You are on page 1of 9

Forensic Chemistry 3 (2017) 8189

Contents lists available at ScienceDirect

Forensic Chemistry
journal homepage: www.elsevier.com/locate/forc

Cryofocusing capillary microextraction of volatiles (Cryo-CMV) as a


novel headspace extraction device for the analysis of volatile organic
compounds and smokeless powders
Anamary Tarifa, Natasha M. Kreitals, Jerome Mulloor, Sigalit Gura, Jos R. Almirall
Department of Chemistry and Biochemistry and International Forensic Research Institute, Florida International University, 11200SW 8th Street, Miami, Florida 33126, United States

a r t i c l e i n f o a b s t r a c t

Article history: A novel field sampling device for fast, sensitive, dynamic headspace sampling of volatile organic com-
Received 16 December 2016 pounds (VOCs) and smokeless powders from air is described. The capillary microextraction of volatiles
Received in revised form 10 February 2017 (CMV) is an inexpensive, preconcentration technique that results in fast sampling times (as low as
Accepted 16 February 2017
2 min) with high efficiency and sensitivity. Analysis of the CMV with GCMS is achieved with a thermal
Available online 20 February 2017
desorption probe to introduce the CMV directly into the injection port of a GC. A cryofocusing device con-
sisting of a Peltier-cooled module that operates down to 10 C was employed to improve the extraction
Keywords:
efficiency of the CMV for sampling both VOCs and smokeless powders.
VOC
Smokeless powders
The method performance using the CMV for the analysis of 17 VOCs, including the BTEX compounds
Capillary microextraction of volatiles (CMV) (Benzene, Toluene, Ethylbenzene, and Xylenes) is demonstrated using the criteria established by the
Cryo-CMV Environmental Protection Agency (EPA) for sorbent tube sampling of toxic organic compounds in ambient
Thermal separation probe with gas air (EPA TO-17). Experimental results with known standards yield absolute mass detection limits ranging
chromatography-mass spectrometry from 2.0 to 4.6 ng for the target compounds, in a 2 L air sample. A 35100% improvement in the analytical
performance was achieved by sampling at 10 C, in comparison to 20 C, for most compounds. Using
cryofocusing conditions, precision of 853 %RSD was achieved for headspace extraction for the majority
of the compounds with recoveries of as much as 37% and low breakthrough. Furthermore, the headspace
calibration curves show a linearity of 0.951 or better for the target compounds, suggesting the quantita-
tion capabilities of the CMV device.
The CMV method performance was further demonstrated for the detection of markers above the head-
space of smokeless powder. Experimental results with known standards, yield method detection limits of
2.0 ng and 1.3 ng for nitroglycerine (NG) and diphenylamine (DPA), respectively, in a 2 L air sampling vol-
ume followed by simultaneous SIM mode GCMS analysis. Using cryofocusing conditions for the head-
space extraction of Hodgdon smokeless powder, a 2938% improvement in the analytical performance
of NG was achieved with 18 %RSD precision, by sampling at 2.5 C to 10 C, in comparison to room tem-
perature at 20 C.
2017 Elsevier B.V. All rights reserved.

1. Introduction example. The list of VOCs thought to be toxic organic compounds


is extensive and has been compiled by the Environmental Protec-
The detection of volatile organic compounds (VOCs) present in tion Agency (EPA) to indicate chemicals of concern that may be
ambient air is of great concern to air quality monitoring programs detectable in areas where air pollution is present, such as indus-
because of the potential hazards to human health and the environ- trial sites [3]. While the severity of the hazard posed by different
ment [1,2], but also of interest to forensic scientists interested in contaminants varies, many VOCs present in ambient air have the
detecting VOCs associated with smokeless powders, gunshot resi- potential to act as mutagens and carcinogens [1,2]. Therefore, the
dues (GSR), explosives, drugs, and ignitable liquid residues for detection and quantitation of VOCs is important to managing and
mitigating health impacts from toxic compounds in ambient air.
In an effort to address this issue, the EPA has published the Com-
Corresponding author.
pendium of Methods for Toxic Organic Air Pollutants since 1984
E-mail addresses: atarifa@fiu.edu (A. Tarifa), nkreital@fiu.edu (N.M. Kreitals),
(TO-1TO-17). These are a series of reports describing the most
jmull008@fiu.edu (J. Mulloor), sgura@fiu.edu (S. Gura), almirall@fiu.edu (J.R.
Almirall).

http://dx.doi.org/10.1016/j.forc.2017.02.007
2468-1709/ 2017 Elsevier B.V. All rights reserved.
82 A. Tarifa et al. / Forensic Chemistry 3 (2017) 8189

current methods and guidelines to be followed for the monitoring No. 61/779,690) [9] (see Fig. 1). PDMS is a non-polar polymer that
of VOCs in ambient air or polluted environments. provides a hydrophobic coating over the glass filter, thus improv-
The analysis of VOCs in ambient air is currently performed with ing extraction of VOCs in humid conditions [7,8].
sorbent tubes following the guidelines from the EPA method TO- In addition, the use of PDMS has been demonstrated to be more
17. The commercially available sorbent tubes consist of a thin effective than Tenax TA for the analysis of large injection volumes
cylinder that can be made out of glass or stainless steel and packed and retention efficiency with good precision [7]. With a total sur-
with sorbent material. The commonly used sorbent materials face area of 0.05 m2 and a phase volume of 100 mm3, the CMV
include: several variations of Tenax, Carbotrap, and Car- device offers 5000 times greater absorption capacity compared to
bopackTM, with the possibility of using multiple sorbents in a single SPME [9]. It also facilitates dynamic sampling, thus increasing sam-
tube [4]. The sorbent material selected depends largely on the tar- pling efficiency and decreasing sampling time.
get compounds properties, specifically the volatility or vapor pres- In order to further improve the trapping efficiency of the target
sure of the molecules of interest. Each sorbent material is classified compounds vapor, a cryofocusing CMV technique was developed
according to its strength, which is described as the affinity of the by cooling the device down to 10 C during the dynamic extrac-
compounds to the sorbent. A strong sorbent will allow greater tion, using a newly designed single-stage Peltier-cooled module
sampling volumes for all or most of the targeted VOCs. The with a temperature controller (see Fig. 2, patent pending). Similar
strength of a sorbent tube is related to the surface area of the sor- thermoelectric devices have been reported in analytical applications
bent material. In general, a weak sorbent has a surface area less for the preconcentration of analytes using cold fiber SPME [13], cool-
than 50 m2/g (e.g. Tenax TA), a medium sorbent has a surface area ing assisted liquid extraction, and thin film microextraction [14].
in the range of 100500 m2/g, and a strong sorbent has a surface The aim of this work was to demonstrate the utility of CMV
area around 1000 m2/g [3]. In general, stronger sorbents are used samplers, including while being coupled to the novel cryofocusing
for highly volatile compounds. Some of the limitations observed device, for the enhancement in extraction capabilities of VOCs in
for the analysis of VOCs with sorbent tubes include: long head- indoor air, with its potential in forensic chemistry applications
space extraction times (1 h) with low flow rates, and the use of demonstrated herein with smokeless powder. The performance
expensive thermal desorption units coupled with a transfer line of Cryo-CMV for VOCs was evaluated following the reported guide-
to the gas chromatography-mass spectrometry (GCMS) instru- lines of the EPA Compendium Method TO-17 based on four perfor-
ment, which can result in poor recoveries [57]. mance criteria: 1) a method detection limit of 0.5 ppbv or less, 2)
This study describes for the first time the use of capillary an analytical precision of 20%, 3) a precision for distributed volume
microextraction of volatiles (CMV), an innovative headspace pair of 25% or less and 4) an audit accuracy within 30% for concen-
extraction technique [8], while being coupled to a cryofocusing trations expected in contaminated ambient air (0.525 ppbv). In
device (a Peltier-cooled module, Cryo), to enhance the sampling addition to these 4 criteria, the EPA TO-17 mentions that the break-
and pre concentration capabilities of the CMV. The CMV has through of the sorbent tubes should be less than 5% during the
already demonstrated to result in improved sensitivity and selec- sampling time.
tivity for the extraction of volatiles in the headspace of explosives
compared to SPME [9], for the detection of gunshot residues on 2. Materials and methods
swab samples from the hands of shooters [10], and for marijuana
markers from air and small plastic bags [11]. Herein, the enhanced 2.1. Materials
performance of the Cryo-CMV is demonstrated for air monitoring
of VOCs with its potential in forensic applications. Standard compounds for 17 hazardous air contaminants
The CMV consists of an open-ended 2 cm glass capillary tube, (Table 1) were sourced from Fisher Scientific (Pittsburgh, PA),
packed with approximately seven 2 cm  2 mm glass filter strips, Sigma-Aldrich (St. Louis, MO), TCI America (Tokyo, Japan), Acros
each coated with vinyl-terminated polydimethylsiloxane (New Jersey, USA), and Aldrich (Milwaukee, WI). Purities of
(vt-PDMS) [8,12], at overall average weight of 0.23 g (USPTO patent stock standards exceeded 97.0% with the exception of

Fig. 1. The dimensions of a CMV device (left) with an inner diameter of 2 mm and length of 2 cm, and (right) a CMV device inserted into the thermal separation probe for
injection into the GCMS.
A. Tarifa et al. / Forensic Chemistry 3 (2017) 8189 83

Fig. 2. Prototype single-stage Peltier-cooled cryofocusing device with a programmable temperature controller (top) showing the CMV module connected to the sampling
tube and a pump, and inserted into a Peltier-cooled aluminum block cooled down to 10 C (bottom).

Table 1
Compound list in order of elution time (tR) for the headspace extraction at 20 C, the quantifier and qualifier ions, method limits of detection (MDL), method quantitation limits
(MQL), precision (as % RSD, n = 3) for both direct spike onto a CMV and the headspace extraction using CMV, and recoveries (as %) for the headspace extraction using the CMV.

Compounds tR min Quantifier ion Qualifier Direct spike on CMV %RSDa Headspace %RSDa %Recoverya
ions extraction with CMV
Q1 Q2 MDL ng MQL ng MDL ng MQL ng
Methylene chloride 2.61 84 49 86 4.1 14 25 2.7b
Benzene 3.44 78 77 51 4.6 15 19 37 123 13 11
Pyridine 4.35 79 52 78 3.3 11 13 10b
Toluene 4.54 91 92 65 2.9 10 23 17 55 25 18
Furfural 5.31 96 95 39 3.7 12 15 74 247 30 8.0
Ethylbenzene 5.67 91 106 77 2.0 6.6 11 33 110 7 26
m-Xylene/p-Xylene 5.77 91 106 77 3.0 10 15 23 75 15 18
o-Xylene 6.05 91 106 105 2.9 10 12 24 81 14 17
Benzaldehyde 6.87 105 106 77 2.5 8.4 12 52 173 20 16
Phenol 6.92 94 66 65 3.1 10 15 0.6b
Benzonitrile 7.09 103 76 50 2.6 8.8 14 53 177 16 11
1,2,4-Trimethylbenzene 7.19 105 120 91 2.4 8.1 12 47 155 7 14
1,2,3-Trimethylbenzene 7.47 105 120 91 2.6 8.7 14 88 292 4 12
Acetophenone 7.88 105 77 120 2.8 9.3 11 83 276 29 4.5
Nonanal 8.15 57 67 81 5.6 19 9 95 315 33 1.2
Naphthalene 8.88 128 127 102 2.5 8.4 12 56 186 15 11
a
Precision (%RSD) from the calibration curve at 15 ng (direct spike) and 200 ng (headspace), %Recovery from headspace extraction at 1 min equilibrium of 200 ng inside a
quart can.
b
%Recoveries were calculated for these compounds at 500 ng.

1,2,3-trimethylbenzene (90.0%), benzonitrile (95.0%), and nonanal (Fisher Scientific, Pittsburgh, PA). The standard mixture solution
(95.0%). A 300 ng mL 1 mixture of the standard compounds was was further diluted to concentrations ranging from 1.0 to
prepared in methanol and stored in a 10 mL amber glass vial 250 ng mL 1.
84 A. Tarifa et al. / Forensic Chemistry 3 (2017) 8189

Standard compounds for smokeless powders, nitroglycerin (NG) TSP with the CMV in the injection port of the GC system is shown
at 1 mg mL 1 acetonitrile and diphenylamine (DPA) at 5 mg mL 1 in Fig. 3.
methanol, were sourced from Sigma-Aldrich (St. Louis, MO). Both VOCs analysis: A 29.17 m  0.25 mm  0.25 mm DB-5 ms Ultra
standards were diluted in methanol to 2100 ng mL 1 for NG and Inert column (Agilent Technologies, Santa Clara, CA) was used for
0.525 ng mL 1 for DPA. For each solution, 1 ml was spiked directly chromatographic separation. The GC oven ramp temperature
on the CMV device for the generation of GCMS calibration curves. started at 35 C with a 1 min hold. The oven was then ramped to
The double-base classification smokeless powder, BLC-2, was 120 C at 15 C/min, to 220 C at 30 C /min with a 1.5 min hold,
sourced from Hodgdon (KS, USA). and subsequently to a final temperature of 280 C at 30 C /min
with a 1 min hold. The total time for the chromatographic separa-
2.2. Fabrication of the cryofocusing device tion was 14.50 min. The injector temperature was set at 180 C in
split mode (split ratio 5:1) with a column flow of 1.2 mL/min. The
A newly designed 5 V single-stage Peltier-cooled module was EI source was kept at 230 C, the transfer line to the mass spec-
attached to a heat sink assembly on its hot surface (Adafruit, trometer was set to 280 C and the quadrupoles were maintained
New York, NY), and to an aluminum block configured with a tem- at 150 C. The scan mass range was set at 45300 amu. The resolu-
perature controller and thermocouple probe on its cold surface. tion of the mass analyzer is 0.1 amu. A list of the target compounds
The latter was attached to an orifice large enough to fit the CMV with their respective retention times and ion peaks is presented in
(see Fig. 2). Two small vacuum pumps capable of drawing 0.2L/ Table 1 and used to confirm the presence of the target compounds
min for VOCs (Bailey Nurture III) or 1 L/min for smokeless powders in the field samples.
(Escort Elf Air Sampling Pump, Zefon International Inc, Ocala, FL) Smokeless powder analysis: A 5.8 m  0.25 mm  0.25 mm DB-
were used to sample the air through the CMV devices. A commer- 5 ms Ultra Inert column (Agilent Technologies, Santa Clara, CA)
cially available thermal separation probe (Agilent, Santa Clara, CA) was used for chromatographic separation. The GC oven ramp tem-
was used to insert the CMV into the inlet port (Fig. 3) of the GCMS perature started at 40 C with a 0.5 min hold, then to 240 C at
(Agilent, Santa Clara, CA). 15 C/min and to 280 C at 30 C/min with a 1 min hold. The total
time for the chromatographic separation was 21.16 min. The injec-
tor temperature was set at 180 C in split mode (split ratio 5:1)
2.3. Instrumentation with a column flow of 1.2 mL/min. The EI source was kept at
230 C, the transfer line to the mass spectrometer was set to
All analyses were conducted using a 7890A gas chromatograph 280 C and the quadrupoles were maintained at 150 C. The total
coupled to a 5975C inert XL MSD with a triple axis detector (Agi- ion chromatogram (TIC) scan mass range was set at 45500 amu
lent Technologies, Santa Clara, CA). The CMV was introduced and at selected ion monitoring (SIM) simultaneously. Applying this
directly into the GC injection port for thermal desorption using method respective retention times of 6.1 min. and 8.1 min. were
an Agilent Thermal Separation Probe (TSP) with a Sky 4 mm ID obtained for NG (46, 76 m/z) and DPA (169 m/z) and the SIM
single taper inlet liner (Restek, Bellefonte, PA). A schematic of the response was used for quantification.
The instrument was tuned before each experiment using the
autotune feature as recommended by the manufacturer.

2.4. Method development

Prior to sampling, conditioning of the CMV was performed in an


oven at 250 C for 30 min. The CMV was then desorbed as a blank
sample in the GCMS. Liquid standards containing a known mass
of the analytes were spiked (1 lL) directly onto the CMV and ther-
mally desorbed in the GC inlet (see Fig. 3). Thermal desorption of
the CMV devices was performed inside the GC inlet preheated at
180 C and through the entire runtime. For all cryofocusing mea-
surements, the CMV device was suspended 2 min inside the mod-
ule, to cool down prior to sampling, as was confirmed by the
thermocouple probe.
The headspace extractions were performed using a portable
vacuum pump connected to a flow meter with a sampling rate of
0.2 L/min for VOCs or 1 L/min for smokeless powder applications.
For VOCs 1 lL standard containing a known amount of the target
analyte(s) was spiked onto a Kimwipe (Kimberly-Clark Global Sales
LLC, Roswell, GA) that was placed inside a quart can (0.95 L) (All-
American Containers, Miami, FL, USA) and sealed. For smokeless
powder, 100 mg of powder was placed inside a quart can and
sealed. All cans were previously baked to 250 C in an oven over
three days prior to use, to remove any volatile contaminants from
the manufacturing process. Holes were fashioned on the can lids
and sealed using a rubber septum (Capitol Scientific Inc. Austin,
TX) to facilitate sampling through the holes. The extraction results
are reported in total mass extracted (ng), with values for the head-
space calibration curve reported as the calculated amount of mass
extracted.
Fig. 3. Schematic of the CMV device inserted in the injection port of a GC using the The evaluation of CMV devices for the detection of VOCs using
thermal separation probe. the EPA compendium TO-17 method was performed using the stan-
A. Tarifa et al. / Forensic Chemistry 3 (2017) 8189 85

dard solutions with the following optimized parameters: 1 min correspond with the highest concentration of compounds that
equilibrium time, 10 min extraction time, 0.2 L/min sampling flow are commonly detected in polluted air [3]. The linearity of the cal-
and total sampling volume of 2 L at a room temperature of 20 C. ibration curves was determined by plotting concentrations with a
The headspace calibration curves were also constructed using those response of a signal-to-noise ratio (SNR) of 10 or larger. The linear-
parameters. Optimization of the method using the cryofocusing ity obtained for the calibration curves were greater than 0.958,
device at 10 C and at room temperature (20 C) was tested for 0.969, and 0.951 for direct liquid injection, direct spike on CMV
30 min equilibrium time at different extraction times (15 s, 30 s, and headspace extraction, respectively.
1, 1.5, 2.5, 5, 10, 15, 20, 25, 30 min) to determine recovery and vol- The sensitivity of the method was tested by determining the
ume extraction capability at a sampling flow of 0.2 L/min. The method detection limits (MDL) and the method quantitation limits
experiments showed that 10 min extraction was optimum for most (MQL) for all analytes by direct spike on CMV and by headspace
VOCs. Recovery experiments were performed at room temperature extraction with the CMV device (Table 1). The reported MDL and
(20 C) and using the cryofocusing device ( 10 C) applying the MQL in Table 1 from the calibration curves, represent the minimal
same sampling and pumping parameters for comparison. sample concentration that can be spiked on a can for the signal-to-
The evaluation of the cryofocusing for the detection of smoke- noise ratio (SNR) of the response to be at least 3. Calibration curves
less powders was performed using previously reported optimized for headspace extraction of methylene chloride, pyridine, and phe-
parameters for headspace sampling [9,10]: 24 h equilibrium time nol were not feasible due to their low detection. These results can
at ambient conditions, 2 min extraction time and 1 L/min sampling be attributed to different physicochemical properties of each com-
flow for a total sampling volume of 2 L air. Optimization of the pound [2931]. Methylene chloride is the most volatile compound
method using the cryofocusing device was performed at different (vapor pressure = 414 Torr at 25 C) in the mixture, which lead to a
extraction temperatures, 10 C, 2.5 C and 20 C for enhanced tendency to incur breakthrough. Pyridine is the most polar com-
pre-concentration, all at three replicates. pound (log P = 0.65 and m = 2.2 Debye) in the mixture, which may
Equivalent blank samples were analyzed, through all method have led to an incompatibility in its physical interaction with the
development stages and headspace samplings, both for VOC and hydrophobic PDMS adsorption phase of the CMV sampler. Overall,
smokeless powder applications. Blank samples were prepared fol- an extremely low extraction recovery of only 0.6% was obtained for
lowing the exact parameters and sample treatments while exclud- phenol, which may be attributed to a combination of low volatility
ing target compounds. (vapor pressure = 0.325 Torr at 25 C) and the polarity of the com-
pound (m = 1.2 Debye) when sampling the headspace. In a separate
experiment a spiked amount of 500 ng yield extracted amounts of
2.5. Field sampling
13 ng, 50 ng, 3 ng for methylene chloride, pyridine and phenol,
respectively.
To demonstrate the utility of air monitoring in the field, tripli-
Detection limits for each compound were confirmed using two
cate samples were collected in a chemical laboratory, a lecture hall
methods. In the first method analysis of 10 replicates of a methanol
and a nail and hair salon. CMVs were conditioned and analyzed
blank sample was performed. The standard deviation of the inte-
prior to field sampling to ensure no previous contamination or car-
grated signal (or noise) at the known analyte retention times was
ryover. All samples were extracted for 10 min at a flow rate of
multiplied by 3. The second method (described in EPA TO-17) con-
0.2 L/min, at 20 C. The GCMS chromatographic peaks of the
sisted of conducting an analysis on 7 replicates at a concentration
detected compounds in the samples were first identified using
close to the expected MDL. The concentrations selected were 1 ng
the NIST mass spectral reference library and confirmed by compar-
and 5 ng. The standard deviation of the replicates was multiplied
ing the mass spectra of the individual standard compounds, under
by 3.14 (Students t value for confidence interval of 99%). The mass
the same instrumental conditions.
detected for headspace MDL and MQL were calculated using the
regression line results from the direct spike calibration curves.
3. Results and discussion The percent recovery for headspace extraction using a 200 ng stan-
dard ranged from 2.726%, with the exception of phenol (0.6%) and
3.1. CMV and Cryo-CMV for VOCs sampling nonanal (1.2%).
Table 2 summarizes the results obtained for the target com-
3.1.1. Method validation pounds following the additional EPA TO-17 performance criteria:
A rapid GCMS temperature program (14.5 min) was developed analytical precision, distributed volume pair, accuracy and break-
for the detection of 17 volatile organic compounds commonly through. The analytical precision calculated demonstrates lower
found in the air of polluted environments (Table 1) that represent performance for acetophenone and nonanal. However, all other
a wide range of boiling points (40.0 C217.9 C) [15]. The reten- compounds have an analytical precision of 20% or less. The dis-
tion time for each compound was determined by injecting 1 mL of tributed volume pair precision is calculated by performing several
a 20 ppm of the standard into the GCMS. Chromatographic sepa- measurements at different volumes. The amount (ng) of VOCs
rations were achieved for all the compounds except for m-xylene extracted at different sampling volumes and sampling flow
and p-xylene therefore, a combined quantitation was used for (extraction time constant) should result in a linear response. For
these compounds. this purpose, two different sampling volumes (2 L and 3 L) were
The parameters and performance of the developed method as evaluated. The distributed volume pair precision obtained ranged
an extraction technique for air sampling was established by follow- from 136% for all compounds. Factors that can affect the precision
ing the criteria specified in the EPA TO-17 [3]. The criteria include are artifact formation, and breakthrough of the compounds [3]. The
method detection limit (MDL) of 0.5 ppbv (35 ng), analytical compounds with distributed volume pairs higher than the
precision of replicate measurements within 20%, precision for the expected (25%) were benzonitrile and acetophenone. The % audit
distributed volume pair of 25% or less, and an audit accuracy of accuracy was less than the expected percentage value (30%) for
30% or better for the expected concentration range 0.525 ppbv the compounds. Calculations for % audit accuracy take into account
(5300 ng). the % recovery for each compound to properly represent the
Calibration curves for direct liquid injections into the GC, direct extraction efficiency of the CMV. All compounds were found to
injection on the CMV, and headspace extraction were developed meet the four method criteria specified in the EPA method TO-17
with the standard mixture across a range of 1 ppm-300 ppm to except benzonitrile, nonanal, and acetophenone.
86 A. Tarifa et al. / Forensic Chemistry 3 (2017) 8189

Table 2
Compounds listed in order of elution time following the EPA TO-17 criteria for the headspace extraction with CMV: analytical precision (n = 3), distributed volume pair (as %),
audit accuracy (as %) and calculated breakthrough (as %). In addition recoveries (as %) for headspace extraction above 30 ng target compounds, at sampling flow of 0.2 L/min and
10 min extraction time at room temperature (RT) and 10 C using the cryofocusing device.

Compounds Analytical precision % Distributed volume pair % Accy % Breakthrough % Recovery %RT Recovery% 10 C
Methylene chloride 16 21
Benzene 15 8.9 46 20
Pyridine
Toluene 20 12 8.2 35 5 37
Furfural 4 9.2
Ethylbenzene 3 16 7.4 27 20 25
m-Xylene/p-Xylene 18 17 8.2 22 24 29
o-Xylene 14 11 8.3 21 18 36
Benzaldehyde 1 11 8.4 10 2 8
Phenol
Benzonitrile 2 36 8.9 7 1 3
1,2,4-Trimethylbenzene 3 6 8.6 12 7 19
1,2,3-Trimethylbenzene 3 10 8.8 15 3 8
Acetophenone 42 34 9.5 28 2
Nonanal 60 1.3 9.9 19 12 5
Naphthalene 18 10 8.9 5 1 1

Analytical precision in percent, precision for the distributed volume pair and audit accuracy for a 200 ng standard solution. Breakthrough at 0.2 L/min and 10 min extraction
time at ambient temperature, 1 min equilibrium time for a 30 ng standard solution. Percent recoveries at ambient temperature and 10 C, 30 min equilibrium time for a
30 ng standard solution and 2 L extraction volume. Values not reported are from concentrations below the method detection limit.

Breakthrough experiments were conducted applying 0.2 L/min the CMV while recovering 14% and up to 67% for the BTEX com-
pumping flow through two CMV devices connected in series, to pounds, as well as for most detected compounds except naph-
extract the headspace above 30 ng of the target compounds main- thalene (1%). At 5 and 6 L sampled air the percent recovery starts
tained at 20 C, and the two CMVs were analyzed separately. As to plateau for some compounds and show a gradual decrease to
shown in Table 2, the breakthrough calculated as a percentage 7% up to 44% for the latest. However, at an extraction temperature
from the second CMV, under the present sampling conditions, of 20 C, sampling volume can be drawn through the CMV up to 2 L
was higher than that specified in the EPA TO-17 method (5%). volume only, followed by a gradual decrease for BTEX down to 6 L,
The only compounds with similar breakthroughs were benzonitrile and complete loss of benzene is observed at 3 L due to break-
(7%) and naphthalene (5%). Nonetheless, the compounds were through. An improvement between 35100% in recoveries resulted
detected at the expected detection limits. in the use of the novel Cryo-CMV device with lower temperature
In order to improve breakthrough and recoveries of the target extraction for all compounds, notably for benzene, allowing its
compounds, a thermoelectric cooler was constructed to accommo- detection in the tested experimental concentration even at high
date the CMV devices and perform cryofocusing of the extracted sampling volumes. Generally, the decrease in recoveries with the
headspace. Breakthrough experiments were conducted using the increased extracted volume of air, while a continuous vapor source
cryofocusing device operated at 10 C and at room temperature is sampled, is attributed to saturation of the sampler adsorption
(20 C) at 0.2 L/min. Compared to room temperature results, the phase leading to breakthrough. The experimental setup developed
calculated breakthrough at 10 C were generally lower for most herein, is based on a finite amount of compound generated inside a
compounds. In a different experiment, a higher flow rate of (1 L/ semi-closed system, and up to 37% of BTEX compounds were sam-
min) was used and resulted in an increase in breakthrough and pled in a short time scale, followed by a moderate loss with
reduced recoveries, for some compounds. However, a sampling increased extracted air. These results demonstrate the efficient
flow of 1 L/min at 10 C favors the extraction of the compounds irreversible trapping of the CMV adsorption phase for the volatile
eluting after 6.8 min (Table 1). compounds, at 20 C and all the more at -10 C extraction
Recovery experiments (Table 2) were conducted using the cry- conditions.
ofocusing device operated at 10 C and at room temperature
(20 C) at 0.2 L/min and 30 min equilibrium time. Compared to 3.1.2. Dynamic headspace sampling of indoor air with CMV
room temperature results, well defined higher recoveries were As a proof of concept, CMV sampling was applied for collection
measured for the Cryo-CMV extraction for all volatile compounds, of indoor air VOCs using the optimized parameters: sampling flow
excluding naphthalene and nonanal. A 25%640% enhanced recov- rate (0.2 L/min) and 10 min extraction time. Three replicates of
eries were achieved with 10 C extraction temperature over indoor air samples were extracted from different rooms including
20 C, for sampling the headspace generated inside a quart can a laboratory (918 ft2), a classroom (1694 ft2), and a nail and hair
containing 30 ng of target compounds, indicating on the significant salon (1053 ft2). For air extraction in the laboratory the portable
higher detection sensitivities attributed to the novel cryofocusing pump and the CMV were placed on top of a table in the center of
add-on CMV device (Table 2). Notably, the detection of the volatile the room, with the CMV positioned in an upward direction. The
compounds benzene and acetophenone, became feasible in this air extraction in the classroom was performed in one of the corners
experimental concentration with cryo extraction, although for pyr- of the room. The air extraction in the hair and nail salon was per-
idine, phenol and furfural higher concentrations are required. For formed towards the middle of the room between the hair and nail
the less volatile compounds, naphthalene and nonanal, the Cryo- sections. All replicates were performed in the same location for
CMV advantage was less apparent. each room. The extraction was performed at room temperature
Fig. 4 shows the response curves of % recovery with different (21.0 C).
extracted volumes (0.05, 0.1, 0.2, 0.3, 0.5, 1, 2, 3, 4, 5, 6 L) sampled An example of the chromatogram obtained from the hair and
at 10 C (fig. 4 top) in comparison to the response curves at 20 C nail salon is shown in Fig. 5. At all three indoor air spaces, the com-
(fig. 4 bottom). The figure shows that, at the extraction tempera- pounds identified were detected at least in two of the three repli-
ture of 10 C, up to 4 L of sampling volume can be drawn through cates obtained from each room. From the compounds calibrated in
A. Tarifa et al. / Forensic Chemistry 3 (2017) 8189 87

70% -10 C CMV extraction


temperature
60%
50%

% Recovery
40% Benzene
Toluene
30%
Ethylbenzene
20%
m-Xylene/p-Xylene
10%
o-Xylene
0%
0 1 2 3 4 5 6 7
Extracted volume, L

70% 20C CMV extraction


temperature
60%
50%
% Recovery

Benzene
40%
Toluene
30%
Ethylbenzene
20% m-Xylene/p-Xylene
10% o-Xylene
0%
0 1 2 3 4 5 6 7
Extracted volume, L

Fig. 4. Response curves for the recovery (as %) at 30 min equilibrium time and different extracted volumes at 10 C (top) and 20 C (bottom) for the BTEX compounds
showing improvements of up to 100% in the recovery of the compounds of interest.

this study, toluene, m, p-xylenes and o-xylenes were detected 1-chloro-4-(trifluoromethyl)benzene is used as a solvent in com-
below the MQL in the laboratory and the classroom. In the hair mercial surface finishes [21]. Limonene is a common monoterpene,
salon, the amount of toluene detected was only 3.25  10 3 mg/ detected in indoor air and is found in products such as, citrus fruits,
m3 (overall 6.5 ng). According to the Occupational Safety & Health air refresheners, household cleaning agents, hair dye, and waxes
Administration (OSHA), the permissible exposure limit (TWA) for [2227].
toluene is 37 mg/m3, which is orders of magnitude higher than
the amount detected in the nail salon [16]. 3.2. CMV and Cryo-CMV for smokeless powders sampling
The presence of other compounds detected in the samples
including hexane, camphor, tridecane, tetradecane, pentadecane, Calibration curves for direct injection on the CMV, across the
and diethyl phthalate were confirmed by obtaining the retention range of 2 ppm100 ppm for nitroglycerine (NG) and 0.5 ppm
time and mass spectrum of a 10 ppm standard solution. Quantita- 25 ppm for diphenylamine (DPA), were generated to correspond
tion of these compounds was not conducted. Other compounds with the concentration range detected above the smokeless pow-
such as acetone, ethyl acetate, ethyl methacrylate, butyl ester der samples. The linearity obtained for the calibration curves were
acetic acid, 1-chloro-4-(trifluoromethyl)benzene, and limonene greater than 0.976 and 0.998 for NG and DPA, respectively. The
were identified using the mass spectra by comparison with the percent recovery for headspace extraction, generated by 1000 ng
NIST library. Presence of acetone in nail salon is expected since this (spiked in 10 mL methanol) standard solutions in a quart can after
is the main ingredient of nail polish remover. Ethyl acetate, ethyl 10 min. incubation time at ambient conditions, was determined
methacrylate, and butyl ester acetic acid are also ingredients of nail to be 1% or less for 2 L extraction air at a flow rate of 1 L/min,
polish, acrylic nails, and acrylic removers [1720]. In addition, which corresponds with previously reported and expected results

4.5E+05 1 4 5 13
10 11 CMV blank
3.0E+05 7 9
8 12 Hair and nail
salon
1.5E+05
2 3 6
0.0E+00
2 3 4 5 6 7 8 9 10 11 12
Retention time, min

Fig. 5. Chromatogram obtained by CMV-GCMS showing the compounds identified in the hair and nail salon. Peak identities: (1) acetone; (2) ethyl acetate; (3) toluene; (4)
ethyl methacrylate; (5) butyl ester acetic acid; (6) 1-chloro-4-(trifluoromethyl)benzene; (7) benzaldehyde; (8) limonene; (9) camphor; (10) tridecane; (11) tetradecane; (12)
pentadecane; (13) diethyl phthalate.
88 A. Tarifa et al. / Forensic Chemistry 3 (2017) 8189

[9]. These two compounds, known to be detected in the headspace and NG (2 L). Therefore, optimization of the extraction parameters
of smokeless powder samples [9], were also the only organic com- is crucial to obtain higher recoveries.
ponents detected in the headspace of swabs collected from the The well-known marker of smokeless powder, DPA, was
hands of 50 gun shooters [10]. detected in all the smokeless powder samples at the different
To reach enhanced sensitivity in the detection of smokeless extraction temperatures tested. The amount of DPA detected at
powders and gunshot residues, the novel Cryo-CMV device devel- room temperature was 1.6 ng and the amount extracted at
oped in this research, was applied for the detection of NG and 2.5 C and 10 C was not significantly higher for this semi-
DPA vapors. Fig. 6 demonstrate the amounts of NG detected at volatile compound (vapor pressure 6.4 4 torr, 25 C). These results
three different extraction temperatures using the operated tem- indicate on the low availability of DPA vapors inside the cans for
perature Cryo-CMV at 10 C, 2.5 C and 20 C for comparison. extraction with the CMV and the short time available during
All measurements were performed for the headspace extraction extraction for the vapors to rebuild at the experimental setup
of a quart can incubated at ambient conditions for 18 h with applied.
100 mg double-base smokeless powders, sampled for 2 min at
1 L/min flow rate. Enhanced signals were measured for the cryofo-
cusing extractions. An average (3 replicates) of 38% and 29% rela- 4. Conclusions
tive increase in recoveries was measured at the 2.5 C and
10 C, respectively, compared to ambient temperature extraction. The method proposed facilitates a more rapid sampling and
For NG, 90 15.5 ng was extracted at 2.5 C and 84 12.5 ng at detection method for VOCs using a novel technique, CMV-GC
10 C, compared to 65 12 ng at ambient extraction for 2 min MS. The CMV extraction method has been shown to be a fast and
sampling time. Although a minor overlap in amount extracted sensitive technique for the headspace extraction of organic vola-
was observed between the ambient and both cooled extraction tiles present in the air. One of the major advantages of CMV devices
temperatures, improved recoveries were still measured for the over sorbent tubes for air monitoring purposes is the ability to con-
Cryo-CMV. The amounts detected are at least 30-fold higher than duct dynamic headspace sampling in less than 10 min, compared
the MDL. The higher recovery achieved for 2.5 C extraction over to 1 h for sorbent tubes. The CMV is also more cost effective, cost-
10 C, can be attributed to the reduced condensation, while still ing less than $1 to manufacture compared to ($100/each) for
keeping the CMV cooled during sampling. The condensation sorbent tubes. The low cost facilitates the disposable nature of
observed inside the CMV at 10 C, modify the performance of the CMV when multiple uses are not feasible. However, the CMV
the adsorption surface while sampling. In order to address conden- is reusable after desorption in the GC injection port with experi-
sation effects, using the same experimental setup, silica gel desic- mental results demonstrating a life span of the CMV used for more
cant were placed inside the quart can during incubation. Compared than 100 injections and headspace extractions, comparable to sor-
to ambient temperature, at 10 C a 62% higher recoveries were bent tubes such as Tenax TA [27]. Furthermore, the CMV can be
observed for NG, confirming the contribution of the cryo extraction directly introduced into the injection port of the GC using a low-
for enhanced recoveries. This apparent increase suggests that fur- cost thermal separation probe. The advantages of direct injection
ther improvements in the setup will lead to better results for cry- in the GC port include reduced sample loss and faster elution com-
ofocused extraction of smokeless powders. pared to other thermal desorption units that contain a transfer line.
Generally, higher sampling flows are preferred to enhance the The significance of this study was to demonstrate the utility of
detection sensitivity of semi-volatile compounds, while avoiding CMV devices for the extraction of volatile organic compounds in
the condensation effects generated from cooling the device at ambient air through method validation following a comparison
sub-ambient temperatures. Unlike the sampling flow rate (0.2 L/ to the EPA method TO-17. The results indicate that the CMV
min) used for the extraction of VOCs to yield higher recoveries method meets the criteria described in the EPA method, depending
and avoid breakthrough, the sensitivity of NG was lower at 0.2 L/ on the analyzed compound. The four criteria consisted of: method
min than when extraction was performed at a higher sampling detection limit (MDL) of 0.5 ppbv (5 ng), analytical precision
flow of 1 L/min, while sampling the same volume of air for VOCs within 20%, precision for the distributed volume pair of 25% or less,
and an audit accuracy of 30% or better for the expected concentra-
tion range 0.525 ppbv (5300 ng) of the TO-17 method outlined
by EPA [28]. All compounds were found to meet the four method
criteria specified in the EPA method TO-17 except benzonitrile,
nonanal, and acetophenone. The data also suggests that improving
the chemistry of the CMV sorbent coatings will enhance the molec-
ular weight range of compounds that can be extracted as well as
improve on the extraction performance, further enhancing the ana-
lytical figures of merit described in the EPA TO-17 method.
A cryofocusing device consisting of a single-stage Peltier cooled
module was designed and constructed for this study, to perform
headspace extraction of target compounds by cooling the CMV
device to 10 C during the extraction. The application of the
Cryo-CMV for VOCs sampling improved recoveries for all com-
pounds and remained similar for methylene chloride, toluene,
and naphthalene. Similarly, the application of the Cryo-CMV for
double-base smokeless powder sampling improved the extraction
recovery of NG, suggesting this additional application within
forensic chemistry as well as many other potential applications
such as fire debris analysis, volatiles in clandestine laboratories
Fig. 6. NG headspace extraction above 100 mg BLC-2 double base smokeless
and others. Further improvement of the cryofocusing device sam-
powder incubated for 18 h inside a quart can using Cryo-CMV device operated at pling setup to include insulating the sampling tube and to the
2.5 C and 10 C in comparison to 20 C (room temperature). CMV sampling geometry, both of which are currently pursued as
A. Tarifa et al. / Forensic Chemistry 3 (2017) 8189 89

lines of investigation in our group, may provide significant microextraction of volatiles (CMV) coupled to gas-chromatography mass
spectrometry (GC-MS), Forensic Chem. 2 (2016) 1.
improvements in the extraction capability to retain high volatility
[12] P. Guerra, H. Lai, J.R. Almirall, Analysis of the volatile chemical markers of
compounds and small molecules, while minimizing condensation explosives using novel solid phase microextraction coupled to ion mobility
effects of water when sampling at sub-zero temperatures. spectrometry, J. Sep. Sci. 31 (2008) 28912898.
The use of CMV at 20 C has already been demonstrated by Fan [13] S.H. Haddadi, J. Pawliszyn, Cold fiber solid-phase microextraction device based
on thermoelectric cooling of metal fiber, J. Chroma. A 1216 (2009) 27832788.
and Almirall for the analysis of explosives in air [9], by Tarifa and [14] A.R. Ghiasvand, S. Hajipour, N. Heidari, Cooling-assisted microextraction:
Almirall for the analysis of organic GSR from the hands of shooters comparison of techniques and applications, Trends Anal. Chem. 77 (2016) 54
[10] and by Wiebelhaus et al. for the analysis of marijuana markers 65.
[15] W.M. Haynes, CRC Handbook of Chemistry and Physics, in: W.M. Haynes (Ed.),
[11]. It is anticipated that cryofocusing CMV extraction operated at ninety-sixth ed., CRC Press/Taylor and Francis, 2016, Internet Version 2016.
sub-ambient temperatures will further improve these applications [16] OSHA (2007) (Sections 142.3 and 144.6), retrieved from <http://www.dir.
as well as other forensic chemistry applications. Additional foren- ca.gov/title8/ac1.pdf> (accessed October 12.10.15), Occupational Safety &
Health Administration.
sic applications for cryofocusing CMV headspace extraction under [17] Cover girl continuous color nail polish; MSDS [Online]; The Procter & Gamble
study in our research group include the detection of VOCs from fire Company: Hunt Valley, MD, October, 2010, <https://www.pg.com/
debris, either from collected samples in a can or directly at the productsafety/msds/beauty_care/color_cosmetics/Cover_Girl_Continuous_
Color_Nail_Polish_(95223489).pdf> (accessed 12.10.15).
scene of a fire, in order to detect VOCs associated with the presence [18] OPI nail lacquer; MSDS No. 062G [Online]; OPI Products Inc: Hollywood, CA,
of ignitable liquid residues. March 04, 2010, <http://www.dpcdsb.org/NR/rdonlyres/14A811B1-E388-
4651-BCCD-01D5B03DC68D/83018/OPINailLacquerMSDS.pdf> (accessed
12.10.15).
[19] Nail liquid; MSDS [Online]; Tammy Taylor Nails, Inc.: Santa Ana, CA, June 24,
References
2013, <http://tammytaylornails.com/Professional/ttown/msds/Nail_Liquid_
MSDS.pdf> (accessed 12.10.15).
[1] G.K.S. Wong, S.J. Ng, R.D. Webster, Quantitative analysis of atmospheric [20] Nail polish; MSDS [Online]; National Salon Supplies: Victoria, Australia, March
volatile organic pollutants by thermal desorption gas chromatography mass 26, 2010, <http://www.nationalsalonsupplies.com.au/upload/products/1MSDS
spectrometry, Anal. Methods 5 (2013) 219230, http://dx.doi.org/10.1039/ %20Nail%20Polish%205100TFF%20EU16.pdf> (accessed 12.10.15).
c2ay25982f. [21] Topcoat/clear polyurethane; MSDS [Online]; Rust-Oleum Corporation: Vernon
[2] B. Dou, J. Li, Y. Wang, H. Wang, C. Ma, Z. Hao, Adsorption and desorption Hills, IL, November 26, 2014, <http://www.rustoleum.com/MSDS/ENGLISH/
performance of benzene over hierarchically structured carbon-silica aerogel 286769.pdf> (accessed 12.10.15).
composites, J. Hazard. Mater. 196 (2011) 194200, http://dx.doi.org/10.1016/j. [22] T. Wainman, J. Zhang, C.J. Weschler, P.J. Lioy, Ozone and limonene in indoor
jhazmat.2011.09.019. air: a source of submicron particle exposure, Environ. Health Perps. 108 (2000)
[3] EPA, Compendium of Methods for the Detection of Toxic Organic Compounds 11391145.
in Ambient Air, Method TO-17, Center for Environmental Research [23] E. Vartiainen, M. Kulmala, T.M. Ruuskanen, R. Taipale, J. Rinne, H. Vehkamki,
Information, Office of Research and Development, Environmental Protection Formation and growth of indoor air aerosol particles as a result of D-limonene
Agency, U.S, 1999. oxidation, Atmos. Environ. 40 (2006) 78827892.
[4] E. Gallego, F.X. Roca1, J.F. Perales, X. Guardino, Comparative study of the [24] G. De Gennaro, L. de Gennaro, A. Mazzone, F. Porcelli, M. Tutino, Indoor air
adsorption performance of a multi-sorbent bed (Carbotrap, Carbopack X, quality in hair salons: screening of volatile organic compounds and indicators
Carboxen 569) and a Tenax TA adsorbent tube for the analysis of volatile based on health risk assessment, Atmos. Environ. 83 (2014) 119126.
organic compounds (VOCs), Talanta 81 (2010) 916924. [25] J. Madureira, I. Pacincia, J. Rufo, E. Ramos, H. Barros, J.P. Teixeira, E. de
[5] K.D. Oliver, J.R. Adams, E.H. Daughtrey Jr., Technique for monitoring toxic VOCs Oliveira, Fernandes, Indoor air quality in schools and its relationship with
in air: sorbent preconcentration, closed-cycle cooler cryofocusing, and GC/MS childrens respiratory symptoms, Atmos. Environ. 118 (2015) 145156.
analysis, Environ. Sci. Technol. 30 (1996) 19391945. [26] S. Youssefi, M.S. Waring, Indoor transient SOA formation from ozone + a-
[6] E.H. Daughtrey Jr, K.D. Oliver, J.R. Adams, K.G. Kronmiller, W.A. Lonneman, W. pinene reactions: impacts of air exchange and initial product concentrations,
A. McClenny, A comparison of sampling and analysis methods for low-ppbC and comparison to limonene ozonolysis, Atmos. Environ. 112 (2015) 106115.
levels of volatile organic compounds in ambient air, J. Environ. Monit. 3 (2001) [27] A. Tsigonia, A. Lagoudi, S. Chandrinou, A. Linos, N. Evlogias, E.C. Alexopoulos,
166174, http://dx.doi.org/10.1039/b007158g. Indoor air in beauty salons and occupational health exposure of
[7] G. Flores, M. Herraiz, G.P. Blanch, M.L. Ruiz del Castillo, Polydimethylsiloxane cosmetologists to chemical substances, Int. J. Environ. Res. Public Health 7
as a packing material in a programmed temperature vaporizer to introduce (2010) 314324, http://dx.doi.org/10.3390/ijerph7010314.
large-volume samples in capillary gas chromatography, J. Chromatogr. Sci. 45 [28] E. Woolfenden, Monitoring VOCs in air using sorbent tubes followed by
(2007) 3337. thermal desorption-capillary GC analysis: summary of data and practical
[8] J.R. Almirall, W. Fan, Capillary Microextraction of Volatiles. U.S. Patent guidelines, J. Air Waste Manag. Assoc. 47 (1997) 2036, http://dx.doi.org/
9,267,866 B2, February 23, 2016. 10.1080/10473289.1997.10464411.
[9] W. Fan, J.R. Almirall, High-efficiency headspace sampling of volatile organic [29] L.K. Schnackenberg, R.D. Beger, Whole-molecule calculation of log P on molar
compounds in explosives using capillary microextraction of volatiles (CMV) volume, hydrogen bonds, and simulated 13C NMR spectra, J. Chem. Inf. Model.
coupled to gas chromatography-mass spectrometry (GC-MS), Anal. Bioanal. 45 (2005) 360365.
Chem. 406 (2013) 21892195, http://dx.doi.org/10.1007/s00216-013-7410-3. [30] J. Ng Lee, C. Park, G.M. Whitesides, Solvent compatibility of poly
[10] A. Tarifa, J.R. Almirall, Fast detection and characterization of organic and (dimethylsiloxane)-based microfluidic devices, Anal. Chem. 75 (2003) 6544
inorganic gunshot residues on the hands of suspects by CMV-GC-MS and LIBS, 6554.
Sci. Justice 55 (2015) 168175, http://dx.doi.org/10.1016/j.scijus.2015.02.003. [31] W.M. Haynes, CRC handbook of chemistry and physics: a ready-reference book
[11] N. Wiebelhaus, D. Hamblin, N. Kreitals, J.R. Almirall, Differentiation of of chemical and physical data, CRC Press, Boca Raton, 2016.
marijuana headspace of plants and hemp products using capillary

You might also like