You are on page 1of 15

1510

Cardiac Potential and Potential Gradient Fields


Generated by Single, Combined, and Sequential
Shocks During Ventricular Defibrillation
J. Marcus Wharton, MD; Patrick D. Wolf, MS; William M. Smith, PhD; Peng-Sheng Chen, MD;
David W. Frazier, MD; Seitaro Yabe, MD; Ned Danieley, MS; and Raymond E. Ideker, MD, PhD

Background. Potential gradient field determination may be a helpful means of describing the effects of
defibrillation shocks; however, potential gradient field requirements for defibnrllation with different
electrode configurations have not been established.
Methods and Results. To evaluate the field requirements for defibrillation, potential fields during
Downloaded from http://circ.ahajournals.org/ by guest on January 6, 2017

defibrillation shocks and the following ventricular activations were recorded with 74 epicardial electrodes
in 12 open-chest dogs with the use of a computerized mapping system. Shock electrodes (2.64 cm2) were
attached to the lateral right atrium (R), lateral left ventricular base (L), and left ventricular apex (V).
Four electrode configurations were tested: single shocks of 14-msec duration given to two single
anode-single cathode configurations, R: V and L: V, and to one dual anode-single cathode configuration,
(R+L):V; and sequential 7-msec shocks separated by 1 msec given to R:V and L:V (R:V-*L:V).
Defibrillation threshold (DFT) current was significantly lower for R:V--L:V than for the other
configurations and markedly higher for L:V. Despite these differences, the minimum potential gradients
measured at DEFI were not significantly different (approximately 6-7 V/cm for each electrode configura-
tion). Potential gradient fields generated by the electrode configurations were markedly uneven, with a
15-27-fold change from lowest to highest gradient, with the greatest decrease in gradient occurring near
the shock electrodes. Although gradient fields varied with the electrode configuration, all configurations
produced weak fields along the right ventricular base. Early sites of epicardial activation after all
unsuccessful shocks occurred in areas in which the field was weak; 87% occurred at sites with gradients
less than 15 V/cm. Ventricular tachycardia originating in high gradient areas near shock electrodes
followed 11 of 67 successful shocks.
Conclusions. These data suggest that 1) defibrillation fields created by small epicardial electrodes are
very uneven; 2) achievement of a certain minimum potential gradient over both ventricles is necessary for
ventricular defibrillation; 3) the difference in shock strengths required to achieve this minimum gradient
over both ventricles may explain the differences in DIFTs for various electrode configurations; and 4) high
gradient areas in the uneven fields can induce ectopic activation after successful shocks. (Circuaion
1992;85:1510-1523)
KEY WORDS * defibrillation * fields, gradient

T odefibrillate with the least amount of energy, the myocardium; thus, potential gradients may explain the
potential field generated by the defibrillating relative efficacy of different electrode configurations.
shock must be optimized. The potential gradi- Though a few studies have measured potential gradients
ent has been reported to be closely related to the and gradient fields during electrical shocks in sinus
success or failure of defibrillation.' If a certain mini- rhythm or ventricular fibrillation,2-5 systematic evalua-
mum potential gradient is necessary over the myocar- tion of field requirements for multiple electrode config-
dium to achieve defibrillation, shocks that do not pro- urations to determine the minimum magnitude and
duce this minimum potential gradient will not be spatial requirements of potential gradient fields has not
successful. Furthermore, the defibrillation threshold been performed. In addition, mapping of potential
(DFT) should be determined by the shock strength gradient fields may help to explain the increased efficacy
required to obtain this minimum gradient across the of some sequential shock electrode configurations. The

All editorial decisions for this article, including selection of grants HL-17670, HL-28429, HL-33637, HL-07063, HL-44066,
reviewers and the final decision, were made by a guest editor. This and HL-42760; American Heart Association North Carolina Af-
procedure applies to all manuscripts with authors from the Uni- filiate grant-in-aid 1985-86 A-06; and a grant from the Fannie E.
versity of California San Diego or UCSD Medical Center. Ripple Foundation. J.M.W. received the North American Society
From the Departments of Medicine and Pathology (J.M.W., of Pacing and Electrophysiology Young Investigator Award for a
P.D.W., W.M.S., D.W.F., S.Y., N.D., R.E.I.), Duke University Med- presentation of a preliminary version of this article.
ical Center, Durham, N.C., and the Cardiology Division (P.-S.C.), Address for correspondence: J. Marcus Wharton, MD, PO Box
University of California Medical Center, San Diego, Calif. 3816, Duke University Medical Center, Durham, NC 27710.
Supported in part by National Institutes of Health research Received January 22, 1991; revision accepted November 21, 1991.
Wharton et al Defibrillation Gradients 1511

present study was designed to 1) describe the potential lateral wall of the right atrium near the junction of the
and potential gradient fields generated by electrode superior vena cava (electrode R in Figures 1A and 1B),
configurations with different defibrillation efficacies; 2) the lateral free wall of the left ventricle adjacent to the
correlate the field characteristics with the efficacy of atrioventricular groove (electrode L), and the left ven-
each electrode configuration; 3) describe the field char- tricular apex (electrode V). The location of the L and V
acteristics necessary for defibrillation; and 4) define electrodes was varied slightly to avoid placement over or
further the relation of field strength to the site of initial near coronary arteries. Two pairs of stainless steel wires
ventricular activation after successful and unsuccessful were inserted into the right ventricular free wall approx-
defibrillation shocks. imately 1 cm apart for ventricular pacing and sensing.

Methods Defibrillation Protocol


Surgical Preparation After induction of ventricular fibrillation with a short
burst of 60-Hz alternating current in the open-chest
In 12 mongrel dogs (mean weight, 19.74.7 kg), preparation, single truncated exponential defibrillation
anesthesia was induced with intravenous pentobarbital shocks with a duration of 14 msec were given with a
30-35 mg/kg and maintained with a dose of 0.05 specially designed 750-microfarad defibrillator (Inter-
mg/kg/min.6 Skeletal muscle paralysis was obtained medics, Inc.) through one of three anode-cathode
with intravenous succinylcholine 1 mg/kg and main- electrode configurations, abbreviated R:V, L:V, and
tained with a dose of 0.25-0.50 mg/kg/hr as needed to (R+L):V. The last abbreviation indicates that the R
prevent skeletal muscle contraction.7 The dogs were and L electrodes are combined as a common anode.
intubated and mechanically ventilated (model 607, Har- One sequential shock (abbreviated R:V--L:V) was
vard Apparatus Company, Inc.). A midline sternotomy
Downloaded from http://circ.ahajournals.org/ by guest on January 6, 2017

tested in which a 7-msec R: V was followed by a 7-msec


was performed, and the heart was suspended in a L: V separated by a pause of 1 msec.9 The 14-msec
pericardial cradle. The nylon sock, which was placed monophasic waveform had a 16% tilt when delivered
over the ventricles for mapping purposes (see below), across a test resistor of 100 Ql.
helped prevent desiccation of the exposed portions of DFTs were determined using a modified Purdue
the heart; warm saline was also applied repeatedly to method.10 The initial shocks were given at an estimated
the sock to keep it moist. Warm lactated Ringer's strength of 4 J (2 J to each shock for the sequential pair)
solution was infused continuously through a peripheral to each of the electrode configurations in random order.
intravenous cannula, and blood pressure was monitored The shock energy to be delivered was estimated by
through a femoral artery cannula connected to a trans- measuring the current during a 100-V shock delivered
ducer (model 1280, Hewlett-Packard Co.). Arterial in sinus rhythm from an electrode configuration, calcu-
blood was sampled every 30-60 minutes (or more lating the impedance, and then calculating the voltage
frequently if indicated) and pH, Pco2, Po2, HCO2-, Na+, necessary to deliver 4 J (or 2 J sequentially) assuming no
K+, and Ca24 levels were monitored and corrected as change in impedance. Because impedances for R: V and
necessary to maintain these parameters within normal L: V were different, potentials required to deliver 2 J to
limits. Core temperature was monitored with a tongue each component of the sequential shock were not the
thermistor and maintained at 37C using a heating same. Each shock was given approximately 15 seconds
blanket. A heating lamp warmed the open chest. All after the induction of ventricular fibrillation; when it
studies were performed in accordance with the guide- was unsuccessful, a 10-30-J rescue shock (defibrillator
lines for use of experimental animals established by the model 7802B, Hewlett-Packard) was given through
American Physiological Society. hand-held paddles placed on the pericardium over the
right and left ventricles. Temporary pacing and/or brief
Recording and Shock Electrodes cardiac massage was given during periods of asystole
A nylon sock with 72 evenly spaced button electrodes8 after cardioversion. Another shock was not attempted
was applied over the heart and secured to the atria, for at least 5-10 minutes or until baseline heart rate and
atrioventricular groove fat, or pericardium to ensure blood pressure were reattained.6,11,12
stability of location. The button electrodes were placed When a defibrillation attempt failed, the energy level
in seven rows from apex to base (Figure 1A), with the for the subsequent shock was increased by 20% above the
basal row extending to the atrioventricular groove. Each delivered energy from the preceding unsuccessful shock
button encases two electrodes with centers 1.5 mm apart for that electrode configuration, and defibrillation was
(1.0 mm edge to edge). Unipolar or bipolar recordings reattempted. Estimation of the voltage required to deliver
were obtained from these electrodes depending on the the desired energy for each electrode configuration was
setting of the mapping system (see below). A clamp based on the impedance measured from the preceding
electrode attached to the left leg served as the reference shock with the same electrode configuration. When a
for unipolar recordings. defibrillation shock was successful, the energy for the next
Three solid circular titanium mesh electrodes with a shock for that electrode configuration was decreased by
diameter of 2.0 cm (total surface area, 3.14 cm2) and 10% until defibrillation was not obtained. Each electrode
with a bipolar button recording electrode (surface area configuration was tested in random order for each incre-
of plastic button, 0.50 cm2) centered underneath were mental change in energy. The lowest energy resulting in
used to deliver shocks (Figures 1A and 1B). The total defibrillation was called the DFT.
area of the shock electrode exposed to myocardium and
not bounded by insulator was a circular ring with a Electrode Recordings
width of 0.6 cm (surface area, 2.64 cm2). Shock elec- Before and after each attempted defibrillation, epi-
trodes were sewn securely to the epicardium of the cardial activation was recorded by the 75 bipolar button
1512 Circulation Vol 85, No 4 April 1992

FIGURE 1. Schematic representation of an-


terior (panel A) and right lateral (panel B)
views of dog heart illustrating locations ofthe
three shock electrodes (large cross-hatched or
unfilled circles labeled L, V and R). The right
atrial shock electrode (R) in the anterior view
(panel A) and the left lateral free wall shock
electrode (L) in the right lateral view (panel
B) are not cross-hatched to indicate that they
are located behind the surfaces illustrated.
Bipolar recording electrode within an insu-
lated button is shown in the center of each
shock electrode. Smallfilled circles represent-
A. . ^
ing locations of recording electrodes across
the ventricles are shown in the anterior view
(panelA) only. In the right lateral view (panel
B), an imaginary plane formed by the three
shock electrodes is outlined by straight lines
between the electrodes. Note that right ven-
tricular outflow tract is farther away from this
Downloaded from http://circ.ahajournals.org/ by guest on January 6, 2017

imaginary plane than is the inferiorportion of


the ventricles. Panel C: Simultaneous record-
C. ing of an unsuccessful 252-V (R+L): V (dual
RVOT anode-single cathode configuration) shock
displaying recordings from two electrodes,
one on the right ventricular outflow tract
I lOmV (RVOT) and the other on the left ventricular
apex. Solid bar at the top offigure indicates
APEX ' I 20V 25-msec period that the mapping system is
recording unipolar with a set of gains and
filter settings appropriate for the shock and
lOOmnsec II.. with voltage attenuator switched on. Four
loomw ~~~~~~~~11 intervals can be identified: A, sharp initial
I>VIII deflection as the voltage attenuator is
switched on; B, period offlat baseline lasting
approximately 6 or 7 msec from which the
magnitude of the shock potentials is refer-
enced; C, abrupt square deflection (with
spiked leading and trailing edges caused by
stray capacitance) representing the 14-msec
truncated exponential shock; and D, another
abrupt, high-amplitude deflection after
switching off voltage attenuators. Both before
and after the shock period, the system is
recording bipolar with a different set of gains
and filters. Vertical scales on right side of
figure indicate 10 mVand 20 Vduring bipolar
electrogram and unipolar shock recording,
respectively.

electrodes (74 of which were on the ventricles). Ampli- resistor at the front end of the amplifier was simultane-
fier gains were set for each channel for optimal signal ously used with the other changes to reduce the elec-
recording. Electrogram signals were filtered to pass trode potential by a factor of 1,000. During the shock,
0.1-500 Hz before and after attempted defibrillation.13 epicardial potentials were measured by the 75 button
Ten milliseconds before a shock, recording was switched electrodes, which included the three electrodes located
to unipolar at a second set of gains (previously deter- beneath the shock electrodes. The potential delivered to
mined to be appropriate for the voltage of the shock to the shock electrodes was also directly measured. Cur-
be delivered) with a low-pass filter of 500 Hz and rent delivered to each electrode pair for single, com-
high-pass filter direct-current coupled14 (Figure 1C). bined, and sequential electrode configurations was mea-
Ten milliseconds allowed adequate time to establish a sured on an oscilloscope. The voltage attenuator was
baseline for measuring potentials generated by the switched off 1 msec after the shock, and recording of
shocks.14 A voltage attenuator produced by a 1-Gfl bipolar electrograms was resumed (Figure 1C).
Wharton et al Defibrillation Gradients 1513

TABLE 1. Global and Field Parameters at Defibrillation Threshold


R:V L:V (R+L):V R:V--L:V
Global parameters
Voltage (V) 272.733.6 430.2102.9t 256.6+64.7 238.745.6
Current (A) 2.24+0.54* 4.101.47t 2.700.92* 1.900.36
Energy (J) 8.72.8* 26.417.7t 10.26.4 6.42.2
Field parameters
Minimum gradient (V/cm) 6.71.1 7.42.0 5.81.3 6.61.81
Maximum gradient (V/cm) 129.388.1 188.892.4t 136.9103.4 109.462.2t
R:V and L:V, single anode-single cathode configurations; (R+L):V, dual anode-single cathode configuration;
R:V-*L:V, sequential 7-msec shocks separated by 1 msec given to R:V and L:V.
*p<0.05 with respect to R:V--L:V.
tpO0.005 with respect to R:V->L:V.
:Represents field parameter from maximum combined gradient field (see text).

After completion of the study, the dog was killed by Potential gradients were calculated as described pre-
induction of ventricular fibrillation. The button and shock viously.4 Briefly, the nearest neighbors of each recording
electrode locations were marked with color-coded pins. electrode were determined, and the interelectrode dis-
Their locations were subsequently transcribed to a com- tance of each electrode to its neighbors was calculated
from their x-y-z coordinates. From the potential differ-
Downloaded from http://circ.ahajournals.org/ by guest on January 6, 2017

puter-generated, three-dimensional display of the dog


heart for generation of surface maps. ences and interelectrode distances between an elec-
trode and each of its nearest neighbors, a method that
Data Analysis minimized the summed squared error was used to
Data were digitized at 1,000 Hz and recorded on estimate the potential gradient at the recording elec-
videocassette for subsequent analysis.15,16 Shock poten- trode. The minimum and maximum gradient for each
tials at the median time of the shock were determined defibrillation electrode configuration for each dog was
for every recording electrode by a computer program determined from the lowest and highest gradient, re-
spectively, from any recording electrode for a given
with known calibration signals. Total delivered potential shock.
was calculated from the measured shock electrode Postshock epicardial activation times at each record-
potentials as the anodal potential or mean anodal ing electrode, referenced to the onset of the defibrilla-
potential for (R+L):V, minus the cathodal potential. tion shock, were determined by using the peak of
Total energy delivered was calculated for all dogs monophasic waveforms or the maximum potential
from the equation change of biphasic or multiphasic waveforms as the
J=VIt
point of activation. The earliest site of postshock ven-
tricular activation was defined as the recording elec-
where J is energy in joules, V is mean delivered trode site that first recorded activation after the shock.
potential in volts, I is mean current in amperes, and t is Other early sites were defined as electrode sites with
the duration of the shock in seconds. In five dogs, a activation times earlier than surrounding neighbor elec-
digital oscilloscope was used to store the measured trodes. Epicardial activation times, potentials, and po-
current and potential waveforms, and the power curve tential gradients were projected onto two-dimensional
was integrated over time to obtain the delivered energy. representations of the dog heart for visual analysis.
Energy measurements by this means were nearly iden- Data are expressed as meanSD. Repeated-mea-
sures analysis of variance17 was performed to compare
tical to those made with the above equation (r=0.9998; measured and calculated values between the multiple
slope, 0.99). electrode configurations. Pearson correlation coeffi-
Impedance was calculated from Ohm's law, assuming cients were used to compare different means of energy
a primarily resistive element to impedance in the range calculation and to compare potential fields to test the
of potentials used in this study (see "Appendix"): rule of superposition. A probability value less than 0.05
R= AE/I was considered significant.

For total impedance, AE equals the total delivered Results


potential. For electrode-myocardial interface imped- Total Voltage, Current, and Energy Required at
ance, AE is the potential difference from shock elec- Defibrillation Threshold
trode to the button electrode underneath it. Cardiac Of the 12 dogs studied, eight had their lowest DFT in
impedance was calculated as the difference between terms of energy with R: V->L: V, two with R: V, and
total and electrode-myocardial interface impedance. two with the (R+L):V electrode configuration. The
Because intracardiac and shunt current were not mea- sequential shock configuration required significantly
sured in the parallel circuit of (R+L):V, direct mea- lower current (p<0.05) for defibrillation than any other
surement of the associated cardiac impedance was not configuration and tended to require less energy: p= 0.07
possible; it was estimated by averaging the potentials compared with (R+L):V and pc0.05 compared with
measured underneath the anodal shock electrodes and R: V and L: V (Table 1). Voltage requirements for
assuming a single circuit impedance. defibrillation with the sequential shock were not signif-
1514 Circulation Vol 85, No 4 April 1992

icantly different from R:V and (R+L):V. R:V and Table 1. There was a 19.5-+f13.2-, 27.6 + 17.7-,
(R+L): V were similar in terms of energy and voltage, 23.3 15.3-, and 15.7+4.4-fold change between the min-
but (R+L): V required significantly more current than imum and maximum measured potential gradient at the
R:V (p<0.05). On the other hand, L:V was markedly DFT for R: V, L: V, (R+L): V, and the composite field
worse in all parameters. At DFT, mean voltage and (see below) of (R: V)->(L: V), respectively. Paralleling
current for L:V were both approximately 1.5 times the changes seen in the potential fields, potential gra-
higher and energy was approximately three times higher dient decline was most rapid near the shock electrodes
than for the other electrode configurations (pc0.005 and more gradual at sites more distant from the shock
for all parameters). electrodes. Relatively weak potential gradient fields
There were 329 shocks delivered: 81 R:V, 114 L:V, (less than 10 V/cm) were produced across the base of
80 (R+L):V, and 54 R:V->L:V. The mean numbers of the right ventricle by all electrode configurations used.
shocks necessary to obtain DFT for each of these R: V also generated weak fields across most of the base
electrode configurations were 72, 102, 73, and of the left ventricle and L: V across the posterior base of
51, respectively. Determination of the DFT for L:V the left ventricle. For (R+L): V, the gradient field was
required a significantly greater number of shocks than less than 10 V/cm across the base of both ventricles
for the other electrode configurations. except for a relatively small area surrounding the L
electrode (Figure 3B). Except for the area enhanced by
Potential and Potential Gradient Fields the L anode, the gradient field was not greatly different
The potential fields, potential gradient fields, and the from that of R: V, which may explain their similar
postshock activation sequences were mapped for all DFTs. Figure 4 graphically illustrates the distribution of
Downloaded from http://circ.ahajournals.org/ by guest on January 6, 2017

unsuccessful and successful shocks. For each electrode low gradient areas for each electrode configuration.
configuration, the potential and potential gradient fields As can be anticipated from the individual gradient
generated were similar among different experimental fields for R: V and L: V, neither component of the
animals. A representative example of maps of the sequential shocks obtained high gradients across the
potential field and potential gradient field for each right ventricular base or across portions of the base of
component of the sequential shock configuration is the left ventricle anteriorly and posteriorly near the
shown in Figure 2. Each component shock of right ventricle and distant from the L electrode. To
R:V->L:V was similar to R:V and L:V shocks of illustrate this, composite maps were generated by using
equivalent magnitude given singularly. Potential mea- the maximum gradient measured from either of the two
sured at the recording electrode immediately under- component shocks at each electrode site (Figure 2E).
neath the shock electrode was approximately two thirds This is justifiable because both 7- and 14-msec trun-
of the measured delivered potential for both compo- cated exponential shocks occur on the relatively flat
nents of the sequential shock. Mean potential decrease portion of the strength-duration curve10 and presum-
at the electrode-myocardial interface ranged from 27% ably should have approximately the same potential
to 29% of the applied potential for all of the electrode gradient field requirements for defibrillation. Such com-
configurations tested. Decline in potential across the posite maps demonstrated that neither of the sequential
myocardium was relatively symmetrical with respect to shocks created potential gradients greater than 10 V/cm
the interelectrode axis of each configuration and was across the base of the right ventricle. In some experi-
most rapid near the shock electrodes. As can be seen in ments, gradients greater than 10 V/cm were also not
Figures 2A and 2B, potential change was small across created by the sequential shocks across the anterior
the base of the right ventricle for both R: V and L: V. and/or posterior base of the left ventricle. However,
Potential change was also small across the entire base of 52+19% of recording electrodes with gradients less
the left ventricle with R: V shocks and across the than or equal to 10 V/cm during the R: V component of
posterior left ventricular base with L: V shocks. the sequential shock at DFT recorded gradients greater
When the combined anodal configuration (R+L):V than 10 V/cm during the L:V component. The mini-
was used, the potential field was similar to R: V except mum gradient for each component of the sequential
for a rapidly decreasing component which rapidly de- shock at DFT was 5.3+0.8 V/cm and 4.40.9 V/cm,
creased with distance from the left ventricular anode respectively. These were significantly less (p=0.012)
(Figure 3A). From the rule of superposition, the poten- than the minimum gradient of 6.61.8 V/cm for the
tial field generated by (R+L): V should be equal to the composite field. The ability of two sequential fields of
sum of the potential fields generated by R: V and L: V lower minimum gradient to generate a composite field
if current flow were equal between each component of with higher minimum gradient presumably accounts for
(R+L): V and the corresponding R: V and L: V shocks. the greater efficacy of the sequential shock compared
Equal currents (0.1 A) among all necessary compo- with either component alone. The minimum gradients
nent shocks were available for four shock strengths in at DFT for the R: V and L: V components of the
two dogs. The correlation between the actually mea- sequential shock were significantly less (p<0.01) than
sured (R+L): V potential field and the field calculated for those at DFI' for the respective single 14-msec
from superpositioning equicurrent R: V and L: V fields shocks. However, the minimum gradients at DFT for
for these four shocks was excellent (r=0.993-0.998). the composite field of the sequential shock and for the
The small size of the shock electrodes resulted in very fields of the single R: V and L: V shocks were similar.
inhomogeneous potential gradient fields (Figures 2C
and 2D). The unevenness of the potential gradient fields Minimum Gradient and Defibrillation Parameters
for each electrode configuration is quantitated by com- R:VL:V required significantly less and L: V re-
parison of minimum and maximum gradients at DFT in quired significantly more current to defibrillate than
Wharton et al Defibrillation Gradients 1515

A B
R:V -.L:V R:V -* L: V
+27
3
-II
6 -5 i`ij
10

-ill 14
a -B
24 1 - 25
7--0 -

_e X7/ -'37~
~ ~ ~ ~ ~

RI5-81
POTENTIAL FIELD POTENTIAL FIELD

2? 29 26 26
C S 467 -
Downloaded from http://circ.ahajournals.org/ by guest on January 6, 2017

GRADIENT FIELD GRADIENT FIELD

E
18
41T
Io
1~ ~ ~ ~ ~ ~
1~~~~~~~~~~~~~~~~~~~~I

II

'43~ ~~lr

MAXIMUM COMBINED GRADIENT FIELD POST-SHOCK ACTIVATION


FiGURE 2. Maps ofpotentialfieldsforsequential 208-VR:V(two single anode-single cathode configuration) (panel A) and 205-V
L: V (two single anode-single cathode configuration) (panel B) shocks are iUlustrated for an unsuccessful defibriUation attempt.
Left-hand figure of each panel represents anterior half and right-hand the posterior half ofthe ventricles. Numbers indicate location of
recording epicardial electrodes andpotential in volts (relative to the left leg) recorded at each site. Boxed-in numbers represent measured
potential delivered to shock electrodes: number at upper left-hand corner in panelA represents the anode at the right atrium (R); number
at left base in panel B represents the left basal anode (L); nuwberat left-hand apex ofbothpanelsA and B represent the left ventricular
apical cathode (V). Numbers immediately beneath the right atrial or to the left ofthe left basal anodalpotentials and above the cathodal
potentials indicate potentials measured by electrode immediately beneath shock electrodes. Isopotential lines are separated by 20 V Note
relative symmetry of isopotential lines; they approximately describe circles perpendicular to the long axis of the heart during the R: V
shock. During the L: V shock, there is also relative symmetry of the potential field along the axis between L and V Fields generated
between each 7-msec duration component of the sequential shock were similar to the corresponding 14-msec single shock of similar
deliveredpotentiaL Panels C and D: DerivedpotentialgradientfieldsforR: Vand L: Vcomponents ofsequential shock illustrated above
are shown. Numbers represent potential gradient at each site in V/cm. Lowest gradients ('10 V/cm) are enclosed in cross-hatched area.
Isogradient lines are separated by 20 V/cm and parallel the symmetry ofpotential fields. During R: V shock (panel C), gradient at
electrode beneath apical cathode was 48 V/cm. During the L: Vshock (panel D), gradient at electrode underneath left ventricular basal
anode was 48 Vlcm and underneath apical cathode was 43 V/cm. Panel E: To observe the effect of sequential shocks on improving
gradient distribution across the epicardium, maximum gradient from either shock at each electrode site was displayed. "Maximum
combined gradientfield" for sequential shock illustrated in panels C and D is shown in panel E. Note that extent of myocardium not
obtaining a gradient greater than 10 V/cm is substantially reduced compared with gradientfields for component shocks. Panel F: Map
ofepicardial activation after unsuccessful sequential shock. Numbers represent times in milliseconds when epicardial activation occurs
at the recording electrode, with time zero being the start of the shock. Filled circles represent sites of electrodes in which adequate biolar
recording of activation was not obtained. Isochronal lines are separated by 20 msec. Epicardial activation map shows three sites ofearly
epicardial activation arising along the atrioventricular groove in which neither shock electrode generated a strong field. Thick arrow
indicates site of earliest postshock epicardial activation; thin arrows indicate other somewhat later sites of epicardial activation.
Subsequently, there is rather uniform spread across the ventricles from these foci.
1516 Circulation Vol 85, No 4 April 1992

A R: V or (R+L): V (Table 1). In addition, the potential


(R+ l) :V fields generated were different for each electrode con-
figuration. Despite these differences, the minimum po-
+18 tential gradients measured at DFT for each of these
RV LV LV RV
electrode configurations were approximately 6-7 V/cm
and were not significantly different (Table 1). This
relation is also true for the composite field of the
sequential shock configuration; however, the individual
component shock fields had minimum gradients that
X
-~I
1 20-1-4 were somewhat lower (see above). The mean maximum
gradient for L:V was significantly larger than for the
other electrode configurations (p<O.OO9), although the
latter were not different from each other (Table 1).
Thus, despite the differences in electrode configuration,
FIELD in generated potential fields, and in voltage, current and
B energy required for defibrillation, there was not a
significant difference in the minimum potential gradient
generated by any of the electrode configurations at their
DFTs. This implies that there is a minimum potential
gradient that must be obtained over both ventricles for
electrical defibrillation in normal myocardium. For se-
Downloaded from http://circ.ahajournals.org/ by guest on January 6, 2017

quential shocks, this minimum gradient must be ob-


tained by at least one of the two shocks at each
ventricular site.
GRADIENT FIELD Sites of Epicardial Activation After
Subthreshold Shocks
There were 262 unsuccessful shocks, with an average
1211 of 2.31.0 early epicardial sites per shock. The number
04 1~~~~~~ of early sites did not differ significantly with the elec-
trode configuration used, as can be seen in Table 2.
Initial activation occurred where gradients were weak
for each electrode configuration (Figures 2F and 3C). In
general, early sites occurred where the gradients were
less than 10 V/cm (Figure 4). There was no evidence of
early epicardial activation occurring in high gradient
POST-SHOCK ACTIVATION areas after unsuccessful shocks. The distribution of
FIGURE 3. Maps. Panel A: Potentialfieldfor 220-Vsubthresh- potential gradients at the sites of early activation for all
old shock with the (R+L):V (dual anode-single cathode) unsuccessful shocks was approximately normal with a
electrode configuration from the same experimental animal as skew to the left (Figure 5). The mean potential gradient
in Figure 2 Potential delivered to right atrial anode was +35 V at sites of early activation after unsuccessful shocks was
(upper left-hand corner offigure), to the left ventricular anode, 9.55.4 V/cm; median was 7.9 V/cm. Ninety-eight per-
+36 V, and to the apical cathode, -184 V Potentials measured cent had gradients less than 25 V/cm, and 87% had
underneath respective shock electrodes were +18, +21, and gradients less than or equal to 15 V/cm. Fifty-two
-124 V Potential drop across anodes was similar. Potential
percent of the earliest sites were within 1 V/cm of the
minimum measured potential gradient for the corre-
drops rapidly around the L shock electrode with little distortion sponding shock. The largest discrepancy between min-
otherwise of the field generated by R: V (single anode-single imum measured gradient and the gradient at the site of
cathode configuration) (Figure 2A). Isopotential lines are thus earliest epicardial activation was 21.6 V/cm (7.7 and
relatively symmetrical across apical half of myocardium. Panel 29.3 V/cm, respectively), occurring with an unsuccessful
B: Gradientfieldforthe same (R+L): Vshock. Largest gradient sequential shock. Mean gradient at the sites of early
(82 Vkcm) is located beneath apical cathode. Gradient measured activation for each electrode configuration after the
underneath left ventricular anode was 34 V/cm. Gradientfield is highest subthreshold shocks is shown in Table 2.
weak across most of the base of the ventricles except for area
immediately surrounding the L electrode. Panel C: Map of Sites of Epicardial Activation After Successful Shocks
epicardial activity after shock illustrated in panel A reveals two Of the 67 successful shocks, 11 (16%) were followed
electrodes recording the same postshock activation time of 63 by the immediate resumption of supraventricular
msec, presunably resultingfrom an area of earlier activation at rhythm (type A defibrillation) and 55 (83%) were
a site between the two recording electrodes (thick arrow) prop- followed by one to several ventricular ectopic complexes
agating out to recording electrodes at approximately the same before the resumption of supraventricular rhythm (type
time. Two other earl sites (thin arrows) ofpostshock activation B defibrillation).5"18"19 Type A defibrillation occurred in
are seen, one at the base of the left ventricle and the other along six of 19 R: V shocks, none of 14 L: V shocks, four of 18
the right ventricular free wall both in areas in which gradient (R+L):V shocks, and one of 16 R:V->L:V shocks. Of
field is weak. the 55 defibrillations followed by transient ventricular
Wharton et al Defibrillation Gradients 1517

A. c
R R:V R (R+L):V

D
B. R R:V -> L:V
L :V
Downloaded from http://circ.ahajournals.org/ by guest on January 6, 2017

v v
FIGURE 4. Maps indicate sites of epicardial activation after all unsuccessful shocks between 4 J and the highest subthreshold
shockforR: V (single anode-single cathode configuration) (panel A), L: V (single anode-single cathode configuration) (panel B),
(R+L): V (dual anode-single cathode configuration) (panel C), and R: V-L: V (sequential 7-msec shocks separated by 1 msec
given to R: Vand L: V) (panel D). Relative locations of shock electrodes are indicated by double cross-hatched oval areas. Site(s)
of earliest activation are indicated with filled circles; other sites of early activation are indicated with filled triangles. Small branch
arteries are for illustrative purposes only and do not correspond to actual location of branch arteries. Cross-hatched and stippled
areas on each figure represent epicardial surface in which gradients less than 10 V/cm were measured during the highest
subthreshold shocks for the majority (greater than six) and the minority (less than or equal to six) of experimental animals,
respectively. Solid line represents border of area in which at least one dog had gradients less than 10 V/cm for the initial 4-J shocks.
Distance between solid line and apical border ofstippled area illustrates decrease in low gradient area needed to reach the highest
subthreshold shock. The much greater difference for L: Vshocks (panel B) than for other configurations between the solid line (4-I
shocks) and apical border of the stippled area (highest subthreshold shocks) is because of the much greater defibrillation threshold
for L: Vshocks. Figures graphically illustrate that early sites of activation for all electrode configurations occur mainly in areas in
which gradients are less than 10 V/cm and demonstrate similarity of generated fields with each electrode configuration within all
experimental animals. Note that despite the fact that L: Vhad the highest defibrillation threshold, epicardial surface area of lowest
gradients is smallest. Also note that for R: V shocks (panel A) but not (R+L): V shocks (panel C) that gradients are frequently
somewhat greater than 10 V/cm on the posterior right ventricle and that early sites of activation do not occur in this area for R: V
shocks but do for (R+L): V shocks. This is less clearly demonstrated for R: V-L: V shocks (panel D) perhaps because the
compositefield is used to generate the figure. Occurrence of several early sites of activation outside of cross-hatched or stippled areas
is caused primarily by their occurrence after shocks of considerably less total delivered potential (and larger area of low gradient)
than the highest subthreshold shock.

ectopic activity, there were 94 early sites or a mean of One episode occurred approximately 8 seconds after a
1.70.8 per successful shock. Ninety (96%) of these successful type B shock, with 6.5 seconds of intervening
early sites occurred in areas in which the gradient was sinus rhythm before the initiation of ventricular tachy-
weak; however, four (4%) occurred in high gradient cardia. The remaining two episodes occurred 81 and 328
areas (mean gradient, 7140 V/cm) at the perimeter of msec after type A defibrillating shocks. All episodes of
a shock electrode after L:V defibrillations only. Mean ventricular tachycardia arose from or near the perime-
gradient at the site of origin of epicardial activity after ter of ventricular shock electrodes in which gradients
all successful shocks followed by ectopic activity was were high (Figure 6). Mean gradient at the site of
15.4+15.4 V/cm (12.9+6.9 V/cm if the four high gradi- earliest epicardial activation of these ventricular tachy-
ent sites are omitted). cardias was 47.3+15.5 V/cm. Evidence supporting mac-
Eleven episodes of nonsustained, monomorphic ven- roreentry was not identified for any of these episodes of
ventricular tachycardia.
tricular tachycardia occurred after defibrillating shocks.
Eight of these episodes occurred after a pause of Myocardial and Electrode Impedance
167-852 msec after one to seven postshock, ectopic Electrode impedance accounted for approximately
activations originating in low gradient areas (Figure 6). one third to one fourth of total impedance for all
1518 Circulation Vol 85, No 4 April 1992

TABLE 2. Potential Gradients and Number of Sites of Initial Activation for Subthreshold Shocks 10% or Less Than
Defibrillation Threshold Energy
R:V L:V (R+L):V R:V->L:V
Minimum gradient (V/cm) 5.81.6 7.31.6 5.71.5 6.61.7*
Number of early sites 1.90.9 1.40.5 2.21.0 2.00.6
Gradient at sites of 9.04.5 12.26.7 11.85.9 12.66.7*
initiation (V/cm)
Maximum gradient (V/cm) 105.966.0 181.897.1t 130.197.5 99.352.4*
R:V and L:V, single anode-single cathode configurations; (R+L):V, dual anode-single cathode configuration;
R V-L: V, sequential 7-msec shocks separated by 1 msec given to R: V and L: V.
*Represents field parameter from maximum combined gradient field (see text).
tp<0.05 with respect to R:V-*L:V.

electrode configurations. The mean electrode, cardiac, from multiple sites, which was not done in the above
and total impedance at DFT was generally less with studies. Chen et a14 have attempted to directly measure
(R+L):V compared with the other configurations (Ta- the distribution of potentials generated across most of
ble 3), although the impedance for (R+L):V was the ventricles by defibrillation electrodes; however, the
indirectly calculated (see "Methods"). The smaller elec- fields recorded were from 1-2-V shocks delivered dur-
trode impedance for (R+L): V was presumably due to ing the fully repolarized period of sinus rhythm. Com-
the larger combined anodal surface area. At DFT, parison of the potential gradient field generated by the
Downloaded from http://circ.ahajournals.org/ by guest on January 6, 2017

electrode, cardiac, and total impedance were similar low-voltage shocks with the activation sequences re-
between R: V and the R: V component of the sequential corded immediately after high-voltage defibrillation
shock; however, all components of impedance were shocks showed that the continuation of ventricular
significantly higher for the L: V component of the fibrillation after unsuccessful shocks arose in areas in
sequential shock than for L: V shocks alone (p<O.O5). which the field was weak during low-voltage shocks.
Because impedance decreased as shock strength was Because recordings were not made during high-voltage
increased (see "Appendix"), the lower mean imped- shocks, quantitative field requirements for defibrillation
ances for the L: V shocks, compared with the L: V could not be defined. Furthermore, impedance and the
component of the sequential shock, may simply reflect fraction of delivered potential appearing across the
the larger shock strength at DFT for the single L: V ventricles are markedly different for 1-2-V shocks than
shocks. When L V shocks of similar voltage to the L: V for shocks of hundreds of volts (see "Appendix").
component of the sequential shock were compared, Mapping system modifications now allow the mea-
impedances were not significantly different. surement of high-voltage defibrillation shocks as well as
myocardial activation before and after shocks.5"4 Using
Discussion such a device, Witkowski et a15 recently showed that
Defibrillation presumably depends on achieving a sites of early activation after unsuccessful high-voltage
sufficient transsarcolemmal potential change across all, shocks occurred in areas in which the field is weak. By
or most, of the ventricular myocardium. The transsar-
colemmal potential change should be related to the

.rfl-1
extracellular potential gradient or current density.120 140
There is some indirect evidence to support the theory
that a minimum potential gradient or current density is
required across all or most of the myocardium to 120
achieve defibrillation. The efficacy of different paddle
sizes in transchest defibrillation and of two different
external shock electrode locations was shown to be
) uu
innj
r
proportional to the generated intracardiac potential u- 80)I
0
U)
gradients.2'3 However, the limited number of electrodes W
used in these studies did not allow determination of the :~> 60
myocardial distribution of gradients or of the possibility z
of a minimum gradient necessary for defibrillation.
Geddes et al,2' using a presumed uniform current 40
density field across an excised, perfused, whole heart in
vitro, showed that DFTs in terms of current density 20
were similar for three different waveforms of equal
duration. From these studies, Geddes et al estimated V_ W
0 41
A
8 12
1-
16 20 24
^ .
>28
-

that a minimum potential gradient of approximately 10 GRADIENT (V/cm)


V/cm was necessary for defibrillation with 10-msec
rectangular or trapezoidal waveforms, comparable with FIGURE 5. Bar graph shows distribution of gradients from all
the waveform used in this study. electrode configurations measured at sites of earliest postshock
Because the heterogeneous and anisotropic proper- activation (shaded area) and from ohr earl sites of epicardial
ties of the heart preclude uniform changes in potential, activation (unshaded area) aftershocks. Eariest and other earl
determination of the potential field produced across the sites were included from all unsuccessful shocks. Group
ventricles by the shock requires recording potential mean+SD=9.5+5.4 V/cm; range, 1. 7-35.2 V/cm.
Wharton et al Defibrillation Gradients 1519

A TABLE 3. Electrode, Cardiac, and Total Impedance at


Defibrillation Threshold
(R+L): V
RVOT Electrode Cardiac Total
-~
Anode: Cathode (fQ) (Q1) (fQ)
1 T----t- -r- 1

R:V 34.4+9.3 88.8+23.9 123.328.0


APEX L:V 28.410.1 79.712.2 108.1 18.5
(R+L):V 27.3 7.0 71.915.0 99.2 19.6
R: V-> 32.97.7 89.520.2 122.424.0
L: V 37.5 + 11.4 93.4 16.6 130.8+22.4
500ms.c
R:V and L: V, single anode-single cathode configurations;
(R+L):V, dual anode-single cathode configuration; R:V->L:V,
B sequential 7-msec shocks separated by 1 msec given to R : V and L: V.

gradient field is necessary over both ventricles for


14 defibrillation. Despite the marked differences in the
electrode configurations used and the differences in
their corresponding fields, all obtained similar minimum
gradients of approximately 6-7 V/cm at their DFTs.
This is similar to the minimum required potential
Downloaded from http://circ.ahajournals.org/ by guest on January 6, 2017

gradient estimated by Geddes et a121 and to that re-


ported by Witkowski et al.5 Despite the marked differ-
GRADIENT FIELD ence in total current, voltage, and energy required for
defibrillation by L: V compared with the other configu-
c
rations, the minimum gradients were not significantly
different. Also, the efficacy of R: V->L: V appears to be
related to its ability to achieve this minimum gradient
over both ventricles during at least one component of
..A 4. 474 5- the sequential shock, although the minimum gradient of
either component shock may be considerably less. Thus,
it appears that a minimum potential gradient is neces-
sary across all of the ventricular myocardium to achieve
4~~~~~~~~~6 defibrillation. Unfavorable electrode configurations re-
WRIEST SITE
quire large expenditures of energy or current to obtain
EPICARDIAL ACTIVATION this minimum gradient at sites in which their field
FIGURE 6. Panel A: Two-channel recording from right ven- strength is weak.
The fields generated by electrodes with small surface
tricular outflow tract (RVOT) and left ventricular apex adjacent areas relative to that of the heart are inhomogeneous, as
to cathode is shown for a successful 284-V shock with the illustrated by the fields generated by all of the electrode
(R+L): V (dual anode-single cathode configuration) shock configurations used in this study. Because most of the
from a different experimental animal from the one shown in potential is lost near the shock electrodes, high poten-
Figures 2 and 3. Recordings show approximately 500 msec of tials (and thus high energy or current) must be deliv-
ventricular fibrillation, which is interrupted by the shock. A ered to generate gradients distant to the shock elec-
single repetitive complex occurs after the shock that had its origin trodes that are sufficient for defibrillation. Electrodes
in the base of the right ventricle near the RVOT electrode. This with larger surface areas, such as those used clinically,
is followed by a regular uniform complex tachycardia with a should improve the distribution of gradients during
cycle length of approxunately 250 msec and with its origin in the shocks and improve defibrillation parameters.2 Small
left ventricular apex. Although recordings are bipolar after electrodes were used in this study to generate uneven
shock, note prominent ST segment elevation at apical recording gradient fields to enhance discrimination of defibrilla-
site, which is suggestive of myocardial injury. Panel B: Map tion field requirements and to allow sufficient distribu-
shows potential gradientfield for the same shock. Panel C: Map tion of epicardial recording electrodes for precise field
of second activation sequence after defibrillation shock shown in and activation mapping. The finding that a minimum
panel A shows origin of ventricular tachycardia from electrode gradient of 6-7 V/cm is needed for defibrillation should
adjacent to apical shock electrode. Isochronal lines are 10 msec also hold true for larger electrodes.
apart. Calculated gradient at this site was 57 V/cm (panel B). The DFT was not improved by coupling two anodes
Evidence of macroreentry in epicardial electrodes was not together with (R+L):V shocks. Chang et at22 reported
similar results with the use of an intracardiac catheter
identified. Rapid activation of most of the apex of the heart and either an epicardial or subcutaneous patch elec-
suggests an origin deep within ventricles such as from endocar- trode. In the present study, the potential gradient field
dium or septum. of the (R+L):V shocks was weak at the base of the
anterior right ventricle, as was true for the fields of the
determining the potential gradient field at DFT for R: V and L: V shocks. Thus, this particular combination
several different electrode configurations, the present of electrodes did not improve the distribution of gradi-
study provides direct evidence that a minimum potential ents across the ventricles. It is also possible that during
1520 Circulation Vol 85, No 4 April 1992

(R+L):V shocks, vector components of current flow plained by a number of factors, such as measurement
between the R and V electrodes and the L and V error and the lack of intramyocardial and endocardial
electrodes were oppositely directed, particularly be- recording sites.
tween the two anodes in the area of the atrioventricular Another factor explaining why earliest activation
groove, and that this may have resulted in vector after an unsuccessful shock is not always in the region
cancellation of these components. exposed to the weakest potential gradient is variation in
Sequential shocks with short shock separation times the electrophysiological state of the myocardium at the
to two electrode configurations using an intravascular time of the shock. Besides the spatial distribution of
catheter and patch-electrode configuration or a three- potential gradients and the temporal waveform of the
patch-electrode configuration have been shown to have shock, major determinants of the effect of a shock are
greater efficacy than shocks to single pairs of elec- probably the state of myocardial activation, refractori-
trodes.9'22-24 However, not all sequential shock elec- ness, and wavefront propagation at the time of the
trode configurations significantly alter the energy re- shock. Such factors may prevent the postshock appear-
quirements for defibrillation.22,25'26 The present study ance of ventricular fibrillation activation fronts at many
confirmed the greater efficacy of sequential shocks to sites whether or not the potential gradient of the shock
two different epicardial electrode configurations. The exceeds a minimum threshold. For example, when the
mechanism by which sequential shocks may lower the region in which the shock field is weakest is in its
DFI' is presumably due to improvement in the distribu- absolute refractory period at the time of the shock,
tion of potential gradients across the ventricles. The fibrillation will not reappear from this region even when
L: V component of the sequential shocks at the DFT the potential gradient is below threshold. At another
significantly decreased (by 52%) the number of record- point in time during fibrillation, this region may not be
Downloaded from http://circ.ahajournals.org/ by guest on January 6, 2017

ing electrodes that measured a gradient less than or absolutely refractory, and fibrillation activation fronts
equal to 10 V/cm during the preceding R: V component may originate from this region after the shock,19'27
of the sequential shock. Thus, the overall surface area of assuming that adjacent sites are capable of conducting
the ventricles that was exposed to a weak gradient field the impulse and that surrounding wavefronts do not
during the sequential shocks was substantially de- collide and terminate the reentrant circuit. Consider-
creased compared with the R: V shock alone. There was ation of these interactions may help to explain the
no evidence that the initial R: V shock changed imped- observation that there exists a range of voltages or
ance to the subsequent L: V shock to enhance current energies near threshold in which defibrillation is best
delivery of the latter. However, the minimum gradients described in terms of a probability function.28'29 The
of the composite gradient field of the sequential shocks occurrence of a shock strength with which defibrillation
were greater than the minimum gradients of each is always achieved and an upper limit of shock strength
component shock but were similar to the minimum that will not induce fibrillation no matter when it is
gradients produced by all of the other single-shock introduced into the vulnerable period27'30'31 suggests
electrode configurations at the DFT. This again suggests that there should exist a potential gradient above which
the necessity of obtaining a certain minimum gradient fibrillation will be terminated regardless of the state of
over all of the heart for defibrillation. Greater increases the myocardium. However, because the shock strength
in defibrillation efficacy may be obtained by designing that is always successful is usually greater than that at
electrode configurations in which one shock field is the DFT,28,29,31 the minimum gradient measured in this
strong where the other is weak, and neither is weak in study is probably less than that required to always
the same area. defibrillate.
None of the electrode configurations examined in this The occurrence of a minimum gradient at the DFT is
study produced adequate fields along the right ventric- most consistent with either the total ventricular depo-
ular base, especially along the right ventricular outflow larization (or extinction) hypothesis of defibrillation32 or
tract. One explanation for this finding is that the three the upper limit of vulnerability hypothesis,19'27'30 be-
electrodes describe a plane that is located posteriorly cause both theories predict a minimum required poten-
(Figure 1B), so that the right ventricular outflow tract, tial gradient across all of the ventricular myocardium.
which bulges anteriorly, lies the greatest distance from The finding of a required minimum gradient across all
this plane to produce an area in which the field is weak. of the ventricular myocardium is less consistent with the
This effect may not be as pronounced in the human critical mass hypothesis.18'33 The latter hypothesis pre-
heart because the right ventricular outflow tract does dicts that the minimum necessary gradient for different
not project as far anteriorly relative to this plane. electrode configurations should be equal, not over the
The present study confirms and expands on the entire ventricular myocardium but at the boundary
finding that ventricular fibrillation first appears after enclosing the critical mass. The absolute minimum
unsuccessful shocks in areas in which the field is weak, gradient would not necessarily have to be equal for
regardless of the electrode configuration used.4'5 different electrode configurations, because it would
Eighty-seven percent of gradients at the sites of early depend only on the rate of decline in the gradient field
epicardial activation after unsuccessful shocks were less beyond the border of the critical mass. The critical mass
than 15 V/cm, and 52% were within 1 V/cm of the for defibrillation has been estimated to be approxi-
minimum measured gradient. Sites of initiation of ven- mately 75% of the ventricular myocardium with pro-
tricular fibrillation in high gradient areas after unsuc- longed depolarization with potassium administration.33
cessful shocks were not seen. The lack of a closer If so, equal potential gradients should be found at the
correlation between the absolute minimum potential border of the regions enclosing 75% of the ventricular
gradient generated by a field and the gradient at the site myocardium for different electrode configurations, not
of resumption of ventricular fibrillation may be ex- at the minimum gradient value over all of the ventricles.
Wharton et al Defibrillation Gradients 1521

The findings in this study that 1) the L: V shock had a ventricles have the same field requirements were not
smaller surface area of low gradients compared with the assessed in this study.
other shocks (Figure 4), despite the fact that it was When only a certain minimum gradient is needed for
markedly less efficient in defibrillation, and 2) the defibrillation, electrode configurations that must gener-
minimum potential gradients are not significantly differ- ate very high gradients near the shock electrodes to
ent for the different lead configurations both run con- achieve this minimum gradient distantly not only waste
trary to the predictions of the critical mass hypothesis. energy but increase the risk of inducing variable degrees
However, because endocardial and septal gradient mea- of myocardial injury. High potential gradients may
sures are not available for determination of the three- cause cell injury and death, decreased contractility,
dimensional potential gradient field, it is possible that asystole, conduction block, and dysrhythmias.'37-42 Ex-
intramyocardial potential gradient distributions do not tremely high gradients may even generate intractable
parallel in a relatively symmetrical manner the distribu- ventricular fibrillation secondary to induced injury.30,40
tion measured on the epicardium and that a critical Despite the great unevenness of the fields generated by
the electrode configurations used in this study, post-
mass less than 100% existed but was not identified in shock arrhythmias occurred relatively infrequently.
this study. We hope that further developments in the There was no evidence that the high gradients at the
calculation of gradients across the entire volume of the shock electrodes directly initiated ventricular fibrillation
heart will help to resolve this issue.34 after unsuccessful shocks, although repetitive activity
A potential gradient greater than approximately 5 occurred infrequently in high gradient areas after suc-
V/cm for a 3-msec, low-tilt, monophasic shock delivered cessful shocks. Specifically, this occurred only after
during the relative refractory period of a passing wave- successful L: V shocks, which generated the highest
Downloaded from http://circ.ahajournals.org/ by guest on January 6, 2017

front induces conduction block, presumably by prolong- gradients for defibrillation, perhaps reaching a specific
ing refractoriness; gradients less than this do not pro- injury threshold. High-voltage shock induction of ven-
duce conduction block.35 This critical interaction tricular fibrillation (type II ventricular fibrillation)
between refractoriness and decreasing field strength probably requires voltages higher than those used in this
may lead to the induction of reentry and ventricular study.3040 Although shock-induced arrhythmias may oc-
fibrillation. According to the upper limit of vulnerability cur clinically41 and have been shown to occur more
hypothesis, subthreshold shocks will halt all fibrillation frequently after high-voltage shocks in vitro,'37'40 the
wavefronts but reinitiate ventricular fibrillation, pre- present study is the first to map the origin of postshock
sumably by a mechanism similar to that just described.19 ventricular tachycardia from the site of the shock elec-
The similarity between the minimum gradient of 6-7 trode, which in all cases was recorded from electrodes at
V/cm for defibrillation found in this study and the the perimeter of the shock electrode possibly reflecting
critical gradient of 5 V/cm for electrical induction of the higher current density at the electrode edge.43 The
fibrillation in the vulnerable period of regular rhythm is mechanism of the postshock ventricular tachycardia
consistent with but does not prove this hypothesis. seen in this study could not be ascertained; evidence of
There are several limitations to this study. Wave- epicardial macroreentry was not seen.
forms of depolarization are obscured during defibrilla- The ability to measure shock potentials also allows
tion by the shock and immediately after the shock by the measurement of various components of impedance,
baseline deflection generated by switching off the map- such as that at the electrode-myocardial interface or
ping system modification; thus, depolarization during or across the heart itself (see "Appendix"). The imped-
immediately after the shock cannot be identified or ance measured in this study is generally greater than
mapped. Furthermore, the lack of septal, intramyocar- that reported by others for transmyocardial shocks,44,45
dial, and endocardial recording sites limits the resolu- presumably reflecting the small size of the shock elec-
tion of the mapping and the determination of all trodes used in this study. Unlike a previous study,27 the
components of the gradients. Specifically, the lack of present study did not show significant changes in im-
transmural recordings may decrease the absolute value pedance with increasing shock voltage for R: V,
of the measured gradients. For low-voltage shocks to an (R+L): V, or the components of R:V-L: V. This may
electrode array resembling R: V, transmural gradients be due to the random sequence of shock delivery and/or
in the right ventricular outflow tract are approximately to the relatively small range of shock voltages delivered
half the tangential gradient.4 The exposure of the to these configurations. There was a significant change
anterior surface of the heart to air in the open-chest dog with L: V, which, because of its decreased efficacy,
preparation may increase the anterior gradient mea- required a larger number of shocks over a greater range
surements due to a boundary effect; however, the of voltages to achieve a DFT. When R: V shocks were
greater tangential current flow would improve gradient given over a greater range of voltages, marked changes
calculation using epicardial recording electrodes. Mea- in impedance were noted (see "Appendix"). However,
surement of rapidly changing gradients near shock Lawrence et al44 were not able to show a change in
electrodes requires small interelectrode distances to impedance between shocks of different strengths when
maintain accuracy; irregularities in high gradient areas more than 5 minutes elapsed between shocks and
in this study may reflect inadequate spacing of recording showed only a very small change when the shocks were
electrodes or inhomogeneous current distribution separated by a few seconds.
around the perimeter of the shock electrode. Depolar- The ability to map the potential gradient field gener-
ization of ventricular myocardium is dependent on fiber ated by shock electrodes should lead to marked im-
orientation relative to the vector of the potential gradi- provements in their design and function. If the desired
ent.36 Whether fiber orientation affects the field require- goal is a uniform field slightly exceeding a certain
ments for defibrillation or whether different areas of the minimum gradient, the possible electrode combinations
1522 Circulation Vol 85, No 4 April 1992

are almost limitless; however, application of the rule of


superposition may lead to a more efficient way of testing 400
potential fields for numerous electrode configurations
with a relatively limited number of shocks. Further
description of factors involved in electrode impedance : 3000
may also result in marked improvement in electrode
design. If electrodes, electrode configurations, and U
z 4V
shock sequences can be designed that will generate a < 200
more even field that only slightly exceeds the minimum
Lin
gradient required for reliable defibrillation without in-
creasing electrode impedance, implantable defibrilla- 100
F
1
tors then can be designed that will require less energy to 2 -25V
operate. This in turn will have major benefits, such as 3 -
o 4-7
increasing the lifetime and decreasing the size of the
battery of implanted units and decreasing the pain, 1 V10 100 100J
potential cardiac dysfunction, and dysrhythmias associ-
V
1VOL 100(V 1000 v

VOLTAGE (V)
ated with defibrillation shocks.
FIGuRE 7. Graph of calculated total impedance or mean
Appendix change in potential divided by total delivered current between
Effect of Shock Strength on Impedance successive rows of electrodes across the ventricles as a function
In an earlier study4 with low-voltage shocks from 1 to 2 V of logarithm of delivered potential from R: V (single anode-
Downloaded from http://circ.ahajournals.org/ by guest on January 6, 2017

given to R:V using 4.5-cm2 titanium mesh electrodes, only single cathode configuration) shocks of 1-1,000 Logarith- V

173% of the delivered shock potential was measured on the mic scale wasused to emphasize abrupt changes at low
ventricles, whereas in the present study, recorded myocardial delivered potential. Line labeled "Total" represents relation of
potential was 76.46.3% of that delivered with 2.64-cm2 total impedance and delivered potential. Lines labeled 1-7
electrodes. The addition of recording electrodes beneath the represent changes between successive rows of electrodes pro-
shock electrodes and inclusion of the voltage drop across the gressively cephalad to apical shock electrode. Thus, line 1
atria in this study may explain some but not all of this
difference. To see if part of this discrepancy was caused by represents impedance between shock electrode and recording
differences in impedance for low- and high-voltage shocks, an electrode immediately underneath (row 1). Line 2 represents
additional dog was studied with the same surgical preparation estimated impedance changes between recording electrode un-
as described in "Methods." However, 4.5-cm2-round titanium derneath apical shock electrode and ring of electrodes at
mesh shock electrodes with a recording button electrode perimeter of apical shock electrode (row 2). Lines 3-7 represent
underneath were attached in an R: V configuration. Shocks of estimated impedance changes between successive rows progres-
14-msec duration were delivered once each at increasing sively cephalad to recording electrodes at perimeter of shock
strengths as follows: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 12.5, 15, 20, 25, electrode. For further discussion, see 'Appendix."
50, 75, 100, 150, 200, 300, 400, 500, 750, and 1,000 V. Shocks
from 1 to 25 V were delivered by a constant voltage device and
from 25 to 1,000 V by the device used in the present study. Two
calibrated input dynamic ranges (10 V and 500 V) to the diately encircling (row 2) the apical shock electrode, respec-
amplifiers were used to record potentials. Current was mea- tively, and the electrode beneath the right atrial anode. These
sured on an oscilloscope. measurements were not significantly changed at 4 V and were
With 1-4-V shocks, total impedance increased markedly from in accordance with the previous study.4 As delivered potential
254 to 434Q1, decreased abruptly to 273Q1 at 5 V, and then increased from 5 V, 25 V, 100 V, and 1,000 V, the percent of
gradually declined with increasing potential down to 106 fl at delivered potential measured in row 1 steadily increased to
1,000 V. A plot of the logarithm of potential delivered versus the 48%, 66%, 73%, and 75% and in row 2 from 20%, 32%, 34%,
calculated total impedance illustrates the marked change in and 43%, respectively.
impedance at low voltages (Figure 7). Beyond approximately 10 The current waveform for shocks less than V showed a 5
V, the relation between impedance and the logarithm of deliv- logarithmic increase in current flow during the period of the
ered voltage became more linear. shock, suggesting a large contribution of capacitance to the
To define where the change in impedance occurred, mean overall impedance. This capacitive element may explain the
differences in potential between successive rows of electrodes initial increase in total impedance. As potential was increased
starting at the apical shock electrode were calculated and above 5 V, the current waveform showed progressively less
divided by the total measured current as an estimate of distortion and impedance decreased, primarily at or near the
effective impedance of the layers of ventricular myocardium electrode-myocardial interface. This strong capacitive element
between each row of electrodes. The greatest impedance was to impedance may be generated to some extent by electrode
calculated between the shock electrode and the recording polarization at the myocardial interface during low-voltage
electrode underneath it (row 1) for shocks of 1-20 V (Figure shocks. However, the occurrence of similar changes in esti-
7). At 25 V, however, the impedance at row 1 and the next mated impedance distant to the apical shock electrode implies
most cephalad row (row 2) equalized with further increases in myocardial changes occurring as a function of increasing
delivered potential causing the greatest impedance change at potential delivered. This capacitive component of impedance
row 2, principally because impedance at row 1 continued to progressively decreased with increasing delivered potential so
decrease, whereas it was relatively stable at row 2. The marked that in the range of potentials necessary for defibrillation with
increase in impedance at 4 V is reflected in the impedance epicardial electrodes, impedance appears to be primarily
curves at all levels of the ventricles, although most of this resistive, confirming the work of others.45
occurs in the apical third of the myocardium in rows 1-3
(Figure 7). Acknowledgments
For a 1-V R: V shock, 36% and 15% of delivered potential We would like to thank Joseph C. Greenfield Jr. for his
was measured between electrodes beneath (row 1) and imme- support and advice; and Ellen G. Dixon, Sharon B. Melnick,
Wharton et al Defibrillation Gradients 1523
Yohannes Afework, Alton T. Ledford, and Cloyce M. Lassiter 23. Jones DL, Klein GJ, Guiraudon GM, Sharma AD, Kallok MJ,
for their technical assistance. Bourland JD, Tacker WA Jr: Internal cardiac defibrillation in man:
Pronounced improvement with sequential pulse delivery to two
References different lead orientations. Circulation 1986;73:484-491
24. Bourland JD, Tacker WA Jr, Wessale JL, Kallok MJ, Graf JE,
1. Lepeschkin E, Jones JL, Rush S, Jones RE: Local potential Geddes ME: Sequential pulse defibrillation for implantable
gradients as a unifying measure for thresholds of stimulation,
standstill, tachyarrhythmia and fibrillation appearing after strong defibrillators. Med Instrum 1986;20:138-142
capacitor discharges. Adv Cardiol 1978;21:268-278 25. Wetherbee JN, Chapman PD, Bach SM Jr, Troup PJ: Sequential
2. Hoyt R, Grayzel J, Kerber RE: Determinations of intracardiac shocks are comparable to single shocks employing two current
current in defibrillation: Experimental studies in dogs. Circulation pathways for internal defibrillation dogs. PACE 1988;11:696-703
1981;64:818-823 26. Mehra R, Maracaccini S: Comparison of sequential and simulta-
3. Kerber RE, Hoyt R, Aronson A, Kieso R, Melton J: Evaluation of neous pulse defibrillation threshold with a non-epicardial elec-
a new defibrillation pathway: The tongue-epigastric route: trode system (abstract). Circulation 1986;74(suppl II):II-184
I. Experimental studies in dogs. JAm Coll Cardiol 1983;2:966-972 27. Chen P-S, Shibata N, Dixon EG, Martin RO, Ideker RE: Com-
4. Chen P-S, Wolf PD, Claydon FJ, Dixon EG, Vidaillet HJ Jr, parison of the defibrillation threshold and the upper limit of
Danieley ND, Pilkington TC, Ideker RE: The potential gradient ventricular vulnerability. Circulation 1986;73:1022-1028
field created by epicardial defibrillation electrodes in dogs. Circu- 28. Gold JH, Schuder JC, Stoeckle H: Contour graph for relating
lation 1986;74:626-636 percent success in achieving ventricular defibrillation to duration,
5. Witkowski FX, Penkoske PA, Plonsey R: Mechanism of cardiac current, and energy content of shock. Am Heart J 1979;98:207-212
defibrillation in open-chest dogs with unipolar DC-coupled simul- 29. Davy JM, Fain ES, Dorian P, Winkle RA: The relationship
taneous activation and shock potential recordings. Circulation between successful defibrillation and delivered energy in open-
1990;82:244-260 chest dogs: Reappraisal of the "defibrillation threshold" concept.
6. Babbs CF: Effect of pentobarbital anesthesia on ventricular Am Heart J 1987;113:77-84
defibrillation threshold in dogs. Am Heart J 1978;95:331-337 30. Lesigne C, Levy B, Saumont R, Birkui P, Bardou A, Rubin B: An
7. Galindo AH, Davis TB: Succinylcholine and cardiac excitability. energy-time analysis of ventricular fibrillation and defibrillation
Downloaded from http://circ.ahajournals.org/ by guest on January 6, 2017

Anesthesiology 1962;23:32-40 thresholds with internal electrodes. Med Biol Eng 1976;14:617-622
8. Worley SJ, Ideker RE, Mastrotatoro J, Smith WM, Vidaillet HJ Jr, 31. Wharton JM, Richard VJ, Murry CE Jr, Dixon EG, Reimer KA,
Chen P-S, Lowe JE: A new sock electrode for recording epicardial Meador J, Smith WM, Ideker RE: Electrophysiologic effects in
activation from the human heart: One size fits all. PACE 1987;10: vivo of monophasic and biphasic stimuli in normal and infarcted
21-31 dogs. PACE 1990;13:1158-1172
9. Jones DL, Klein GJ, Kallok MJ: Improved internal defibrillation 32. Wiggers CJ: The physiologic basis for cardiac resuscitation from
with twin pulse sequential energy delivery to different lead orien- ventricular fibrillation: Method for serial defibrillation. Am Heart J
tations in pigs. Am J Cardiol 1985;55:821-825 1940;20:413-422
10. Bourland JD, Tacker WA Jr, Geddes LA: Strength-duration 33. Zipes DP, Fischer J, King RM, Nicoll A, Jolly WW: Termination
curves for trapezoidal waveforms of various tilts for transchest of ventricular fibrillation in dogs by depolarizing a critical amount
defibrillation in animals. Med Instrum 1978;12:38-41 of myocardium. Am J Cardiol 1975;36:37-44
11. Babbs CF, Whistler SJ, Yim GKW: Temporal stability and preci- 34. Tang ASL, Wolf PD, Claydon FJ III, Smith WM, Pilkington TC,
sion of ventricular defibrillation threshold data. Am J Physiol Ideker RE: Measurement of defibrillation shock potential distri-
1978;235:H553-H558 butions and activation sequences of the heart in three dimensions.
12. Hoffman BF, Suckling EE, Brooks CM: Vulnerability of the dog Proc IEEE 1988;76:1176-1186
ventricle and effects of defibrillation. Circ Res 1955;3:147-151 35. Frazier DW, Wolf PD, Wharton JM, Tang ASL, Smith WM,
13. Barr RC, Spach MS: Sampling rates required for digital recording Ideker RE: Stimulus-induced critical point: Mechanism for the
of intracellular and extracellular cardiac potentials. Circulation electrical initiation of reentry in normal canine myocardium. J Clin
1977;55:40-48 Invest 1989;83:1039-1052
14. Wolf PD, Wharton JM, Wilkinson CD, Smith WM, Ideker RE: A 36. Frazier DW, Krassowska W, Chen P-S, Wolf PD, Dixon EG, Smith
method of measuring cardiac defibrillation potentials, in Proceed- WM, Ideker RE: Extracellular field required for excitation in
ings of the 39th Annual Conference on Engineering in Medicine and three-dimensional anisotropic canine myocardium. Circ Res 1988;
Biology, 1986, p 4 63:147-164
15. Wolf PD, Danieley ND, Ideker RE, Smith WM: A digital tape 37. Jones JL, Lepeschkin E, Jones RE, Rush S: Response of cultured
recorder for electrophysiologic waveforms, in Proceedings of the myocardial cells to countershock-type electric field stimulation.
38th Annual Conference on Engineering in Medicine and Biology, Am J Physiol 1978;235:H214-H222
1985, p 124 38. Jones JL, Proskauer CC, Paull WK, Lepeschkin E, Jones RE:
16. Smith WM, Ideker RE: Computer techniques for epicardial and Ultrastructural injury to chick myocardial cells in vitro following
endocardial mapping. Prog Cardiovasc Dis 1983;26:15-32 "electric countershock." Circ Res 1980;46:387-394
17. SAS/STAT User's Guide, Release 6.03 Edition. Cary, NC, SAS Insti- 39. Koning G, Veefkind AH, Schneider H: Cardiac damage caused by
tute, Inc, 1988 direct application of defibrillation shock to isolated Langendorff-
18. Mower MM, Mirowski M, Spear JF, Moore EN: Patterns of perfused rabbit heart. Am Heart J 1980;100:473-482
ventricular activity during catheter defibrillation. Circulation 1974; 40. Peleska B: Cardiac arrhythmias following condenser discharges
49:858-861 and their dependence upon strength of current and phase of
19. Chen P-S, Shibata N, Dixon EG, Wolf PD, Danieley ND, Sweeney cardiac cycle. Circ Res 1963;13:21-32
MB, Smith WM, Ideker RE: Activation during ventricular defibril- 41. Donoso E, Cohn LH, Friedberg CK: Ventricular arrhythmias after
lation in open-chest dogs: Evidence of complete cessation and precordial electric shock. Am Heart J 1967;73:595-601
regeneration of ventricular fibrillation after unsuccessful shocks. 42. Yabe S, Smith WM, Daubert JP, Wolf PD, Rollins DL, Ideker RE:
J Clin Invest 1986;77:810-823 Conduction disturbances caused by high current density electric
20. Krassowska W, PiLkington TC, Ideker RE: Periodic conductivity as fields. Circ Res 1990;66:1190-1203
a mechanism for cardiac stimulation and defibrillation. IEEE Trans 43. Kim Y, Fahy B, Tupper BJ: Optimal electrode designs for elec-
Biomed Eng 1987;34:555-560 trosurgery, defibrillation, and external cardiac pacing. IEEE Trans
21. Geddes LA, Niebauer MJ, Babbs CF, Bourland JD: Fundamental Biomed Eng 1986;33:845-853
criteria underlying the efficacy and safety of defibrillation current 44. Lawrence JH, Brin KP, Halperin HR, Platia EV, Tsitlik JE, Levine
waveforms. Med Biol Eng Comput 1985;23:122-130 JH, Guarnieri T: The characterization of human transmyocardial
22. Chang MS, Inoue H, Kallok MJ, Zipes DP: Double and triple impedance during implantation of the automatic internal cardio-
sequential shocks reduce ventricular defibrillation threshold in verter defibrillator. PACE 1986;9:745-755
dogs with and without myocardial infarction. J Am Coll Cardiol 45. Kugelberg J: The interelectrode resistances at defibrillation. Scand
1986;8:1393-1405 J Thorac Cardiovasc Surg 1972;6:274-277
Cardiac potential and potential gradient fields generated by single, combined, and
sequential shocks during ventricular defibrillation.
J M Wharton, P D Wolf, W M Smith, P S Chen, D W Frazier, S Yabe, N Danieley and R E
Ideker
Downloaded from http://circ.ahajournals.org/ by guest on January 6, 2017

Circulation. 1992;85:1510-1523
doi: 10.1161/01.CIR.85.4.1510
Circulation is published by the American Heart Association, 7272 Greenville Avenue, Dallas, TX 75231
Copyright 1992 American Heart Association, Inc. All rights reserved.
Print ISSN: 0009-7322. Online ISSN: 1524-4539

The online version of this article, along with updated information and services, is located on
the World Wide Web at:
http://circ.ahajournals.org/content/85/4/1510

Permissions: Requests for permissions to reproduce figures, tables, or portions of articles originally
published in Circulation can be obtained via RightsLink, a service of the Copyright Clearance Center, not the
Editorial Office. Once the online version of the published article for which permission is being requested is
located, click Request Permissions in the middle column of the Web page under Services. Further
information about this process is available in the Permissions and Rights Question and Answer document.

Reprints: Information about reprints can be found online at:


http://www.lww.com/reprints

Subscriptions: Information about subscribing to Circulation is online at:


http://circ.ahajournals.org//subscriptions/

You might also like