Professional Documents
Culture Documents
MB35, MB45
Description: ........................................ Service Tool
Service Notes Release Date: ............................................. 5/1/2006
Released By: ........................................ M. Albright
Requirements:
1. File: MB45 Service Tool Install.zip
2. Application Software File:
a. MB45 Global 11670614L.mot (Latest version is V1.21)
b. MB35 Europe 11670806D.mot (Latest version is V 2.00)
c. MB35 Latin America 802509712.mot (Latest version is V 2.00)
d. MB35 US 11670787D.mot (Latest version is V1.21)
3. A PC running Windows 95 to XP
4. Interface cable Ohaus part number 80500525 (Serial extension cable)
8. Match the serial port parameters to the balance. Balance defaults are shown above.
9. Click on the tab Main Board Replacement
Note: The serial number and TDNR number are printed on a label on the back of the balance.
13. CAUTION Confirm the information entered above. This programming can only be
done once. If you have made and error, click the button Abort Data Entry then reenter
the data.
14. When the above information has been confirmed click the Start button.
The following screen will pop up to indicate the new main board has been configured.
When the software has been downloaded the following screen will be shown.
Error Messages
Indicates the main board has already been configured. A main board can be configured only once
and can not be reconfigured.
Indicates the TDNR number does not match the balance configuration selected. Click Abort
Data Entry then re-enter the correct data.