NOTE: The following exercises should be performed while wearing Smart CapnoLine
Adult non-intubated sampling line and a pulse oximetry sensor. Please perform exercises
for hypoventilation, hyperventilation and apnea while sitting in a chair.

BASELINE ASSESSMENT
1. Turn on the Physio-Control monitor.
2. Plug in the Smart CapnoLine Plus non-intubated oral-nasal sampling line.
Smart CapnoLine
sampling line with O2

The purpose of this exercise is


to gain a better understanding
of the practical application
of the FilterLine sampling
line, the etCO2 value and the
waveform and documentation
with the Physio-Control
LifePak 12 or LifePak 15
monitor/defribrillators. The
exercises will provide you with
hands on experience with
Microstream capnography
and the responsiveness of
the technology with changing
patient conditions.

3. Apply the Smart CapnoLine Plus sampling line to your face.


4. Put the pulse oximeter sensor on your index finger.
a. Monitor your breathing for 60 seconds or until a respiratory rate appears
on the Physio-Control LifePak 12 or LifePak 15 monitor/defibrillator.
b. What is your etCO2 value?___________mmHg
c. What is your RR?_____________
d. What is your SpO2%?_____________%
These etCO2 and SpO2 values are considered your baseline and are important
to capture as the starting point with each patient you are monitoring.
IMPORTANT NOTE: If the Physio-Control LifePak 12 or LifePak 15 monitor/
defibrillator does not have a capnography waveform on Channel 2 or 3, use the
selector knob to choose the capnography waveform.
5. Observe the capnography waveform on the screen and press print to collect
a CO2 waveform strip.

HYPOVENTILATION SIMULATION
1. Decrease respiratory rate in order to increase your baseline etCO2 level
by 5 mmHg.
2. What is your etCO2 value?___________mmHg
3. Observe waveform and capture the event by pressing print.

HYPERVENTILATION SIMULATION
1. Increase your respiratory rate by breathing rapidly and trying to decrease your
baseline etCO2 level by 5 mmHg.
2. What is your etCO2 value?___________mmHg
3. Observe the CO2 waveform and capture the event by pressing print.

APNEA SIMULATION
1. With the SpO2 probe on finger, hold breath to trigger CO2 apnea alarm (30 seconds).
2. Observe CO2 waveform, etCO2 value and SpO2 value. Capture the event by pressing
print.
3. Observation:___________________________________________________________

SAMPLING LINE OCCLUSION SIMULATION


1. Kink FilterLine sampling line until Physio-Control LifePak 12 or LifePak 15 monitor/
defibrillator indicates that the CO2 module is performing a purge. To attempt to
clear an occlusion, these monitors purge whenever an occlusion in the line or airway
adaptor is detected, and will display PURGING while this function is performed. If the
CO2 module cannot clear the occlusion within 30 seconds, the monitor/defibrillator
will issue an equipment alert and display the message: FILTERLINE OCCLUSION.
2. Observe message, then remove the kink from FilterLine sampling line.
3. Do you observe a CO2 waveform and etCO2 value? If not, disconnect the FilterLine
sampling line and reconnect to reactivate the CO2 pump.
4. If reconnecting fails to produce a CO2 waveform and etCO2 value, replace the
FilterLine sampling line.
5. Observation:___________________________________________________________

2016 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are
trademarks of Medtronic. * and * Third party brands are trademarks of their respective
owners. All other brands are trademarks of a Medtronic company. 15-PM-0180(1)

6135 Gunbarrel Avenue


Boulder, CO 80301

800.635.5267

medtronic.com

MEDTRONIC
MEDICAL AFFAIRS
covidien.com/PACE

You might also like