A typical schematic diagram of uv-vis instrument will have 5 important components which

consist of; light source, monochromator, sample and reference cuvette, detector, and signal processor
and readout (UV-Visible Absorption Spectroscopy, 2017). The straightforward instrument started from
where the ultraviolet (uv) and visible radiation act as light sources which provides a combination of
tungsten and hydrogen-deuterium lamps, covering the range from 200 to about 800 nm. From the
RSC (2009) notes, a beam of light from the light source will be separated based on its components
colour of wavelengths by monochromator.

As shown in figure above, monochromator is composed of prisms, slits and diffraction grating.
Diffraction grafting plays the same role as the prism, but more efficient. The separated beam reflected
from the grafting will be selected by the slit and further divided by the half-mirrored prism into two
beam. As for the sample, it will be held in an optical flat, transparent container called cell or cuvette.
One of the beam will passes through the sample solution, and the other one will passes through the
reference solution. The intensities of the beam pass through both cuvette will be measured by the
electronic detectors and compared. The intensity of the reference beam is defined as I0, while for the
sample beam is known as I. if I is found less than I0, it means that the sample has absorbed some of
the light.

The absorbance of the sample shows how it obeys the Beers Law, which state that the
absorbance is proportional to the concentration of the substance in the solution as a result of uv-vis
spectroscopy. It can be expressed by:
0
, = log10

Based on the equation from a series of sample absorbance, a graph of absorbance versus
concentration plotted should linear if it obeys the law. Also, the absorbance can be measured in range
from 0 to 1. At 0 absorbance, no light of the wavelength has been absorbed, hence the ratio is 1 and
the absorbance, A will be 0. This can happened vice versa when the absorbance of is 90%. (Jim C.,
2007)

UV-Visible Absorption Spectroscopy. (2017). Www2.chemistry.msu.edu. Retrieved 1 October 2017,


from https://www2.chemistry.msu.edu/faculty/reusch/virttxtjml/spectrpy/uv-vis/uvspec.htm

Ultraviolet-Visible Spectroscopy (UV). (2009). Royal Society of Chemistry. Retrieved from


https://edisciplinas.usp.br/pluginfile.php/1695367/mod_resource/content/1/UV-
Vis_Student%20resource%20pack_ENGLISH.pdf

Jim, C. (2007). Double beam UV-visible absorption spectrometer. Chemguide.co.uk. Retrieved 1


October 2017, from http://www.chemguide.co.uk/analysis/uvvisible/spectrometer.html#top

You might also like