Professional Documents
Culture Documents
Motivation
Saturated CT cores and their behavior during fault
conditions have been and still are an important
topic for protection engineers, in particular when
fault currents with transient components occur.
Many papers have been written and presented
about this. Typically, the CT cores contain some
residual magnetism when the current flow is
interrupted. Especially at reclosing, this residual
magnetism may cause undesired effects.
In this context, the question "How will the
magnetization in the CT develop when power is
restored and the CT is exposed to normal
operating currents?" is frequently raised.
Surprisingly, the answer is often based on opinion,
it is even said this may be a "matter of philosophy".
Statements like "some say the saturation level
remains there forever, some say it will go away
over time" can be heard. Much of the said is based
on assumptions, models and simulations.
But for engineers, there must be a better way than
relying "on philosophy". The recognized method is
to look for plain facts and align them with proven
theories. For this paper, systematic measurements
have been made to put some light on this matter.
Practical Implications
Measuring residual flux can be a cumbersome
procedure [6]. In the past, it was hardly feasible to
perform a large number of such measurements in
reasonable time.
Also, for establishing a relationship between a
measured residual flux and the conditions that
caused this residual flux, the full previous history
(the anamnesis) of the CT must be known. Even if
the fault records were available for some arbitrary
cases of saturated CTs, this is not a systematic
approach for reliably determining a correlation of
effects.
Another problem is that the established methods
for determining the residual flux result also in a
destruction of the residual flux [3]. It is not possible
to resume the test at the most recent point.
Thus, every single measurement has to start from
the de-magnetized state, first bringing the CT into
a defined state of magnetization (applying a
defined residual flux). Then, some other condition
which may alter the residual flux may be applied.
Then, this residual flux is to be measured.
Establishing a
Defined Magnetization
The first step for performing measurements with
residual magnetism is to apply a defined initial
magnetization to the CT core. It is assumed that
we start with a demagnetized core. In our case,
this condition is fulfilled without further efforts,
since the measurement device for the residual flux
leaves the core demagnetized after the
measurement.
Demagnetization
The goal of this work is to investigate how the
magnetization of the saturated core is affected by
normal operating currents. Therefore, conditions as
under nominal currents shall be applied.
Demagnetization
With Injected Current
The current source injects a sinusoidal current into
the secondary winding. As long the compliance
voltage of the current source is not exceeded, the
voltage adapts to the load impedance and the
current is kept constant. As the secondary
impedances are almost negligible compared to the
impedance of the main inductance, the voltage
measured at the terminals is almost identical to the
voltage at the main inductance and according to
hysteresis curve.
It must be noted that this injection of a defined
current is different from the procedures described
in most standards as IEC 60044-1, where a
sinusoidal voltage is applied and a current with
potentially non-sinusoidal waveform is obtained.
This current
was now applied to the CT for a
few seconds and the remaining residual
magnetism was measured. It was significantly
reduced from the value established through the
magnetization. This was a first finding, but further
measurements with longer exposure time of the
nominal current showed the same result, the
residual magnetism did not drop further.
Consequently, the exposure time was reduced until
a change of the measured residual magnetism
could be observed. It turned out that the time
interval to be considered is less than a second or
only a few cycles. The result is shown in the
following figure.
Frequency Dependency
of the Effect
For the measurements described above it was
implicitly assumed and therefore not explicitly
mentioned that the current applied for changing the
magnetization of the core had nominal frequency,
in this case 50 Hz.
But it turns out that the reduction of the
magnetization is essentially more significant if the
frequency of this current is lowered. Fig. 7 shows
how the residual magnetism develops for different
frequencies of the applied current.
Conclusions
Literature
Dedication
[1]
[2]
[3]
[4]
[5]
[6]
[7]
[8]
[9]
Americas
OMICRON electronics Corp. USA
12 Greenway Plaza, Suite 1510
Houston, TX 77046, USA
Phone: +1 713 830-4660
+1 800-OMICRON
Fax:
+1 713 830-4661
info@omicronusa.com
OMICRON
Asia-Pacific
OMICRON electronics Asia Limited
Suite 2006, 20/F, Tower 2
The Gateway, Harbour City
Kowloon, Hong Kong S.A.R.
Phone: +852 3767 5500
Fax:
+852 3767 5400
info@asia.omicron.at
www.omicron.at www.omicronusa.com