E / 11157 / 1 1/5
www.first-sensor.com
www.sensortechnics.com
Signal conditioning circuitry for the RCO and HDO pressure sensors
A LT1014DN
Z1 LT1004CZ-2.5
R1 J1
R1, R2, R4 10 kOhm SIP resistor array
R3 100 kOhm SIP resistor array
RO C1 10 F / 35 V Tantalum
V+ + RO 10 kOhm multiturn pot
Zero adjust
R1
A2 RS, RG select see text
-
+ R5 2 kOhm resistor
A1 R1 J2 R4 R4
+ - V
2.5 Z1 Ref Gain
adjust
R4 R3
R R
+ R2 R2 R2 S G
- C1
J3 Rgain
R2 J4 V R3 R3
B
GND
- R3
RCO A3 -
+ - + A4 Vout
Sensor +
Vin V- R5
J5
Figure 1: Signal conditioning circuitry for the RCO series pressure sensors
E / 11157 / 1 2/5
www.first-sensor.com
www.sensortechnics.com
Signal conditioning circuitry for the RCO and HDO pressure sensors
Vout R3
AV = = 2 1 +
Vin R gain
A LT1014DN
Z1 LT1004CZ-2.5
R1
R1, R2, R4 10 kOhm SIP resistor array
R3 100 kOhm SIP resistor array
RO C1 10 F / 35 V Tantalum
V+ + RO 10 kOhm multiturn pot
Zero adjust
= 15 V
R1 A2 RS 3.24 kOhm
-
+ RG 500 Ohm multiturn pot
A1 R1 R4 R4
+ - V
2.5 Z1 Ref Gain
adjust
R3
R R
+ R2 R2 R2 S G
- C1
Rgain
R2 V = 10 V R3 R3
B
GND
- R3
A3 -
RCO -
+ + A4 Vout
Sensor +
= 1...6 V
Vin
Figure 2: RCO signal conditioning circuitry for 1 V to 6 V output for 0 to 100 psi input pressure
E / 11157 / 1 3/5
www.first-sensor.com
www.sensortechnics.com
Signal conditioning circuitry for the RCO and HDO pressure sensors
A LT1014DN
Z1 LT1004CZ-2.5
R1, R2, R4 10 kOhm SIP resistor array
R3 100 kOhm SIP resistor array
RO C1 10 F / 35 V Tantalum
V+ + RO 10 kOhm multiturn pot
Zero adjust
= 12 V
R1 A2 RS 3.74 kOhm
-
+ RG 500 Ohm multiturn pot
A1 R1 R4 R4
+ - V
2.5 Z1 Ref Gain
adjust
R4 R3
R R
+ R2 R2 R2 S G
- C1
Rgain
R2 V = 10 V R3 R3
B
GND
- R3
RCO A3 -
+ - + A4 Vout
Sensor + = 5 2.5 V
Vin
Figure 3: RCO signal conditioning circuitry for 5 V 2.5 V output for 500 mmHg input pressure
E / 11157 / 1 4/5
www.first-sensor.com
www.sensortechnics.com
Signal conditioning circuitry for the RCO and HDO pressure sensors
Voffset =
(5000 mV 500 mV ) 25 mbar + 500 mV
125 mbar + 25 mbar The 0.5 V output signal at -25 mbar can be checked
by applying +25 mbar on port A of the RCOP001D...
Voffset = 1.25 V sensor. Thus, no vacuum is required.
A LT1014DN
Z1 LT1004CZ-2.5
R1
R1, R2, R4 10 kOhm SIP resistor array
R3 100 kOhm SIP resistor array
RO C1 10 F / 35 V Tantalum
V+ + RO 10 kOhm multiturn pot
Zero adjust
=9V
R1 A2 RS 680.1 Ohm
-
+ RG 100 Ohm multiturn pot
A1 R1 R4 R4
+ - V
2.5 Z1 Ref Gain
adjust
R3
R R
+ R2 S G
- C1
Rgain
R2 V=5V R3 R3
B
GND
- R3
RCO A3 -
+ - + A4 Vout
Sensor +
= 0.5...5 V
Vin
Figure 4: RCO signal conditioning circuitry for 0.5 V to 5 V output for -25 to +125 mbar input pressure
This information is based on a Sensym application note. First Sensor reserves the right to make changes to any products herein. First Sensor does not
assume any liability arising out of the application or use of any product or circuit described herein, neither does it convey any license under its patent
rights nor the rights of others.