Professional Documents
Culture Documents
Not Applicable
Issue Authorization
Donald Smith,
Quality Assurance Director
Digitally signed by Donald Smith
DN: cn=Donald Smith, o=Mistras
Services, ou=Quality Assurance Director,
email=donald.smith@mistrasgroup.com
, c=US
Table of Contents
1.0 Scope ............................................................................................................................................. 3
2.0 Specific Applicability ...................................................................................................................... 3
3.0 Reference Documents .................................................................................................................... 4
4.0 Definitions...................................................................................................................................... 4
5.0 Personnel Requirements ................................................................................................................ 5
6.0 General Requirements ................................................................................................................... 5
7.0 Pre-inspection Analyzer Start-up and Calibration ......................................................................... 6
8.0 Field Testing Requirements ............................................................................................................ 8
9.0 Welds New and Existing Inspection ............................................................................................ 8
10.0 Undersize or Non-contact Samples ................................................................................................ 9
11.0 Instrument Measurement Quality Control................................................................................... 10
12.0 Certification Stamps ..................................................................................................................... 10
13.0 When and How to Use an Approved Stamp ................................................................................. 10
14.0 Resolution of Material Testing Discrepancies .............................................................................. 11
1.0 Scope
1.1 This procedure establishes a standardized methodology for the Positive Material Identification (PMI)
analysis of materials, such as stainless and alloy steels, utilizing wavelength dispersive X-ray
Fluorescence Spectrometry for the determination of elements with an atomic weight greater than that
of Titanium (Ti).
NOTE: Instruments such as the Niton GOLDD may extend the range of detectable elements down to
Aluminum (Al).
1.2 This procedure may be used to assist clients in the rapid, multi-element determination of a materials
composition for the purpose of:
a. Material identification
b. Procurement compliance
c. Product quality and specification compliance
d. Manufacturing control
e. Assisting in a Clients PMI Program
1.3 This procedure meets the requirements of ASTM E572-02a (06).
2.0 Specific Applicability
2.1 Mistras Services conducts PMI testing services as a means to provide Clients the necessary information
for the purposes identified in paragraph 1.2. Mistras Services does not maintain a Clients PMI Program.
The Client is responsible for identifying on the purchase order or contract document(s) the items
requiring PMI. The following items typically require PMI however the Client shall identify the actual
items on the purchase order or contract document.
a. All pressure-containing welds
b. All pressure-containing piping components, including welds
c. Pressure-containing components and welds of heat exchangers and other pressure vessels
d. Tubular products used in the fabrication of equipment such as heat exchangers, boilers, and heaters
e. Furnace heaters, and boiler internals (tube hangers, tube supports, baffles, and tube-sheets)
f. Plate material
g. Pressure-containing machinery casings such as pumps, compressors, and process expander drivers
h. Pressure-containing instrument housing
i. Internal metallic linings & cladding, and weld overlay used for protection against corrosive
environments
j. Expansion joints and bellows
k. Tubing (except as excluded in Paragraph 2.2, item G)
l. All stud bolts included in the ASTM A-I 93 and A-320 classification
m. Any other components or materials specifically designated for PMI on the purchase order
2.2 The following items are typically exempted from PMI testing but can be included if so directed by the
Client.
a. Gaskets
b. Internal machinery parts.
c. Internal instrument parts.
d. Internal non-pressure containing baffles, trays, tray clips, supports, pallrings, support rings, etc.
e. Non-pressure containing welds.
f. Internal valve components.
g. Type 304 SS tubing in outside diameter and less when properly identified by paint stencil
(exchanger tubing is not included in this exemption).
h. Compression-type ferrules and fittings for use with in outside diameter and smaller tubing.
3.0 Reference Documents
The following documents of the issue in effect form a part of this procedure to the extent specified herein:
3.2 ASME Boiler and Pressure Vessel Code, Section II A, B, C, 2010 edition, 2011 addenda.
3.3 ASTM- E572 -02a(06): Standard Test Method for Analysis of Stainless and Alloy Steels by X-ray
Fluorescence Spectrometry.
3.4 API RP 578, First edition: Material Verification Program for New and Existing Alloy Piping Systems.
4.0 Definitions
4.1 C of C: Certificate of Compliance or Certificate of Conformance
4.2 Client: The purchaser/ requestor of PMI services from Mistras Services. This also includes the Clients
representative i.e. Inspector.
4.3 Inspector: The term inspector, as used in this standard, refers to the Client's representative.
4.4 Inspection Lot: An inspection lot is a group of components from the same heat number from which a
sample is drawn.
4.5 Material Nonconformance: A positive material identification (PMI) test result that is not consistent with
the selected or specified alloy.
4.6 Mill Test Report: A certified document that permits each component to be identified according to the
original heat of material from which it was produced and identifies the applicable material specification
(including documentation of all test results required by the material specification).
4.7 PMI: Positive Materials Identification: A procedure used to assure that specified metallic alloy materials
for use in the refineries, chemical plants, and other facilities are properly identified as to their true
elemental composition and are marked according to specific industry standards and Client's procedures.
4.8 Representative Sample: A representative sample consists of one or more units (pieces) selected at
random from the inspection lot, which will be examined to determine the acceptability of the inspection
lot.
4.9 Certified Reference Materials: Sample materials for which laboratory chemical analysis data are available
and are used in demonstrating test instrument accuracy and reliability.
b. For lots greater than five units, the greatest of five units or 5% shall be analyzed unless
otherwise dictated by the client.
6.7 Equipment and materials
6.7.1 XRF Spectrometry equipment shall meet the requirements of reference 3.3. The Client shall
have final approval of equipment used on their product.
6.7.2 Equipment used for testing shall have factory calibration certificates on file.
6.7.3 Rental equipment shall be supplied with a factory calibration certificate. Copies of calibration
certificates shall be kept on file at Mistras Services.
6.7.4 Reference standards shall be Certified Reference Materials, such as those available from
Analytical Reference Materials International (ARMI), and shall be supplied with their respective
certified test reports. The reference standards shall be homogenous and free of voids, porosity,
and surface contamination.
7.0 Pre-inspection Analyzer Start-up and Calibration
7.1 Start-up and Stabilization
The Analyzer will be turned on and user will log in with the designated password. The instrument should
be allowed to run 1 minute to stabilize the electronics before proceeding.
7.2 Perform Instrument Calibration
7.2.1 A daily detector calibration should be performed after the 1 minute initial instrument
stabilization period. Calibrate the analyzer by pressing the Utilities screen icon, then the
Calibrate Detector button. This will start the instrument auto-calibration routine. When the
instrument completes the auto-calibration routine, return to the menu and press the Test
screen icon and then measure an approved alloy test standard, such as the 1 Cr - Mo alloy
standard supplied with the analyzer by the manufacturer. If the sample is correctly identified
and the Major alloying elements concentration values (Cr and Mo primarily) read within the
calculated acceptance limits (within 3-Sigma of the average value of the factory readings),
testing may commence.
7.2.2 If the check standard values do not read correctly, then re-calibrate the instrument as specified
in Section 7.2.1. If after re-calibrating the readings still do not fall within the 3-Sigma accuracy
boundaries, contact a supervisor and discontinue use of the instrument.
7.3 The need for a Hot Surface Adapter should be determined before analyzing any unknown samples. If
surface temperatures of components to be tested will exceed 315C/600F, but are below 427C/800 F,
the operator may operate the analyzer so there is a 6 mm (~ inch) air gap between the front of the
instrument measuring plate and the sample. When testing hot surfaces with the Weld Adapter in place,
the operator should use double the standard measuring time to maintain measurement precision (this
compensates for signal loss due to the air gap and the smaller 3mm Weld Adapter aperture).
Measurement Accuracy compensation is handled automatically by the software.
7.4 If measurements are to be taken on vibrating surfaces, like a hot sample surface, the analyzer should be
held a few millimeters (~in) away from the vibrating surface, and the measurement time doubled to
maintain precision. All measurement concentration accuracy compensation (for the air gap) is handled
automatically by the software (up to a maximum of 6mm).
7.5 The analyzer should be checked for material-specific performance prior to doing any testing. Typically
this should be done using a Certified Reference Material (CRM), such as an ARMI1 CRM. The CRM(s)
used in this Procedure should be representative of each of the materials expected to be encountered in
the inspection process.
9.3 At the discretion of the inspector, as in the case of large diameter welds or at any other time deemed
necessary to ensure correctness of the weld material, additional locations on the weld may be tested.
9.4 Care must be taken in testing welds to position the analyzer window to read only the weld material, and
not any adjacent piping component material(s), to avoid false readings.
9.4.1 If not equipped with the optional IsoWeld feature, and if the weld bead is smaller than 10 mm,
insert the Weld Adapter and activate the Weld Mode3 with the Weld icon. Alternatively, if
the IsoWeld option is installed, select the IsoWeld icon to reduce the measuring area to 3mm by
software function. Whichever aperture reduction method is used, be sure it reduces the
measurement aperture width to a size smaller that the width of the weld to be tested, or;
9.4.2 Isolate the weld by masking adjacent metal with a non-readable material such as several layers
of pure aluminum tape (care should be taken that the aluminum mask material contains no
interfering elements, is not an aluminum alloy or has surface contamination, and the mask
material is thick enough to block any x-rays coming from the underlying material).
9.4.3 Quadruple the measurement time lengths when using the Weld Adapter; e.g., increase the
measurement time from 5 seconds to 20 seconds for material ID and from 10 seconds for
chemistry to 40 seconds, to compensate for the added air gap and/or reduced analysis area.
9.5 Procedures for Testing of Welding Consumables
9.5.1 Each and every box of welding rod to be used for the job will be prepared by first puddling a
sample, using at least two passes, and grinding the resulting bead smooth for testing.
9.5.2 Documentation for Weld Rod Testing should include:
a. Rod designation
b. Rod Manufacturers Name
c. Box Lot Number
d. Elemental Composition of Samples
9.5.3 Rod Box to be Marked PMI Confirmed with:
a. Name of authorized person who performed the testing
b. Date of Testing
c. Other Information Specified (e.g., vendor or contactor ID)
9.5.4 GTAW (TIG) wire can often be tested directly from the box either individually with a Weld
aperture or IsoWeld, or by using several bundled wire pieces to cover more of the aperture
window of the analyzer. If satisfactory results cannot be obtained, then puddling and grinding
of the wire shall be done as specified in Section 9.5.1 above.
10.0 Undersize or Non-contact Samples
10.1 For samples that do not fully cover the measurement aperture, increase the testing time by increasing
the time in inverse proportion to the decrease in percentage of aperture covered. For example, if a rod
only covers of the aperture area, increase the measurement time by a factor of two; i.e., from 10
seconds to 20 seconds for chemistry.
10.1.1 Undersize samples should be placed in a radiation-safe, test stand (optional) or similar test
fixture.
NOTE: Do Not Test an undersized sample by laying it on another substrate material, such as a wood, plastic, or
metal surface (table, stand, stool, or benchtop), for erroneous measurements may occur due to the
interfering elemental signals coming from the substrate(s). It is highly recommended a suitable
radiation-safe Test Stand be acquired or fabricated for safely measuring undersized samples.
10.2 Non-contact sample measurements include those measured with an air gap of up to a maximum of in
(~6mm) from the measurement aperture. Such measurements require double exposure times; e.g.,
from 5 to 10 seconds for ID tests and from 10 to 20 seconds for chemistry tests.
10.3 Inside pipe measurements on pipes that are less than two feet in diameter will result in non-contact
measurements. Follow directions in Section 10.2 above.
11.0 Instrument Measurement Quality Control
11.1 Measure the certified 1 Cr - Mo alloy calibration check standard that is supplied with the analyzer At
Least once for every two hours of operation. If correct, continue testing. If incorrect, re-calibrate the
detector and re-take the previous two hours of measurements.
12.0 Certification Stamps
12.1 All stamping should be done with a specified/approved stamp.
12.2 All stamps purchased for special marking should be marked with a control number. A list indicating the
number and locations of these stamps is to be held by the PMI Coordinator.
12.3 Typically a stamp is assigned to each individual analyzer, and the stamp is carried in the respective
analyzers field carry bag.
12.4 Report any loss of a stamp immediately to the PMI Coordinator or designate.
12.5 Stamp shall include the letters PMIV
12.6 Stamp shall be of the low-stress, round-face type
13.0 When and How to Use an Approved Stamp
13.1 Materials and components should be tested prior to fabrication.
13.2 Items checked prior to fabrication and before the final as-installed condition will only be marked with an
approved paint pen as follows: PMIV.
13.3 Chemistry documentation may not be necessary for this phase of PMI inspection.
13.4 Number of components and type tested should be documented for the job.
13.5 Discrepancies are to be noted and handled as specified in an appropriate procedures document.
13.6 Shop made spool pieces may be stamped, providing their final field location in the piping system can be
tracked and documented. Final field installation must be confirmed by visual inspection of the stamp.
13.7 Usage of an approval stamp should be in accordance with an appropriate company procedures
document for PMI of existing equipment.
13.8 Usage of an approval stamp does not negate the need for any written documentation required.