You are on page 1of 8

Reviews

Pulse Oximetry: Principles and Limitations


JAMES E. SINEX, MD

The pulse oximeter has become an essential tool in the modern practice the blood was carried by its colored component. 3 This
of emergency medicine. However, despite the reliance placed on the component was isolated the following year by Hoppe-
information this monitor offers, the underlying principles and associated Seyler, who further noted that the pattern of light absorption
limitations of pulse oximetry are poorly understood by medical practitio- of this substance he dubbed "haemoglobin" changed when
ners. This article reviews the principles of pulse oximetry, with an eye shaken in air. 4 Over the following decades, the optical
toward recognizing the limitations of this tool. Among these are perfor-
mance limitations in the settings of carboxyhemoglobinemia, methemo- spectra of the reduced and oxygenated species of hemoglo-
globinemia, motion artifact, hypotension, vasoconstriction, and anemia. bin (Hb) were further studied and elucidated.
The accuracy of pulse oximetry is discussed in light of these factors, with The first system to use transillumination to measure in
further discussion of applications for pulse oximetry in emergency vivo oxygen saturation was developed in 1935 by Matthes,
medicine, including both oximetric and plethysmographic operation. The though difficulties with calibration made it an unwieldy
pulse oximeter is an invaluable instrument for emergency medicine device. Wood in 1948 employed a refined version of this
practice, but as with any test the data it offers must be critically appraised "oximeter" as an anesthesia monitoring device. Unfortu-
for proper interpretation and utilization. (Am J Emerg Med 1999;17:59-67. nately, it too was fraught with practical limitations, requiring
Copyright 1999 by W.B. Saunders Company) individual calibration and a dedicated technician to operate. 5
The greatest step forward in the development of the
Pulse oximetry has become an essential tool in the current modern pulse oximeter occurred in 1974, interestingly
practice of emergency medicine. Detection of hypoxia by enough as an incidental finding. Before this, in order to
clinical indicators alone is notoriously unreliable, whereas isolate arterial blood for transillumination, oximeters had
the pulse oximeter offers real-time assessment of oxygen- relied on compression and heating of the ear lobe to remove
ation status on a moment-to-moment basis, reflecting effi- venous and capillary blood and to "arterialize" the tissue, at
cacy of interventions as well as progression of disease times with resultant second-degree burns. In 1974 Aoyagi,
processes. This technology is noninvasive, widely available, while trying to develop a dye-dilution technique for cardiac
and simple to use. Proper interpretation of pulse oximetry output determination, noted that the washout curves he was
data is dependent on a basic understanding of the underlying measuring with an ear densitometer were muddied by
principles and inherent limitations of the instrument. Unfor- pulsatile variations. In working to eliminate these variations,
tunately, studies have shown a lack of knowledge among Aoyagi recognized that the absorbency ratios of these
both residents and medical staff concerning the limitations pulsations at different wavelengths varied with oxygen
and interpretation of pulse oximetry results? ,2 This impor- saturation. Employing the principles discussed in the next
tant topic has not been widely addressed in the emergency section, Aoyagi developed and marketed the first pulse
medicine literature. This article reviews the principles and oximeter in 1975. 3
limitations of pulse oximetry, with discussion of the accu- The later development of reliable light-emitting diodes
racy and appropriate clinical uses of this invaluable tool. (LEDs), photodetectors, and microprocessors ushered in the
current era of modern pulse oximetry. The availability and
HISTORY use of this instrument boomed through the 1980s, initially in
the hands of anesthesiologists, and then throughout medical
The development of pulse oximetry has been based on practice.
more than a hundred years of experimental and engineering
antecedents. Stokes in 1864 first recognized that oxygen in
PRINCIPLES
The principle of pulse oximetry is that of spectral
From the Departmentof EmergencyMedicine, Universityof Cincin- analysis: the detection and quantitation of components in
nati, Cincinnati, OH, and the Department of Emergency Medicine,
Medical College of Georgia,Augusta, GA. solution by their unique light absorption characteristics. The
Manuscript receivedAugust 1, 1997; acceptedAugust 21, 1997. central tenet of spectral analysis is the Beer-Lambert law,
Address reprint requests to Dr Sinex, Department of Emergency which states that the concentration of an absorbing sub-
Medicine, AF-2018, Medical College of Georgia, 1120 15th St, stance in solution can be determined from the intensity of
Augusta, GA30912. light transmitted through that solution, given the intensity
Key Words:Oximetry,physiologic monitoring, anoxia, plethysmog-
raphy. and wavelength of incident light, the transmission path
Copyright 1999 by W.B. Saunders Company length, and the characteristic absorbance of that substance at
0735-6757/99/1701-0019510.00/(3 a specific wavelength (the extinction coefficient).5 Restated,
59
60 AMERICAN JOURNAL OF EMERGENCY MEDICINE Volume 17, Number 1 January 1999

absorbance is dependent on solute concentration (eg, oxygen- van absorption


ated and reduced Hb), and absorbance can be determined by (Due arterialpulsatile)
transmitting light of a specific wavelength across a solution --Arte~ bloodabsorption
(or tissue) and measuring its intensity on the other side. =~ ' . " #~ ".='~" .=..,#..=..~=.~,.==
To arrive at oxygen saturation, the relative concentrations t,.,- ~ . = . . , ~ . # . = . . ~ ; .~ . . . ~ . ~ . . ~ . ~
--Ven( bloodabsorption
of reduced Hb and oxyhemoglobin (O~Hb) must be known,
e~
and a wavelength of light must be used that each chromo-
phobe will preferentially absorb. Figure 1 illustrates the
~--0the ~sueabsorption
difference in absorbance spectra for reduced Hb and O2Hb
within the red to near-infrared wavelengths. The substantial
variance demonstrated within this range is fortunate for two
Time
reasons. First, red and near-infrared light readily penetrate
tissue, whereas blue, green, and yellow light, as well as FIGURE 2. Tissue composite demonstrating comparative light
longer-wavelength infrared light, tend to be significantly absorption of dynamic and static components. (Courtesy of
absorbed by tissue and water. Second, LEDs, which can Ohmeda.)
reliably emit a specific wavelength of light, are widely
available within the red to near-infrared range for use as work properly, the system must be able to isolate the
light sources. 6 The two wavelengths commonly used are absorbance of arterial blood components from that of venous
those at roughly 660nm (red) and 940nm (near-infrared), blood, connective tissue, and other absorbers. The pulse
although some pulse oximeters use slightly different wave- oximeter does this by exploiting the pulsatile nature of
lengths. Note from the absorbance spectra (Figure 1) that at arterial blood and determining the pulse-added component
660nm, reduced Hb absorbs light roughly ten times more of the signal (Figure 2). By measuring transmitted light
readily than does O2Hb (with extinction or absorbance several hundred times per second, the pulse oximeter is able
shown on a logarithmic scale). The reverse is true at 940nm, to distinguish the variable, pulsatile component of arterial
where O2Hb absorbs more light. blood (which can be referred to as the alternating current
Two LEDs in the pulse oximeter probe each emit light of signal, or AC) from the unchanging, static component of the
specific wavelength through a cutaneous vascular bed, such signal (ie, the direct current signal, or DC) made up of the
as that of the digits or the ear lobe. A photodiode detector at tissue, venous blood, and nonpulsatile arterial blood. The
the far side measures the intensity of transmitted light at pulsatile component, generally comprising 1% to 5% of the
each wavelength, from which oxygen saturation can be total signal] can then be isolated by canceling out the static
derived. This would be adequate in a simple solution components. Dividing the AC level by the DC level at each
containing only reduced and oxygenated Hb. However, wavelength also compensates for changes in incident light
several obstacles to measuring arterial oxygen saturation intensity, removing another complicating variable from the
arise in practice, such as the scatter, reflection, and absor- equation. Performing this operation at each wavelength will
bance of light by other tissue and blood components. To isolate the relative absorbance of reduced and oxygenated
Hb, the major absorbers in arterial blood. The ratio of the red
1 signal (AC 660nm/DC 660nm) to the infrared signal (AC
i
940nm/DC 940nm) reflects a similar ratio of reduced to
(Red) (Infrared) oxygenated Hb, which can then be converted to oxygen
660nm 940nm saturation.

LIMITATIONS
The inherent limitations of pulse oximetry in various
clinical settings must be recognized in order to appropriately
interpret results. The major limitations can be divided into
~02Hb three categories: those arising from calibration assumptions,
optical interference, and signal artifact. Other sources of
potential error also will be examined.
~-1
Calibration Assumptions
Initially the conversion from absorbancy ratios to arterial
oxygen saturation was based directly on Beer-Lambert
calculation, but the effects of reflection and scattering of
-2 I I I I . I light even within the pulsatile fraction of arterial blood led to
600 700 800 900 1000 gross overestimation of oxygen saturation. 3 Better results
Wavelength (nm) have come from using experimentally derived calibration
curves (Figure 3). These curves are based on measurements
FIGURE 1. Hemoglobin extinction curve of oxyhemoglobin in healthy young volunteers after induction of hypoxemia
(HbO~) and reduced hemoglobin (Hb). (Courtesy of Ohmeda, with coincident determination of oxygen saturation by both
Louisville, CO.) pulse oximeter and in vitro laboratory co-oximeter.
JAMES E. SINEX PULSE OXIMETRY: PRINCIPLES AND LIMITATIONS 61

100 determine the relative absorbances of reduced Hb and O2Hb,


the pulse oximeter is unable to distinguish COHb and
80
MetHb. As such, the pulse oximeter measures a functional
saturation:
70 -

6o Functional SaO2 = [O2Hb/(OaHb + Hb)]


o"
o 50 100%. (Equation 1)
[,3
40
A multiwavelength co-oximeter, also known as a labora-
30 tory co-oximeter, uses four different wavelengths in vitro to
20 measure all four species of Hb. When quantities of all four
10
species are known, a fractional saturation can be calculated:
0 Fractional SaO 2
o o o o o o o o o o o o o o o o
Q o o o o o o o o o o o o o o o
= [OzHb/(OzHb -t- Hb + COHb + MetHb)]
c; o c; ,: .-: ,-: ,: ~: & & ~i & & ~ ~ ,4
100%. (Equation 2)
IR
Although levels of COHb and MetHb should be negli-
FIGURE 3. Relationship of red/infrared ratio to oxygen satura- gible in the normal nonsmoking patient, making a functional
tion, as typical pulse oximeter calibration curve. (Reprinted with saturation an adequate approximation of "true" Hb satura-
permission. 6 Courtesy of Ohmeda.) tion, such is not always the case. Figure 4 demonstrates that
both COHb and MetHb will absorb light within the red to
An unavoidable limitation, therefore, is that pulse oxim- near-infrared range used by pulse oximetry. By absorbing
eters can only be as accurate as their empirical calibration light incorrectly attributed to reduced Hb and O2Hb, these
curves. Understandably, researchers were limited in the false absorbers will lead to error in the determination of
degree of hypoxemia inducible in these volunteers, to an relative absorptions of Hb and O2Hb by the pulse oximeter,
Sao2 of approximately 75% to 80%. Therefore, the shape of with propagated error in the displayed functional saturation.
the curve below these levels must be extrapolated, with Therefore, oxygen saturation by pulse oximetry ( S p O 2 ) is
obvious implications for the accuracy of pulse oximetry at only equivalent to functional saturation when insignificant
low oxygen saturations. Early accuracy studies showed such levels of COHb and MetHb are present, and never equiva-
great inaccuracy and bias at low oxygen saturation that lent to fractional saturation, which can only be determined
manufacturers revised early calibration curves and soft- by an in vitro multiwavelength co-oximeter.
ware. 8 More recent studies, however, continue to show The direction and degree of error for each of the
significant bias, increasing as oxygen saturation decreases, 9 dyshemoglobins can be anticipated by examining the Hb
although it has been justifiably pointed out that few, if any, extinction curves. In most nonsmokers, COHb levels will be
clinical treatment decisions will hinge on whether the less than 2%. Heavy smokers have shown levels as high as
oxygen saturation is actually 50% o r 6 0 % . 3 10% to 20%. Levels can be much higher still in significant
Healthy young adults have usually comprised the subject carbon monoxide exposure. Figure 4 shows that COHb
groups used for calibration. One manufacturer was even said absorbs very little light at 940nm. However, at 660nm,
to have used two Olympic athletes in calibration trials. 4 As COHb absorbs light much like O2Hb. Thus by two wave-
such, the applicability of data from such a narrow population length pulse oximetry, COHb will "look like" OzHb in the
to patients at the extremes of ages and with various medical
problems has been questioned.
lo I
Optical Interference
Many substances in the blood, both endogenous and
exogenous, can interfere optically with pulse oximetry, even 1: methemogloNn
when controlling for the contributions of the static compo-
nents of arterial blood. This interference generally takes the
O
form of false absorbers, or components besides reduced Hb
and O2Hb that will absorb light within the red and near- .1 reduced
oglobin
infrared wavelengths used in pulse oximetry.

Dyshemoglobinemias
~ " - - - - - - . - . - . . ~ r earboxyhemaglobfn
b o x y he
The most significant potential false absorbers in the .01 I I I I I I I~""~ I I
600 64O 680 720 760 80O 84O 880 920 960 1000
circulation are carboxyhemoglobin (COHb) and methemo-
Logf Wavelength (nm)
globin (MetHb). To understand the effects of these Hb
species on oximetry, the difference between functional and FIGURE 4. Hemoglobin extinction curve, demonstrating relative
fractional Hb saturation must first be defined. absorptions of the four major classes of hemoglobin. (Reprinted
At the wavelengths of light commonly employed to with permission. 6 Courtesy of Ohmeda.)
62 AMERICAN JOURNAL OF EMERGENCY MEDICINE Volume 17, Number 1 January 1999

red range, with essentially no impact in the infrared. nemia can have high true COHb levels, as carbon monoxide
Therefore, the effect on SpO2 would be to overestimate true is produced with bilirubin in the breakdown of Hb. ~2 This is
oxygen saturation. In experimental carbon monoxide expo- particularly relevant in hemolytic jaundice. Bilirubin can
sure in animals, SpO2 overestimated fi'actional saturation by cause artifactual increases in both COHb and MetHb levels
an amount roughly proportional to the COHb level (or as measured by in vitro co-oximeter because of the overlap
SpO2 = SaO2 + %COHb). ~ In this study, the oximeter read of its absorption spectra with these Hb species. 21 Similar to
up to 90% SpO2 in the presence of a COHb level of 70% and the presence of fetal Fib, then, hyperbilirubinemia is a
an O2Hb level of only 30%. Human case reports support situation in which pulse oximetry can be more accurate than
these findings. 11,I2 More recently, a study of pulse oximetry the in vitro co-oximeter.
in 16 adults with carbon monoxide exposure showed the Intravenous dyes are known to have potentially profound
same degree of overestimation, referred to by the authors as effects on pulse oximetry readings, resulting in falsely low
the "pulse oximetry gap" of CO intoxication.13 measured oxygen saturations. Methylene blue, with very
Likewise, MetHb levels are less than 1% in normal high absorbance at 660nm, causes an impressive spurious
subjects. MetHb is formed as the iron in hemoglobin is decrease in SpO2 .23-25 As little as 5 mL of methylene blue
oxidized from the ferrous to the ferric state. MetHb levels administered to human subjects can decrease SpO2 dramati-
may be high congenitally or as a result of exposure to a cally, in one subject to as low as 1%. 24Lesser decreases from
number of agents, most notably anesthetics, sulfa drugs, and baseline SpOz occur after intravenous administration of
nitrites? 4,15 Figure 4 shows that at 660nm, MetHb looks indocyanine green and indigo carmine. 24 Onset of these
much like reduced Hb. However, more importantly, at changes occurs 30 to 45 seconds after dye administration,
940nm the extinction or absorbance of MetHb is markedly with recovery to baseline oxygen saturation within 3 min-
greater than that of either Hb or O2Hb. As a result, MetHb utes. Dye clearance is prolonged in debilitated, very young,
will contribute greatly to the perceived absorption of both or elderly subjects. In animal studies, administration of
these species, and will increase both the numerator and the fluorescein has been observed to have no effect on pulse
denominator of the ratio of relative absorbances the oxim- oximetryY False desaturation with these intravenous dyes is
eter calculates, driving this ratio towards 1. Note in Figure 3 also observed with the in vitro co-oximeter.
The effects of methylene blue are of interest for other
that when the ratio of pulse-added red absorbance to infrared
reasons as well. This dye is used in the treatment of
absorbance is 1, the calibrated saturation is roughly 85%. 5
methemoglobinemia, which has its own effect on pulse
What are the expected and observed results of this? An
oximetry. Additionally, methylene blue administration can
animal study showed that SpO2 decreased with increasing
cause an initial increase and then decrease in cardiac output,
MetHb levels of up to 40% to a plateau near a pulse oximetry
with resultant independent effects on oxygen deliveryY
reading of 85%. 16As a human example, Eisenkraft reported
a patient with a MetHb level of 5% and a Pao2 of 587 mmHg
on 100% oxygen who showed an SpOz of only 92% to 93%, Signal Artifact
which he attributed solely to methemoglobin effects. ~5 Most commonly, problems in pulse oximetry arise from
Again, these results are consistent with what would be signal artifact. The presence of a sharp pulsatile waveform
expected based on the extinction curves. Furthermore, at a displayed on those oximeter models featuring a plethysmo-
low "true" oxygen saturation in the presence of methemoglo- graph is no guarantee against signal artifact. Signal artifact
binemia, it could be theorized that SpO2 would increase results from false sources of signal or from a low signal-to-
towards 85%, though such has not been clearly demon- noise ratio. False signal can arise from detection of nontrans-
strated. The observation has been made that methemoglobin- mitted light (ambient sources or optical shunt) or from
emia could be suspected based solely on a persistent nonarterial sources of alternating signal. A low signal-to-
unexplained low pulse oximetry saturation. 14,16 noise ratio results from inadequate signal complicated by an
In summary, both COHb and MetHb can have significant excess of physiological or technical noise.
effects on pulse oximetry readings when present in supranor-
real concentrations. When the presence of these abnormal False Signal
Hb species is suspected, pulse oximetry should be supple-
mented by in vitro multiwavelength co-oximetry. However, The simple oximetry system as outlined above incorrectly
it is worth recognizing that in the absence of changing levels assumes that the sum of the light absorbed and the light
of these dyshemoglobins, pulse oximetry maintains its transmitted is equal to the incident fight, with no other light
utility as a monitor of oxygen saturation trends. loss or gain affected the detector. Ambient light, however, is
Fetal Hb has been found to have no significant effect on potentially a major source of interference. Recognizing this,
pulse oximetry. 17,18 This is consistent with the observation designers divided the LED and detector activities into three
that fetal Hb absorbances are the same as those of adult Hb sensing periods, cycling at hundreds of times per second.
values at the wavelengths utilized by pulse oximetry. 19 Two of these periods use light emitted by the LEDs at each
of the two incident wavelengths. In the third period neither
Interestingly, fetal Hb may be misread as COHb by the
LED is activated, and the photodiode detector measures only
particular wavelengths used by the in vitro co-oximeter?
ambient light, the influence of which is subsequently
eliminated from the LED-illuminated sensing periods.
Bilirubin and Intravenous Dyes However, cases of ambient light interference still oc-
Hyperbilirubinemia itself does not appear to interfere with cur. 3'26'27 Implicated sources include fluorescent lighting,
pulse oximetry. 1721,22 However, patients with hyperbilirubi- surgical lamps, fiberoptic instruments, and sunlight. 4,28 Coy-
JAMES E, SINEX PULSE OXIMETRY: PRINCIPLES AND LIMITATIONS 63

ering the probe with an opaque shield offers a simple Low Signal-To-Noise Ratio
solution.
The "pulse search" period when a pulse oximeter probe is
Optical shunting occurs when light from the LED reaches
first placed is spent not only in acquiring a good pulse but
the photodetector without passing through an arterial bed. 29,3
This occurs most commonly with inappropriate probe selec- also in finding an intensity of light sufficient to transmit
tion, as when a digit probe is placed on the ear lobe, or with through the tissue in question? The pulse oximeter will
probe malposition, with a finger probe being "cocked" on increase the amplitude of incident light in a stepwise fashion
the finger. by up to a billion times. 5 Recall that under the best of
In both ambient light interference and optical shunt, the circumstances the pulsatile or AC component of signal
end result is the addition of false signal to both wavelengths comprises only 1% to 5% of total signal. As incident light
tested. 31 As with methemoglobinemia, the result will be a amplitude increases in efforts to pick up AC signal, artifac-
ratio of relative absorbances approaching unity, with a tual signal or noise will likewise be amplified. Although
corresponding displayed SpO2 of 85%. At normal satura- most modern pulse oximeters have complex algorithms to
tions this will manifest as a falsely lowered saturation. More distinguish noise from signal, and will only accept a certain
dangerously, errors caused by ambient light interference and signal-to-noise ratio, at times this noise can be processed and
optical shunt will lead to overestimations of saturation in presented as signal. 34
patients with a true SaO2 of less than 85%. This overestima- Low signal-to-noise ratio results from absent or low-
tion has been demonstrated with probe malposition in amplitude pulses, arising from a multitude of causes includ-
hypoxic patients. 32 ing hypotension, hypovolemia, hypothermia, and peripheral
Nonarterial sources of alternating signal most commonly vascular disease. Iatrogenic causes, including blood pressure
result from motion artifact. Most motion has been said to cuff inflation and administration of vasoconstrictors, will
originate in movement of the oximeter probe in relation to likewise decrease signal. In other words, low signal-to-noise
skin, with the common sense suggestion offered that the ratio can be a potential problem in the majority of seriously
cable be taped to the dorsum of the hand when using digit ill patients, in whom accurate information is most needed.
probes. 33 Shivering is widely believed to interfere, 34 al- The low-perfusion limits of pulse oximetry have been
though pulse oximetry has been found to remain effective studied by inducing extremity hypotension and vasoconstric-
during tonic-clonic seizure. 35 False readings caused by tion in volunteers by various methods, al The failure thresh-
patient motion during cardiopulmonary resuscitation have olds were found to be significantly lower than those
been described. 36 Repetitive cough and the cycling of commonly seen clinically, leading the investigators to
ventilators offer other common potential sources of interfer- suggest that failure to detect signal is more often caused by
ence. Of particular interest to aeromedical practice, several local vasoconstriction rather than systemic hypotension.
oximeters have been tested during helicopter flight, with no So the question becomes what to do in such cases,
apparent effect from vibration. 37-39When interference does
particularly those in which the oximeter is unable to find a
occur from motion artifact it again tends to be additive into
pulse. Obviously, potential underlying cardiorespiratory
both the red and infrared channels, with a resultant absor-
problems need to be addressed first. In an effort to acquire a
bance ratio near 1 and an SpO2 approaching 85%.
suitable pulse oximeter signal, a simple change in probe
Although different oximeter models employ different data
location may suffice. Ear lobe probes have been found in
processing, the dozens of saturation values acquired per
some cases to produce better signal as compared with digit
second are typically averaged over a period of 2 to 16
seconds before a reading is given, serving in part to limit the probes. 3 This is supported by evidence that the ear lobe is
impact of motion artifact. 7 However, this averaging period less vasoactive than the finger pad or nail bed, and so is less
will also delay presentation of data, extending time to first susceptible to vasoconstrictive effects.42
reading as well as response time to saturation changes. The As another alternative, digital blocks have been found to
trade-off between limiting artifact and extending response be quite effective in regaining signal in cases of peripheral
time can be adjusted on many oximeters by variable vasoconstriction.43-45 Such peripheral "clamping down" is
averaging period settings. Oximeters also attempt to mini- probably most often seen in emergency practice in the
mize motion artifact with internal algorithms that eliminate hypothermic patient, including the majority of seriously ill
or differentially weigh outlier readings which are likely due patients who typically are exposed fully for assessment.
to motion. 5 In a similar fashion, some instruments couple Ostensibly, the block works by blocking digital sympathetic
oximetry readings to electrocardiograph signal. 34 innervation, with subsequent relative vasodilatation. The
In addition to extracorporeal motion, there are other digital block itself should not interfere with the detection or
sources of false alternating signal. Arterial pulsations can be correction of any systemic problems that could also underlie
transmitted into the venous circulation in venous congestion, loss of signal. Sometimes simple local heat or massage will
causing artificially lowered oxygen saturation readings. 4 accomplish the same effect. Anecdotally, topical nitroglycer-
This same phenomenon occurs in tricuspid regurgitation. ine has been employed with variable effect.34
Arterial pulsations with a large dicrotic notch, as in aortic To the same end, intraarterial vasodilators have also been
regurgitation, combined aortic stenosis and regurgitation, or used, instilled through ipsilateral radial lines. 46 Diluted
idiopathic hypertophic subaortic stenosis, can be processed boluses of nitroglycerine (10 ~g) and hydralazine (1 nag)
incorrectly at twice the true heart rate. This will have little have been reported to restore pulse oximeter function with
impact on oxygen saturation but has obvious effects on no adverse hemodynamic effect. Clearly, however, this is not
displayed heart rate. 34 a practical intervention in the majority of patients.
64 AMERICAN JOURNAL OF EMERGENCY MEDICINE Volume 17, Number 1 January 1999

Other Limitations supplementary oxygen at high Fio2 and showing a high SpO2
(99% to 100%), there can be a dramatic decrease in Pao2
Anemia appears to adversely affect the accuracy of pulse
before a corresponding decrease in oxygen saturation is
oximetry, although the mechanism is unclear, and it may do
manifested due to the shape of the oxygen-Hb dissociation
so only in the presence of hypoxia. In theory, anemia should
curve.
not affect pulse oximetry at all, as the ratio of relative
absorbances should be preserved and unchanged by changes
in total hemoglobin concentration within the sample. How- ACCURACY
ever, a retrospective study found an underestimation of
oxygen saturation by pulse oximetry in anemic subjects, Hundreds of studies have examined the accuracy of pulse
inversely proportional to Hb concentration and most pro- oximeters, with fair consistency in findings. Reviews have
nounced at an SpO2 less than roughly 80%. 47An experimen- summarized these studies nicely and have illustrated the
tal in vitro study48 and an animal study49 both showed a difficulty in answering the seemingly simple question of
similar trend with coincident anemia and hypoxemia. The accuracy.3,5,6,34,58
observation has been made that the direction of error is In general, manufacturers presently report pulse oximeter
fortuitous, presenting an exaggeration of apparent hypox- accuracy to one standard deviation of <+__3% at arterial
emia, as anemic patients are more susceptible to the oxygen saturations of >70%. These claims are largely
deleterious effects of desaturation. 47 With normal oxygen supported in the most recent reviews of studies addressing
saturation, an earlier dog study found pulse oximetry to be accuracy. 34,58It should be recognized, however, that this bias
consistently accurate with a near-zero bias in subjects with a or error is reported at one standard deviation only, and so
hematocrit of at least 10, but found a bias of 5.4% in subjects must be doubled or tripled to express a 95% or 99%
with a hematocrit less than 10. 5 Encouragingly, a recent confidence interval, respectively. This may seem a surpris-
study of nonhypoxic human patients with acute anemia from ingly wide range of error and is probably contrary to most
hemorrhage down to a Hb value as low as 2.3 g/dL showed clinician's preconceptions of accuracy. This only under-
good accuracy with pulse oximetry.51 scores the importance of widespread recognition of that
Skin pigmentation has shown variable effects on pulse range. Again, this range of error must be reassessed in every
oximetry.5244 Primarily, dark pigmentation appears to make situation in light of the limitations as previously discussed.
adequate light penetration more difficult, with significantly Methodological issues should be briefly discussed to put
more signal detection failures. It has been suggested that the the results of these studies into proper context. Subject
less pigmented nail beds may offer particularly good probe groups have been highly variable, ranging from healthy
sites in these patients. 34More concerning are some studies of volunteers to the critically ill. Correspondingly, study set-
pigmentation effects which have shown overestimations of tings have varied from rigidly controlled laboratories in
oxygen saturation, both in models 54 and in human subjects. 5a some studies to intensive care units and emergency depart-
The mechanism of such artifact, if present, is unclear, and ments in others.
could arise from optical interference, low signal-to-noise The "gold standard" of comparison now is generally
effects, or as a reflection of the relative lack of darkly recognized to be the in vitro four-wavelength co-oximeter,
pigmented subjects in calibration trials. Studies with larger with a reported accuracy for fractional saturation of ___1% at
patient groups are indicated. two standard deviations. 3 Some studies, however, have
Similarly, there is evidence for the effects of nail polish on utilized two wavelength in vitro co-oximeters as standards,
pulse oximetry, but little consensus on occurrence or de- with the same limitations inherent in use of only two
gree. 55-57When demonstrated, most interference appears to wavelengths as pulse oximeters. Still others have used
arise from blue or black polish. It is probably advisable to calculated SaO2 from arterial blood gas samples. The
avoid possible interference by using other probe sites, problem with such calculated oxygen saturations is that,
removing the nail polish, or even simply placing the probe because of shifts in the O2Hb dissociation curve, correction
sideways on the digit to remove the nail from the transmis- factors for variables like temperature, pH, and 2,3-
sion path. diphosphoglycerate concentration must be inserted for accu-
Although LEDs are reliable sources of narrow wavelength rate results. Most often, however, standard corrections are
light in most cases, there can be slight variability between used with potential resulting impact on accuracy.
LEDs in actual wavelength emitted. 5 A very small error in Data analyses also have been variable, with early studies
wavelength can translate into a large error in absorbance employing primarily regression analysis and correlation
interpretation, particularly at the steep portions of the curve coefficients. Such measures were found to be inadequate,
in the red range (Figure 1). In general, LEDs that emit light reflecting a known association between Sao2 and SpO2
too far from standard are rejected by the manufacturer. rather than addressing accuracy. A better measure has been
However, some manufacturers have dealt with this problem bias, or the mean value of (Sa02 - SpO2), in conjunction
in the past by compensatory changes in software and data with precision, or the standard deviation of bias. 3,5,34 Bias
processing. The danger arises in switching probes in such represents systematic differences between methods of mea-
cases, coupling an LED with processing software calibrated surement, with precision reflecting variability.
to a different wavelength of incident light. This problem Revisions in manufacturer algorithms and software, at
most likely will not arise in modern oximeters, but users times prompted by the findings of the studies themselves,
should be aware nonetheless. further complicate accuracy assessments. As such, results
Finally, it should be recognized that in patients receiving can become dated almost by nature of their own existence.
JAMES E. SINEX PULSE OXIMETRY: PRINCIPLES AND LIMITATIONS 65

APPLICATIONS in which other methods of assessing blood pressure are often


difficult or impossible. 38,39It must be stressed, however, that
Potential applications of pulse oximetry in emergency
inflation or deflation must be done very slowly because of
medicine are widespread. In general, the pulse oximeter has
some delay in response from the pulse oximeter (2 to 3
found a role in every situation in which information on
mmHg per second).
oxygenation status would prove helpful. The utility of the
oximeter in this role is fairly obvious. However, some have
used the plethysmographic, or pulse detection, feature of the FUTURE TRENDS
oximeter for various situations as well. At present, the pulse oximeters generally avai!able for use
in critical care and emergency department settings all
Oximetric employ transmitted light. Reflectance oximeters, which
detect "back-scatter" of light from LEDs placed side-by-
The pulse oximeter has found its greatest utility as a side with detectors, are already in use in anesthesia and in
monitor for desaturation or hypoxia, and in this capacity the obstetrics (with placement of a probe on the fetal presenting
oximeter is an excellent tool. With continuous monitoring in part). Potential advantages of reflectance operation include
the emergency setting, the pulse oximeter can rapidly detect the ability to place the probe on virtually any site on the
progression of hypoxia and reflect efficacy of treatment. 59 body, and to demonstrate an earlier response to hypoxia with
Oximeter monitoring has been shown to reduce both the placement of a probe more centrally, on the torso. Further-
frequency of hypoxic episodes and the duration of hypoxia more, reflection oximetry of retinal blood has been exam-
during emergency endotracheal intubation. 6 Unfortunately, ined as a measure of cerebral oxygenation,73 and an instru-
the tendency in all but the most critically ill patients is to use ment that continuously measures both saturation and
the pulse oximeter as a "spot-check" of oxygen saturation. perfusion has been developed by combining a reflectance
The pulse oximeter certainly can be used in this fashion, and oximeter with a laser Doppler flowmeter.74 However, cur-
has been shown to detect clinically unrecognized hypoxia in rently available reflectance oximeters suffer in comparison
emergency department patients. 61-63 However, in this role, with transmission oximetry in terms of accuracy and suscep-
questions of accuracy loom larger, and as discussed the tibility to noise. 58,75,76As such, they are not yet appropriate
oximeter is subject to many limitations and sources of for emergency practice.
artifact. It cannot be stressed enough that the clinician must
critically assess the results given by the pulse oximeter in REFERENCES
every situation. It further bears mentioning that the pulse
oximeter assesses the adequacy of oxygenation, but in no 1. Rodriguez LR, Kotin N, Lowenthal D, Kattan M: A study of
way measures adequacy of ventilation. pediatric house staff's knowledge of pulse oximetry. Pediatrics
1994;93:810-813
2. Stoneham MD, Saville GM, Wilson IH: Knowledge about pulse
Plethysmographic oximetry among medical and nursing staff. Lancet 1994;344:1339-
1342
The use of the oximeter as a plethysmograph is much 3. Kelleher JF: Pulse oximetry. J Clin Monit 1989;5:37-62
more controversial. Its ability to detect nonpalpable pulses 4. Grace RF: Pulse oximetry. Gold standard or false sense of
secu rity? Med J Aust 1994; 160:638-644
has been used as an assessment of perfusion in a number of 5. Tremper KK, Barker SJ: Pulse oximetry. Anesthesiology 1989;
situations, ranging from verification of adequacy of chest 70:98-108
compressions during cardiopulmonary resuscitation 64 and 6. Wukitsch MW, Petterson MT, Tobler DR, Pologe JA: Pulse
assessing ulnar collateral flow in a modified Allen's test 65-67 oximetry: analysis of theory, technology, and practice. J Clin Monit
1988;4:290-301
to assessing vitality of replanted and revascularized extremi-
7. Kidd JF, Vickers MD: Pulse oximeters: essential monitors with
ties.68 limitations. Br J Anaesth 1989;62:355-357
The critical question then is whether a pulsatile signal 8. Severinghaus JW, Naifeh KH: Accuracy of response of six pulse
sufficient to generate a pulse waveform on the oximeter can oximeters to profound hypoxemia. Anesthesiology 1987;67:551-558
serve as a guarantee of adequate perfusion. Most evidence is 9. Thrush D, Hodges MR: Accuracy of pulse oximetry during
hypoxemia. South Med J 1994;87:518-521
only anecdotal, so this question remains largely unan- 10. Barker SJ, Tremper KK: The effect of carbon monoxide
swered. 34 Of particular concern is the experimental finding inhalation on pulse oximetry and transcutaneous PO2. Anesthesiol-
that blood flows determined by laser Doppler to be as low as ogy 1987;66:677-679
4% to 8.6% of baseline were detected as pulsatile signal by a 11. Seidler D, Hirschl MM, Roeggla G: Limitations of pulse
pulse oximeter. 69 At the current state of knowledge, then, oximetry. Lancet 1993;341:1600-1601 (letter; comment)
12. Vegfors M, Lennmarken C: Carboxyhaemoglobinaemia and
such uses of the pulse oximeter should be met with pulse oximetry. Br J Anaesth 1991 ;66:625-626
skepticism and probably avoided. It should be recognized as 13. Buckley RG, Aks SE, Eshom JL, et al: The pulse oximetry gap
well that the pulsatile signal displayed by the oximeter is in carbon monoxide intoxication. Ann Emerg Med 1994;24:252-255
usually automatically scaled as the oximeter increases 14. Anderson ST, Hajduczek J, Barker SJ: Benzocaine-induced
methemoglobinemia in an adult: accuracy of pulse oximetry with
incident light amplitude in an effort to detect signal and so methemoglobinemia. Anesth Analg 1988;67:1099-1101
does not reflect absolute pulse amplitude. 15. Eisenkraft JB: Pulse oximeter desaturation due to methemoglo-
One possible exception is the determination of systolic binemia. Anesthesiology 1988;68:279-282
blood pressure, either by reappearance of pulsatile wave- 16. Barker SJ, Tremper KK, Hyatt J: Effects of methemoglobin-
form during cuff deflation38,39,7 or by disappearance of emia on pulse oximetry and mixed venous oximetry. Anesthesiology
1989;70:112-117
waveform during inflation. 41,71,72 The pulse oximeter may 17. Anderson JV: The accuracy of pulse oximetry in neonates:
find particular utility in this role during helicopter transport, effects of fetal hemoglobin and bilirubin. J Perinatol 1987;7:323
66 AMERICAN JOURNAL OF EMERGENCY MEDICINE Volume 17, Number 1 January 1999

18. Ramanathan R, Durand M, Larrazabal C: Pulse oximetry in 48. Kolesar J, Volgyesi G, Lerman J: Effect of haemoglobin
very low birth weight infants with acute and chronic disease. Pediat- concentration on the accuracy of pulse oximetry. Can J Anaesth
rics 1987;79:612-616 1990;37:$88 (abstr)
19. HarrisAP, Sendak MJ, Donham RT, et al: Absorption character- 49. Vegfors M, Lindberg LG, Oberg PA, Lennmarken C: The
istics of human fetal hemoglobin at wavelengths used in pulse accuracy of pulse oximetry at two haematocrit levels. Acta Anaesthe-
oximetry. J Clin Monit 1988;4:175-177 siol Scand 1992;36:454-459
20. Pologe J, Raley D: Effects of fetal hemoglobin on pulse 50. Lee SE, Tremper KK, Barker SJ: Effects of anemia on pulse
oximetry. J Perinato11987;7:324-326 oximetry and continuous mixed venous oxygen saturation monitoring
21. Beall SN, Moorthy SS: Jaundice, oximetry, and spurious in dogs. Anesth Analg 1988;67:$130
hemoglobin desaturation. Anesth Analg 1989;68:806-807 51. Jay GD, Hughes L, Renzi FP: Pulse oximetry is accurate in
22. Veyckemans F, Baele P, Guillaume JE, et al: Hyperbilirubine- acute anemia from hemorrhage. Ann Emerg Med 1994;24:32-35
mia does not interfere with hemoglobin saturation measured by pulse 52. Cahan C, Decker MJ, Hockje PL, Strohl KP: Agreement
oximetry. Anesthesiology 1989;70:118-122 between noninvasive oximetric values for oxygen saturation. Chest
23. Kessler MR, Eide T, Humayan B, Poppers PJ: Spurious pulse 1990;97:814-819
oximeter desaturation with methylene blue injection. Anesthesiology 53. Ries AL, Prewitt LM, Johnson JJ: Skin color and ear oximetry.
1986;65:435-436
Chest 1989;96:287-290
24. Scheller MS, Unger RJ, Keiner MJ: Effects of intravenously
administered dyes on pulse oximetry readings. Anesthesiology 1986; 54. Volgyesi GA, Spahr-Schopfer I: Does skin pigmentation affect
65:550-552 the accuracy of pulse oximetry? An in vitro study. Anesthesiology
25. Sidi A, Paulus DA, Rush W, et al: Methylene blue and 1991 ;75:A406
indocyanine green artifactually lower pulse oximetry readings of 55. Cote CJ, Goldstein EA, Fuchsman WH, Hoaglin DJ: The effect
oxygen saturation. J Clin Monit 1987;3:249-256 of nail polish on pulse oximetry. Anesth Analg 1988;67:683-686
26. Hanowell L, Eisele JH, Downs D: Ambient light affects pulse 56. Kataria BK, Lampkins R: Nail polish does not affect pulse
oxirneters. Anesthesiology 1987;67:864-865 oximetry saturation. Anesth Analg 1986;65:824 (letter)
27. Poets CF, Seidenberg J, von der Hardt H: Failure of pulse 57. Rubin AS: Nail polish color can affect pulse oximetry. Anesthe-
oximeter to detect sensor detachment. Lancet 1993;341:244 (letter; siology 1988;68:825 (letter)
comment) 58. Severinghaus JW: History and recent developments in pulse
28. Costarino AT, Davis DA, Keon TO: Falsely normal saturation oximetry. Scand J Clin Lab Invest 1993;53:214:105-111 (suppl)
reading with the pulse oximeter. Anesthesiology 1987;67:830-831 59. Jones J, Heiselman D, Cannon L, Gradisek R: Continuous
29. Southall DP, Samuels M: Inappropriate sensor application in emergency department monitoring of arterial saturation in adult
pulse oximetry. Lancet 1992;340:481-482 (letter) patients with respiratory distress. Ann Emerg Med 1988;17:463-468
30. Gardosi JO, Damianou D, Schram CM: Inappropriate sensor 60. Mateer JR, Olson DW, Stueven HA, Aufderheide TP: Continu-
application in pulse oximetry. Lancet 1992;340:920 (letter; comment) ous pulse oximetry during emergency endotracheal intubation. Ann
31. Kelleher JF, Ruff RH: The penumbra effect: Vasomotion- Emerg Med 1993;22:675-679
dependent pulse oximeter artifact due to probe malposition. Anesthe- 61. Holburn CJ, Allen MJ: Pulse oximetry in the accident and
siology 1989;71:787-791 emergency department. Arch Emerg Med 1989;6:137-142
32. Barker SJ, Hyatt J, Shah NK, Kao J: The effect of sensor 62. Lambert MA, Crinnion J: The role of pulse oximetry in the
malpositioning on pulse oximeter accuracy during hypoxemia. Anes- accident and emergency department. Arch Emerg Med 1989;6:211-
thesiology 1993;79:248-254 215
33. Hanning CD, Alexander-Williams JM: Pulse oximetry: A practi- 63. Mower MR, Sachs C, Nicklin EL, et al: Effect of routine
cal review. BMJ 1995;311:367-370 emergency department triage pulse oximetry screening on medical
34. Severinghaus JW, Kelleher JF: Recent developments in pulse management. Chest 1995;108:1297-1302
oximetry. Anesthesiology 1992;76:1018-1038 64. Narang VPS: Utility of the pulse oximeter during cardiopulmo-
35. James MR, Marshall H, Carew-McColl M: Pulse oximetry nary resuscitation. Anesthesiology 1986;65:239-240 (letter)
during apparent tonic-clonic seizures. Lancet 1991 ;337:394-395 65. Levinsohn DG, Gordon L, Sessler DI: The Allen's test: Analysis
36. Moorthy SS, Dierdorf SF, Schmidt SI: Erroneous pulse oxim- of four methods. J Hand Surg [Am] 1991 ;16A:279-282
eter data during CPR. Anesth Analg 1990;70:339 (letter) 66. O'Mara K, Sullivan B: A simple bedside test to identify ulnar
37. Short L, Hecker RB, Middaugh RE, Menk EJ: A comparison of collateral flow. Ann Intern Med 1995; 123:637
pulse oximeters during helicopter flight. J Emerg Med 1989;7:639- 67. Fuhrman TM, McSweeney E: Noninvasive evaluation of the
643 collateral circulation of the hand. Acad Emerg Med 1995;2:195-199
38. Talke P, Nichols RJ, Traber DL: Evaluation of pulse oximetry 68. Graham B, Paulus DA, Caffee HH: Pulse oximetry for vascular
for intraoperative blood pressure measurement and vital sign monitor- monitoring in upper extremity replantation surgery. J Hand Surg [Am]
ing during patient transport via life flight. Anesth Analg 1988;67:$229
1986;11A:687-692
(abstr) 69. Lawson D, Noriey I, Korbon G, et al: Blood flow limits and
39. Talke P, Nichols RJ, Traber DL: Monitoring patients during
pulse oximeter signal detection. Anesthesiology 1987;67:599-603
helicopter flight. J Clin Monit 1990;6:139-140 70. Talke P, Nichols RJ Jr, Traber DL: Does systolic blood
40. Schnapp LM, Cohen NH: Pulse oximetry--uses and abuses.
pressure measured using a pulse oximeter correlate with conven-
Chest 1990;98:1244-1250
tional methods? J Clin Monit 1990;6:5-9
41. Severinghaus JW, Spellman MJ: Pulse oximeter failure thresh-
71. Wallace CT, Baker JD 3d, Alpert CC, et al: Comparison of
olds in hypotension and vasoconstriction. Anesthesiology 1990;73:
blood pressure measurement by Doppler and by pulse oximetry
532-537
42. Evans ML, Geddes LA: An assessment of blood vessel techniques. Anesth Analg 1987;66:1018-1019
vasoactivity using photoplethysmography. Med Instrument 1988;22: 72. Chawla R, Kumarvel V, Girdhar KK, et al: Can pulse oximetry
29-32 be used to measure systolic blood pressure? Anesth Analg 1992;74:
43. Bourke DL, Grayson RF: Digital nerve blocks can restore 196-200
pulse oximeter signal detection. Anesth Analg 1991 ;73:815-817 73. de Kock JP, Tarassenko L, Glynn CJ, Hill AR: Reflectance
44. Gentili ME, Chevaleraud E, Viel E: Digital block of the flexor pulse oximetry readings from the retinal fundus. IEEE Trans Biomed
tendon sheathe can restore pulse oximeter signal detection. Reg Eng 1993;40:817-823
Anesth 1995;20:82-83 (letter) 74. Dougherty G, Lowry J: Design and evaluation of an instrument
45. Grayson RF, Bourke DL: Digital block for pulse oximetry to measure microcirculatory blood flow and oxygen saturation simul-
failure. Anesthesiology 1991 ;75:A406 (abstr) taneously. J Med Eng Technol 1992;16:123-128
46. Holroyd K, Lui M, Beattie C: Intraarterial vasodilator adminis- 75. Jorgensen JS, Schmid ER, Konig V, et a1: Limitations of
tration to restore pulse oximeter function. Anesthesiology 1993;79: forehead pulse oximetry. J Clin Monit 1995;11:253-256
388-390 76. Lindberg LG, Lennmarken C, Vegfors M: Pulse oximetry--
47. Severinghaus JW, Koh SO: Effect of anemia on pulse oximeter clinical implications and recent technical developments. Acta Anaes-
accuracy at low saturation. J Clin Monit 1990;6:85-88 thesiol Scand 1995;39:279-287

You might also like