CONTENTS
Chapter 1. Introduction ..........................................................................................................................................1
1-1. Configurations ............................................................................................................................................1
1-2. Software Requirements..............................................................................................................................1
1-3. Hardware Requirements ............................................................................................................................2
1-4. Autochro Data Module................................................................................................................................2
1-4-1. Functions .........................................................................................................................................2
1-4-2.Specifications....................................................................................................................................3
Chapter 2. Installation ............................................................................................................................................4
2-1. Installing the software ................................................................................................................................4
Chapter 3. Overview .............................................................................................................................................10
3-1. Features of Autochro-3000.......................................................................................................................10
3-1-1. Control of System ..........................................................................................................................10
3-1-2. Acquisition and display of data ......................................................................................................10
3-1-3. Analysis of data and report ............................................................................................................11
3-1-4. The others......................................................................................................................................12
3-2. Before use Autochro-3000 .......................................................................................................................12
3-2-1. Data File.........................................................................................................................................12
3-2-2. Result Table ...................................................................................................................................13
3-3. User Policy ...............................................................................................................................................16
3-3-1. Type of Task...................................................................................................................................16
3-3-2. Type of User and rights for data processing ..................................................................................16
3-3-3. User Login and Logout ..................................................................................................................17
3-3-4. User Registration ...........................................................................................................................18
3-3-5. Setup for user information and password......................................................................................18
Chapter 4. Structure of Autochro-3000 ..............................................................................................................21
4-1. Autochro-3000 Main Program ..................................................................................................................21
4-1-1. Analysis Window............................................................................................................................22
4-1-2. Chromatogram Overlay .................................................................................................................24
4-1-3. Statistics View ................................................................................................................................25
4-1-4. System Window .............................................................................................................................26
4-2. Young Lin DataDaemon Sub-Program.....................................................................................................27
- i -
- ii -
APPENDIX
1. How to Use Autochro Data Module (ADM) .....................................................................................................81
1-1. How to connect ADM................................................................................................................................81
1-2. How to add ADM ......................................................................................................................................82
1-3. How to set up ADM ..................................................................................................................................82
2. How to use Acme 9000 HPLC..........................................................................................................................83
2-1. How to use UV730D.................................................................................................................................83
2-1-1. How to connect UV730D ...............................................................................................................83
2-1-2. How to add UV730D ......................................................................................................................83
2-1-3. How to set up UV730D ..................................................................................................................84
2-2. How to use SP930D.................................................................................................................................86
2-2-1. How to connect SP930D ...............................................................................................................87
2-2-1-1. Connect to PC(Direct control) ....................................................................................................87
2-2-1-2. Connect to UV730D(Control through UV730D) .........................................................................87
2-2-2. How to add SP930D ......................................................................................................................88
2-2-3. How to set up SP930D ..................................................................................................................88
2-3. How to use CTS30 ...................................................................................................................................92
2-3-1. How to connect CTS30..................................................................................................................92
2-3-2. How to add CTS30 ........................................................................................................................92
2-3-3. How to control CTS30....................................................................................................................93
2-4. How to use RI750F (Refractive Index Detector) ......................................................................................95
2-4-1. How to connect RI750F .................................................................................................................95
2-4-2. How to add RI750F........................................................................................................................95
2-4-3. How to set up RI750F ....................................................................................................................95
2-5. How to use MIDAS Autosampler..............................................................................................................96
- iii -
- iv -
- v -
Chapter 1. Introduction
Young Lin Autochro-3000 Chromatograph data system is the software that is designed to control various type
of chromatograph system for data acquisition, analysis, operation and control. It provides independent space to
manage data (Analysis) and control instruments (System) and is fully compatible with MS Windows XP for
optimized analysis condition.
1-1. Configurations
Autochro-3000 data system comes with Autochro Data Module for analog and digital data conversion.
Table 1-1 Product configuration
Young Lin Autochro-3000 Software
- 1 -
Specifications
CPU
RAM
Graphic capability
4 GB or above
COM Ports
Printer
Monitor
- 2 -
1-4-2.Specifications
Table1-3. Specifications of Autochro Data Module (ADM)
Specification
Resolution
20
Dynamic range
1000000 or above
Input range
-1.25 ~ +1.25
No of channel
Sampling rate
Events
2 Channel Operations
Baud rate
Power
220V/60Hz or 110V/60Hz
Dimension
21011050
Unit
Bit
VDC
Channel
- 3 -
Hz
BPS
mm
Chapter 2. Installation
2-1. Installing the software
1. Insert the Autochro-3000 CD into the CD-ROM driver, then, the screen shows up as below.
Click
- 4 -
Ref.) If Window doesnt automatically run CD-ROM, you may not able to see the initial screen. Run
PreInstall.exe in the folder through Windows Explorer to run Autochro-3000.
- 5 -
3. Read the License Agreement and select I Agree. Then, click on Next
- 6 -
5. Enter the user information and serial No. on the back of CD case. Then click on Next.
When entering serial No, distinguish between upper- and lowercase letters.
The below message shows up when the serial no. is not correct.
- 7 -
7. The installation is preceded as below. If you want to cancel the installation, click on Cancel.
- 8 -
Ref.) The shortcut of Autochro3000 will be created on the wallpaper as well as start-menu.
- 9 -
Chapter 3. Overview
3-1. Features of Autochro-3000
Autochro-3000 is the software that is 32 bit and fully compatible with fully compatible with MS Windows XP. It
provides independent module to manage data (Analysis) and control instruments (System) for optimized analysis
condition.
- 10 -
Analysis report, Calibration report , Statistics report and Audit trail report
Compatible with other program since data transform to Industrial standard ASCII CDF automatically.
Description
List of raw files and the information to analyze them.
File where raw data is saved
Acquisition File
- 12 -
Click a right button of the mouse on the crossed area between each column.
Select [Add Column]-Choose one what you want to add on a pop-up menu.
Delete columns
Click a right button of the mouse on the column that you want to delete.
Insert Columns
Click a right button of the mouse on the column where youre going to insert.
Select [Add Column]-Choose one what you want to insert on a pop-up menu.
Modification is allowed
- 14 -
- 15 -
Read
Overlay
Statistic
Editing
Changing information of raw data such as name and kinds of sample or Scale factor
Analysis
Printing
Conversion
Signature
Read
Overlay
Statistic
Editing
Analysis
Printing
Conversion
Signature
Guest
Administrator
General User
Prohibited User
Unknown User
- 17 -
User Log-out
1) Go to the menu [Analysis]-[Log Out].
2) Guest is shown on the status bar.
- 18 -
- 19 -
- 20 -
Description
Autochro-3000.exe
DataDaemon.exe
The sub-program that controls user information, raw data and records
CtrlDaemon.exe
Description
Analysis window
Chromatogram view
Overlay view
Statistics view
System window
Acquisition view
- 21 -
Workspace
Double-click or right-click and select [Information] the below items of selected CRM file. A dialog box pops up.
Integration: Set up integration parameters and events.
Calibration: Check for peaks and draw calibration curves.
Analysis report: Set up report information such as printing setup and contents.
Sample Table
Sample table edit : The owner of raw data can edit columns below
Sample name
Sample type: Standard or Unknown
Owner: The user who can edit and set analysis of raw data
Inject(ul): Injected amount of sample for the analysis
Vial No.
Scale: Correlation constant to compensate concentration after analysis
Comments: Remark or description note
- 22 -
- 23 -
Overlay view
Title bar: Click on Print Option to set up print setting.
Multiple Chromatogram
Overlay information
Title : Title of the overlaid chromatogram when it comes to printing
Table: Set up color of chromatogram and space where its printed out
Comments : Remark note on print-out
How to Overlay
- On the analysis window, select Name of sample want to compare.
- Click right mouse button and then select To overlay view -> New.
- On the left window, select Overlay View
- 24 -
Statistics View
Title bar : Click on Print Option to set up print setting.
Statistic table tab
Statistic table 1: Compare the analysis results by samples vs components
Statistic table 2: Compare and take statistics from analysis results by specific components
Statistics Information
Title of Statistics: Title of page when it comes to printing
Table
Comments : Remark note on print-out
- 25 -
Workspace
List of devices: Add or delete a device and control of devices
Control of devices: Double-click on the devices or right-click on the devices - [Setup Device]. A dialog box
pops up.
Run table
Status of devices : Status of devices which are running is shown in real time.
Add a sample youre going to analyze
Edit run table: Editing below items is available to only a sample in ready
Sample name
Control : Control file you want to use. If not selected, present analysis condition will be applied.
CRM : CRM file that will include the raw data. Before you inject sample, you must select one CRM file.
Type : Type of sample ( Unknown or Standard)
Vial No. : It is used for autosampler.
Inj. Count : No. of injection per each sample
Inject(ul): Amount of injected sample
Post Run : Auto integration, Quantification and Report printing out after data acquisition
Acquisition view
It shows the chromatogram of raw data on acquisition in real time. You can view dual channel at same time
and selection for each channel is available by acquisition channel box. There is a bar in the middle for scaling
of graphs.
- 26 -
All data are automatically saved at every set-time. DataDaemon is designed to manage data in safe and efficient
in case of unexpected system error. Even though the main program is closed, this sub-program keeps running.
To close it, click on the tray icon-[Exit].
This program controls all process of analysis except of devices set-up. It performs actual tasks that Autochro3000 (main program) needs. Even though the main program is closed, this sub-program keeps running for the
safest analysis condition.
This CtrlDaemon sub-program was combined with DataDaemon sub-program at present version.
- 27 -
is added in task bar when you first starts the Autochro-3000. These sub-programs are
not terminated with the main program, even though you close the Autochro-3000 main program.
ACME_HPLC.CRM
ACME_HPLC
HPLC0001.RAW
HPLC0002.RAW
HPLC0003.RAW
HPLC0004.RAW
Select location/directory for new Analysis and create a folder, type the CRM file name and then click an
open button in new Analysis dialog box.
Even though you select an existing Analysis, new Analysis does not overwrite an existing Analysis.
Open CRM
1) Select [Analysis]-[Open CRM] of Menu or click
2) Click open button after selecting CRM file on the folder that you will open in an Open CRM dialogue box.
- 29 -
Close CRM
Select [Analysis]-[Close CRM] of Menu after selecting the CRM file name that you will close.
Extension Name
RAW
CDF
RAW
- 30 -
remark
- 31 -
(2) Drag it and drop to Sample Table of workspace using right mouse button
(3) Select [ Copy here] in Pop-up Menu
Movement of Raw data
Autochro-3000 data system does not allow user to move location of raw data file of operating system level in
order to maintain the data validity. The raw data is copied if you are not the user of raw data.
Movement by using Menu
(1) Select the raw data to move in Sample Table
(2) Select [Sample Table]-[Cut] of Menu
(3) Select the folder of Sample Table to paste
(4) Select [Sample Table]-[Paste] of Menu
Movement by using a mouse
(1) Select the raw data to move in Sample Table
(2) Drag it and drop to target sample by using a left mouse button
Export of Raw data
Autochro-3000 data system supports ANDI format file to use raw data with another program. The raw data
can export to CSV file that MS Excel can read, as well as, Analysis report can export to MS word. Raw data
file can be changed into original file name [channel collected] detector.
Table 5-2. Raw data Export (C:\A\HPLC0001.RAW )
Export file
Export method
ANDI
CSV
MS WORD
- 32 -
Double click integration item under CRM file name of Analysis window in a workspace.
Edit integration information dialogue box
- 33 -
button
of toolbar.
3) Select [Sample Table]-[Integration] or click
of toolbar.
Integration of User
1) Put a conversion factor and integral factor to integral information form
2) Move a mouse to start time of integral event set up on a chromatogram graph
3) Drag the mouse by stop time pressing a left mouse button with SHIFT key
User need not press the SHIFT key in case the graph is an Event mode
4) Select OK after selecting integral event and putting value in an integral event dialogue box
5) Press the button Now to integrate using changed information in integral information form if necessary.
6) Change the collected channel in a collected channel box and repeat the process from 1) to 5).
Applying information of user integration to sample table
Relevant raw data and other raw data in sample table can be integrated using same integral information if
integral information is applied to sample table after user integration.
1) Click Apply to CRM button in Sample Table of integral information form.
2) Choose one between [Apply to default] or [Apply to this channel].
Apply to default : Applying to integral information default of sample table.
Apply to this channel : Applying to integral information of relevant collected channel of sample table.
- 34 -
button in toolbar.
- 35 -
- 36 -
- 37 -
Sample
Name of Variable
Comment
$Injection
Volume of injection
$Scale
Correction factor
$Sum[Area]
$Sum[Area], Calc
$Sum[Area], Uncal
$Sum[Height]
$Sum[Height] , Calc
$Sum[Height], Uncal
$Sum[Amount]
Peak
$RT
- Sample Peak
$Area
-Reference Peak
$Area%
- ISTD Peak
Retention time
Ratio of Area
$Height
$Height%
Ratio of Height
$Amount
$Amount%
Ratio of amount
$Width
$TP
$Resolution
Constant
$TF
Trailing Factor
$Const 1
$Const 2
In case of Peak Variables , a type of each peak is set up prior to variables. (If you didnt setup, sample
peak is assigned automatically.)
Ex) Area of Sample Peak : $Sample Peak. Area
Amount of Reference Peak : $Reference Peak. Amount
Table 5-4. Function of user defined form
Function
Comment
ABS(X)
Absolute value
EXP(X)
Exponential Function
LOG(X)
LN(X)
POW(X1,X2)
X1^X2
MIN(X1,X2)
MAX(X1,X2)
button in a toolbar.
- 39 -
Comment
Title of report
Analyst
Analysis
Acquisition
Chromatogram
Result
Logo
Comments
Explanation
Audit Trail
- 40 -
Header setup : Setup a form and content of header that will print out top of page.
Range, margin : Printing section of header of inner range except margin of page setup.
Print Line : Indication of boundary line between header and body.
Details : Details that you print out in header ( Page No, the number of all pages, Date, time, User ID
and so on.)
- 41 -
Body setup : Setup a form of body that is between header or footer of page.
Body margin : Setup margin of body.
Sub-title : Setup font that is used in title of contents.
Section : Inner margin of contents and tap distance.
Footer setup : Setup form and details of footer that is printed bottom side of page.
Print line : Whether boundary between body and footer is indicated or not.
- 42 -
- 43 -
Audit Trail : Print out history of raw data until before report is printed.
Acquisition : History of collection start, collection success/fail during collection, communication error
and instrument error.
Analysis : History of integration and weight.
Editing : History of sample information that user change.
- 44 -
- 45 -
- 46 -
Print Preview
(1) Choose raw data to print out in sample table window.
(2) Click [Analysis]- [Print preview] in a menu.
Print
(1) Choose raw data to print out in sample table window.
(2) Click [Analysis]- [Print] in a menu or
button in a toolbox.
- 47 -
Load Method
1) Choose analysis to apply in Analysis Window.
2) Click [Analysis]-[Load Method] in a menu.
3) Find a (.MTD) file in Load Method dialog box and click open button.
Transfer of chromatogram
1) Choose [Pan] mode of title line.
2) Click mouse left button and drag.
Setup of integration event.
Choose [Event] mode of title line and drag your mouse.
Drag a mouse while SHIFT key is pressed.
in a popup menu.
Additional comparison
It is added chromatogram to compare with a existing Overlay View.
1) Click sample that you want to compare additionally in sample table window.
2) Click [Sample table]-[To overlay View] [Append] in menu.
TIP) If you drag samples want to compare using mouse right button to Overlay View item of workspace, you
- 49 -
Choose several chromatogram in Y offset[mV] column, and click [Set Y offset..] in a popup menu of a
mouse right button.
Horizontal direction
Input directly X offset[min] column of sample table in Overlay View window.
Choose several chromatogram in X offset [min] column, and click [Set X offset..] in a popup menu of
mouse right button.
Color change
Input directly Color column of sample table in Overlay View window.
Choose several chromatogram in Color column, and click [Set Color Gradient..] in a popup menu of
mouse right button.
- 50 -
- 51 -
Print
Print preview : Click in the menu [Analysis]-[Print Preview]
Print : Click in the menu [Analysis]-[Print]
add sample.
Deletion of analytical result
1) Choose analytical result that you want to delete in sample table of [Overlay View] form.
2) Click [Delete] in a popup menu of mouse right button.
- 52 -
- 53 -
Start of analysis
After sample injection, start of collection
Stop of collection, stop of sequence of control
Analysis of next sample
Stop of analysis
Forced end
button in a toolbar.
2) Set a location and a name of System that you want to make in a dialog box and click Open button.
- 54 -
button in a toolbar.
System closing
1) Choose System in a workspace.
2) Click [System]-[Close System] in a menu.
If you end a Autochro-3000 program, System file that is different from Analysis file is remembered by
CtrlDaemon supporting program. When you start again Autochro-3000, former opened System file is added
in a workspace automatically.
- 55 -
ADM0[A]*
Model
Reference
Device No.
Link ID or Route
We recommend that after all device is added to System file, in advance System file close and
reopen once because in some case immediate use is not available.
Deletion of device
1) Click device that you want to delete in a workspace.
2) Click [System]- [Del Device] in a menu.
Change of device information
Change model type and information related with communication.
1) Click device that you want to change in a workspace.
2) Click [System]-[Information] in a menu.
- 56 -
Setup of device.
You can setup device by using setup dialog box. And you can setup a lot of device simultaneously by opening
many dialog box.
1) Click device that you want to setup.
2) Click [System]- [ Setup Device] in a menu or mouse double click.
Trace Graph
By using this ability, you can monitor real time working condition of device.
Open of trace graph
(1) Click device that you want to open in workspace.
(2) Click [Trace Graph] in popup menu of mouse right button.
Information of trace graph
Selection of information : Above a trace graph, click in a popup menu of mouse right button.
Section of display : It is displayed device state during last 10 minute.
Color of line : Color is changed depend on device state.
Not connected section : Gray
Run section : lime-green
Communication failure section : Red
Normal section except for run section : white
- 57 -
the index is increased in each time. If you want to change index, click check box.
File Prefix
File name of File cant be assigned at users pleasure in Autochro-3000. Each time program collect data,
additional file which index is increased is checked whether exist or not. After that, this file is automatically
assigned vacant file name to prevent collected raw data from changing or losing.
When error occurred , Run
If error is reported in a device of [System], Autochro 3000 will stop operating of that device. If you want to use
another system setup when error is reported, you could assign control setup file in this route.
Background Chromatogram
When you want to compare collecting data with already existing data, you can setup background
chromatogram that you want to compare.
- 59 -
- 60 -
Set location and name of Run Table file and click save button.
Save Run Table
Click [Run Table]-[Save SRS] in menu.
Set location and name of Run Table file and click save button.
Setup of analysis condition
You can setup analysis condition of a sample that you plan to analyze. Other sample should be renewed and
then can setup. You can setup below list.
Name
CRM
This ability assign analysis file which analyzed raw data will be added in. If that is not assigned, recent
- 61 -
Analysis file is assigned automatically. If renewal is done, this element isnt renewed. To initialize, click CRM
column that you want to initialize and click [Clear CRM] in a popup menu of mouse right button.
Control
This ability setup a control way to analyze sample. If you dont setup, control way isnt changed.
Likewise CRM if you initialize sample, this element is not initialized. To initialize, click Control column that
you want to initialize and click [Clear CTL] in a popup menu of mouse right button.
Type : Click between [Standard] sample and [Unknown] sample.
Vial no.
It can setup vial number that is used by autosampler. If you assign invalid vial number, analysis will stop in
that sample. After choosing several samples in vial number column, if you choose [Set Vial no] in popup
menu of mouse right button, you can easily setup a lot of vial number.
Inj .count
This ability can setup the number of injection. If the number of injection is larger than 1, when analysis is
started, automatically same sample column is added and the number of injection is decreased..
Inject [l]
This ability can setup sample amount that is used in analysis. Autosampler inject sample depends on this
value. If this value is 0, analysis is operated by default setup of autosampler.
Post Run : Choose among integration, weight, and report.
Comments : Simple explanation of raw data
Ready / Run / Stop of analysis
Ready
After finish of registration of sample and setup of condition, or when previous analysis is ended abnormally,
before run of analysis you should initialize appropriately depends on next analysis.
Click [Run Table]-[Ready] or
Run : Click [Run Table]-[Run] or
Stop : Click [Run Table]-[Stop] or
- 62 -
- 63 -
- 64 -
- 65 -
You can see or print used history of run item by clicking [Sample Table]- [Audit trail] in a menu.
- 66 -
Seeing signature
All working that change data is prohibited in a already signed raw data except for changing ID of owner.
And when you sign , tag that monitor changing data by using illegal way is inserted at doing signature to check
file. If you want to see content of signature, click [Signature]-[Verify].
- 67 -
File Check
If file is not changed, OK is displayed.
If file is changed, Records Changed is displayed.
- 68 -
When you input Peak Width in Set integration parameters dialog box, software calculates the data-bunching
factor using the following equation:
WxS
B = --------15
Where:
B = Bunching value(integer, 1 to 10)
W = Measured width(in seconds) of the narrowest peak you want to detect
S = Sampling rate (points/second)
2) Determining Peak Start
The following summarizes the process for determining peak start:
1. The detection algorithm searches for the start of a peak by performing a threshold test on the signal.
2. The detection algorithm compares the signal rise to the user-specified Threshold parameter to locate the
peak liftoff point. The Threshold parameter defines the rate of change of the signal.
3. Signal slope is averaged across two data bunch intervals and compared to the threshold to determine the
peak start point. A possible peak start point is determined when the averaged slope of the signal between
bunches B1 and B3 is greater than or equal to the threshold value. When this requirement is met, the
detection algorithm flags B1 as the possible peak start.
4. Examines the individual points in the B1 bunch to determine the actual liftoff point.
- 70 -
- 71 -
1. Searches for a peak apex by comparing the slope of the signal to the user-specified Threshold parameter
(or Set Touchdown event). When two consecutive slopes are less than the Threshold value, the algorithm
flags the last data point in the last bunch as the possible peak end.
2. Examines the individual data points in the next bunch to find the actual peak end.
3. Checks for a change in sign during testing for peak end. A Change in sign of the slope before the
touchdown is determined indicates a peak valley.
4. From this valley point, the peak detection algorithm proceeds to the peak apex test and continues until peak
touchdown is successfully determined.
- 72 -
- 73 -
The total peak area is calculated by adding the areas for each RAW data point interval between peak start
and peak end. The region below the constructed baseline is subtracted from the total area of the peak. This
yields the peak area above the constructed baseline (Ap).
- 74 -
Forced Peak
Controls peak integration based on the start and end time window you specify.
Set Threshold
Redefines the threshold value to which the signal slope is compared in order to determine peak in a specific
range.
Set minimum peak width
Redefines the minimum peak width within the start and end times of the event.
Set Minimum Peak Area
Redefines the minimum peak area within the start and end times of the event. Enter start and end times to
change and do not enter in the Value.
- 75 -
- 76 -
- 77 -
- 78 -
Shoulder peak
Detects shoulder peaks in a specific range. Enter times for the range and do not enter in the Value
- 79 -
Tangential Peak
Resolves rider or shoulder peaks from the front or rear of a parent peak within the start and end times of the
event. Enter start and end times and a value for tangential skim in the Value.
Group Peak
Determine all peaks in a peak within the start and end times of the event. Enter start and end times to change
and do not enter in the Value.
- 80 -
- 81 -
2. Enter the number of acquisition channel where the acquired data is saved. Entering 0(zero) will make
no data acquisition for the appropriate channel
- 82 -
3.
- 83 -
- 84 -
Time Events
- 85 -
4. Clink on Apply or OK to set all parameter of detector. You can cancel the setup by clicking on Cancel.
- 86 -
- 87 -
- 88 -
Const. Pressure
The solvents are delivered in constant pressure and mixing ratio.
Upper Limit : Upper flow limit for normal performance. Standard of upper flow limit error.
Lower Limit : Lower flow limit for normal performance. Standard of lower flow limit error.
Flow, Mixing ratio, Event : Setting up for a pump operation.
- 89 -
Gradient
SP930D can be set depending on time.
Enter the time on the end of lines and press ENTER key. New line is added.
You can add a column such as Switch 1, Switch 2, Switch 3, buzzer, marker, upper pressure limit and
lower pressure limit by clicking right button on the title bar.
- 90 -
Drag a vertical bar on the right to see a graph for temperature programming. Click a right button on the
mouse to change line color.
4. Drag a vertical bar on the right to see a graph for temperature programming. Click a right button on the mouse
to change line color.
- 91 -
- 92 -
Gradient
Young Lin column oven can be controlled depending on time
- 93 -
Enter the time on the end of lines and press ENTER key. New line is added.
You can add a column such as buzzer, marker, upper temperature limit and door open limit by clicking
right button on the title bar.
Drag a vertical bar on the right to see a graph for temperature programming. Click a right button on the
mouse to change line color.
4. Clink on Apply or OK to set all parameter of column oven. You can cancel the setup by clicking on Cancel.
- 94 -
- 95 -
Connect COM port of PC to RS232 IN port(Communication port) on the rear side of the instrument by a
RS232 cable that comes with the instrument.
(*) RS232 cable for use with MIDAS is a general one not a crossed one like a null modem cable.
2. Connect P1 port of the instrument to both Start and DGND terminal of SP930D pump by an Injector marker
cable.
- 96 -
- 97 -
- 100 -
- 101 -
- 103 -
3-3-1. Oven
- Click Oven icon
- Oven control window is displayed
- According to the analysis condition, select isothermal or oven program mode on the mode selection.
- 104 -
- 105 -
- 106 -
- 107 -
- Select Makeup gas and input the Reference gas/Makeup gas flow.
- After the reference and makeup gas are increased to setup value, input the temperature.
- After the detector temperature is increased to setup value, turn on the Filament and then input the TCD
sensitivity value on the TCD SEN.
- If you want to change the polarity, select Polarity Change and change Initial Polarity.
- 108 -
3-3-8. Signal
- On this window, you can set the sensitivity and signal zero value.
- If you want to change signal from Det 1 to Det 2 or vice versa on the same channel, select Signal Change.
After that, select Initial Status and set the signal change program.
- 109 -
- Using mouse, drag the sample table and then the color of table is changed blue.
- Click the Control column and then Find control window is displayed.
- If you select one control file, the control file name is displayed on the Control column.
- 111 -
- 112 -
3-3-10. Autosampler(Headspace)
- If you want to use the Headspace autosampler, turn on the autosampler first and then turn on GC.
- If the autosampler is connected with GC, the main control window is diplayed as below.
- 114 -
- 115 -
(*) The LAN Hub used for communication should be switching type.
- 116 -
Pump condition
On the condition part, it displays flow rate, pressure, and solvent composition at present.
If you click right mouse button on this window, you can select the condition to display in the device
monitor window.
- 117 -
Prime Start/Stop
It is used for filling the solvent into the solvent tubing or remove the air bubbles.
Diagnosis
If you click this button, the system checks the voltage, operation status, sensor status to confirm that
the system operation is normal. At first, select the item wan to test and then click start button for
diagnosis.
Reset
If the error is happened, this button is activated and reset the system using this button.
Pump configuration setup
Config
Setup the pump configuration.
- 118 -
Pump Type
Select the type of pump. You must select proper pump type that is connected with software.
Head volume
Displays the pump head volume. This value can not be changed by the operator.
Max Flow
Set the maximum flow rate for operation. This value is used for limitation of pump flow.
Max Pressure
Set the maximum pressure for operation. This value is used for limitation of pressure.
Conversion Frequency
Set the maximum frequency to operate the motor.
Valve Calibration Factor
Correct the Valve control value, set the correction value for precision of gradient profile.
Compressibility Factor 1, 2
Set the correction factor for precision and accuracy of flow that affected by the compressibility of
solvent and coefficient of elasticity of Seal.
Pressure Zero
Set the zero point of pressure data.
- 119 -
(*) The pump configuration values are related with performance of pump. Before change the values, have
to understand each function. It is recommended to use the pump with default values.
Details
Set the External input/output signal, and also initial value of switch and marker.
Pulse width
If you select the pulse as a external output signal, set the duration time of output signal.
Isocratic Mode
Deliver the solvent under the constant flow and solvent ratio during analysis.
Lower Limit(Pressure)
Set the lower pressure limit that the system operates under the normal condition. This value is a
guide value for error displaying, the pump stops if the value is lower than limit.
Upper Limit(Pressure)
Set the upper pressure limit that the system operates under the normal condition. This value is a
guide value for error displaying, the pump stops if the value is higher than limit.
Flow
Set the flow rate of pump.
Mix
Set the solvent ratio for mixing, can use up to 4 solvents if the pump is YL9110 quaternary.
- 120 -
Constant Mode
Deliver the solvent under the constant pressure during analysis.
Lower Limit(Flow)
Set the lower flow limit that the system operates under the normal condition. This value is a guide
value for error displaying, the pump stops if the value is lower than limit.
Upper Limit(Flow)
Set the upper flow limit that the system operates under the normal condition. This value is a guide
value for error displaying, the pump stops if the value is higher than limit.
Pressure
Set the pressure to operate.
Gradient Mode
Set the different flow and solvent ratio according to the analysis time to get the proper separation.
The gradient graph is displayed if you drag the right side to left using mouse.
You can set the graph condition using right mouse click on the graph.
- 121 -
- 122 -
Detector Condition
On the condition part, it displays wavelength, light intensity of SAM/REF, and absorbance at present.
If you click right mouse button on this window, you can select the condition to display in the device
monitor window.
Auto Zero
Set the absorbance value zero.
- 123 -
Diagnosis
If you click this button, the system checks the voltage, operation status, sensor status to confirm that the
system operation is normal. At first, select the item wan to test and then click start button for diagnosis.
Scan
Using scan function, can get the spectrum data of sample on the specific wavelength range.
Set the start/stop wavelength range and file path to save the scanned data.
- 124 -
Verification
Using verification function, can verify the detector performance of wavelength accuracy.
The Holmium filter is built-in the detector.
Fault Reset
If the error is happened, this button is activated and reset the system using this button.
Acquisition setup
Set the channel, sampling rate, and run time. Run time can be changed during analysis, but the sampling rate
should not be changed during analysis to get the meaningful result. The sampling rate(Hz) is the number of point
per second to make a chromatogram.
If the channel number is 0, that channel can not acquire the data.
Detector control setup
Config
Set the configuration of detector.
- 125 -
- 126 -
Time Constant
Set the time constant value of filter for baseline.
Data Filter
Select the type of filter.
Auto Zero
Select auto zero applying point, acquisition start and/or wavelength changing.
Auto Offset
Set the baseline offset to connect the baseline when the wavelength or polarity is changed.
External Input/Output
Set the type of external signal input/output.
- 127 -
Pulse width
If you select the pulse as a external output signal, set the duration time of output signal.
Analog Signal
Set the analog signal output. Analog signal output can be set for two acquisition channels.
Single wavelength or Dual wavelength
Lamp
Set the status of each lamp ON/OFF.
It takes time to turn on the D2 lamp for pre-heating time. Lamp status can be checked on the condition
window.
Wavelength for Channel A/B
Set the wavelength of channel A and B.
Cut Filter
Select the type of Cut filter.
Time Program
The detector condition as like wavelength can be changed according to analysis time.
Using right mouse click on the table, edit the column item.
- 128 -
- 129 -
- 130 -
Pulse width
If you select the pulse as a external output signal, set the duration time of output signal.
Min. Temperature
Set the minimum temperature that the system operates under the normal condition. This value is a guide
- 131 -
value for error displaying, the column comp. stops if the value is lower than limit.
Max. Temperature
Set the maximum temperature that the system operates under the normal condition. This value is a guide
value for error displaying, the column comp. stops if the value is higher than limit
Isothermal
Set and maintain constant temperature during analysis.
Set temp.
Set the temperature of column Comp..
Gradient
Set the temperature program according to time and also make a event program for valve.
The gradient graph is displayed if you drag the right side to left using mouse.
You can set the graph condition using right mouse click on the graph.
- 133 -
- 134 -
7. After finished analysis time(Run time) for each sample, the Task is changed from Run to File. Inject next
sample.
8. If you autosampler, the autosampler will inject all of samples according to setup condition of autosampler.
- 136 -
5. On the left side of Analysis window, double click Calibration and then the calibration table is displayed.
6. Input the Amount unit and select each compound name, and then the calibration result is displayed.
If you want to use this calibration result to analyze unknown sample, click OK button.
7. Pressing Ctrl+Shift key, select all of the unknown samples. The colors of selected lines are changed blue.
Click icons on the top of Analysis window step by step(Initialize -> Integrate -> Calibrate -> Quantification).
- 137 -
- 138 -
- 139 -
- 140 -
- 141 -
- 142 -
- 143 -
- 144 -