Pulse Oximeter Using Arduino


Ms. Chaitali G. Gavhale1, Ms.Pooja G. Karhale2, Ms.Komal B. Patil3, Ms.Priyanka K. Jadhav4,
Ms.Mayuri R. Pankar5
Department of Electronics and Telecommunication Engineering,PLITMS,Buldana1,2,3,4,5
Email: chaitaligavhale27@gmail.com1;poojakarhale@gmail.com2;komalpatil0695@gmail.com3;
jadhavpriyanka9995@gmail.com4; pankar.mayuri@gmail.com5;

Abstract-A pulse oximeter system is implemented, thats why system power consumption is very low.
which is used for real time monitoring of various We show that the suggested pulse oximetry system
health parameters of a patient. Oxygen saturation, works properly and that it is applicable for
heart rate and blood pressure are the vital monitoring and controlling a health parameter in
parameters calculated and transmitted to a receiver. real time.
This paper relates the working principle of the Keywords Pulse oximeter,non-invasive,oxygen
pulse oximeter which is an non-invasive medical saturation,arduino uno.
equipment which is capable of measuring pulse
rate, arterial blood oxygenation, hypertension and
hypotension. This system uses ARDUINO UNO as
a microcontroller having ultra-power capability
during the circulation of blood over the another body
1. INTRODUCTION parts is called as blood pressure.
Currently, there are many different types of pulse
Persons health is very vital factor in a world like oximeters are present in the market. They differ in
ours, so this paper is presented about a biodevice for size, quality and cost. The complex ones used in
health. Non-invasive pulse oximeter is a safe, reliable, hospitals are typically large and not portable.
real time and continuous measurement method and it However, there are various hand held pulse oximeters
is widely used in patient monitoring. Pulse oximeter for home use which are very small and easy to use.
measure not two, but three very important patient The most familiar model is the finger model, which
parameter : blood SPO2, heart rate ,low blood displays the blood oxygenation level (SpO2) and heart
pressure and high blood pressure. Pulse oximeters rate[3]. A wrist-finger model is very common in the
measure such parameters to a high degree of market.
accuracy. People with heart related or pulmonary
disease have to monitor their blood oxygenation 2. LITERATURE SURVEY
continuously, especially while jogging or
exercising[1]. The measurement of o2 saturation in arterial by
In a blood vessels, the hemoglobin with oxygen transmiting light through a translucent area of a
are called oxygenated hemoglobin (oxy Hb), whereas patient,is called oximetry[1]. In early days , a painful
the hemoglobin without oxygen are called arterial blood gas technique is utilized for
deoxygenated hemoglobin (deoxy Hb).A low blood computation of oxygen saturation. But this process
oxygen saturation is a signal of many bodily illness, was quite time consuming that it took 20 to 30
specifically respiratory problems. Allow oxygen minutes to obtain the results. This delay is not
saturation can be from poor lung compliance due to acceptable as severe brain damage can occur within 5
pneumonia, ADRS, fibrosis. It can be signal of airway minutes of low oxygenation. According to survey,
restriction or obstacles as a result of asthma chronic every year 2,000 to 10,000 patients died because of
bronchitis, cystic fibrosis, choking secretions from undetected hypoxemia and without estimation of
intubation, or obstructive sleep apnea. A low blood patient morbidity.
saturation can even be used to diagnose lead In 1935 Karl Matthe have invented the first pulse
poisoning and congenital heart disease[2]. The basic oximeter device which is used to measure the oxygen
operation of our system depends on the use of LED saturation in blood[4]. At that time the values of O2
light, which is used for computation of the oxygen saturation is not directly displayed on LCD but
saturation. The spo2s in the blood is determine by display in the from of graph & that graph one can
using red light took up by the blood and detector. The calculate O2 saturation.
second sensor used in the system is the blood pressure In 1937 the scientist name Hertzman have made
sensor. The pressure exerted on the blood arteries modification of the pulse oximeter invented by Karl
matthe ,Hertzan have made usage of the photoelectric

169
International Journal of Research in Advent Technology (IJRAT)
(E-ISSN: 2321-9637)
Special Issue National Conference CONVERGENCE 2017, 09th April 2017
technique & device like photodiode & LED using this Further signal processing and computing are
technique the plethsmograph is obtain and from done in arduino uno which gets analog input from the
plethsmograph one can made assumption of the O2 finger probe and pressure sensor. The result can be
saturation. showed on the LCD display and buzzer gives the
Later on in 1972 Takuo Aoyagi have made indication of problem if any. In this paper, we focus
modification on pulse oximeter that was developed by on the description of the pressure sensor, finger probe
Hertzman.Takuo Aoyagi pulse oximeter measures the and arduino uno.
arterial oxygen saturation using pulsatile changes in
absorption of red & infra-red light[6]. when changes 3.1 FINGER PROBE:
in oxygen in blood then there is a changes in red and The sensor probe contains the Light Emitting
infra-red light intensity & according to that one can Diode and a photo-detector to receive the light
measure the oxygen saturation after 3 years Japan coming from our finger.
have introduce their pulse oximeter.
A notice was published by FDA in 2007, in Federal
Registers named as "Draft Guidance for Industry and
Food and Drug Administration Staff; Pulse Oximeters
Premarket Notification Submissions.s;
Availability".Pulse oximeters appeared on the market
after approval of FDA[7].
Recently: Pulse oximeter has become affordable and
widely available for domestic use. Fig. 2 Finger Probe

3. SYSTEM DESIGN In particular, the required condition of red LED,


is because of its place on the light extermination
Fig.1 gives the schematic of pulse oximeter curve, unpredictability in the Light Emitting Diode
system which consist of amplifier unit, finger probe, wavelength can vary the exactness of the final SPO2
pressure sensor, LCD display, buzzer and arduino computation. In that case, having a narrow bandwidth,
uno. or high intensity, is necessary. Having a wavelength
of 660nm is important, because almost all the
investigation has been concluded with this
wavelength, as its the easy way work with. Behind
the finger touch, there should be a photo-detector used
to convert the light signal into electrical signal. The
amplitude of these electrical signal is extremely low.
Therefore, an amplifier is added directly after the
photo-detector, to form the reading detectable. This
amplifier is called as comparator amplifier that is used
not only to amplify but also to convert the current into
voltage[7].

3.2 PRESSURE SENSOR:

The pressure exerted by the blood upon the


walls of the blood vessels specifically the arteries is
known as blood pressure. After pulsation of heart, it
Fig.1 System design diagram. contracts and pull down blood over the arteries to the
remaining parts of your body. This force causes
pressure on the arteries. Blood pressure is of two
The amplifier unit uses operational op-amp as a types as systolic pressure (as the heart beats) and
comparator which compares blood particles and diastolic pressure (as the heart relaxes between beats).
oxygen and gives voltage difference in the amplified The unit which measures this is
form. called Sphygmomanometer.
Finger probe is used to acquired original light
signal reflected from fluctuating blood and tissue.
pressure sensor that can measures a persons blood
pressure and heart rate through an inflatable hand
cuff.

170
International Journal of Research in Advent Technology (IJRAT)
(E-ISSN: 2321-9637)
Special Issue National Conference CONVERGENCE 2017, 09th April 2017

Fig.3 Pressure Sensor

Normally when the physician takes the patient's Fig.4 Graph between Times v Spo2/Heart Rate
blood pressure, he pumps the air into the cuff. Air is
pumped to be above the systolic value at start. Then Table 1 Output Comparison of Proposed Device and
the pressure is evolved slowly, at one point, the Existing Device
pressure in the cuff relates to the systolic pressure.
Then the pressure in the cuff will decreases furthers.
The diastolic pressure will be recorded at a point
when the variation starts to disappear.

4 .FLOW CHART

Our primary objective in designing a new pulse


oximeter was to reduce its power consumption as far
as possible without sacrificing any of its critical
medical properties. Therefore, we carefully examined
5. RESULT ANALYSIS the power requirement of various stages of our pulse
oximeter. Though our produced pulse oximeter
Table 1 gives the combined analysis of pulse currently dissipates considerably less power than
rate, oxygen saturation, high blood pressure and low other actual oximeters, it is possible to reduce the
blood pressure of device. Fig.4 shows the graphical power dessipation of our oximeter even further to
analysis of these demonstrated data. Here both oxygen around 1 mW. Beyond this optimal point, our
saturation and pulse rate measured at the same time. comparator amplifiers would have to burn more
After comparing these values with desired values, power than our LEDs to sense their reduced input
there will be some difference between the measured light intensity so that the total power requirement of
values and actual values[6]. the oximeter would rise again[9].
To remove these errors, ring sensor and
combination of low pass filter, notch filter and high 6. CONCLUSION
pass filter are used in proposed device. These
variations occur due to hand movement and small The reflection oximeter we design can obtain
portion of ac signal representing the arterial pulse. oxygen saturation and heart rate information
continuously, non-invasively and real-time through a
pressure sensor and finger probe sensor.

171
International Journal of Research in Advent Technology (IJRAT)
(E-ISSN: 2321-9637)
Special Issue National Conference CONVERGENCE 2017, 09th April 2017
Our low-power pulse oximeter achieved more
than an order-of-magnitude reduction in power
consumption over the best commercial pulse
oximeters, and yielded similar performance as we Ms.Priyanka K. Jadhav.
confirmed with experimental measurements. Most of Pursuing Bachelors Degree in
this reduction was because of the use of a energy- final year Electronic &
efficient arduino uno and comparator amplifier that Telecommunication. Engg from
was inherently sensitive to the signals required for PLITMS Buldana, SGBAU,
pulse oximetry computations, and which was built
with distributed amplification, adaptive loop gain
Amravati, (MS), India.
control. Consequently, the highest source of power Email:jadhavpriyanka9995@g
dessipation, LED powers could be significantly mail.com
reduced. Our oximeter is suited for use in emerging
portable and wearable medical applications.

REFERENCES
Ms. Mayuri R. Pankar.
[1]Baily, M., Katz, B., & West, D. Brookings. Pursuing Bachelors Degree in
Retrieved on:20th of November of 2012, URL: final year Electronic &
http://www.brookings.edu/research/papers/2011/05/go Telecommunication. Engg from
wth innovation PLITMS Buldana, SGBAU,
[2] J.G.Webster: Medical Instrumentation, IIIed.,
John Wiley & Sons.
Amravati, (MS), India.
[3] J. G. Webster, Design of Pulse Oximeters. Bristol, Email:pankar.mayuri@gmail.co
U.K.: Instrumentation Physician 1997. m@gmail.com
[4] C.A. Mead, Analog VLSI and Neural Systems.
Reading, MA: Addison-Wesley, 1989.
[5] A. Milenkovic, C. Otto, and E. Jovanov, Wireless
Sensor Network for Personal Health Monitoring: Ms. Pooja G. Karhale.
Issues and an Implementation, in Computer Diploma in (E & TC Engg) from
Communications, vol. 29 (13- 14), August 2006.
MSBTE, Mumbai. Pursuing B.E.
[6] F. Tay, D. Guo, L. Xu, M. Nyan, and K. Yap,
MEMS Wear bio monitoring System for Remote (E&TC Engg) final year from
Vital Signs Monitoring, in Journal of the Franklin PLITMS, Buldana. SGBAU
Institute, 346(6), 531-542, August 2009 Amravati, (MS), India.
[7] Sarussi, ``Device for use with reluctive pulse Email:poojakarhale46@gmail.co
oximetry,''U.S. Patent 20 070 038 050, Feb. 15, 2007. m@gmail.com
[8] Y. Shieh, Y. Tsai, A. Anavim, M. Shieh, and M.
Lin, Mobile Healthcare: Opportunities and
Challenges, in International Journal of Electronic Ms. Komal B. Patil
Healthcare, 4(2), 208-219, 2008. Diploma in (E & T Engg) from
[9] Jinling Zhang,Jun Wang, Xin-sheng L v, and MSBTE, Mumbai. Pursuing
Yunpeng Han, Design of Health Monitoring
Platform Base on Medical Wisdom, Applied
B.E. (E&TC Engg) final year
Mechanics and Materials (in Chinese). from PLITMS, Buldana.
SGBAU Amravati, (MS),India.
Authors Profile: Email:komalpatil0695@gmail.c
om.
Ms. Chaitali D. Gavhale
Diploma in (E & T Engg) from
MSBTE, Mumbai. Pursuing
B.E. (E&TC Engg) final year
from PLITMS, Buldana.
SGBAU Amravati, (MS),India.
Email:
chaitaligavhale27@gmail.com

172

You might also like