Professional Documents
Culture Documents
| Quick Quide
SYSTEM VERSION
COPYRIGHT
No part of this documentation may be reproduced in any form, nor processed, copied or
distributed by means of electronic systems, without prior permission of HUMAN in writ-
ing. Since all precautionary measures were taken into account in producing these operating
instructions, the manufacturer accepts no responsibility for any errors or omissions. This
includes any liability for damage that could arise from possible incorrect operation based on this
information. Subject to changes without notice as result of technical development.
TABLE OF CONTENTS
1PREPARATION 3
1.1 UNPACKING THE HUMACYCLER 3
1.2 LOCATING THE HUMACYCLER 4
1.3 ENVIRONMENTAL CONDITIONS 4
1.4 CONNECTING THE INSTRUMENT TO THE COMPUTER 4
1.5 SOFTWARE INSTALLATION 5
4 RUN PROGRAM 13
5 RESULT ANALYSIS 15
6CAUTION 21
Preparation 3
1 PREPARATION
Before starting, the Fixing Pin shall be taken out from of the instrument by turn-
ing it anticlockwise (see figure 1).
Figure 1
Insert the pin into the unlocker port at its back and fully tighten it clockwise (see
figure 2).
Figure 2
Three communication boxes are provided: RS232C, USB and Bluetooth (see
figure 3). If using USB, the USB driver of the provided software must be installed.
If using Bluetooth, Bluetooth equipment and software is needed. The communi-
cation box RS232C can be used directly.
Figure 3
RS232C, Bluetooth and
USB Converter Box.
To install the software, insert the provided CD into the disc drive of the
computer, click onto the HCy4 software file and follow the instructions to install
it.
Figure 4
6
Figure 5
After the software started, the User Login interface appears. Please enter
User name: admin, Password: admin and click OK (see figure 6).
8
Figure 6
Click Service User Management Add User and enter your login data
The accessories provide an Anti-Condensate Water Seal Ring for improving the
accuracy of the experiment results and preventing the tubes at and near the
edges of the heating block to generate condensate, which would lead to a re-
duced fluorescent signal. In addition, the seal ring insulates the heat after it is
placed in position. It is applicable for the use of single PCR tubes, 8-strip PCR
tubes, half- skirted and non skirted PCR plates. Full skirted plates cannot be used
with the seal ring. Before starting the experiment, the seal ring should be placed
around the heating block (see figure 7)
Figure 7
Figure 8
1. Set up Detectors
Click Detector
Enter the experiment properties, experiment name (is generated automatically
but can be modified), the user name and add comments to the experiment.
Click Add Detector to add a new target. Enter the name of the target and choose
the reporter dye. If you are using a Human Real-Time PCR Assay (HIV-1, HCV,
HBV, CMV), choose FAM.
Click Add Detector to add another target. Enter the name of the target and
choose the reporter dye. If you are using a Human Real-Time PCR Assay (HIV-1,
HCV, HBV, CMV), choose HEX for the Internal Control (IC) (see figure 9).
Figure 9
10
Select the assay item and set up the properties of your sample (Unknown, Stand-
ard, Non-template Control and Positive Control) and the Concentration Units.
For Human Real-Time PCR assays select IU/ml (see figure 10).
Figure 10
Figure 11
Add Stage
Hold Stage
PCR Stage
Melting Stage
Infinite Stage
Add Step (before or after an existing step)
Cycles
Liquid Quantity of the Test
Temperature
Time
Measuring Step
Delete
12
Figure 12
4 RUN PROGRAM
To ensure getting better application curves, it is necessary to adjust the gain
of the instrument before the amplification reaction. View the gain in the Gain
Settings under Tools. Select Custom Gain to set the gain by yourself. The default
gain is 7.
To adapt users reagents, the Custom Gain is adopted. Prepare the sample set up
like it is described in the User Manual. Edit the program, sample settings and the
temperature profile as follows:
Temperature 55C
Time 10 sec
Cycles 10
Click Run and Start Run. The gain setting interface appears (see figure 13).
Choose Custom Gain and click OK. Observe the fluorescence curves of the sam-
ples.
Figure 13
The fluorescence values should be between 1000 and 2000. If the values are too
high, repeat the run and decrease the gain volume for the corresponding dye. If
the values are lower than expected, repeat the run and increase the gain volume
for the corresponding dye. If the fluorescence values are in the expected range,
set up the program, samples and plate for the testing.
Fluorescence, temperature curves and the program can be observed during the
run by choosing the corresponding item. Furthermore, the run status is shown,
e.g. start and end time, current cycle and step (see figure 14).
Figure 14
14
5 RESULT ANALYSIS
After the run is completed, an interface to save the experiment will pop up. Click
Yes to save or No to view the experiment but not to save it.
The fluorescence curves and the plate are shown automatically (see
figure 15). To modify the analysis settings, click Analysis Settings. After modifi-
cation click Apply Analysis Settings to save it.
Figure 15
Figure 16
16
Click Result Summary to view all values for the corresponding wells.
Figure 17
Figure 19
Figure 20
18
4. SNP analysis:
Click SNP to show the Allele Scatter Diagram or click Amplification Plot to show
the fluorescence curves (see figure 21).
Figure 21