains a minimum of four pages and consists of the following
files:
M8124-3-multi.htm (page 1)
Best Curves.png (page 1)
Trend.png (page 1)
VT6 Curves.png (page 1)
M8124-5-multi.htm (page 2)
PreEnvelopeCurves.png (page 2)
PostEnvelopeCurves.png (page 2)
M8124-6-multi.htm (page 3)
PreFlowVolumes.png (page 3)
PostFlowVolumes.png (page 3)
All of these files are needed to display the full final report, but you can also choose to select files for certain page(s) of the
final report. Drag and drop the files to an e-mail message as an attachment and you can send the Final Report in an e-mail to
the recipient.
10.3.2
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 58 of 81
11.2
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 59 of 81
12.1
Import
The import consists of three steps: receiving an input file, checking imported person's identity and selecting a session for the
imported person.
12.1.1
12.1.2
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 60 of 81
The following identification errors are possible (because the database must have unique PID and PatientCode values):
Several persons found in the database (the match must be unique)
Another person with the given PID exists in the database (this is possible if the criteria does not include PID)
Another person with the given Patient Code exists in the database (this is possible if the criteria does not include
Patient Code)
The input file has neither Personal ID nor Patient Code for the imported person (every person must have either
Personal ID or Patient Code in the database)
The errors are recognized before saving data to the database and they are shown in the dialog window. If there is an
identification error, the acceptance of the person is not possible without corrections.
After accepted person identification, the input data is saved to a database. If the input file includes several persons, each of
them is checked and the data saved separately.
The XML input file is deleted if all persons were accepted and there were no other errors during import. If the identification
of a person is not accepted, the importing is interrupted and will be started again in the next log on (as explained in the
previous chapter).
12.1.3
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 61 of 81
12.1.4
Handling errors
For a successful import the input file must be formally correct (the XML file is a valid XML document, the GDT file has
right check sums). Additionally XML input file must have proper values for the variables which determine the predicted
values. The variables are shown below:
Height
Weight
Gender
Date of Birth
If errors are found in a XML input file, they are reported and the file is moved to the "..\Medikro\HIS Log\Incorrect inputs"folder. Note: if there is a previous input file with the same name, it is overwritten without a warning.
12.1.5
12.2
12.2.1
Export
XML Export
The HIS export will be accomplished when closing a session (or it can also be omitted if the export option is "No Export").
The export settings (see the chapters 4.4.7 HIS Output files and 4.4.8 How and What to Export) define whether and how the
output file is produced.
12.2.2
GDT Export
GDT export is possible only for a new session of the person last imported (and Spiro2000 software has not been closed after
that). Export is accomplished when displaying a report. The GDT input file is deleted after the successful export.
12.3
Logging
This chapter applies only if the XML interface has been chosen; for GDT there is no logging.
All import and export operations are logged to a file located in "..\Medikro\HIS Log" folder. Information older than one week
will not be preserved. The log file is named "SpiroHISLog.xml" and it will include following information:
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 62 of 81
Operator
Operation type (import or export)
Input or output file name
Success info
Reason for error in unsuccessful operations
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 63 of 81
13 Keyboard Shortcuts
Keyboard Shortcut
Ctrl+Shift+C
Ctrl+I
F5
F6
F7
F8
F9
F10
Ctrl+M
Ctrl+V
Ctrl+6
Ctrl+F
Ctrl+E
Ctrl+A
Ctrl+B
Ctrl+T
Ctrl+R
Ctrl+P
Ctrl+H
Ctrl+X
Enter
Esc
Function in Spiro2000
Start Calibration
Open Personal Information
Start/Finish TV manoeuvre
Start/Finish VC manoeuvre
Start/Finish FEVC manoeuvre
Start/Finish FIVC manoeuvre
Start/Finish FVC manoeuvre
Start/Finish MVV manoeuvre
Start Measurement
Show Volume/time Curve(s)
Show 6s Volume/time Curve(s)
Show Flow/volume Curve(s)
Show Envelope f/v Curve(s)
Show Results of Measurements
Show Representative Results
Show Trend
Show Final Report
Print Current Window
Show Help
Exit
Positive response - Log On, OK, Yes etc.,
if a dialog window is open
1) Negative response - Cancel, Close, Exit, No etc., if
a dialog window is open
2) Stop Measurement, if a measurement is going on
3) Exit, otherwise
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 64 of 81
14.1
Database
Device
Help
14.1.1
Database
14.1.1.1
Personal Information
Opens the Personal Information window.
14.1.1.2
14.1.1.3
Page Setup
Opens the Page Setup window and gives the opportunity to adjust paper settings for printing.
14.1.1.4
Exit
Opens the Log On window, which gives the choice to exit the program or to log on.
14.1.2
Device
14.1.2.1
Calibrate
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 65 of 81
Settings
Allows you to define all Spiro2000 settings as described in the chapter 4 Setup.
14.1.3
Help
14.1.3.1
Show Help
Displays this Spiro2000 User's Manual.
14.1.3.2
About
Displays a system summary window (Figure 14-2). The table inside the window contains information about:
Information of the About window is useful if the user has a problem when using the software. Especially when the
Spiro2000 software user needs to contact Medikro Technical Support to solve a problem, it is highly recommended
that this information be sent to Technical Support.
The whole content of the About window can be saved to a separate report file by selecting Save to file from the upper right
corner of the window. This generates an XML-formatted file containing the window information and also the INI-file
content. The file produced can be then easily attached to e-mail and sent to Medikro Technical Support should assistance be
needed.
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
14.2
Page 66 of 81
Session
Measure
Device
Show
Help
14.2.1
Session
14.2.1.1
Personal Information
Opens the Personal Information window.
14.2.1.2
Delete Signal
Allows you to delete individual measurements as described earlier in the chapter 7.9 Deleting an Individual Measurement.
14.2.1.3
Page Setup
Opens the Page Setup window and gives the opportunity to adjust paper settings for printing.
14.2.1.4
14.2.1.5
Exit
Opens the Log On window, which gives the choice to exit the program or to log on.
14.2.2
Measure
14.2.2.1
Start
Starts the spirometric measurement. Note: With this selection the measurement begins and curves are drawn on the screen,
but no results are calculated until you press some of the manoeuvre buttons. This is intended mainly for the MVV manoeuvre
to monitor the signal before the start of the manoeuvre.
14.2.2.2
Stop
Stops any spirometry measurement manoeuvre earlier than the maximum time of 60 seconds.
14.2.3
Device
14.2.3.1
Calibrate
Allows you to calibrate the spirometer as described in the chapter 5 Calibration.
14.2.3.2
Settings
Allows you to define all Spiro2000 settings as described in the chapter 4 Setup.
14.2.3.3
Normalize Graph
Returns the magnified graph to its original size.
14.2.4
Show
In this menu you will have the following choices:
Volume/time Curve(s)
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 67 of 81
6 s Volume/time Curve(s)
Flow/volume Curve(s)
Envelope flow/volume Curve(s)
Results of the Measurements
Representative (Best) Results
Trend
Final Report
The functions of these choices are the same as when using Show Curves and Show Results buttons (see the chapter 8
Analyzing Results).
14.2.5
Help
14.2.5.1
Show Help
Displays this Spiro2000 User's Manual.
14.2.5.2
About
Displays the system summary window, see the chapter 14.1.3.2 About.
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 68 of 81
15 Maintenance
The Flow Transducer is for single use only. Change a clean Flow Transducer for every patient.
Check the pressure tube and replace it if it is leaking or unclean. It is recommended to change the pressure tube after 300
patients or four times a year.
Calibrate the spirometer carefully with a calibration syringe as described earlier in the chapter 5 Calibration. Perform a new
calibration daily and always when starting to use a new lot of Flow Transducers. In addition, calibration should be repeated in
case there is a notable change during the day in environmental conditions (ambient temperature, pressure or humidity).
The spirometer unit does not normally need any maintenance.
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 69 of 81
16 Trouble Shooting
Problems with spirometry measurements are related to instrument failures and erroneous manoeuvres. Instrument failures are
typically either mechanical or software related.
16.1
Leaks
The Spirometer can sometimes experience air leakage. Leaks may occur if pressure tubing is damaged or if any connection
between the flow transducer, pressure tube and spirometer is not properly fastened. Leaks can be detected by daily calibration
of the spirometer. If the gain value after calibration is not in the acceptable gain range, the software will inform the user with
the following message (Fig. 16-1).
Make sure that the pressure tube is properly fastened into the transducer and into the spirometer unit.
16.2
16.3
Operating Environment
The minimum system requirements should be fulfilled to guarantee that the software functions properly. The minimum
requirements can be found in the Installation Guide.
16.4
Software
Software related problems are system dependent. Memory allocation, processor speed, operating system, other devices and
applications play a significant role. The user is advised either to contact Medikro Oy Technical Support using e-mail (e-mail
address: technicalsupport@medikro.com) or to search the Medikro Oy homepage (www.medikro.com) which also provides
useful information.
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 70 of 81
17 Certification
17.1
EC Declaration of Conformity
17.2
Certificate of Conformity
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 71 of 81
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 72 of 81
18 Accessories
Product Name
Software Upgrade
Product Code
M9832
M9228-100
M9227-10
M9220
M9230
M9424
M91130
Description
Use this product code to order an upgrade from your current
Spiro2000 software version to the latest version
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 73 of 81
19 Release Information
You most likely are interested to know what is new in your Spiro2000 software.
Click the link below and you will see
new features,
enhancements and
bug fixes
for not only this version but for all versions released since the first one.
ReleaseInformation Spiro2000.htm
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 74 of 81
20 Additional Information
For additional information, please contact us via e-mail:
technicalsupport@medikro.com
or visit our Homepage:
www.medikro.com
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 75 of 81
21 Disclaimer
Our policy of continuous product improvement makes all Medikro instruments, accessories, brand names, specifications,
product documentation and model availability subject to change without notice.
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 76 of 81
22 Trademarks
Windows is a trademark of Microsoft Corporation, Redmond, WA, USA
Internet Explorer is a trademark of Microsoft Corporation, Redmond, WA, USA
Netscape Navigator is a trademark of Netscape Communications Corporation, USA
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 77 of 81
23 References
1.
2.
3.
4.
5.
6.
American Thoracic Society, Standardization of Spirometry, 1994 Update, Am J Respir Crit Care Med, 1995, Vol 152,
pp 1107-1136.
Quanjer PH, Tammeling GJ, Cotes JE, Pedersen OF, Peslin R, Yernault JC. Lung volumes and forced ventilatory
flows. Official Statement of the European Respiratory Society, Eur Respir J, 1993, Vol 6, Suppl. 16, pp 5-40.
Miller MR, Crapo R, Hankinson J, et al. General considerations for lung function testing. Eur Respir J, 2005, Vol 26
pp 153161. No. 1 in SERIES ATS/ERS TASK FORCE: STANDARDISATION OF LUNG FUNCTION
TESTING.
Miller MR, Hankinson J, Brusasco V, et al. Standardisation of spirometry. Eur Respir J, 2005, Vol 26, pp 319338.
No. 2 in SERIES ATS/ERS TASK FORCE: STANDARDISATION OF LUNG FUNCTION TESTING.
Pellegrino R, Viegi G, Brusasco V, et al. Interpretative strategies for lung function tests. Eur Respir J, 2005, Vol 26,
pp 948-968. No. 5 in SERIES ATS/ERS TASK FORCE: STANDARDISATION OF LUNG FUNCTION
TESTING.
Sovijarvi A, Uusitalo A, Lansimies E & Vuori I: Kliininen fysiologia. Duodecim. Helsinki 1994.
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 78 of 81
24 APPENDIX: Variables
Abbreviation
Name
Unit
TIME
Time
hh:mm
2
3
4
5
6
7
8
9
TV
FR
MV
VC
FVC
FEV0.5
FEV1
FEV6
Tidal volume
Breathing frequency
Minute ventilation
Vital capacity
Forced vital capacity
Forced expiratory volume at 0.5s
Forced expiratory volume at 1.0s
Forced expiratory volume at 6.0s
L
1/min
L/min
L
L
L
L
L
10
11
12
13
14
15
16
17
FEV0.5%(VC)
FEV0.5%(FVC)
FEV0.5%(FEV6)
FEV1%(VC)
FEV1%(FVC)
FEV1%(FEV6)
PEF
PEF
%
%
%
%
%
%
L/s
L/min
18
FEF25
19
FEF50
20
21
22
23
24
25
26
27
FEF75
FEF25-75%
METT
AEFV
FET
VEXT
LAGE
FIVC
FEV0.5 / VC
FEV0.5 / FVC
FEV0.5 / FEV6
FEV1 / VC
FEV1 / FVC
FEV1 / FEV6
Peak expiratory flow
Peak expiratory flow
Forced expiratory flow 25% VC
[MEF75]
Forced expiratory flow 50% VC
[MEF50]
Forced expiratory flow 75% VC
[MEF25]
Forced expiratory flow 25-75% [MMEF]
Mean transit time
Area of expiratory FV-curve
Forced expiratory time
Extrapolated volume
Lung age
Forced inspiratory vital capacity.
28
29
30
31
32
33
34
35
36
37
38
FIV0.5
FIV1
FIV6
FIV0.5%(VC)
FIV0.5%(FIVC)
FIV0.5%(FIV6)
FIV1%(VC)
FIV1%(FIVC)
FIV1%(FIV6)
PIF
PIF
L
L
L
%
%
%
%
%
%
L/s
L/min
39
FIF25
40
FIF50
41
42
43
44
45
46
FIF75
FIF25-75%
MITT
AIFV
FIT
VIEXT
47
48
49
MVV
MVVFR
MVVT
L/min
1/min
s
50
51
VCDIFF
FVCDIFF
Difference to best VC
Difference to best FVC
L
L
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
L/s
L/s
L/s
L/s
s
L*L/s
s
L
a
L
L/s
L/s
L/s
L/s
s
L*L/s
s
L
11/17/2009
FEV1DIFF
PEFDIFF
PEFDIFF
MVVDIFF
Page 79 of 81
Difference to best FEV1
Difference to best PEF
Difference to best PEF
Difference to best MVV
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
L
L/s
L/min
L/min
11/17/2009
Page 80 of 81
Introduction
The Spiro2000 software highlights a value in red if its reproducibility criteria are not fulfilled. . Reproducibility is determined
for both single measurements and session best values. Note that reproducibility is phase related, which means that pre- and
post- measurements are not compared with each other. For detailed information see the ATS 1993 [1], ERS 1994 [2] and
ATS/ERS 2005 [3-5] recommendations.
25.2
25.3
ATS 1994
ml
%best
200
200
200
-
ATS/ERS 2005
ml
%best
150
150 *)
150 *)
PEF
ERS 1993
%best
ml
10
ATS 1994
ml
%best
-
ATS/ERS 2005
ml
%best
-
(If the ATS 1994 or ATS/ERS 2005 recommendations are selected, the best FEFs are obtained from a forced vital capacity
manoeuvre, which has the largest sum of FVC plus FEV1).
25.4
FVC
ERS 1993
%best
ml
5
ATS 1994
ml
%best
-
ATS/ERS 2005
ml
%best
-
(If the ATS 1994 or ATS/ERS 2005 recommendations are selected, the best FEF25-75% is obtained from a forced vital
capacity manoeuvre, which has the largest sum of FVC plus FEV1).
25.5
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009
Page 81 of 81
Indicating reproducibility for Session Best values: the best value is 1) the largest PEF or 2) PEF obtained from a forced
vital capacity manoeuvre, which has the largest sum of FVC plus FEV1. Whichever the choice, the highest value should not
differ next highest one by more than the criterion. If the criterion is not fulfilled, the difference between the two largest values
is flagged red.
Indicating reproducibility for single measurements: if PEF differs from its largest value by more than its criterion, it is
flagged red. The respective difference PEFDIFF is also flagged red.
PEF
ERS 1993
%best
L/s
-
ATS 1994
L/s
%best
-
ATS/ERS 2005
L/s
%best
0.67
-
25.6
MVV
ERS 1993
%best
L/min
-
ATS 1994
L/min
%best
-
ATS/ERS 2005
L/min
%best
20
file://C:\Medikro\Spiro2000\Language\English\Help\Spiro2000Help.htm
11/17/2009