Machine diagnostics
and data collection
Operating instructions
Dear Customer,
If you have any suggestions for improving
this instruction manual or the instrument
itself or any of its accessories, please let us
know.
PRFTECHNIK Condition Monitoring
Fax: +49 (0)89 99616-300
eMail: info@pruftechnik.com
2
Introduction
VIBSCANNER 09.2003
3
Introduction
Foreword
This manual describes the main functions of the VIBSCANNER as well as the most important steps of its operation.
The optional firmware modules FFT Analysis, Balancing
and Signal Analysis are described in the manual VIBSCANNER - FFT-Analysis & Balancing (VIB 9.664.G).
An online help function is integrated in the instrument that
can be called up at any time via the menu.
This manual is valid for firmware version 1.60 and higher.
Safety
VIBSCANNER must be treated in such a way that it is not
subject to any mechanical knocks.
The housing is made of an electrically conducting plastic.
Consequently, the instrument may only be used to measure
small signal voltage (30V) or small signal current
(20mA).
VIBSCANNER is only dustproof and watertight (IP65) when
the battery is plugged into the instrument. Please note that,
during the battery change, the instrument and the battery
only comply with protection class IP 50 (dustproof).
For signal measurement and for data transmission, only use
the connecting cable provided for this (see VIBSCANNER
Product Catalog VIB 9.661-4DG).
The relevant safety regulations must be observed during
measurements on machines.
Authorized use
- VIBSCANNER may only be used to record machine
signals within industrial environments.
- Transducers and cables may only be used for their
respective intended tasks as described in the corresponding sales leaflets.
VIBSCANNER 09.2003
CAUTION!
4
Introduction
VIBSCANNER 09.2003
CAUTION!
5
Introduction
VIBSCANNER 09.2003
6
Contents
Contents
Description ............................................................ 8
Interfaces and functional elements ................................ 8
Transducer module ................................................................ 9
Operation ............................................................ 13
Operating elements .................................................... 13
Program structure ....................................................... 14
Getting started ........................................................... 16
Setup menu ........................................................................ 17
Changing parameters .......................................................... 18
Instrument setting (setup) ................................................... 19
Limits ................................................................... 53
Transducers ......................................................... 58
External vibration transducers ............................................. 58
Setting up the transducer .................................................... 59
External temperature probes ............................................... 60
External RPM transducers .................................................... 61
Creating / deleting transducers ............................................ 62
VIBSCANNER 09.2003
7
Contents
Appendix ............................................................. 63
Text editor ........................................................................... 63
Options in the File Manager ................................................ 64
Tips & Tricks ........................................................................ 66
Data exchange with the PC ................................................. 68
ISO 10816-3 ........................................................................ 69
VIBSCANNER 09.2003
8
Description
Description
Interfaces and functional elements
Transducer
module
(see next page)
LED indicators
Display
Escape key
Function key
Joystick
Battery with
charging socket
LED indicators
Blue LED ...
... lights up after the measurement if the result is valid.
VIBSCANNER 09.2003
9
Description
Transducer module
(view from above)
Protective caps
Yellow channel
Blue channel
Temperature
probe
IR transmitter
Trigger (IR sensor)
LED
(red)
Blue channel: Input for vibration transducers, Pt100 temperature sensor and small signal voltage /currents (
30V / 20mA). The relevant connection cables are
marked with a blue protective rubber sleeve at the
connector.
Yellow channel: Interface for data transmission (RS 232),
connection for external trigger and output for analog
signals. The relevant connection cables are marked with
a yellow protective rubber sleeve at the connector.
Internal vibration transducers: Permanently-installed transducers for the measurement of machine vibrations above
10 Hz and high-frequency cavitation and bearing shock
pulse signals.
VIBSCANNER 09.2003
10
Description
Power supply
The VIBSCANNER is supplied with a NiMH battery that is
charged by a battery charger (VIB 5.420). During the
charging process, the VIBSCANNER is switched off and the
battery remains inside the instrument. Reserve batteries
can also be charged outside the instrument.
Note
The NiMH battery loses 1- 2% of its charge (selfdischarge) per day when the VIBSCANNER is switched
off. If you do not need the VIBSCANNER for a longer
period of time (> 3-4 weeks), please remove the
battery and leave it connected to the charger (not in
the intrinsically safe version). Before you reinsert it,
recharge the battery.
When VIBSCANNER is switched on and the battery
charger is connected, the charging process is interrupted and...
VIBSCANNER is supplied with mains power.
VIBSCANNER (intrinsically safe) is supplied with
battery power.
Measurements are not allowed with the charger connected because this can result in measurement errors.
After approx. 10 minutes, the VIBSCANNER loses the
time and date setting if the battery is removed.
Rechargeable batteries must not be charged in explosive environments!
VIB 5.420
VIBSCANNER 09.2003
CAUTION!
11
Description
Protective cap
Level of charge
LED (red, green)
Level of charge
(black= battery full)
12
Description
Battery empty
Note
VIBSCANNER 09.2003
13
Operation
Operation
Operating elements
The VIBSCANNER has three operating elements:
Joystick
The joystick moves the cursor either horizontally or vertically on the screen. By pressing down on the joystick
(clicking), the selection is confirmed (ENTER function).
Function key
The function key is used to move the cursor from the
display field into the menu column and vice-versa. As a
result, you can delete incorrect entries in the text editor.
Escape key
The Escape key is used to cancel a selection. It enables you
to return to the higher menu from a submenu.
VIBSCANNER 09.2003
Escape key
Function key
14
Operation
Program structure
The VIBSCANNER application program is graphic-oriented
and menu-driven.
Graphic-oriented
a) The measurement tasks are started via corresponding
icons on the screen.
b) In the basic version the measurement tasks are stored in
two screens that are opened via the screen icon S1 or S2:
The S1 screen contains measurement tasks for overall
vibration and other important variables (see page 16).
The S2 screen contains measurement tasks for overall
vibration that are optimized for specific machine types
and RPM ranges. The first line depicts the machine types
and the left column defines the RPM range (see page
32).
c) A corresponding selection screen exists for the optional
program modules FFT analysis, Balance and Signal
Analysis (details of the optional modules are given in
the VIBSCANNER - FFT analysis and balancing operating manual - VIB 9.664.G).
VIBSCANNER 09.2003
15
Operation
Menu-driven
a) All functions (e.g. Setup, Help, Save , File,...) are called
up via the menu column.
The menu column is hidden in some submenus due to
lack of space. To display it, either press the function key
or move the joystick to the right.
b) All the measurement and instrument parameters are
listed in menus and can be changed with the joystick.
Some menus branch further into submenus to provide a
clearer overview (e.g. VIBSCANNER Setup).
Menu
column
Meas.
parameters
VIBSCANNER 09.2003
16
Operation - Getting started
Getting started
Push the joystick upwards towards the display for one to
two seconds. Leave go of the joystick and the VIBSCANNER
main window appears:
Cursor
(reverse video)
Display field
Menu column
Status line
Main window1
The following measurement tasks can be started directly
via the relevant icon in the display field:
Preset measurement tasks (line 1):
- Vibration velocity
- Shock pulse measurement (bearing condition)
- Cavitation (e.g. in pumps)
Preset measurement tasks (line 2):
- RPM
- Temperature
- Vibration displacement
- Vibration acceleration
User-defined measurement tasks (line 3)
- Manual input of measured values
- Measurement of small signal currents / voltages (DC)
- Measurement of small signal currents / voltages (AC)
S: Icon for the measurement of characteristic overall values. Clicking on it calls up the next screen that has
measurement tasks.
VIBSCANNER setup: Instrument settings
Menu column:
FILE: to open the File manager
ROUTE: to start the Route operating mode
SETUP: Settings for measurement, transducer, evaluation, recording, machine
HELP: Online help on the highlighted icon
OFF: to switch off the instrument
VIBSCANNER 09.2003
Status line: The transducer that is selected for the measurement task is displayed.
17
Operation - Getting started
Setup menu
In the Setup menu, you can check the parameters of the
selected measurement task and change them if necessary.
The Setup menu is task-specific, i.e. the settings only
apply to the selected measurement task.
F
Note
VIBSCANNER 09.2003
18
Operation - Getting started
Changing parameters
Click on the parameter to open an entry mask, the text
editor or a submenu:
Parameter type Number: Press the joystick up/ down to
change the value; Press the joystick to the left/ right to
select the units, 10s, 100s, 1000s digits (multidigit
numbers can be set up more quickly in this way).
Parameter type Text: Enter the text in the text editor
(see page 63)
Parameter type Selection: The settings are selected
from a submenu (i.e. Units). If only a few parameters
are available for selection, the submenu does not appear. The parameters can be set simply by clicking on
them (e.g. tool tips = yes/no).
The cursor for variable parameters is displayed in reverse
video. A frame appears around non-variable parameters.
Example of parameter types
in the transducer setup
- Num: 'Power off'
- Text: Company
- Selection: 'Display'
Show menu
Scroll forwards
VIBSCANNER 09.2003
*Alternatively:
Move the cursor across the
right edge of the window.
19
Operation - Getting started
RATE:
VIBSCANNER 09.2003
20
Operation - Getting started
*xx: AM or PM
SELECT AVAILABLE SENSORS: VIBSCANNER contains a comprehensive transducer database. Since, usually, only particular transducers are used, it is possible to select favorite
transducers here. Only these transducers then appear
for selection in the list of the setup.
VIBSCANNER 09.2003
21
Operation - Getting started
22
Operation - Multimeter
Multimeter measurement
VIBSCANNER can be used as a measuring instrument for
characteristic overall values. This section describes the
operating sequence of these measurements in the multimeter mode.
Vibration measurement using an int. transducer
You can measure machine vibrations above 10 Hz as well as
high frequency shock pulse signals of bearings and cavitation using the built-in vibration transducers.
F
Note
Countersink the measurement location (e.g. with special countersink bit VIB 8.610) and clean it.
To measure the vibration level (effective vibration velocity, 10 Hz), select the vibration symbol shown on the left
of the main window.
Press the transducer vertically against the measurement
location with sufficient pressure (see figure above).
Click the joystick to start the measurement.
Unit
RMS / 0-peak-value
Number of averages
Transducer
VIBSCANNER 09.2003
Measurement screen
for vibration level
23
Operation - Multimeter
F
Note
VIBSCANNER 09.2003
24
Operation - Multimeter
* Exception:
Trend measurement (see page
33)
File manager
free memory in %
Directory
File
F
Note
VIBSCANNER 09.2003
Text editor
25
Operation - Multimeter
F
Note
F
Note
26
Operation - Multimeter
Temperature measurement
The temperatures of surfaces and liquids can be measured
with the built-in temperature probe (-50 to +100 C).
External transducers must be used for wider temperature
ranges (see page 60)
F
Note
VIBSCANNER 09.2003
27
Operation - Multimeter
RPM measurement
For a reliable RPM measurement, a reflecting mark
(e.g. Reflective tape, VIB 3.306) must be attached to
the shaft.
If you attach more than one measurement mark to the
circumference of the shaft, you must enter the number
in the transducer setup (Sens. [no/rpm] parameter).
F
Note
VIBSCANNER 09.2003
28
Operation - Multimeter
F
Note
Note
VIBSCANNER 09.2003
29
Operation - Multimeter
Press the transducer vertically to the prepared measurement location with sufficient force (see page 22).
If the transducer is permanently installed, connect the
transducer cable to the instrument.
Click on Start SIM to start the measurement.
VIBSCANNER 09.2003
30
Operation - Multimeter
New transducer:
Click on the first line (->) to display the list of manual
transducers.
Press the function key and click on New in the menu.
Enter a name in the text editor.
Set up the following parameters:
SENS. NAME: Name of transducer (text editor)
MEAS.TYPE: Measured quantity depends on the signal type
SIGNAL TYPE: Set to manual
EDIT QUANTITY: Description of the measured quantity
ENGIN. UNIT: Unit (text editor)
DECIMAL: Decimal places of the results display
VIBSCANNER 09.2003
31
Operation - Multimeter
VIBSCANNER 09.2003
32
Operation - Multimeter
Adapted
vibration measurements:
2nd window (S2)
Machines
(general)
Gears
Bearings
High RPMs
Medium RPMs
Note
VIBSCANNER 09.2003
33
Operation - Multimeter
Trend
The development of the machine condition can be recorded
and documented with the aid of a trend measurement. In
this case, characteristic values are always regularly recorded at the same measurement location under the same
conditions. The results are saved in a file.
Trend display
Mark a measurement task in the selection screen (S1,
S2), and click on File in the menu to open the file
manager (see Note, page 25):
File Manager
(see page 64)
Joystick functions:
VIBSCANNER 09.2003
34
Operation - Multimeter
1 measurement from
09.09.02
Procedure:
Move the cursor to the section to be zoomed.
Click once with the joystick:
VIBSCANNER 09.2003
35
Operation - Multimeter
Trend information
The measured value, date and time for each individual
measurement are saved.
Press the function key, and click on Info in the menu.
VIBSCANNER 09.2003
36
Operation - Route
F
Note
VIBSCANNER 09.2003
37
Operation - Route
If the next task at the measurement location is performed using the same transducer, the measurement
starts automatically after the result has been saved. If
you measure this with one of the built-in transducers,
hold the instrument to the measurement location
before you click on Save for this reason.
F
Note
38
Operation - Route
Info:
Displays the complete path in the OMNITREND database
for the selected entry.
Skip:
Enables you to skip the selected route element*. Thereafter, the route is considered to have been processed.
Example: If an aggregate is temporarily switched off, no
measurements can be performed on it.
Skipped elements are marked by an S symbol.
Displaying setup:
Press the function key in the task screen (see below) and
click on Setup in the menu.
Delete:
You can delete a route in the Route selection window
(page 36 above).
Mark the route, press the function key, and click on
Rem. (remove) in the menu.
<<- : Return
Pool:
If reserve measurement tasks for a route are stored in a
pool, you can start these at any time.
Press the function key, and click on Pool in the menu.
The VIBSCANNER changes to the Pool mode and, in the
same way as the route mode, displays the aggregates,
machines and measurement locations in the pool. The
sequence of measurements corresponds to that in the
route. To change back to Route mode, click on Route in
the menu.
VIBSCANNER 09.2003
Measured
task
39
Operation - Route
In adaptive measurement tasks a further diagnostic measurement starts automatically when a limit has been exceeded. In this way, the sequence of the route adapts itself
to the local machine status (adaptive). This diagnostic
measurement may be a spectrum or further characteristic
overall value and was defined in the OMNITREND software.
The limit that triggers the diagnostic measurement
(prewarning, warning, alarm), is defined as an 'adaptive trigger' in the OMNITREND software.
If no 'adaptive trigger' is defined, the diagnostic measurement is triggered when the smallest limit is exceeded.
Adaptive measurement tasks are marked by an arrow in the
task screen:
Adaptive task
F
Note
VIBSCANNER 09.2003
40
Operation - Route
Event
You can store additional information on special incidents
on the machine together with the measuring result. You
can select these data from a predefined event list (EVENT)
and add a comment (COMM) to the data record if necessary.
F
Note
VIBSCANNER 09.2003
41
Operation - Route
Visual inspection
In addition to electrically measurable machine signals, you
can also record and further process inspection data with the
VIBSCANNER. The data are either recorded as an event or
selected from a list with predefined results. For example,
you can be asked to check the oil level at a pump:
VIBSCANNER 09.2003
42
Operation - Route
VIBSCANNER 09.2003
43
Operation - Route
The aggregate selected here (Exhaust fan EX-1559) consists of a ventilator (Ventilator 12) and a drive motor
(Motor B145).
Aggregate (graphical overview)
The selected machine is surrounded by a frame; To navigate, move the joystick in the corresponding direction
(here: to the left).
VIBSCANNER 09.2003
44
Operation - Route
Location and
measurement direction
Top view
vertical
horizontal
horizontal
vertical
axial
axial
Measurement direction
45
Operation - Route
Note
VIBSCANNER 09.2003
46
Operation - Route
RPM-dependent parameters
The RPM is required as an additional parameter in the
following measurements:
Shock pulse measurement ->
for the normalization of carpet and maximum value
FFT spectrum ->
for the display of RPM-dependent (damage) frequencies
in OMNITREND
So that the RPM does not always need to be recorded
before each of these measurements, VIBSCANNER together
with OMNITREND provides the following functions for the
route mode:
Aggregate RPM
Production line speed
Aggregate RPM
For a single aggregate (e.g. motor-gear-pump), a reference
RPM is always recorded first. The RPM value is then
converted for all measurement locations on the aggregate
and saved together with the results of the measurements
listed above. The conversion factors are defined in OMNITREND and take into account the different RPM relations in
the aggregate (e.g. in gearing).
Reference RPM
Motor spectrum
VIBSCANNER 09.2003
Shock pulse
(roller bearing)
Gear
spectrum
47
Operation - Route
VIBSCANNER 09.2003
Before recording a spectrum or a shock pulse measurement on the aggregate, first measure the reference
RPM.
F
Note
48
Operation - Route
n1
n2
n3
VIBSCANNER 09.2003
ni: RPM
49
Operation - Route
F
Note
If you leave the entry screen with ESC, and want to start a
measurement, the following warning appears:
VIBSCANNER 09.2003
F
Note
50
Operation - Signal output
Signal output
The recorded vibration signal is available at the analog
output (yellow socket). It can be tapped with the VIBSCANNER cable - VIB 5.431 - and displayed on a further
processing measuring instrument (oscilloscope, vibration
analyzer,...). The signal can be listened to via a headphone
that is directly connected (VIB 6.670).
VIB 6.670
VIB 5.431
(vibration acceleration)
(vibration acceleration)
(vibration acceleration/
envelope for shock pulse)
Headphones
The headphone is used to listen to the machine vibration
signal (vibration acceleration) or the bearing signal (shock
pulse).
VIBSCANNER 09.2003
Measuring procedure
51
Operation - Signal output
F
Note
VIBSCANNER 09.2003
VIBSCANNER 09.2003
52
empty page
53
Limits
Limits
The VIBSCANNER uses limits that comply with ISO norms
or user-defined settings to evaluate the measurement results. Depending on the limits selected, one of the four
LEDs above the display lights up after measurement:
Blue=OK;
Green=Prewarning;
Yellow=Warning;
Red=Alarm
For vibration measurements, limits corresponding to the
new ISO 10816-3 can be set up (page 66).
For the shock pulse measurement (bearing condition), the
upper alarm value is preset:
- Carpet value = 15 dBn
- Max. value = 35 dBn
User-defined limits must be set up for all further measurement tasks.
VIBSCANNER 09.2003
F
Note
54
Limits
Setup menu
VIBSCANNER 09.2003
55
Limits
VIBSCANNER 09.2003
56
Limits
User-defined limits
The new setup and changing of user-defined limits is shown
using the example of a temperature measurement:
F
Note
VIBSCANNER 09.2003
57
Limits
1 = Alarm
2 = Alarm & warning
3 = Alarm & warning &
Pre-warning
F
Note
58
Transducer
Transducers
External vibration transducers
... are required for vibration measurements on
Note
VIBSCANNER 09.2003
59
Transducer
Transducer setup
F
Note
60
Transducer
F
Note
VIBSCANNER 09.2003
61
Transducer
F
Note
VIBSCANNER 09.2003
62
Transducer
VIBSCANNER 09.2003
63
Appendix
Appendix
Text editor
Text line
Character field
Menu
VIBSCANNER 09.2003
To cancel the input and close the text editor, use the Escape
key.
F
Note
64
Appendix
1. To display a file:
Click on the required file in the list, or on Load in the
menu to open the marked file.
2. To create a directory:
Click on the higher directory to open it.
Click on MkDir (Make directory) and enter the directory name in the text editor.
Note
VIBSCANNER 09.2003
65
Appendix
5. To copy/move a file:
Mark the file and click on Copy in the menu.
Click on the required directory to open it.
VIBSCANNER 09.2003
yes/ no
file size / save date
file size & date / time
yes / no
yes / no
66
Appendix
VIBSCANNER 09.2003
b. Tooltips
In addition to the online help, a brief text can be displayed
for each symbol marked on the screen. The Tooltips
function can be switched on/off in the VIBSCANNER setup
(Option Display).
67
Appendix
SERNO: 00105
MAIN: 150 (0220)
04.02.2002 13:04:35
BOOT: 106 (0028)
* = Subversion
VIBSCANNER 09.2003
68
Appendix
Note
Attention!
VIBSCANNER 09.2003
69
Appendix
ISO 10816-3
D
11
7.1
4.5
3.5
2.8
2.3
1.4
A
rigid
elastic
rigid
elastic
0.71
rigid
elastic
rigid
elastic
mm/s
Foundation
Intermediate gear/
belt drive
Motors
160<H<315 mm
Group 3
Group 2
Motors
315 mm < H
Group 1
Machine type
Group
140
113
71
56
45
36
28
22
90
18
VIBSCANNER 09.2003
11
Newly commissioned
70
Appendix
Technical data
Hardware
Outputs
RS 232 (PC conn., up to 115 kbaud), Headphone, Analog signal (4Vpp; Rout= 200 Ohm)
Operating elements
1 joystick (Cursor & ENTER function)
2 keys
(Menu and Escape)
LED indicators
Four LEDs for device status / signal evaluation
Display
Graphical pixel display (backgr. illumination)
Dimensions
54 x 27mm/ 2.1 x 1.1inch
64 x 128 pixels
Contrast
Adjustable
Illumination Adjustable
Power supply
NiMH rechargeable pack (7.2V / 1.5 Ah)
Charg. dur. < 6 hours (intr. safe: <10 hours)
Operat. dur. > 10 hours in intermit. use
> 6 hours in continuous use
with illumination
Charg. display 2 LEDs (green, red)
Charg. temp. 10 ... 40C / 50 ... 104F
Sleep mode Adjustable
Internal sensors
Vibration/ shock pulse (bearing condition)
RPM (IR sensor with light point for adjustment)
Temperature (NiCrNi)
Signal processing
RMS, 0-P, P-P, max/carpet, envelope curve,
rectifying
Filter:
Highpass (2 /10 Hz; 1 / 5 kHz)
Lowpass (1 / 5 / 40 kHz)
Integrat.: Two selectable stages
Sampling frequencies: Up to 64kHz (depending
on measuring range)
Memory
4 MB
Housing
ABS strengthened with steel fiber
Environmental protection
IP 65
Relative humidity
10 ... 90%
Temperature range
Operation
0... +60C / 32...140F
Storage
-20... +65C / -4...160F
Dimensions (HxWxD)
250 x 100 x 55 mm / 9.8 x 3.9 x 2.2 inch
Weight
approx. 690 g / 24 oz.
Transport case
Polypropylene, black
Dimensions
390 x 340 x 90 mm
(HxWxD)
15.3 x 13.4 x 3.5 inch
Measurement range / Accuracy
RPM
60 ... 60000 min-1 / 0.1
Temperature
Pt 100
-50..+600C / 1+ sensor%
(-58 ...+1112F)
NiCrNi
-50...+100C / 0.5 + 3%
(internal) (-58...+212F)
(external) -50...+100C / 0.5+ sensor%
(-58...+212F)
(external) +100..+1000C / 1+ sensor%
(212...1832F)
Extra-low
-9...+9V / 2% (Ri=30kOhm,
voltage
with cable VIB 5.440)
(AC/DC)
-30...+30V / 2% (Ri=100kOhm,
with cable VIB 5.433)
Extra-low
-20...+20mA / 2%;
current
4...20mA / 2% (Rshunt =200 Ohm
(AC/DC)
with cable VIB 5.434)
For internal sensors, external sensors (1A/ms-
CLD*; 100mV/g ICP) and external measurement devices (1mV/ms-), the following applies:
Displacement
Velocity
Acceleration
Shock pulses
<
<
<
<
9000 m (p-p) / 1%
9000 mm/s (p-p) / 1%
6000 m/s (p-p) / 1%
81 dBsv / 3dB
Fulfilled standards
Frequency response according to ISO 2954 other parameters and measured variables
according to DIN 45662 class 1
Noise, internal sensor (from 10 Hz)
Velocity
0.1 mm/s eff.
Displacement 2m eff. (instr.+sensor)
Shock pulse < 0dBsv , peak
Compatibility (external transducer)
Vibration
CurrentLineDrive (CLD*) transducer
VIBCODE for measurement studs
with location recognition;
TIPTECTOR hand-held probe for
vibration & shock pulse;
VIBSCANNER 09.2003
Measurement channels
Analog: Vibration signals (LineDrive, ICP)
Temperature (Pt 100, NiCrNi)
Transducer and instrument outputs
AC ( 30V; 0-20mA)
DC ( 30V; 0-20mA)
Digital: Trigger (5V TTL)
71
Appendix
Firmware
Compatibility (external transducer)
Quick fit transducer for SPM
measurement studs;
Industrial transducer with magnetic,
threaded, adhesive mounting, with
adapter and hand-held probe
ICP transducer
Velocity detection (mV/mms-1)
Displacement detection (mV/m)**
RPM
Optical sensor (passive/ active)
5V TTL (opt. or induct. transducer)
Temp.
NiCrNi (magnetic/ probe)
IR probe
Pt100
CE conformity (for cable length < 3m)
Interference EN 50081-1
emission
(Residential area)
Interference
sensibility
EN 50082-1
(industrial area)
< 4% of meas. value or
< twice noise value
VIBSCANNER 09.2003
II 2 G
Measurement functions
Velocity / displacement / acceleration in machine-specific measurement tasks;
Shock pulse (bearing condition);
Temperature;
RPM
Process parameters
Manual input;
User-defined DC: 30V; -20...+20mA
tasks
AC: 30V; -20...+20mA
(Extra-low voltage/ current)
Data processing
Evaluation functions for characteristic overall
value;
Bearing diagnosis using shock pulse;
Machine-condition evaluation according to ISO
standards (vibration according to the new
ISO 10816-3);
Data collection functions for characteristic
overall value and for machine inspection;
Measurement parameters
Averaging:
linear, peak hold, exponential,
time-synchronous;
Adjustable averaging no. & time
Meas. time: adjustable
Ampl. range Adjustable, fixed or autorange
Measurement setups
Predefined, knowledge based measurement
setups for machines, bearing and gear diagnosis;
Freely selectable multimeter measurement
functions;
User interface
Graphic-oriented and cursor controlled:
Icons for measurement tasks;
Machine pictures for measurement
location scanning;
Graphic route guide;
Integrated help function
Units
ISO and US units, switchable
Comments
User-defined events with comments
Language
English, German, French, Italian
72
Troubleshooting
Troubleshooting
Symptom:
After the power has been switched on, the blue LED flashes
slowly.
Meaning:
The device is in boot mode. This can have two reasons:
1. Boot mode was manually entered by the user
2. The application firmware is missing or corrupted.
What to do:
If the battery charger is connected to the device, disconnect
it.
Reset the device.
If the device still goes into boot mode, update the firmware
using the VIBSCANNER update tool.
Symptom:
When switching on the device, the red LED flashes quickly
for approx. 1 second.
Meaning:
The rechargeable battery is empty.
What to do:
Charge the battery.
FOR SERVICE ONLY: the battery check during power on
can be bypassed by pressing the left button (ESC key)
during power on.
Symptom:
When switching on the device the backlight of the displays
lights up but nothing appears on the display.
What to do:
Reset the device.
If this doesnt work, update the firmware using the VIBSCANNER update tool.
VIBSCANNER 09.2003
Meaning:
The software is not running properly.
73
Troubleshooting
Symptom:
Battery does not charge (red LED on the battery does not
light up) when the charger is connected and VIBSCANNER
is switched off.
Meaning:
VIBSCANNER erroneously sends a starting signal to the
battery.
What to do:
Remove the battery from VIBSCANNER and charge it
separately.
Symptom:
Charging process in the case of a deeply discharged battery
terminates prematurely (5min 1h), green LED lights up
on the battery.
Meaning:
Deeply discharged battery.
What to do:
Switch on VIBSCANNER without the charger connected
and leave it switched on until the battery is empty again
and the instrument switches off by itself. Then repeat the
charging process. If necessary, charge the battery separately.
Symptom:
Charging process does not start, red LED on the storage
battery flashes continuously (>20min).
Meaning:
Error condition identified during the charging process, or
battery defective.
What to do:
Ambient temperature 1040C? Discharge the battery
again. Charge the battery separately. If charging is still not
possible, return the battery for repair.
VIBSCANNER 09.2003
Symptom:
Battery level inaccurate appears when switching off or
the display of the level of charge is obviously wrong.
Meaning:
Charge electronics must be newly programmed.
What to do: see page 11
74
Troubleshooting
** Select Action **
Format
Abort
Meaning:
A serious error occurred while accessing the internal
VIBSCANNER hard disc.
What to do:
Try to abort the current action by pressing the right button
(function key).
If this doesnt help, try to reset the device.
If this doesnt help, try to format the VIBSCANNER hard
disc by pressing the left button (ESC key). Formatting the
hard disc may take a few minutes.
If the format is successful, all files on the VIBSCANNER are
deleted. In order to restore these files, use the VIBSCANNER UpdateTool and download the files by choosing the
Filetransfer button in step 4. Make sure that you download
the files from the same firmware version as that already on
the VIBSCANNER.
If the format is not successful, return the device for repair.
Meaning:
A serious error occurred while accessing the internal
VIBSCANNER hard disc
What to do:
Retry the current action by pressing the left button (ESC
key).
VIBSCANNER 09.2003
** Filesystem Error **
Media status == ....
Volume == .....
Serial # == ....
Volume is ....
.....
** Select Action **
Retry
Clear
Abort
75
Troubleshooting
Meaning:
A serious error occurred while accessing the internal
VIBSCANNER hard disc.
What to do:
Retry the current action by pressing the left button (ESC
key). If this doesnt help, try to abort by pressing the right
button (function key). If this doesnt help, clear the current
action by pressing the joystick. If this doesnt help, try to
reset the device.
If the problem persists, return the device for repair.