You are on page 1of 7

Article

pubs.acs.org/ac

Inaccuracies of Nitric Oxide Measurement Methods in Biological


Media
Rebecca A. Hunter, Wesley L. Storm, Peter N. Coneski, and Mark H. Schoensch*
Department of Chemistry, University of North Carolina at Chapel Hill, Chapel Hill, North Carolina 27599, United States
*
S Supporting Information

ABSTRACT: Despite growing reports on the biological


action of nitric oxide (NO) as a function of NO payload,
the validity of such work is often questionable due to the
manner in which NO is measured and/or the solution
composition in which NO is quantied. To highlight the
importance of measurement technique for a given sample type,
NO produced from a small-molecule NO donor (N-
diazeniumdiolated L-proline, PROLI/NO) and a NO-releasing
xerogel lm were quantied in a number of physiological
buers and uids, cell culture media, and bacterial broth by the
Griess assay, a chemiluminescence analyzer, and an ampero-
metric NO sensor. Despite widespread use, the Griess assay proved to be inaccurate for measuring NO in many of the media
tested. In contrast, the chemiluminescence analyzer provided superb kinetic information in most buers but was impractical for
NO analysis in proteinaceous media. The electrochemical NO sensor enabled greater exibility across the various media with
potential for spatial resolution, albeit at lower than expected NO totals versus either the Griess assay or chemiluminescence. The
results of this study highlight the importance of measurement strategy for accurate NO analysis and reporting NO-based
biological activity.

N itric oxide (NO), an endogenous free radical produced by


a collection of enzymes known as nitric oxide synthases
(NOS), is a physiological mediator of the cardiovascular,
for the indirect determination of NO totals, the limit of
detection for Griess is only 0.5 M.12,22,23 In contrast,
chemiluminescence detection via a commercially available NO
immune, and nervous systems.1 For example, NO produced in analyzer is a more costly method primarily due to the
the vasculature by endothelial NOS serves as a vasodilator and instrumentation, but the NO is measured more directly via
blood pressure regulator.2,3 The immune system produces NO reaction with ozone.24 Since NO is also redox-active,
at high concentrations via inducible NOS to serve as a signaling
electrochemical sensors have also been developed to quantify
molecule4,5 and potent antimicrobial agent.6 In the brain, NO
produced by neuronal NOS functions as a neurotransmitter NO from sources as small as single cells made possible via
and is involved in memory formation.7 The physiological sensor miniaturization.23
signicance of NO has led to increased research on NO and While each of these techniques has inherent advantages for
NO-releasing scaolds as potential therapeutics.814 Given that measuring NO, it is important to consider the environment
the location and concentration of NO governs its biological (i.e., solution or sample) in which the measurement takes place.
activity, the validity of analytical methods for measuring NO is Since NO readily reacts with free radical species (e.g.,
critical. superoxide, thiyl radicals, and lipid peroxyls), metal-containing
While understanding the behavior of NO in vivo is proteins (e.g., hemoglobin), and (to a lesser extent) thiols, the
important, this task is far from trivial. Nitric oxide is reactive available (i.e., free) NO will vary depending on the sample
and short-lived, with a lifetime on the order of seconds to medium.2527 While recent reviews have focused on the
minutes.15 Additionally, the diusion of NO is rapid, with methodologies for measuring NO in biological sam-
diusion coecients approaching 3300 m2s1 in physiological ples,23,24,2835 an understanding of the inuence of the
buer.16,17 Complicating matters further, the physiological technique and sample milieu on the validity of such
concentration of NO spans 6 orders of magnitude (picomolar
measurements is lacking. Herein, we evaluate the accuracy of
to micromolar).18 Sensitive analytical tools for measuring NO
over wide concentration ranges are thus a necessity.19 To date, NO analysis with respect to measuring NO in physiological
three analytical techniques account for the majority of NO buers and uids, cell culture media, and bacterial broth using
measurements used in the literature to quantify NO.12,20 First
described in 1864,21 the Griess assay (Griess) allows for the Received: December 28, 2012
estimation of total NO concentrations via nitrite analysis. Accepted: January 3, 2013
Although inexpensive, readily available commercially, and useful Published: January 3, 2013

2013 American Chemical Society 1957 dx.doi.org/10.1021/ac303787p | Anal. Chem. 2013, 85, 19571963
Analytical Chemistry Article

the Griess assay, a chemiluminescence NO analyzer, and an Synthesis of NO-Releasing Xerogels. Nitric oxide-
amperometric NO sensor. releasing xerogels were fabricated by a procedure similar to

EXPERIMENTAL SECTION
Reagents and Materials. (Heptadecauoro-1,1,2,2-
one described previously.40 Briey, 378 L of BTMOS was
prehydrolyzed in 633 L of ethanol, 190 L of Milli-Q water,
and 31.7 L of 0.5 M hydrochloric acid for 1 h with agitation.
tetrahydrodecyl)trimethoxysilane (17FTMS), isobutyltrime- Next, 255 L of AHAP3 was added and the resulting
thoxysilane (BTMOS), and N-(6-aminohexyl)- aminosilane/alkylsilane sol was mixed for an additional 1 h.
aminopropyltrimethoxysilane (AHAP3) were purchased from Glass micro slides (Gold Seal, Portsmouth, NH) were cut into
Gelest (Morrisville, PA). Methyltrimethoxysilane (MTMOS) 9 25 mm sections and sonicated successively in water,
was purchased from Fluka (Buchs, Switzerland). Griess assay ethanol, and acetone. The substrates were dried under a stream
reagents were purchased from Promega (Madison, WI). of nitrogen and then cleaned in an ultravioletozone cleaner
Tryptic soy broth (TSB) and brain heart infusion (BHI) (BioForce Nanosciences; Ames, IA) for 20 min. Afterward, 40
broth were purchased from BD Biosciences (San Jose, CA). L aliquots of the sol were cast onto each substrate, dried for 1
Dulbeccos modied Eagle medium (DMEM), McCoys h in ambient conditions and cured at 70 C for 3 days. To
medium 5A modied, fetal bovine serum (FBS), dipotassium convert the secondary amines in the matrix to N-diazenium-
ethylenediaminetetraacetic acid (K2EDTA), nicotinamide ad- diolate NO donors, the xerogels were placed in a stainless steel
enine dinucleotide phosphate (NADPH), and nitrate reductase 500 mL Parr bomb, purged copiously with argon gas, and held
(from Aspergillus niger) were purchased from Sigma (St. Louis, under 10 atm NO gas for 3 days. After additional argon purging
MO). Opti-MEM I (a reduced serum medium) was purchased to remove unreacted NO, the lms were sealed under nitrogen
from Life Technologies (Grand Isle, NY). Leibovitz medium and stored at 20 C until use.
(L-15; a carbon dioxide-free cell culture medium) was Griess Assay. To quantify NO via the Griess assay,41 50 L
purchased from Lonza (Basel, Switzerland). Porcine blood of a 2 mgmL1 solution of PROLI/NO in 100 mM sodium
was obtained from the Francis Owen Blood Research hydroxide (NaOH) was added to 15 mL of desired media and
Laboratory (University of North Carolina, Chapel Hill, NC). incubated at room temperature for at least 24 h. Aliquots (50
Blood serum was obtained by collecting porcine blood without L) of this sample were added to a sulfanilamide solution (50
the addition of anticoagulant. After the blood was allowed to L, 46.5 mM water) and incubated in the dark at room
clot, it was centrifuged at 2500 rpm for 15 min and the temperature for 5 min. N-(1-naphthyl)ethylenediamine (50 L,
supernatant (i.e., serum) was removed. To obtain blood 3.9 mM in 5% (v/v) phosphoric acid in water) was added to
plasma, porcine blood was drawn into a tube with K2EDTA the mixture to form a colorimetric product with concomitant
(1.8 mgmL1), mixed immediately, and then centrifuged at absorbance measured in each well at 540 nm by use of a
2500 rpm for 15 min. After centrifugation, three layers were LabSystems MultiSkan RC microplate reader (Helsinki,
present (from top to bottom, plasma, leukocytes, and Finland). Sodium nitrite standards were used to normalize
erythrocytes); the top layer was removed. A Millipore Milli-Q the assay reactivity and associated absorbance.
UV gradient A10 system (Bedford, MA) was used to purify For analysis of blood constituents (i.e., plasma and serum),
distilled water to a nal resistivity of 18.2 Mcm and a total NADPH (25 L) and nitrate reductase (2 L) were added to
organic content of 6 ppb. Nitrogen and argon gases were the sample and the mixture was allowed to incubate for at least
purchased from AirGas National Welders (Raleigh, NC). Nitric 30 min prior to addition of the Griess reagents. Headspace
oxide gas was purchased from Praxair (Danbury, CT). Other studies via Griess were conducted in the same manner, but the
solvents and chemicals were analytical-reagent grade and used volume of medium added to the 20 mL scintillation vial was
as received. varied (10, 15, and 20 mL). Concentration dependence studies
Phosphate-buered saline (PBS; 10 mM, pH 7.4), were conducted by varying the concentration of the stock
physiosol36 (pH 6.0), articial urine37 (pH 5.9), and articial PROLI/NO solution and injecting aliquots into 15 mL of
saliva38 (pH 6.7) were prepared as described in Supporting medium to yield nal PROLI/NO concentrations of 0.67, 6.7,
Information. A saturated NO solution (1.9 mM NO) was made and 67 gmL1.
by purging 20 mL of PBS with argon for 30 min to remove Chemiluminescence Detection. Real-time NO release
oxygen, followed by NO gas for 20 min. was monitored by a Sievers 280 chemiluminescent NO analyzer
Synthesis of PROLI/NO. N-diazeniumdiolated L-proline (Boulder, CO). The instrument was calibrated with a 25.6 ppm
(PROLI/NO) was prepared following a previously published gas standard (balance N2) and an atmospheric sample that had
protocol.39 Briey, L-proline (2.05 g) was dissolved in a been passed through a NO zero lter. Samples were prepared
solution of methanol (25 mL) and sodium methoxide (2.00 g). by adding 10 L of a 2 mgmL1 solution of PROLI/NO in 100
The solution was then placed in a stainless steel reaction vessel mM NaOH to 30 mL of desired medium that had been
and ushed with Ar six times (three in succession, three for 10 degassed in a sample vessel for at least 20 min. Nitric oxide
min each), then charged with NO at a pressure of 10 atm for 3 produced in the vessel was carried to the NO analyzer by a
days with constant stirring. Six additional Ar purges were stream of nitrogen gas bubbled into the solution (80
performed after 3 days. The solution was then precipitated by mLmin1) and across the headspace of the ask (120
the addition of diethyl ether (150 mL) at 20 C for 4 h. The mLmin1), equivalent to 200 mLmin1 ow to the instrument.
white precipitate was isolated by vacuum ltration and dried in Concentration dependence studies were conducted in the same
vacuo to yield PROLI/NO, which was stored at 20 C until manner, but the concentration of the stock PROLI/NO
use. Ultraviolet spectra of a 14.9 gmL1 solution of the solutions was varied to yield nal concentrations of 0.167,
product (in 1.0 M sodium hydroxide) was acquired on a 1.67, and 16.7 gmL1 in 30 mL. Release of evolved NO from
Thermo Scientic evolution array UVvisible spectrophotom- 40% AHAP (balance BTMOS) xerogels was measured similarly
eter. The molecular weight of pure PROLI/NO was taken to be following placement of the substrates directly into deoxy-
251 gmol1.39 genated PBS by use of the vessels described above.
1958 dx.doi.org/10.1021/ac303787p | Anal. Chem. 2013, 85, 19571963
Analytical Chemistry Article

Electrochemical Detection. Inlaid 2 mm diameter To accurately quantify NO, calibration curves are constructed
polycrystalline platinum (Pt) disk electrodes sealed in Kel-F by use of a standard nitrite solution in the sample medium.41
(CH Instruments, Austin, TX) were mechanically polished with In theory, 1 molecule of PROLI/NO decomposes to release
successively ner grades of deagglomerated alumina slurries 2 molecules of NO, for a theoretical total NO release of 7.9
down to 0.05 m particles (Buehler, Lake Blu, IL). Residual molmg1.39 Following synthesis, PROLI/NO was charac-
alumina was removed with an ultrasonic cleaner (in water) and terized by UV/vis spectroscopy (Supporting Information,
the electrodes were dried with nitrogen. A uorinated NO- Figure S1). An observed max at 252 nm conrmed N-
selective xerogel membrane was applied to the electrode as diazeniumdiolate formation from the L-proline precursor,
previously described to minimize response to common consistent with prior reports.39 A 14.9 gmL1 solution was
interferents.42,43 Briey, a silane solution was prepared by prepared in 1.0 M sodium hydroxide and the molar absorptivity
mixing MTMOS (60 L) in ethanol (300 L). To this solution coecient () reported previously by Saavedra et al.39 at 252
was added 17FTMS (15 L), resulting in a 20% (v/v) nm (8.4 mM1cm1) was used to determine a PROLI/NO
uoroalkoxysilane (balance MTMOS) mixture. The silane concentration of 55.7 M. If total purity is assumed, the
solution was subsequently mixed with water (80 L) and 0.5 concentration of this solution would be 59.5 M, implying a
M HCl (5 L) for 1 h. The resulting sol (1.5 L) was cast onto relative purity of 93.5%. This value is also supported by the
Pt working electrodes and allowed to cure for 24 h under chemiluminescent NO release totals in buers lacking
ambient conditions. To evaluate the analytical performance of scavenging components (i.e., PBS). Estimated NO release
the NO sensors, amperometric measurements were performed totals derived from the total nitrite present in each type of
by use of a CH Instruments 730B bipotentiostat (Austin, TX). medium as determined by the Griess assay are given in Figure 1
The electrode assembly (three-electrode conguration) con-
sisted of the xerogel-modied Pt working electrode, a Pt-coiled
counter electrode, and a Ag/AgCl reference electrode (3.0 M
KCl; CH Instruments). Electrooxidation currents were
recorded at an applied potential of +700 mV (vs Ag/AgCl).
When NO was measured in the various media with PROLI/
NO as the NO source, a 2 mgmL1 solution of the NO donor
(i.e., PROLI/NO) was added to a constantly stirring 30 mL
solution to achieve a nal concentration of 0.1 mgmL1. Of
note, the larger volume of medium was necessary to
accommodate the working, reference, and counter electrodes
in the ask. Concentration dependence studies were conducted
in the same manner, but the concentration of the stock
PROLI/NO solutions was varied to yield nal PROLI/NO
concentrations of 0.67, 6.7, and 67 gmL1. Xerogel lm
studies were conducted by placing the lm directly below a
working electrode and using a micromanipulator to adjust the
electrode distance from the NO-releasing surface.

RESULTS AND DISCUSSION


Diverse biological media were chosen for these experiments to
properly represent environments in which NO measurements
are relevant (e.g., in vivo, cell/tissue culture, bacteria culture). Figure 1. Total NO released from PROLI/NO in several types of
Solutions included simple physiological buers (PBS and biological media, determined via Griess assay (solid bars) and
physiosol), simulated biological uids (saliva, urine, and wound chemiluminescence (striped bars). Theoretical NO release from
PROLI/NO is 7.9 molmg1. Asterisks denote a signicant dierence
uid), cell culture media (DMEM, McCoys, L-15, and Opti-
(p < 0.05) relative to PBS.
MEM), bacterial broth (TSB and BHI), whole blood, plasma,
and serum. The salt, amino acid, protein, and vitamin content
of these media vary signicantly (Tables S1S4 in the along with totals derived from chemiluminescence detection.
Supporting Information) and thus were predicted to impact The total NO measured in PBS via Griess was only 5.67 0.27
the results. molmg1, roughly 30% lower than the theoretical amount and
NO Determination via Griess Assay. The Griess assay that detected via the chemiluminescence method (7.24 0.04
allows for indirect measurement of NO by quantifying nitrite, molmg1). Of note, care was taken to add fresh PROLI/NO
NOs reaction product in oxygenated media (eqs 13): solution to the media quickly. In addition, multiple samples
were analyzed to obtain standard deviations. Similar depressed
2NO + O2 2NO2 (1) NO levels were observed in all of the medium types tested by
Griess. We attribute the deviation between Griess and
NO + NO2 N2O3 (2) theoretical to lost (e.g., escaped) NO gas (from solution)
that subsequently was not accounted for in the total nitrite
N2O3 + H 2O 2NO2 + 2H+ (3) levels. Indeed, when the headspace in a 20 mL scintillation vial
was reduced by adding 20 mL of PBS versus 15 mL, the total
Once formed, the nitrite is reacted with the Griess reagents nitrite detected was increased by 44% 12%. Conversely, when
(i.e., sulfanilamide and N-(1-naphthyl)ethylenediamine) to the total buer volume was decreased to 10 mL, the total nitrite
produce an azo dye with an absorbance maximum at 540 nm. recovered was reduced by 32% 21% (data not shown). The
1959 dx.doi.org/10.1021/ac303787p | Anal. Chem. 2013, 85, 19571963
Analytical Chemistry Article

most signicant deviations of measured versus theoretical NO of NO release were undetectable with chemiluminescence after
totals were observed in bacterial broth (i.e., TSB and BHI), 2 weeks, the accumulation of nitrite was still measurable by the
McCoys and L-15 cell culture media, and plasma and serum. Griess assay and indicated a continued release of NO from the
These results were somewhat expected, as both cell culture and substrates.
bacterial growth media consist of complex mixtures of amino Chemiluminescence NO Analyzer. To facilitate real-time
acids, proteins, sugars, and vitamins of various concentrations measurement of NO with greater time resolution and lower
(Tables S1S4 in the Supporting Information). Proteins and limits of detection relative to the Griess assay, chemilumi-
other additives have been shown to interfere with the Griess nescence-based instrumentation is a standard tool for solution-
assay previously,4446 acting as either positive or negative based analysis.20 According to measurement principles from its
interferents. For example, positive interferents such as NOS original use (i.e., respiratory applications), NO is carried from
and hemoglobin absorb light around 540 nm. Other the sample (e.g., exhaled breath, solution vessel, etc.) by an
interferents including cysteine, tyrosine, ascorbate, and inert gas (e.g., nitrogen, argon) to an instrument where the NO
NADPH react with nitrite to negatively skew NO totals. is reacted with ozone to form an excited-state nitrogen dioxide
Although interfering proteins could be removed by chemical species that emits a photon upon relaxation back to the ground
precipitation or ultraltration,28 such solution conditioning is state. This light, ranging from 600 to 875 nm, is detected by a
tedious and would be at the expense of biological relevance. As photomultiplier tube and is proportional to the NO present in
anticipated, NO was not quantiable in whole blood via Griess the sample. The reaction of NO with ozone is both specic for
due to the opacity of blood and presence of interfering and sensitive to NO, with a detection limit approaching 0.5 ppb
proteins.45 Additionally, NO and nitrite are readily oxidized to (0.66 pM in 100 mL) in PBS.12 Detection limits in solutions
nitrate in whole blood by oxyhemoglobin.47,48 Nevertheless, other than PBS have not been reported previously.
nitrite levels were measurable in plasma and serum samples Analogous to the Griess experiments, samples of PROLI/NO
extracted from the blood, with the caveat that nitrate reductase (0.67 gmL1) were introduced into a solution but analyzed
and NADPH were used to revert nitrate that formed back to by chemiluminescence detection to quantify NO. Unexpect-
nitrite. Despite this, NO lost by its conversion to nitrosylheme edly, the media compatible with this method were limited to
via the reaction with deoxyhemoglobin could not be recovered solutions that did not foam as a result of purging with nitrogen
via this assay. In both plasma and serum, the NO totals were gas (necessary to carry the NO to the instrument and eliminate
>15% lower (4.75 1.1 and 4.18 0.85 molmg1, autoxidation to nitrite by removing oxygen). Foaming due to
respectively) than those in PBS, further indicating measure- nitrogen bubbling was most prominent for culture media,
ment inaccuracies in protein-rich samples. Of note, nitrite bacterial broth solutions, and blood (i.e., whole, plasma, and
recovery is highly variable and dependent on enzyme serum) due to their high protein content.50 Nitric oxide release
activity.28,33,49 totals via chemiluminescence for the compatible media are also
The accurate detection of NO via the Griess assay is also shown in Figure 1, allowing for direct comparison of total NO
concentration-dependent in certain types of media. While no detected with Griess as a function of media type. Of note, the
signicant dierences were observed for nonproteinaceous 7.24 0.04 mol of NO mg1 measured in PBS via
media, detection in the more complex media (i.e., cell culture chemiluminescence was near the predicted NO payload for
media and bacterial broth) varied signicantly depending on PROLI/NO based on the theoretical total and calculated
the amount of PROLI/NO added to solution (Figure S2 in relative purity. The NO totals measured in low protein content
Supporting Information). For example, while the addition of a media by chemiluminescence were not signicantly dierent
low PROLI/NO concentration (0.67 gmL1) in DMEM from theoretical NO payloads, indicating high accuracy of this
yielded nitrite totals that were only 48% 9% of those technique for measuring total NO. While L-15 cell culture
achieved at the highest concentration of PROLI/NO (67 medium contains biomolecular components that may scavenge
gmL1), an intermediate concentration (67 gmL1) yielded NO, such reactivity was limited as the NO was rapidly moved
nitrite amounts that nearly matched the highest concentration out of the sample vessel to the instrument by the carrier gas
(97% 1%). The same trend held true for the bacterial broth after formation.
TSB. This can be attributed to the eect of scavenging, as the A key advantage of chemiluminescence NO detection is the
concentration of such scavengers in a given medium is constant, ability to obtain kinetic information about NO release from
so adding more NO allows this eect to be overcome to some NO-donor systems. In addition to total NO amount, the
extent. maximum NO release rate, time to the maximum release (tmax),
Given these results, the use of Griess for quantifying NO in and NO-release half-life (t1/2) may be determined for a
most biological media leads to questionable results. Never- particular NO source (e.g., small molecule, macromolecular
theless, the potential for high-throughput analysis via microtiter scaold). These parameters were extracted from the NO-
plates and readers makes Griess useful for initial screening of release prole of PROLI/NO in PBS, physiosol, articial saliva,
NO-release materials in less complex media (e.g., PBS) so long and L-15 (Table 1). Little variation in the NO-release kinetics
as such data are supplemented with more rigorous analysis in for PROLI/NO in these media was noted for chemilumines-
relevant milieu prior to drawing conclusions regarding clinical cence, with two exceptions. In articial saliva, the maximum
utility. In addition, some kinetic data are obtainable via Griess NO release was signicantly lower than that in PBS (73 360
by taking aliquots from a sample solution at set periods or 1100 versus 86 200 3700 pmols1mg1, respectively). This
moving bulk (i.e., larger) substrates (e.g., polymer-coated behavior may be attributed to greater scavenging of NO by
slides) in and out of a soak solution and subsequently sampling components in saliva relative to PBS. Likewise, the half-life of
those solutions. For example, AHAP lms soaked in select PROLI/NO in L-15 cell culture media was smaller (reduced by
media yielded totals similar to those obtained under the same 8 s) than that measured in PBS. These variances are not
conditions by chemiluminescent detection (Table S5 in surprising given the complex and varying nature of biological
Supporting Information) after 1 week. Although the low levels media (Supporting Information).
1960 dx.doi.org/10.1021/ac303787p | Anal. Chem. 2013, 85, 19571963
Analytical Chemistry Article

Table 1. Kinetic Parameters of NO Release from PROLI/ proportional to the moles of analyte oxidized.64 More accurate
NO in Dierent Mediaa predictions of NO totals in the bulk solution were then
obtained by relating these values to a calibration curve created
medium tmax (s) max NO release (pmols1mg1) t1/2 (s)
by use of aliquots of a saturated NO solution (1.9 mM) in PBS.
PBS 42 1.9 86 200 3700 65 3.0 As shown in Figure 2, the integrated totals yielded NO
physiosol 40 1.7 81 200 36 64 1.1
L-15 40 1.2 96 600 3100 57 0.7*
saliva 47 5.2 73 400 1100* 74 3.5
a
Asterisks denote a signicant dierence (p < 0.05) relative to PBS.

Analogous to the Griess assay, the accurate measurement of


NO totals from PROLI/NO was dependent on the
concentration of the NO donor introduced into the sample
ask (Figure S2 in Supporting Information). As expected, no
dierence was observed between concentrations in the buer
solution (PBS) or the simulated biological solution tested
(saliva), as these are simply composed of various salts.
However, a slight decrease of 9% 1% was observed for the
lowest concentration relative to the 100 more concentrated
sample for the cell culture media tested (L-15 in this case, as
DMEM foams too signicantly to be used with this technique).
Despite the advantage of kinetic NO-release parameters, the
required purging of the sample vessel remains problematic for
measuring NO via chemiluminescence in biological samples
due to undesirable frothing. Furthermore, the reaction vessel
should be free of oxygen to minimize any side reactions that
would decrease NO concentrations entering the instru- Figure 2. Nitric oxide totals measured with an amperometric NO
ment.12,24 Unfortunately, mammalian cells and tissues that sensor and PROLI/NO as the NO source as a function of biological
produce NO require an aerobic environment for survival, thus medium, with (black) and without (gray) oxygen present. Theoretical
excluding chemiluminescence for measuring NO production NO release from PROLI/NO is 7.9 molmg1. Asterisks indicate
from cells and/or tissue in their native environment for signicant dierence (p < 0.05) relative to PBS; (#) signicant
extended periods. dierence (p < 0.05) relative to deoxygenated solution.
Amperometric NO Sensor. In contrast to Griess and
chemiluminescence, electrochemical sensors allow for the
measurement of NO in almost any biological setting down to concentrations signicantly lower than the theoretical amount
the single cell level.5153 The advantages of electrochemical of NO produced from PROLI/NO regardless of sample
detection of NO include superior spatial and temporal medium. These results were somewhat expected since the
resolution, excellent limits of detection and dynamic ranges, electrode is able to oxidize (i.e., measure) only a portion of the
and the ability to miniaturize the sensor. Both the selectivity NO in its vicinity due to probe geometry, despite vigorous
and sensitivity of the sensor are tunable by changing the applied stirring. While electrochemical sensors are clearly less suitable
potential to the working electrode and/or modifying the for characterization of NO donors or drugs that produce/
working electrode with a catalyst and/or permselective consume NO, useful information may be deduced by
membrane.12,54 Due to these inherent advantages, ampero- comparing the relative NO totals in dierent media and
metric NO sensors have been used extensively to examine both when more localized NO generation (e.g., from a surface or
endogenous11,52,53,5558 and exogenous5961 NO production. single cell) is quantied. The solutions resulting in the greatest
Nitric oxide measured by an amperometric sensor results in total NO detected when oxygenated were PBS and physiosol
current proportional to the NO concentration upon the (1.07 0.09 molmg1), not surprising due to the
oxidation of NO via a three-electron process (eqs 46):62,63 similarities of these solutions; only their salt concentration
and pH dier slightly.
NO NO+ + e (4) The NO totals in oxygenated L-15 and Opti-MEM cell
culture media were also in line with that in PBS despite
NO+ + OH HNO2 (5) signicant protein content. In assessing trends, we noted that L-
HNO2 + H 2O NO3 + 2e + 3H+ 15 and Opti-MEM solutions do not contain fetal bovine serum
(6)
(FBS) as do the DMEM and McCoys solutions, the latter
Depending on the thickness of the permselective membrane, resulting in NO scavenging via sulfhydryl-containing proteins
the response to NO can be fast (subsecond) with limits of (e.g., albumin, brinogen, macroglobulins, glycoproteins).65,66
detection approaching 83 pM.51 For this study, a NO-selective The amount of NO measured in simulated wound uid (10%
uorosilane-based xerogel-coated platinum working electrode42 v/v FBS in water) with the electrochemical sensor was also
was used to measure the electrooxidation of NO as current at signicantly lower (0.40 0.05 molmg1). In whole blood,
an applied potential of +700 mV versus Ag/AgCl reference undoubtedly the most complex of all the media tested, the
electrode in a well-stirred solution. Total NO release was measured NO total was the lowest [(12.0 3.2) 104
determined by integrating the current versus time response for molmg1] due to the anticipated reaction of NO with blood
each sample, since the charge at the electrode surface is proteins including oxyhemoglobin.67 Despite the lower totals in
1961 dx.doi.org/10.1021/ac303787p | Anal. Chem. 2013, 85, 19571963
Analytical Chemistry Article

whole blood, NO concentrations were still measurable with the considered with respect to sample, desired data (e.g., NO totals,
amperometric sensor, as was construction of a calibration curve ux, kinetics and bioavailability), and result integrity.
(in whole blood). Nitric oxide totals in plasma and serum
measured greater than in blood (0.27 0.09 and 0.11 0.04
molmg1, respectively) as would be predicted due to the

*
ASSOCIATED CONTENT
S Supporting Information
reduced hemoglobin concentration in these samples. Additional text, describing methods for preparing biological
While testing of deoxygenated media was limited to solutions (articial saliva, articial urine, and physiosol); ve
biological solutions that did not foam upon purging with tables, showing concentrations of key components in solutions
nitrogen, the measured NO levels were greater, as would be and NO release from a xerogel lm; and three gures, showing
expected upon eliminating NOs propensity to react with UV-visible spectra of PROLI/NO, concentration-dependent
oxygen.68,69 In blood and other media containing high protein trends of NO totals, and distance dependence of NO release
content and cells, fouling of the electrode resulting from from xerogel lms. This material is available free of charge via
protein/cell adsorption may also impact the analytical accuracy the Internet at http://pubs.acs.org.
of the measurement for electrochemical sensors.7073 In the
short experiments described herein, such fouling accounted for
only a 35% decrease in sensitivity (data not shown) and
AUTHOR INFORMATION
Corresponding Author
therefore did not contribute signicantly to the dierences *E-mail schoensch@unc.edu.
observed in the NO totals. Notes
Analogous to the Griess assay and chemiluminescent The authors declare no competing nancial interest.


detection, the accurate electrochemical detection of NO was
also dependent on the concentration of the NO donor. ACKNOWLEDGMENTS
However, a concentration-dependent eect was observed in all
We acknowledge research support from the National Institutes
media (Figure S2 in Supporting Information). We attribute this
of Health (NIH Grant AI097539) and the Francis Owen Blood
eect to the nite surface area of the working electrode,
Lab for porcine blood.


whereby greater concentrations of the NO donor readily alter
the local NO concentrations. REFERENCES
As the amount of NO donor near the electrode surface
(1) Epstein, F. H.; Moncada, S.; Higgs, A. N. Engl. J. Med. 1993, 329,
inuences the concomitant NO measured, the electrode
20022012.
distance from an NO source should have a similar eect on (2) Ignarro, L. J.; Buga, G. M.; Wood, K. S.; Byrns, R. E.; Chaudhuri,
NO measurement. To examine this methodically, a NO- G. Proc. Natl. Acad. Sci. U.S.A. 1987, 84, 92659269.
selective electrode was placed 25, 50, and 100 m above a NO- (3) Moncada, S.; Radomski, M. W.; Palmer, R. M. J. Biochem.
releasing surface (xerogel polymer cast on glass) under ambient Pharmacol. 1988, 37, 24952501.
conditions in PBS. The amperometric signal (current) obtained (4) Thomas, D. D.; Ridnour, L. A.; Isenberg, J. S.; Flores-Santana,
when the working electrode was placed 50 m above the W.; Switzer, C. H.; Donzelli, S.; Hussain, P.; Vecoli, C.; Paolocci, N.;
surface at 95 min was 45% of that recorded at 25 m (Figure Ambs, S.; Colton, C. A.; Harris, C. C.; Roberts, D. D.; Wink, D. A. Free
S3 in Supporting Information). Similarly, the signal was Radical Biol. Med. 2008, 45, 1831.
(5) Snyder, S. H.; Bredt, D. S. Sci. Am. 1992, 266, 7477.
reduced by 73% at 100 m above the surface (relative to
(6) Fang, F. C. J. Clin. Invest. 1997, 99, 28182825.
the 25 m working electrode placement). While such NO (7) Snyder, S. H. Curr. Opin. Neurobiol. 1992, 2, 323327.
source/electrode distance dependence will clearly impact the (8) Carpenter, A. W.; Schoenfisch, M. H. Chem. Soc. Rev. 2012, 41,
analytical accuracy of a measurement, it may be useful for 37423752.
measuring the diusion of NO from surfaces or cells as a (9) Riccio, D. A.; Schoenfisch, M. H. Chem. Soc. Rev. 2012, 41,
function of biological medium. 37313741.
Despite the lower NO totals quantied electrochemically, the (10) Nichols, S. P.; Storm, W. L.; Koh, A.; Schoenfisch, M. H. Adv.
use of amperometric sensors remains advantageous due to Drug Delivery Rev. 2012, 64 (12), 11771188.
unparalleled spatial and temporal resolution.74 As such, (11) Patel, B. A.; Galligan, J. J.; Swain, G. M.; Bian, X.
Neurogastroenterol. Motil. 2008, 20, 12431250.
electrochemistry remains the method of choice for the
(12) Hetrick, E. M.; Schoenfisch, M. H. Annu. Rev. Anal. Chem. 2009,
characterization of NO generation from cells11,52,53,5558 and 2, 409433.
conned surface locations.75 Unfortunately, both commercial (13) Damodaran, V. B.; Joslin, J. M.; Wold, K. A.; Lantvit, S. M.;
and home-built sensors are generally not robust, requiring Reynolds, M. M. J. Mater. Chem. 2012, 22, 59906001.
frequent (e.g., after analysis of 12 samples) calibration and (14) Seabra, A. B.; Martins, D.; Simoes, M. M. S. G.; Da Silva, R.;
performance testing to maintain data integrity. Brocchi, M.; De Oliveira, M. G. Artif. Organs 2010, 34, E204E214.

(15) Lowenstein, C. J.; Dinerman, J. L.; Snyder, S. H. Ann. Intern.


Med. 1994, 120, 227237.
CONCLUSIONS (16) Lancaster, J. R. Nitric Oxide 1997, 1, 1830.
The analytical measurement of NO is clearly complex,12 (17) Malinski, T.; Taha, Z.; Grunfeld, S.; Patton, S.; Kapturczak, M.;
requiring careful scrutiny of the analysis method with respect to Tomboulian, P. Biochem. Biophys. Res. Commun. 1993, 193, 1076
performance. While prior work has described the utility of the 1082.
Griess assay, chemiluminescence analyzers, and electrochemical (18) Wink, D. A.; Mitchell, J. B. Free Radical Biol. Med. 1998, 25,
434456.
sensors for measuring NO in solution, performance discrep- (19) Moncada, S.; Palmer, R. M.; Higgs, E. A. Pharmacol. Rev. 1991,
ancies as a function of sample type have been disregarded. The 43, 109142.
data reported here clearly demonstrate the signicant variations (20) Coneski, P. N.; Schoenfisch, M. H. Chem. Soc. Rev. 2012, 41,
between analysis technique and sample composition. As such, 37533758.
the caveats of the analytical method employed must be carefully (21) Griess, P. Philos. Trans. R. Soc. London 1864, 154, 667731.

1962 dx.doi.org/10.1021/ac303787p | Anal. Chem. 2013, 85, 19571963


Analytical Chemistry Article

(22) Sun, J.; Zhang, X. J.; Broderick, M.; Fein, H. Sensors 2003, 3, (59) Riccio, D. A.; Nutz, S. T.; Schoenfisch, M. H. Anal. Chem. 2012,
276284. 84, 851856.
(23) Bryan, N. S.; Grisham, M. B. Free Radical Biol. Med. 2007, 43, (60) Zhang, X.; Cardosa, L.; Broderick, M.; Fein, H.; Davies, I. R.
645657. Electroanalysis 2000, 12, 425428.
(24) Bates, J. N. Neuroprotocols 1992, 1, 141149. (61) Hou, Y.; Wu, X.; Xie, W.; Braunschweiger, P. G.; Wang, P. G.
(25) Williams, R. J. P. Chem. Soc. Rev. 1996, 25, 7783. Tetrahedron Lett. 2001, 42, 825829.
(26) Moller, M. N.; Li, Q.; Lancaster, J. R.; Denicola, A. IUBMB Life (62) Malinski, T.; Bailey, F.; Zhang, Z. G.; Chopp, M. J. Cereb. Blood
2007, 59, 243248. Flow Metab. 1993, 13, 355358.
(27) Hall, C. N.; Garthwaite, J. Nitric Oxide 2009, 21, 92103. (63) Trevin, S.; Bedioui, F.; Devynck, F. Talanta 1996, 43, 303311.
(28) Tsikas, D. J. Chromatogr. B 2007, 851, 5170. (64) Bard, A. J.; Faulkner, L. R. Electrochemical Methods:
(29) Taha, Z. H. Talanta 2003, 61, 310. Fundamentals and Applications, 2nd ed.; John Wiley and Sons, Inc.:
(30) Allen, B. W.; Liu, J.; Piantadosi, C. A.; Lester, P.; Enrique, C. In Hoboken, NJ, 2001.
Methods in Enzymology; Academic Press: San Diego, CA, 2005; Vol. (65) Elsadek, B.; Kratz, F. J. Controlled Release 2012, 157, 428.
396, pp 6877. (66) Stamler, J. S.; Simon, D. I.; Osborne, J. A.; Mullins, M. E.; Jaraki,
(31) Bedioui, F.; Villeneuve, N. Electroanalysis 2003, 15, 518. O.; Michel, T.; Singel, D. J.; Loscalzo, J. Proc. Natl. Acad. Sci. U.S.A.
(32) Yao, D.; Vlessidis, A. G.; Evmiridis, N. P. Microchim. Acta 2004, 1992, 89, 444448.
147, 120. (67) Kim-Shapiro, D. B.; Schechter, A. N.; Gladwin, M. T.
(33) Giustarini, D.; Rossi, R.; Milzani, A.; Dalle-Donne, I. Nitric Arterioscler. Thromb. Vasc. Biol. 2006, 26, 697705.
Oxide 2008, 440, 361380. (68) Solc, M. Nature 1966, 209, 706706.
(34) Tsikas, D. Free Radical Res. 2005, 39, 797815. (69) Lewis, R. S.; Deen, W. M. Chem. Res. Toxicol. 1994, 7, 568574.
(35) Guevara, I.; Iwanejko, J.; Dembinska-Kiec, A.; Pankiewicz, J.; (70) Wisniewski, N.; Moussy, F.; Reichert, W. M. Fresenius J. Anal.
Wanat, A.; Anna, P.; Golabek, I.; Bartus, S.; Malczewska-Malec, M.; Chem. 2000, 366, 611621.
Szczudlik, A. Clin. Chim. Acta 1998, 274, 177188. (71) Wisniewski, N.; Klitzman, B.; Miller, B.; Reichert, W. M. J.
(36) In DailyMed; U.S. National Library of Medicine, National Biomed. Mater. Res. 2001, 57, 513521.
Institutes of Health: Bethesda, MD, 2006. (72) Frost, M.; Meyerhoff, M. E. Anal. Chem. 2006, 78, 73707377.
(37) Kark, R. M.; Lawrence, J. R.; Pollack, V. E.; Pirani, C. L.; (73) Voskerician, G.; Anderson, J. In Wiley Encyclopedia of Biomedical
Muehrcke, R. C.; Silva, H. A Primer of Urinalysis, 2nd ed.; Hoeber Engineering; John Wiley & Sons, Inc.: New York, 2006.
Medical Division, Harper & Row: New York, 1964. (74) Wightman, R. M. Science 2006, 311, 15701574.
(38) Arvidson, K.; Johansson, E. G. Eur. J. Oral Sci. 1985, 93, 467 (75) Oh, B. K.; Robbins, M. E.; Schoenfisch, M. H. Analyst 2006,
473. 131, 4854.
(39) Saavedra, J. E.; Southan, G. J.; Davies, K. M.; Lundell, A.;
Markou, C.; Hanson, S. R.; Adrie, C.; Hurford, W. E.; Zapol, W. M.;
Keefer, L. K. J. Med. Chem. 1996, 39, 43614365.
(40) Marxer, S. M.; Rothrock, A. R.; Nablo, B. J.; Robbins, M. E.;
Schoenfisch, M. H. Chem. Mater. 2003, 15, 41934199.
(41) Schmidt, H.; Kelm, M. In Methods in Nitric Oxide Research;
Feelisch, M., Stamler, J. S., Eds.; Wiley: New York, 1996; pp 491497.
(42) Shin, J. H.; Privett, B. J.; Kita, J. M.; Wightman, R. M.;
Schoenfisch, M. H. Anal. Chem. 2008, 80, 68506859.
(43) Shin, J. H.; Weinman, S. W.; Schoenfisch, M. H. Anal. Chem.
2005, 77, 34943501.
(44) Indika, P. N.; Bayachou, M. Anal. Bioanal. Chem. 2004, 379,
10551061.
(45) Tsikas, D.; Caidahl, K. J. Chromatogr. B 2005, 814, 19.
(46) Fox, J. B. Anal. Chem. 1979, 51, 14931502.
(47) Moshage, H.; Kok, B.; Huizenga, J. R.; Jansen, P. L. Clin. Chem.
1995, 41, 892896.
(48) Miranda, K. M.; Espey, M. G.; Wink, D. A. Nitric Oxide 2001, 5,
6271.
(49) Tsikas, D.; Gutzki, F.-M.; Rossa, S.; Bauer, H.; Neumann, C.;
Dockendorff, K.; Sandmann, J.; Frolich, J. C. Anal. Biochem. 1997, 244,
208220.
(50) Kitabatake, N.; Doi, E. J. Food Sci. 1982, 47, 12181221.
(51) Privett, B. J.; Shin, J. H.; Schoenfisch, M. H. Chem. Soc. Rev.
2010, 39, 19251935.
(52) Amatore, C.; Arbault, S.; Bouton, C.; Drapier, J.-C.; Ghandour,
H.; Koh, A. C. W. ChemBioChem 2008, 9, 14721480.
(53) Cha, W.; Tung, Y.-C.; Meyerhoff, M. E.; Takayama, S. Anal.
Chem. 2010, 82, 33003305.
(54) Davies, I. R.; Zhang, X.; Robert, K. P. In Methods in Enzymology;
Academic Press: San Diego, CA, 2008; Vol. 436, pp 6395.
(55) Lee, Y.; Yang, J.; Rudich, S. M.; Schreiner, R. J.; Meyerhoff, M.
E. Anal. Chem. 2004, 76, 545551.
(56) Patel, B. A.; Arundell, M.; Parker, K. H.; Yeoman, M. S.; OHare,
D. Anal. Chem. 2006, 78, 76437648.
(57) Amatore, C.; Arbault, S.; Bouton, C.; Coffi, K.; Drapier, J.-C.;
Ghandour, H.; Tong, Y. ChemBioChem 2006, 7, 653661.
(58) Amatore, C.; Arbault, S. p.; Koh, A. C. W. Anal. Chem. 2010, 82,
14111419.

1963 dx.doi.org/10.1021/ac303787p | Anal. Chem. 2013, 85, 19571963

You might also like