You are on page 1of 10

- Journal of New Remedies & Clinics Vol. 59 No.

7 2010 -

Assessment of the Analytical Performance of MEDISAFE FIT, a


Simplified System for Self-monitoring of Blood Glucose (SMBG)
Using a New double-wavelength Colorimetric Method

- Assessment of the Analytical Performance According to ISO15197 and


the Japanese Approval Standard for SMBG Devices

Tokyo Saiseikai Central Hospital


Shu Meguro Masuomi Tomita
Chisato Murata Sonoko Anazawa
Kiyoe Kato Kenpei Matsuoka
Yoshihito Atsumi Yuko Takeda
Nahoko Sato Kunio Koka

Introduction measurement decreased substantially.


In May 2003, ISO15197 was enacted for
SMBG systems developed in early 1970s the assessment of performance of
successfully changed the blood glucose blood-glucose monitoring systems for
monitoring site from medical institutions to the self-testing in managing diabetes mellitus. In
homes of patients. As reported by Sonksen et March 2007, the Approval Standard for
al.1) and Ikeda et al.,2) the clinical application SMBG Devices was issued in Japan as a
of SMBG systems developed in late 1970s MHLW notification (PFSB Notification No.
played important roles in the early days. In 0302006), which obligate the applicants to
Japan, a symposium titled Self-monitoring of obtain data for [1] the assessment of
Blood Glucose and Diabetes Management repeatability (within-run repeatability and
was held in 1981, while in the U.S., the between-run repeatability) and [2] the
American Diabetes Association (ADA) started assessment of precision (comparison with the
to sponsor SMBG Symposium in 1986. In the standard measurement method).
same year, health insurance started to cover The authors had a chance to assess the
home SMBG in patients undergoing insulin performance of a newly developed SMBG
therapy in Japan. During the period, SMBG system based on the double-wavelength
system made steady advancements to the colorimetric method, MEDISAFE FIT
models operable by anyone using a smaller (Terumo Corporation) according to the
amount of blood for a broader range of blood Japanese Approval Standard for SMBG
glucose level. The time necessary for the Devices.

1
- Journal of New Remedies & Clinics Vol. 59 No. 7 2010 -

I. Subjects and methods installed, which enables the correction of


measurement data with Ht level and efficient
1. Subjects colorimetric analysis of blood glucose.
The study was performed using the blood The measurement using MEDISAFE FIT
samples from 116 outpatients who were needs 0.8 L of whole blood samples. The
undergoing diabetes treatment in Department measurement time is as short as 9 seconds and
of Internal Medicine, Tokyo Saiseikai Central the Ht level needs to be between 20 and 60%
Hospital, fulfilled the pre-defined inclusion and the ambient temperature needs to be
criteria, and consented in writing to participate between 5 and 40C. Measurement using
in the study. MEDISAFE FIT is verified unaffected by the
2. Collection of samples and treatment substances specified in MHLW notifications
method regarding safety measures (maltose (PFAB/SD
Venous blood collected into the vacuum Notification No. 0207007) and pralidoxime
blood collection tube containing heparin methyl iodide (PFAB/SD Notification No.
sodium was refrigerated promptly in 0907001)).
consideration of the risk of glycolysis. In 4. Control comparison method
principle, to perform the measurement by the As the control method, the automatic analysis
end of the day of collection, samples were system, LABOSPECT 008 and L Type Wako
subdivided into two tubes, ice-cooled until Glu2, which is the test method routinely used
immediately before the measurement, and in the hospital, and the measurement reagent
measured using MEDISAFE FIT. The based on the principle of hexonase/glucose-6-
temperature of the samples was returned to the phosphate dehydrogenase method
room temperature immediately before the (HK-G6PDH method), which is the glucose
measurement. The other sample was measurement method recommended by the
centrifuged (3000 rpm, 10 minutes) and the Japan Society of Clinical Chemistry were used.
supernatant was measured using LABOSPECT As for the assessment of precision of the
008 (Hitachi High-Technologies Corporation) measurement data obtained using
and L Type Wako Glu2 (Wako Pure Chemical LABOSPECT 008 and L Type Wako Glu2,
Industries, Ltd.). the equipment were calibrated using the
3. Overview of the SMBG device and regular standard substance of
measurement principles nitrogen-contained glucose of Reference
MEDISAFE FIT, a SMBG system with Material Institute for Clinical Chemistry
which the chromogen whose color is produced Standards (JCCRM 521) as specified in
by the glucose oxidase method is determined MHLW Notification No. 120 as the standard
by the double-wavelength measurement substance for the verification of performance
system using the test paper, was used. Photo 1 for the application for the in vitro diagnostics.
shows the appearance of the main unit of the JCCRM 521 is the primary standard substance
SMBG device and Fig. 1 shows the reagent determined using the isotope dilution mass
reaction and the measurement principles. The spectrometry to assure the precision of the
measurement sensitivity was improved by the glucose concentration. The certified glucose
enhancement of color production of the test concentration is determined by the method
strip by the use of the reagent recommended by JCCRM.
(N-ethyl-N-(2-hydroxy-3-sulfopropyl)-3,5-dim
ethyl aniline sodium and the reagent newly
II. Study methods
synthesized by Terumo Corporation,
1-(4-sulphonyl)-2, 1. Within-run repeatability
3-dimethyl-4-amino-5-pyrazolone). In the Blood samples were prepared at 7 glucose
main body of the SMBG device, two light levels ([1] 24 mg/dL, [2] 41 mg/dL, [3]
sources, LED1 with the central wavelength 109 mg/dL, [4] 141 mg/dL, [5] 185 mg/dL, [6]
located at 630 nm for reading the optical 272 mg/dL, [7] 561 mg/dL) as specified in
absorption of the color production reagent and ISO15197 and the Approval Standard for
LED2 with the central wavelength located at SMBG Devices using the venous blood
520 nm for detecting the blood color and obtained from patients and collected into the
determining the hematocrit (Ht) level, are vacuum blood collection tube containing

2
- Journal of New Remedies & Clinics Vol. 59 No. 7 2010 -

heparin sodium. Blood samples containing glucose by 7 concentrations were measured

Main unit Appearance of test


tip
Test strip (Color
production area)

Test tip (Aspiration nozzle)

Photo 1: Appearance of the SMBG device using the double-wavelength colorimetric method, MEDISAFE FIT

Reaction principle of the reagent


Glucose oxidase

-D-glucose+O2 + H2O D-glucono--lactone + H2O2

1-(4-sulphonyl)-2,3-dimethyl-4-amin
o-5-pyrazolone

Perodixase

N-ethyl-N-(2-hydroxy-3-sulfopropyl)-3,5-dimethyl
aniline sodium
Light blue production
Framed compound: Active ingredient of the test paper

Measurement principle of the main unit: Cross section image diagram


LED1: Light-emitting diode with the wavelength
located at 630 nm (for the assessment using the color
production reagent)
LED2: Light-emitting diode with the wavelength
located at 520 nm (for the determination of blood/Ht
concentrations)

Fig. 1 Reaction principle of the reagent and measurement principle of the main body

3
- Journal of New Remedies & Clinics Vol. 59 No. 7 2010 -

Table 1 Range of glucose concentration specified in ISO15197 and Japanese Approval Standard for SMBG
Devices

[1] Acceptance criteria for within-run repeatability


Sample Range of glucose concentration (mg/dL) Acceptance criteria
1 20-29 SD, not more than 5 mg/dL
2 30-50 SD, not more than 5 mg/dL
3 51-110 CV, not more than 5%
4 111-150 CV, not more than 5%
5 151-250 CV, not more than 5%
6 251-400 CV, not more than 5%
7 401-600 CV, not more than 5%

[2] Acceptance criteria for between-run repeatability


Sample Range of glucose concentration (mg/dL) Acceptance criteria
1 20-29 Total SD, not more than 7.5 mg/dL
2 30-50 Total SD, not more than 7.5 mg/dL
3 96-144 Total CV, not more than 7.5%
4 280-420 Total CV, not more than 7.5%
5 421-600 Total CV, not more than 7.5%

using ten MEDISAFE FIT each by 10 MEDISAFE FIT. The stability of the samples
repetitions and the mean, standard deviation associated with glycolysis was assessed before
(SD), and coefficient of variation (CV) were and after the measurement in comparison with
calculate to obtain within-run repeatability. the previously determined acceptable glucose
Table 1-(1) shows the range of glucose variation range and the data obtained from the
concentrations (Samples [1] to [7]) and the samples assessed deviating from the
acceptance criteria for within-run repeatability. acceptable range were excluded. The
2. Between-run repeatability. acceptable variation between the first
Between-run repeatability was assessed measurement data and the last measurement
also using the blood samples at 5 glucose data was designated as less than 4% for the
levels ([1] 24 mg/dL, [2] 40 mg/dL, [3] samples of which glucose concentration is
127 mg/dL, [4] 315 mg/dL, [5] 545 mg/dL) as higher than 100 mg/dL and less than 4 mg/dL
specified in the Approval Standard for SMBG for the samples of which glucose concentration
Devices. Using ten MEDISAFE FIT, the is not higher than 100 mg/dL. As a result, 14
control solutions at 5 glucose concentrations of the 116 samples were assessed deviating
were measured for 10 days and the mean, from the acceptable variation and excluded
standard deviation (SD), and coefficient of from the correlation analysis and dual
variation (CV) were calculated to obtain measurement data obtained from 102 samples
between-run repeatability. Table 1-(2) shows (from 204 measurements) were subject to the
the range of glucose concentrations (Samples analysis.
[1] to [5]) and the acceptance criteria for Meanwhile, the same samples were
between-run repeatability. centrifuged and the correlation between the
3. Precision assessment (Control data obtained by the recommended glucose
comparison method and correlation measurement method (HK-G6PDH method)
with MEDISAFE FIT) using LABOSPECT 008 and L Type Wako
Dual measurements of 116 samples of Glu2 (control method) and the measurement
venous blood obtained from patients and data obtained using MEDISAFE FIT was
collected into the vacuum blood collection assessed.
tube containing heparin sodium were
performed using two of randomly selected

4
- Journal of New Remedies & Clinics Vol. 59 No. 7 2010 -

III. Results performance in ISO15197 and the Japanese


Approval Standard for SMBG Devices.
1. Within-run repeatability As a result, at the low glucose range at
Table 2 shows the result of assessment of which within-run repeatability data tend to be
within-run repeatability. The standard relatively more dispersed, the standard
deviation of the samples containing glucose at deviation was between 1.4 and 1.8 mg/dL,
the concentration of not more than 50 mg/dL showing that the data were acceptable in terms
(Samples [1] and [2]) was 1.8 and 1.4 mg/dL, of within-run repeatability. Also at the range
respectively, which was not higher than the of clinical concentration of not less than 100
acceptance criteria of 5 mg/dL, and the result mg/dL, the total CV was as small as 2.5 to
suggests that the data from the samples were 3.3%, showing that the data were adequately
adequately acceptable in terms of within-run acceptable in terms of the precision. The result
repeatability. As for the samples containing of assessment of between-run repeatability
glucose at the concentration of 51 mg/dL or fulfills the acceptance criteria, suggesting that
over (Samples [3] to [7]), the total CV was as measurement using the device is expected to
satisfactory as between 2.5 and 3.3% and less produce stable data. Furthermore, in the
than the acceptance criterion of not more than assessment of precision, a favorable
5%. correlation was seen between the measurement
2. Between-run repeatability data obtained using the device and the data
Table 3 shows the result of assessment of obtained by the glucose measurement method
between-run repeatability. The total SD of (HK method) recommended by the Japanese
Control Solutions [1] and [2] was 0.96 and 1.3 Committee for Clinical Laboratory Standards
mg/dL, respectively, which was within the (JCCLS) using LABOSPECT 008 and L Type
acceptable criterion of 7.5 mg/dL or lower. Wako Glu2, and the device is considered to
Also as for Control Solutions [3] to [5], the have basic performance and be usable
total CV was as satisfactory as between 2.6 according to the global blood glucose
and 4.8% and less than the acceptable criterion management standards. These days, various
of not more than 7.5%. SMBG devices have many advantages, such as
3. Precision assessment which are simple and easy to operate and can
Fig. 2 shows the correlation with the be handled easily by anyone, so that they are
control comparison method (x) and the plots of used widely in medical institutions and at
differences. The linear regression formula of y homes of patients and contribute to the routine
= 1.00x 4.11 and the correlation coefficient blood glucose control in diabetic patients.
(R) of 0.995 substantiate the satisfactory MEDISAFE FIT is also usable widely for
correlation. Plots of differences between all blood glucose control not only in medical
measurement data obtained from 204 samples institutions but in elderly care facilities and at
and the measurement data obtained by the homes of diabetic patients and is considered
control comparison method were within the highly useful.
acceptable range of error (within 15 mg/dL Meanwhile, as stricter blood glucose
for the samples containing glucose by less than control is demanded, the devices are expected
75 mg/dL, within 20% for the samples to be more accurate and precise. Also along
containing glucose by not less than 75 mg/dL) with the diversification of therapeutic means in
and fulfilled the requirements of ISO15197 clinical sites, clinicians are required to more
and the Japanese Approval Standard for carefully select and use the SMBG devices,
SMBG Devices. especially in consideration of the influences of
the substances specified in MHLW
IV. Discussions notifications regarding safety measures
The authors assessed a SMBG systems (maltose (PFAB/SD Notification No.
newly developed by Terumo Corporation and 02507007) and pralidoxime methyl iodide
based on the double-wavelength colorimetric (PFAB/SD Notification No. 0907001)).
method, MEDISAFE FIT, in terms of the Measurement using MEDISAFE FIT is
within-run repeatability, between-run verified unaffected by maltose or pralidoxime
repeatability, and precision in accordance with methyl iodide and MEDISAFE FIT is
the criteria for the assessment of analytical considered usable in various healthcare

5
- Journal of New Remedies & Clinics Vol. 59 No. 7 2010 -

environments. various restrictions and conditions in mind as


Furthermore, clinicians need to always bear

Table 2 Result of assessment of within-run repeatability


Glucose concentration in Sample [1]: 24 mg/dL
Device No. 1 2 3 4 5 6 7 8 9 10
Mean data (mg/dL) 28 27 27 26 27 26 26 28 26 27
SD (mg/dL) 1.1 2.3 1.8 1.5 1.5 1.5 1.4 2.2 2.1 1.5
95% confidence interval (Mean) 0.79 1.7 1.3 1.1 1.1 1.1 1.0 1.5 1.5 1.1
CV (%) 3.9 8.8 6.9 5.7 5.7 5.9 5.6 7.8 8.0 5.6

Total mean (mg/dL) 27


Total variation 274
Total SD (mg/dL) 1.8
95% confidence interval (Total mean) 0.37
Total CV (%) 6.9

Glucose concentration in Sample [2]: 41 mg/dL


Device No. 1 2 3 4 5 6 7 8 9 10
Mean data (mg/dL) 43 43 42 41 42 42 40 43 42 44
SD (mg/dL) 1.2 1.1 1.2 1.7 0.88 0.95 0.95 1.0 1.1 0.85
95% confidence interval (Mean) 0.83 0.77 0.89 1.2 0.63 0.68 0.68 0.71 0.79 0.61
CV (%) 2.7 2.5 3.0 4.1 2.1 2.3 2.4 2.3 2.6 2.0

Total mean (mg/dL) 42


Total variation 112
Total SD (mg/dL) 1.4
95% confidence interval (Total mean) 0.27
Total CV (%) 3.2

Glucose concentration in Sample [3]: 109 mg/dL


Device No. 1 2 3 4 5 6 7 8 9 10
Mean data (mg/dL) 94 93 91 90 93 92 90 96 92 93
SD (mg/dL) 1.7 1.6 2.3 1.6 1.6 2.1 1.4 1.2 1.4 1.0
95% confidence interval (Mean) 1.2 1.1 1.6 1.1 1.2 1.5 1.0 0.86 1.0 0.74
CV (%) 1.8 1.7 2.5 1.7 1.8 2.3 1.6 1.2 1.5 1.1

Total mean (mg/dL) 92


Total variation 241
Total SD (mg/dL) 2.3
95% confidence interval (Total mean) 0.46
Total CV (%) 2.5

(Tables continued to the following page)

6
- Journal of New Remedies & Clinics Vol. 59 No. 7 2010 -

(Tables continued from the previous page)

Glucose concentration in Sample [4]: 141 mg/dL


Device No. 1 2 3 4 5 6 7 8 9 10
Mean data (mg/dL) 128 123 119 121 124 124 120 125 120 120
SD (mg/dL) 2.5 3.0 1.4 2.6 2.8 3.3 2.7 3.3 1.8 2.0
95% confidence interval (Mean) 1.8 2.1 1.0 1.8 2.0 2.4 1.9 2.3 1.3 1.4
CV (%) 2.0 2.4 1.2 2.1 2.3 2.7 2.3 2.6 1.5 1.7

Total mean (mg/dL) 122


Total variation 610
Total SD (mg/dL) 3.6
95% confidence interval (Total mean) 0.71
Total CV (%) 2.9

Glucose concentration in Sample [5]: 185 mg/dL


Device No. 1 2 3 4 5 6 7 8 9 10
Mean data (mg/dL) 171 171 166 163 168 164 165 167 168 168
SD (mg/dL) 3.4 5.0 3.0 4.5 4.5 3.7 4.2 4.7 5.4 4.3
95% confidence interval (Mean) 2.4 3.6 2.2 3.2 3.2 2.7 3.0 3.3 3.9 3.1
CV (%) 2.0 2.9 1.8 2.8 2.7 2.3 2.5 2.8 3.2 2.6

Total mean (mg/dL) 167


Total variation 1688
Total SD (mg/dL) 4.8
95% confidence interval (Total mean) 0.96
Total CV (%) 2.9

Glucose concentration in Sample [6]: 272 mg/dL


Device No. 1 2 3 4 5 6 7 8 9 10
Mean data (mg/dL) 245 252 244 242 244 246 242 253 245 240
SD (mg/dL) 9.5 7.4 3.9 6.1 5.5 8.3 4.7 5.0 4.4 5.8
95% confidence interval (Mean) 6.8 5.3 2.8 4.4 4.0 5.9 3.3 3.6 3.2 4.2
CV (%) 3.9 2.9 1.6 2.5 2.3 3.4 1.9 2.0 1.8 2.4

Total mean (mg/dL) 245


Total variation 3590
Total SD (mg/dL) 7.2
95% confidence interval (Total mean) 1.4
Total CV (%) 2.9

(Tables continued to the following page)

7
- Journal of New Remedies & Clinics Vol. 59 No. 7 2010 -

(Table 2 Result of assessment of within-run repeatability continued from the previous page)

Glucose concentration in Sample [7]: 561 mg/dL


Device No. 1 2 3 4 5 6 7 8 9 10
Mean data (mg/dL) 578 572 559 541 555 571 551 583 565 558
SD (mg/dL) 14.7 21.6 13.0 14.3 16.5 16.3 14.1 11.2 10.2 12.2
95% confidence interval (Mean) 10.5 15.5 9.3 10.2 11.8 11.6 10.1 8.0 7.3 8.7
CV (%) 2.5 3.8 2.3 2.6 3.0 2.8 2.6 1.9 1.8 2.2

Total mean (mg/dL) 563


Total variation 19533
Total SD (mg/dL) 18.7
95% confidence interval (Total mean) 3.7
Total CV (%) 3.3

Table 3 Result of assessment of between-run repeatability


Glucose concentration in Sample [1]: 24 mg/dL
Device No. 1 2 3 4 5 6 7 8 9 10
Mean data (mg/dL) 28 27 28 27 27 27 28 28 27 28
SD (mg/dL) 0.94 0.82 0.52 0.52 1.1 1.1 0.97 1.1 1.1 0.88
95% confidence interval (Mean) 0.67 0.59 0.37 0.37 0.76 0.76 0.69 0.76 0.75 0.63
CV (%) 3.4 3.0 1.9 1.9 3.9 3.9 3.5 3.7 3.9 3.1

Total mean (mg/dL) 28


Total variation 75
Total SD (mg/dL) 0.96
95% confidence interval (Total mean) 0.19
Total CV (%) 3.5

Glucose concentration in Sample [2]: 40 mg/dL


Device No. 1 2 3 4 5 6 7 8 9 10
Mean data (mg/dL) 46 45 46 46 46 46 45 46 45 46
SD (mg/dL) 1.3 2.0 0.99 0.67 1.3 0.97 1.1 1.2 0.85 1.2
95% confidence interval (Mean) 0.89 1.5 0.71 0.48 0.90 0.69 0.77 0.88 0.61 0.83
CV (%) 2.7 4.5 2.2 1.5 2.8 2.1 2.4 2.7 1.9 2.5

Total mean (mg/dL) 46


Total variation 130
Total SD (mg/dL) 1.3
95% confidence interval (Total mean) 0.26
Total CV (%) 2.8

(Tables continued to the following page)


(Tables continued from the previous page)

8
- Journal of New Remedies & Clinics Vol. 59 No. 7 2010 -

Glucose concentration in Sample [3]: 127 mg/dL


Device No. 1 2 3 4 5 6 7 8 9 10
Mean data (mg/dL) 128 128 126 125 127 126 125 128 125 129
SD (mg/dL) 2.2 3.0 2.5 3.5 2.9 2.2 2.7 4.2 4.1 2.8
95% confidence interval (Mean) 1.6 2.2 1.8 2.5 2.1 1.6 1.9 3.0 2.9 2.0
CV (%) 1.7 2.4 2.0 2.8 2.3 1.8 2.2 3.3 3.2 2.2

Total mean (mg/dL) 127


Total variation 854
Total SD (mg/dL) 3.3
95% confidence interval (Total mean) 0.65
Total CV (%) 2.6

Glucose concentration in Sample [4]: 315 mg/dL


Device No. 1 2 3 4 5 6 7 8 9 10
Mean data (mg/dL) 311 310 305 304 301 300 304 307 300 303
SD (mg/dL) 12.7 13.1 7.1 9.1 13.0 14.3 13.7 11.5 11.4 9.1
95% confidence interval (Mean) 9.1 9.3 5.1 6.5 9.3 10.2 9.8 8.2 8.2 6.5
CV (%) 4.1 4.2 2.3 3.0 4.3 4.8 4.5 3.7 3.8 3.0

Total mean (mg/dL) 305


Total variation 12334
Total SD (mg/dL) 11.8
95% confidence interval (Total mean) 2.3
Total CV (%) 3.9

Glucose concentration in Sample [5]: 545 mg/dL


Device No. 1 2 3 4 5 6 7 8 9 10
Mean data (mg/dL) 547 564 547 535 539 524 532 545 545 543
SD (mg/dL) 25.8 26.3 26.3 23.1 20.2 26.2 27.6 23.2 32.0 17.2
95% confidence interval (Mean) 18.5 18.8 18.8 16.5 14.5 18.7 19.7 16.6 22.9 12.3
CV (%) 4.7 4.7 4.8 4.3 3.8 5.0 5.2 4.3 5.9 3.2

Total mean (mg/dL) 542


Total variation 56731
Total SD (mg/dL) 26.0
95% confidence interval (Total mean) 5.2
Total CV (%) 4.8

(Tables continued to the following page)

9
- Journal of New Remedies & Clinics Vol. 59 No. 7 2010 -

[Correlation]

Measurement data obtained using


MEDISAFE FIT

Measurement data obtained using LABOSPECT 008 and L Type Wako Glu2

[Plot of differences from the control method]


Difference from the control method

Measurement data obtained using LABOSPECT 008 and L Type Wako Glu2

Fig. 2 Result of precision assessment

exemplified by the fact that most of the users of MEDISAFE FIT broadens the usability in
SMBG systems are old and have concomitant blood glucose control and team medicine as a
diseases. Simple and easy operating and SMBG system which has undergone
handling procedures are important points to remodeling with adequate consideration and is
bear in mind when selecting the device model. useful for the blood glucose management of
MEDISAFE FIT has superior properties which patients under home care.
existing products do not have as exemplified by
the main body which elderly patients can easily Summary
carry, the considerations for the appearance of
the test tip, dot LED display of operating The authors assessed a SMBG system based
procedures and other information, on the double-wavelength colorimetric method,
easy-to-understand precautions and error MEDISAFE FIT, and has reached the
messages. The main body of MEDISAFE FIT conclusion that the system has the analytical
can save up to 500 test results and the data are performance fulfilling the criteria of ISO15197
uploaded to PC or other devices by Near Field and the Japanese Approval Standard for SMBG
Communication System (FeliCa System), a Devices and is useful for the blood glucose
function introduced for the first time in the management in diabetic patients in future.
world.

10

You might also like