You are on page 1of 4

Repairs

DD+DIS015.01E Codes and fuses

Section 6.4

Contents
1 Switches and codes (only on control board GS1, type 106/206/846)......................................1

2 Fuses (only on distributor board GS3, type 106/206/846)........................................................2

3 Fuses (type 105/155/165/205/285/845) .......................................................................................3

4 Voltages (type 105/155/165/205/285/845)...................................................................................3

Edition 2, Revision 0 CURIX 60 / CP 1000 (type 9462) Section 6.4 / I


Repairs
DD+DIS015.01E Codes and fuses

1 Switches and codes (only on control board GS1, type 106/206/846)

ST8

ST7

ST9

S1

946264cm.cdr

Firgure 6.4 - 1

Switch Description Possible choices


S1 Adjusting full-degree temperatures 29 36 C,
for the developer solution see section 6.6
ST 7 Adjusting circulation pump speed var / fix, do not change
ST 8 Adjusting main drive speed var / fix, do not change
ST 9 Adjusting developer temperature var / fix, see section 6.6

Edition 2, Revision 0 CURIX 60 / CP 1000 (type 9462) Section 6.4 / 1


Reparatur
Kodierungen und Sicherungen DD+DIS015.01E

2 Fuses (only on distributor board GS3, type 106/206/846)

SI5
D4

SI6
D3

D2
SI7
SI1

SI2
D1 SI8
946264cm.cdr
SI3
GS2
SI4
SI9
SI10

946264bm.cdr

GS3
Firgure 6.4 - 2

Fuse Value Description Control diode


SI1 6.25 A slow Mains connection 230*
SI2 6.25 A slow Mains connection 208 V*
SI3 6.25 A slow Mains connection 200 V*
SI4 6.25 A slow Mains connection 120 V*
SI10 6.25 A slow Mains connection 100 V*
SI5 6.25 A slow IR heater D4 on GS2
SI6 400 mA slow Lower dryer fan D2 on GS2
SI7 400 mA slow Upper dryer fan
SI8 700 mA slow External replenishment D3 on GS2
SI9 2.5 A slow Developer heater D1 on GS2
*Only one position for fuses SI1 to SI4 and SI10 is occupied.

Section 6.4 / 2 CURIX 60 / CP 1000 (type 9462) Edition 2, Revision 0


Repairs
DD+DIS015.01E Codes and fuses

3 Fuses (type 105/155/165/205/285/845)

Fuse Type 165/845 Type 105/155/205/285 Component Plug LED


at 120 V at 230 V
Si 1 T 2,5 A (slowbl.) T 1,6 A Heater DEV St 5 D8
Si 2 T 0,5 A (slowbl.) T 0,4 A Drive St 6
Si 3 T 2,5 A (slowbl.) T 1,6 A Heater Fix St 7
Si 4 T 6,25 A (slowbl.) T 4 A Dryer heater St 17/8/5 D9
Si 5 T 1 A (slowbl.) T1A Transformer

Si 6 T 0,5 A (slowbl.) T 0,4 A Lower fan St 4 D 10


Si 7 T 0,5 A (slowbl.) T 0,4 A Upper fan St 3
Si 8 T 1,25 A (slowbl.) T 0,7 A Opt. ext. replenishm. St 10 D 12

4 Voltages (type 105/155/165/205/285/845)

Voltage Measuring point Consumers


~120 V GS1.XK1/1, 1.XK1/2 Drive motor
Infrared dryer
Overheat cut-out
Bath heater
Blower
~230 V GS1.XK1/1, 1.XK1/2 Drive motor
Infrared dryer
Overheat cut-out
Bath heater
Blower
+12 V / +22 V* GS1 MP3 +12 V / +22 V DC Replenishment pump
MP2 GND
+5 V, stabilized GS1 MP1 +5 V Electronics
MP2 GND Surface scanning unit
* at a speed of 360 rpm ---> 12 V
at a speed of 720 rpm ---> 22 V

Edition 2, Revision 0 CURIX 60 / CP 1000 (type 9462) Section 6.4 / 3