You are on page 1of 77

IQ Documentation

OQ Documentation
PQ Documentation

For products from the company Memmert GmbH + Co. KG

Version: 2012-05/1

Memmert GmbH + Co. KG


uere Rittersbacher Str. 38
91126 Schwabach, Germany
Phone 09122/925-0
E-Mail:sales@memmert.com
www.memmert.com
www.atmosafe.com
www.facebook.com/memmert.family

Overview
Table of Contents
Version: Version: 2012-05/1

Applicable chapter based on appliances


Appliances
Chapter
U/I/SNB/E
1. Introduction / General
1.1
x
1.2
x
1.3
x
1.4
1.4.1
x
1.4.2
x
1.4.3
1.4.4
1.4.5
x
1.4.6
2. IQ
2.1
x
2.2
x
2.3
x
2.4
x
2.5
2.5.1
x
2.5.2
x
2.6
x
3. OQ
3.1
x
3.2
3.2.1
x
3.2.2
3.2.3
3.2.4
3.2.5
3.2.6
3.2.7
3.3
3.3.1
3.3.2
x
3.3.3
3.3.4
3.3.5
3.3.6
3.3.7
3.3.8
3.3.9
3.4
3.4.1
x
3.4.2
x
3.5
3.6
3.7
3.8
3.8.1
3.8.2
3.9
3.10
3.11
x
4. PQ
4.1
4.1.1
x
4.1.2
x
4.2
4.3
4.4
4.5
4.5.1
4.5.2
4.6
4.7
4.8
x
5.
x
Attachment

U/I/SNP

U/I/SFB/E

U/I/SFP

HPP

INCO2

HCP

ICP

ICH

IPP

IPS

CTC/TTC

VO

x
x
x

x
x
x

x
x
x

x
x
x

x
x
x

x
x
x

x
x
x

x
x
x

x
x
x

x
x
x

x
x
x

x
x
x

x
x

x
x

x
x

x
x

x
x

x
x

x
x

x
x

x
x

x
x

x
x

x
x

x
x

x
x

x
x

x
x

x
x
x
x

x
x
x
x

x
x
x
x

x
x
x
x

x
x
x
x

x
x
x
x

x
x
x
x

x
x
x
x

x
x
x
x

x
x
x
x

x
x
x
x

x
x
x
x

x
x
x

x
x
x

x
x
x

x
x
x

x
x
x

x
x
x

x
x
x

x
x
x

x
x
x

x
x
x

x
x
x

x
x
x

x
x

x
x

x
x
x
x
x

x
x

x
x

x
x

x
x
x

x
x
x
x

x
x

x
x

x
x
x

x
x
x
x
x

x
x
x

x
x

x
x
x
x

x
x

x
x

x
x
x

x
x
x
x

x
x

x
x

x
x

x
x
x
x
x

x
x
x
x
x

x
x
x

x
x

x
x

x
x

x
x
x
x

x
x

x
x

x
x
x

x
x
x
x

x
x

x
x

x
x

Testing personnel ........................................................................................... 6

Overview
Table of Contents
Version: Version: 2012-05/1

Revision history .............................................................................................. 6


Distributor...................................................................................................... 6
1

Introduction / General ............................................................................ 7


Purpose of qualifications .......................................................................... 7
Measuring equipment............................................................................... 7
Test object ................................................................................................ 8
Test data of initial test .............................................................................. 8
Electrical test ............................................................................................ 8
Function test and calibration oven, incubator, climate chamber........... 10
Function test and calibration vacuum oven .......................................... 11
Calibration light intensity ........................................................................ 12
Test of the overtemperature protection device oven, incubator, climate
chamber ................................................................................................. 14
Test of overtemperature protection device vacuum oven ..................... 14

IQ ......................................................................................................... 16
Packaging/ appliance .............................................................................. 16
Documentation / accessories ................................................................... 16
Installation of the appliance.................................................................... 16
Installation of software ........................................................................... 17
Assessment of the qualification .............................................................. 17
Damaged or missing parts ...................................................................... 17
Preparing for operation / instruction of operating personnel .................. 19
Release for next qualification level .......................................................... 19

OQ ........................................................................................................ 20
Testing the display elements ................................................................... 20
Test of the alarm- /overtemperature protection function ........................ 22
Universal oven, incubator, steriliser (Uxx/ Ixx /Sxx) ................................... 22
Constant climate chamber (HPP) ............................................................. 24
CO2 incubator / humidity chamber (INC / HCP) ....................................... 26
Cooled incubator (ICP / ICH / IPP) ............................................................ 26
IPS Temperature control storage chamber (IPS)....................................... 29

Overview
Table of Contents
Version: Version: 2012-05/1

Climatic / Temperature test chamber (CTC / TTC) ..................................... 31


Vacuum oven (VO) .................................................................................. 33
Functional tests....................................................................................... 34
Internal card reader ................................................................................ 34
Air vent controls, mechanical .................................................................. 35
Air vent controls, electrical ...................................................................... 36
Active humidity control or humidity limit control .................................... 37
Control air circulation motor................................................................... 39
Solenoid valves VO: OUT, IN1 and / or IN2 .................................... 40
VO leak test ............................................................................................ 41
Solenoid valves INCO2: IN1 and IN2 ................................................. 42
Functional testing option: Light module ................................................. 43
Testing temperature distribution
in compliance with DIN 12880: 2007-05 ................................................. 44
Measurement results for
temperature distribution / humidity measurement .................................. 45
Measurement results on the temperature fluctuation ............................. 50
Measurement result for calibrating humidity .......................................... 51
Measurement result for calibrating CO2 ................................................. 51
Measurement result for calibrating O2 ................................................... 51
Testing the thermoshelves ...................................................................... 52
Measurement results for area temperature distribution (VO) .................. 53
Measurement results on the temperature fluctuation (VO) ..................... 55
Calibration of the vacuum controller....................................................... 56
Measurement of light intensity ............................................................... 58
Release of the tested appliance for use ................................................... 59

PQ ........................................................................................................ 60
Testing temperature distribution
in compliance with DIN 12880: 2007-05 ................................................. 60
Measurement results for
temperature distribution / humidity measurement .................................. 61
Measurement results on the temperature fluctuation ............................. 65
Measurement result for calibrating humidity .......................................... 66

Overview
Table of Contents
Version: Version: 2012-05/1

Measurement result for calibrating CO2 ................................................. 66


Measurement result for calibrating O2 ................................................... 66
Testing the thermoshelves ...................................................................... 67
Measurement results for area temperature distribution (VO) .................. 68
Measurement results on the temperature fluctuation (VO) ..................... 69
Calibration of the vacuum controller....................................................... 70
Measurement of light intensity ............................................................... 72
Release of the tested appliance for use ................................................... 73

Attachment .......................................................................................... 74

Notes 75

Appliance Qualification
Version: Version: 2012-05/1

Date: 30.08.2012

Testing personnel
Company-internal qualification
Created by:
Memmert
GmbH + Co. KG
Reviewed by:
Memmert
GmbH + Co. KG

Name

Position

Signature

Date

---

---

---

---

---

---

Name

Position

On site qualification

Signature

Date

Performed by:

Performed by:

Performed by:

Revision history
Version
01

Date

Reason for revision


New facility / first installation

02

Installation after successful repair

03

New application field for appliance

Signature

Distributor
Department

Function

Number/ copies
Original
1. copy
2. copy

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 6 of 76

Appliance Qualification
Introduction / General
Date: 24.05.2012

Version: Version: 2012-05/1

Introduction / General

Chapter not applicable must be marked NA

Purpose of qualifications

IQ: This qualification is used to review the appliance listed under test object for
the completeness of the delivery and for its fundamental readiness for operation.
It also documents all important specifications and parameters stipulated by the
manufacturer, whether the safety regulations are fulfilled and which persons have
been instructed in the basic functions of the appliance and its operation.
OQ: This qualification is used for the verification / testing of the function and
production readiness of the appliance listed under test object.
The following is tested on site:
- All quality or process relevant alarm and switching functions
- All quality or process-relevant control functions
- Temperature distribution as well as measurement of humidity,
/CO2/O2/pressure and light
The settings for the setpoints (parameters) necessary for the test must be made as
described in the respective operating manuals
PQ: This documentation is used as a support for the testing of the function and
production readiness under process-dependent conditions of the appliance listed
under test object.
The following is recorded on site with the test specimens:
- All quality or process relevant alarm and switching functions
- All quality or process relevant control functions
- Temperature distribution as well as measurement of humidity, /CO2
O2/pressure and light

Measuring equipment

The calibration certificates for the measuring instruments used for temperature,
humidity, pressure, CO2, O2 and lux and W/m can be found in the attachment.

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 7 of 76

Appliance Qualification
Introduction / General
Version: Version: 2012-05/1

Date: 24.05.2012

Test object

Manufacturer:

MEMMERT GmbH + Co. KG, 91107 Schwabach, Germany

Designation:
Type:
Voltage:
Nominal
temperature:
Serial No.

----- ----- V, 50 / 60 Hz
--- C

yes
yes
yes

no
no
no

yes

no

VO Shelves
Shelf 1
Shelf 2
Shelf 3
Shelf 4

Serial number
# --# --# --# ---

Variant
---------

N/A
yes
yes
yes
yes

no
no
no
no

Test data of initial test

Electrical test
PE conductor resistance
High voltage test
Power input

Limit values

Measured values

max. 200 m with a minimum 10 A


at least 1.25 kV at max. 3000 A

--- m at --- A
--- kV at --- A
--- Watt

(for VO-thermoshelves position 1 and 2 for models 400 and 500, position 1 for model 200)

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 8 of 76

Appliance Qualification
Introduction / General
Date: 24.05.2012

Only VO:
N/A
PE conductor resistance

High voltage test

Version: Version: 2012-05/1

Limit values

Measured values

max. 200 m with a minimum 10 A

--- m at Fehler!
Verweisquelle konnte
nicht gefunden werden.
A
--- kV at Fehler!
Verweisquelle konnte
nicht gefunden werden.
A
--- Watt

at least 1.25 kV at max. 3000 A

Power consumption
(for VO-thermoshelves position 3 and 4 for models 400 and 500, position 2 for model 200)

Only VO:
N/A
PE conductor resistance

High voltage test

Limit values

Measured values

max. 200 m with a minimum 10 A

--- m at Fehler! Es
wurde kein
Textmarkenname
vergeben. A
--- kV at Fehler! Es
wurde kein
Textmarkenname
vergeben. A
--- Watt

at least 1.25 kV at max. 3000 A

Power consumption
(for VO-thermoshelves position 3 for model 200)

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 9 of 76

Appliance Qualification
Introduction / General
Date: 24.05.2012

Version: Version: 2012-05/1

Func
Function test and calibration oven,
oven, incubator, climate chamber
N/A
Function test
Tested at a room temperature of 22C +/-2K
Warm-up time up to 40 C
cool down time from 40 C to 5 C
Maximum working temperature reaches 50 C

--- C
--- Minutes
--- minutes
yes
no

Calibration setpoint 1
Setpoint temperature 1
Actual value on the controller
Reference value measured in the working chamber (acc. to
DIN 12880 and 50011), permissible: actual value +/-0,5 K
Calibration setpoint 2
N/A
Setpoint temperature
Actual value on the controller
Reference value measured in the working chamber (acc. to
DIN 12880 and 50011), permissible: actual value +/-0,5 K
Calibration set value 3
N/A
Setpoint temperature
Actual value on the controller
Reference value measured in the working chamber (acc. to
DIN 12880 and 50011), permissible: actual value +/-0,5 K
Humidity calibration at 30 C
Humidity setpoint
Actual value on the controller
Reference value measured in the working chamber,
permissible: actual value +/-3% rh

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

--- C
--- C
--- C

--- C
--- C
--- C

--- C
--- C
--- C

N/A

Ref.: D24374_IQ_OQ_PQ_eng

--- %rh
--- %rh
--- %rh

Page 10 of 76

Appliance Qualification
Introduction / General
Date: 24.05.2012

Version: Version: 2012-05/1

CO2 calibration
Setpoint value CO2
Actual value on the controller
Reference value measured in the working chamber,
permissible: actual value +/-0.3 % CO2
Setpoint value CO2
Actual value on the controller
Reference value measured in the working chamber,
permissible: actual value +/-0.5 % CO2
Setpoint value CO2
Actual value on the controller
Reference value measured in the working chamber,
permissible: actual value +/0.7 % CO2

N/A

O2 calibration
N/A
Setpoint value O2
Actual value on the controller
Reference value measured in the working chamber,
permissible: actual value +/-0.5 % O2

5 % CO2
--- % CO2
--- % CO2
10 % CO2
--- % CO2
--- % CO2
15 % CO2
--- % CO2
--- % CO2

5 % O2
--- % O2
--- % O2

Function test and calibration vacuum oven


N/A
Function test
Tested at a room temperature of 22C +/- 2K
Heating up to 200 C at 50 mbar
(on 1st shelf from bottom)

--- C
yes

no

Calibration
Setpoint on shelf at 50 mbar
--- C
Actual value on the controller
--- C
Thermoshelf 1 = reference value measured on 1st shelf
--- C
from bottom,
permissible: stainless steel shelf actual value +/-5 K
aluminium shelf shelf actual value +/2 K
The other thermoshelves can be ignored since controlling for all thermoshelves
is from the same controller.
Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 11 of 76

Appliance Qualification
Introduction / General
Date: 24.05.2012

Version: Version: 2012-05/1

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 12 of 76

Appliance Qualification
Introduction / General
Date: 24.05.2012

Version: Version: 2012-05/1

N/A
ICP
(measurement is made vertically, in parallel with and at a distance of 10 cm from the
light module, on the middle level, see measurement arrangement Nxxx in the
attachment)

Calibration pressure
Setpoint pressure
Actual value on the controller
measured reference value, permissible: actual value +/- 3
mbar

--- mbar
--- mbar
--- mbar

Calibration light intensity


intensity
N/A
To record the light intensity, the light modules must be switched on for at least 30
minutes

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 13 of 76

Appliance Qualification
Introduction / General
Date: 24.05.2012

Version: Version: 2012-05/1

Measured reference value


--- Lux
(at least ICP500 1,000 Lux, ICP600 2,500 Lux, ICP700 1,000 Lux, ICP800 2,000 Lux)
HPP108L
N/A
(measurement is made vertically, in parallel with and at a distance of 10 cm from the
light module,
3. bead from below, see measurement arrangement Nxxx in the attachment)
cold white (light performance set to 100%) N/A
Measured reference value (at least 8,000 Lux)
--- Lux
cold white /warm white (light performance set to 100%)
N/A
Measured reference value (at least 8,000 Lux)

--- Lux

ICH256L
N/A
ICH256L (Standard light cartridge)
(measurement is made on the 4th bead from the bottom, centrally, beneath the light
cartridge,
see measurement arrangement Nxxx in the attachment)
cold white
Measured reference value (at least 8,000 Lux)
--- Lux
UV light
Measured reference value (at least 4 W/m)

--- W/m

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 14 of 76

Appliance Qualification
Introduction / General
Date: 24.05.2012

Version: Version: 2012-05/1

Test of the overtemperature protection device oven, incubator, climate


chamber
N/A
For appliances of the P-class a selection of the temperature protection class (SK 3.x
/SK 2) is necessary (see operating instructions basic appliance settings chapter).
The acoustic alarm (for P class only) must also be activated (see operating
instructions basic appliance settings chapter.)
Over/ undertemperature protection class
(delete if not applicable)
Setpoint set to
Switch off point overtemperature protection (display
temperature controller) 3.1 /3.3 TWW and TWB 2
Limit value 5K above set setpoint
Switch off point overtemperature protection (display
temperature controller) 3.1 /3.3 TWW
Limit value 5K below set setpoint

Class 3.1/3.3 TWW


Class 2 TWB
--- C
--- C

--- C

Test of overtemperature protection device vacuum oven


N/A
A selection of the temperature protection class SK 3.1/SK 2 is necessary (see
operating instructions basic appliance settings chapter). The acoustic alarm
must also be activated (see operating instructions basic appliance settings
chapter.)

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 15 of 76

Appliance Qualification
Introduction / General
Date: 24.05.2012

Version: Version: 2012-05/1

Setpoint set to 70C.


Thermoshelf 1 = 1st shelf from bottom
Switch off point < 73 C
--Switch on point > 68 C
--The other thermoshelves can be ignored since the overtemperature control is the same
for all thermoshelves.

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 16 of 76

Appliance Qualification
IQ
Date: 30.08.2012

Version: Version: 2012-05/1

IQ
Packaging/ appliance

Packaging undamaged?

yes

no

Are all appliance parts undamaged as far as can be seen on outside?

yes

no

Completeness of the scope of delivery, according to the order and


order confirmation (e.g. steel grids/ thermoshelves)

yes

no

yes

no

yes

no

yes

no

yes

no

yes

no

yes

no

yes

no

yes

no

Documentation / accessories

Operating manual available?


Control software CELSIUS on CD-ROM is at hand?
(only with PERFECT appliances and INC with communication
module)
Operating manual for CELSIUS at hand?
(only with PERFECT appliances and INC with communication
module)
STERI-card at hand? (INP / IFP / INC / HCP)
MEMory-card at hand?
(only with PERFECT appliances, not with INC)
Calibration certificate available?(VO: one for each thermoshelf)
Installation of the appliance

Uxx / Ixx / Sxx / VO appliances


N/A
Compliance with the following environmental conditions?
- Ambient temperature +5 C to +40 C
- rel. humidity max. 80 %, not condensing
INC / HCP appliances
N/A
Compliance with the following environmental conditions?
- Ambient temperature +5 C to +35
- rel. humidity max. 80 %, not condensing

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 17 of 76

Appliance Qualification
IQ
Date: 30.08.2012

Version: Version: 2012-05/1

IPP / ICP / IPS / ICHICH- appliances


N/A
Compliance with the following environmental conditions?
- Ambient temperature
+12 C to +28 C (IPP / IPS-models)
+16 C to +28 C (ICP-models)
+ 5 C to +28 C (ICH-models)
- rel. humidity max. 80%, not condensing

yes

no

yes

no

Does the local mains voltage match with the nameplate?

yes

no

Mains voltage connected according to manufacturer specifications?

yes

no

yes
yes
yes

no
no
no

yes
yes

no
no

HPP / CTC / TTC appliances


N/A
Compliance with the following environmental conditions?
- Ambient temperature +16 to +28
- rel. humidity max. 70 %, not condensing
Mains voltage

Installation of software

Installation der CD: Celsius (falls Vorhanden)


Has the software been successfully installed?
Does the software start without an error message?
Can the chamber connection (online) be established?

N/A

Assessment of the qualification

Damaged or missing parts

The appliance is operational


Corrective action necessary
The necessary corrective action must be performed by:
Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

________________

Page 18 of 76

Appliance Qualification
IQ
Date: 30.08.2012

Version: Version: 2012-05/1

The action itself must be documented and confirmed in the document:


_______________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 19 of 76

Appliance Qualification
IQ
Date: 30.08.2012

Version: Version: 2012-05/1

Preparing for operation / instruction of operating personnel


Appliance was correctly installed and is operational:
Instruction of operating personnel:

yes
yes

Name of operator instructed


Signature
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
Instruction of operating personnel on the appliance was performed by:
Name

Company

Date

no
no
Date

Signature

Release for next qualification


qualification level

The release of the appliance for the next qualification level (OQ) is done by:
Name

Company / Department

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Date

Ref.: D24374_IQ_OQ_PQ_eng

Signature

Page 20 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

OQ
Testing the display elements

Setting the normal operation mode (see operating manual)


Does display light up?
yes
no
Setpoint energy controller 10 C above displayed actual temperature
Is the yellow heating icon flashing?
yes
no
Setpoint energy controller on maximum
Is the yellow heating icon lit permanently?
yes
no
Setpoint of the overtemperature controller to minimum (low actual temperature)
Is the red alarm icon lit up and the heating icon is off?
yes
no
Setpoint of the overtemperature controller to maximum (high actual temperature)
Is the yellow heating icon lit up and the alarm icon is off? (Nur bei
yes
no
Schutzklasse 3.1 TWW)

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 21 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

In addition for HPP / IPP / ICP / ICH / CTC / TTC appliances


N/A
Setpoint energy controller minimum
Is the cooling icon permanently lit up?
yes
no
Setpoint of the overtemperature controller to minimum (low actual temperature)
Is the red alarm icon lit up and the heating icon and cooling icon are
yes
no
off?
Setpoint of the overtemperature controller to maximum (high actual temperature)
Is the cooling icon lit up (flashing) and the alarm icon is off? (Only
for protection class 3.3 TWW)
yes
no

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 22 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Test of the alarmalarm- /overtemperature protection function


The functional readiness of the overtemperature protection function must be tested according to
predefined fixed limit values.
First, the specified test temperature must be set. 30 minutes after the setpoint is stable the
temperature protection must be adjusted in 1 C steps downwards or upwards until the alarm is
triggered or switched off.
For appliances of the P-class, a selection of the temperature protection class is necessary (see
operating instructions basic appliance settings chapter). The acoustic alarm (for P class
appliances/INC/IPS/VO/ICH only) must also be activated (see operating instructions basic appliance
settings chapter.)

Universal oven, incubator, steriliser (Uxx/ Ixx /Sxx)


N/A
Electronic overtemperature protection class
Class 3.1 TWW
(delete where not applicable)
Class 2 TWB
Test temperature Uxx/Sxx 70 C; Ixx 40 C, working temperature settled
Switch off temperature (TWW/TWB) > 68 C (39C)
Alarm icon on
yes
no
Acoustic alarm on (P-class only)
yes
no
Switch on temperature (TWW) < 72 C (41C)
Alarm icon off
yes
no
Acoustic alarm off (P-class only)
yes
no

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 23 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Test temperature Uxx/Sxx 250 C, Ixx 70 C


(working temperature settled)
Abschalttemperatur (TWW/TWB) > 248 C (69 C)
Alarm icon on
Acoustic alarm on (P-class only)
Switch on temperature (TWW) < 252 (71 )
Alarm icon off
Acoustic alarm off (P-class only)

yes
yes

no
no

yes
yes

no
no

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 24 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Additionally for protection class 2 TWB


N/A
Is the alarm icon flashing?
(even if setpoint / actual value is changed)?
Is the activated alarm icon, switched off when appliance is switched
off and on again (TWB reset)?

yes

no

yes

no

Constant climate chamber (HPP)


N/A
Electronic over - /undertemperature protection class
(delete where not applicable)
Test temperature10.0 C (working temperature settled)
Switch off temperature (TWW/TWB) > 9.7
Alarm icon on
Acoustic alarm on
Switch on temperature (TWW) < 10.3
Alarm icon off
Acoustic alarm off
Test temperature 37.0 C (working temperature settled)
Switch off temperature (TWW/TWB) > 36.7
Alarm icon on
Acoustic alarm on
Switch on temperature (TWW) < 37.3
Alarm icon off
Acoustic alarm off
Additionally for protection class 2 TWB
N/A
Is the alarm icon flashing (even if setpoint / actual value is
changed)?

Class 3.3 TWW


Class 2 TWB

yes
yes

no
no

yes
yes

no
no

yes
yes

no
no

yes
yes

no
no

yes

no

Comments: _________________________________________________________________
Is the activated alarm icon, switched off when appliance is switched
yes
no
off and on again (TWB reset)?

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 25 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 26 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

CO2 incubator
incubator / humidity chamber (INC / HCP)
N/A
Electronic overtemperature protection class
Class 3.1 TWW
INC / HCP - test temperature 37 HCP (working temperature stabilised)
Switch-off temperature (TWW) < 36.7
Alarm icon on
yes
no
Acoustic alarm on
yes
no
Switch on temperature (TWW) < 37.3
Alarm icon off
yes
no
Acoustic alarm off
yes
no
HCP - test temperature 160 (working temperature stabilised)
Switch-off temperature (TWW) < 158
Alarm icon on
Acoustic alarm on
Switch on temperature (TWW) < 162
Alarm icon off
Acoustic alarm off

yes
yes

no
no

yes
yes

no
no

Cooled incubator (ICP / ICH / IPP)


N/A
Electronic over - /undertemperature protection class
(delete where not applicable)
Test temperature10.0 C (working temperature settled)
Switch off temperature (TWW/TWB) > 9.5
Alarm icon on
Acoustic alarm on
Switch on temperature (TWW) < 10.5

Class 3.3 TWW


Class 2 TWB

yes
yes

no
no

Comments: _________________________________________________________________
Alarm icon off
yes
no
Acoustic alarm off
yes
no

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 27 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 28 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Test temperature 37.0 (working temperature stabilised)


Switch off temperature (TWW/TWB) > 36.5
Alarm icon on
Acoustic alarm on
Switch on temperature (TWW) < 37.5
Alarm icon off
Acoustic alarm off
Zustzlich bei Schutzklasse 2 TWB
N/A
Blinkt das Alarm-Symbol (auch bei nderung des Sollwertes /
Istwertes)?
Wird das aktivierte Alarm-Symbol durch Aus Einschalten des
Gertes wieder abgeschaltet (TWB zurckgesetzt)?

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

yes
yes

no
no

yes
yes

no
no

yes

no

yes

no

Page 29 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

IPS Temperature control storage chamber (IPS)


N/A
Electronic overtemperature protection class
(delete where not applicable)
Test temperature 20.0 C (working temperature settled)
Switch off temperature (TWW/TWB) > 19.5
Alarm icon on
Acoustic alarm on
Switch on temperature (TWW) < 20.5
Alarm icon off
Acoustic alarm off

Class 3.3 TWW


Class 2 TWB

yes
yes

no
no

yes
yes

no
no

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 30 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Test temperature 37.0 (working temperature settled)


Switch off temperature (TWW/TWB) > 36.5
Alarm icon on
Acoustic alarm on
Switch on temperature (TWW) < 37.5
Alarm icon off
Acoustic alarm off
Additionally for protection class 2 TWB
N/A
Is the alarm icon flashing (even if setpoint / actual value is
changed)?
Is the activated alarm icon, switched off when appliance is switched
off and on again (TWB reset)?

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

yes
yes

no
no

yes
yes

no
no

yes

no

yes

no

Page 31 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Climatic / Temperature test chamber (CTC / TTC)


N/A
Electronic over - /undertemperature protection class
(delete where not applicable)
Test temperature -20 (working temperature settled)
Switch off temperature (TWW/TWB) > 20.5
Alarm icon on
Acoustic alarm on
Switch on temperature (TWW) < 19.5
Alarm icon off
Acoustic alarm off
Test temperature 30 (working temperature settled)
Switch off temperature (TWW/TWB) > 29.5
Alarm icon on
Acoustic alarm on
Switch on temperature (TWW) < 30.5
Alarm icon off
Acoustic alarm off

Class 3.3 TWW


Class 2 TWB

yes
yes

no
no

yes
yes

no
no

yes
yes

no
no

yes
yes

no
no

Comments: __________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 32 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Test temperature 160 (working temperature settled)


Switch off temperature (TWW/TWB) > 159.5
Alarm icon on
Acoustic alarm on
Switch on temperature (TWW) < 160.5
Alarm icon off
Acoustic alarm off
Additionally for protection class 2 TWB
N/A
Is the alarm icon flashing (even if setpoint / actual value is
changed)?
Is the activated alarm icon, switched off when appliance is switched
off and on again (TWB reset)?

yes
yes

no
no

yes
yes

no
no

yes

no

yes

no

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 33 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Vacuum oven (VO)


N/A
Electronic overtemperature protection class
(delete where not applicable)

Class 3.1 TWW


Class 2 TWB

Test temperature 70 , working temperature settled


Settings setpoint: Overtemperature controller 72C / working chamber 70C
Thermoshelf 1 = 1st shelf from bottom
Switch off point < 73
yes
no
Switch on point > 69
yes
no
Alarm icon on
yes
no
Acoustic alarm on
yes
no
The other thermoshelves can be ignored since the controls for all thermoshelves
are the same.
Test temperature 160 , working temperature settled
Settings setpoint: Overtemperature controller 162 / working chamber 160
Thermoshelf 1 = 1st shelf from bottom
Switch off point < 163
yes
no
Switch on point > 159
yes
no
Alarm icon on
yes
no
Acoustic alarm on
yes
no
The other thermoshelves can be ignored since the controls for all thermoshelves
are the same.

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 34 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Functional tests

Internal card reader


N/A
MEMory Card
N/A
Switch appliance to normal operation mode (see operating manual Selecting
operating modes UIS PERFECT/HPP/IPP/ICP/ICH/HCP/CTC/TTC), insert MEMory card into
internal card reader
Display of temperature controller shows 32KB MEMCARD OK?

yes

no

Automatic switching to ramp timer?

yes

no

SteriCard
N/A
Insert SteriCard into the card reader (see operating manual Sterilisation chip card)
(only for INC / INP / IFP /HCP appliances)
Display of temperature controller shows STERI CARD OK?

yes

no

USER ID
N/A
Set appliance to Normal operation mode (see operating manual selecting operating
mode), insert USER ID Card into the internal card reader
(only PERFECT appliances)
Display of temperature controller shows ID OK?

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

yes

no

Page 35 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Air vent controls, mechanical


N/A
(only for UxB/E IxB/E SxB/E appliances)
Switch appliance to Normal operation mode (see operating
manual selecting operating mode chapter)
Set slider to air vent position MAX (right), is air vent (rear of
appliance) fully open?
Set slider to air vent position MIN (left), is air vent (rear of
appliance) fully closed?

N/A
yes

no

yes

no

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 36 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Air vent controls, electrical


N/A
(only for UxP/ IxP/ SxP appliance)
Switch appliance to Normal operation mode (see operating
manual selecting operating mode chapter)
Select FLAP display icon and set air vent position to MAX (right),
is air vent (rear of appliance) fully open?
Select FLAP display icon and set air vent position to MIN (right),
is air vent (rear of appliance) fully closed?

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

N/A
yes

no

yes

no

Page 37 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Active humidity control or humidity limit control


N/A
Fill container as described in the operating manual under Initial operation
(for the limit control, a water tray, for active humidity control, a can).
N/A
(bei HPP / ICH / HCP / CTC-Gert)
Set working temperature to 40.0 C
(working temperature settled)
Active humidity control
Humidifying: Humidity setpoint to 60 % rh
After a few minutes the steam process icon lights up (display for
steam supply)
After a maximum of 30 minutes the setpoint is reached
(60 % rh +/- 3%rh)
Dehumidifying: Humidity setpoint to 40 % rh
After a few minutes the dehumidifying icon lights up (display for
steam supply)
After a maximum of 30 minutes the setpoint is reached
(40 % rh +/- 3%rh)

yes

no

yes

no

yes

no

yes

no

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 38 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Set working temperature to 37.0 -working temperature settled


Active humidity control:
N/A
Humidifying: Humidity setpoint to 95 % rh
Does the steam process icon light up after a few minutes?
Is the set relative humidity of 95 % rh +/- 3% rh reached after max.
30 minutes?
Dehumidifying: Humidity setpoint to 60% rh
After max. 30 minutes the setpoint of 60 % rh +/- 3% rh is reached
Humidity limit control:
N/A
Humidifying: Humidity setpoint to 95 % rh
Is the set relative humidity of 95 % rh +/- 3% rh reached after max.
90 minutes?
Dehumidifying: Humidity setpoint to 88 % rh
Nach max. 30 Minuten ist der Sollwert von 88 % rh +/- 3% rh
erreicht
N/A
(for INC appliance)

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

yes
yes

no
no

yes

no

yes

no

yes

no

Page 39 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Control air circulation motor


N/A
(for UFx/ IFx/ SFx/ ICP / ICH / CTC / TTC appliance)
Switch appliance to Normal operation mode (see operating
manual selecting operating mode chapter)
Select FAN display icon and set air speed to MAX, does fan run
at maximum speed?
Select FAN display icon and set air speed to MIN, does fan run
at minimum speed?
Select FAN display icon and set air speed to Off, does fan stop
running? (test does not apply for ICP / ICH / CTC / TTC!)

yes

no

yes

no

yes

no

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 40 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Solenoid valves VO: OUT, IN1 and / or IN2


N/A
Set appliance to Normal operation mode (see operating manual
selecting operating mode chapter), connect vacuum supply
Set the vacuum setpoint smaller than the actual value (e.g. 500
mbar), does OUT output icon light up and is OUT solenoid valve
opened? (pressure value is reduced)

yes

no

Activate ventilation input IN1 (see operating manual chapter


Basic device settings)
Set the vacuum setpoint to maximum, does IN1 input icon light
yes
no
up and is IN1 solenoid valve opened? (pressure value increases)
Only if equipped with switch to inert gas
N/A
Activate ventilation input IN2 (see operating manual chapter basic device settings),
set vacuum setpoint to 500 mbar and wait for conditions to settle
Set the vacuum setpoint to maximum (larger than actual value),
yes
no
does IN2 input icon light up and is IN2 solenoid valve opened?
Is ventilation via IN1 quicker than ventilation via IN2?
yes
no

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 41 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

VO leak test
N/A
Switch appliance to Normal operation mode (see operating manual selecting
operating mode chapter)
Set vacuum setpoint to 50 mbar, wait for conditions to settle (max. 20 minutes)
Is actual value, the same as setpoint and fluctuation less than +/- 3
yes
no
mbar
Switch of vacuum source (pump module external vacuum supply),
disconnect solenoid valve (as described in Service instructions) and seal off OUT
Pressure display after one hour < 60 mbar
yes
no
(Vacuum loss < 10 mbar)

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 42 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Solenoid valves INCO2: IN1 and IN2


N/A
INCO2 (only for corresponding model)
Activate CO2 connection IN1 (see operating manual chapter basic device settings),
the CO2 supply must be connected
Set the CO2 setpoint to maximum (larger than actual value), does
IN1 input icon light up and is IN1 solenoid valve opened? (CO2
yes
no
actual value increases)
Interrupt CO2 connection IN1, activate gas input IN2 (see operating manual
chapter basic device settings), the CO2 supply must be connected
Set the CO2 setpoint to maximum (larger than actual value), is IN1
input icon off, does IN2 input icon light up and is IN2 solenoid
yes
no
valve opened? (CO2 actual value increases)
Interrupt CO2 gas supply IN1 and IN2
Set CO2 setpoint to maximum (larger than actual value), alarm icon
yes
no
flashes and display shows CO2 EMPTY after max. 15 min.

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 43 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Functional testing option:


option: Light module
N/A
(only for ICP/ICH256L)
Can the light module be switched on and off (depending on device
model)
N/A
Daylight
UV light (only for ICH256L with UV lighting)

yes
yes

no
no

(only for HPP108L)


Can the light module be dimmed?
Light module can be dimmed in stages (stages: Off to 10)
10

yes

no

N/A

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 44 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Testing temperature distribution in compliance with DIN 12880: 20072007-05

N/A
The distribution of the temperature in the working chamber is recorded via 9
measurement points for an interior volume below 50 litres, above 50 litres is tested
with 27 measurement points.
Apart from determining the lowest and highest temperature inside the chamber
(distribution of temperature), the maximum fluctuations in temperature must be
recorded.
To do this, the lowest and chronologically following highest temperature at same
measurement point (middle of chamber) is recorded and documented after a state
of inertia has been reached.
Measurements performed by the company MEMMERT are documented in the
attachment. (see Attachment No. III)
OQ measurements performed at the customer must be entered in the table below.
The temperature sensors for testing temperature distribution and temperature
fluctuation, must be placed in accordance with DIN 12880: 2007-05 (see
Attachment No. II).

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 45 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Measurement results for temperature distribution / humidity measurement


Measurement 1
N/A
Set setpoint temperature
Displayed actual temperature (wait for state of inertia - at least
120 min. after setpoint temperature has been reached
Set humidity setpoint N/A
Displayed humidity actual value (a humidity setpoint +/- 3% rh)
N/A
Level 0
Sensor # 1 --- C
Sensor # 2 --- C
Sensor # 3 --- C
Sensor # 4 --- C
Sensor # 5 --- C
Sensor # 6 --- C
Sensor # 7 --- C
Sensor # 8 --- C
Sensor # 9 --- C

Level 1
Sensor # 11 --- C
Sensor # 12 --- C
Sensor # 13 --- C
Sensor # 14 --- C
Sensor # 15 --- C
Sensor # 16 --- C
Sensor # 17 --- C
Sensor # 18 --- C
Sensor # 19 --- C

--- C
--- C
--- %rh
--- %rh
Level 2
Sensor # 21 --- C
Sensor # 22 --- C
Sensor # 23 --- C
Sensor # 24 --- C
Sensor # 25 --- C
Sensor # 26 --- C
Sensor # 27 --- C
Sensor # 28 --- C
Sensor # 29 --- C

Maximum temperature difference (spatially)


permissible: <70 = +/- 0.5K; 70 - 180 = +/- 1.5K; >180 = +/3K
Humidity sensor middle of chamber (permissible: actual value +/3% rh)
N/A

--- K
--- %rh

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 46 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Measurement 2
N/A
Set setpoint temperature
Displayed actual temperature (wait for state of inertia - at least
120 min. after setpoint temperature has been reached
Set humidity setpoint N/A
Displayed humidity actual value (a humidity setpoint +/- 3% rh)
N/A
Level 0
Sensor # 1 --- C
Sensor # 2 --- C
Sensor # 3 --- C
Sensor # 4 --- C
Sensor # 5 --- C
Sensor # 6 --- C
Sensor # 7 --- C
Sensor # 8 --- C
Sensor # 9 --- C

Level 1
Sensor # 11 --- C
Sensor # 12 --- C
Sensor # 13 --- C
Sensor # 14 --- C
Sensor # 15 --- C
Sensor # 16 --- C
Sensor # 17 --- C
Sensor # 18 --- C
Sensor # 19 --- C

--- C
--- C
--- %rh
--- %rh
Level 2
Sensor # 21 --- C
Sensor # 22 --- C
Sensor # 23 --- C
Sensor # 24 --- C
Sensor # 25 --- C
Sensor # 26 --- C
Sensor # 27 --- C
Sensor # 28 --- C
Sensor # 29 --- C

Maximum temperature difference (spatially)


permissible: <70 = +/- 0.5K; 70 - 180 = +/- 1.5K; >180 = +/3K
Humidity sensor middle of chamber (permissible: actual value +/3% rh)
N/A

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

--- K
--- %rh

Page 47 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 48 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Measurement 3
N/A
Set setpoint temperature
Displayed actual temperature (wait for state of inertia - at least
120 min. after setpoint temperature has been reached
Set humidity setpoint N/A
Displayed humidity actual value (a humidity setpoint +/- 3% rh)
N/A
Level 0
Sensor # 1 --- C
Sensor # 2 --- C
Sensor # 3 --- C
Sensor # 4 --- C
Sensor # 5 --- C
Sensor # 6 --- C
Sensor # 7 --- C
Sensor # 8 --- C
Sensor # 9 --- C

Level 1
Sensor # 11 --- C
Sensor # 12 --- C
Sensor # 13 --- C
Sensor # 14 --- C
Sensor # 15 --- C
Sensor # 16 --- C
Sensor # 17 --- C
Sensor # 18 --- C
Sensor # 19 --- C

--- C
--- C
--- %rh
--- %rh
Level 2
Sensor # 21 --- C
Sensor # 22 --- C
Sensor # 23 --- C
Sensor # 24 --- C
Sensor # 25 --- C
Sensor # 26 --- C
Sensor # 27 --- C
Sensor # 28 --- C
Sensor # 29 --- C

Maximum temperature difference (spatially)


permissible: <70 = +/- 0.5K; 70 - 180 = +/- 1.5K; >180 = +/3K
Humidity sensor middle of chamber (permissible: actual value +/3% rh)
N/A

--- K
--- %rh

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 49 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Measurement 4
N/A
Set setpoint temperature
Displayed actual temperature (wait for state of inertia - at least
120 min. after setpoint temperature has been reached
Set humidity setpoint N/A
Displayed humidity actual value (a humidity setpoint +/- 3% rh)
N/A
Level 0
Sensor # 1 --- C
Sensor # 2 --- C
Sensor # 3 --- C
Sensor # 4 --- C
Sensor # 5 --- C
Sensor # 6 --- C
Sensor # 7 --- C
Sensor # 8 --- C
Sensor # 9 --- C

Level 1
Sensor # 11 --- C
Sensor # 12 --- C
Sensor # 13 --- C
Sensor # 14 --- C
Sensor # 15 --- C
Sensor # 16 --- C
Sensor # 17 --- C
Sensor # 18 --- C
Sensor # 19 --- C

--- C
--- C
--- %rh
--- %rh
Level 2
Sensor # 21 --- C
Sensor # 22 --- C
Sensor # 23 --- C
Sensor # 24 --- C
Sensor # 25 --- C
Sensor # 26 --- C
Sensor # 27 --- C
Sensor # 28 --- C
Sensor # 29 --- C

Maximum temperature difference (spatially)


permissible: <70 = +/- 0.5K; 70 - 180 = +/- 1.5K; >180 = +/3K
Humidity sensor middle of chamber (permissible: actual value +/3% rh)
N/A

--- K
--- %rh

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 50 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Measurement results on the temperature fluctuation


N/A
Recorded temperature at sensor position 11 / level 1 (middle of chamber)
N/A
Measuremen Measuremen Measuremen Measuremen
Measurement
t1
t2
t3
t4
Temperature minimum
--- C
--- C
--- C
--- C
Temperature maximum
--- C
--- C
--- C
--- C
Temperature fluctuation
max. +/- 0.5 K

--- K

--- K

Recorded humidity at sensor position (middle of chamber)


Humidity minimum
--- %rh
--- %rh
Humidity maximum
--- %rh
--- %rh
Humidity fluctuation
max. +/- 3% rh

--- % rh

--- % rh

--- K

--- K

--- %rh
--- %rh

N/A
--- %rh
--- %rh

--- % rh

--- % rh

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 51 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Measurement result for calibrating humidity

N/A
Humidity calibration
Humidity setpoint
Actual value on the controller
Reference value measured in the working chamber (actual
value +/-3% rh)

--- % rh
--- % rh
--- % rh

Measurement result for calibrating CO2

N/A
CO2 calibration
Setpoint value CO2
Actual value on the controller
Reference value measured in the working chamber (actual
value +/-0.7 % CO2)

--- % CO2
--- % CO2
--- % CO2

Measurement result for calibrating O2

N/A
O2 calibration
Setpoint value O2
Actual value on the controller
Reference value measured in the working chamber (actual
value +/-0.5 % O2)

--- % O2
--- % O2
--- % O2

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 52 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Testing the thermoshelves

N/A
(only VO appliances)
The distribution of temperature for each installed thermoshelf via 5 measuring
points (see Attachment I for arrangement of measuring sensors on thermoshelves).
Apart from determining the lowest and highest temperature on the thermoshelves
(distribution of temperature), the maximum fluctuations in temperature must be
recorded.
To do this, the lowest and chronologically following highest temperature at one
measurement point (middle of respective thermoshelf) is recorded and documented
after a state of inertia has been reached.

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 53 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Measurement results for area temperature distribution (VO)


N/A
Test conditions to record temperature distribution:
- for a precise description of positions of the sensors see Attachment
Attachment No. I:
- Close doors, switch on oven, set temperature to the desired calibration
setpoint
- Wait for state of inertia to be reached at least 120 min. after setpoint
temperature has been reached

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 54 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

The temperature values determined must be entered in


in the table below.
Set setpoint temperature
Thermoshelf (serial No.)
Actual temperature displayed

---C
N
--- C

N ----- C

N ----- C

N
--- C

Measurement position
front left T1
rear left T2
rear right T3
front right T4

--- C
--- C
--- C
--- C

--- C
--- C
--- C
--- C

--- C
--- C
--- C
--- C

--- C
--- C
--- C
--- C

Thermoshelf, middle T5

--- C

--- C

--- C

--- C

Maximum temperature
difference
- Aluminium sheet max. +/- 2%
- stainless steel sheet max. +/5%
(the actual temperature
displayed)

--- K

--- K

--- K

--- K

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 55 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Measurement results on the temperature fluctuation (VO)


N/A
Test conditions to record temperature fluctuation:
- Positioning of thermoshelves in the vacuum oven
- Positioning of the reference sensor at position T5, middle
- Close doors, switch on oven, set temperature to the desired
calibration setpoint
- Wait for state of inertia to be reached at least 120 min. after setpoint
temperature has been reached
- Documentation of the highest and lowest value
The temperature values determined must be entered in the table below.
below.
Thermoshelf
Set setpoint temperature
Actual temperature displayed

N
--- C
--- C

N
--- C
--- C

N
--- C
--- C

N
--- C
--- C

Temperature minimum
Recorded at sensor position

--- C
--- C
Thermoshelf, middle T5

--- C

--- C

Temperature maximum
Recorded at sensor position

--- C
--- C
Thermoshelf, middle T5

--- C

--- C

--- K

--- K

Temperature fluctuation
max. +/- 0.3 K

--- K

--- K

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 56 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Calibration of the vacuum controller

N/A
Test conditions:
- Positioning of the thermo shelves in the vacuum oven (see Chapter 0 Function
Function
test, calibration - vacuum oven)
- Connection of vacuum calibration system to the IN1 connection on the rear of
the appliance
- Close doors and switch on oven with pump, set vacuum to the desired
calibration setpoint
- Wait until the vacuum value has stabilised (at least 60 minutes)
- Set vacuum value to the maximum (1100 mbar) a notice to connect the
calibration system with the vacuum chamber (solenoid valve for IN1 inlet is
opened)
- Wait until the display of the vacuum controller and of the calibration system has
stabilised (at least 5 minutes)
The final steps must be repeated in each case if more than one vacuum value is to
be tested!
Tested at a room temperature of 22C +/- 2K
Setpoint for vacuum
Actual value vacuum controller
Measured reference value (permissible: actual value +/- 5
mbar)
N/A
Setpoint for vacuum
Actual value vacuum controller
Measured reference value (permissible: actual value +/- 5
mbar)
N/A
Setpoint for vacuum

--- C
--- mbar
--- mbar
--- mbar

--- mbar
--- mbar
--- mbar

--- mbar

Comments: _________________________________________________________________
Actual value vacuum controller
--- mbar
Measured reference value (permissible: actual value +/- 5
--- mbar
mbar)

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 57 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 58 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Measurement of light intensity

N/A
To measure the light intensity, the light modules must be switched on for at least
30 minutes

ICP
Measured reference value

N/A

HPP108L
Cold white N/A
Measured reference value

N/A

--- Lux

--- Lux

Cold white / warm white N/A


Measured reference value

--- Lux

ICH256L (Standard light cartridge)


Cold white N/A
Measured reference value

N/A
--- Lux

UV light N/A
Measured reference value

--- W/m

Comments: ______________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 59 of 76

Appliance Qualification
OQ
Date: 24.05.2012

Version: Version: 2012-05/1

Release of the tested appliance for use

The original copies of the recorded temperature values entered in the qualification
documents (recording strips, PC printouts, etc.) must be enclosed with the
qualification documents.
The measurement of temperature distribution/temperature fluctuation was made
by
Name

Company

Date

Signature

The release of the appliance for the intended use was done by:
Name

Company

Date

Signature

Comments: ______________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 60 of 76

Appliance Qualification
PQ
Date: 30.08.2012

Version: Version: 2012-05/1

PQ
Testing temperature distribution
distribution in compliance with DIN 12880: 20072007-05

N/A
The distribution of the temperature in the working chamber is recorded via 9
measurement points for an interior volume below 50 litres, above 50 litres is tested
with 27 measurement points.
Apart from determining the lowest and highest temperature inside the chamber
(distribution of temperature), the maximum fluctuations in temperature must be
recorded.
To do this, the lowest and chronologically following highest temperature at same
measurement point (middle of chamber) is recorded and documented after a state
of inertia has been reached.
Measurements performed by the company MEMMERT are documented in the
attachment. (see Attachment No. III)
OQ measurements performed at the customer must be entered in the table below.
The temperature sensors for testing temperature distribution and temperature
fluctuation, must be placed in accordance with DIN 12880: 2007-05 (see
Attachment No. II).

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 61 of 76

Appliance Qualification
PQ
Date: 30.08.2012

Version: Version: 2012-05/1

measurement
ement
Measurement results for temperature distribution / humidity measur
Measurement 1
N/A
Set setpoint temperature
Displayed actual temperature (wait for state of inertia - at least
120 min. after setpoint temperature has been reached
Set humidity setpoint N/A
Displayed humidity actual value (a humidity setpoint +/- 3% rh)
N/A
Level 0
Sensor # 1 --- C
Sensor # 2 --- C
Sensor # 3 --- C
Sensor # 4 --- C
Sensor # 5 --- C
Sensor # 6 --- C
Sensor # 7 --- C
Sensor # 8 --- C
Sensor # 9 --- C

Level 1
Sensor # 11 --- C
Sensor # 12 --- C
Sensor # 13 --- C
Sensor # 14 --- C
Sensor # 15 --- C
Sensor # 16 --- C
Sensor # 17 --- C
Sensor # 18 --- C
Sensor # 19 --- C

--- C
--- C
--- % rh
--- % rh
Level 2
Sensor # 21 --- C
Sensor # 22 --- C
Sensor # 23 --- C
Sensor # 24 --- C
Sensor # 25 --- C
Sensor # 26 --- C
Sensor # 27 --- C
Sensor # 28 --- C
Sensor # 29 --- C

Maximum temperature difference (spatially)


permissible: <70 = +/- 0.5K; 70 - 180 = +/- 1.5K; >180 = +/3K
Humidity sensor middle of chamber (permissible: actual value +/3% rh)
N/A

--- K
--- % rh

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 62 of 76

Appliance Qualification
PQ
Date: 30.08.2012

Version: Version: 2012-05/1

Measurement 2
N/A
Set setpoint temperature
Displayed actual temperature (wait for state of inertia - at least
120 min. after setpoint temperature has been reached
Set humidity setpoint N/A
Displayed humidity actual value (a humidity setpoint +/- 3% rh)
N/A
Level 0
Sensor # 1 --- C
Sensor # 2 --- C
Sensor # 3 --- C
Sensor # 4 --- C
Sensor # 5 --- C
Sensor # 6 --- C
Sensor # 7 --- C
Sensor # 8 --- C
Sensor # 9 --- C

Level 1
Sensor # 11 --- C
Sensor # 12 --- C
Sensor # 13 --- C
Sensor # 14 --- C
Sensor # 15 --- C
Sensor # 16 --- C
Sensor # 17 --- C
Sensor # 18 --- C
Sensor # 19 --- C

--- C
--- C
--- % rh
--- % rh
Level 2
Sensor # 21 --- C
Sensor # 22 --- C
Sensor # 23 --- C
Sensor # 24 --- C
Sensor # 25 --- C
Sensor # 26 --- C
Sensor # 27 --- C
Sensor # 28 --- C
Sensor # 29 --- C

Maximum temperature difference (spatially)


permissible: <70 = +/- 0.5K; 70 - 180 = +/- 1.5K; >180 = +/3K
Humidity sensor middle of chamber (permissible: actual value +/3% rh)
N/A

--- K
--- % rh

Comments: ______________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 63 of 76

Appliance Qualification
PQ
Date: 30.08.2012

Version: Version: 2012-05/1

Measurement 3
N/A
Set setpoint temperature
Displayed actual temperature (wait for state of inertia - at least
120 min. after setpoint temperature has been reached
Set humidity setpoint N/A
Displayed humidity actual value (a humidity setpoint +/- 3% rh)
N/A
Level 0
Sensor # 1 --- C
Sensor # 2 --- C
Sensor # 3 --- C
Sensor # 4 --- C
Sensor # 5 --- C
Sensor # 6 --- C
Sensor # 7 --- C
Sensor # 8 --- C
Sensor # 9 --- C

Level 1
Sensor # 11 --- C
Sensor # 12 --- C
Sensor # 13 --- C
Sensor # 14 --- C
Sensor # 15 --- C
Sensor # 16 --- C
Sensor # 17 --- C
Sensor # 18 --- C
Sensor # 19 --- C

--- C
--- C
--- % rh
--- % rh
Level 2
Sensor # 21 --- C
Sensor # 22 --- C
Sensor # 23 --- C
Sensor # 24 --- C
Sensor # 25 --- C
Sensor # 26 --- C
Sensor # 27 --- C
Sensor # 28 --- C
Sensor # 29 --- C

Maximum temperature difference (spatially)


permissible: <70 = +/- 0.5K; 70 - 180 = +/- 1.5K; >180 = +/3K
Humidity sensor middle of chamber (permissible: actual value +/3% rh)
N/A

--- K
--- % rh

Comments: ______________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 64 of 76

Appliance Qualification
PQ
Date: 30.08.2012

Version: Version: 2012-05/1

Measurement 4
N/A
Set setpoint temperature
Displayed actual temperature (wait for state of inertia - at least
120 min. after setpoint temperature has been reached
Set humidity setpoint N/A
Displayed humidity actual value (a humidity setpoint +/- 3% rh)
N/A
Level 0
Sensor # 1 --- C
Sensor # 2 --- C
Sensor # 3 --- C
Sensor # 4 --- C
Sensor # 5 --- C
Sensor # 6 --- C
Sensor # 7 --- C
Sensor # 8 --- C
Sensor # 9 --- C

Level 1
Sensor # 11 --- C
Sensor # 12 --- C
Sensor # 13 --- C
Sensor # 14 --- C
Sensor # 15 --- C
Sensor # 16 --- C
Sensor # 17 --- C
Sensor # 18 --- C
Sensor # 19 --- C

--- C
--- C
--- % rh
--- % rh
Level 2
Sensor # 21 --- C
Sensor # 22 --- C
Sensor # 23 --- C
Sensor # 24 --- C
Sensor # 25 --- C
Sensor # 26 --- C
Sensor # 27 --- C
Sensor # 28 --- C
Sensor # 29 --- C

Maximum temperature difference (spatially)


permissible: <70 = +/- 0.5K; 70 - 180 = +/- 1.5K; >180 = +/3K
Humidity sensor middle of chamber (permissible: actual value +/3% rh)
N/A

--- K
--- % rh

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 65 of 76

Appliance Qualification
PQ
Date: 30.08.2012

Version: Version: 2012-05/1

Measurement results on the temperature fluctuation


N/A
Recorded temperature at sensor position 11 / level 1 (middle of chamber)
N/A
Measuremen Measuremen Measuremen Measuremen
Measurement
t1
t2
t3
t4
Temperature minimum
--- C
--- C
--- C
--- C
Temperature maximum
--- C
--- C
--- C
--- C
Temperature fluctuation
max. +/- 0.5 K

--- K

--- K

Recorded humidity at sensor position (middle of chamber)


Humidity minimum
--- % rh
--- % rh
Humidity maximum
--- % rh
--- % rh
Humidity fluctuation
max. +/- 3% rh

--- % rh

--- % rh

--- K

--- K

--- % rh
--- % rh

N/A
--- % rh
--- % rh

--- % rh

--- % rh

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 66 of 76

Appliance Qualification
PQ
Date: 30.08.2012

Version: Version: 2012-05/1

Measurement result for calibrating humidity

N/A
Humidity calibration
Humidity setpoint
Actual value on the controller
Reference value measured in the working chamber (actual
value +/-3% rh)

--- % rh
--- % rh
--- % rh

Measurement result for calibrating CO2

N/A
CO2 calibration
Setpoint value CO2
Actual value on the controller
Reference value measured in the working chamber (actual
value +/-0.7 % CO2)

--- % CO2
--- % CO2
--- % CO2

Measurement result for calibrating O2

N/A
O2 calibration
Setpoint value O2
Actual value on the controller
Reference value measured in the working chamber (actual
value +/-0.5 % O2)

--- % O2
--- % O2
--- % O2

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 67 of 76

Appliance Qualification
PQ
Date: 30.08.2012

Version: Version: 2012-05/1

Testing the thermoshelves

N/A
(only VO appliances)
The distribution of temperature for each installed thermoshelf via 5 measuring
points (see Attachment I for arrangement of measuring sensors on thermoshelves).
Apart from determining the lowest and highest temperature on the thermoshelves
(distribution of temperature), the maximum fluctuations in temperature must be
recorded.
To do this, the lowest and chronologically following highest temperature at one
measurement point (middle of respective thermoshelf) is recorded and documented
after a state of inertia has been reached.

Comments: ______________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 68 of 76

Appliance Qualification
PQ
Date: 30.08.2012

Version: Version: 2012-05/1

Measurement results for area temperature distribution (VO)


N/A
Test conditions to record temperature distribution:
- for a precise description of positions of the sensors see Attachment No. I:
- Close doors, switch on oven, set temperature to the desired calibration
setpoint
- Wait for state of inertia to be reached at least 120 min. after setpoint
temperature has been reached
reached
The temperature values determined must be entered in the table below.
Set setpoint temperature
Thermoshelf (serial No.)
Actual temperature displayed

---C
N ----- C

N ----- C

N ----- C

N
--- C

Measurement position
front left T1
rear left T2
rear right T3
front right T4

--- C
--- C
--- C
--- C

--- C
--- C
--- C
--- C

--- C
--- C
--- C
--- C

--- C
--- C
--- C
--- C

Thermoshelf, middle T5

--- C

--- C

--- C

--- C

Maximum temperature
difference
- Aluminium sheet max. +/- 2%
- stainless steel sheet max. +/5%
(the actual temperature
displayed)

--- K

--- K

--- K

--- K

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 69 of 76

Appliance Qualification
PQ
Date: 30.08.2012

Version: Version: 2012-05/1

Measurement results on the temperature fluctuation (VO)


N/A
Test conditions to record temperature fluctuation:
- Positioning of thermoshelves in the vacuum oven
- Positioning of
of the reference sensor at position T5, middle
- Close doors, switch on oven, set temperature to the desired
calibration setpoint
- Wait for state of inertia to be reached at least 120 min. after setpoint
temperature has been reached
- Documentation of the highest and lowest value
The temperature values determined must be entered in the table below.
Thermoshelf
Set setpoint temperature
Actual temperature displayed

N
--- C
--- C

N
--- C
--- C

N
--- C
--- C

N
--- C
--- C

Temperature minimum
Recorded at sensor position

--- C
--- C
Thermoshelf, middle T5

--- C

--- C

Temperature maximum
Recorded at sensor position

--- C
--- C
Thermoshelf, middle T5

--- C

--- C

--- K

--- K

Temperature fluctuation
max. +/- 0.3 K

--- K

--- K

Comments: __________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 70 of 76

Appliance Qualification
PQ
Date: 30.08.2012

Version: Version: 2012-05/1

Calibration of the vacuum controller

N/A
Test conditions:
- Positioning of the thermo shelves in the vacuum oven (see Chapter 0 Function
Function
test, calibration - vacuum
vacuum oven)
- Connection of vacuum calibration system to the IN1 connection on the rear of
the appliance
- Close doors and switch on oven with pump, set vacuum to the desired
calibration setpoint
- Wait until the vacuum value has stabilised (at least 60 minutes)
- Set vacuum value to the maximum (1100 mbar) a notice to connect the
calibration system with the vacuum chamber (solenoid valve for IN1 inlet is
opened)
- Wait until the display of the vacuum controller and of the calibration system has
stabilised (at least 5 minutes)
The final steps must be repeated in each case if more than one vacuum value is to
be tested!
Tested at a room temperature of 22C +/- 2K
Setpoint for vacuum
Actual value vacuum controller
Measured reference value (permissible: actual value +/- 5
mbar)
N/A
Setpoint for vacuum
Actual value vacuum controller
Measured reference value (permissible: actual value +/- 5
mbar)

--- C
--- mbar
--- mbar
--- mbar

--- mbar
--- mbar
--- mbar

N/A
Comments: _________________________________________________________________
Setpoint for vacuum
--- mbar
Actual value vacuum controller
--- mbar
Measured reference value (permissible: actual value +/- 5
--- mbar
mbar)

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 71 of 76

Appliance Qualification
PQ
Date: 30.08.2012

Version: Version: 2012-05/1

_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 72 of 76

Appliance Qualification
PQ
Date: 30.08.2012

Version: Version: 2012-05/1

Measurement of light intensity

N/A
To measure the light intensity, the light modules must be switched on for at least
30 minutes

ICP
Measured reference value

N/A

HPP108L
Cold white N/A
Measured reference value

N/A

--- Lux

--- Lux

Cold white / warm white N/A


Measured reference value

--- Lux

ICH256L (Standard light cartridge)


Cold white N/A
Measured reference value

N/A
--- Lux

UV light N/A
Measured reference value

--- W/m

Comments: ______________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 73 of 76

Appliance Qualification
PQ
Date: 30.08.2012

Version: Version: 2012-05/1

Release of the tested appliance for use

The original copies of the recorded temperature values entered in the qualification
documents (recording strips, PC printouts, etc.) must be enclosed with the
qualification documents.
The measurement of temperature distribution/temperature fluctuation was made
by
Name

Company

Date

Signature

The release of the appliance for the intended use was done by:
Name

Company

Date

Signature

Comments: _________________________________________________________________
_____________________________________________________________________________
_____________________________________________________________________________

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 74 of 76

Appliance Qualification
Attachment
Date: 30.08.2012

Version: Version: 2012-05/1

Attachment

Exists:
I. VO cable allocation 5 points distribution measurement

- Document N D09032VO 200


D09032-A.pdf
- Document N D09033VO 400
D09033-B.pdf
- Document N D09034VO 500
D09034-A.pdf
II. 9/27 points distribution measurement acc. to DIN 12880: 2007-05

-Document N D30540.pdf

27 point measurement

-Document N D30541.pdf

9 point measurement

III. Test records

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 75 of 76

Appliance Qualification
Attachment
Date: 30.08.2012

Version: Version: 2012-05/1

Notes

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng

Page 76 of 76

Appliance Qualification
Notes
Date: 24.05.2012

Version: Version: 2012-05/1

Test performed by

Date

Signature

Reviewed by

Date

Signature

by MEMMERT GmbH + Co. KG

Ref.: D24374_IQ_OQ_PQ_eng