You are on page 1of 9

ANALYTICAL BIOCHEMISTRY 132, 353-361 (1983)

Correction for Light Absorption in Fluorescence Studies


of Protein-Ligand Interactions

BERRY BIRDSALL,* RODNEY W. KING,* MIRIAM R. WHEELER,?


CHARLES A. LEWIS, JR.,? SCOTT R. GOODE,? R. BRUCE DUNLAP,?
AND GORDON C. K. ROBERTS*
*Division of Molecular Pharmacology, National Institute for Medical Research, Mill Hill, London, NW7 IAA,
United Kingdom, and TDepartment of Chemistry, University of South Carolina, Columbia, South Carolina 29208

Received January 11, 1983

It is shown that absorption of the excitation light can lead to substantial systematic errors in
fluorescence measurements of equilibrium constants for formation of protein-ligand complexes.
The assumptions about the optical arrangement of the fluorescence spectrometer involved in
the calculation of tbe correction of this absorption am discussed. A general semiempirical correction
procedure which can be used for (calculated) absorbance values as high as 5 is described. The
importance of choosing the excitation wavelength so as to minimize the necessity for these
corrections is emphasized.
KEY WORDS: fluorescence; protein-ligand complexes; inner filter effect; equilibrium constants;
absorption correction.

Fluorescence spectroscopy has proven to be have described the use of primary absorption
a widely useful technique foi the study of pro- effect corrections in biochemical systems (1 l-
tein-ligand interactions ( l-2). In particular, 17), often using approximate corrections valid
it often affords a simple method for the mea- only under specific conditions, such as rela-
surement of equilibrium constants for for- tively low absorbance.
mation of protein-ligand complexes. A num- The purpose of this paper is to describe a
ber of instrumental and photophysical vari- simple, general procedure for correcting for
ables affect the accuracy of fluorescence primary absorbance effects in fluorescence
measurements (2-6), of which the most im- measurements of equilibrium constants for
portant in the context of equilibrium constant protein-ligand complex formation.
determinations is the absorption of the exci-
tation beam, the primary absorption or inner- THEORY
filter effect ((7); see also (8)). If a significant
part of the total absorption is due to the spe- We shall consider the case where the only
cies-usually the ligand-whose concentra- instrumental correction required is that for
tion is varied in the course of the experiment, the primary absorption. We therefore assume
the primary absorption effect will distort the that (a) any absorption of emitted fluorescence
concentration dependence of the fluorescence, is negligible and (b) effects of light scattering,
leading to an incorrect value for the equilib- refractive index changes, and emission an-
rium constant. isotropy are negligible. It is usually possible
The form of the correction factor required to ensure the validity of the first assumption
to compensate for this absorption effect has by a suitable choice of emission wavelength;
been discussed in detail by Parker (3), Holland if necessary a correction factor analogous to
et al. (9), and Lloyd ( 10). A number of authors that for primary absorption can be applied

353 0003-2697183 $3.00


Copvnght 63 1983 by Academlr Press. Inc
All nght5 of reprcduct~on m any form rewwd
354 BIRDSALL ET AL.

(10). The emission from fluorophores in or represents the efficiency with which this flu-
bound to a macromolecule is likely to be an- orescence is detected. If we define the ab-
isotropic, and this can lead to substantial errors sorption correction factor, C, by Fobs = F,,,
in quantum yield and in the shape of the X C, then
emission spectrum (( 18) and refs. therein). W2
c=ws4w)E(
WdW
However, for the present limited purposes of
estimating the equilibrium constant for for-
mation of a protein-ligand complex, these ef- PI
(W2 - WI) *
fects can be largely ignored, since the con-
centration dependence of the observed inten- From the Beer-Lambert law,
sity will only be slightly affected.
The right-angle geometry adopted in the Z(w) = lJO-AW [31
majority of commercial fluorescence spec-
where A is the absorbance of the sample at
trometers is shown in Fig. 1. The excitation
the excitation wavelength for the path length
beam is directed along the x axis, and the
used. (Note that this absorbance should be
fluorescence detected is that emitted in the y
measured using the same bandwidth and
direction from within some observation win-
spectral distribution as used for the excitation
dow between the points Xi and X2. The dis-
beam in the fluorescence measurements.)
tances of those points from the front face of
The detection efficiency function, E(W), is
the cell, as fractions of the total path length,
much less easily defined. It will depend on the
are denoted W, and IV,.
collimation of the excitation beam, on the ob-
The corrected fluorescence, F,, , is defined servation window, and on the uniformity of
as the observed fluorescence, Fobs, divided by sensitivity across the detector surface. E(W)
the average fluorescence yield across the will thus depend critically on the sample-
observation window Xi to X2.
compartment optics, and will vary consider-
WZ ably from one spectrometer to another, as well
4 w)E( w)dw as varying with the width of the emission slit(s).
s
F corn= Fobs w In view of this uncertainty, three distinct ap-
1 (W, - W,) * [13 proaches to the evaluation of Eq. [2] have
The term Z( IV) represents the intensity of the been used:
excitation beam at a point W on the x axis (i) Assume that the observed fluorescence
in the observation window, the intensity of originates from a single point in the sample
fluorescence emitted from point W being cell. We then simply need to correct for the
taken to be proportional to Z(w). E(w) then intensity of the excitation beam at this point,
giving
c= l(yQ [41
Y
where W. is the fraction of the total path length
Excitation the excitation beam must travel to reach the
.-.I _..._.__..; ___.. ..-X point from which the fluorescence is observed.
Clearly the assumption of point-source fluo-
I
21 j xc rescence can never be strictly justified, par-
ticularly since, for maximum sensitivity, rel-
Emission atively large slit-widths are commonly used
1 on the emission side.
(ii) Assume that the efficiency of detection
FIG. 1. The experimental arrangement for the mea- of the fluorescence is uniform across the ob-
surement of fluorescenceconsidered in the text. servation window (i.e., E is independent of
PROTEIN-LIGAND FLUORESCENCE CORRECTIONS 355

IV). Equation [2] then becomes (4,9) curve fitting. The parameter a is related to,
but not numerically equal to, tL, the molar
c = lo-AW - lo-4w2
151 absorption coefficient of the ligand at the ex-
2.303,4(W, - IV,) citation wavelength. With optimum values of
and the correction factor can be calculated a and d, Eq. [7] gives values of C within 0.1%
from a knowledge of the absorbance and of of those given by Eq. [5] up to an absorbance
the position and size of the observation win- of 2.0. In addition, Eq. [7] provides a satis-
dow. If the dominant contribution to the ab- factory empirical description of the variation
sorbance comes from the ligand, A can be of C with LT under conditions where neither
replaced by eLLT, where LT is the total ligand the point-source nor the uniform sensitivity
concentration and tL its molar absorption assumptions are likely to be valid (see below).
coefficient at the excitation wavelength. For low absorbance, setting d = 0, Eq. [7]
Uniform efficiency of detection requires, becomes identical to Eq. [6], with aL,r
among other things, a uniform observation = 2.3034.
angle across the window; again, this can never In fact, for sufficiently low absorbance, all
three approaches give closely similar values
be strictly true. Holland et al. (9) have de-
scribed an optical arrangement which satisfies for C and provide equivalent procedures for
the uniform efficiency criterion to within l%, applying the primary absorption correction;
but the optics of most commercial instruments for A < 0.2 Eqs. [4], [5], [6], and [7] give
are quite different. For example, the Perkin- values for C which agree within 1%. At higher
absorbance, on the other hand, the applica-
Elmer instrument used in the present work
bility of the assumptions about E( IV) to the
has a lens adjacent to the sample cell on the
emission side. particular spectrometer being used will deter-
For sufficiently small values of (IV,- IV,), mine whether Eq. [4], [5], or [6] can be used.
Eq. [5] gives values of C identical to those
MATERIALS AND METHODS
given by Eq. [4]. For low absorbance, Eq. [5]
can be simplified to Materials. Dihydrofolate reductase was pu-
c = 1 - 10-A rified from Lactobacillus casei MTX/R ( 19).
2.3034 .
[61 The enzyme concentration was determined
by absorption measurements and by fluori-
This is equivalent to Eq. (51 with IV, = 0, IV, metric titration with methotrexate (19). Nic-
= 1; for low enough absorbance, the absorp- otinamide 1,N6-ethenoadenine dinucleotide
tion outside the window IV,- IV, becomes in- phosphate (eNADP+), folate, tryptophan, and
significant. For A < 0.3, Eq. [6] gives values p-aminobenzoyl-L-glutamate were from
of C within 2% of those given by Eq. [5]. Sigma.
(iii) Determine the form of E( W) empiri- Fluorescence spectroscopy. Measurements
cally by curve fitting. This approach was used were carried out using a Perkin-Elmer MPF-
by Ehrenberg et al. ( 1 l), who described E( IV) 44A fluorescence spectrometer. Experiments
by a polynomial. Rather than determine E( IV) were conducted at 25C using l-cm-path
for each instrument setting, we have applied length quartz cells; ligands were added as
the curve fitting directly to the determination microliter volumes of concentrated stock so-
of C, as described under Results and Discus- lutions. Ligand binding to dihydrofolate re-
sion. For this purpose we use the equation ductase was studied in 15 mM Bistris [2-[bis(2-
hydroxyethyl)amino] -2-(hydroxymethyl)- 1,3-
c = (e-aLTd - e-aL~)
[71
ALT(l - d) Abbreviations used: cNADP+, nicotinamide I ,N-
ethenoadenine dinucleotide phosphate: Bistris. 2-[bis-(2-
where a and d are parameters determined by hydroxyethyl)aminol-2-(hydroxymethyl)-1,3-propanediol.
356 BIRDSALL ET AL.

propanediol], pH 6.0, containing 0.5 M KCl, Marquardt (22). Initial values of K were es-
unless otherwise noted. timated from l/K = LSo - ET/2, where LSo
Data analysis. This was carried out in two is the ligand concentration at which 50% of
stages. The raw data were first corrected for the fluorescence change has occurred, and ini-
dilution, by calculating the true enzyme and tial values of F, either from a double-recip-
ligand concentrations at each data point, and rocal plot ( l/F vs 1/L-J or simply from the
for the primary absorption effect, as discussed measured fluorescence at the highest value of
below. The second stage was the estimation LT used. (With a robust minimization routine
of the equilibrium constant for complex for- and a sufficient data set, very accurate initial
mation from the corrected fluorescence data values are not required.) The standard errors
(20,21). For a simple binding process quoted for K are those estimated by the
regression program; as discussed earlier (20),
E+L=EL 181 these are close to those estimated from rep-
K = [EL]/[E][L]. [91 licate determinations.
For measurements of protein fluorescence,
RESULTS AND DISCUSSION
WI F0 - F
-= PO1 Choice of Experimental Conditions to
ET Fo - Fm, Minimize Inner-Filter Efects
where [EL] is the concentration of complex In an experiment designed to measure an
at a ligand concentration giving fluorescence equilibrium constant, the enzyme and ligand
intensity F, ET is the total enzyme concen- concentrations to be used are governed by the
tration, and F. and F, denote the fluorescence magnitude of the equilibrium constant. If a
intensity at zero and saturating ligand con- good estimate of K is to be obtained, one
centrations, respectively. Similarly, when li- must ensure ET d 3/K, and ligand concen-
gand fluorescence is measured trations up to at least 1O/K must be employed.
WI F - FL The experimental variables which can be ma-
-= [Ill nipulated to minimize the primary absorption
ET Fm-FL
effect are thus limited to the path length and
where FL is the fluorescence of an equal con- the excitation wavelength. Short path length
centration of ligand in the absence of protein. cells can be valuable (25), although some care
From the mass-action equation [Eq. 91, is necessary to ensure accurate and reproduc-
ible positioning of the cell.
[EL] = Yz[l/K + ET + LT However, the most convenient way to min-
- ({l/K + ET + LT}* - 4ETLT)l*] [12] imize primary absorption effects is by an ap-
propriate choice of excitation wavelength
and by combining Eq. [ 121 with Eq. [lo] or (25,26). If the absorbing species is the fluo-
[ 111, an expression for the corrected fluores- rophore itself, the use of an excitation wave-
cence as a function of LT can be obtained.* length on the red edge of the longest wave-
The two unknowns, K and F,, were deter- length absorption band will often be more sat-
mined by nonlinear regression, using lo-30 isfactory than excitation at the absorption
values of F and LT. The regression analysis maximum. This is illustrated in Fig. 2, which
employed modifications of the programs of shows the concentration dependence of the
Batchelor (23), based on the algorithm of fluorescence of N-(p-aminobenzoyl)-L-gluta-
mate for several excitation wavelengths. The
*If there is more than one binding site per protein
measured fluorescence is shown directly in Fig.
molecule, not only will Eq. [ 121 require modification but 2A, while in Fig. 2B it has been normalized
Eqs. [IO] and [ 1 I] may also (24). by the molar absorption coefficient at the ex-
PROTEIN-LIGAND FLUORESCENCE CORRECT IONS 357

However, by exciting further from the ab-


sorption maximum, at 335 nm, a more nearly
linear concentration dependence is observed
(Fig. 2) and the binding constant can be mea-
sured by working at this wavelength (20).
When using this approach, precautions must
be taken to ensure that the fluorescence ob-
served is still that of the compound of interest,
and not that of a fluorescent impurity with a
longer wavelength absorption.

Correction for Primary Absorption Eflects


Using the semiempirical Eq. [7] for the ab-

*'
D 330
sorption correction term C, we have the gen-
eral expression

.'

&
325
where F&,k represents fluorescence from
320 sources other than the compound of interest.
k
0 2 4 6 8 10 The use of Eq. [ 131 can be illustrated by the
CONCENTRATION mM correction of the data in Fig. 2. Here F&,,,k
FIG. 2. The fluorescence ofpaminobenzoyl-L@namate = 0 and, since the primary absorption effect
as a function of its concentration, using various excitation comes from the fluorophore itself, F,, = kLT,
wavelengths. (A) Fluorescence (arbitrary units) at exci- where k is a coefficient of proportionality.3
tation wavelengths of 320 (0), 325 (Cl), 330 (V), and 335 Values of k, a, and d were determined for each
nm (0). The lines are the best fits of Eq. [ 131. (B) The
solid lines represent the data from (A), expressed as flu- excitation wavelength by nonlinear regression
orescence divided by e, the extinction coefficient at the analysis of the data in Fig. 2. Having deter-
excitation wavelength. In the absence of the inner-filter mined a set of a and d values in this way, the
effect, these curves would follow the dashed line. The correction factor for any concentration and
points (symbols as in (A)) show the experimental data any excitation wavelength can be calculated.
after correction for the inner-filter effect as described in
Provided that Eq. [7] is able to describe the
the text.
primary-absorption effect accurately, the cor-
rected data points should fall on the straight
citation wavelength; in the absence of inner- line in Fig. 2B. It is clear that this does in fact
filter effects, the data for all excitation wave- happen. The corrected data point which lies
lengths would fall on the straight (dashed) line furthest from the theoretical line is that for
in Fig. 2B. The concentration range shown, 19 mM N-( p-aminobenzoyl)-L-glutamate ex-
0.1-10 mM, is that relevant to measurements cited at 320 nm; here the calculated absor-
of the binding of this compound to dihydro- bance was 5.78, and yet correction by Eq. [7]
folate reductase (K = 830 M- (20)). On ex- gives a value within 3.2% of theory.
citation at the absorption maximum, 274 nm Equation [7] is thus able to correct accu-
(eZT4= 1.45 X lo4 M- cm-), essentially no rately even for very large primary absorption
fluorescence is detectable at concentrations
This assumption of linearity of lluorescence with fluo-
> 0.5 mM (where A > 7.25). Even with ex- rophore concentration may not be justified in some sys-
citation at 320 nm, the concentration depen- tems, if there is self-association of the lluorophore in the
dence of fluorescence is still grossly nonlinear. relevant concentration range.
358 BIRDSALL ET AL.

effects. However, it is clear from Fig. 2 that 6


it is experimentally much more satisfactory
to choose an excitation wavelength at which
a much less drastic correction is required (here
335 nm).
The practical application of this correction
procedure to measurements of binding con-
stants is best illustrated by examples. Two
kinds of experiments can be distinguished, de-
pending on whether changes in ligand or pro-
tein fluorescence are measured. In each case,
the ligand is added in small aliquots to the
protein solution and is the absorbing species
which necessitates the correction for primary
absorption effects; the procedure used for de-
termining the correction factor depends on
whether the ligand is also the fluorophore.

Changes in Ligand Fluorescence

Figure 3 shows the data from an experiment or> L I


0 0.1 0.2 0.3
in which the binding of cNADP+ to dihydro-
CONCENTRATION mM
folate reductase was measured by following
the increase in fluorescence of the ethenoad- PIG. 3. Measurement of the binding of cNADP+ to
enine fluorophore on binding (cf. (2 1)). Figure dihydrofolate reductase by using the increase in its flu-
orescence (excitation, 340 urn; emission, 400 nm) on
3A shows the fluorescence as a function of binding. Solution conditions: 500 mM KCl, 50 mM citrate,
ligand concentration for the ligand alone and pH 5.1. (A) Fluorescence in the absence (0) and presence
in the presence of the enzyme. In the absence (0) of the enzyme, uncorrected for inner-filter effects.The
of any primary absorption effect, the fluores- line through the dam obtained in the presence of the en-
cence in the absence of enzyme will be linear zyme is calculated for the best fit value of K, assuming
no inner-filter effect. This assumption implies that the
with ligand concentration, and the data in Fig. data in the absence of enzyme should be adequately de-
3A can be analyzed on this assumption, using scribed by a straight line. (B) Pluorescence in the absence
Eq. [ 1 l] and [ 121, to obtain a value for the (0) and presence (0) of the enzyme after correction for
binding constant. The value obtained is 2.32 the inner-filter effect as described in the text (correction
(f0.75) X lo4 M-; the theoretical curve cal- factors calculated by fitting the open circles in (A) to Po.
[ 131).The line through the solid circles is calculated using
culated with this value of K is shown in Fig. the best fit value of K; the open circles, having been
3A. Although the fit of this curve to the data corrected, are now adequately described by a straight line.
is probably acceptable, it is not good. It is in (C) Data points derived by a point-by-point subtraction
fact clear that the fluorescence in the absence of data in the absence of enzyme from that in the presence
of enzyme is not simply a linear function of of enzyme (open and closed circles of part A). The line
is calculated using the best fit value of K. In each case
ligand concentration-there is a significant the brackets indicate the 95% confidence limits derived
primary absorption effect. By fitting the data from the nonlinear regression analysis; in (B), the brackets
in the absence of enzyme to Eq. [ 131, the val- are smaller than the symbols.
ues of a and d can be calculated, allowing the
appropriate correction to be applied to the
data obtained in the presence of the enzyme. using the best fit value of K, 5.83 (+O. 10)
The corrected data are shown in Fig. 3B, to- X lo3 M-. The corrected data are clearly in
gether with the theoretical curve calculated much better agreement with the theoretical
PROTEIN-LIGAND FLUORESCENCE CORRECTIONS 359

curve, and it is notable that the binding con- should not itself form a complex with the li-
stant is not only much better determined but gand(s) of interest. In view of the appreciable
also differs by a factor of 4 from that estimated bandwidth often used for the excitation beam,
from the uncorrected data. Failure to take ac- it is also desirable that the standard should
count of the primary absorption effect can have an excitation spectrum of similar shape
thus lead to substantial systematic errors in to that of the protein, and it is convenient if
the binding constant derived from fluores- they have similar emission wavelengths. For
cence data. these reasons, tryptophan has often been used
One approach to the analysis of data such as a standard in experiments of this kind.
as this which has sometimes been used is to However, its photochemical lability is a sig-
subtract, point by point, the fluorescence ob- nificant disadvantage in this context, and we
served in the absence of enzyme from that have sometimes found it more convenient to
observed in its presence. In the absence of any use N-( p-aminobenzoyl)-L-glutamate. Al-
primary absorption effect, this procedure is of though this compound has an excitation spec-
course correct (cf. Eq. [ 1 l]), but its application trum which is less similar to that of the average
to uncorrected data will give erroneous results. protein, the same emission wavelength can be
(Absorption of the excitation beam produces used and it is photochemically stable.
the same proportional decrease in the lluo- The procedure can be illustrated with ref-
rescence with and without enzyme, not the erence to Fig. 4, which shows the quenching
same absolute decrease.) Figure 3C shows data of the fluorescence of dihydrofolate reductase
calculated by this subtraction procedure from by folate. The uncorrected data, shown in Fig.
that shown in Fig. 3A. The binding constant 4B, can be fitted tolerably well by a theoretical
calculated by regression analysis is 1.29 (+_O.1) binding curve with K = 3.45 (~L0.4) X lo5
X lo4 Mp, a factor of 2.2 greater than that MI. However, the decrease in fluorescence of
obtained by analysis of the properly corrected the standard shown in Fig. 4A indicates that
data. there is a significant primary absorption effect.
In addition to the filter effect, the fluorescence
Changes in Enzyme Fluorescence of the standard (here tryptophan) will be de-
creased by simple dilution on addition of the
When binding is followed by means of the ligand. The open circles in Fig. 4A show the
quenching of the enzyme fluorescence accom- standard fluorescence corrected for this di-
panying complex formation, it is obviously lution. These data can now be used to estimate
important to correct for that part of the ob- the parameters a and cl in the correction term
served decrease in fluorescence which is due by curve fitting using Eq. [ 131, but now with
simply to the absorption of the excitation F co* = constant (since the concentration of
beam by the added ligand. When the ligand the standard is constant). The curve obtained
fluorescence is monitored, values of a and d with the optimum values of a and d is shown
in Eqs. [7] and [ 131 can be obtained from the in Fig. 4A. The appropriate correction factors
measurements made in the absence of enzyme can then be calculated for each ligand con-
(see above). However, when the protein flu- centration. The corrected enzyme fluorescence
orescence is followed, a separate calibration is shown in Fig. 4B, and the standard fluo-
curve must be obtained by measuring the rescence in Fig. 4A; the latter data points are,
effects of light absorption by the ligand on as expected, scattered about a horizontal line.
the fluorescence of a standard fluoro- The corrected enzyme fluorescence gives a
phore. The only essential criteria for the choice significantly better fit to the theoretical curve,
of this standard are that it should fluoresce with a value of K = 1.49 (20.13) X 10 M- .
when excited at the wavelength to be used in a factor of 2.3 less than that obtained from
the ligand-binding experiment and that it the uncorrected data.
360 BIRDSALL ET AL.

900

T
3 800
E
E 700
s
&
600

500
E
8
8 400

3
Lt. 300

t 200
E
c
100

6
f
2 700 3
5
s
I5
600

8 500
5
Bu
E 400
s
u.
g 300
E
1= 200

100

0
0 10 20 30
Concentration of Folate [MicromolZJ

FIG. 4. The binding of folate to dihydrofolate reductase as measured by the quenching of the enzyme
tluorescence (excitation, 290 nm; emission, 340 nm). (A) Standard (tryptophan) fluorescence, shown as
raw data (0) and after correction for dilution (0). The solid line is the beat fit to Bq. [ 131, and using
the correction factors obtained from this fit the corrected data points (m) are seen to be scattered about
the theoretical (dashed) fine. (B) Enzyme fluoresence. Uncorrected (0) and corrected (O)data, with brackets
indicating the 95% confidence limits from the nonlinear regression analysis.

CONCLUSIONS effect can lead to appreciable errors in the


estimated equilibrium constant. These errors
The results presented here show that failure will tend to be greater in experiments involving
to correct properly for the primary absorption enhancement of ligand fluorescence than in
PROTEIN-LIGAND FLUORESCENCE CORRECTIONS 361

those where the quenching of protein fluo- 4. Guilbault, C. C. (1973) Practical Fluorescence, Dek-
ker, New York,
rescence is measured. In the former case, the
5. Wehry, E. L. (ed.) (1976) Modem Fluorescence Spec-
largest primary absorption corrections must troscopy, Vols. 1 and 2, Dekker, New York.
be applied to the largest signals, whereas in 6. Miller J. N. (ed.) (1981) Standards in Fluorescence
the latter case, if there is a substantial quench- Spectrometry (UV Spectrometry Group), Chapman
ing, the large correction is applied to a small & Hall, London.
7. Parker C. A., and Rees, W. T. (1962) Analyst 87,83.
signal, and the absolute change in fluorescence
8. Stokes, G. G. (1852) Phil. Trans. 142, 463.
due to the primary absorption effect is much 9. Holland, J. F., Teets, R. E., Kelly, P. M., and Timnick.
less. The primary absorption effect can be A. (1977) Anal. Chem. 49, 706-710.
minimized by proper choice of excitation 10. Lloyd, J. B. F. (1981) In Standards in Fluorescence
wavelength, and if the absorbance is less than Spectrometry (UV Spectrometry Group), (Miller,
0.2, any one of several equations can be used J. N., ed.), Chapman & Hall, London.
11. Ehrenberg, M., Cronvall, E., and Rigler. R. ( 197 1)
to calculate the correction factor from the ex- FEBS Lett. 18, 199-203.
tinction coefficient at the excitation wave- 12. Brand, L., and Witholt, B. (1967) in Methods in En-
length. With larger absorption effects, the ap- zymology (Him., C. H. W., ed.), Vol. 11. pp. 776-
plicability of the assumptions involved in de- 856, Academic Press, New York.
riving the equations to the optical arrangement 13. Engel, G., Heider, H., Maelicke, A., von der Haar,
F., and Cramer, F. (1972) Eur. J. Biochem. 29,
of the spectrometer must be considered. The 251-262.
simplest and most general solution is to use 14. Helene, C., Brun, F., and Yaniv, M. (1971) J Mol.
the semiempirical Eq. [7] in conjunction with Biol. 58, 349-365.
calibration curves. 15. Lamola, A. A., Eisinger, J., Blumberg, W. E.. Ko-
metani, T., and Bumham, B. F. (1977) J. Lab.
C/in. Med. 89, 881-890.
ACKNOWLEDGMENTS 16. Mertens, M. L., and Kagi, J. H. R. ( 1979) Anal.
Biochem. 96,448-455.
17. Chen, R. F. (1967) Anal. Biochem. 19. 374-387.
We are grateful to Dr. J. G. Batchelor and Mr. B. J. 18. Roberts, G. C. K. (198 1) In Standards in Fluorescence
Hammond for substantial help at various stages in the Spectrometry (UV Spectrometry Group). (Miller,
development of the set of programs currently used at
J. N., ed.), Chapman & Hall, London.
NIMR and to P. C. Turner for valuable technical assis- 19. Dann, J. G., Ostler, G., Bjur R. A., King, R. W..
tance. The part of the work carried out at the University
Scudder, P., Turner, P. C., Roberts, G. C. K., Bur-
of South Carolina was supported by NIH through Grant
gen, A. S. V., and Harding, N. G. L. ( 1976)
CA 12842 from the National Cancer Institute. R.B.D. is
B&hem. J. 157, 559-571.
recipient of a Faculty Research Award (FRA-144) from 20. Birdsall, B., Burgen, A. S. V., Rodrigues de Miranda,
the American Cancer Society.
J., and Roberts, G. C. K. (1978) Biochemistry 17,
2102-2110.
21. Birdsall, B.. Burgen A. S. V., and Roberts, G. C. K.
REFERENCES (1980) Biochemistry 19, 3723-373 1.
22. Batchelor, J. G. (1977) Brit. J. Pharmacol. 59, 521P.
1. Chen, R. L., and Edelhoch, H. (eds.) (1975-6) Bio- 23. Marquardt, D. W. (1963) J. Sot. Ind. Appi. Math.
chemical Fluorescence: Concepts, Vols I and 2, 11, 431.
Dekker, New York. 24. Holbrook, J. J. (1972) Biochem. J. 128, 921-931.
2. Udenfriend, S. (1969) Fluorescence Assay in Biology 25. Chen, R. F., and Hayes, J. E. (1965) Anal. Biochem.
and Medicine, Vol. 2. Academic Press, New York. 13, 523-529.
3. Parker, C. A. ( 1968) Photoluminescence of Solutions, 26. Parker, C. A., and Barnes, W. J. (1957) Ana/.vst 82,
Elsevier, Amsterdam. 606-6 18.

You might also like