You are on page 1of 2

Performance Sheet

DEEP LIGHTING STERILIZABLE HANDLE COLD LIGHT DIGIT CONTROL


D 400

HIGHLIGHTS
BTE97 DEVICE RESERVE BULB B AT T E R Y G R O U P

OPTIONALS

D 400 MOBILE
180 300 350 360

D 400
mobile

540
880

1670
130
530

380
480 180

660

Technical Data: D 400


Light intensity at 1 mt. distance with colour temperature of 4.200K 5% - at voltage 22,8V Klux 100
Light intensity at 1 mt. distance with color temperature of 4.200K 5% - at voltage 24V Klux 120
Main bulb W 100
Reserve bulb (optional, order code: DLE-device) W (100)3
Automatic switch to reserve bulb when main bulb fails optional
with intermittent led indicator (electronic DLE-device)2
Average lamp life of halogen bulb approximately at 22,8V1 h.10% 1.000
Average lamp life of halogen bulb approximately at 24V1 h.10% 500
CRI (Colour rendering Index) Ra 92<>94
Light field size adjustable from - to by handle mm. 130 - 220
Dept of working light without focus IEC 60601-2-41 mm. 1.140
Focusable at a distance from - to cm. 60 - 200
Absorption infra-red rays % 98
Total radiation (visible and infrared) at 100 Klux W/m2 360
UV-irradiance below 400 nm. at max light intensity W/m2 5
Infra-red filter RMS97
Temperature increase on operating field at 100 Klux C 12
Temperature increase at head height C 2
Elliptic reflector diameter mm. 360
Light body diameter mm. 400
Useful lighting surface of the cupola cm2 698
Sterilizable handle No. 1
Electronic control of light intensity positioned on vertical arm with ergonomic handle rotating dimming
Colour RAL 7035
Weight (ceiling, mobile, wall) Kg. 20-18-26
Power requirements V 230/110
Minimum height of the room mm. 2.300
Standards:
IEC 60601-2-41, MDD 93/42 CEE, medical device of Class I
EMC 89/336 CEE
Battery group with automatic charger autonomy 1 hour, free maintenance battery optional

1 It could change in consideration of different suggested voltage, peak of tension and frequency of utilization
2 With DLE-device (reserve bulb), light intensity reduce of 10%
3 optional

You might also like