You are on page 1of 5

[Downloaded free from http://www.annals.in on Thursday, October 27, 2016, IP: 115.178.239.

182]

Original
44 Sujatha etArticle
al. Thermodilution versus Pulse Contour Cardiac Output Annals of Cardiac Anaesthesia 2006; 9: 4448

Comparison of Cardiac Output in OPCAB: Bolus Thermodilution


Technique versus Pulse Contour Analysis
P Sujatha, Y Mehta, A Dhar, D Sarkar, ZS Meharwal, N Trehan
Department of Anaesthesiology and Critical Care and Department of Cardiac Surgery
Escorts Heart Institute and Research Centre, New Delhi

The study was designed to evaluate the clinical agreement between intermittent bolus thermodilution
technique and pulse contour analysis technique. Sixty patients with normal left ventricular function
undergoing elective off-pump coronary bypass surgery were included in this prospective study. In addition
to routine monitoring, a 7.5F pulmonary artery thermodilution catheter via right internal jugular vein
and a 4F arterial thermodilution catheter into femoral artery were also placed. Cardiac output measurements
were compared before induction, after induction, after sternotomy, during the various anastomoses, post-
protamine and post-sternal closure. Statistical analysis was performed using analysis of agreement to
assure bias distribution of differences between the two methods by using Bland and Altman analysis. The
cardiac output values obtained at preinduction, post-induction, and post-sternal closure time points showed
good agreement, whereas the values obtained during the various anastomoses showed significant differences
(p <0.05). Therefore it was concluded that pulse contour analysis cannot be relied upon completely whenever
there is a change in the position of heart or alteration in systemic vascular resistance. But the trends in
cardiac output were in complete agreement during the entire procedure. (Annals of Cardiac Anaesthesia
2006; 9: 4448)

Key words: Off-pump coronary artery bypass, Cardiac output, Thermodilution, Pulse contour
analysis

C ardiac output (CO) is one of the important


haemodynamic variables during and after
cardiac surgery. CO estimation using thermo-
suited to track the abrupt haemodynamic changes
that occur during off-pump coronary bypass
(OPCAB) Surgery. Continuous CO monitoring
dilution pulmonary artery catheter (PAC) remains devices using the pulse contour CO principle have
the current clinical standard1 despite its pitfalls2,3 been described by Wesseling et al,7 which appears
and potential hazards.4 There has been consensus to be accurate in measuring CO and in tracking its
regarding the indications for use of PAC despite changes. This method measures CO continuously
the absence of prospective randomized studies by measuring the area under the systolic portion
showing improvement in patient outcome. 5 of the arterial pressure waveform. We studied the
Continuous measurement of CO is preferable to results of measurements using the pulse contour
intermittent measurements, especially during CO monitoring system and compared them with
rapid changes in cardiovascular function. Many the conventional pulmonary bolus thermodilution
experimental and clinical studies have method in cardiac surgery patients.
demonstrated excellent agreement between
intermittent bolus and continuous CO. 6 The Meterials and Methods
response time of continuous CO by pulmonary
thermodilution is slow and may not be ideally After obtaining approval from the ethics
committee and written informed consent, we
Address for Correspondence: Dr. Yatin Mehta, Senior Consultant and Head, studied 60 patients aged 40-75 years with normal
Department of Anaesthesiology and Critical Care, Escorts Heart Institute
and Research Centre, Okhla Road, New Delhi, Phone : 011-26825000, left ventricular and pulmonary functions
26825001, Fax: 011-51628442, 26825013. Email: yatinmehta@hotmail.com undergoing OPCAB surgery. Patients with

ACA-04-126- OA.p65 44 12/24/2005, 12:31 PM


[Downloaded free from http://www.annals.in on Thursday, October 27, 2016, IP: 115.178.239.182]

Annals of Cardiac Anaesthesia 2006; 9: 4448 Sujatha et al. Thermodilution versus Pulse Contour Cardiac Output 45

valvular heart disease, left ventricular ejection was facilitated with vecuronium bromide 0.08 mg/
fraction <30% and peripheral vascular disease were Kg. Anaesthesia was maintained with oxygen + air
excluded. Individual cardiac medications including + isoflurane mixture, intermittent doses of fentanyl
antihypertensive agents, nitroglycerin and beta 1-2 g/Kg, midazolam 0.02 mg/Kg and
blockers were continued until the day of surgery. vecuronium bromide 0.04 mg/Kg. Heparin
Routine haemodynamic monitoring was 2 mg/kg was administered prior to commencing
established, which included electrocardiography, myocardial revascularisation to achieve an
pulse oximetry, central venous and radial artery activated clotting time of 250-300 seconds.
catheterization. A 7.5F PAC (Baxter Health Myocardial revascularisation was performed with
Corporation, Edwards Critical care division, Irvine the aid of octopus III (Medtronic Minneapolis,
CA, USA) was inserted into the right internal USA) myocardial stabilizer. All patients were
jugular vein and a 4F arterial thermodilution studied during periods of haemodynamic stability
catheter (PiCCO, Pulsion Medical Systems, and without cardiac arrhythmias. Heparin was
Munich, Germany) was inserted via femoral artery neutralised with protamine at the completion of
using Seldinger technique for clinical monitoring revascularisation. Normothermia was maintained
of arterial pressure and continuous arterial in all patients intraoperatively with forced air
thermodilution CO measurements. All invasive warmers, warm intravenous fluids and circulating
cannulations for monitoring were performed under warm water mattress beneath the patient. On
local anaesthesia. The arterial catheter was completion of surgery, patients were shifted to
connected to a computer for pulse contour analysis. intensive care unit. Intraoperative intravenous fluid
To calibrate this system, individual arterial infusion was individually adjusted for each patient
impedance to arterial pressure is calculated by to maintain a cardiac index preferably > 2 L/min/
simultaneously determining the area under this m2 with normal haemodynamic filling pressures
systolic portion of the arterial pulse wave and (central venous pressure 4-10 mm Hg and
measuring arterial thermodilution CO. pulmonary capillary wedge pressure 8-12 mm Hg)
and urine output 1-2 ml/Kg/hour. Haemoglobin
The mean of three subsequent measurements, concentration of < 9 gm/dl and platelet < 60,000/
randomized within the respiratory cycle, was used. mm3 were the triggers for transfusion of packed
They were performed by injection of 10 ml of iced red cells and platelet concentrate respectively.
saline solution via a central venous catheter with
subsequent detection by the thermister embedded Statistical analysis was performed using analysis
in the wall of the arterial catheter. Thereafter a beat of agreement to assure bias distribution of
to beat analysis of CO based on pulse pressure differences between two methods by using Bland
waveform is obtained. Conventional pulmonary and Altman analysis.8 Bias is defined as the average
bolus thermodilution CO was also measured and difference between 2 measures. Precision is defined
compared with pulse contour CO values on 8 as limits of agreement or the variation (standard
occasions. Before and after five minutes of deviation) of the difference between 2
induction of anaesthesia(T1,T2), post sternotomy measurements. Additional descriptive analysis was
(five minutes after the start of LIMA dissection, T3) performed using SPSS 10.0 Chicago, USA software
five minutes after commencement of left anterior package.
descending artery (T 4) left obtuse marginal/
diagonal, (T5) and right sided, (T6) (right coronary/ Table 1. Demographics
posterior descending artery) anastomoses, five
Age (years) 59.2 7.8
minutes after completion of protamine (T7) and post Male 45/60 (75%)
sternal closure (T8). Number of distal anastomoses 3.4 0.2
BSA 1.74 0.16
Anaesthesia was induced with fentanyl citrate Diabetes 20/60 (33.3%)
Hypertension 30/60 (50%)
5 g/Kg, midazolam 0.05 mg/Kg, titrated doses
of thiopentone sodium and endotracheal intubation BSA: body surface area

ACA-04-126- OA.p65 45 12/24/2005, 12:31 PM


[Downloaded free from http://www.annals.in on Thursday, October 27, 2016, IP: 115.178.239.182]

46 Sujatha et al. Thermodilution versus Pulse Contour Cardiac Output Annals of Cardiac Anaesthesia 2006; 9: 4448

Results error percentage (2SD/mean CO100) was 20%.


The mean error percentage at T2 and T8 were 22%
There were 15 females and 45 males with mean and 24% respectively. The error percentage during
age of 59 7 years, mean weight of 68 12 Kgs rest of time frames were more than 35%, which may
and height of 164 6 cms (table 1). Values are be considered above the levels of clinical acceptance
presented as mean standard deviation and p (table 3). The pattern observed in systemic vascular
value <0.05 was considered statistically significant. resistance measurements made with both methods,
Differences between methods during each time was the same as observed in CO measurements
points are given in table 2. There was no significant (table 4).
difference between CO values assessed by
intermittent bolus thermodilution and pulse Discussion
contour analysis of cardiac output at T1, T2 and T8
points but showed statistically significant Bolus intermittent thermodilution was chosen
difference during T3, T4,T5, T6, and T7 time points as the CO comparison standard because it is a
(p< 0.05). The mean difference between widespread clinical method of measuring CO.
thermodilution CO and pulse contour CO at Rapid changes in CO may not be detected with
preinduction (T 1) was 0.23 L/min (bias). The intermittent technique. Previous studies have
precision (2SD of bias) was 1L/min and the mean reported a small mean difference (bias) and
clinically acceptable limits of agreement (bias 2 x
Table 2. Analysis of comparison of different methods of SD).9,10 Transpulmonary thermodilution using
cardiac output (L/min) estimation pulse contour analysis has been shown to be an
Variable T1 T2 T3 T4 T5 T6 T7 T8
easy and reliable technique for assessing CO and
is increasingly being used for haemodynamic
TDCO 51 41 4.81 51 41 4.21 4.01 61.5
PICCO 5 0 41 3.41 41 31 3.51 3.51 5.81 assessment in the operating room and intensive
P NS NS <0.05 <.01 <0.05 <0.01 <0.01 NS care unit. 12 Our results comparing these two
T1- preinduction, T2- 5 minutes after induction, T3- 5 minutes after the start of left
internal mammory artery (LIMA) dissection T4- 5 minutes after commencement of
methods did not agree well under conditions
LIMA to left anterior descending artery anastomosis T5 5 minutes after where heart was displaced to reach the target site
commencement of lateral wall anastomosis, T6- 5 minutes after commencement of
right wall anastomosis, T7- 5 minutes after completion of protamine, T8- post-sternal of anastomoses.
closure
TDCO: Thermodilution cardice output, PICCO: pulsian continuous cardiac output
Here we have tried to evaluate the agreement
Table 3. Statistical analysis of comparison of between pulmonary bolus thermodilution CO and
thermodilution cardiac output and pulsian continuous pulse contour CO, which delivers beat to beat CO
cardiac output values leading to early detection of CO changes.
Time No.of CO (L/mt) Bias SD of L.O.A Mean error
Frame data pairs (L/mt) Bias (L/mt) (L/mt) %
The concept of arterial pulse contour analysis
1 60 4.940.97 0.23 0.5 1.0 20
2 60 4.070.86 0.29 0.45 0.9 22 for the determination of continuous CO has been
3 60 4.150.85 0.39 0.91 1.8 43
4 60 4.221.1 0.51 0.91 1.8 46 the subject of investigation for a number of years.
5 60 3.971.9 0.55 0.74 1.4 35
6 60 3.831.01 0.51 0.95 1.8 47
Basic algorithm for calculation of CO was put
7
8
60
60
5.11.4
5.1.3
1.04
.23
1.22
0.6
2.4
1.2
47
24
forward by Wesseling.7 According to this, stroke
CO: cardiac output; SD: standard deviation; LOA: limits of agreement (2SD), mean error % volume is computed by measuring the area under
LOA/mean CO100
systolic portion of arterial pressure waveform

Table 4. Comparison of systemic vascular resistance (dynes. sec. m2 cm-5) measured by 2 methods
Variable T1 T2 T3 T4 T5 T6 T7 T8

TDCO 1437254 1468255 1593374 1325356 1400360 1311328 1430270 1091317

PICCO 1550264 1638381 1872326 1598355 1766383 1690446 1080292 149 270

P NS NS <0.01 <0.01 <0.01 <0.01 <0.01 NS


T1- preinduction, T2- 5 minutes after induction, T3- 5 minutes after the start of LIMA dissection T4- 5 minutes after commencement of LIMA to LAD anastomosis T5
5 minutes after commencement of lateral wall anastomosis, T6- 5 minutes after commencement of right wall anastomosis, T7- 5 minutes after completion of
protamine, T8- post sternal closure
TDCO: thermodilution cardiac output, PICCO: pulsian continuous cardiac output

ACA-04-126- OA.p65 46 12/24/2005, 12:31 PM


[Downloaded free from http://www.annals.in on Thursday, October 27, 2016, IP: 115.178.239.182]

Annals of Cardiac Anaesthesia 2006; 9: 4448 Sujatha et al. Thermodilution versus Pulse Contour Cardiac Output 47

divided by aortic impedance. This value multiplied various anastomoses may be producing some sort
by heart rate gives the pulse contour CO. To adjust of descending thoracic aortic compression which
for aortic impedance, which differs from patient might be causing the discrepancy in the values. The
to patient , a thermodilution measurement of CO difference in the values were statistically significant
for calibration is performed. The new software (p <0.05, table 2). The algorithm for computation
analyses the shape of entire pressure waveform in of CO by pulse contour CO includes three major
addition to area under systolic portion of pressure properties of arterial system compliance,
wave.11 It takes into account individual aortic peripheral vascular resistance and impedance.
compliance and systemic vascular resistance based Peripheral vascular wave reflects interplay between
on following consideration. During systole, more left ventricular output and capacitance of vascular
blood is ejected from left ventricle than what tree. Systemic vascular changes caused by
actually leaves aorta. During subsequent diastole vasopressors and protamine can have more impact
blood flows from aorta into arterial network at a on pulse contour CO than thermodilution bolus
rate depending on aortic wall compliance, systemic CO. This also explains the discrepancy during
vascular resistance and blood pressure. Calibrating various anastomoses and post-protamine CO
procedure for pulse contour CO device necessitates measurements. During T1, T2, T8 where these factors
a transpulmonary measurement of CO. Ten ml of do not come into play, CO measurements showed
isotonic saline at 7o to 9oC is injected into central good agreement. Bland and altmann analysis
venous line, which is subsequently detected by a revealed a bias value between thermodilution CO
thermistor embedded in the wall of the arterial and pulse contour CO of 0.23 L/min, 0.29 L/min
catheter. The mean of three consecutive and 0.23 L/min at T1, T2 and T8 time points, which
measurements randomized within the respiratory shows good agreement. Use of pulse contour CO
cycle was used. CO is computed by multiplying alone provides no information on pulmonary
stroke volume by heart rate. A moving average of arterial pressure, and mixed venous oxygen
the preceding 30 sec is calculated and displayed saturation, which are useful variables in
on the monitor. management of OPCAB.12,13

The difference in CO during left internal Conclusion


mammary artery dissection could be due to the use
of sternal retractor which may cause compression We conclude that assesment of continuous CO
of subclavian artery or descending aorta leading by pulse waveform analysis and by intermittent
to a decrease in CO measured by pulse contour thermodilution technique provided comparable
analysis as compared to the normal value when measurements during OPCAB surgery as long as
measured using thermodilution CO. The octopus singnifcant changes in systemic vascular resistance
stabiliser used for optimal positioning during did not occur.

References

1. Zhao X, Mashikian JS, Panzika P, et al. Comparison of effectiveness of right heart catheterization in initial care
thermodilution bolus cardiac output and Doppler cardiac of the critically ill patients. JAMA 1996; 276: 889-897
output in the early post cardiopulmonary bypass period. 5. Chernow B. Pulmonary artery flotation catheters. A
J Cardiothorac Vasc Anesth 2003; 17: 193-198 statement by the American College of Chest Physicians
2. Wallace DC, Winslow EH. Effect of iced and room and the American Thoracic Society. Chest 1997; 111: 261-
temperature injectable on cardiac output measurements 66
in critically ill patients with low and high cardiac output. 6. Yelderman M. Continuous cardiac output by
Heart Lung 1993; 22: 55-631 thermodilution. Int Anaesthesiology Clin 1993; 3: 127-140
3. Nishikawa T, Shuji Dohi. Errors in the measurement of 7. Wesseling R, de Wit B, Ty Smith N. A simple device for
cardiac output by thermodilution. Can J Anaesth 1993; 40: the continuous measurement of cardiac output. Adv
142-53 Cardiovas Phys 1983; 5: 16-52
4. Connors AF, Seperoff T, Dawson NV, et al. The 8. Bland JM, Altman DG. Statistical methods for assessing

ACA-04-126- OA.p65 47 12/24/2005, 12:31 PM


[Downloaded free from http://www.annals.in on Thursday, October 27, 2016, IP: 115.178.239.182]

48 Sujatha et al. Thermodilution versus Pulse Contour Cardiac Output Annals of Cardiac Anaesthesia 2006; 9: 4448

agreement between two methods of clinical measurement. by pulse contour analysis during hemodynamic
Lancet 1986; 1: 307-310 instability. Crit Care Med 2002; 30: 52- 58
9. Rodig G, Prasser C, Keyl C, et al. Continuous cardiac 12. Sakka SG, Ruhl CC, Pfeiffer UJ, et al. Assessment of
output measurement pulse contour analysis vs cardiac preload and extravascular lung water by
thermodilution technique in cardiac surgical patients. Br transpulmonary thermodilution. Intensive Care Med 2000;
J Anaesth 1999; 82: 525-30 26: 180-187
10. Chassot PG, van derlinden P, Zaugg M, et al. Off Pump 13. Michard F, Perel A. Management of circulatory and
coronary artery bypass surgery: physiology and respiratory failure using less invasive hemodynamic
anaesthetic management. Br J Anaesth 2004; 92: 400-413 monitoring in Vincent JL (ed): Yearbook of intensive care
11. Goedje O, Hoeke K, Goetz AE, et al. Reliability of a new and emergency medicine, Springer, berlin, 2003, pp 508-
algorithm for continuous cardiac output determination 520

ACA-04-126- OA.p65 48 12/24/2005, 12:31 PM

You might also like