Professional Documents
Culture Documents
Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
--````,,`,,,``,```,```,,`,-`-`,,`,,`,`,,`---
ground noise might interfere with hearing the audible output of 9.2 Calibration/Sensitivity Validation EquipmentUse the
a speaker. following equipment for calibration of the test system:
8. Apparatus NOTE 1This equipment serves a dual function; either to calibrate the
ultrasonic instrument for leakage rate approximation as in 11.3, or to
8.1 Ultrasonic Leak Detection System:
verify the sensitivity for detection and location as in 11.1.
8.1.1 The system shall consist of an instrument, probes,
focusing probe accessory and earphones. (A speaker may or 9.2.1 Leak Standard, with a preset flow rate of 6.7 3 105
may not be utilized.) mol/s (1.5 std. cm3/s at 0C) 65 % for both Class I and Class
8.1.2 The system shall provide for detection of acoustic II instruments. The orifice size shall be approximately 0.2 mm
energy in the ultrasonic range from 20 000 to 100 000 Hz and [0.008 in.].
shall translate this energy into an audible signal that can be 9.2.2 Regulator, for the nitrogen supply with output pres-
heard by use of earphones or speaker, or both. sure and flow gages. The tank pressure gage is optional.
8.1.3 The detected energy shall be indicated on a signal 9.3 Air Probe Sensitivity Validation:
level indicator. 9.3.1 For Class I instruments, locate the detection probe a
8.2 Minimum Instrument RequirementsThe instrument distance of 10.0 m (60.1 m) from the calibrated leak standard
shall meet the following requirements: specified in 9.2.1. For Class II instruments, place the detection
8.2.1 The detected ultrasonic energy shall be indicated on a probe a distance of 5.0 m (60.1 m) from the calibrated leak
signal level indicator with a change above normal background specified in 9.2.1.
fluctuations.
9.3.2 Check to see that the detector probe and leak source
8.2.2 The audible response shall consist of the down-
are aligned to obtain the peak response (see Fig. 1).
converted heterodyned ultrasonic signal. This audio signal will
be representative of the amplitude or frequency characteristics, 9.3.3 A signal level indication above normal background
or both, of the original ultrasonic signal. Heterodyned signals fluctuations should be observed.
could allow the operator to discriminate audible background 9.3.4 Place a sound absorbing barrier in front of the micro-
interference as in 7.2. phone, blocking out the calibrated leak source; the meter
8.2.3 The instrument shall be equipped with a sensitivity reading should zero with a corresponding absence of an audible
control or gain adjustment, or both, to achieve the conditions of signal.
9.3.3. 9.4 Recalibration:
9.4.1 Recheck or validate the sensitivity of the equipment at 10.2.2 Check safe pressure rating to be sure it will not be
the beginning of each shift or designated work period interval. exceeded during test.
This test must be performed at the same sensing frequency as 10.3 Temperature of Vessel and Testing Medium:
the initial test. 10.3.1 The temperature of the pressurizing gas must not be
9.4.2 Recheck the equipment when abnormalities are ob- at a level that would be injurious to the part or its components.
served in its operation. 10.3.2 The test gas is dry nitrogen. (Compressed air can
9.5 Ultrasonic Transmitter Sensitivity Validation, should contain oil or water droplets which can seal leaks.)
have the generated amplitude validated before each initial use. 10.4 Pressure:
This could be done by placing the ultrasonic transmitter a fixed 10.4.1 Unless otherwise specified the gage pressure shall be
distance from the detection instrument to obtain a reading or by at least 70 kPa [10 psig]. Normal safety precautions should be
taking a NIST-traceable reading with a sound pressure level observed when pressurizing the part under test with the gas to
detector. Another method would be to place the ultrasonic avoid test part ruptures.
transmitter in a container with a known leak that is equivalent 10.4.2 Gradually increase the pressure in the part to final
to the leaks that are to be detected. To eliminate vessel or test pressure.
container vibrations from influencing the readings, the interior 10.5 Ultrasonic Transmitter MethodInspect test object to
should be lined with closed cell foam, or rubber, or both. The be sure it is free of debris and liquids at the test sites.
orifice size shall be equivalent to the leaks that are to be
detected. 11. Detection and Location of Pressurized Gas Leaks
11.1 Detection:
10. General Considerations 11.1.1 Set the sensitivity at maximum.
10.1 Openings: 11.1.2 Begin to scan by pointing the probe towards the test
10.1.1 Seal all openings using plugs, covers or other suit- area. The procedure is to go from high sensitivity to low
able materials that can be readily and completely removed after sensitivity, reducing the sensitivity as the leak is approached.
the completion of the test. 11.1.3 Note the fluctuations in meter readings and the
10.1.2 Provide a gas inlet by attaching a valve to one of the volume from the earphones or speaker.
test covers on all items to be pressurized. 11.1.4 If there is too much ultrasound in the area, reduce the
10.2 Check of Test Parts: sensitivity and continue to scan.
10.2.1 Examine the part and test equipment before pressure 11.1.5 If it is difficult to isolate the leak due to competing
is applied to ensure that it is tight and all appurtenances that ultrasound, place the focusing probe accessory over the scan-
should not be subjected to the test pressure have been discon- ning probe. This increases the directional response character-
nected or isolated by valves or other suitable means. istics of the probe.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).
--````,,`,,,``,```,```,,`,-`-`,,`,,`,`,,`---