General
1
2
3
4
C-1
AL-1
A-1
AD-1
Rev Date
Drw
TTH
Proposal
Chk
Proj. phase
Description
Cover sheet
Table of contents
ID-16-00108
Sheet Subject
Revision history
H
G
F
E
D
C
B
A
Semarang , Indonesia
Subject
Achieva 1.5T
RSUD Kota Semarang
Type
Project
Sheet
Drawing
Important notes
Philips assumes no liability nor offers any warranty for the fitness
or adequacy of the premises or the utilities available at the premises
in which the equipment is to be installed, used or stored.
Philips contact
Isnadi Syukri, +628111001819, isnadi.syukri@philips.com
Philips Healthcare
Customer ID
Technical ID
A3
Quote No
Order No
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.
Table of contents
mm
C-1
Sheet
1/4
Maximum Cable lengths between the Cable in- and Outlets of the Cabinets
From:
(**)
To:
DACC (*)
LCC (*)
Gradient Amplifier (**)
Gradient Amplifier (**)
LCC (*)
LCC (*)
Gradient Amplifier (**)
LCC (*)
All other components
Meter:
7
7
7
7
7
7
10
13
10
Technical room:
Temperature 15C-24C
Maximum temperature change 5C per 10min
Relative humidity 30% - 70%, no condensation
Required Air Conditioning Capacity: 7 kW
- Standby heat dissipation: 2 kW
- Heat dissipation to Water: 7 - 40kW
- Min. flow 4200L/h.
- Pressure drop (with 100% water) 1.8bar.
Examination area
A
1.1
1.2
Patient table
4630
165
1.3
80 1000*
1.4
1.5
1.6
1.7
300
1.8
51
1.9
1.10 Injector
63
76
65
Control area
A
2.1
Operating station
2.2
2.3
2.4
100
2.5
4.2
100*
2.6
11
Technical area
A
3.1
135
200*
3.2
300 1000*
3.3
467
200*
3.4
265 7000*
3.5
3.6
Mains switch
3.7
General
B
4.1
RF door
4.2
RF window
Responsibilities
A Delivered and installed by philips
B Delivered and installed by customer/contractor
C (Pre) Delivered by Philips, installed by customer
D (Pre) Delivered by customer and installed by Philips
E Existing
Control room:
Temperature for human comfort 18C-24C
Temperature required for X-ray films15C-30C
Temperature MR equipment 10C-35C
Maximum temperature change 5C per 10min
Relative humidity 30% - 70%, no condensation
Required Air Conditioning Capacity: 0.5 kW
* Heat load indicated is peak dissipation for each cabinet measured individually. Peak room
heat dissipation as indicated in Site Requirements column will be less than the sum of each
individual cabinet in a given room due to the fact that not all cabinets will run peak heat
loads at the same time.
The equipment proposed in this room layout produces a strong magnetic field. It is the
customer's responsibility to comply with the local regulation. Normally this field has to be
contained in a controlled area that is only accessible by authorised personnel and screened
patients. Consult the Philips contact for details and support.
Drawing
Project notes MR
The feasibility of this project is based on site information (like drawings) provided to
Philips. Philips cannot assume any liability for the accuracy of this information and the
consequences.
A3
mm
AL-1
Sheet
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
Semarang , Indonesia
(**)
For 230V Y-configuration ( = fase to neutral ), type AM or GL according DIN 43620, VDE
0636 part 1 and 2a
Heat Noise
[W] [dB(A)]
Quote No
Order No
Technical ID
For the square area of 3000 mm x 3000 mm symmetrically around magnet isocentre,
ferro-magnetic reinforcement must be:
- NO less than 50 mm below finished floor.
- NO greater than 25 kg/m average concentration between 50 mm and 250 mm
below the finished floor. Reinforcements below 250 mm can be ignored.
- Evenly distributed.
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.
(*)
If an UPS for the complete MR system is required, provide the UPS supplier that not only
the required kVA but also the required above-mentioned peak current. This can result in a
higher (kVA) specification of the UPS needed.
Mass
[kg]
Achieva 1.5T
RSUD Kota Semarang
Philips strongly recommends the RF door opens outwards to aid opening in the
event of a possible quench. If the door opens inwards an overpressure feed through
with minimum dimensions of 600 mm x 600 mm must be installed.
Description
Site requirements
Equipment legend
Passive shielding or fencing may be required to contain the 0.5mt field line within the
back and side wall of the examination room.
Resp No
03 Aug 2016
NDI
Passive shielding or fencing may be required to contain the 0.5mt field line within the
floor and ceiling of the examination room.
ID-16-00108
Dimensions for building works in the examination room (like RF-cage construction
and wall openings) are for indication and bidding only! Refer to the drawing of the RF
cage vendor for actual construction dimensions. For ranges refer to AD-1 sheet.
Mains power
Supply configuration:
Star, 3 phase+neutral+protective earth (PE).
Delta is allowed for the 480V version.
MDU version to be ordered via order questionnaire,
Nominal Line Voltage:
230/400V +/-10% 50/60Hz +/-1Hz
480V +/-10% 60Hz +/-1Hz
Mains impedance (at nominal voltage):
< 200 mOhm
Sheet Subject
Site requirements
Project
Drawing alerts:
2/4
5m
5500
Customer Acceptance
2500
AHU CHILLER
1m
Designation: ______________________________________________
Date:
______________________________________________
Signature:
______________________________________________
1850
3.6
3.3
3.4
3.7
3.5
Technical Room
3.1
10mT
3.2
1.3
1.6
1.5
UPS Room
1.10
1.4
6500
R.
1.1
GANTI
3mT
MAIN
ENTRANCE
1.8
1.2
MRI SUITE
0.3mT
Project
0.5mT
1mT
1.7
4.2
Semarang , Indonesia
Examination Room
Achieva 1.5T
RSUD Kota Semarang
2500
2.2
0.1mT
4.1
2000
2.2
Site layout
Sheet Subject
RECOVERY AREA
Control Room
2.5 2.3 2.1
2.4
A3
Quote No
Order No
Technical ID
03 Aug 2016
NDI
ID-16-00108
Important Note:
Drawing
2.6
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.
2300
1.9
1 : 50
mm
A-1
Sheet
3/4
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
A
0.3mT
0.1mT
A3
Sheet
Quote No
Order No
Technical ID
Suspended ceiling
Semarang , Indonesia
Achieva 1.5T
RSUD Kota Semarang
0.1mT
Project
(4000)
0.5mT
3000
0.3mT
3000
2520
2800
(4000)
1m
03 Aug 2016
NDI
B
4
2520
2800
1000
Sheet Subject
1000
1490
ID-16-00108
1490
2395
Drawing
2395
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.
0
5m
Suspended ceiling
3mT
1mT
2
X
3mT
1mT
1 : 50
mm
AD-1
4/4