Automatic Autoclave, Sterilizer


Operation Manual

Model : J-NAS103

Please read this Operation Manual carefully before use.


Operation Manual

Use of Operation Manual

z Please read through and understand this Operation Manual before operating the product.
After reading, always keep the manual nearby so that you may refer to it as needed.
When moving the product to another location, be sure to bring the manual as well.

z If you find any incorrectly arranged or missing pages in this manual, they will be replaced.
If the manual it gets lost or soiled, a new Operation Manual can be purchased.
In either case, please contact your JISICO agent, and provide the "Model No." given on cover.

z This manual has been prepared with the utmost care; however, if you have any questions, or note
any errors or omissions, please contact your JISICO distributor or agent.

Reproduction and reprinting of this operation manual, whole or partially, without our permission is prohibited.
Both unit specifications and manual contents are subject to change without notice.

Copyright 2004-2006 JISICO Co., Ltd.


Printed in Republic of Korea.

J-NAS103-M-E-09-001(REV.3) Page 2
Operation Manual

To SUPERVISOR in charge of this equipment.

z If the operator does not read the language used in this manual,
translate the manual into appropriate language.

z Help the operator in understanding this manual before use in


operation.

z Keep this manual near the tester for easy access by the operator.

J-NAS103-M-E-09-001(REV.3) Page 3
Operation Manual

SAFETY SYMBOLS

This operation manual and this product use the following safety symbols.
Please keep in mind the meaning of each of the symbols to ensure safe use of the product and to
prevent any damages, accidents, or problems resulting from negligence of these safety
symbols. (As using symbols may depend on the product, not all of the symbols may be used.)

SAFETY SYMBOLS
z Indicates the possibility of personal injury or death.
Never fail to follow the operating procedure.

z Do not proceed beyond a WARNING sign until the noted


conditions are fully met to satisfactory and understood.

z Indicates the existence of possibility of damage to the product


or to the operator in operation.

z Never fail to follow the operating procedure.

z Indicates the existence of possibility of damage to the product


or to the operator in operation when cautions are not observed
and followed properly.

z Indicates additional information such as general warning,


caution, risk of danger, and operating procedure.

z When this mark is indicated on the product, refer the relevant


section of the Operation Manual.

J-NAS103-M-E-09-001(REV.3) Page 4
Operation Manual

Safety Precautions

The following safety precautions must be observed and followed to avoid fire hazard, electrical
shock, accidents, and other failures. Keep them in mind and make sure that all the precautions
are observed and followed properly. JISICO Co., Ltd. assumes no liability nor any responsibility
against any damages, accidents, or problems resulting from negligence of the precautions.

Users
z This product must be used only by qualified personnel who understand the
contents of this operation manual.
z If it is handled by disqualified personnel, personal injury may result.
Be sure to handle it under supervision of qualified personnel.

Purpose of use
z If the product is to be used for purposes not described in this manual,
contact your JISICO agent in advance.

Input Power
z Use the product with the specified input power voltage.
z For applying power, use the AC power cable provided. The shape of the
plug differs according to the power voltage and areas. Use the cable
which is suitable for the line voltage used.

Fuse
z With products with a fuse holder on the exterior surface, the fuse can be
replaced with a new one.
z When replacing a fuse, use the one which has appropriate shape, ratings,
and specifications.

Cover
z There are parts inside the product which may cause physical hazards. Do
not remove the external cover. If the cover must be removed, contact your
JISICO agent in advance.

J-NAS103-M-E-09-001(REV.3) Page 5
Operation Manual

Installation
z When installing products be sure to observe Installation Precaution
described in this manual.
z To avoid electrical shock, connect the protective ground terminal to
electrical ground (safety ground).
z When applying power to the products from a switchboard, be sure work
is performed by a qualified and licensed electrician or is conducted
under the direction of such a person.
z Be sure to use the AC power cable provided.

Relocation
z Turn off the power switch and then disconnect all cables when
relocating the product.
z Use two or more persons when relocating the product which weights
more than 20 kg. The weight of the products can be found on the rear
panel of the product and/or in this operation manual.
z Use extra precautions such as using more people when relocating into
or out of present locations including inclines or steps. Also handle
carefully when relocating tall products as they can fall over easily.
z Be sure the operation manual be included when the product is relocate

Maintenance
z To avoid electrical shock, be absolutely sure to unplug the AC power
cable or stop applying power before performing maintenance or
checking.
z Do not remove the cover when performing maintenance or checking. If
the cover must be removed, contact your JISICO agent in advance.
z To maintain performance and safe operation of the product, it is
recommended that periodic maintenance, checking, cleaning, and
calibration be performed.

SERVICE
z Internal service is to be done by JISICO service engineers.
z If the product must be adjusted or repaired, contact your JISICO agent.

J-NAS103-M-E-09-001(REV.3) Page 6
Operation Manual

Contents

Use of Operation Manual 2


To SUPERVISOR in charge of this equipment 3
SAFETY SYMBOLS 4
Safety Precautions 5

1. Intended Use of the Equipment 8


2. Feature of Product 9
3. Appearance 10
4. Installation 11
5. Name and functions of controller 13
6. How to operate 16
7. Maintenance and service 22
8. Trouble shooting 25
9. After Service 26
10.Specification
10.1 Specification 27
10.2 Component List 27
10.3 Assembly Drawing 28
10.4 Circuit Diagram 29
11.Certification
11.1 ISO9001 Certificate 30
Warranty Certificate 31

J-NAS103-M-E-09-001(REV.3) Page 7
Operation Manual

1. Intended Use of the Equipment

Autoclaves and sterilizers are used for applications requiring constant pressure and
temperature for long periods of time. Common applications include the sterilization of
instruments and polymerization of rubbers, plastics, glass, ceramic, water and culture
medium.

Autoclaves and sterilizers in general may use a number of technologies to perform


equipment sterilization, including steam heating, dry heating, chemical sterilization, or
ultraviolet light, but majority of autoclaves function using some form of heat.

Steam autoclaves are far more common, using heated, vaporized water to kill pathogens,
and this equipment, J-NAS series Automatic Autoclaves use steam heating technology to
perform sterilization.

J-NAS series Automatic Autoclaves which are Laboratory autoclaves, are used in a variety of
thermal processing applications including general lab work, component and stability testing,
core hardening, drying glassware, and sterilizing.

J-NAS103-M-E-09-001(REV.3) Page 8
Operation Manual

2. Feature of Product

Open/close one touch type


Adopting four arm holder, one touch type switch opens and closes the lid conveniently.

Fully automated sterilization process.


One time setting enables fully automated sterilization process. Also exhaust pressure in
vessel after sterilizing for setting time. And finish-melody rings after pressure exhausting
procedure is completed.

JISICO-developed temperature controller


JISICO-developed temperature controller precisely regulates water-temperature in basket,
and this guarantees sterilization performance and safety of operation.

Internal air exhaust function


Internal air exhaust fountain is adjusted to exhaust air generated during the pre-heating
period.

Dual safety valves


In case of overpressure which may hardly occur, the initial safety valve activates and
decompresses the pressure inside the vessel of autoclave. When initial safety valve fails,
the secondary safety valve works to decompress pressure, and this dual safety valve
system guarantees safety of operation and operator.

Safety backup unit


The safety backup unit operates to cut off the power supply to heater when water
volume in the basket is below the minimum.

Installation of timer to enable automation


Timer is attached to set operation start time so as to automate turn-on and shutdown timing.

J-NAS103-M-E-09-001(REV.3) Page 9
Operation Manual

3. Appearance

Lid Handle
Lid Arm
Lid Cover

Pressure Gauge

Manual Pressure -
Valve

Exhaust bottle

Drain Port
for steam

Caster

[ FRONT VIEW ]

Controller

Main Switch
Solenoid Valve

Safety valves

Pressure exhaust
control valve

Drain Valve
for main vessel

[ UP SIDE VIEW ]
[ SIDE VIEW ]

J-NAS103-M-E-09-001(REV.3) Page 10
Operation Manual

4. Installation

4.1 Precautions for Installation

z Take precaution when moving the equipment to another location as the equipment is
heavy.
z This equipment is designed and manufactured to operate properly only in use of rated
voltage.
Check the rated voltage before installation.(Refer to Specification section of this
manual.)
z Use AC power cable that comes with (or installed to) the product and NEVER use an
AC power cable that comes with other equipment.
z Do not touch the Power Cord wet handed.
z Avoid locations where the equipment is exposed to direct sunshine.
z Do not place the equipment in a corrosive atmosphere.
z Do not use the equipment in a flammable atmosphere.
z Do not use the equipment where ventilation is poor.
z Do not locate the equipment in a dusty location.
z Do not use in an unstable place that are subject to vibrations.
z Avoid locations of high humidity.
z Avoid locations where the equipment is exposed to high temperature.
z Recommended temperature range : 20 to 30
z Use the equipment indoors only.
z Secure adequate space around the power plug

z This equipment is not designed nor manufactured explosion-proof, so NEVER insert any
explosive material nor flammable material inside the vessel.

4.2 How to install

z Leave intervals of above 20cm in front and back side of the equipment.
z Locate in a flat place.
z Avoid installing the equipment directly under a fire detector. When hot steam is
exhausted during the operation, it may activate the fire detector.

J-NAS103-M-E-09-001(REV.3) Page 11
Operation Manual

z Avoid installing the equipment in a location where there are electric devices or
equipments which may be affected by hot steam is exhausted during the operation.
z Never fail to ground the earth wire to prevent any electrical accidents or damages.
z Connect the power cord to an outlet with earthling point.
z In case of connecting to an outlet without earthling point, connect lead wire to the
earth after connecting grounding adapter.
z Grab the handle of the exhaust bottle, pull outward and take out the exhaust bottle
from the main body of the equipment.
z Fill water into the exhaust bottle about one third depth through filling port.
z Pour out water from the exhaust bottle when too much water is filled.
(Fill water about one third depth.)
z Put back the exhaust bottle where it was placed in the main body.

4.3 How to ground Earth wire

z Earth wire must be grounded to prevent any electrical accidents.


z Improper or no grounding may cause electrical shock
z Grounding terminal is included in the power plug.
z Therefore, there is no need to earth if plug in connected to a consent with grounding
terminal.
z If grounding terminal with consent is not available, install adapter to plug to ground
earth wire from green terminal or install earth wire to the grounding terminal of the
equipment.

Grounding earth wire to a water tap is not recommendable,


since water suppling pipes are usually connected by plastic tubes.
Don't ground earth wires to gas pipes.
Don't ground earth wires to telephone line.
Don't ground earth wires to lightning rod.

J-NAS103-M-E-09-001(REV.3) Page 12
Operation Manual

5. Name and functions of controller.

5.1 Name and functions of controller.

TEMPERATURE display window : displays temperature and parameter MODE


TIME display window : displays remaining time and data value of each MODE
READY display lamp
STERILIZE display lamp
EXHAUST display lamp
FINISH display lamp
AUTO- TUNING activation display lamp
HEATER output power display lamp
ALARM activation display lamp
MODE setting switch
Switch for moving decimal point of setting value.
Switch for increasing setting value of each point.
Switch for displaying AUTO-TUNING and time, temperature.
RUN/STOP switch

J-NAS103-M-E-09-001(REV.3) Page 13
Operation Manual

5.2 Characteristics of Lamp function.

RDY (ready) display lamp


After pressing RUN key switch, the Green lamp is turned on until the temperature reaches
TOP TEMPERATURE of temperature set-value, and afterwards the lamp turns into Red
colour.
STERILIZE display lamp
Green light is turned on as operation proceeds, and turns into Red light when finished.
EXHAUST display lamp
Green light is turned on as operation is completed for the setting time, and turns into red
when temperature fall below 100.
FINISH display lamp
Green light is turned on after temperature falls below 100 and alarm is activated, and
turns into Red light when operation is finished.
AUTO- TUNING activation display lamp
red light is turned on during AUTO-TURNING process.
OUT (=HEATER output power) display lamp
Lamp turns on only when Heater generates output power.
(After Exhaust display lamp has been activated, heater does not activate even though the
lamp is turn on.)
ALARM activation display lamp
Turns on only when there is an alarm output such as malfunction of the equipment.
Stop and other MODE => LAMP is turned off. (ALARM Relay off)

5.2 Characteristics of Key Switch function.

z Pressing RUN key once will START the equipment. To STOP, press the key once more.
z MODE KEY
(1) Setting DATA value of each MODE can be done during START (=in operation).
(2) Pressing once and releasing the key in normal display status changes into setting
MODE of set-value for temperature and time.
(3) Press the key for 6 seconds in normal display status, the MODE changes into
PARAMETER display MODE.
(The set-values are factory default values, so do NOT change them.)
(4) Pressing the key once in each MODE changes the PARAMETER. Press the key for more
than 2 seconds, it will end MODE and will return to normal status.

J-NAS103-M-E-09-001(REV.3) Page 14
Operation Manual

z KEY
(1) Pressing the key for 6 seconds in normal display status changes into FACTORY MODE.
(The set-values are factory default values, so do NOT change them.)
(2) Moves decimal point of setting value in each mode.
z KEY
Increases set-value of each mode.
z AT KEY
(1) Press and release the KEY in normal display mode, and temperature set-value or
remaining time of operation will be displayed each time you press the KEY.
(During AUTO-TUNING process, only temperature will be displayed.)
(2) Pressing the key for more than 4 seconds in normal status starts AUTO-TUNING, and
pressing the key for more than 4 seconds during AUTO-TUNING stops AUTO-TUNING
process.
(3) In each MODE, changes the DATA(= set-value) to 00.00

* If it is out of set-value range, the data display window will start to flash and the parameter
will not be changed when and MODE KEY is pressed, so it should be set within set-value
range. (CAUTION)

* When there is no input for 60 seconds in any MODE, it will return to normal status.

J-NAS103-M-E-09-001(REV.3) Page 15
Operation Manual

6. How to operate.

z Read this manual thoroughly before using J-NAS series Autoclaves, and keep in mind and
observe precautions and instructions given when handling for safe operation.
z If you have any questions, or note any errors or omissions, please contact your JISICO
distributor/agent.

6.1 How to operate.

Turn ' ON' the power switch on left panel of upper part of main body.
Release holder, turn lid handle counterclockwise completely and open the lid cover.
Pour water into vessel of autoclave until water levels just above the holes of bottom plate.
Put the sample to sterilize into the vessel of autoclave.

z Use distilled water or primary refined water if possible.


z Be careful that hands do not get caught when closing the lid cover.
z Before closing lid cover, confirm whether any foreign material is caught in lid gasket.
The foreign material caught may cause leakage of steam.

z When sterilizing long hollow objects, place them horizontally so that the steam may fully
pass through freely. Placing them vertically may result in poor sterilization.
z When sterilizing waste disposal bag, open the opening of bag as wide as possible so that
steam may be soaked into bag. When the opening of bag is closed during sterilizing,
complete sterilization will not be achieved and may result in poor sterilization.
z Do not pile sterilizing objects on top of one another, as steam may not penetrate into the
sterilizing objects and may result in poor sterilization.
z When sterilizing liquid such as chemicals and medium, consider the capacity of vessel or
container containing the objects to be sterilized, and fill reasonable amount.

J-NAS103-M-E-09-001(REV.3) Page 16
Operation Manual

For example, as of Erlenmeyer, fill about 3/4 depth and as of test tube, fill about 1/2 depth.
If too much amount is filled, it will result in overflow as temperature rises.
z When melting Agar medium, fill no more than 2 liters per every vessel or container.
If more than 2 liters are filled, it will not melt completely.

Lower the lid cover by turning the lid handle clockwise by using both hand until lid gasket
touches the surface. Turn the lid handle until it does not turn any more to prevent leakage
of steam during sterilization process.
Check and make sure that there is water about half depth in the exhaust bottle before
starting operation.
Automatic Operation programmed as default, it is programmed to reach the pressure of
2
1.2Kg/cm when temperature reaches 121, and the timer function activates.
So set the temp. value at 121. (Temp. set-value can be adjusted by 1(one) degree.
Set up the time for sterilization.
For details, refer to Section 6.3 Setting Temp. and Time MODE of this MANUAL .
This Autoclave does NOT run if there is no water in the vessel. (Alarm display lamp ON.)
Press RUN/STOP key once to start the operation.
To Exhaust the cold air which had been trapped inside the vessel, the equipment exhaust
steam for a fixed period of time set as factory-default value. After exhaustion of steam has
been completed, the temperature rises to temperature set-value.
z Cold Steam Exhaustion Time for Autoclaves :
J-NAS3 = 150 sec, J-NAS4 = 180 sec, J-NAS5 =200 sec, J-NAS6 = 250 sec
2
Timer starts to function when the pressure inside the vessel reaches 1.2Kg/cm .
After time or period set as time set-value has passed, EXHAUST display lamp turns on and
it wills tart to exhaust steam.
To exhaust steam immediately in need or in case of emergency, press RUN/STOP key to
halt the equipment.
After the steam has been exhausted completely, FINISH display lamp turns on and melody
sound will start for 30seconds.
To repeat or continue the test under the same conditions, put sample to sterilize in the
vessel and press RUN/STOP key, and it will start to perform operation at same condition
(same temperature and time) as previous operation.
J-NAS series Autoclaves are set to operate at factory default set-value of pressure of
2
1.2 Kg/cm at 121 only.
To operate at a different temperature (meaning different pressure), you have to adjust
set-value to other than factory default set-value.
For details, refer to Section 6.3 Setting Temp. and Time MODE of this MANUAL .

J-NAS103-M-E-09-001(REV.3) Page 17
Operation Manual

When the sterilization is completed, the equipment will exhaust steam completely and the
temperature drops, Finish display lamp turns on, and finish melody sound will start to ring.
Then turn OFF main power switch.

z Do not take out exhaust bottle nor drain water in the vessel of autoclave until the pressure
gauge installed on the controller indicates 0 Kg/cm2. Drained hot water and hot steam may
cause severe damage such as BURN or INJURY to the OPERATOR.
z Always wear heat-proof gloves when taking sterilized samples from the vessel of autoclave
to protect your hands from burn or injury.

6.2 Operation status procedure.

Before RUN (= start )


- ALARM RELAY is OFF, START RELAY is ON, all status display lamps are OFF.
- Time set-value will be displayed on the TIME display window
When RUNNING (= in operation)
- STERILIZE display lamp is turned ON
- Heater output power controlled by P.I.D control.
- START RELAY is OFF
(when reaching the START RELAY ON TEMPERATURE VALUE (=set-value),
START RELAY will be turned ON once for a period of time set as time set-value)
- ALARM RELAY is OFF when there are no ALARM generated.
ALARM will turn ON when ALARM is generated on. (ALARM display lamp will turn ON)
When present temperature reaches top temperature (=relative value)
- STERILIZE display lamp is turned ON (Green light turned into Red light)
Proceeds for time set as set-value, and displays remaining time.
(When 00.00 is set as time set-value, --.-- will be displayed as time at TIME display
window, and will proceed to operate permanently.
- HEATER output power and RELAY motion are same as those of When RUNNING (= in
operation) described above.
When operation for set time has been completed
- 00.00 is displayed on TIME display window (operation for set time has been completed)
- STERILIZE display lamp turns ON(Red light),
EXHAUST display lamp turns ON(Green light).

J-NAS103-M-E-09-001(REV.3) Page 18
Operation Manual

- HEATER output power is OFF, and START RELAY is ON.


- ALARM RELAY is OFF when there are no ALARM generated.
ALARM will turn ON when ALARM is generated on. (ALARM display lamp will turn ON)
- Operation will proceed until temperature reaches 100.
When present temperature is below 100
- STERILIZE display lamp is OFF (Red light),
EXHAUST display lamp is ON (Red light),
FINISH display lamp is ON (Green light)
- MELODY sound will ring for the period of time configured through the speaker.
- HEATER output power is OFF, and START RELAY is ON, ALARM RELAY is OFF
When MELODY proceeding time has been completed
- STERILIZE display lamp is OFF (Red light),
EXHAUST display lamp is ON (Red light),
FINISH display lamp is ON (Red light)
- HEATER output power and RELAY motion are same as those of When present
temperature is below 100 described above.

z In case power supply is cut off during operation


- If power supply is cut off while Autoclave is in operation, operation in progress is cancelled.
- When power returns, Autoclave is returned back to READY status, you should start operation
from the beginning again.

* AUTO -TUNING

- Press AT key for more than 4 seconds at normal status, AT display lamp will start to flicker
and AUTO-TUNING process will start. When AUTO-TUNING is completed, AT LAMP will be
turned OFF.

- If you wish to STOP the AUTO-TUNING process in progress by force, press AT key for more
than 4 seconds and AUTO-TUNING process will be stopped.

- This AUTO-TUNING can be done only when the autoclave is running and in process of
When RUNNING (= in operation) or When present temperature reaches top temperature
described in Section 6.2 Operation status procedure of this MANUAL.

J-NAS103-M-E-09-001(REV.3) Page 19
Operation Manual

- The OPERATING TIME when autoclave is running is NOT counted when the equipment is in
the middle of AUTO-TUNING process and will be stopped. And when AUTO-TUNING is
completed the time will increase continuously.

z Pay special attention to while operating this equipment.


Only the person in charge who is fully aware of Precautions, Warnings, Operation Methods
described in this MANUAL should have access to this equipment.
The person in charge should perform checkups on this equipment often and regularly.

z When operation is done, NEVER fail to exhaust the pressure (= hot steam captured inside
the vessel of Autoclave) through the pressure discharger completely until the pressure gauge
installed on the controller indicates 0 Kg/cm2, before opening the COVER, as it may result in
an EXPLOSION caused by the pressure.

z In case of over-pressure caused by mal-function or trouble of safety valve in use during


operation, TURN OFF the Main Switch immediately and exhaust the pressure (= hot steam
captured inside the vessel of Autoclave) through the pressure discharger completely until the
pressure gauge installed on the controller indicates 0 Kg/cm2 by opening Manual Pressure
Valve installed on the front side panel of the equipment.
Then contact your JISICO distributor or agent for the repair.

z Heater should NEVER be exposed in air and should ALWAYS be under water during
operation. If heater is exposed to air, it may lead to a SERIOUS DAMAGE or BREAK-DOWN
of the equipment.
So please perform often and regular checkup before use, and supply enough water as
stated in this manual.

J-NAS103-M-E-09-001(REV.3) Page 20
Operation Manual

6.3 Setting Temp. and Time MODE.

Normal Status

* Press MODE Key


1) Temperature setting range
tEmP - PT TYPE : 0 ~ 150
(Temperature Setting) 2) Time Setting Range
- Day, Time MODE (TIME SCALE : d.h)
* Press MODE Key
: 0~99 days, 23 hours
* Adjust with Key, Key - Hour, Minute MODE (TIME SCALE : h.m)
: 0~99 hours, 59 minutes
tImE - Minute, Second MODE (TIME SCALE : m.s)
(Time Setting) : 0~99 minutes, 59 seconds

MODE Key * If TIME set-value is 00.00 , it will operate


continuously.
*Press MODE Key for more than
2 seconds, it will return to
NORMAL Status

J-NAS103-M-E-09-001(REV.3) Page 21
Operation Manual

7. Maintenance and service.

z NEVER carry out MAINTENANCE or SERVICING before the temperature of the main body of
the equipment has been cooled completely, or may result in severe damage or injury.

7.1 Taking out exhaust bottle

Water gets collected and gathered into exhaust bottle as you continue to use the
equipment. If exhaust bottle gets filled up with water, pour water out to half depth of the
exhaust bottle.

z NEVER try to touch, grab, take out the exhaust bottle until the water inside the exhaust
bottle has been cooled down enough to a temperature safe enough to touch.

7.2 Draining water from the vessel.

To utilize the equipment without trouble, failure, or damage and prolong the lifespan, NEVER
fail to completely drain the water inside the vessel of the autoclave, to prevent water pipes
from being clogged by foreign materials such as dust.

Make sure that the exhaust bottle has cooled to safe temperature, and carry out water
draining process as described herein ;

Open the lid cover.


Connect one end of drain hose to the nipple of drain-valve installed on the lower left
panel of the main body of the equipment.
Dip the other end of drain-valve into the exhaust bottle.
Turn drain-valve knob open installed on the lower left panel of the main body of the
equipment.
Drain the water until the water is completely drained,
Close drain-valve by turning the drain-valve knob.

J-NAS103-M-E-09-001(REV.3) Page 22
Operation Manual

z NEVER try to touch, grab, take out the exhaust bottle or try to drain out the water from
the autoclave until the temperature has been cooled down enough to a safe
temperature and the pressure has dropped completely, or drained hot water and hot
steam may cause severe damage such as BURN or INJURY to the OPERATOR.

7.3 In case of cleaning the inside of vessel.

z As temperature sensor is installed to heater, be careful not to damage the temperature


sensor.
z In order to prolong lifespan of the equipment or to decrease the chances of contamination
to the equipment, ALWAYS clean the equipment after use.

- Remove power plug from power outlet.


- Take out bottom plate to clean heater.
- Separate lid and gasket (if necessary).
- Open the drain-valve, drain the water in the vessel completely, and then clean the vessel,
lid and heater by using neutral washer and refined water (or water for injection) with soft
brush.
- Rinse fully by spraying with refined water or water for injection lest cleaner should be
remained.
-Clear fully with neutral cleaner and refined water and rinse with refined water (or water for
injection) lest cleaner should be remained. (In case of separating gasket)
-Take action lest water in the vessel should be remained and close drain valve after drying
water.
- In case of separating gasket, initial step is to assemble gasket and keep it with closing lid.
- Confirm clean condition of equipment prior to use it and when equipment is polluted , use
equipment after it is cleaned .

7.4 In case of cleaning the exterior of the equipment.

- Remove power plug from power outlet.


- Clean with sponge or soft clothes getting wet by neutral cleaner.

J-NAS103-M-E-09-001(REV.3) Page 23
Operation Manual

- Clean with dry clothes.


- Do not use organic solvent (Especially, at parts of control panel and door handle)

z Do not use flammable or volatile chemicals such as benzene. alcohol to clean the
equipment.
z This equipment is not designed nor manufactured Explosion-proof, so never place highly
flammable material or explosive material inside the chamber, which may cause an explosion
or a fire.

7.4 Cleaning electric parts.

- Clean that with dry clothes.

7.5 When the equipment is not used for long time.

- Remove power plug from power outlet.


- Clean with soft clothes.
- Pack the equipment in a appropriate way and store in a safe place for storage.

J-NAS103-M-E-09-001(REV.3) Page 24
Operation Manual

8. Trouble shooting.

8.1 When the machine's not operating with Power switch on.
Is the machine's plug properly plugged in the power consent?
Are rating and capacity of the power consent proper?
Is the power enough?
Is the breaker on the side of outer part open?

8.2 In case of Heater trouble


Check the heater for possible damage.

8.3 When temperature does not rise


Check the temperature set-value.
Check the heater for possible damage.

8.4 In case of steam leakage from the piping temperature and setting temperature of the
Fix the fittings and piping tightly.
Is the machine inclined?
Install the machine horizontally all the time.

8.5 In case of water leakage


Check and close the water-drain valve installed on the lower left side of the
equipment

J-NAS103-M-E-09-001(REV.3) Page 25
Operation Manual

9. After Service.

9.1 Free repair

Faults which is responsible for manufacture in normal condition can be repaired with free for
1year from purchase date (only, pressure vessel is zero year), and it is desirous to check
following items when requesting A/S.
. Part and condition generated fault
( It is necessary to explain in detail within limit of possibility )
. Model name
. Serial Number
. Purchase date (year, month, date)

9.2 Onerous repair

In following case, it is required to repair with compensation in spite of guarantee period.


- Fault by user mistake, unsuitable repair or modification
- Fault by negligence for carrying and handling during use
- Fault due to natural disasters such as fire, flood disaster and abnormal voltage
- Fault by using against direction for u

J-NAS103-M-E-09-001(REV.3) Page 26
Operation Manual

10. Specification

10.1 Technical Specification

Model J-NAS103
Overall size(WxDxH) mm 800 x 600 x 880/1165
Inner size(xH) mm 450 x 650
Capacity 103
Method sterilization High pressure system
Using temperature 121
Using pressure kg/ 1.15
Temperature regulator SSR type
Temp controller P.I.D. Controller
Internal steam exhaust About 30 minutes
Heater(Sheath) W 4,500
Timer 99hr 59min or infinity
Pressure gauge kg/ Mechanical type, 0~2
Basket 2ea
Power source AC 110V, 50/60Hz, 1Phase

10.2 Component List

No Description Specification Model No. Q'ty Manufacturer


1 Temp. Controller P.I.D. Controller JISICO-PID-AT 1 JISICO CO., LTD.
2 Temp. Sensor PT100, PT 1/4''x150(L) JISICO-A/T-PT 1 JISICO CO., LTD.
UNION ELECOM
3 SSR Unit AC110V 50A Load WYG1C-50Z4 1
CO., LTD.
4 Exhaust Fan AC110V SF23080AT 1 SUNON CORPORATION
5 Magnetic Contact AC110V, GMC-48 1 LS Industrial Systems
6 E.L.B 50A EBE52Fb 1 LS Industrial Systems
7 Relay AC110V HR707N-2P 2 Han Kuk Relay
2
8 Solenoid Valve 1/4'', 0.3~0.7 Kg/cm VPS2135 1 SMC Corporation
9 Needle V/V SUS304, 1/4'' FNF-80A 1
10 Safety V/V 1/8'', 3/8'' 2
11 Ball V/V 1/4'' 1
12 Train V/V 3/8'' 1
2
13 Pressure Gauge 2 Kg/cm 308004 1 Nisshin
14 Heater SUS304, AC110V 1
15 Trust Bearing I.D 20 2

J-NAS103-M-E-09-001(REV.3) Page 27
Operation Manual

10.3 Circuit Diagram

J-NAS103-M-E-09-001(REV.3) Page 28
Operation Manual

10.4 Assembly Drawing

J-NAS103-M-E-09-001(REV.3) Page 29
Operation Manual

11. Certification

11.1 ISO9001 certificate

J-NAS103-M-E-09-001(REV.3) Page 30
Operation Manual

Warranty Certification.

Thank you for purchasing a JISICO product. Please take a moment to fill in the following and return it
to JISICO CO., LTD. for product registration.

z JISICO products are warranted against defect in material and workmanship for a period of one(1) year
after delivery to the first purchaser for use.
z This warranty does not apply to the damage resulting from misuse, alteration, transportation, abuse,
accidents, tampering, servicing attempted or performed by unauthorized service agency.
z Before use, read the manual thoroughly. JISICO can not be held responsible for the malfunctioning
resulting from the use of the product other than specified in operation manual.
z Call or fax Customer Service for a Return Material Authorization (RMA) number before returning the
product or any specific material.
z Put the RMA number on the shipping label, and a written detailed description of the problem.

Product Name
Model No. Serial No.
Delivery Date / / ( dd/mm/yy) Warranty One(1) year
JISICO Company
Agent/ Distributor Phone No.
Company Contact
Custom Address
Phone No. Fax. No.

JISICO only warrants the safety and proper operation of product under the condition that :
- The product is properly and in accordance with the instructions stated in the operation manual.
- No modifications are made.
- Only original spare parts are used. The use of other parts carries unknown risks and will void unit's
warranty.
- The back of the unit has not been opened.