Professional Documents
Culture Documents
Basic Concepts of
Measurement Methods
What Is Measurements?
Measurement:
an estimation of a physical (chemical or biological)
variable by a measurement device.
Environment parameter (P, T, rh etc.)
Instrument
Data
operator
Measurement
Method
Physical
parameter
Measured
stimulus
variable
(Input signal
or
measurand)
Sensor or
Transducer
State
Signal
Conditioning
State
-amplifier
-filter
Output
State
System
output
-meter
-oscilloscope
-recorder
A bulb thermometer
Thermal
Electrical
Magnetic
Radiant
Chemical
Reference
value, Td
Error
signal
Room
Temperatrue, Ta
Heater
Room
Td - Ta
Ta
Temp.
sensor
System
System
input
Measurement
System
Self-generating
output
input
External
power
Measurement
System
Modulating
output
A spring balance
Null-type Method:
a null-type device attempts to maintain
deflection at zero by suitable application of a
known effect opposing the generated by the
measured quantity. (a null detector and a
means of restoring balancing are necessary).
An equal arm balance
VS
Vx
c
V
Vx
+ Im
Galvanometer
a
Measuring unknown voltage using a
voltmeter
Vx
Unknown
voltage
Potentiometer voltage
measurement
Experimental Variables
Independent variable
a variable that can be changed independently of other variables.
Dependent variable
a variable that is affected by one or more other variables.
Controlled variable
a variable that can be held at constant value during the
measurement process.
Extraneous variable
a variable that are not or can not be controlled during
measurement but can affected the value of the measured
variable.
Extraneous Variables
Interference
An undesirable deterministic trends on the measured value
because of extraneous variables.
Noise
a random variation of the value of the measured signal as a
consequence of the variation of the extraneous variables.
Calibration
Calibration:
A test in which known values of the input are applied to a
measurement system (or sensor) for the purpose of observing the
system (or sensor) output.
Static calibration:
A calibration procedure in which the values of the
variable involved remain constant (do not change with time).
Dynamic calibration:
When the variables of interest are time dependent and
time-based information is need. The dynamic calibration
determines the relationship between an input of known dynamic
behavior and the measurement system output.
Static Characteristics
Static Sensitivity: Incremental ratio of the output signal (y) to the desired input
signal (x).
S=
y
x
Static Characteristics
Measurand range, operating range, full-scale range, span: the range of input
variable (xmax xmin) that produces a meaningful output.
ri = (xmax xmin)
Full scale output (FSO): Difference between the end points of the output. The
upper limit of output over the measurand range is called the full scale (FS)
ro = (ymax ymin)
Offset: The output of a sensor, under room temperature condition unless
otherwise specified, with zero measurand applied.
Static Characteristics
Accuracy: the difference between the true (expected) and measured values
from the measurement system or sensor. Normally, it is quoted in as a
fractional of the full scale output.
Percentage of reading
a (%) =
( ym yt )
100
yt
( ym yt )
100
yFSO
Absolute error
= indicated value- true value
Static Characteristics
Precision: The ability of the system to indicate a particular value upon repeated
but independent applications of a specific value of input. The precision error is a
measure of the random variation found during repeated measurements.
measured value
measured average
bias error
Static Characteristics
Example: Three load cells are tested for repeatability. The same 50-kg
weight is placed on each load cell 10 times. The resulting data are given in the
following table. Discuss the repeatability and accuracy of each sensor. If the
expected output of these load cells is 10 mV.
Trail
no.
1
2
3
4
5
6
7
8
9
10
Maximum
Average
Minimum
10.02
10.96
11.20
9.39
10.50
10.94
9.02
9.47
10.08
9.32
11.20
10.09
9.02
11.50
11.53
11.52
11.47
11.42
11.51
11.58
11.50
11.43
11.48
11.58
11.49
11.42
10.00
10.03
10.02
9.93
9.92
10.01
10.08
10.00
9.97
9.98
10.08
9.99
9.92
Static Characteristics
Load cell A
Load cell B
not repeatable
Load cell C
Accurate and repeatable
11.6
11.6
Max.
11.2
11.0
10.6
Output (mV)
Output (mV)
10.8
10.4
Ave.
10.0
11.6
11.4
Min.
11.2
10.2
11.4
11.2
11.0
11.0
10.8
10.8
10.6
Output (mV)
11.4
10.4
10.2
10.0
10.6
10.4
9.8
9.8
9.6
9.6
9.6
9.4
9.4
9.2
9.2
9.2
Min
9.0
x
0
5
Trial no.
9.0
8
10
10.0
9.8
9.4
Max.
10.2
10
Min.
9.0
0
5
Trial no.
10
Trial no.
A transducer or sensor that is repeatable but not overly accurate may still be quite usable in
a measurement or control application. As long as the transducer or sensor is repeatable, you
will get consistent results. We may correct this inaccuracy by the recalibration this transducer
or sensor.
Static Characteristics
Resolution: the smallest increment in the value of the measurand that
results in a detectable increment in the output. It is expressed in the
percentage of the measurand range
Resolution (%) =
x
100
xmax xmin
Static Characteristics
Hysteresis: Difference in the output of a sensor or instrument for a given input
value x, when x is increased and decreased or vice versa. (expressed in % of
FSO) (indication of reproducibility)
upscale
maximum
hysteresis
output (%FSO)
100
downscale
0
0
100
measurand (% range)
Static Characteristics
Linearity: (also called Nonlinearity) A measure of deviation from linear of a
sensor or instrument, which is usually described in terms of the percentage of
FSO.
(1) best-fit straight line
(2) terminal-based straight line
(3) independent straight line
yL(x) = a0 + a1x
A predicted output based on linear relation:
output (%FSO)
best-fit line
terminal-base line
0
100
measurand (% range)
maximum
nonlinearity
100
output (%FSO)
maximum
nonlinearity
100
100
measurand (% range)
maximum
100 nonlinearity
output (%FSO)
Linearity error:
independent line
100
measurand (% range)
Static Characteristics
Operating conditions: Ambient conditions may have profound effects on
the sensor or instrument operation. These include temperature, acceleration,
vibration, shock, pressure, moisture, corrosive materials, and electromagnetic
field.
Temperature zero drift: the change in the output level of a sensor or
instrument due to temperature variation when the input is set to zero.
100
measurand (% range)
100
measurand (% range)
Nominal
desired
temp.
output (%FSO)
output (%FSO)
output (%FSO)
Temperature
zero error
0
Temperature
change
100
100
Temperature
sensitivity error
100
Total error
Zero drift
100
measurand (% range)
Static Characteristics
Overall Performance: An estimate of the overall sensor error is made based
on all known errors. An estimate is computed from
The worst case approach:
ec = e1 + e2 + e3 + L + en
The root of sum square approach:
0-1000 cm H2O
15 V dc
0-5 V
0-50oC nominal at 25oC
0.5%FSO
Less than 0.15%FSO
0.25%of reading
0.02%/oC of reading from 25oC
0.02%/oC FSO from 25oC
500 cm H2O
Pressure
Linearity error eL
25 mV
5 cm H2O
Hysteresis error eh
7.5 mV
1.5 cm H2O
Sensitivity error eS
6.25 mV
1.25 cm H2O
- 3.5 mV
-0.7 cm H2O
- 7.0 mV
-1.4 cm H2O
9.9 cm H2O
5.6 cm H2O
Basic specifications
Input range
Output range
Offset
Sensitivity
Performance specifications
Input
Input
Sensor
Accuracy
Resolution
Repeatability
Hysteresis
Linearity
environmental parameter
etc.
Output
Static Characteristics
Example: A load cell is a sensor used to measure weight. A calibration record table is
given below. Determine (a) accuracy, (b) hysteresis and (c) linearity of the sensor. If we
assume that the true or expected output has a linear relationship with the input. In
addition, the expected output are 0 mV at 0 kg load and 20 mV at 50 kg load.
Output (mV)
Increasing
0.08
0.45
1.02
1.71
2.55
3.43
4.48
5.50
6.53
7.64
8.70
9.85
11.01
12.40
13.32
14.35
15.40
16.48
17.66
18.90
19.93
Decreasing
0.06
0.88
2.04
3.10
4.18
5.13
6.04
7.02
8.06
9.35
10.52
11.80
12.94
13.86
14.82
15.71
16.84
17.92
18.70
19.51
20.02
20
15
Output (mV)
Load (kg)
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
Decreasing
10
Increasing
0
0
20
40
60
Load (kg)
80
100
Static Characteristics
(a) Accuracy
Desired output = 0.2mV/kg x load
20
True Output
Actual
(mV)
Output (mV) Error (mV)
0.08
0
0.08
-0.55
1
0.45
-0.98
2
1.02
-1.29
3
1.71
-1.45
4
2.55
-1.57
5
3.43
-1.52
6
4.48
-1.50
7
5.5
0.30
0.10
0.20
0.65
0.90
0.50
0.20
-0.60
0.30
0.06
0.02
0.04
0.13
0.18
0.10
0.04
-0.12
0.06
8.06
7.02
6.04
5.13
4.18
3.1
2.04
0.88
0.06
8
7
6
5
4
3
2
1
0
40
35
30
25
20
15
10
5
0
%FSO
%reading
0.40
a
-2.75
-55.00
-4.90
-49.00
-6.45
-43.00
-7.25
-36.25
-7.85
-31.40
-7.60
-25.33
-7.50
-21.43
0.75
0.29
0.67
2.60
4.50
3.33
2.00
-12.00
a
15
Output (mV)
Load (kg)
0
5
10
15
20
25
30
35
Decreasing
10
Increasing
%FSO
0
0
20
%reading
40
60
Load (kg)
80
100
Static Characteristics
(b) Hysteresis
20
Output (mV)
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
Increasing
0.08
0.45
1.02
1.71
2.55
3.43
4.48
5.50
6.53
7.64
8.70
9.85
11.01
12.40
13.32
14.35
15.40
16.48
17.66
18.90
19.93
Decreasing
0.06
0.88
2.04
3.10
4.18
5.13
6.04
7.02
8.06
9.35
10.52
11.80
12.94
13.86
14.82
15.71
16.84
17.92
18.70
19.51
20.02
Hysteresis
(%FSO)
0.10
2.15
5.10
6.95
8.15
8.50
7.80
7.60
7.65
8.55
9.10
9.75
9.65
7.30
7.50
6.80
7.20
7.20
5.20
3.05
0.45
15
Output (mV)
Load (kg)
10
11.80 mV 9.85 mV
100% = 9.75%FSO
20 mV
0
0
20
40
60
80
100
Load (kg)
Static Characteristics
(c) Linearity: Terminal-based straight line (endpoint straight line)
Actual Output
(mV)
Linearity
(%FSO)
0
5
10
15
20
25
30
35
40
0
1
2
3
4
5
6
7
8
0.08
0.45
1.02
1.71
2.55
3.43
4.48
5.50
6.53
0.40
-2.75
-4.90
-6.45
-7.25
-7.85
-7.60
-7.50
-7.35
8
7
6
5
4
3
2
1
0
40
35
30
25
20
15
10
5
0
8.06
7.02
6.04
5.13
4.18
3.10
2.04
0.88
0.06
-0.30
-0.10
-0.20
-0.65
-0.90
-0.50
-0.20
0.60
-0.30
y ~ 0.20 mV/kg x
20
endpoint
15
Output (mV)
Load (kg)
Endpoint
line (mV)
10
5
3.43 mV 5.00 mV
100% = 7.85%FSO
20 mV
0
0
20
endpoint
40
60
80
100
Load (kg)
Static Characteristics
(c) Linearity: Best-fit straight line
Least square method: minimizes the sum of the square of the vertical
deviations of the data points from the fitted line.
Here, we will estimate y by y = mx + b
m=
N xy x y
2
N x 2 ( x )
y
x
m
b=
N
N
N = Total number of data points
Static Characteristics
x = Load (kg)
y = Load cell output (mV)
x
0
5
10
15
20
25
30
35
40
45
50
0.08
0.45
1.02
1.71
2.55
3.43
4.48
5.50
6.53
7.64
8.70
0.00
25.00
100.00
225.00
400.00
625.00
900.00
1225.00
1600.00
2025.00
2500.00
xy
0
2.25
10.2
25.65
51
85.75
134.4
192.5
261.2
343.8
435
x2
45
40
35
30
25
20
15
10
5
0
9.35
8.06
7.02
6.04
5.13
4.18
3.10
2.04
0.88
0.06
2025.00
1600.00
1225.00
900.00
625.00
400.00
225.00
100.00
25.00
0.00
420.75
322.4
245.7
181.2
128.25
83.6
46.5
20.4
4.4
0
2100
409.89
143500
28499.45
m = 0.2079 mV/kg
b = -0.6368 mV
Obtained eq.
Static Characteristics
(c) Linearity: Best-fit straight line
0.08
0.45
1.02
1.71
2.55
3.43
4.48
5.50
6.53
7.64
8.70
9.85
-3.45
-0.23
2.03
3.71
4.67
5.44
5.39
5.48
5.53
5.19
5.09
4.56
11.84
10.80
9.76
8.72
7.68
6.64
5.60
4.56
3.52
2.48
1.44
0.40
-0.64
12.94
11.80
10.52
9.35
8.06
7.02
6.04
5.13
4.18
3.10
2.04
0.88
0.06
-5.30
-4.82
-3.66
-3.04
-1.83
-1.83
-2.12
-2.74
-3.17
-2.97
-2.88
-2.30
-3.35
20
15
Output (mV)
-0.64
0.40
1.44
2.48
3.52
4.56
5.60
6.64
7.68
8.72
9.76
10.80
Linearity
(%FSO)
60
55
50
45
40
35
30
25
20
15
10
5
0
Actual Output
(mV)
Load (kg)
0
5
10
15
20
25
30
35
40
45
50
55
best-fit line
(mV)
10
6.53 mV 7.68 mV
6.53 mV 7.68 mV
100% = 5.70%FSO
100% = 5.70%FSO
20.79
20.15
mV mV
0
0
20
40
60
80
100
Load (kg)
Static Characteristics
(c) Linearity: Independent straight line
20
Output (mV)
15
10
5
4.48 mV 5.65 mV
100% = 5.85%FSO
20 mV
0
0
20
40
60
80
100
Load (kg)