EMG Measuring System

Measuring System to Record Muscle Activity

Compatible with the


motion analysis device
CMS20S, CMS-HSL
and CMS-HS

Accurate and reliable


registration of the
EMG signals

Active electronic circuit


in the electrode cables

High signal-to-noise ratio,


no cable artifacts
The zebris EMG measuring system By using the latest integrated circuits,
registers the action potential by the measuring unit for eight channels, Precise specification of the
means of skin surface electrodes. together with the complete electro- electrical characteristics
This is done in a similar manner to nics for patient safety, have been in-
that of an ECG, where the action corporated in an adaptor with the di- Light and small basic unit
potential of the cardiac muscle is mensions 70x44(33)x115 mm. This for 8 or 4 channels
registered. can be worn comfortably on the body.

Through connection to the zebris The EMG differential electrode cables


motion analysis measuring systems, contain an active electronic circuit,
CS20S, CMS-HSL or CMS-HS, the which ensures a high signal-to-noise
muscle action potentials can be ratio. As a consequence, no cable
directly superimposed on the motion artifacts can occur.
data. The EMG activity of small and large
The measuring results of the gait muscles can be precisely and reliably
analysis, the analysis of the move- detected.
ments of the hand and arm system, The characteristic of every EMG
or the spinal column, are displayed amplifier is precisely specified.
in real time on a PC and can be
subsequently analysed.

For further information


please contact

zebris Medical GmbH

Max-Eyth-Weg 42
D-88316 Isny i. Allgu
Germany

Tel.: +49 7562 / 9726-0


Fax: +49 7562 / 9726-50
E-mail: zebris@zebris.de
Internet: www.zebris.de

zebris 09/2005
zebris
Surface EMG System: Technical Data

Options for CMS20 Options for CMS-HSL


devices and CMS-HS devices

Analog-digital converter
(has to be installed into the device; necessary for the operation of EMG 4 or EMG 8 basic units)

number of channels per 4 single channels or versions with 8 and 16


AD-converter 2 differential channels differential available
max. number of channels per channels
device 4 (1x4) 16 (CMS-HSL System)
32 (CMS-HS System)
measuring rate max. 4 kHz max. 64 kHz
resolution 8 / 12 bit (selectable) 12 bit
input voltage range K5 V (1 digit=2.4 mV) K2.047 V (1 digit=1 mV)
Input impedance (per entrance) 220 kS against GND 50 kS against GND

EMG 4 EMG 8
Basic unit
(for the connection to AD-converters)

mechanical data - dimensions: 70 mm 70 mm


width 33 mm 44 mm
height 115 mm 115 mm
depth 160 g 210 g
weight 4,5 m 4,5 m
cable length (standard)

Active differential electrode cable


(versions with or without a neutral-electrode available)

electrical data: K5 V bis K15 V


supply voltage 10 E + 12 S
input impedance 110 dB
CMRR 0.28 :V pp
noise referred to input 1000, 2500, 5000
voltage gain
7.5 to 1000 Hz
bandwidth 1 V < supply voltage
output voltage

mechanical data - dimensions: 23 mm


width 9 mm
height 30 mm
depth 1,45 m
cable length

technical modifications reserved