Professional Documents
Culture Documents
The pulse oximeter has become an essential tool in the modern practice the blood was carried by its colored component. 3 This
of emergency medicine. However, despite the reliance placed on the component was isolated the following year by Hoppe-
information this monitor offers, the underlying principles and associated Seyler, who further noted that the pattern of light absorption
limitations of pulse oximetry are poorly understood by medical practitio- of this substance he dubbed "haemoglobin" changed when
ners. This article reviews the principles of pulse oximetry, with an eye shaken in air. 4 Over the following decades, the optical
toward recognizing the limitations of this tool. Among these are perfor-
mance limitations in the settings of carboxyhemoglobinemia, methemo- spectra of the reduced and oxygenated species of hemoglo-
globinemia, motion artifact, hypotension, vasoconstriction, and anemia. bin (Hb) were further studied and elucidated.
The accuracy of pulse oximetry is discussed in light of these factors, with The first system to use transillumination to measure in
further discussion of applications for pulse oximetry in emergency vivo oxygen saturation was developed in 1935 by Matthes,
medicine, including both oximetric and plethysmographic operation. The though difficulties with calibration made it an unwieldy
pulse oximeter is an invaluable instrument for emergency medicine device. Wood in 1948 employed a refined version of this
practice, but as with any test the data it offers must be critically appraised "oximeter" as an anesthesia monitoring device. Unfortu-
for proper interpretation and utilization. (Am J Emerg Med 1999;17:59-67. nately, it too was fraught with practical limitations, requiring
Copyright 1999 by W.B. Saunders Company) individual calibration and a dedicated technician to operate. 5
The greatest step forward in the development of the
Pulse oximetry has become an essential tool in the current modern pulse oximeter occurred in 1974, interestingly
practice of emergency medicine. Detection of hypoxia by enough as an incidental finding. Before this, in order to
clinical indicators alone is notoriously unreliable, whereas isolate arterial blood for transillumination, oximeters had
the pulse oximeter offers real-time assessment of oxygen- relied on compression and heating of the ear lobe to remove
ation status on a moment-to-moment basis, reflecting effi- venous and capillary blood and to "arterialize" the tissue, at
cacy of interventions as well as progression of disease times with resultant second-degree burns. In 1974 Aoyagi,
processes. This technology is noninvasive, widely available, while trying to develop a dye-dilution technique for cardiac
and simple to use. Proper interpretation of pulse oximetry output determination, noted that the washout curves he was
data is dependent on a basic understanding of the underlying measuring with an ear densitometer were muddied by
principles and inherent limitations of the instrument. Unfor- pulsatile variations. In working to eliminate these variations,
tunately, studies have shown a lack of knowledge among Aoyagi recognized that the absorbency ratios of these
both residents and medical staff concerning the limitations pulsations at different wavelengths varied with oxygen
and interpretation of pulse oximetry results? ,2 This impor- saturation. Employing the principles discussed in the next
tant topic has not been widely addressed in the emergency section, Aoyagi developed and marketed the first pulse
medicine literature. This article reviews the principles and oximeter in 1975. 3
limitations of pulse oximetry, with discussion of the accu- The later development of reliable light-emitting diodes
racy and appropriate clinical uses of this invaluable tool. (LEDs), photodetectors, and microprocessors ushered in the
current era of modern pulse oximetry. The availability and
HISTORY use of this instrument boomed through the 1980s, initially in
the hands of anesthesiologists, and then throughout medical
The development of pulse oximetry has been based on practice.
more than a hundred years of experimental and engineering
antecedents. Stokes in 1864 first recognized that oxygen in
PRINCIPLES
The principle of pulse oximetry is that of spectral
From the Departmentof EmergencyMedicine, Universityof Cincin- analysis: the detection and quantitation of components in
nati, Cincinnati, OH, and the Department of Emergency Medicine,
Medical College of Georgia,Augusta, GA. solution by their unique light absorption characteristics. The
Manuscript receivedAugust 1, 1997; acceptedAugust 21, 1997. central tenet of spectral analysis is the Beer-Lambert law,
Address reprint requests to Dr Sinex, Department of Emergency which states that the concentration of an absorbing sub-
Medicine, AF-2018, Medical College of Georgia, 1120 15th St, stance in solution can be determined from the intensity of
Augusta, GA30912. light transmitted through that solution, given the intensity
Key Words:Oximetry,physiologic monitoring, anoxia, plethysmog-
raphy. and wavelength of incident light, the transmission path
Copyright 1999 by W.B. Saunders Company length, and the characteristic absorbance of that substance at
0735-6757/99/1701-0019510.00/(3 a specific wavelength (the extinction coefficient).5 Restated,
59
60 AMERICAN JOURNAL OF EMERGENCY MEDICINE Volume 17, Number 1 January 1999
LIMITATIONS
The inherent limitations of pulse oximetry in various
clinical settings must be recognized in order to appropriately
interpret results. The major limitations can be divided into
~02Hb three categories: those arising from calibration assumptions,
optical interference, and signal artifact. Other sources of
potential error also will be examined.
~-1
Calibration Assumptions
Initially the conversion from absorbancy ratios to arterial
oxygen saturation was based directly on Beer-Lambert
calculation, but the effects of reflection and scattering of
-2 I I I I . I light even within the pulsatile fraction of arterial blood led to
600 700 800 900 1000 gross overestimation of oxygen saturation. 3 Better results
Wavelength (nm) have come from using experimentally derived calibration
curves (Figure 3). These curves are based on measurements
FIGURE 1. Hemoglobin extinction curve of oxyhemoglobin in healthy young volunteers after induction of hypoxemia
(HbO~) and reduced hemoglobin (Hb). (Courtesy of Ohmeda, with coincident determination of oxygen saturation by both
Louisville, CO.) pulse oximeter and in vitro laboratory co-oximeter.
JAMES E. SINEX PULSE OXIMETRY: PRINCIPLES AND LIMITATIONS 61
Dyshemoglobinemias
~ " - - - - - - . - . - . . ~ r earboxyhemaglobfn
b o x y he
The most significant potential false absorbers in the .01 I I I I I I I~""~ I I
600 64O 680 720 760 80O 84O 880 920 960 1000
circulation are carboxyhemoglobin (COHb) and methemo-
Logf Wavelength (nm)
globin (MetHb). To understand the effects of these Hb
species on oximetry, the difference between functional and FIGURE 4. Hemoglobin extinction curve, demonstrating relative
fractional Hb saturation must first be defined. absorptions of the four major classes of hemoglobin. (Reprinted
At the wavelengths of light commonly employed to with permission. 6 Courtesy of Ohmeda.)
62 AMERICAN JOURNAL OF EMERGENCY MEDICINE Volume 17, Number 1 January 1999
red range, with essentially no impact in the infrared. nemia can have high true COHb levels, as carbon monoxide
Therefore, the effect on SpO2 would be to overestimate true is produced with bilirubin in the breakdown of Hb. ~2 This is
oxygen saturation. In experimental carbon monoxide expo- particularly relevant in hemolytic jaundice. Bilirubin can
sure in animals, SpO2 overestimated fi'actional saturation by cause artifactual increases in both COHb and MetHb levels
an amount roughly proportional to the COHb level (or as measured by in vitro co-oximeter because of the overlap
SpO2 = SaO2 + %COHb). ~ In this study, the oximeter read of its absorption spectra with these Hb species. 21 Similar to
up to 90% SpO2 in the presence of a COHb level of 70% and the presence of fetal Fib, then, hyperbilirubinemia is a
an O2Hb level of only 30%. Human case reports support situation in which pulse oximetry can be more accurate than
these findings. 11,I2 More recently, a study of pulse oximetry the in vitro co-oximeter.
in 16 adults with carbon monoxide exposure showed the Intravenous dyes are known to have potentially profound
same degree of overestimation, referred to by the authors as effects on pulse oximetry readings, resulting in falsely low
the "pulse oximetry gap" of CO intoxication.13 measured oxygen saturations. Methylene blue, with very
Likewise, MetHb levels are less than 1% in normal high absorbance at 660nm, causes an impressive spurious
subjects. MetHb is formed as the iron in hemoglobin is decrease in SpO2 .23-25 As little as 5 mL of methylene blue
oxidized from the ferrous to the ferric state. MetHb levels administered to human subjects can decrease SpO2 dramati-
may be high congenitally or as a result of exposure to a cally, in one subject to as low as 1%. 24Lesser decreases from
number of agents, most notably anesthetics, sulfa drugs, and baseline SpOz occur after intravenous administration of
nitrites? 4,15 Figure 4 shows that at 660nm, MetHb looks indocyanine green and indigo carmine. 24 Onset of these
much like reduced Hb. However, more importantly, at changes occurs 30 to 45 seconds after dye administration,
940nm the extinction or absorbance of MetHb is markedly with recovery to baseline oxygen saturation within 3 min-
greater than that of either Hb or O2Hb. As a result, MetHb utes. Dye clearance is prolonged in debilitated, very young,
will contribute greatly to the perceived absorption of both or elderly subjects. In animal studies, administration of
these species, and will increase both the numerator and the fluorescein has been observed to have no effect on pulse
denominator of the ratio of relative absorbances the oxim- oximetryY False desaturation with these intravenous dyes is
eter calculates, driving this ratio towards 1. Note in Figure 3 also observed with the in vitro co-oximeter.
The effects of methylene blue are of interest for other
that when the ratio of pulse-added red absorbance to infrared
reasons as well. This dye is used in the treatment of
absorbance is 1, the calibrated saturation is roughly 85%. 5
methemoglobinemia, which has its own effect on pulse
What are the expected and observed results of this? An
oximetry. Additionally, methylene blue administration can
animal study showed that SpO2 decreased with increasing
cause an initial increase and then decrease in cardiac output,
MetHb levels of up to 40% to a plateau near a pulse oximetry
with resultant independent effects on oxygen deliveryY
reading of 85%. 16As a human example, Eisenkraft reported
a patient with a MetHb level of 5% and a Pao2 of 587 mmHg
on 100% oxygen who showed an SpOz of only 92% to 93%, Signal Artifact
which he attributed solely to methemoglobin effects. ~5 Most commonly, problems in pulse oximetry arise from
Again, these results are consistent with what would be signal artifact. The presence of a sharp pulsatile waveform
expected based on the extinction curves. Furthermore, at a displayed on those oximeter models featuring a plethysmo-
low "true" oxygen saturation in the presence of methemoglo- graph is no guarantee against signal artifact. Signal artifact
binemia, it could be theorized that SpO2 would increase results from false sources of signal or from a low signal-to-
towards 85%, though such has not been clearly demon- noise ratio. False signal can arise from detection of nontrans-
strated. The observation has been made that methemoglobin- mitted light (ambient sources or optical shunt) or from
emia could be suspected based solely on a persistent nonarterial sources of alternating signal. A low signal-to-
unexplained low pulse oximetry saturation. 14,16 noise ratio results from inadequate signal complicated by an
In summary, both COHb and MetHb can have significant excess of physiological or technical noise.
effects on pulse oximetry readings when present in supranor-
real concentrations. When the presence of these abnormal False Signal
Hb species is suspected, pulse oximetry should be supple-
mented by in vitro multiwavelength co-oximetry. However, The simple oximetry system as outlined above incorrectly
it is worth recognizing that in the absence of changing levels assumes that the sum of the light absorbed and the light
of these dyshemoglobins, pulse oximetry maintains its transmitted is equal to the incident fight, with no other light
utility as a monitor of oxygen saturation trends. loss or gain affected the detector. Ambient light, however, is
Fetal Hb has been found to have no significant effect on potentially a major source of interference. Recognizing this,
pulse oximetry. 17,18 This is consistent with the observation designers divided the LED and detector activities into three
that fetal Hb absorbances are the same as those of adult Hb sensing periods, cycling at hundreds of times per second.
values at the wavelengths utilized by pulse oximetry. 19 Two of these periods use light emitted by the LEDs at each
of the two incident wavelengths. In the third period neither
Interestingly, fetal Hb may be misread as COHb by the
LED is activated, and the photodiode detector measures only
particular wavelengths used by the in vitro co-oximeter?
ambient light, the influence of which is subsequently
eliminated from the LED-illuminated sensing periods.
Bilirubin and Intravenous Dyes However, cases of ambient light interference still oc-
Hyperbilirubinemia itself does not appear to interfere with cur. 3'26'27 Implicated sources include fluorescent lighting,
pulse oximetry. 1721,22 However, patients with hyperbilirubi- surgical lamps, fiberoptic instruments, and sunlight. 4,28 Coy-
JAMES E, SINEX PULSE OXIMETRY: PRINCIPLES AND LIMITATIONS 63
ering the probe with an opaque shield offers a simple Low Signal-To-Noise Ratio
solution.
The "pulse search" period when a pulse oximeter probe is
Optical shunting occurs when light from the LED reaches
first placed is spent not only in acquiring a good pulse but
the photodetector without passing through an arterial bed. 29,3
This occurs most commonly with inappropriate probe selec- also in finding an intensity of light sufficient to transmit
tion, as when a digit probe is placed on the ear lobe, or with through the tissue in question? The pulse oximeter will
probe malposition, with a finger probe being "cocked" on increase the amplitude of incident light in a stepwise fashion
the finger. by up to a billion times. 5 Recall that under the best of
In both ambient light interference and optical shunt, the circumstances the pulsatile or AC component of signal
end result is the addition of false signal to both wavelengths comprises only 1% to 5% of total signal. As incident light
tested. 31 As with methemoglobinemia, the result will be a amplitude increases in efforts to pick up AC signal, artifac-
ratio of relative absorbances approaching unity, with a tual signal or noise will likewise be amplified. Although
corresponding displayed SpO2 of 85%. At normal satura- most modern pulse oximeters have complex algorithms to
tions this will manifest as a falsely lowered saturation. More distinguish noise from signal, and will only accept a certain
dangerously, errors caused by ambient light interference and signal-to-noise ratio, at times this noise can be processed and
optical shunt will lead to overestimations of saturation in presented as signal. 34
patients with a true SaO2 of less than 85%. This overestima- Low signal-to-noise ratio results from absent or low-
tion has been demonstrated with probe malposition in amplitude pulses, arising from a multitude of causes includ-
hypoxic patients. 32 ing hypotension, hypovolemia, hypothermia, and peripheral
Nonarterial sources of alternating signal most commonly vascular disease. Iatrogenic causes, including blood pressure
result from motion artifact. Most motion has been said to cuff inflation and administration of vasoconstrictors, will
originate in movement of the oximeter probe in relation to likewise decrease signal. In other words, low signal-to-noise
skin, with the common sense suggestion offered that the ratio can be a potential problem in the majority of seriously
cable be taped to the dorsum of the hand when using digit ill patients, in whom accurate information is most needed.
probes. 33 Shivering is widely believed to interfere, 34 al- The low-perfusion limits of pulse oximetry have been
though pulse oximetry has been found to remain effective studied by inducing extremity hypotension and vasoconstric-
during tonic-clonic seizure. 35 False readings caused by tion in volunteers by various methods, al The failure thresh-
patient motion during cardiopulmonary resuscitation have olds were found to be significantly lower than those
been described. 36 Repetitive cough and the cycling of commonly seen clinically, leading the investigators to
ventilators offer other common potential sources of interfer- suggest that failure to detect signal is more often caused by
ence. Of particular interest to aeromedical practice, several local vasoconstriction rather than systemic hypotension.
oximeters have been tested during helicopter flight, with no So the question becomes what to do in such cases,
apparent effect from vibration. 37-39When interference does
particularly those in which the oximeter is unable to find a
occur from motion artifact it again tends to be additive into
pulse. Obviously, potential underlying cardiorespiratory
both the red and infrared channels, with a resultant absor-
problems need to be addressed first. In an effort to acquire a
bance ratio near 1 and an SpO2 approaching 85%.
suitable pulse oximeter signal, a simple change in probe
Although different oximeter models employ different data
location may suffice. Ear lobe probes have been found in
processing, the dozens of saturation values acquired per
some cases to produce better signal as compared with digit
second are typically averaged over a period of 2 to 16
seconds before a reading is given, serving in part to limit the probes. 3 This is supported by evidence that the ear lobe is
impact of motion artifact. 7 However, this averaging period less vasoactive than the finger pad or nail bed, and so is less
will also delay presentation of data, extending time to first susceptible to vasoconstrictive effects.42
reading as well as response time to saturation changes. The As another alternative, digital blocks have been found to
trade-off between limiting artifact and extending response be quite effective in regaining signal in cases of peripheral
time can be adjusted on many oximeters by variable vasoconstriction.43-45 Such peripheral "clamping down" is
averaging period settings. Oximeters also attempt to mini- probably most often seen in emergency practice in the
mize motion artifact with internal algorithms that eliminate hypothermic patient, including the majority of seriously ill
or differentially weigh outlier readings which are likely due patients who typically are exposed fully for assessment.
to motion. 5 In a similar fashion, some instruments couple Ostensibly, the block works by blocking digital sympathetic
oximetry readings to electrocardiograph signal. 34 innervation, with subsequent relative vasodilatation. The
In addition to extracorporeal motion, there are other digital block itself should not interfere with the detection or
sources of false alternating signal. Arterial pulsations can be correction of any systemic problems that could also underlie
transmitted into the venous circulation in venous congestion, loss of signal. Sometimes simple local heat or massage will
causing artificially lowered oxygen saturation readings. 4 accomplish the same effect. Anecdotally, topical nitroglycer-
This same phenomenon occurs in tricuspid regurgitation. ine has been employed with variable effect.34
Arterial pulsations with a large dicrotic notch, as in aortic To the same end, intraarterial vasodilators have also been
regurgitation, combined aortic stenosis and regurgitation, or used, instilled through ipsilateral radial lines. 46 Diluted
idiopathic hypertophic subaortic stenosis, can be processed boluses of nitroglycerine (10 ~g) and hydralazine (1 nag)
incorrectly at twice the true heart rate. This will have little have been reported to restore pulse oximeter function with
impact on oxygen saturation but has obvious effects on no adverse hemodynamic effect. Clearly, however, this is not
displayed heart rate. 34 a practical intervention in the majority of patients.
64 AMERICAN JOURNAL OF EMERGENCY MEDICINE Volume 17, Number 1 January 1999
Other Limitations supplementary oxygen at high Fio2 and showing a high SpO2
(99% to 100%), there can be a dramatic decrease in Pao2
Anemia appears to adversely affect the accuracy of pulse
before a corresponding decrease in oxygen saturation is
oximetry, although the mechanism is unclear, and it may do
manifested due to the shape of the oxygen-Hb dissociation
so only in the presence of hypoxia. In theory, anemia should
curve.
not affect pulse oximetry at all, as the ratio of relative
absorbances should be preserved and unchanged by changes
in total hemoglobin concentration within the sample. How- ACCURACY
ever, a retrospective study found an underestimation of
oxygen saturation by pulse oximetry in anemic subjects, Hundreds of studies have examined the accuracy of pulse
inversely proportional to Hb concentration and most pro- oximeters, with fair consistency in findings. Reviews have
nounced at an SpO2 less than roughly 80%. 47An experimen- summarized these studies nicely and have illustrated the
tal in vitro study48 and an animal study49 both showed a difficulty in answering the seemingly simple question of
similar trend with coincident anemia and hypoxemia. The accuracy.3,5,6,34,58
observation has been made that the direction of error is In general, manufacturers presently report pulse oximeter
fortuitous, presenting an exaggeration of apparent hypox- accuracy to one standard deviation of <+__3% at arterial
emia, as anemic patients are more susceptible to the oxygen saturations of >70%. These claims are largely
deleterious effects of desaturation. 47 With normal oxygen supported in the most recent reviews of studies addressing
saturation, an earlier dog study found pulse oximetry to be accuracy. 34,58It should be recognized, however, that this bias
consistently accurate with a near-zero bias in subjects with a or error is reported at one standard deviation only, and so
hematocrit of at least 10, but found a bias of 5.4% in subjects must be doubled or tripled to express a 95% or 99%
with a hematocrit less than 10. 5 Encouragingly, a recent confidence interval, respectively. This may seem a surpris-
study of nonhypoxic human patients with acute anemia from ingly wide range of error and is probably contrary to most
hemorrhage down to a Hb value as low as 2.3 g/dL showed clinician's preconceptions of accuracy. This only under-
good accuracy with pulse oximetry.51 scores the importance of widespread recognition of that
Skin pigmentation has shown variable effects on pulse range. Again, this range of error must be reassessed in every
oximetry.5244 Primarily, dark pigmentation appears to make situation in light of the limitations as previously discussed.
adequate light penetration more difficult, with significantly Methodological issues should be briefly discussed to put
more signal detection failures. It has been suggested that the the results of these studies into proper context. Subject
less pigmented nail beds may offer particularly good probe groups have been highly variable, ranging from healthy
sites in these patients. 34More concerning are some studies of volunteers to the critically ill. Correspondingly, study set-
pigmentation effects which have shown overestimations of tings have varied from rigidly controlled laboratories in
oxygen saturation, both in models 54 and in human subjects. 5a some studies to intensive care units and emergency depart-
The mechanism of such artifact, if present, is unclear, and ments in others.
could arise from optical interference, low signal-to-noise The "gold standard" of comparison now is generally
effects, or as a reflection of the relative lack of darkly recognized to be the in vitro four-wavelength co-oximeter,
pigmented subjects in calibration trials. Studies with larger with a reported accuracy for fractional saturation of ___1% at
patient groups are indicated. two standard deviations. 3 Some studies, however, have
Similarly, there is evidence for the effects of nail polish on utilized two wavelength in vitro co-oximeters as standards,
pulse oximetry, but little consensus on occurrence or de- with the same limitations inherent in use of only two
gree. 55-57When demonstrated, most interference appears to wavelengths as pulse oximeters. Still others have used
arise from blue or black polish. It is probably advisable to calculated SaO2 from arterial blood gas samples. The
avoid possible interference by using other probe sites, problem with such calculated oxygen saturations is that,
removing the nail polish, or even simply placing the probe because of shifts in the O2Hb dissociation curve, correction
sideways on the digit to remove the nail from the transmis- factors for variables like temperature, pH, and 2,3-
sion path. diphosphoglycerate concentration must be inserted for accu-
Although LEDs are reliable sources of narrow wavelength rate results. Most often, however, standard corrections are
light in most cases, there can be slight variability between used with potential resulting impact on accuracy.
LEDs in actual wavelength emitted. 5 A very small error in Data analyses also have been variable, with early studies
wavelength can translate into a large error in absorbance employing primarily regression analysis and correlation
interpretation, particularly at the steep portions of the curve coefficients. Such measures were found to be inadequate,
in the red range (Figure 1). In general, LEDs that emit light reflecting a known association between Sao2 and SpO2
too far from standard are rejected by the manufacturer. rather than addressing accuracy. A better measure has been
However, some manufacturers have dealt with this problem bias, or the mean value of (Sa02 - SpO2), in conjunction
in the past by compensatory changes in software and data with precision, or the standard deviation of bias. 3,5,34 Bias
processing. The danger arises in switching probes in such represents systematic differences between methods of mea-
cases, coupling an LED with processing software calibrated surement, with precision reflecting variability.
to a different wavelength of incident light. This problem Revisions in manufacturer algorithms and software, at
most likely will not arise in modern oximeters, but users times prompted by the findings of the studies themselves,
should be aware nonetheless. further complicate accuracy assessments. As such, results
Finally, it should be recognized that in patients receiving can become dated almost by nature of their own existence.
JAMES E. SINEX PULSE OXIMETRY: PRINCIPLES AND LIMITATIONS 65
18. Ramanathan R, Durand M, Larrazabal C: Pulse oximetry in 48. Kolesar J, Volgyesi G, Lerman J: Effect of haemoglobin
very low birth weight infants with acute and chronic disease. Pediat- concentration on the accuracy of pulse oximetry. Can J Anaesth
rics 1987;79:612-616 1990;37:$88 (abstr)
19. HarrisAP, Sendak MJ, Donham RT, et al: Absorption character- 49. Vegfors M, Lindberg LG, Oberg PA, Lennmarken C: The
istics of human fetal hemoglobin at wavelengths used in pulse accuracy of pulse oximetry at two haematocrit levels. Acta Anaesthe-
oximetry. J Clin Monit 1988;4:175-177 siol Scand 1992;36:454-459
20. Pologe J, Raley D: Effects of fetal hemoglobin on pulse 50. Lee SE, Tremper KK, Barker SJ: Effects of anemia on pulse
oximetry. J Perinato11987;7:324-326 oximetry and continuous mixed venous oxygen saturation monitoring
21. Beall SN, Moorthy SS: Jaundice, oximetry, and spurious in dogs. Anesth Analg 1988;67:$130
hemoglobin desaturation. Anesth Analg 1989;68:806-807 51. Jay GD, Hughes L, Renzi FP: Pulse oximetry is accurate in
22. Veyckemans F, Baele P, Guillaume JE, et al: Hyperbilirubine- acute anemia from hemorrhage. Ann Emerg Med 1994;24:32-35
mia does not interfere with hemoglobin saturation measured by pulse 52. Cahan C, Decker MJ, Hockje PL, Strohl KP: Agreement
oximetry. Anesthesiology 1989;70:118-122 between noninvasive oximetric values for oxygen saturation. Chest
23. Kessler MR, Eide T, Humayan B, Poppers PJ: Spurious pulse 1990;97:814-819
oximeter desaturation with methylene blue injection. Anesthesiology 53. Ries AL, Prewitt LM, Johnson JJ: Skin color and ear oximetry.
1986;65:435-436
Chest 1989;96:287-290
24. Scheller MS, Unger RJ, Keiner MJ: Effects of intravenously
administered dyes on pulse oximetry readings. Anesthesiology 1986; 54. Volgyesi GA, Spahr-Schopfer I: Does skin pigmentation affect
65:550-552 the accuracy of pulse oximetry? An in vitro study. Anesthesiology
25. Sidi A, Paulus DA, Rush W, et al: Methylene blue and 1991 ;75:A406
indocyanine green artifactually lower pulse oximetry readings of 55. Cote CJ, Goldstein EA, Fuchsman WH, Hoaglin DJ: The effect
oxygen saturation. J Clin Monit 1987;3:249-256 of nail polish on pulse oximetry. Anesth Analg 1988;67:683-686
26. Hanowell L, Eisele JH, Downs D: Ambient light affects pulse 56. Kataria BK, Lampkins R: Nail polish does not affect pulse
oxirneters. Anesthesiology 1987;67:864-865 oximetry saturation. Anesth Analg 1986;65:824 (letter)
27. Poets CF, Seidenberg J, von der Hardt H: Failure of pulse 57. Rubin AS: Nail polish color can affect pulse oximetry. Anesthe-
oximeter to detect sensor detachment. Lancet 1993;341:244 (letter; siology 1988;68:825 (letter)
comment) 58. Severinghaus JW: History and recent developments in pulse
28. Costarino AT, Davis DA, Keon TO: Falsely normal saturation oximetry. Scand J Clin Lab Invest 1993;53:214:105-111 (suppl)
reading with the pulse oximeter. Anesthesiology 1987;67:830-831 59. Jones J, Heiselman D, Cannon L, Gradisek R: Continuous
29. Southall DP, Samuels M: Inappropriate sensor application in emergency department monitoring of arterial saturation in adult
pulse oximetry. Lancet 1992;340:481-482 (letter) patients with respiratory distress. Ann Emerg Med 1988;17:463-468
30. Gardosi JO, Damianou D, Schram CM: Inappropriate sensor 60. Mateer JR, Olson DW, Stueven HA, Aufderheide TP: Continu-
application in pulse oximetry. Lancet 1992;340:920 (letter; comment) ous pulse oximetry during emergency endotracheal intubation. Ann
31. Kelleher JF, Ruff RH: The penumbra effect: Vasomotion- Emerg Med 1993;22:675-679
dependent pulse oximeter artifact due to probe malposition. Anesthe- 61. Holburn CJ, Allen MJ: Pulse oximetry in the accident and
siology 1989;71:787-791 emergency department. Arch Emerg Med 1989;6:137-142
32. Barker SJ, Hyatt J, Shah NK, Kao J: The effect of sensor 62. Lambert MA, Crinnion J: The role of pulse oximetry in the
malpositioning on pulse oximeter accuracy during hypoxemia. Anes- accident and emergency department. Arch Emerg Med 1989;6:211-
thesiology 1993;79:248-254 215
33. Hanning CD, Alexander-Williams JM: Pulse oximetry: A practi- 63. Mower MR, Sachs C, Nicklin EL, et al: Effect of routine
cal review. BMJ 1995;311:367-370 emergency department triage pulse oximetry screening on medical
34. Severinghaus JW, Kelleher JF: Recent developments in pulse management. Chest 1995;108:1297-1302
oximetry. Anesthesiology 1992;76:1018-1038 64. Narang VPS: Utility of the pulse oximeter during cardiopulmo-
35. James MR, Marshall H, Carew-McColl M: Pulse oximetry nary resuscitation. Anesthesiology 1986;65:239-240 (letter)
during apparent tonic-clonic seizures. Lancet 1991 ;337:394-395 65. Levinsohn DG, Gordon L, Sessler DI: The Allen's test: Analysis
36. Moorthy SS, Dierdorf SF, Schmidt SI: Erroneous pulse oxim- of four methods. J Hand Surg [Am] 1991 ;16A:279-282
eter data during CPR. Anesth Analg 1990;70:339 (letter) 66. O'Mara K, Sullivan B: A simple bedside test to identify ulnar
37. Short L, Hecker RB, Middaugh RE, Menk EJ: A comparison of collateral flow. Ann Intern Med 1995; 123:637
pulse oximeters during helicopter flight. J Emerg Med 1989;7:639- 67. Fuhrman TM, McSweeney E: Noninvasive evaluation of the
643 collateral circulation of the hand. Acad Emerg Med 1995;2:195-199
38. Talke P, Nichols RJ, Traber DL: Evaluation of pulse oximetry 68. Graham B, Paulus DA, Caffee HH: Pulse oximetry for vascular
for intraoperative blood pressure measurement and vital sign monitor- monitoring in upper extremity replantation surgery. J Hand Surg [Am]
ing during patient transport via life flight. Anesth Analg 1988;67:$229
1986;11A:687-692
(abstr) 69. Lawson D, Noriey I, Korbon G, et al: Blood flow limits and
39. Talke P, Nichols RJ, Traber DL: Monitoring patients during
pulse oximeter signal detection. Anesthesiology 1987;67:599-603
helicopter flight. J Clin Monit 1990;6:139-140 70. Talke P, Nichols RJ Jr, Traber DL: Does systolic blood
40. Schnapp LM, Cohen NH: Pulse oximetry--uses and abuses.
pressure measured using a pulse oximeter correlate with conven-
Chest 1990;98:1244-1250
tional methods? J Clin Monit 1990;6:5-9
41. Severinghaus JW, Spellman MJ: Pulse oximeter failure thresh-
71. Wallace CT, Baker JD 3d, Alpert CC, et al: Comparison of
olds in hypotension and vasoconstriction. Anesthesiology 1990;73:
blood pressure measurement by Doppler and by pulse oximetry
532-537
42. Evans ML, Geddes LA: An assessment of blood vessel techniques. Anesth Analg 1987;66:1018-1019
vasoactivity using photoplethysmography. Med Instrument 1988;22: 72. Chawla R, Kumarvel V, Girdhar KK, et al: Can pulse oximetry
29-32 be used to measure systolic blood pressure? Anesth Analg 1992;74:
43. Bourke DL, Grayson RF: Digital nerve blocks can restore 196-200
pulse oximeter signal detection. Anesth Analg 1991 ;73:815-817 73. de Kock JP, Tarassenko L, Glynn CJ, Hill AR: Reflectance
44. Gentili ME, Chevaleraud E, Viel E: Digital block of the flexor pulse oximetry readings from the retinal fundus. IEEE Trans Biomed
tendon sheathe can restore pulse oximeter signal detection. Reg Eng 1993;40:817-823
Anesth 1995;20:82-83 (letter) 74. Dougherty G, Lowry J: Design and evaluation of an instrument
45. Grayson RF, Bourke DL: Digital block for pulse oximetry to measure microcirculatory blood flow and oxygen saturation simul-
failure. Anesthesiology 1991 ;75:A406 (abstr) taneously. J Med Eng Technol 1992;16:123-128
46. Holroyd K, Lui M, Beattie C: Intraarterial vasodilator adminis- 75. Jorgensen JS, Schmid ER, Konig V, et a1: Limitations of
tration to restore pulse oximeter function. Anesthesiology 1993;79: forehead pulse oximetry. J Clin Monit 1995;11:253-256
388-390 76. Lindberg LG, Lennmarken C, Vegfors M: Pulse oximetry--
47. Severinghaus JW, Koh SO: Effect of anemia on pulse oximeter clinical implications and recent technical developments. Acta Anaes-
accuracy at low saturation. J Clin Monit 1990;6:85-88 thesiol Scand 1995;39:279-287