You are on page 1of 3

Einstein II DGS Plus TFT

SPECIFICATION
Test Range 10 mm to 5 meter (0.500 in to 200 in) 10 mm to 5 meter (0.500 in to 200 in) 10mm to 5 meter (in steel) adjustable 10 mm to 5 meter (in steel)
(in steel). In Hot Key mode it has 13 (in steel). In Hot Key mode it has 13 in 1mm or 10mm step adjustable in 1mm or 10mm
preset values. Fine mode adjustable in preset values. Fine mode adjustable in step.
step of 1mm (0.2 in). step of 1mm (0.2 in).
Velocity 1000 m/sec to 9999 m/sec (0.0400 1000 m/sec to 9999 m/sec (0.0400 1000 m/s to 9999 m/s adjustable in 1 1000 m/s to 9999 m/s adjustable
in/s to 0.4000 in/s). In Hot Key mode in/s to 0.4000 in/s). In Hot Key mode m/s or 10 m/s step. Direct velocity in 1m/s or 10 m/s step.
it has 7 preset values. Fine mode it has 7 preset values. Fine mode setting for longitudinal and shear
adjustable in step of 1 met/sec (0.0001 adjustable in step of 1 met/sec (0.0001 waves for steel.
in/s). in/s).
Delay Variable from 0 to 3000 mm (0 to 100 Variable from 0 to 3000 mm (0 to 100 Up to 3 meter adjustable in 0.5mm or Up to 3 meter adjustable in 0.5mm
in). In Hot Key mode 5 mm (0.2 in). in). In Hot Key mode 5mm (0.2 in). 5mm step or 5mm step.
Fine mode adjustable in step of 0.5 mm Fine mode adjustable in step of 0.5mm
(0.02 in). (0.02 in).
Gain 100 dB calibrated gain adjustable in 100 dB calibrated gain adjustable in 100 dB calibrated gain adjustable in 100 dB calibrated gain adjustable
0.5, 1, 2, 6, 12 or 20 dB step. 0.5, 1, 2, 6, 12 or 20 dB step. 0.5, 1, 2, 6, 12 or 20 dB step. in 0.5, 1, 2, 6, 12 or 20 dB step.
Rejection 0 to 100% FSH with LED indicator 0 to 100% FSH with LED indicator. 0 to 100% FSH with LED indicator 0 to 100% FSH with LED indicator.
Rectification Full-wave rectified with filtering Full-wave rectified with filtering. Fullwave rectified with filtering. Fullwave rectified with filtering.
Frequency Broad Band amplifier 0.5 MHz to 15 Broad Band amplifier 0.5 MHz to 15 0.5 MHz to 15 MHz (Wideband) 1 MHz to 6 MHz.
MHz. MHz.
Test Modes Pulse echo and transmit/receive. Pulse echo and transmit/receive. Pulse echo and transmit/receive. Pulse echo and transmit/receive.
Pulser Transmission pulse Negative spike with Transmission pulse Negative spike with
selectable high or low power selectable high or low power.
Damping Damping high / low is selectable. Damping high or low is selectable.
Freeze/Peak Freeze A- Scan Freeze and Peak Freeze A-Scan Freeze and Peak Freeze
available. In Peak freeze it holds echo available. In Peak freeze it holds echo
dynamic pattern which is useful during dynamic pattern which is useful during
angle beam probing to locate peak angle beam probing to locate peak
signal. signal.
Connectors BNC and LEMO(Size 1) both are BNC and LEMO (Size 1) both are BNC and LEMO(Size 1) both are BNC and LEMO (Size 1) both are
provided. provided. provided provided.
Monitor Dual gate adjustable in 1% of Screen Dual gate adjustable in 1% of Screen Dual gate adjustable in1%of Screen Dual gate, adjustable in 1% of
width with Positive/Negative logic, Gate width with Positive/Negative logic, width with Positive/Negative logic, Screen width with
Expand modes. Gate Expand modes. Gate Expand modes Positive/Negative logic, Gate
Expand modes.
Gate Expand Expands Range to width of the gate. Expands Range to width of the gate. Expands Range to width of the gate Expands Range to width of the
gate.
A-Scan memory 200 Trace Patterns can be stored (with 200 Trace Patterns can be stored (with 200 Trace Patterns can be stored, 150 Trace Patterns can be stored,
Note/Detail) recalled, printed or Note/Detail), recalled, printed or recalled, printed or transferred to PC recalled, print or transferred to PC
transferred to PC via RS-232 serial port transferred to PC via RS-232 serial via RS-232 serial port. via RS-232 serial port.
port.
Calibration Set-up 50 different calibration set-ups can be 50 different calibration set-ups can be 50 different calibration set-ups can be 50 different calibration set-ups
Stored and Recalled Stored and Recalled. Stored and Recalled can be Stored and Recalled.
Software EinSoft DGS Interface software for EinSoft Plus Interface software for EinSoft Interface software for EinSoft Interface software for
transferring A-Scan from Einstein-II transferring A-Scan from Einstein- II transferring A-Scan from Einstein- II transferring A-Scan from Einstein-
DGS to PC is supplied. PLUS to PC is supplied. TFT to PC is supplied II(R) to PC is supplied.
Labeling Any PC Printer with serial port can
be attached for instant print-out of
Trace Pattern(A-Scan with Set-up
data).
Printer Attachment IBM compatible Printer having serial Any PC Printer with serial port can be Any PC Printer with serial port can be
port can be directly attached to main attached for instant print-out of Trace attached for instant print-out of Trace
unit for printing of stored A-Scan Pattern(A-Scan with Set-up data). Pattern (A-Scan with Set-up data).
waveform with calibration data
Display High brightness Colour LCD/TFT 122x92 mm (320x240 pixel) Colour 122x92 mm (320x240 pixel) Colour 122x92 mm (320x240 pixel)
Display. Display area 320x240 LCD/TFT display for better visibility LCD/TFT display for better visibility Colour LCD/TFT display for better
pixel(122x92mm). Eleven different with option of eleven different colour with option of eleven different colour visibility with option of eleven
color scheme option. sets. sets. Different colour sets or Black and
White LCD.
Full Screen By pressing Zoom keyA-Scan can be By pressing Zoom key A-Scan can be By pressing Zoom key A-Scan can be By pressing Zoom key A-Scan can
displayed in Full Screen area. displayed in Full Screen area. displayed in Full Screen area. be displayed in Full Screen area.
Reference A-Scan Reference A-Scan pattern of standard Reference A-Scan pattern of standard Reference A-Scan pattern of
test object can be saved and recalled test object can be saved and recalled standard test object can be saved
Reference A-Scan pattern of standard in the background for easy in the background for easy and recalled in the background for
test object can saved and recalled in comparision during testing. comparision during testing. easy comparison during testing.
the background for easy comparision
during testing.

DAC Dynamic DAC curve can be Digitally DAC curve can be Digitally plotted DAC curve can be Digitally plotted on DAC curve can be Digitally
plotted on screen with -6dB and -14dB (Smooth parabolic curve) on screen screen with -6dB and -14dB DAC plotted on screen. DAC curve can
DAC curves. DAC curve can be set as with -6dB and -14dB DAC Curves. DAC curves. DAC curve can be set as flaw be set as flaw monitor gate.
flaw monitor gate. Using minimum 2 to curve can be set as flaw monitor gate. monitor gate with -6dB and -14dB DAC
maximum 10 points DAC curve can be curves.
plotted.
DGS Defect size evaluation can be
performed using 18 predefined probe
settings and custom probe set-ups.
Defect size is directly displayed in ERS
value (Equivalent Reflector Size).
Digital Readout Thickness/Depth can be displayed in Thickness/Depth can be displayed in Thickness/Depth can be displayed in Thickness/Depth can be displayed
And Trig. Function digital readout when using a normal digital readout when using a normal digital readout when using a normal in digital readout when using a
probe and Sound path, Surface probe and Beam path, Surface probe and Beam path, Surface normal probe and Beam path,
Distance and Depth are directly Distance and Depth are directly Distance and Depth are directly Surface Distance and Depth are
displayed when angle probe is in use. displayed when angle probe is in use. displayed when angle probe is in use. directly displayed when angle
Measurement point can be selected to Measurement point can be selected to probe is in use.
be Peak or Flank. Echo height, ERS be Peak or Flank.
value, dB diff of DGS curve to signle
height, Echo height w.r.t. DAC can be
measured.
Measurement Unit Metric or British unit of measurement is Metric or British unit of measurement
selectable. is selectable.
Video Output Composite video signal (PAL) output for Composite video signal (PAL) output for
connection to a TV. connection to a TV.
PRF High or low PRF is selectable. High or low PRF is selectable
Update Rate 50Hz. 50Hz. 50Hz 50Hz.
Power Lithium-Ion Battery pack 14.4VDC, Lithium-Ion Battery pack 14.4VDC, Lithium-Ion Battery pack 14.4VDC, Lithium-Ion Battery pack 14.4VDC,
4AH, gives 8 hours continuous 4AH, gives 8 hours continuous 4AH, gives 8 hours continuous 4AH, gives 8 hours continuous
operation from fully charged battery. operation from fully charged battery. operation from fully charged battery. operation from fully
charged battery. Full charging is
done within 4 hours.
Charger Input Voltage 100 to 240 VAC. Charge Input Voltage 100 to 240 VAC. Charge Input Voltage 100 to 240 VAC. Charge Input Voltage 100 to 240 VAC.
status indicator is provided. status indicator is provided. status indicator is provided. Charge status indicator is
provided.
Temperature 0 to 55 C. (For operation) 0 to 55 C.(For operation) 0 to 55 C 0 to 55 C.
Size 152 x 255 x 70 mm (HxWxD). 152 x 255 x 70mm (HxWxD). 152 x 255 x 70mm (HxWxD). 152 x 255 x 70 mm (HxWxD).
Weight 2.3 Kgs. with Battery. 2.3 Kgs. with Battery 2.3 Kgs. with Battery. 2.3 Kgs. with Battery.

You might also like